J Med Dent Sci 2006; 53: 93–101 Original Article Influence of Number of Dental Autoclave Treatment Cycles on Rotational Performance of Commercially Available Air-Turbine Handpieces Masahiro Nagai and Kazuo Takakuda Department of Biodesign, Division of Biosystems, Institute of Biomaterials and Bioengineering, Tokyo Medical and Dental University The influence of number of autoclave treatment cycles (N ) on rotational speed and total indicated run-out of commercially available air-turbine handpieces from five manufacturers was investigated at N =0, 50, 100, 150, 200, 250 and 300 cycles, and the significance in the test results was assessed by Dunnett’’s multiple comparison test. Some air-turbine handpieces showed the significant differences in rotational speed at N =300 cycles, however, the decreases of the rotational speeds were only 1 to 3.5 percent. Some air-turbine handpieces showed the significant differences in total indicated run-out, however, the respective values were smaller than that at N =0 cycle. Accordingly, it can be considered that the ball bearing in the air-turbine handpieces is not affected significantly by autoclave. To further evaluate rotational performance, this study focused on the rotational vibration of the ball bearing components of the air-turbine, as measured by Fast Fourier Transform (FFT) analysis; the power spectra of frequency of the ball’’s revolution, frequency of the cage’’s rotation and frequency of the ball’’s rotation were comparatively investigated at N =0, 150 and 300 cycles, and the influence of autoclave was evaluated qualitatively. No abnormalities in the ball bearings were recognized. Key words: Autoclave sterilization, Rotational speed, Total indicated run-out, Air-turbine handpieces. Corresponding Author: Masahiro Nagai 2-3-10 Kanda-Surugadai, Chiyoda-ku, Tokyo 101-0062, Japan Received December 6, 2005; Accepted March 17, 2006 Introduction Social awareness of infectious disease such as hepatitis B, hepatitis C and acquired immune deficiency syndrome (AIDS) is increasing. In dental fields, there has long been concern that patients may contract such infectious diseases via contact with poorly sterilized instrument or appliances, particularly because patients commonly bleed during dental treatment. Furthermore, dentists, dental hygienists and assistants are also at risk of contracting such diseases via contact with patient blood and saliva. Therefore, sufficient measures to protect against infection must be taken1,2. In light of this social background, manufacturers of air-turbine handpieces have actively improved their products to facilitate autoclave sterilization. Dentists also understand the necessity of sterilizing air-turbine handpieces after each patient. However, there is some concern among dentists that if air-turbine handpieces are sterilized frequently under severe environmental conditions, such as those present in an autoclave, the rotational performance is adversely affected. In order to alleviate such misgivings, we performed a recent literature search regarding the durability of airturbine handpieces with sterilization, but were able to find only three reports3-5. Two reports3,4 did not focus on the rotational speed or eccentricity of the test mandrel; the testing criteria focused on the number of cycles until use of the air-turbine handpiece was impossible. Accordingly, no correlations between changes in rotational performance and sterilization treatment could be determined. The third report5 focused on the rotational speed, the stall torque, the bearing resistance, the 94 M. NAGAI and K. TAKAKUDA noise and the light output of fiber-optic of the air-turbine handpieces during the course of daily use in general dental practice. However, the report5 did not focus on the eccentricity of a rotational tool. In the present study, we investigated the influence of the number of autoclave cycles on the rotational performance (rotational speed and eccentricity) of the most commonly used commercially available air-turbine handpieces in Japan. Furthermore, we diagnosed the rotational conditions of ball bearing in the air-turbine handpieces by using the Fast Fourier Transform (FFT) analyzer. MATERIALS AND METHODS Air-turbine handpieces The air-turbine handpieces used in this study were commercially available products. They were made by four Japanese manufacturers and one German manufacturer, as shown in Table 1, with two products coming from each manufacturer. The air-turbine handpieces made by Osada and Yoshida used steel ball bearings, while the air-turbine handpieces made by Morita, Nakanishi and Kavo used ceramic ball bearings. In addition, the air-turbine handpieces made by Kavo were equipped with an automatic air pressure regulator. Furthermore, all the air-turbine handpieces included a device in the head that prevented oral fluids and other foreign substances from entering. J Med Dent Sci Autoclave sterilization conditions The sterilization conditions for small medical auto6 claves are specified in JIS T7324-2005 . In addition, the sterilization conditions for air-turbine handpieces, as specified in “Dental handpiece - high-speed air-turbine 7 handpieces” (JIS T5906 ), call for a sterilization temperature of 132±2°C and a sterilization pressure of 0.2 MPa for 5 min. Rotational performance must not deteriorate after autoclaving, even if sterilization treatment is repeated for 250 cycles. In this study, the autoclave used was the Super Clave HF260 (Hillson Dec Co.) and the sterilization temperature and pressure were set at 132±2°C and 0.2 MPa, respectively, as specified in JIS T59067. However, sterilization was performed for 15 min and the number of cycles was 300. These conditions were severer than the JIS regulations7. Testing procedure For the repeated sterilization test, one cycle (N=1 cycle) was as follows. First, the air-turbine handpieces were lubricated with turbine spray using nozzles provided by the manufacturers for 2 s, were immediately rotated without applied load for 30 s after inserting a test mandrel (Í1.6 × 19 mm), and were sterilized for 15 min in the autoclave after removing the test mandrel. When setting the air-turbine handpieces into the autoclave chamber, all air-turbine handpieces were tilted at about 20°, with the heads placed upward, in stainless steel container. The container was then set in the autoclave chamber. In addition, the positions of the air-turbine handpieces in the container were changed after Table 1. Commercially available air-turbine handpieces used in this test. INFULUENCE OF DENTAL AUTOCLAVE ON AIR-TURBINE every cycle in order to minimize the influence of air-turbine handpiece position. After each cycle, the container in which the air-turbine handpieces were placed was immediately removed from the autoclave chamber and was allowed to cool to room temperature. Lubrication was repeated as described above, and this was repeated for 300 cycles. None of the air-turbine handpieces were subjected to load during testing. This is because the present study was conducted in order to investigate the influence of autoclave on the ball bearing parts of the air-turbine handpiece. Evaluation method Rotational performance of the air-turbine handpieces was evaluated based on rotational speed, eccentricity of test mandrel in rotation and power spectral analysis of the ball bearings supporting the body of the rotating air turbine, without applied load. Rotational speed and eccentricity were measured every 50 cycles, and power spectral analysis was performed every 150 cycles using sound level meter (NA40, Rion Co.) and FFT analyzer (SA-74, Rion Co.). Rotational performance tests were carried out as follows: 1) Air-turbine handpieces were rotated without applied load; however, at cycle N=0, when the rotational speed of each air-turbine handpiece reached the manufacturer’s recommended rotational speed (R* in Table 2) by adjusting the compressor air valve, the supply air pressure, po, was measured using a digital pressure gauge connected at the coupling joint of the airturbine handpiece. This value (po) was used when supplying the air pressure used to rotate the respective air-turbine handpieces in the rotational performance Table 2. Each manufacturer’s recommended rotational speed, R * and the supply air pressure, po at N=0 cycle. 95 tests. In Table 2 the measured values of R* and po for each air-turbine handpiece are shown. 2) When arriving at the predetermined number of autoclave cycles, the air pressure of each air-turbine handpiece was adjusted to po via the compressor air valve (Table 2). 3) After the rotational speed of each air-turbine handpiece became stable, as the rotational performance tests, the rotational speed, RN, and the total indicated run-out, δN, were measured simultaneously according to the testing methods depicted in Fig. 1. The tests were performed continually from 30 to 50 s and was performed three times. Power spectrum was measured based on the sound generated from the airturbine head. Test mandrels (φ1.6×19 mm) having the dimensions specified in the International Standard (ISO8) and JIS7 documents were prepared. The test mandrels were magnetized using a permanent magnet and were inserted into the spring-type chuck of the air-turbine handpieces. The necessity of such magnetization depends on the type rotational speed sensor used. For measurement of rotational speed and total indicated run-out, as shown in Fig. 1, after fixing the air-turbine handpiece with a fastening device in order to maintain the test mandrel in a horizontal position, the rotational speed sensor was set below the test mandrel. The laser beam of the laser displacement meter was focused on the test mandrel 6 mm from the air-turbine head, and both tests were carried out after the air-turbine handpiece was activated. The laser displacement meter measures the distance between specific points on the test mandrel. And it was able to follow the movement of the test mandrel in direction of the laser beam axis, because the laser displacement meter shown in Fig. 1 had a maximum response frequency of 10 kHz. The “peak-to-peak” output mode was thus selected for the laser displacement meter. This output mode measures the difference between the maximum and minimum values over the measured distance, and thus it was possible to measure the total indicated run-out change of the test mandrel. The data shown in Fig. 1 were transferred to a personal computer via the sensor interface and were converted to rotational speed and a total indicated run-out‐time diagrams, respectively. For power spectral analysis, the air-turbine handpiece was hung in balance from the ceiling. The rotational axis of the test mandrel coincided with the direction of gravity. A microphone of the sound level meter to record sound levels was set at a position 0.45 m away from the air-turbine head. The power spectra of 96 M. NAGAI and K. TAKAKUDA J Med Dent Sci Fig. 1. Measuring methods of rotational speed and total indicated run-out. [Rotational speed meter: JADAS-701, Ogura jewel Ind., Laser sensor: LC-2440, Keyence Co., Laser displacement meter: LC-2400, Keyence Co. and Sensor interface: PCD-320A, Kyowa E. I. Co.] predicted frequencies described later and a major spectrum with comparatively large power value were observed scientifically in the screen of 10-kHz and 20kHz display modes of the FFT analyzer, and then the two images were recorded immediately. Furthermore, those images were compared with those at N=0 cycle. FFT analysis In the case of ball bearings in the dental air-turbine handpieces, the outer ring is fixed and the inner ring is rotated at a predetermined rotational speed. The rotational speed of the air-turbine handpiece, fr, the frequency of the ball’s revolution, fa, the frequency of the cage’s rotation, fb, and the frequency of the ball’s rotation, fc, were calculated from the following equations: (1) (2) (3) where, d is the ball diameter, D is the pitch circle of the ball and α is the contact angle of the ball. If the ball bearing has a dent or a scratch on the outer ring, and when number of balls is z, the frequency, fd, that these balls pass over the dent or scratch is given by the following equation: (4) In addition, if the ball bearing has a dent or a scratch on the inner ring, because the relative rotational speed of the inner ring for the frequency of the ball’s revolution is (fr ‐ fa), where, the number of balls is z, the frequency, fe, that the balls pass over the dimple or scratch is given by the following equation: (5) The frequencies, fa, fb, fc, fd and fe of each air-turbine handpiece in five manufacturers were predicted by using Eqs.(1) to (5), where, fr converted the R* values shown in Table 2 to Hz, d=1 mm, D=4.76 mm, α=0° and z=7. Here, the numerical values for the ball bearing were provided by the five manufacturers. Statistical analysis The significances of the rotational speed (RN) and the total indicated run-out (δN) at N=50, 100, 150, 200, 250 and 300 cycles were assessed by Dunnett’s multiple comparison test; free software:R9 (significance level: α =0.05) in comparison with RN and δN at N=0 cycles for control group, in order to examine influence of autoclave sterilization of an air-turbine handpiece. INFULUENCE OF DENTAL AUTOCLAVE ON AIR-TURBINE RESULTS Rotational speed and total indicated run-out In the time-series diagrams, there were almost no changes in rotational speed and total indicated run-out during the course of measurement in all air-turbine handpieces. Table 3 shows the means of three test values of the rotational speed (RN) measured at the predetermined number of sterilization treatment cycles (N=0, 50, 100, 150, 200, 250, 300 cycles) with all airturbine handpieces used in this study. The numerical values in parenthesis under each RN value indicate standard deviation. The value of RN for each air-turbine handpiece fluctuated as sterilization cycles increased. However, the standard deviations for all RN values were extremely small. This may be because establishing equal air pressure at cycle 0 was difficult due to fluctuations in the digital air pressure gauge reading. When the air pressure was fixed, rotational speed was stable, even when measurement was repeated 3 times. Consequently, small standard deviations were obtained. Table 4 shows the mean total indicated run-out val- 97 ues (δN) calculated by the same procedure as for rotational speed. The numerical values in parenthesis under each δN value indicate standard deviation. The values for δN were around 2 µm and were extremely small. Incidentally, as specified in ISO8, eccentricity should not exceed a total indicated run-out of 0.03 mm. However, the air-turbine handpieces used in this study gave significantly smaller values than the permissible values specified in ISO. However, the values fluctuated somewhat as sterilization cycles increased, regardless of the type of air-turbine handpiece. Furthermore, the standard deviations of all δN values were very large, as shown in Table 4. This demonstrates that the δN data, as measured 3 times for each test, varied widely. Such scattering may have been due to problems in the dynamic measurement system, including the resolution (0.2 µm) of the laser displacement meter, laser sensor installation, fixing of the airturbine handpiece and stiffness of the table that instruments were placed on. The mean values shown in Tables 3 and 4 were plotted as RN‐N and δN‐N diagrams in Fig. 2. The symbols for each product in Fig. 2 correspond to those shown in Table 1. On the whole, the rotational speed of Table 3. Mean values of rotational speed measured at predetermined number of sterilization treatment cycles. 98 M. NAGAI and K. TAKAKUDA J Med Dent Sci Table 4. Mean values of total indicated run-out measured at predetermined number of sterilization treatment cycles. each air-turbine handpiece seems not to be affected by increasing N values, except those of Nakanishi. The details are determined by Dunnett’s multiple comparison test to be described next. However, both the rotational speeds of N1 and N2 showed a tendency to gradually decrease by increasing N values. The results assessed by the Dunnett’s multiple comparison test were shown in Tables 5 and 6. In case of rotational speed, as shown in Table 5, some of results of O1-1 and Y2 decreased significantly. In O1-1 which is similar to O1-2, although O1-2 showed no significant difference in all cycles of sterilization, the significant difference was observed at N=150, 200 and 300 cycles. Similarly, Y1 showed no significant difference at all cycles of sterilization. However, the significant difference was observed at N=100, 200, 250 and 300 cycles in Y2. Although the Y1 and the Y2 were of different types, only the ball bearing part was same. Meanwhile, in case of RN values in N1 and N2, the significant differences were observed at all cycles of sterilization. As shown in Table 6, in case of Y2 and K1-1, the significant differences were observed in δN values at N=200 and 250 cycles, and in δN values at N=150, 200, Fig. 2. RN and δN - N diagrams in each air-turbine handpiece, where the symbols for each product correspond to those shown in Table 1. INFULUENCE OF DENTAL AUTOCLAVE ON AIR-TURBINE 99 Table 5. The results of Dunnett’s multiple comparison test (α=0.05) of RN. Table 6. The results of Dunnett’s multiple comparison test (α=0.05) of δN. Fig. 3. Representative power spectra of the air-turbine handpiece, which were measured after predetermined number of sterilization treatment cycles, where the respective arrows indicate fr=6.2 kHz, fa= fb=2.4 kHz, fc=14.0 kHz and fd=17.1 kHz. 250 and 300 cycles, respectively. However, as shown in Table 4, the δN values in Y2 or K1-1 were smaller than those N=0 cycle. N1 and N2 showed no significant differences in δN values, although both the rotational speeds showed a tendency to gradually decrease and differed significantly. This is noticeable result and suggests that it is not a problem of ball bearing. FFT analysis Figure 3 shows representative power spectra of airturbine handpiece. The frequency domain of noise in the measurement room was equal to or less than 2 kHz. The power spectral diagrams were recorded respectively when the air-turbine handpiece: M1-1 was tested at N=0, 150 and 300 cycles. The arrows in Fig. 3 indicate each predicted frequency calculated by Eqs.(1) to (4): fr=6.2 kHz, fa= fb=2.4 kHz, fc=14.0 kHz and fd=17.1 kHz. As shown in Fig. 3, M1-1 after sterilized did not exhibit characteristic difference in the power spectra in comparison with those at N=0 cycle, because similar power spectra were observed at different stages of the sterilization treatment and none of the power spectra varied in both sides on frequencyaxis when the air-turbine handpiece after sterilized was tested. Additionally, an abnormal power spectrum was not observed besides the predicted frequencies. Such similarities were observed for all air-turbine handpieces tested. DISCUSSION Rotational speed and total indicated run-out Monaghan and others5 discussed the influence of 100 M. NAGAI and K. TAKAKUDA autoclave on ball bearing by means of the bearing resistance. A wear condition of the ball bearing might be investigated from such bearing resistance. However, their method is not commonly used. The evaluation of the ball bearing is considered of value in the case of air-turbine handpiece that the dentist used in dental practice. In case of this study, it was selected to evaluate by means of the total indicated run-out and FFT analysis, because we employed the air-turbine handpieces without tooth cutting in this test. The significance regarding rotational speed and total indicated run-out is discussed as follows. In O1-1, the RN values at N=150, 200 and 300 cycles, as shown in Table 3, decreased 1 to 3 percent in comparison with that at N=0 cycle, although the significant differences were observed at those N values as shown in Table 5. However, O1-1 showed no significant differences in total indicated run-out as shown in Table 6. Similar to O1-1, the RN values in Y2 decreased 1 to 2 percent in comparison with that at N=0 cycle. And the Y2 showed the significant differences in the δN values. However, the δN values, as shown in Table 4, were smaller than that at N=0 cycle. Incidentally, the δN values of K1-1 that the significant differences were shown in total indicated run-out were small than that at N=0 cycle. If the ball bearing was damaged by autoclave, the total indicated run-out of test mandrel would increase suddenly even if the damage of the ball bearing was a little. Therefore, such the result may have depended on the quality of the equipment. In regard to the measurement of a total indicated run-out, the ball bearing rotates at super-high-speed and sufficient measurement technology has not been provided at present. The development of new measuring equipment is expected in future. In case of N1 and N2, the significant differences were observed in all RN values, furthermore, both the RN values gradually decreased as shown in Fig. 2. Especially, both the RN values decreased instantly to 3.5 and 2.2 percent at N=50 cycles, respectively. A hypothesis that the ball bearing was affected by autoclave at the smaller N was rejected, because a cage of the ball bearing is made by heat-resistant resin. Accordingly, it was speculated that such the phenomenon depended on other problems aside from ball bearing itself. On the contrary, Monaghan and others5 reported that the rotational speed of several air-turbine handpieces increased at early use stage. They suggested that initial conditions or variation of quality of ball bearing might become a problem as the reason for this. However, the dentist will try enough lubrication and J Med Dent Sci free-running of the air-turbine handpiece before used in clinical practice for the first time, and then will sterilize the air-turbine handpieces. Accordingly, their consideration is hard to be understood. Hence, it is considered that such phenomena would depend on account of others. And it is hard to make a clear consideration for the early decrease in this paper. The aim of this paper is to evaluate the affect of an autoclave on the rotational performance of ball bearing. So, the ratio of RN at N=300 cycles to that at N=50 cycles in N1 and N2 was calculated, respectively. As the result, the decreases were 5 percent together. From such the decrease rates, it may be evaluated that the durability of the air-turbine handpieces were deteriorated inconsiderably. However, it was suggested that the ball bearing did not become dynamically the serious situation at N=300 cycles, because N1 and N2 showed no significant differences in the δN values in all cycles of sterilization. If the ball bearing was influenced by autoclave, the δN values must have shown a tendency to increase. In the actual clinical case, small decrease of rotational speed has no influence on the tooth cutting technique for a dentist. FFT analysis Among the components of the ball bearing, the cage is made of heat-resistant resin. The ball bearing may thus be gradually influenced by the heat of sterilization, even if heat-resistant resins are used. This may result in reduced rotational performance. However, from the former results of total indicated run-out, it was guessed that the ball bearing would not be influenced by autoclave. Furthermore, as another test for evaluating the ball bearing, a qualitative diagnosis of ball bearing was performed while observing scientifically major power spectra displaying on the screen of FFT analyzer, on the basis of each predicted frequency of the components in the ball bearing. Each arrow in Fig. 3 indicates the frequencies for fr, fa (=fb), fc and fd as a reference. Thus, each predicted frequency, as shown for each result after a predetermined number of sterilization treatment cycles (Fig. 3), agreed with almost each peak value from the spectra. However, the spectra for fc and fd were masked by noise because the amplitudes of both frequencies were extremely small. This suggests that an abnormal sound was generated in the frequencies of fc and fd because the ball bearing showed no signs of damage after N=300 cycles. Consequently, this suggests that the ball bearing did not receive any type of damage by autoclave. Comprehensively, it was demonstrated that INFULUENCE OF DENTAL AUTOCLAVE ON AIR-TURBINE all air-turbine handpieces rotated normally after repeated sterilization and were not affected by the autoclave. From the above-mentioned results, we may speculate that the rotational performance of air-turbine handpieces would not be influenced by sterilization treatment until N=900 cycles, because the sterilization period was 3 times longer in this study than that recommended by the JIS. One air-turbine handpiece is sterilized 2-3 times per day, and five days per week at a typical Japanese dental clinic. This means that the airturbine handpieces were subjected to the equivalent of more than two years of sterilization treatment in the present study. 101 frequency of the ball’s rotation were comparatively investigated at N=0, 150 and 300 cycles, and the influence of autoclave was determined. No abnormalities in the ball bearings were recognized. ACKNOWLEDGEMENTS On the occasion of implementation of this research, authors wish to thank Morita Mfg. Corp., Nakanishi Inc., Osada Electric Co., Yoshida Dental Mfg. Co., and Shirokusu Dental Supply Works for supplying those respective air-turbine handpieces shown in Table 1, and Mr. Masayuki Kobayashi for his assistance in sterilization treatment cyclic test for air-turbine handpiece. CONCLUSION REFERENCES The influence of number of autoclave treatment cycles (N) on rotational speed and total indicated runout of commercially available air-turbine handpieces from five manufacturers was investigated at N=0, 50, 100, 150, 200, 250 and 300 cycles, and the significance in the test results was assessed by Dunnett’s multiple comparison test. Some air-turbine handpieces showed the significant differences in rotational speed at N=300 cycles, however, the decreases of the rotational speeds were only 1 to 3.5 percent. Therefore, it is not concluded that those ball bearings received a considerable damage by autoclave. Some air-turbine handpieces showed the significant differences in total indicated run-out, however, the respective values were smaller than that at N=0 cycle. Therefore, it was suggested to be influenced variation of data. Accordingly, it can be considered that the ball bearing in the air-turbine handpieces is not affected significantly by autoclave. To further evaluate rotational performance, this study focused on the rotational vibration of the ball bearing components of the air-turbine, as measured by FFT analysis; the power spectra of frequency of the ball’s revolution, frequency of the cage’s rotation and 1 2 3 4 5 6 7 8 9 Kohn WG, Collins AS, Cleveland JL et al. Guideline for Infection Control in Dental Health-Care Setting—2003. Centers for Disease Control and Prevention. Recommendations and Reports 2003/52(RR17);1-61. Yamaga M, Kiryu T, Ohfuchi Y et al. Internal contamination of dental air turbine system—The protective effect of a newly design handpiece on bacterial invasion (in Japanese, English abstract). J J Conser Dent 1995;38(2):472-478. Clinical Research Associates Newsletter (in Japanese). Advanced high-speed handpieces—Ten years after the sterilization procedure was enabled. 2000;4(9):1-2. Clinical Research Associates Newsletter, Additional study. High speed handpieces: A. Use-life sterilized vs. non-sterilized HPs. The October 2000 Newsletter. Monaghan DM, Wilson NHF, Darvell BW. The performance of air-turbine handpieces in general dental practice. Operative Dentistry 2005;30(1):16-25. JIS T 7324. High-pressure steam sterilizers for medical use (small size) (in Japanese), Japanese Standards Association 2005. JIS T 5906. Dental handpieces—Part 1: High-speed air turbine handpieces (in Japanese), Japanese Standards Association 2001. ISO 7785-1. Dental handpieces—Part 1: High-speed air turbine handpieces. the International Organization for Standardization 1997. R, Copyright 2005, The R foundation for statistical computing Version 2.1.0 (2005-04-18), ISBN 3-900051-07-0, Multiple comparison by Dunnett’s method.
© Copyright 2026 Paperzz