Free, Total and Double SO2 in Wine DOC316.52.93088 Endpoint potentiometric titration with imposed current Scope and application: Wines 1. Introduction This working procedure refers to the titration of sulfur dioxide in wines. A double platinum electrode is used: an alternating current is imposed and the resulting potential is measured. 2. Principle The protocol rests on the redox reaction between sulfur dioxide and iodine: SO2 + I2 + 2 H2O H2SO4 + 2 HI According to the Ripper method, free SO2 is directly titrated in acidic media (sulfuric acid is added to the sample). Total SO2 is the sum of free SO2 and combined SO2. The latter is released by hydrolysis in alkaline media (sodium hydroxide is added to the sample prior to sulfuric acid). Then, iodine solution is added until the measured potential drops abruptly, which corresponds to the equivalence. The SO2 concentration is expressed in mg/L. 3. Electrode and reagents Electrodes: Intellical double platinum electrode with integrated temperature sensor, MTC695 Titrant: Iodine I2 0.0282 eq/L (part number 2333353) Acid reagent: Sulfuric acid H2SO4 10 eq/L (part number 93153); install this reagent on PUMP 1 Alkali reagent: Sodium hydroxide NaOH 5 eq/L (part number 245053); install this reagent on PUMP 2 Solid KI: With dispenser (part number 2834100) Standard for titrant calibration: Sodium thiosulfate Na2S2O3 0.00564 eq/L (part number 2408853) Deionized water 4. Ranges and settings 4.1. Default parameters The working procedure is described using the following parameters: • • V sample = 20 mL Syringe volume = 10 mL 4.2. Working range For most samples, 1 syringe (10 mL) of titrant should be sufficient to reach the equivalent point. It provides the following range: Veq (mL) SO2 (mg/L) 0.1 4.5 © Hach Company/Hach Lange GmbH, 2014-2015. All rights reserved. 9.5 430 Page 1 of 8 04/2015, Edition 2 For low SO2 concentrations below 45 mg/L, (equivalent volumes under 1 mL) the titration typically lasts less than 1 minute. For SO2 concentrations up to 110 mg/L (equivalent volume around 2.5 mL), the titration typically lasts then than 2 minutes. For SO2 concentrations between 110 and 430 mg/L, it is recommended to add a predose in order to keep titration duration to around 2-3 minutes. Orders of magnitude are given in the table below: Expected SO2 concentration (mg/L) Expected equivalent volume (mL) 3.3 4.4 5.5 6.6 7.7 8.9 150 200 250 300 350 400 Recommended predose (mL) 0.8 1.9 3.0 4.1 5.2 6.4 4.3. Settings Name Sample Name Amount Amount min Amount max Titrant Name Titrant concentration Syringe Probe Recommended probe Leveling Active Time Default parameter Unit Wine 20 0 22 [mL] [mL] [mL] I2 0.0282 Syringe 1 [eq/L] MTC695 No 30 [s] 4.3.1. Free SO2 Name Step 1: Acid addition Duration Stirrer speed Step 2: Reaction Duration Stirrer speed Step 3: Titration Stirring speed Measured parameter Predose Max volume stop point Stop on last EQP Delay Addition mode Increment size EP Ordinates Result 1 name R1 resolution R1 min R1 max R1 QC min R1 QC max Default parameter Unit 1 20 [s] [%] 15 0 [s] [%] 10 [%] [mV] [mL] [mL] 0 10 True 0 Monotonic 0.08 160 Free SO2 Integer 5 430 5 430 © Hach Company/Hach Lange GmbH, 2014-2015. All rights reserved. [s] [mL] [mV] [mg/L] [mg/L] [mg/L] [mg/L] Page 2 of 8 04/2015, Edition 2 4.3.2. Total SO2 Name Step 1: Alkali addition Duration Stirrer speed Step 2: Reaction Duration Stirrer speed Step 3: Acid addition Duration Stirrer speed Step 4: Reaction Duration Stirrer speed Step 5: Titration Stirring speed Measured parameter Predose Max volume stop point Stop on last EQP Delay Addition mode Increment size EP Ordinates Result 1 name R1 resolution R1 min R1 max R1 QC min R1 QC max Default parameter Unit 3 20 [s] [%] 5 0 [min] [%] 3 20 [s] [%] 15 0 [s] [%] 10 [%] [mV] [mL] [mL] 0 10 True 0 Monotonic 0.08 140 Total SO2 Integer 5 430 5 430 [s] [mL] [mV] [mg/L] [mg/L] [mg/L] [mg/L] 4.3.3. Double SO2 The measurement of double SO2 consists of a first measurement of free SO2 followed directly by a measurement of combined SO2. The sum of the 2 results gives the total SO2. Name Step 1: Acid addition Duration Stirrer speed Step 2: Reaction Duration Stirrer speed Step 3 : Free SO2 titration Stirring speed Measured parameter Predose Max volume stop point Stop on last EQP Delay Addition mode Increment size EP Ordinates Result 1 name R1 resolution R1 min R1 max R1 QC min Default parameter Unit 1 20 [s] [%] 15 0 [s] [%] 10 [%] [mV] [mL] [mL] 0 10 True 0 Monotonic 0.08 160 Free SO2 Integer 5 430 5 © Hach Company/Hach Lange GmbH, 2014-2015. All rights reserved. [s] [mL] [mV] [mg/L] [mg/L] [mg/L] Page 3 of 8 04/2015, Edition 2 R1 QC max Result 2 name Hide Step 4: Alkali addition Duration Stirrer speed Step 5: Reaction Duration Stirrer speed Step 6: Acid addition Duration Stirrer speed Step 7: Reaction Duration Stirrer speed Step 8: Combined SO2 titration Stirring speed Measured parameter Predose Max volume stop point Stop on last EQP Delay Addition mode Increment size EP Ordinates Result 3 name Hide Equation result name Equation result resolution Equation result min Equation result max Equation result QC min Equation result QC max 430 Veq1 True [mg/L] 4 20 [s] [%] 5 0 [min] [%] 3 20 [s] [%] 15 0 [s] [%] 10 [%] [mV] [mL] [mL] 0 10 True 0 Monotonic 0.08 140 Veq2 True Combined SO2 Integer 5 430 5 430 [s] [mL] [mV] [mg/L] [mg/L] [mg/L] [mg/L] 4.4. Modification of the settings The parameters are defined in order to have the best compromise between accuracy and titration time. For higher concentration with a high titrant volume, titration time can be reduced with an addition of titrant (predose) at the beginning of the titration. Enter the predose volume (in mL) and the stirring time after the addition in the application edit window. 5. Titration procedure 5.1. Leveling To use this method, an external pump is required. All elements (probes, tubes from the titrator and the tube from the external pump) have to be well installed on the probe holder. The beaker has to contain a level of sample higher than the position of the tube of the external pump. When the beaker is attached to the probe holder, this method allows the system to automatically remove the excess sample by a defined pump working time, and always keep the same sample volume before launching the analysis. In order to define this volume, autoleveling calibration sequence has to be previously executed (see section 8.2 Autoleveling calibration). When this option is active, the working time of the external pump must be set (default 30 s). The minimum working time must allow the pump to be removing air during the last few seconds of the external pump activation. Note: Do not forget to re-edit the sample amount with the expected value when deactivating the leveling method. © Hach Company/Hach Lange GmbH, 2014-2015. All rights reserved. Page 4 of 8 04/2015, Edition 2 5.2. Titration Rinse the probe with deionized water. If leveling is disabled, use a pipette to collect precisely 20 mL of sample. Pour the sample into the 50 mL polypropylene beaker, put in a magnetic stir bar, dip the probe and the delivery tip in the solution and then start the application. At the end of the titration, a first window displays the result. A second window displays the titration curve and the equivalent point coordinates. After the titration, there are two possibilities: • • 6. Replicate the sample. This is used to study the repeatability by analyzing several samples successively. At the end of each titration, a window displays the average value, the standard deviation (SD) and the relative standard deviation (RSD in %). Analyze a new sample. Another titration can be started but no Standard Deviation and RSD value will be made. Results 6.1. Result calculation The generic equation is: For Free SO2: For Total SO2: For Double SO2: SO2 (mg⁄L) = ne- titrant Ctitrant (eq⁄L) × Vtitrant (mL) × × MSO2 (g/mol) × 1000 ne- titrant × Vsample (mL) ne- SO2 Free SO2 (mg⁄L) = Total SO2 (mg⁄L) = Free SO2 (mg⁄L) = Combined SO2 (mg⁄L) = 2 0.0282 (eq/L) × Veq (mL) × × 64.1 (g/mol) × 1000 2 × 20 (mL) 2 2 0.0282 (eq/L) × Veq (mL) × × 64.1 (g/mol) × 1000 2 × 20 (mL) 2 2 0.0282 (eq/L) × Veq1 (mL) × × 64.1 (g/mol) × 1000 2 × 20 (mL) 2 2 0.0282 (eq/L) × [Veq2 + V'] (mL) × × 64.1 (g/mol) × 1000 × FX 2 × 20 (mL) 2 Where V’ is the volume of excess titrant added after Veq1 during Free SO2 titration. For Combined SO2, an equation is used since the result cannot be calculated directly from the equivalent volume. Therefore, the factor FX (1 by default) has to be changed if the titrant unit is changed, for example: Titrant titer unit Factor FX eq/L 1 meq/L 0.001 6.2. Experimental results These results are indicative and have been obtained for given samples, for 8 to 10 successive determinations. 6.2.1 Free SO2 For 10 determinations on red wine: Measurement Mean concentration Standard deviation Relative standard deviation Mean titration duration © Hach Company/Hach Lange GmbH, 2014-2015. All rights reserved. Parameter 25 0.4 1.7 45 Unit [mg/L] [mg/L] [%] [s] Page 5 of 8 04/2015, Edition 2 For 10 determinations on white wine: Measurement Mean concentration Standard deviation Relative standard deviation Mean titration duration Parameter 4 0.1 1.8 19 Unit [mg/L] [mg/L] [%] [s] Parameter 80 2.8 3.5 108 Unit [mg/L] [mg/L] [%] [s] Parameter 72 1.2 1.7 99 Unit [mg/L] [mg/L] [%] [s] Parameter Unit 6.2.2. Total SO2 For 10 determinations on red wine: Measurement Mean concentration Standard deviation Relative standard deviation Mean titration duration For 10 determinations on white wine: Measurement Mean concentration Standard deviation Relative standard deviation Mean titration duration 6.2.3. Double SO2 For 8 determinations on red wine: Measurement Free SO2 Mean concentration Standard deviation Relative standard deviation Mean titration duration Combined SO2 Mean concentration Standard deviation Relative standard deviation Mean titration duration 9 0.1 0.9 14 [mg/L] [mg/L] [%] [s] 48 0.9 1.8 48 [mg/L] [mg/L] [%] [s] 6.3. Example of a titration curve 7. Recommendations Always rinse the pH probe and the delivery tip between measurements. © Hach Company/Hach Lange GmbH, 2014-2015. All rights reserved. Page 6 of 8 04/2015, Edition 2 8. Appendices 8.1. Titrant calibration The iodine solution can be calibrated: its exact concentration can be determined from a titration using a sodium thiosulfate standard solution. Use a pipette to collect precisely 20 mL of 0.00564 N sodium thiosulfate solution. Pour the standard in the 50 mL propylene beaker and add a stir bar. Dip the probe and the delivery tip into the solution and launch the titrant calibration sequence. At the end of the titrant calibration, titer (eq/L) is displayed and the user can reject, replicate, or save the result. Default settings for titrant calibration Name Titrant Name Titrant concentration Syringe Standard Name Amount EP titration Stirring speed Measured parameter Predose Max volume stop point Stop on last EQP Delay Increment size EP Ordinates Result name Result resolution Result min Result max Default parameter I2 0.0282 Syringe 1 Na2S2O3 20 15 2 8 True 0 0.04 400 Titer 5 decimals 0.02538 0.03102 Unit [eq/L] [mL] [%] [mV] [mL] [mL] [s] [mL] [mV] [eq/L] [eq/L] 8.2. Autoleveling calibration The aim of this method is to calibrate the volume of sample by leveling. The result of this calibration will be used as sample volume for the following titrations. This option is ONLY available from the calibration menu if Method Leveling is set to Active (Yes). Refer to the documentation delivered with the external pump for a correct installation, paying particular attention to the suction tube from the pump. Use a 0.00564 eq/L sodium thiosulfate standard solution and pour a sufficient amount of the solution into a beaker allowing the external pump tube to be immersed in the liquid. In the calibration menu select Autoleveling calibration and then Free SO2, Total SO2 or Double SO2. The result in mL is compared to minimum and maximum amounts defined for the sample volume. The calculation used is: Vsample = Vtitrant (mL) × ne- titrant × Ctitrant (eq/L) ne- sodium thiosulfate × Csodium thiosulfate solution (eq/L) = Vtitrant (mL) × 2 × 0.0282 (eq/L) 2 × 0.00564 (eq/L) © Hach Company/Hach Lange GmbH, 2014-2015. All rights reserved. Page 7 of 8 04/2015, Edition 2 Autoleveling calibration uses the same settings as for a titrant calibration (see section 8.1 Titrant calibration). Default settings for autoleveling calibration Name Sample Amount min Amount max Autoleveling calibration Solution name Concentration Resolution Default parameter Unit 0 22 [mL] [mL] Na2S2O3 0.00564 3 decimals © Hach Company/Hach Lange GmbH, 2014-2015. All rights reserved. [eq/L] Page 8 of 8 04/2015, Edition 2
© Copyright 2025 Paperzz