3M™ True Definition Scanner Quick Start Guide for Incognito™ Appliance System Steps to Complete a Case Prescription 1. Scan the lower arch. 2. Scan the upper arch. 3. Scan both right and left quadrants (recommended). 5. Sign and submit. 4. Complete prescription. Scan Path 1. Isolation and Powder Isolate only the sextant you are starting with. Dry and powder this sextant. 2. Posterior Scan Begin parallel to the occlusal surface at the first premolar. Scan the occlusal table and rotate on the first molar towards the lingual. Rotate over the occlusal to scan the buccal. Finish on the occlusal. 1 2 3 3. Anterior Scan Start on the occlusal and scan lingual surfaces to the next premolar. 4. Anterior Facial Rotate over the occlusal of the premolar to the buccal and scan to the midline in a horizontal orientation. Pause the scan and return to the occlusal of the original premolar. 4 Rotate to the buccal and scan to the midline in a horizontal orientation to meet your last scan. Finish on incisal. Complete full arch by starting at the first premolar and repeat posterior scan. 5. Bite Scan With patient in desired occlusion, scan a box of three upper and lower teeth. At least two bite scans are needed for accuracy. 5 Scan Criteria Acceptable Incognito™ Digital Impression Criteria/Requirements • Capture the entire arch including the terminal molars regardless of Incognito™ Complete or Incognito™ Lite case • All incisal edges and marginal ridges should be clear and sharp • Complete occlusal surfaces present (no holes) • Clean and well defined interproximal embrasures and natural IP spaces • Clear crisp gingival margins with 3–5mm of gingiva • Complete capture of the palatal rugae • At least 3mm of soft tissue capture in the lingual palatal regions • Right and left posterior bite registrations with six to eight teeth Capture the entire arch – full dentitions for Incognito. Capture a minimum of 3-5mm of gingiva lingually. Capture a minimum of 3-5mm of gingiva labially. Obtain Patient’s complete right bite registration scan. Obtain Patient’s complete left bite registration scan. Capture complete occlusal surfaces — no holes. Lab Prescription Form ouch the prescription screen to T access the appropriate selection box. 1.Select the “Prescription” tab to complete the information sent to Unitek TMP Lab. 2.From the “Date Requested” field, choose the current date. Please note that this date does not affect the delivery promise dates for products produced by 3M Unitek from the scan. 1 5 2 6 7 3 8 4 3.Select Lab: “Unitek TMP Lab”. 4.Choose “Dentist”. 5.Select “Restoration”. 6.Choose “Models and Appliances”. 9 7.Choose “Select Type”. 8.Select “Unitek™ Treatment Management Portal”. 10 9.Enter “Notes” if desired. 10. V erify Rx for accuracy, then “Sign” (enter PIN). To Upload Digital Impressions • Please allow up to 90 minutes for digital impression files to be available at the Unitek Lab. •Ensure that the patient name used to create the scan matches the name used on your Treatment Management Portal | TMP. •If you cannot find the patient by name, try using the 3M Case Number (found in 3M Connection Center) instead of the patient’s name to search for the patient’s scan. If you cannot find your patient’s scan file, contact Unitek TMP Customer Support at (800) 265-1943. •F ill out the order form within Unitek TMP with your treatment plan. •A custom setup will be created based on your prescription and will be posted back to your account in the Treatment Management Portal | TMP for your review/approval. •After approval of the setup, the Incognito™ Appliance System indirect bonding trays are custom manufactured and shipped to the doctor’s office. Reminder: Click “Digital Scan” when ordering Incognito in the Unitek TMP Order Wizard. 3M and Incognito™ Appliance System Customer Support 3M™ True Definition Scanner 3M™ Unitek™ Incognito Appliance System Incognito Appliance System or™Appliances or Applicances Unitek Treatment Management ™ UnitekPortal Treatment Management or Unitek TMP Portal or Unitek TMP Hardware Software Wireless Connection Scan Data/Case Upload Scan Protocol 3M True Definition Scanner, 3M Connection Center and 3M Software Questions 1-800-634-2249, Option 3 Incognito System Customer Support 1-800-401-3001 Unitek TMP Customer Support 1-800-265-1943 Customer Care Center: 1-800-634-2249 www.3M.com/TrueDef 3M Oral Care 2510 Conway Avenue St. Paul, MN 55144-1000 USA Phone1-800-634-2249 Webwww.3M.com/TrueDef 3M Canada Post Office Box 5757 London, Ontario N6A 4T1 Canada 1-888-363-3685 3M Unitek Orthodontic Products 2724 South Peck Road Monrovia, CA 91016 USA www.3MUnitek.com 3M, Incognito and Unitek are trademarks of 3M. Used under license in Canada. Please recycle. Printed in U.S.A. © 3M 2015. All rights reserved. 70-2013-0633-2 (Rev. 8/2015)
© Copyright 2026 Paperzz