American Mineralogist, Volume 89, pages 1553–1556, 2004 Piston-cylinder calibration at 400 to 500 MPa: A comparison of using water solubility in albite melt and NaCl melting DON R. BAKER* Earth and Planetary Sciences, McGill University, 3450 rue University, Montréal, QC Canada H3A 2A7 ABSTRACT A crushable alumina-pyrex-NaCl solid-media assembly for the piston-cylinder apparatus was calibrated by NaCl melting at 500 MPa, 920 °C, and used to assess a new calibration technique based upon the solubility of water in albite melt at 400 to 500 MPa, 800 and 1200 °C. Use of the “difference from 100” technique to measure water solubility with the electron microprobe produces a calibration accuracy of ± 25 MPa. Both calibration techniques agree and demonstrate that the real pressure is 50 MPa greater than the nominal pressure. The water solubility in albite calibration is advantageous because it can be used to demonstrate that pressure, temperature, and time have no effect on the pressure calibration. Additionally, a single capsule containing albite + water can act as an in situ pressure monitor in solid-media assemblies containing other experiments. These calibrations demonstrate that the piston-cylinder apparatus can be used reliably at middle-to-upper crustal conditions for long-duration experiments. INTRODUCTION The invention of the piston-cylinder apparatus by Boyd and England (1960) provided us with an easy-to-use, reliable tool to investigate processes occurring in the deep crust and upper mantle. Routine use of the piston cylinder at pressures between 800 MPa and 5.0 GPa in laboratories around the world has resulted in thousands of scientific contributions that have shaped our understanding of igneous petrogenesis and high-temperature geochemistry in the lower crust and upper mantle. However, these machines are not routinely used at lower pressures. At pressures below 800 MPa, the internally heated pressure vessel is commonly the preferred apparatus for high-temperature geochemical studies (Holloway and Wood 1988), but these vessels are expensive and difficult to operate and maintain. In comparison to most internally heated pressure vessels, the piston-cylinder apparatus has the advantage of more-rapid heating and cooling; this rapidity is important in kinetic studies (e.g., H2O and CO2 diffusion) and equilibrium studies involving low-viscosity melts that may not quench to a glass unless cooled very rapidly. Here I show that the piston-cylinder apparatus can be calibrated and used reliably at pressures as low as 400 MPa and temperatures from 800 to 1200 °C, making them useful machines for the study of processes occurring in the middle-to-upper crust. The principle behind the piston-cylinder is that a moderate oil pressure, at most a few hundred megapascals, exerts force on a large diameter piston that then pushes a small diameter piston. The small piston compresses a solid-media assembly inside a cylindrical pressure vessel to high pressure. The solid-media assembly is composed of nested cylindrical bushings, and the sample is situated at the center of the assembly (Fig. 1). One of the bushings is made of graphite through which a current runs that can heat the sample to temperatures in excess of 2000 °C; * E-mail: [email protected] 0003-004X/04/0010–1553$05.00 because of the small thermal mass of the assembly, it heats and quenches rapidly. The nominal pressure on the sample is determined by the ratio of the piston areas and the pressure applied to the large piston. However, it is well known that the sample pressure is typically different from the nominal pressure. The discrepancy between the nominal and actual pressure is attributed to friction effects between the various portions of the assembly, between the assembly and the wall of the pressure vessel, and between the piston and the wall of the pressure vessel; the extent of these effects depends upon the pressurization and heating history (e.g., Boyd and England 1960; Johannes et al. 1971; Mirwald et al. 1975). In addition, differential loading due to bushings of different strength in the assembly also can cause a discrepancy between the nominal and actual pressure on the sample and create pressure differentials across the assembly (Mirwald et al. 1975). The value of the pressure correction necessary to convert from the nominal to the real pressure is determined by comparison of the nominal pressure of a selected phase transition measured in the piston-cylinder with the actual pressure of the transition. The actual pressure of the transition is measured in another type of apparatus, ideally an internally heated pressure vessel (cf., Holland 1980). Commonly used transitions are the albite breakdown reaction (e.g., Johannes et al. 1971; Holland 1980), the melting points of metals (e.g., Mirwald et al. 1975), and of alkali halides (e.g., Bohlen 1984). Each type of phase transition provides a unique calibration point in pressure and temperature space. Importantly, none of these phase transitions allow us to calibrate the piston-cylinder apparatus isobarically at multiple temperatures, thus possible temperature effects on the pressure calibration remain largely unknown. In order to minimize, or even eliminate, measurable differences between the nominal and sample pressures, the solid-media assemblies used inside the piston-cylinder apparatus have been 1553 1554 BAKER: PISTON–CYLINDER CALIBRATION USING WATER SOLUBILITY IN ALBITE Thermocouple Alumina ceramic Tool steel Pyrex glass NaCl Graphite Crushable alumina Samples in noble metal capsules Pb foil Pyrex glass 19.1 mm FIGURE 1. 19.1 mm crushable alumina-pyrex-NaCl assembly used in these experiments. modified by the replacement of bushings originally made of boron nitride, pyrophyllite, and talc by ones composed of NaCl, or other salts or carbonates (Johannes et al. 1971; Mirwald et al. 1975, Boettcher et al. 1981; Bohlen 1984; Rosenbaum et al. 1994). Because the low-pressure limit of the piston-cylinder apparatus is determined by the strength of the materials comprising the solid-media assembly, the use of NaCl bushings allows the piston-cylinder apparatus to reach pressures as low as 500 MPa (Bohlen 1984). However, the maximum sample temperature is limited by the NaCl solidus at the pressure of interest (e.g., 918 °C at 500 MPa, Bohlen 1984). Because many geochemical processes occur at higher temperatures, other materials more refractory than NaCl must used in the assemblies. But these materials are stronger than NaCl and their use might increase the minimum pressure attainable in a piston-cylinder apparatus and create a significant difference between the nominal and the sample pressures due to differential loading. Additionally, the strength of materials is temperature sensitive and a pressure calibration at one temperature may not be applicable to experiments at another. This paper presents the results of low-pressure calibration of the crushable alumina-pyrex-NaCl assemblies normally used in our laboratory. I show that the water solubility in albite melt determined by Behrens et al. (2001) can be used for piston-cylinder apparatus pressure calibration with an easily attainable precision of ±25 MPa. This method of calibration is consistent with that determined from the melting of NaCl, but more flexible in that calibrations can be performed at constant pressure and varying temperature. The calibrations demonstrate that the assemblies have an average friction correction of +50 MPa over the range of temperatures from 800 to 1200 °C and 400 to 500 MPa. EXPERIMENTAL AND ANALYTICAL TECHNIQUES All experiments were performed in 19.1 mm crushable alumina-pyrex-NaCl assemblies (Fig. 1) in a piston-cylinder apparatus. For each run, the assembly was cold pressurized to a nominal pressure of 320 MPa and then simultaneously heated and continuously pressurized to run conditions such that at 500 °C the nominal pressure was 400 MPa, and at 800 °C the nominal pressure was 450 MPa. For 500 MPa experiments, this nominal pressure was maintained until the experiment reached the desired final temperature; for experiments that needed to reach 550 MPa, the sample was additionally pressurized to a nominal pressure of 500 MPa at 1000 °C and this pressure maintained until run temperature was reached. Pressure was controlled to within 32 MPa during the experiments, which in some cases required either episodic pumping or venting of the piston cylinder. Temperatures were measured with type C, W-Re thermocouples and controlled to within 3 °C of the setpoint. Experiments were quenched isobarically to below 600 °C in less than 20 s. I initially calibrated the assemblies using the melting temperature of NaCl at 500 MPa, 918 °C (Bohlen 1984). Reagent-grade NaCl plus a Pt sphere of approximately 0.5 mm diameter were placed into a 3 mm diameter Au75Pd25 capsule with the sphere at the top. The open capsule was dried at 110 °C for a minimum of 1 h before being welded closed. Melting during experiments was detected by falling of the sphere. Based upon 20 minute duration experiments at 920 °C and nominal pressures of 475 MPa, where no melting occurred, and 425 MPa, where melting occurred, I determined the need to add 50 MPa to the nominal pressure to find the real pressure; all pressures for the albite water solubility experiments include this correction. The starting material for albite water-saturation experiments was an albite glass synthesized from carbonates and oxides by repeated melting and grinding; the analyzed composition of the starting material is homogenous and almost pure albite: Na0.97Al1.00Si3.03O8. Powder of this glass plus 11 to 16 wt% distilled H2O were placed inside 3 mm diameter Pt or Au75Pd25 capsules and welded closed without volatile loss. Sealed capsules with albite glass and water were heated at 110 °C for a minimum of one hour to check for leaks and to homogenize the water along the grain boundaries of the powder. The capsules were then placed into holes drilled in the crushable alumina of the solid media assembly and backfilled with pyrophyllite to minimize water loss from the capsules by hydrogen diffusion. Albite water solubility experiments at 500 MPa (AB-31 and -32) were pressurized and heated to 500 MPa, 1200 °C and held at those conditions for 1 hour before quenching. All other experiments were pressurized and heated to 550 MPa, 1200 ˚C, where the melts were water undersaturated, and held at those conditions for 1 hour to allow all the water present in the capsule to dissolve completely into the melt. Following this step, the pressure was reduced at a rate of approximately 1 MPa/s to the pressure of interest and held at the pressure and temperature of interest for the durations listed in Table 1. Experiments whose final temperature was 800 °C (AB-19, -21, -22) were hydrated at 1200 °C, 550 MPa for one hour before isobarically lowering the temperature to 800 °C, followed by decreasing the pressure to 450 MPa. All experiments were highly vesiculated and most contained a free water phase when opened. The water concentration was determined by the “difference from 100” technique. This technique assumes that the difference between the analytical sum determined by electron microprobe analysis and 100% is the concentration of dissolved water in the melt. Although not as accurate as infrared techniques and Karl Fischer analysis (see Behrens et al. 2001), this technique has been shown to be accurate to within approximately 0.5 wt% H2O (Devine et al. 1995; Thomas 2000); this value is similar to the analytical uncertainty based upon multiple analyses of the experiments from this study (Table 1). Electron microprobe analysis at McGill University used albite as the standard for Na, and orthoclase for Al and Si. Counting times for Na were 10 s on the peak and 5 s on each background, whereas Al and Si were each counted for 20 s on the peak and 10 s on each background. A 15 kV accelerating voltage with a 5.5 nA current on the sample and a beam diameter of 20 micrometers was used for the analysis. These conditions were chosen such that Na volatilization during analysis was not observed; such volatilization would produce artificially high water concentration values. The lack of Na volatilization was determined by comparing the stoichiometry of the hydrated glass with the anhydrous starting glass. Our analyses were repeated in at least four locations in the sample and demonstrate the homogeneous distribution of water. Water-undersaturated experiments quenched directly from 550 MPa, 1200 °C with 11.1 wt% added water produced micrometer-sized quench bubbles. These experiments were imaged on the electron microprobe using back-scattered electrons and the bubble areas counted. These experiments produced approximately 1.5 vol% quench bubbles, which when combined with a maximum estimate of the capsule volumes after the experiment indicates that the water loss during quenching was less than our analytical uncertainty. Measured water concentrations were not corrected for water loss during quenching. RESULTS AND DISCUSSION The results listed in Table 1 and displayed in Figure 2 present the measured water solubilities in albite melts at 392, 450, and 500 MPa, 1200 and 800 °C. Comparison between these data and BAKER: PISTON–CYLINDER CALIBRATION USING WATER SOLUBILITY IN ALBITE TABLE 1. Albite water solubility determinations in calibration experiments 15 14 Experiment the albite water solubility measurements of Behrens et al. (2001), obtained from experiments in cold-seal and internally heated pressure vessels, demonstrates excellent agreement. Almost all average water solubilities measured in this study are within 1σ of the 4th-order polynomial fit to the Behrens et al. (2001) measurements (Fig. 2). This relationship was inverted to calculate the pressure from the water solubility: P = –67.061 + 48.777 H2O – 2.2576 H2O2 + 0.42747 H2O3 – 0.019803 H2O4, where P is in MPa and H2O is the wt% water measured in the glass. Comparison of the calculated pressures from the water solubility with the calibrated pressure demonstrates that both the NaCl calibration and the water solubility pressure calibration techniques typically agree to within 25 MPa (Table 1). The average difference between the nominal pressures of the experiments, 50 MPa below those listed in Table 1, and the pressures measured by water solubility indicate that the true pressure in the experiments was equal to the nominal pressure +48 MPa. This difference is well within the uncertainty of the NaCl calibration technique and demonstrates the agreement of both calibration techniques. Replicate albite water-solubility experiments at the same temperature and pressure conditions, but in different assemblies, indicate a maximum difference of 31 MPa between experiments (Table 1). This variation is only slightly above the analytical uncertainty. I performed duplicate experiments in the same assemblies to investigate the possibility of significant pressure differentials in the assembly and to assess the reproducibility of the experiments. These experiments yielded pressures with overlapping uncertainties. The measured pressure differences between the duplicate experiments are attributed to analytical uncertainties rather than a true pressure differential within the assembly. The albite water-solubility technique allows us to calibrate Water solubility in albite melt 13 12 H2O (wt%) T Pressure* Duration No. H2O† Pcalc‡ (MPa) (°C) (MPa) (h) analyses (wt%) 1200 500 1 9 11.29 (0.39) 489 (19) AB-31§ 1200 500 1 10 11.95 (0.57) 519 (26) AB-32§ AB-18 1200 450 32 6 10.35( 0.30) 443 (16) 1200 392 2 8 9.95 (0.31) 422 (16) AB-25§ 1200 392 2 8 9.74 (0.47) 411 (25) AB-26§ AB-27 1200 392 2 8 9.38 (0.47) 391 (25) AB-29 1200 392 8 8 9.40 (0.38) 392 (21) AB-19 800 450 2 5 10.17 (0.41) 433 (22) 800 450 32 4 9.84 (0.42) 416 (22) AB-21§ 800 450 32 6 10.28 (0.46) 439 (24) AB-22§ * Nominal pressure + 50 MPa, based upon calibration with NaCl melting at 500 MPa, 920 °C. † Water solubility in albite melt determined by the difference between the electron microprobe analytical total and 100 wt%. The average water concentration is followed in parentheses by the 1-sigma standard deviation of the electron microprobe analytical totals. ‡ Pressure is calculated from the measured water solubility and a fourth-order polynomial equation fit to the solubilities reported in Behrens et al. (2001). Numbers in parentheses are the uncertainties in the pressure based upon the 1-sigma standard deviation of the water concentration. § These sequentially numbered runs were performed simultaneously in the same 19.1 mm assembly. 1555 11 10 9 8 o 7 1200 C o 800 C Behrens et al. (2001) 6 5 250 300 350 400 450 500 550 Pressure (MPa) FIGURE 2. Water solubility in albite melt as a function of the experimental pressure based upon calibration using the melting of NaCl at 500 MPa. Error bars represent uncertainty in the measured water concentrations based upon the 1σ standard deviation of the analyses from this study and from Behrens et al. (2001). the piston cylinder at constant pressure and temperatures that span the range of most igneous processes. An advantage of using water solubility in albite is its virtual insensitivity to temperature at the investigated pressures (Behrens et al. 2001). The pressures determined from water concentrations in experiments at 450 MPa, 1200 and 800 °C, are within 28 MPa of each other, slightly larger than the measurement uncertainties. This variation is less than that seen in replicate experiments performed in different assemblies at 392 MPa, 1200 °C. Thus, there is no significant temperature effect on the pressure correction for these assemblies. An advantage of using albite water solubility to calibrate pressures in piston-cylinder apparatus is that the rapid kinetics of water diffusion and vesiculation in silicate melts ensures that the pressure measured by albite water solubility represents the final pressure in the capsule. Because of this property, we can easily investigate the possibility of temporal changes in the pressure calibration, which is impossible to do using the falling sphere technique to determine NaCl melting (although more complicated techniques such as electrical conductivity and differential thermal analysis can be used with NaCl melting to investigate temporal changes in pressure calibration). Changes in the pressure calibration might be expected as, with time, the crushable alumina inside the assembly sinters into a hard ceramic and reacts with the pyrex pedestal. Experiments of different durations were performed at 1200 and 800 °C to investigate whether the relationship between the nominal and the real pressure changed with time. However, no such effect was observed and the pressure correction observed in short experiments is applicable to longer ones. DISADVANTAGES AND ADVANTAGES OF THE WATER SOLUBILITY PRESSURE CALIBRATION The albite water solubility calibration technique for the piston-cylinder apparatus is accurate to approximately ±25 MPa when the water concentration in the quenched glasses is determined by electron microprobe analysis. This precision is limited by the ± 0.5 wt% uncertainty in the electron microprobe “difference from 100” technique. This accuracy in pressure is about an order of magnitude less than that attainable in cold-seal 1556 BAKER: PISTON–CYLINDER CALIBRATION USING WATER SOLUBILITY IN ALBITE pressure vessels or internally heated pressure vessels where the sample pressure is measured directly. However, this level of accuracy is sufficient for most petrologic and kinetic studies where experiments are commonly performed at approximately 100 MPa intervals; additionally, the piston-cylinder apparatus has the advantage of a higher temperature limit and more-rapid heating and cooling compared with cold-seal and internally heated pressure vessels. Other, more-accurate and time consuming methods of water analysis (e.g., infrared spectrometry, Karl Fischer titration) may be used to increase the precision of the calibration by a factor of two to three (see Behrens et al. 2001). Such techniques were not used in this study because of my desire to develop a routine, rapid, easy, and inexpensive, technique for piston-cylinder calibration. These results demonstrate that the piston-cylinder apparatus with a crushable alumina-pyrex-NaCl solid media assembly can be calibrated easily by measuring water solubility in albite melt and routinely used for long-duration experiments at pressures as low as 400 MPa. Attempts to perform experiments below 392 MPa failed because of poor electrical contact between the piston and the assembly resulting in furnace failures. The maximum temperature limit of these assemblies is at least 1300 °C at 500 MPa based upon diffusion experiments we have performed in them. Neither temperature nor time affects the pressure calibration of the solid-media assembly used in this study. The albite water solubility technique allows calibration of a piston-cylinder apparatus in a single experiment; this is a slight advantage compared to the minimum of a pair of experiments needed for the NaCl melting calibration with a falling sphere. However, unlike the NaCl melting calibration, the water solubility technique can determine the effects of temperature and duration on the pressure calibration of a solid-media assembly. Additionally, the ability to calibrate with just a single capsule containing albite + water suggests the possibility of including such a capsule in the solid-media assembly for use as an in situ pressure monitor for critical experiments. ACKNOWLEDGMENTS Thoughtful reviews from Dana Johnston and David Walker were very much appreciated and resulted in an improved manuscript. Lang Shi is thanked for his help in choosing the best analytical parameters for electron microprobe analysis and keeping the machine in excellent operating condition. Comments on this manuscript by Carmela Freda were very helpful. Some of the experiments performed in this study were done with the aid of Phyllis Lang, Jean-Francois Bergevin, Leslie Gold, Yanan Liu, and Geneviève Robert. Funding for this study was provided by an NSERC Discovery grant to D.R.B. REFERENCES CITED Behrens, H., Meyer, M., Holtz, F., Benne, D., and Nowak, M. (2001) The effect of alkali ionic radius, temperature, and pressure on the solubility of water in MAlSi3O8 melts (M = Li, Na, K, Rb). Chemical Geology, 174, 275−289. Boettcher, A.L., Windom, K.E., Bohlen, S.R., and Luth, R.W. (1981) Low-friction, anhydrous, low- to high-temperature furnace assembly for piston-cylinder apparatus. Reviews of Scientific Instrumentation, 52, 1903−1904. Bohlen, S.R. (1984) Equilibria for precise pressure calibration and a frictionless furnace assembly for the piston-cylinder apparatus. Neues Jahrbuch für Mineralogie Monatshefte, 9, 404−412. Boyd, J.R. and England, J.L. (1960) Apparatus for phase-equilibrium measurements at pressures up to 50 kilobars and temperatures up to 1750 °C. Journal of Geophysical Research, 65, 741−748. Devine, J.D., Gardner, J.E., Brack, H.P., Layne, G.D., and Rutherford, M.J. (1995) Comparison of microanalytical methods for estimating H2O contents of silicic volcanic glasses. American Mineralogist, 80, 319−328. Holland, T.J.B. (1980) The reaction albite = jadeite + quartz determined experimentally in the range 600–1200 °C. American Mineralogist, 65, 129−134. Holloway, J.R. and Wood, B.J. (1988) Simulating the Earth, 196 p. Unwin Hyman, Winchester, Massachusetts. Johannes, W., Bell, P.M., Boettcher, A.L., Chipman, D.W., Hays, J.F., Mao, H.K., Newton, R.C., and Seifert, F. (1971) An interlaboratory comparison of pistoncylinder pressure calibration using the albite-breakdown reaction. Contributions to Mineralogy and Petrology, 32, 24−38. Mirwald, P.W., Getting, I.C., and Kennedy, G.C. (1975) Low-friction cell for piston-cylinder high-pressure apparatus. Journal of Geophysical Research, 80, 1519−1525. Rosenbaum, J.M., Kyser, T.K., and Walker, D. (1994) High temperature oxygen isotope fractionation in the enstatite-olivine-BaCO3 system. Geochimica et Cosmochimica Acta, 58, 2653−2660. Thomas, R. (2000) Estimation of water contents of granite melt inclusions by confocal laser Raman microprobe spectroscopy. American Mineralogist, 85, 868−872. MANUSCRIPT RECEIVED FEBRUARY 12, 2004 MANUSCRIPT ACCEPTED APRIL 22, 2004 MANUSCRIPT HANDLED BY JOHN AYERS
© Copyright 2026 Paperzz