Journal of Analytical Toxicology 2013;37:227 –232 doi:10.1093/jat/bkt008 Advance Access publication March 6, 2013 Article EtG/EtS in Urine from Sexual Assault Victims Determined by UPLC –MS-MS Solfrid Hegstad1*, Arne Helland1, Cecilie Hagemann2,3, Lisbeth Michelsen1 and Olav Spigset1,4 1 Department of Clinical Pharmacology, St. Olav University Hospital, Trondheim, Norway, 2Department of Public Health and General Practice, Norwegian University of Science and Technology, Trondheim, Norway, 3Department of Obstetrics and Gynecology, St. Olav University Hospital, Trondheim, Norway and 4Department of Laboratory Medicine, Children’s and Women’s Health, Norwegian University of Science and Technology, Trondheim, Norway *Author to whom correspondence should be addressed. Email: [email protected] In cases of sexual assault, victims often present too late for the detection of ethanol in biological samples. An ultra-performance liquid chromatography –tandem mass spectrometry (UPLC–MS-MS) method was developed and validated for the determination of ethyl glucuronide (EtG) and ethyl sulfate (EtS) in urine. Sample preparation prior to UPLC –MS-MS analysis was a simple sample dilution. The calibration ranges were 0.2 –20 mg/L, and between-assay relative standard deviations were in the range of 0.7– 7.0% at concentrations of 0.3, 3.0 and 7.0 mg/L. Urine samples were analyzed from 59 female patients presenting to the Sexual Assault Centre at St. Olav University Hospital in Trondheim, Norway between November 2010 and October 2011. EtG and EtS results were fully concordant, and positive in 45 of the 48 cases with self-reported alcohol intake. In contrast, ethanol was detectable in only 20 of these cases, corresponding to sensitivities of 94 and 42%, respectively. Of the patients reporting no alcohol intake, none had positive EtG/EtS findings. These data show that analysis of EtG and EtS greatly increases the detection window of alcohol ingestion in cases of sexual assault, and may shed additional light on the involvement of ethanol in such cases. The victims’ self-reported intake of alcohol seems to be reliable in this study, according to the EtG/EtS findings. Introduction Ethyl glucuronide (EtG) and ethyl sulfate (EtS) are nonoxidative minor metabolites of ethanol formed by the conjugation of ethanol to glucuronic acid via uridine diphosphate (UDP)-glucuronosyltransferases and to sulfate via sulfotransferases, respectively (1 –3). EtG and EtS are detectable in urine for considerably longer than ethanol. The detection times for EtG and EtS in urine range from hours up to several days, depending on the amount consumed. Therefore, urinary testing of these metabolites can be used for the identification of recent alcohol consumption, even after the ingested ethanol itself has been eliminated from the body (4– 8). Drug and/or alcohol intake is often implied in cases of sexual assault, either voluntary or as a result of suspected spiking of drinks. Ethanol, either alone or together with recreational/illicit drugs, has been the most common finding in previous surveys of alleged drug-facilitated sexual assault (9 –11). Many victims hesitate to seek assistance, which often leads to a considerable delay in obtaining urine samples for analysis. In such cases, the measurement of EtG and EtS can represent a valuable supplement to ethanol analysis, and contribute to a more comprehensive evaluation of the victim’s state of intoxication at the time of the assault. Several liquid chromatography – mass spectrometry (LC – MS) and liquid chromatography – tandem mass spectrometry (LC – MS-MS) methods have been described in the literature for the determination of EtG and EtS in urine (3, 12 – 16). Recently, two ultra-performance liquid chromatography (UPLC) – MS-MS methods have been described for the determination of EtG (17), and EtG and EtS (18), respectively, in urine. The aims of the present study were to develop and validate a fast and selective UPLC –MS-MS method for the determination of EtG and EtS in urine, using a simple sample preparation by means of a Tecan Freedom Evo pipetting robot platform, to apply the method on urine samples from a group of female patients subjected to sexual assault and to explore whether EtG/EtS analysis can aid the investigation of possible alcohol involvement in such cases. Material and Methods Patients and sampling Urine samples were obtained from 59 female patients 12 years who were examined at the Sexual Assault Centre at St. Olav University Hospital in Trondheim, Norway, between November 2010 and October 2011. Self-reported alcohol consumption in relation to the assault, in addition to the time lapse from the end of alcohol intake to urine sampling, were estimated by assessment of the patient files. Intake of alcohol was converted to alcohol units and classified in the following groups: no alcohol intake, intake of 1 –6 alcohol units, intake of .6 alcohol units and intake of an unknown/uncertain amount. Six alcohol units was chosen as a reasonable limit separating social drinking from excessive drinking that may cause drunkenness. The national Norwegian definition of one alcohol unit was used, i.e., 12.8 g of ethanol, which corresponds to approximately one standard can or bottle of 4.5% beer (330 mL), one standard glass of 12% wine (120– 150 mL) or one standard drink with 40% spirits (40 mL) (19). The present study represents a sub-sample of a larger study of 315 patients 12 years of age who presented to the Sexual Assault Centre from July, 2003, to October, 2011, and in which urine and/or blood were obtained for toxicological analyses. Exclusion criteria from the study were male sex, no (suspected) sexual assault or no medical examination performed. According to instructions from the Regional Committee for Research Ethics, which approved the study, these patients received a letter of information about the study. Those who declined to participate on the basis of this letter were also excluded. # The Author [2013]. Published by Oxford University Press. All rights reserved. For Permissions, please email: [email protected] Analysis of EtG and EtS Chemicals and reagents EtG, EtS, EtG-d5 and EtS-d5 were obtained from Lipomed GmbH (Weil am Rhein, Germany). Other chemicals were of LC–MS, high-performance liquid chromatography (HPLC) or analytical grade from various commercial sources. External quality control samples were purchased from Equalis (Uppsala, Sweden: http://www.equalis.se/sv/start.aspx). Preparation of solutions For each compound, stock solutions were prepared to a concentration of 1 mg/mL in methanol and were used for calibrator and quality control (QC) samples. The stock solutions were fortified with methanol (1,000 mg/L) and used for calibration and QC solutions prepared in urine with concentrations of 0.20, 0.50, 1.0, 5.0, 10.0 and 20.0 mg/L, and 0.30, 3.0, 7.0 and 15.0 mg/L, respectively. The internal standards EtG-d5 and EtS-d5 were diluted with water to a concentration of 2.5 mg/L. The stock solutions were stored at –208C; standards in urine and internal standard were stored at 48C. Sample treatment A 100 mL aliquot of a urine specimen was mixed with 100 mL of internal standard and diluted 10 times with water directly into a 96 well plate by Tecan Freedom Evo 100 or 150 pipetting robots (Tecan Nordic, Mölndal, Sweden). Instruments UPLC was performed using a Waters Acquity System (Waters, Manchester, UK). Separation was performed on a Waters HSS T3 column (2.1 100 mm, 1.8 mm) using gradient elution at a flow rate of 0.6 mL/min with the following binary solvent system: 0.1% formic acid in water (A) and 100 % methanol (B). The gradient was run as follows: 0 min, 99% A, 1% B; 1.0 min, 80% A, 20% B; 1.1 min 10% A, 90% B; 1.7 min, 99 % A, 1% B. The time before the next injection of 1 min was sufficient to equilibrate the column. The column temperature was kept constant at 508C. The injection volume was 5 mL performed by partial loop injection using needle overfill as injection technique. A dual wash was applied to the autosampler using 600 mL methanol –isopropanol–acetonitrile –water –ammonia (25:25:25:23:2, v/v), and 200 mL methanol–isopropanol – acetonitrile –water–formic acid (25:25:25:23:2, v/v), denoted as weak and strong wash. A Xevo TQ-S tandem-quadrupole mass spectrometer (Waters) was used, equipped with a Z-spray electrospray interface. Negative ionization was performed in the multiple reaction monitoring (MRM) mode. The capillary voltage was set to 1.0 kV, the source block temperature was 1208C and the desolvation gas nitrogen was heated to 6008C and delivered at a flow rate of 1,000 L/h. The m/z 221.0 . 85.1 and 221.0 . 75.1 transitions (cone voltages: 40 V; collision energy: 15 eV) were monitored for EtG, m/z 125.0 . 97.1 transition (cone voltages: 50 V; collision energy: 15 eV) was monitored for EtS, m/z 226.0 . 85.1 transition (cone voltages: 40 V; collision energy: 15 eV) was monitored for EtG-d5 and m/z 130.0 . 97.8 transition (cone voltages: 50 V; collision energy: 15 eV) was monitored for EtS-d5. System operation and data acquisition were 228 Hegstad et al. controlled using Mass Lynx 4.1 software (Waters). All data were processed with the Target Lynx quantification program (Waters). The analytes were identified by comparing the retention time of the corresponding calibrator and QC samples. The ratio between the two MRM transitions (EtG) was also compared with those of the corresponding calibrator and QC samples and should not deviate more than 20%. Method validation The six-point calibration curves (five replicates of each standard) were based on peak area ratios of the analyte relative to the internal standard using a weighted (1/x) linear line, which included the origin. Within-assay precision was estimated by the analysis of 10 separate replicates of QC samples at four concentrations in a single assay. Between-assay precision and accuracy were determined by the analysis of one replicate at four QC concentrations on 10 different days. Matrix effects (MEs) were evaluated by the method proposed by Matuszewski et al. (20). The analyte signal in the spiked water was compared with the analyte signal in the matrix, and the ME was defined as ME (%) ¼ (matrix area/water area) 100. Five replicates of urine sample extracts (from five different individuals) were analyzed. The concentrations were 0.30 and 7.0 mg/L. Negative urine samples supplemented with various concentrations of analytes (0.01– 0.10 mg/L) were analyzed to determine the limit of quantification (LOQ). Samples of 0.10 mg/L of EtG and 0.080 mg/L of EtS were run in one replicate on six different days, and the concentrations were calculated using a calibration curve in the ranges of 0.10 –5.0 mg/L (EtG) and 0.080 – 5.0 mg/L (EtS). The signal-to-noise (S/N) criteria for the LOQ samples were 10 and precision of calculated concentrations were within + 20%. The limit of detection (LOD) was determined by dilution and evaluation of S/N (3). Analysis of ethanol and creatinine Ethanol and creatinine were determined by the test kits EthanolGen2 (ETOH2) and Creatinine Jaffé Gen.2 (CRJ2U) on a Cobas Intergra 400þ multianalyzer (Roche Diagnostics Norway AS, Oslo, Norway). The analytical cutoff for ethanol was 10 mg/dL. Calculations and statistics The results from the analyses of ethanol, EtG and EtS were compared with the victims’ self-reported ethanol intake to determine the reliability of the method. The analytical efficiency is reported in terms of sensitivity, specificity, positive predictive value, negative predictive value and accuracy. For EtG and EtS, the concentrations in urine are dependent on whether urine is relatively concentrated or diluted. To correct for this factor when assessing the time-dependent elimination of EtG and EtS, the concentrations of EtG and EtS were adjusted to a standard creatinine concentration of 100 mg/dL, by using the following equation: standardized concentrations of EtG or EtS ¼ (measured concentrations of EtG or EtS/measured creatinine concentration in mg/dL) 100 mg/dL. In contrast, ethanol easily passes biologic membranes and the urinary level is rapidly equilibrated with the plasma level; thus, the ethanol concentration in urine is not subject to urinary dilution variances (21). Therefore, no creatinine adjustment is necessary for ethanol. Statistical analyses were performed with SPSS version 18.0. McNemar’s test was used to compare differences in the reliability of urinary ethanol, EtG and EtS for verifying self-reported alcohol intake. Spearman’s rank correlation was used to test for the relationship between concentrations of EtG and EtS, and between EtG/EtS concentrations and time from alcohol intake to urine sampling, respectively. P values , 0.05 were considered statistically significant. Results Method validation Calibration curves were made for each compound in the concentration range, as shown in Table I. The correlation coefficient was above 0.999 for both compounds. In cases with a concentration higher than 20 mg/L, the samples were diluted and reanalyzed. The MRM chromatograms of a real sample are shown in Figure 1. The calibration range, LOD, LOQ, within-assay precision, between-assay precision and bias for EtG and EtS are presented in Table I. The within-assay coefficients of variation (CVs) were 2.7 –4.9%, and the between-assay CVs were 0.7 – 7.0%. The bias was in the range of 21.0 to 4.0%. The observed matrix effects indicated some ion suppression for EtG and an ion enhancement for EtS (Table II). However, when corrected to the internal standard, the observed matrix effects and CVs were reduced significantly for both compounds. External quality control samples, which were run with the presented method, showed good correspondence in the low to medium concentration range. The Z-scores were jZj 1.3 for all measurements when compared to the mean value of the other laboratories using the same methodological platform (i.e., LC– MS-MS) as this study, thus indicating good accuracy for the quantitative results obtained with the method. Application The mean age of the 59 women included in the study was 24 years (range 14– 61 years). Of the women, 48 (81%) reported intake of alcohol in relation to the assault. The mean time from alcohol intake to when the urine samples were obtained was 20.9 h (range 1.5 –108 h). Ethanol was found in 20 samples, whereas EtG and EtS were found in 45 samples. In no cases, ethanol, EtG or EtS were found in urine from patients reporting no intake of alcohol. The positive ethanol concentrations ranged from 37 to 280 mg/dL. The positive EtG concentrations ranged from 0.34 to 1,123 mg/L and the positive EtS concentrations ranged from 0.18 to 322 mg/L (not corrected for creatinine). There was full concordance between the findings of EtG and EtS, and there was also a close and highly significant relationship (r ¼ 0.981; p , 0.001) between the concentrations of EtG and EtS (Figure 2), with a mean EtG/EtS ratio of 3.5. There were, however, large interindividual variations in this ratio, ranging from 1.4 to 7.4. Using self-reported intake of alcohol as a reference, the numbers of true and false positives and negatives, in addition to sensitivity, specificity, positive and negative predictive value and accuracy, are shown in Table III. The sensitivities and accuracies for EtG and EtS were significantly higher than for ethanol ( p , 0.001). The three subjects with false negative EtG/EtS results had the following characteristics: reported intake of three alcohol units, time interval of 15.5 h from alcohol intake to sampling (Case A); reported intake of one alcohol unit, time interval of 51 h from alcohol intake to sampling (Case B); reported intake of an unknown amount of alcohol, time interval of 65 h from alcohol intake to sampling (Case C). There was a significant decrease with time after alcohol intake in creatinine-corrected EtG and EtS concentrations (Figure 3). As shown in the figure, approximate detection times of 2 –3 days could be anticipated. Discussion Method validation Dilution of the urine sample prior to the LC–MS-MS analysis of EtG and EtS has been reported previously (14, 17, 18). Solid-phase extraction (SPE) with an anion-exchange column combined with LC–MS-MS has been shown to be a more reliable method for the quantification of EtG in urine than a direct injection LC– MS-MS method (17). EtG and EtS are highly polar analytes and a selective SPE method is difficult to conduct simultaneously for both compounds. The anion-exchange extraction method is not suitable for EtS due to low extraction recovery (22). Therefore, a simple and time-saving dilution method was developed with a sufficient selectivity and S/N ratio at the lowest calibration level for both compounds. In this study, calibrator and QC samples were prepared from the same vendor. An external QC system was used to assure the validity of the method. Table I Calibration Range, Correlation Coefficient, LOD, LOQ, Within-Assay and Between-Assay Precisions and Bias for EtG and EtS in Urine Analyte Calibration range (mg/L) Correlation coefficient (r value) (n ¼ 5) LOD (mg/L) LOQ (mg/L) Theoretical concentration (mg/L) Within-assay CV (%) (n ¼ 10) Between-assay CV (%) (n ¼ 10) Bias (%) (n ¼ 10) EtG 0.20– 20 0.999 0.030 0.10 EtS 0.20– 20 0.999 0.020 0.080 0.30 3.0 7.0 15.0 0.30 3.0 7.0 15.0 4.7 4.1 3.3 3.9 4.9 3.6 2.7 3.6 7.0 3.8 2.4 0.7 5.8 4.5 2.8 1.7 2.4 4.0 2.8 –0.9 2.5 3.1 3.4 –1.0 EtG/EtS in Urine from Sexual Assault Victims Determined by UPLC– MS-MS 229 Figure 1. MRM chromatograms of an authentic sample: EtG (A) (the peaks at 0.93 and 1.1 min are due to unknown interfering compounds observed to be present in some samples) EtS (B). Determined concentrations: EtG ¼ 1.9 mg/L, EtS ¼ 0.81 mg/L. Table II Evaluation of MEs for EtG and EtS in Urine Analyte EtG EtS Table III Results from the Analyses of Ethanol, EtG and EtS Compared to Self-Reported Intake of Ethanol in 59 Victims of Sexual Assault* Theoretical concentration (mg/L) ME (%) Relative ME, CV (%) ME corrected with internal standard (%) Relative ME to internal standard, CV (%) 0.30 7.0 0.30 7.0 82.9 82.2 159.7 138.9 11.7 11.1 12.7 15.0 98.2 94.3 107.4 99.8 5.1 3.2 2.4 2.6 Variable Ethanol EtG EtS True positives True negatives False positives False negatives Sensitivity Specificity Positive predictive value Negative predictive value Accuracy 20 11 0 28 20/48 ¼ 0.42 11/11 ¼ 1.0 20/20 ¼ 1.0 11/39 ¼ 0.28 31/59 ¼ 0.53 45 11 0 3 45/48 ¼ 0.94† 11/11 ¼ 1.0 45/45 ¼ 1.0 11/14 ¼ 0.79 55/59 ¼ 0.95† 45 11 0 3 45/48 ¼ 0.94† 11/11 ¼ 1.0 45/45 ¼ 1.0 11/14 ¼ 0.79 55/59 ¼ 0.95† *Note: there was a full concordance between the results of EtG and of EtS. † p , 0.001 versus ethanol. Figure 2. Scatterplot showing the relationship between EtG and EtS concentrations in 45 positive urine samples. The correlation is highly significant (r ¼ 0.981, p , 0.001); both axes are logarithmic. Application The results demonstrate that the analysis of EtG and EtS improves the sensitivity and detection times for ethanol in cases of sexual assault. EtG and EtS were detected in 45 out of 48 cases with self-reported intake of alcohol, whereas ethanol 230 Hegstad et al. was detected in only 20 cases, corresponding to sensitivities of 94 and 42%, respectively. This is in accordance with previous studies, which have shown that urinary EtG and EtS remain positive considerably longer than urinary ethanol (6, 7). The improved detection times and sensitivity for EtG and EtS may provide a more complete assessment of toxicological involvement in the investigation of cases of sexual assault, because the delay in reporting is particularly common in cases of drug facilitated sexual assault. The detection of EtG and EtS may corroborate the victims’ self-reported alcohol intake, and may lend credence to possible claims by victims of being assaulted while intoxicated. In studies in healthy volunteers, the excretion profiles for EtG and EtS in urine are well documented, and there is an established dose-response relationship between the amount of alcohol ingested and the detection time (6, 7). The current data on the relationship between reported intake, a time from assault to sampling and urinary concentrations of EtG and EtS (Figure 3) should be interpreted with caution due to the limited sample size and the uncertainties of recall in the selfreported amount of alcohol and time interval since the alcohol intake took place. Nevertheless, these results are in accordance with data from a previous study showing that following an Figure 3. Scatterplots showing the relationship between time since alcohol intake and EtG and EtS urinary concentrations standardized to a creatinine concentration of 100 mg/dL. Samples from patients reporting alcohol intake: 1– 6 units (A); . 6 units (B). The correlations between time after intake and concentrations are statistically significant (1– 6 units: r ¼ – 0.636 for EtG, r ¼ –0.704 for EtS; . 6 units: r ¼ –0.794 for EtG, r ¼ –0.786 for EtS; p , 0.01 for all); the plots are semi-logarithmic. From extrapolation of the regression lines to the intercept with the LOQs for EtG and EtS in this method (0.10 and 0.080 mg/L, respectively), detection times of 2 –2.5 days could be assumed for those ingesting 1– 6 alcohol units, whereas detection times of approximately 3 days could be assumed for those ingesting . 6 alcohol units. The detection times seem to be slightly longer for EtG than for EtS. intake of 0.50 g of ethanol per kg body weight (i.e., 35 g or approximately three alcohol units in a subject weighing 70 kg), the maximum detection time for EtG and EtS would be approximately 48 h (6). For the purpose of detecting alcohol intake in victims of sexual assault, this study suggests that the lowest possible cutoff (e.g., LOQ) should be employed to maximize the sensitivity of the method. The observed EtG concentrations were always higher than the EtS concentrations, with a mean ratio of 3.5. In other studies, the mean EtG/EtS ratios have varied between 2.6 and 4.0 (3, 6, 23). Thus, if the same cutoff values are applied for both EtG and EtS, samples in the lower range could turn out positive for EtG and negative for EtS. To avoid difficulties in the interpretation and reporting of such samples, a lower cutoff can be used for EtS than for EtG. Because the ratio between the LOQs for EtG and EtS in the present study was 1.25, there is still a risk that a sample containing both EtG and EtS could be positive for EtG and negative for EtS by employing this method. From the high concordance between self-reported alcohol intake and positive EtG and EtS findings in urine, it was concluded that the patients’ qualitative reporting of whether they did or did not ingest alcohol in relation to the alleged assault was highly credible in this study, although authors of studies from other Western sexual assault centers claim the contrary (24, 25). In none of the cases were EtG or EtS detected in samples from patients not reporting alcohol intake. In three cases, EtG or EtS could not be detected despite alcohol ingestion being reported by the patient. Two of these cases can easily be explained, in Case B by a low alcohol intake and in Case C by a long duration between intake and urine sampling. In Case A, the self-reported intake was three standard alcohol units and the time from assault to sampling was 15.5 h. This would be expected to yield positive EtG and EtS findings in most cases; however, the negative results could stem from individual peculiarities in ethanol metabolism and excretion, or simply from inaccuracies in the reporting of the amount of ethanol ingested and/or the time of the intake. In conclusion, this study reported the development and validation of a fast and reliable method for the simultaneous quantification of EtG and EtS in urine. The application of this method on a series of sexual assault cases confirms the superiority of EtG/EtS analysis over ethanol analysis for the verification of alcohol intake in conjunction with the assault, and shows a high concordance between victims’ self-reported ethanol intake and EtG/EtS findings. References 1. Foti, R.S., Fisher, M.B. (2005) Assessment of UDP-glucuronosyltransferase catalyzed formation of ethyl glucuronide in human liver microsomes and recombinant UGTs. Forensic Science International, 153, 109–116. 2. Helander, A., Beck, O. (2004) Mass spectrometric identification of ethyl sulfate as an ethanol metabolite in humans. Clinical Chemistry, 50, 936– 937. 3. Helander, A., Beck, O. (2005) Ethyl sulfate: A metabolite of ethanol in humans and a potential biomarker of acute alcohol intake. Journal of Analytical Toxicology, 29, 270–274. EtG/EtS in Urine from Sexual Assault Victims Determined by UPLC– MS-MS 231 4. Dahl, H., Stephanson, N., Beck, O., Helander, A. (2002) Comparison of urinary excretion characteristics of ethanol and ethyl glucuronide. Journal of Analytical Toxicology, 26, 201–204. 5. Wurst, F.M., Dresen, S., Allen, J.P., Wiesbeck, G., Graf, M., Weinmann, W. (2006) Ethyl sulphate: A direct ethanol metabolite reflecting recent alcohol consumption. Addiction, 101, 204–211. 6. Høiseth, G., Bernard, J.P., Stephanson, N., Normann, P.T., Christophersen, A.S., Mørland, J. et al. (2008) Comparison between the urinary alcohol markers EtG, EtS, and GTOL/5-HIAA in a controlled drinking experiment. Alcohol and Alcoholism, 43, 187–191. 7. Helander, A., Böttcher, M., Fehr, C., Dahmen, N., Beck, O. (2009) Detection times for urinary ethyl glucuronide and ethyl sulfate in heavy drinkers during alcohol detoxification. Alcohol and Alcoholism, 44, 55 –61. 8. Wurst, F.M., Metzger, J. (2002) The ethanol conjugate ethyl glucuronide is a useful marker of recent alcohol consumption. Alcoholism: Clinical and Experimental Research, 26, 1114–1119. 9. Jones, A.W., Kugelberg, F.C., Holmgren, A., Ahlner, J. (2008) Occurrence of ethanol and other drugs in blood and urine specimens from female victims of alleged sexual assault. Forensic Science International, 181, 40 –46. 10. Scott-Ham, M., Burton, F.C. (2006) A study of blood and urine alcohol concentrations in cases of alleged drug-facilitated sexual assault in the United Kingdom over a 3-year period. Journal of Clinical and Forensic Medicine, 13, 107– 111. 11. Hindmarch, I., ElSohly, M., Gambles, J., Salamone, S. (2001) Forensic urinalysis of drug use in cases of alleged sexual assault. Journal of Clinical and Forensic Medicine, 8, 197– 205. 12. Stephanson, N., Dahl, H., Helander, A., Beck, O. (2002) Direct quantification of ethyl glucuronide in clinical urine samples by liquid chromatography-mass spectrometry. Therapeutic Drug Monitoring, 24, 645–651. 13. Weinmann, W., Schaefer, P., Thierauf, A., Schreiber, A., Wurst, F.M. (2004) Confirmatory analysis of ethylglucuronide in urine by liquidchromatography/electrospray ionization/tandem mass spectrometry according to forensic guidelines. Journal of the American Society of Mass Spectrometry, 15, 188–193. 14. Politi, L., Morini, L., Groppi, A., Polini, V., Pozzi, F., Polettini, A. (2005) Direct determination of the ethanol metabolites ethyl glucuronide and ethyl sulfate in urine by liquid chromatography/electrospray tandem mass spectrometry. Rapid Communications in Mass Spectrometry, 19, 1321– 1331. 232 Hegstad et al. 15. Crews, B., Latyshev, S., Mikel, C., Almazan, P., West, R., Pesce, A. et al. (2010) Improved detection of ethyl glucuronide and ethyl sulfate in a pain management population using high-throughput LC-MS/MS. Journal of Opioid Management, 6, 415–421. 16. Albermann, M.E., Musshoff, F., Madea, B. (2012) A high-performance liquid chromatographic-tandem mass spectrometric method for the determination of ethyl glucuronide and ethyl sulfate in urine validated according to forensic guidelines. Journal of Chromatographic Science, 50, 51– 56. 17. Helander, A., Kenan, N., Beck, O. (2010) Comparison of analytical approaches for liquid chromatography/mass spectrometry determination of the alcohol biomarker ethyl glucuronide in urine. Rapid Communications in Mass Spectrometry, 24, 1737– 1743. 18. Høiseth, G., Yttredal, B., Karinen, R., Gjerde, H., Christophersen, A. (2010) Levels of ethyl glucuronide and ethyl sulfate in oral fluid, blood, and urine after use of mouthwash and ingestion of nonalcoholic wine. Journal of Analytical Toxicology, 34, 84 –88. 19. Nasjonalt Folkehelseinstitutt. (2008) Fakta om alkohol. http:// www.fhi.no/artikler?id=42834 (accessed 10 October 2012). 20. Matuszewski, B.K., Constanzer, M.L., Chavez-Eng, C.M. (2003) Strategies for the assessment of matrix effect in quantitative bioanalytical methods based on HPLC-MS/MS. Analytical Chemistry, 75, 3019–3030. 21. Jones, A.W. (1998) Lack of association between urinary creatinine and ethanol concentrations and urine/blood ratio of ethanol in two successive voids from drinking drivers. Journal of Analytical Toxicology, 22, 184– 190. 22. Hegstad, S., Johnsen, L., Mørland, J., Christophersen, A.S. (2009) Determination of ethylglucuronide in oral fluid by ultraperformance liquid chromatography– tandem mass spectrometry. Journal of Analytical Toxicology, 33, 204–207. 23. Halter, C.C., Dresen, S., Auwaerter, V., Wurst, F.M., Weinmann, W. (2008) Kinetics in serum and urinary excretion of ethyl sulfate and ethyl glucuronide after medium dose ethanol intake. International Journal of Legal Medicine, 122, 123–128. 24. Du Mont, J., Macdonald, S., Rotbard, N., Bainbridge, D., Asllani, E., Smith, N. et al. (2010) Drug-facilitated sexual assault in Ontario, Canada: toxicological and DNA findings. Journal of Forensic and Legal Medicine, 17, 333– 338. 25. Juhascik, M.P., Negrusz, A., Faugno, D., Ledray, L., Greene, P., Lindner, A. et al. (2007) An estimate of the proportion of drugfacilitation of sexual assault in four U.S. localities. Journal of Forensic Sciences, 52, 1396– 1400.
© Copyright 2026 Paperzz