Analytica Chimica Acta 599 (2007) 256–263 Purity of potassium hydrogen phthalate, determination with precision coulometric and volumetric titration–A comparison Sebastian Recknagel ∗ , Martin Breitenbach, Joachim Pautz, Detlef Lück Federal Institute for Materials Research and Testing (BAM), Division of Inorganic-Chemical Analysis, Reference Materials, Richard-Willstaetter-Str. 11, D-12489 Berlin, Germany Received 8 March 2007; received in revised form 5 July 2007; accepted 20 July 2007 Available online 2 August 2007 Abstract The mass fraction of potassium hydrogen phthalate (KHP) from a specific batch was certified as an acidimetric standard. Two different analytical methods on a metrological level were used to carry out certification analysis: precision constant current coulometric and volumetric titration with NaOH. It could be shown that with a commercial automatic titration system in combination with a reliable software for the end-point detection it is possible to produce equivalent results with the same accuracy in comparison to a definite method handled by a fundamental apparatus for traceable precision coulometry. Prerequisite for titrations are that a high number of single measurement are applied which are calibrated with a high precision certified reference material. © 2007 Elsevier B.V. All rights reserved. Keywords: Potassium hydrogen phthalate; Precision constant current coulometry; Volumetric titration; Certified reference material 1. Introduction Potassium hydrogen phthalate (KHP) is widely used both as acidimetric standard for acid and base titration analysis and as pH standard. Therefore, well characterised certified reference materials are necessary. Within a framework of BAM and Sigma–Aldrich Production GmbH for certification of primary substances for titration and standard anion solutions a batch of 994 bottles of KHP was certified for its acidimetric purity expressed as a mass fraction. It is intended to be used as an acidimetric standard only. For the certification of KHP as well as for all other substances included in this certification program two independent analytical methods are used for certification analysis. Traceability of these measurements is ensured by using well characterised pure substances or certified reference materials for calibration in case there is no absolute method like coulometry available. In the case of KHP precision coulometry was used as primary direct method [1–3] whose results are traceable to the SI-system. ∗ Corresponding author. Tel.: +49 30 8104 1111; fax: +49 30 8104 1117. E-mail address: [email protected] (S. Recknagel). 0003-2670/$ – see front matter © 2007 Elsevier B.V. All rights reserved. doi:10.1016/j.aca.2007.07.062 KHP has been determined by precision coulometric acidimetric titration for about 50 years [4–7]. Since then the procedure was extensively automated [1,8–12]. The purity of KHP was subject of CCQM-P36 and CCQM-K34/K34.1 where seven metrology institutes compared their measurement capability to determine the amount content of acid in solid weak acids. All institutes used constant current coulometry. It was demonstrated that the agreement of the results of all participants was good in general [13]. On the other hand, this study showed again that great care has to be taken during the sample preparation step especially if some impurities remain undetected and the purity of high purity materials is determined by the difference to 100%. Although there has been a lot of work carried out on KHPdetermination using precision coulometry, it is still an extremely extensive method not suitable for high sample throughput. A fundamental apparatus for coulometry is commercially not available. Therefore, individual experimental designs are applied, which are only possible to build up if special know how and knowledge of coulometry procedures and electrochemistry techniques are available. Handling such a complex equipment and performing precise measurements which are based on calibrated SI-quantities, e.g. mass, voltage, electrical resistance and time, and knowledge to avoid chemical errors are only manageable by highly trained specialists. S. Recknagel et al. / Analytica Chimica Acta 599 (2007) 256–263 To overcome these problems a special experimental design should be developed to carry out certification analysis of a large batch of a KHP-standard. Based on a commercially available automatic titration system in combination with a reliable software for the end-point detection it should be possible to produce equivalent results with the same accuracy in comparison to a definite method handled by a fundamental apparatus for traceable precision coulometry. Felber et al. demonstrated the capabilities of titrimetry under carefully controlled conditions for the complexometric determination of Cu2+ with EDTA [14]. Brown et al. developed a high accuracy titrimetric method for HCl-determination [15] and showed the equivalence of the results of this method with coulometric acid–base titration [16]. Prerequisite for high precision titrations are that a high number of single measurements are applied which are calibrated with a high precision certified reference material. In this paper the results of precision coulometric titration are compared with those from precision volumetric titration used for analysis and certification of a specific KHP reference material. 2. Experimental (I) precision coulometry Coulometry is a primary direct method for measuring the amount of substance. It is based on Faraday’s law of electrochemical equivalence. The amount content of substance in a solution is proportional to electricity, which flows between the generating electrodes. The basic requirement of a coulometric 257 titration [17,18] is practically 100% current efficiency of the titrant generation. The high precision of results is based on: (a) a very precise weighing combined with high sample intake (m); (b) the possibility of very precise dosing of electricity; (c) the exact determination of the electricity by precise measurement of the physical quantities: electrical resistance (R), voltage (U) and time (t). Therefore, traceability of results to quantities of SI (système international d’unités) units is given. 2.1. Apparatus The determinations performed in this study were carried out using the apparatus given in Fig. 1. Measuring cell: Vertical arrangement, cathodic compartment and anodic compartment were separated by an intermediate compartment (all made of boron glass), arranged between two glass frits. The glass frit near the cathodic compartment was blocked with a pH-neutral agar–agar gel plug. Generating cathode: Pt-wire (2 cm2 ); generating anode: Ag-rod (16 cm2 ); indicating electrode: pH-glass electrode with fixed ground-joint diaphragm and Ag/AgCl reference system (6.0253.100, Metrohm, Herisau, Switzerland; reference electrolyte: KCl gel (c(KCl) = 3 mol L−1 ); membrane resistance: 80–200 M). Fig. 1. Precision constant current coulometry system (principal sketch). Measuring cell for acid/base titrations, valve manifold and associated pneumatic connections: ) flow resistor, (IC) internal compartment. (1) Pt generating anode, (2) Pt generating cathode, (3) PTFE spray shield, (4) (×) valves, (DT) deaeration tubes, ( indicating glass electrode, (5) agar diaphragma, (6) internal compartment. 258 S. Recknagel et al. / Analytica Chimica Acta 599 (2007) 256–263 The precision coulometer used was an electrochemical computerised system (for details of the titration software see [1]). The assembly consisted of a constant current source model 8011A, an indication unit 7011A, a valve unit 9018A (all Applied Precision Ltd., Bratislava, Slovakia), a piston burette type 765 Dosimat (Metrohm). For measurement we used four calibrated devices: An analytical microbalance MC 5, resolution 1 g with an uncertainty uMC5 of the balance of uMC5 = 50 ppm (Sartorius AG, Göttingen, Germany), a timer with an average relative frequency error and expanded uncertainty of measurement (k = 2) of δf = (9.12 ± 0.08) × 10−7 (Applied Precision Ltd.) to control the electrolysis time with a precision of 0.1 ppm (±1 s). The current was determined measuring the voltage drop on a resistor using a 6½-digit-multimeter DMM 2000 with 0.02 ppm per 90-day basis accuracy (Keithley Instruments, Inc., Cleveland, Ohio, USA) and a certified 1 resistor type 742A (Fluke GmbH, Kassel, Germany) with a minimum uncertainty of 2.5 ppm per 6 months (temperature coefficient: 0.1 ppm K−1 , calibrated value: 1.00000 ± 0.000017 , traceability chain: Fluke 742A – HP 3458A). This system controlled the generator current, switched on/off the titration time, registered the pH value versus time and rinsed the intermediate compartment of the cell automatically. solution using a small silica weighing cup. For the mass value evaluation the air buoyancy correction of the weighing values was performed, taking into account a model equation for the air density which is based on measured values for the air temperature, air pressure and the relative air humidity assuming standard air composition. The coulometric titration consisted of three steps: the initial titration, the main titration and the final titration. Different constant currents were used. The purity p was calculated according to Faraday’s law from the ratio of the experimental coulometric charge (Qtot ) and the theoretically expected charge (zFm/M) using Eq. (1). p= MQtot zFm (1) The amount content v in mol kg−1 was calculated applying Eq. (2): v(KHC8 H4 O4 ) = U1 (t1∗ − t1∗∗ − tcorr ) + U2 t2 + U3 t3 zFmR The mass fraction w in % was calculated using Eq. (3): w(KHC8 H4 O4 ) = 2.2. Procedure Potassium hydrogen phthalate reference material, KHP lot no. 1291630 (Sigma–Aldrich product-no: 60357 [19]) was dried at 110 ◦ C for 4 h and allowed to cool in a desiccator over P4 O10 before analysis. After first drying the loss was <0.01% and after the repeated drying the maximum mass difference was ± 0.001%. As supporting electrolyte a 1 mol L−1 KCl solution was used. The initial titration (2 mA; 2 s pulses) was carried out using potentiometric end-point detection at pH ∼ 6.9. After the initial titration the sample was introduced into the cell using a silica crucible. During the main titration 99.9% of the stoichiometric amount of hydroxyl ions were generated with a generator current of 200 mA. The final titration (10 mA; 3.5 s pulses) was performed using potentiometric end-point detection at pH ∼8.2. The calculation of the purity of KHP was done using Eq. (1). Biases resulting from impurities from the purging gas argon, from the electrolyte and from the agar gel were minimised using 99.9999% argon, neutralisation of the agar gel prior to each determination with an indicator, the use of water with a conductivity of κ < 2 S cm−1 and KCl (≥99.5%, Merck, Darmstadt, Germany). To avoid losses by diffusion and (electro-) migration of hydrogen phthalate ion (HC8 H4 O4 − ) into the intermediate compartment (IC) during the main titration, the IC was filled with supporting electrolyte which was forced slowly into the working compartment during the main titration. Biases due to spraying, rinsing of the intermediate compartment with measuring solution and losses by sample introduction were minimised using the following experimental conditions: Application of a spray shield: This spray shield and the intermediate compartment were repeatedly rinsed with measuring solution during titration. The sample was introduced into the (2) 100M[U1 (t1∗ − t1∗∗ − tcorr ) + U2 t2 + U3 t3 ] zFmR (3) with Quantity Unit p v w Qtot mol kg−1 % C M g mol−1 U1 t1 * t1 ** tcorr U2 t2 U3 t3 z V s s s V s V s – F m R A s mol−1 kg * Description Purity Amount of substance content Mass fraction Amount of electricity between the end-point of initial titration and the end-point of final titration [Q = It], [I = U/R] Molar mass of KHC8 H4 O4 = 204.2212 g mol−1 [20] Average voltage during initial titration Total time of the initial titration Time of the initial titration, up to the end-point Correction of the initial titration time Average voltage during main titration Time of the main titration Average voltage during final titration Time of the final titration, up to the end-point Product of number of transferred electrons and current efficiency Faraday constant: 96485.3399 A s mol−1 [21] KHP mass, corrected for air buoyancy* Resistance of measuring resistor Sample density used for buoyancy correction: 1636 kg m−3 [22]. Calculation of the measurement uncertainty was carried out according to the BAM-Guideline for the evaluation of measurement uncertainties for quantitative measurements [23] based on the ISO Guide for the Expression of Uncertainty in Measurement [24]. It was done by quadratic addition of type A and type B uncertainty contributions. Type A is the statistical part (standard deviation of the mean), type B is the non-statistical part based primarily on the apparatus. For calcu- S. Recknagel et al. / Analytica Chimica Acta 599 (2007) 256–263 lation of the expanded uncertainty the coverage factor k = 2 was used. Uncertainty contributions of type B were calculated using the individual uncertainties of the parameters used in Eq. (3) (see Table 2). The uncertainty contributions from the instruments used for measuring the physical values were at most 0.005%, therefore chemical factors like contributions for electrolyte and inert gas impurities, diffusion, and phthalate reduction which were in the same order of magnitude had to be considered. 259 dioxide from the air. Before starting a titration the sample solution was flushed carefully with argon (only 2 bubbles per second to avoid any splashing) for 5 min to expel any CO2 from the solution. 3.2. Procedure The second analytical method used for the characterisation of the potassium hydrogen phthalate and for homogeneity testing of the whole batch was a volumetric method, acid–base titration with potentiometric end-point detection. Calibration was done using certified reference material NIST SRM 84k acidimetric primary standard. This CRM was characterised by the US National Institute for Standards and Technology using coulometric titration which is a primary method since it is directly traceable to the SI. The certified mass fraction of potassium hydrogen phthalate investigated via volumetric titration in this study therefore is traceable to NIST SRM 84k and not directly to the SI. For homogeneity testing traceability is not important because the total mass fraction of the material investigated is not needed. To check for inhomogeneities an analytical method as precise as possible has to be used. Normally it is impossible to differentiate between variations resulting from inhomogeneities of the sample and variations coming from the analytical method. In most cases it is the sum of both. The spread of the method could only be determined if a sample of ideal homogeneity would be available. Since this ideal sample normally does not exist, the spread of results observed is always the sum of inhomogeneity contributions from the sample material and spread of the method. The higher the spread of the method is, the higher is the estimated inhomogeneity contribution to the uncertainty of the certified mass fraction. Weber et al. showed that volumetric titration with a titration equipment similar to that used for this investigation is precise enough to be used for homogeneity testing [25]. Acid–base titration with 0.05 mol L−1 NaOH solution was performed to determine the mass content of potassium hydrogen phthalate. 175 ± 1 mg of sample were weighed after the material had been dried for 3 h at 120 ◦ C and then cooled in a desiccator over Mg(ClO4 )2 . A correction of the initial weight for buoyancy was not necessary because sample and reference had the same density. The volume of the NaOH solution used was approximately 17.4 mL. This solution was prepared using NaOH pellets (p.a. Merck, Darmstadt, Germany) which were precleaned with water to remove Na2 CO3 from the surface of the pellets. Dissolution was carried out using Milli-Q water (κ < 2 S cm−1 ) degassed first with argon and second with ultrasound. The solution was stored under argon atmosphere in a glass bottle endued with a CO2 -absorber (NaOH fixed on a substrate in an absorber tube). Standard reference material NIST SRM 84k potassium hydrogen phthalate (99.9911%) was used for calibration of the apparatus. Sigma–Aldrich Production GmbH (Buchs, Switzerland) filled a total number of 994 glass bottles each with 50 g of batch material. The sampling of 35 randomly chosen bottles for certification analyses, taken out of the entire batch, was done by Sigma–Aldrich Production GmbH. These bottles were sent to BAM to carry out certification analyses. Analyses were carried out following the analysis scheme given in Fig. 2. Seven subsamples out of each bottle were analysed to determine the mass fraction of potassium hydrogen phthalate and to assess the degree of heterogeneity of the whole batch of 994 bottles. In total 301 determinations were performed which – regarding to expenditure of time – is only possible with this special titration equipment but not with the coulometric method. Each of the seven titration runs consisted of 43 titrations, i.e. 35 samples and eight portions of certified reference material NIST SRM 84k. 3.1. Apparatus 4. Results and discussion Titrations were carried out using a modular automatic titration system (Metrohm, Herisau, Switzerland) consisting of a sample changer 730, a dispensing unit Dosino 700 with a GPTitrino 736 titroprocessor and titration software Metrodata “TiNet©2.4”. For potentiometric end-point detection a combined pH-glass electrode with ceramic pin diaphragm and Ag/AgCl reference system (6.0253.100, Metrohm, Herisau, Switzerland; reference electrolyte: KCl gel (c(KCl) = 3 mol L−1 ); membrane resistance: 150–400 M) was used. End-point detection was an integral part of the titration software used. Calculation was done using the second derivation of the curve d (measurant)/d (volume). Titration was performed under argon atmosphere to avoid any interference of carbon 4.1. Coulometric titration 3. Experimental (II) volumetric titration Table 1a shows the results of the nine coulometric determinations of three different bottles of the investigated KHP batch and the mean as well as the standard deviation of the mean. As a check for a potential bias, eight additional measurements with low (0.5 mg) sample intakes were carried out to establish the recovery mass function. The purity was calculated as the ratio of coulometrically determined amount of substance versus theoretical amount of substance calculated from sample mass. In Fig. 3 17 measurements are included, the results of the high sample intake measurements are given in Table 1a, the results of the measurements with very low sample intake are given in 260 S. Recknagel et al. / Analytica Chimica Acta 599 (2007) 256–263 Fig. 2. Analysis scheme of volumetric titration for certification. Table 1b. The slope of this curve gave a bias corrected mass fraction of 99.990%. This value was used to calculate the certified value of the material by combining it with the volumetric results. The difference between the coulometrically determined mean from Table 1a and the slope calculated from the mass-function (n = 17), Tables 1a and 1b, was the bias of 0.004%. Calculation of the uncertainty is shown in Tables 2 and 5. The uncertainty contribution of type A for the KHP content determined by coulometry could √ either be calculated from the standard deviation divided by m, if the mass fraction w was obtained as mean value of n single results from m independent bottles or it could be taken from regression analysis of the recovery function. Fig. 3. Recovery plot of coulometrically determined amount of substance vs. theoretical amount of substance from sample weight (n = 17). Table 1a Coulometric determination of purity of KHP lot no. 1291630, (Fluka product-no: 60357) Bottle no. Mass (m in g) Acid amount content (v in mol kg−1 ) Acid mass fraction (w in %)* 1 1 1 0.723633 0.732635 0.748235 4.89671 4.89582 4.89649 100.001 99.983 99.997 2 2 2 0.817912 0.506147 0.682642 4.89593 4.89694 4.89561 99.985 100.006 99.979 3 3 3 0.752915 0.514089 0.713291 4.89681 4.89661 4.89634 100.003 99.999 99.994 4.89636 0.00047 99.994 0.010 Mean (n = 9) Standard deviation S (n = 9) w (%) = 100 × v (mol kg−1 ) × M (kg mol−1 ) with MKHP = 0.2042212 kg mol−1 . * 4.2. Volumetric titration The results of the volumetric determination are given in Table 3. For each of the 35 bottles the means of seven single determinations are listed together with the corresponding absolute and relative standard deviations. There was no hint for any inhomogeneity of the material. The mass fraction of Table 1b Low masses measurements for recovery plot Bottle no. Mass (m in g) Acid amount content (v in mol kg−1 ) 1 1 0.000553 0.000541 5.33124 4.69284 2 2 0.000503 0.000510 5.15955 4.98119 3 3 3 3 0.000543 0.000472 0.000466 0.000466 5.01092 4.94627 5.27164 5.05187 S. Recknagel et al. / Analytica Chimica Acta 599 (2007) 256–263 261 Table 2 Uncertainties, type B for coulometric determination of KHP Effect Distribution Sensitivity Uncertainty Product ci Units ui Units ci ui Units 4.9E+02 1.0E+02 1.4E+02 6.0E−02 % V−1 4.0E−06 1.7E−05 3.5E−05 2.0E−04 V g s 2.0E−03 1.7E−03 5.0E−03 1.2E−05 % % % % 5.6E−03 % 4.1E−04 2.0E−03 1.4E−03 1.0E−03 4.1E−04 1.0E−04 % % % % % % 2.8E−03 % 4.2E−04 1.1E−03 % % Constants 1.2E−03 % Total 6.4E−03 % Voltage Resistance Mass Time Normal Normal Normal Normal % −1 % g−1 % s−1 Physical contributions Phthalate reduction Electrolyte impurities Inertgas impurities Diffusion Incomplete rinsing Current efficiency Normal Normal Normal Normal Normal Rectang. 8.2E+01 1.0E+02 1.4E+02 1.0E+02 1.0E+02 1.0E+02 – – – – – – 1.1E−01 5.0E−01 % mol C−1 5.0E−06 2.0E−05 1.0E−05 1.0E−05 4.0E−06 1.0E−06 – – – – – – 3.9E−03 2.2E−03 C mol−1 Chemical contributions F(CODATA) M(KHP) Rectang. Rectang. % mol g−1 KHP was calculated using Eqs. (4) and (5). The certified value of the material was calculated combining the mean value of 99.987% resulting from the volumetric determinations with the coulometric results (see below). f (titrant) = m(RM) × w(RM) V (titrant, RM) w (sample) = f (titrant) × V (titrant, sample) × 100% m (sample) (4) (5) with f (titrant): titration factor in mg mL−1 ; w (sample): mass fraction of sample in %; m (RM): mass of reference material used for titration in mg; w (RM): mass fraction of reference material used for titration in %; V (titrant, RM): volume of titrant used for titration of the reference material in mL; V (titrant, sample): volume of titrant used for titration of the sample in mL; m (sample): mass of sample in mg. Table 4 shows potential contributions to the combined uncertainty of the results received from volumetric titration. 4.2.1. Type A uncertainty (uA ) Random deviations are expressed by the spread of the results which is the repeatability of the determination. Since test samples and reference material were analysed alternately and the volume of titrant was nearly the same for all titration solutions any contribution from the uncertainty of the volume was also included into the repeatability standard deviation. Type A uncertainty consisted of repeatability of the factor and the sample determination. These values were calculated using the mean standard deviation of seven repeated titrations with 35 samples and eight determinations of the factor using NIST SRM 84k with seven repeated single determinations. The result√ ing standard deviation was divided by 7 for the number of independent titrations. g mol−1 The contribution of type A uncertainty decreased due to the high number of repetitions performed for the mass fraction deter√ mination (division by factor n). 4.2.2. Type B uncertainty (uB ) Variations of the concentration of the titrant by absorption of carbon dioxide which would lead to increasing titrant volumes and therefore give a bias were avoided by working under argon atmosphere. A blank correction was not necessary because a possible blank would influence reference material and sample in the same way. Beside uncertainty contributions which could be avoided by careful handling there were some potential sources of uncertainty which had not to be considered because they were compensated by using the same procedure for the calibration substance and the test samples. This was true for a possible influence of temperature on the volume which would affect reference material and test sample in the same way. This was also the case for a possible bias of the end-point detection which would also affect the determination of reference material and test sample in the same way. As mentioned above buoyancy correction was not carried out because sample and reference had the same density. Therefore, any uncertainty contribution based on buoyancy correction was not taken into consideration for volumetric titrations. Contributions to the type B uncertainty came from the weighing of samples and reference material – linearity of the balance ±0.125 mg which could be converted to a standard uncertainty assuming a rectangular distribution – and the purity of the reference material, given in the reference material certificate [26]. As a result main contributions to the combined uncertainty came from the weighing process of sample and reference material. 262 S. Recknagel et al. / Analytica Chimica Acta 599 (2007) 256–263 Table 3 Results of KHP-determination (volumetric titration) Table 4 Potential uncertainty contributions of volumetric titration Sample u(Vtitr, sample ): volume of titrant Mean values of all titration steps 1291630-050 (01) 1291630-094 (02) 1291630-095 (03) 1291630-190 (04) 1291630-248 (05) 1291630-262 (06) 1291630-282 (07) 1291630-291 (08) 1291630-315 (09) 1291630-323 (10) 1291630-376 (11) 1291630-392 (12) 1291630-474 (13) 1291630-481 (14) 1291630-485 (15) 1291630-496 (16) 1291630-516 (17) 1291630-517 (18) 1291630-537 (19) 1291630-563 (20) 1291630-587 (21) 1291630-589 (22) 1291630-616 (23) 1291630-626 (24) 1291630-647 (25) 1291630-667 (26) 1291630-669 (27) 1291630-680 (28) 1291630-683 (29) 1291630-733 (30) 1291630-776 (31) 1291630-812 (32) 1291630-840 (33) 1291630-888 (34) 1291630-960 (35) Mean (%) S.D. (%) R.S.D. (%) n 99.980 99.986 99.975 99.983 99.978 99.989 99.989 100.001 99.986 99.969 99.966 99.994 99.994 99.987 99.976 99.982 100.002 99.997 99.990 99.987 99.984 100.000 99.989 99.990 99.995 100.002 99.996 99.983 99.972 99.971 99.998 99.982 99.981 100.005 99.992 0.028 0.029 0.024 0.022 0.030 0.029 0.024 0.035 0.028 0.013 0.039 0.020 0.024 0.026 0.022 0.019 0.033 0.025 0.026 0.017 0.034 0.021 0.038 0.019 0.030 0.029 0.039 0.024 0.031 0.025 0.038 0.038 0.023 0.020 0.043 0.028 0.029 0.024 0.022 0.030 0.029 0.024 0.035 0.028 0.013 0.039 0.020 0.024 0.026 0.022 0.019 0.033 0.025 0.026 0.017 0.034 0.021 0.038 0.019 0.030 0.029 0.039 0.024 0.031 0.025 0.038 0.038 0.023 0.020 0.043 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 Mean of means S.D. of means R.S.D. of means m 99.987% 0.010% 0.010% 35 4.3. Calculation of the certified value and its combined uncertainty The certified value of KHP, batch no. 1291630 which was investigated in this study, was calculated as the mean value of u(Vtitr, RM ): volume of titrant u(Vend point, bias ) u(Vend point, ) u(MKHP ) u(Vtitr, T ): influence of temperature on volume of titrant u(msample ) u(mRM ) u(ksample ): buoyancy correction u(kRM ): buoyancy correction u(pRM ): purity of reference material Not relevant because same volume for reference material and sample Not relevant because same volume for reference material and sample Same detection of end point for reference material and sample Part of result spread Not relevant because same material as reference material and sample Part of result spread, compensated by alternate determination of reference material and sample Linearity of balance, to be taken into account Linearity of balance, to be taken into account Not relevant because same mass for reference material and sample Not relevant because same mass for reference material and sample To be taken into account the means of both methods (see Fig. 4): (99.987 + 99.990) = 99.989% 2 The combined uncertainty of the certified value resulted from the combination of the uncertainty contributions of both methods and the inhomogeneity contribution: uc = u2Volumetry + u2Coulometry + u2Inhom = 0.0212 + 0.0072 + 0.0102 = 0.024% The inhomogeneity contribution to the uncertainty uInhom was set to the experimentally determined between-bottle standard deviation as the best estimate of the uncertainty due to betweenbottle heterogeneity [27]. Since homogeneity testing was carried out using 175 mg of material, the minimum sample intake for any user of the CRM must be above 175 mg. The expanded uncertainty could then be calculated by multiplication of the combined uncertainty with the coverage factor k = 2, i.e. U = 0.048%. Fig. 4. Measurement results with uncertainties, uc = combined uncertainty, 2uc = expanded uncertainty. S. Recknagel et al. / Analytica Chimica Acta 599 (2007) 256–263 263 Table 5 Comparison between volumetric titration and coulometry Method Volumetric titration Coulometry Coulometry Traceability of results Evaluation of determinations Number of single determinations Sample intake in mg Number of bottles investigated (m) Number of single determinations per bottle (n) Result calculated as Mass fraction in % Standard deviation of mean in %; S (RM); S (sample) SRM 84k (NIST) Mean value 245 175 35 7 (no. of factor determination) MMW mean of means 99.987 S (RM) = 0.007; S (sample) = 0.010; 0.012 Directly to SI Mean value 9 514–818 3 3 MMW mean of means 99.994 S (sample) 0.005 Recovery plot ncoul = f (nweight ) 17 0.5–817.9 3 ≥5 Slope coul. vs. sample intake 99.990 Error of slope 0.004 Uncertainty type A in % (uA ) 0.005 0.003 0.004 Uncertainty type B in % (uB ) Combined uncertainty in % (uc ) Expanded uncertainty in % (U; k = 2) 0.020 0.021 0.041 √S n 5. Conclusions The different results for both methods used for the determination of the mass fraction of KHP are summarised in Table 5. Acidimetric purity of KHP is statistically not different from 100% purity. Very precise volumetric titration with regards to bias and repeatability using optimised procedures shows within uncertainties the same results as measurements with precision coulometry. Whereby the volumetric method is calibrated with a certified reference material and the coulometric determination rely on traceable SI quantities. The purity of KHP is the same with slightly different uncertainties while the throughput of the volumetric titration is significantly higher than that of coulometric titration. This material is well characterised and can be used as an acidimetric standard with low uncertainty. Thus, it forms the basis for traceability of a high number of acid–base titrations. Acknowledgements The authors want to thank J. Wüthrich from Sigma–Aldrich Production GmbH for the successful cooperation and fruitful discussions. References [1] M. Máriássy, L. Vyskocil, A. Mathiasová, Accred. Qual. Assur. 5 (2000) 437–440. [2] P. Taylor, H. Kipphardt, P. DeBievre, Accred. Qual. Assur. 6 (2001) 103–106. [3] M.J.T. Milton, A Marschal, Accred. Qual. Assur. 6 (2001) 270–271. [4] J.K. Taylor, S.W. Smith, J. Res. Natl. Bur. Stand. 63A (1959) 153. [5] E.J. Eckfeldt, E.W. Shaffer Jr., Anal. Chem. 37 (1965) 1534. [6] J. Knoeck, H. Diehl, Talanta 16 (1969) 567. [7] T. Yoshimori, T. Tanaka, Bull. Chem. Soc. Jpn. 52 (1979) 1366. [8] K.W. Pratt, Anal. Chim. Acta 289 (1994) 125–134. 0.006 0.007 0.013 √S m 0.006 0.007 0.014 [9] K.W. Pratt, Anal. Chim. Acta 289 (1994) 135–142. [10] T. Ashakai, Y. Kakihara, Y. Kozuka, S. Hossaka, M. Murayama, T. Tanaka, Anal. Chim. Acta 567 (2006) 269–276. [11] T. Ashakai, M. Murayama, T. Tanaka, Accred. Qual. Assur. 12 (2007) 151–155. [12] Katalog, Unikale Prüfeinrichtungen für Chemie- und Materialtechnik, BAM, Berlin, in press (http://www.bam.de/de/service/publikationen/ publikationen medien/unikale pruef/upe 101 de 1-1.pdf). [13] M. Máriássy, M. Breitenbach, C. Oberroeder, E. Hwang, Y. Lim, K.W. Pratt, A. Hioki, T. Suzuki, M. Liandi, W. Bing, S. Yu, A.V. Skutina, G.I. Terentiev (2006). CCQM-K34 Final Report: Assay of potassium hydrogen phthalate, Metrologia 43, Tech. Suppl., 2006, 08008. [14] H. Felber, S. Rezzonico, M. Mariassy, Metrologia 40 (2003) 249–254. [15] R. Brown, M. Milton, P. Brewer, R. Wielgosz, NPL Report COAM 5, 2001, 36 p. [16] R. Brown, M. Milton, P. Brewer, NPL Report COAM 18, 2003, 25 p. [17] T. Yoshimori, Rev. Anal. Chem. 6 (1982) 13–48. [18] W. Hässelbarth, G. Bauer, U. Beck, M. Breitenbach, J. Ehreke, M. Golze, W. Gräfe, M. Hedrich, M. Krause, C. Müller, S. Noack, A. Nymschefsky, A. Recknagel, S. Recknagel, G. Riebe, M. Scharmach, T. Steiger, C. Ullner, Cataloque of Reference Procedures provided by BAM, 2001, pp. 24, ISSN 1617–6634 (http://www.bam.de/de/fachthemen/ referenzverfahren/referenzverfahren medien/107de.pdf). [19] Fluka-Catalog, 2007/2008: No: 60357. [20] M.E. Wieser, Pure Appl. Chem. 78 (2006) 2051–2066, http://www. chem.qmul.ac.uk/iupac/AtWt/. [21] P.J. Mohr, B.N. Taylor, J. Phys. Chem. Ref. Data 28 (1999) 1713–1852, http://physics.nist.gov/cgi-bin/cuu/Value?f. [22] B.A. Németh, Chemical Tables, Akadémiai Kiadó, Budapest, 1975. [23] W. Hässelbarth, BAM-Forschungsbericht Nr. 266, 2004, ISBN 386509-212-8 (http://www.bam.de/de/service/publikationen/publikationen medien/leitfaden messunsicherheit.pdf). [24] International Standards Organisation, Guide to the Expression of Uncertainty in Measurement, Geneva, 1993. [25] M. Weber, J. Wüthrich, S. Rezzonico, Microchim. Acta 146 (2004) 165–171. [26] NIST Certificate of Analysis SRM 84k, Gaithersburg, USA, 1999, http://www.clarksonlab.com/84k.pdf. [27] S.L.R. Ellison, S. Burke, R.F. Walker, K. Heydorn, M. Månsson, J. Pauwels, W. Wegscheider, B. te Nijenhuis, Accred. Qual. Assur. 6 (2001) 274–277.
© Copyright 2025 Paperzz