RAPID COMMUNICATIONS IN MASS SPECTROMETRY Rapid Commun. Mass Spectrom. 13, 2260–2267 (1999) Analysis of Poly(bisphenol A carbonate) by Size Exclusion Chromatography/Matrix-assisted Laser Desorption/Ionization. 1. End Group and Molar Mass Determination Concetto Puglisi1, Filippo Samperi1, Sabrina Carroccio2 and Giorgio Montaudo2* 1 Istituto per la Chimica e la Tecnologia dei Materiali Polimerici, Consiglio Nazionale delle Ricerche, Viale A. Doria, 6 - 95125 Catania, Italy 2 Dipartimento di Scienze Chimiche, Università di Catania, Viale A. Doria, 6 - 95125 Catania, Italy The determination of molar mass (MM) data for polydisperse polymers by SEC/MALDI involves the fractionation of samples through analytical size exclusion chromatography (SEC). Selected SEC fractions are then analyzed by matrix-assisted laser desorption/ionization time-of-flight (MALDI-TOF) and the mass spectra of these nearly monodisperse samples allow the determination of the average molar masses. The SEC/MALDI procedure has now been applied to two polycarbonate samples, PC1 and PC2. The results show that the MALDI spectra of the SEC fractions allow not only the detection of linear and cyclic oligomers contained in these samples, but also the simultaneous determination of their average molar masses. Two slightly differing SEC calibration plots were obtained, due to the smaller hydrodynamic volume of the polycarbonate cyclic chains with respect to the linear ones. In agreement with theory, the ratio (Mcycle/Mlinear)Ve at a fixed elution volume was found to be 1.22, independent of the molar mass values. Copyright # 1999 John Wiley & Sons, Ltd. Received 17 August 1999; Revised 24 September 1999; Accepted 26 September 1999 Matrix-assisted laser desorption/ionization time-of-flight Mass Spectrometry (MALDI-TOFMS)1,2 allows the desorption and ionization of very large molecules even if present in complex mixtures, and is widely used with increasing success for the characterization of synthetic polymers, particularly for end group analysis. Although MALDI spectra of synthetic polymers can reach masses up to 106 Da and beyond, there is a much lower limit to the mass spectral resolution of individual oligomers, and this depends both on the resolution of the MALDI instruments and on the polymer repeat unit. The mass limit at which peaks corresponding to individual oligomers are no longer mass-resolved reaches a value of about 15 kDa with conventional MALDI-TOF (continuous extraction), whereas with the development of delayed extraction (time-lag focusing) it can be extended to about 50 kDa. For the analysis of polymers it has been found that molar mass (MM) estimates provided by MALDI agree with the values obtained by conventional techniques only in the case of samples with narrow molar mass distributions (MMD), whereas with polydisperse polymers MALDI fails to yield reliable MM values.2 This implies that the relative intensities of the MALDI signals as a function of their m/z values is far from that expected from the actual MMD of the polymer sample. Lighter molecules (which outnumber the larger ones in a most probable distribution) tend to saturate *Correspondence to: G. Montaudo, Dipartimento di Scienze Chimiche Università di Catania, Viale A. Doria, 6 - 95125 Catania, Italy. Contract/grant sponsor: Italian Ministry for University and for Scientific and Technological Research. Contract/grant sponsor: Progetto Finalizzato Materiali per Tecnologie Avanzate II. Contract/grant sponsor: National Council of Research, Italy. CCC 0951–4198/99/222260–08 $17.50 the MALDI detector, suppressing the detection of the laterarriving larger molecules. Furthermore, the small molecules reach the detector with a higher kinetic energy and are thus preferentially revealed in the MALDI process. Even when the detectors currently employed in MALDI instruments are equipped with a post-accelerator, the detection of high masses may still be partially discriminated against.3 Synthetic polymers may show a wide range of molar mass distributions according to the synthetic method used in their preparation, and therefore the direct application of the MALDI technique for the MM measurements appears limited to narrow distributions, although several attempts are currently being performed to extend it to polymers with wider distributions.3–5 A widely accepted method for MM determination in polydisperse polymer samples by MALDI is6–12 the analytical size exclusion chromatography (SEC) fractionation of polydisperse samples. We have proposed this approach earlier,6,7 and the results appear encouraging. Furthering our studies on the off-line SEC/MALDI method, we have now applied it to the characterization of end groups and molar masses in poly(bisphenol A carbonate) (PC). The characterization of poly(bisphenol A carbonate) by mass spectrometry has received continued attention, due to the fact this polymer is an important engineering thermoplastic material.8,13–17 End group analysis and molar mass characterization are among the major concerns in the structural characterization of PC. It is well known that the production of this polymer is accompanied by the formation of sizeable amounts of cyclic oligomers, as is the case for other polymers produced by condensation and ring-opening reactions, and it is of great interest to determine their size and relative abundance.18 The presence of a mixture of linear and cyclic chains in low mass polycarbonates causes Copyright # 1999 John Wiley & Sons, Ltd. ANALYSIS OF POLYCARBONATES BY SEC/MALDI 1 2261 Figure 1. MALDI-TOF spectrum of unfractionated PC1 sample. difficulties in establishing an appropriate SEC calibration curve for such samples.8 It is well known that, for a given polymer, there is a small but well-defined difference between the hydrodynamic volumes of linear and cyclic oligomers of the same molar mass,19 and this difference is reflected in the elution volumes (Ve) of the linear and cyclic species of the same molar mass.19 Thus, it has been shown by SEC-light scattering experiments,19 that cyclic oligomers of polydimethyl siloxane (PDMS) are eluted at slightly higher volumes with respect to linear oligomers of the same MM, owing to the smaller hydrodynamic volume. The ratio (Mcycle/Mlinear)Ve at a fixed elution volume (Ve) was found to be 1.24, independent of MM values. This result was recently confirmed by performing the SEC/MALDI calibration of two PDMS samples (linear and cyclic, respectively), and the ratio (Mcycle/Mlinear)Ve was found to be 1.22.20 In the case of PC a small but well-defined difference was found in an early study21 on linear versus cyclic PC samples, by measurements of molar masses and intrisic viscosities of Figure 2. MALDI-TOF spectrum of unfractionated PC2 sample. Copyright # 1999 John Wiley & Sons, Ltd. Rapid Commun. Mass Spectrom. 13, 2260–2267 (1999) 2262 ANALYSIS OF POLYCARBONATES BY SEC/MALDI 1 Table 1. Structural assignments of the peaks appearing in the inset sections of the MALDI spectra of polydisperse samples PC1 and PC2 reported in Figs 1 and 2 M n Na A 11 12 13 3034 3288 3542 A' 13 14 15 3654 3908 4162 B 12 13 14 3168 3422 3676 B' 14 15 16 3732 3986 4240 11 12 13 14 15 16 3048 3302 3556 3810 4064 4318 12 13 14 15 16 17 3074 3328 3582 3836 4090 4344 Mass series Oligomer structure C D a series of monodisperse BPA-polycarbonate samples. Two Mark-Houwink relationships could be discerned for the linear and cyclic PC samples, but no quantitative estimates of the ratio (Mcycle/Mlinear) were made.21 In a recent investigation,8 different slopes for the SEC/ MALDI calibration lines of linear and cyclic PC chains were reported. This result implies that the ratio (Mcycle/ Mlinear)Ve changes with the MM value, and that the theoretical19 constant value of 1.24 is not valid for all the polymers. However, the conditions used in that SEC experiment8 are likely to promote sizeable self-association of the PC chains,22 and this might affect the SEC/MALDI calibration line. We have applied the SEC/MALDI procedure to two polycarbonate samples, PC1 and PC2, which contain both linear and cyclic oligomers. The results show that the MALDI spectra of the SEC fractions allow the individual detection of linear and cyclic oligomers contained in these samples, and also the simultaneous determination of the average molar masses of the linear and cyclic oligomers. As expected, two parallel SEC calibration lines were obtained for linear and cyclic PC chains, and the ratio (Mcycle/Mlinear) was found to be constant (about 1.22). In earlier works,19–21 in order to investigate the molecular Rapid Commun. Mass Spectrom. 13, 2260–2267 (1999) K 3670 3924 4178 3748 4002 4256 3090 3344 3598 3852 406 4350 dimensions of a set of linear and cyclic polymers, samples of pure linear and pure cyclic polymers were analyzed. This effort was needed in order to check the value of the (Mcycle/ Mlinear) ratio over an extended range of MM. Of course, pure linear and pure cyclic samples are not always available for every polymer, whereas the most common case is that of an essentially linear polymer sample which contains cyclic oligomers. The advantage of MALDI mass spectra is that they yield separate and simultaneous information on the molar masses of both cyclic and linear oligomers present in each SEC fraction. EXPERIMENTAL Materials Poly(bisphenol A carbonate) sample PC1 was supplied by Aldrich Chemical Co.(Italy), whereas PC2 was a L220 GE polycarbonate sample. Both samples were dried at 70 °C, under vacuum, for one week before use. 2-(4-Hydroxyphenylazo)benzoic acid (HABA) and trans-3-indoleacrylic acid (IAA) were supplied by Aldrich Chemical Co. (Italy) and used as supplied. Copyright # 1999 John Wiley & Sons, Ltd. ANALYSIS OF POLYCARBONATES BY SEC/MALDI 1 2263 Figure 3. SEC traces of samples PC1 and PC2 together with the calibration line obtained by the SEC/MALDI method. SEC analysis and fractionation Size exclusion chromatography (SEC) analyses were performed using a Waters 600 A apparatus, equipped with five Ultrastyragel columns (7.8 300 mm; in the order 105, 103, 500, 104 and 100 Å pore size) connected in series, and a Waters R401 differential refractometer. 90 mL of sample solutions (2.5 mg/mL) were injected and eluted at a flow rate of 1 mL/min using CHCl3 as the mobile phase. The fractionation was performed by collecting about 100 equal volume fractions (12 drops each) corresponding to about 0.165 mL. MALDI-TOF analysis Matrix-assisted laser desorption ionization time-of-flight (MALDI-TOF) mass spectra were obtained using VoyagerDE2 STR and the Voyager-DE2 (PE Biosystems) instruments, both equipped with a nitrogen laser emitting at 337 nm with a 3 ns pulse width and working in positive ion mode. The accelerating voltage was 20–25 kV, the grid voltage and delay time (delayed extraction, time-lag) were optimized for each sample to obtain the higher molar mass values. The laser irradiance was maintained slightly above threshold. The MALDI-TOF mass spectra of the unfractionated PC1 and PC2 samples, and those of fractions with molar mass lower than 5000 Da, were acquired in reflectron mode with the Voyager-DE2 STR instrument, which gave a mass resolution greater than 3000, measured from full peak width measured at half-height. MALDI sample preparation HABA (0.1 mol Lÿ1 in tetrahydrofuran (THF)) was used as the matrix for the two unfractionated PC samples and for SEC fractions containing low molar mass polycarbonate Copyright # 1999 John Wiley & Sons, Ltd. (mass range 1000–12000), whereas IAA (0.1 mol Lÿ1 in THF) was used as the matrix for SEC fractions with higher molar mass. Solutions of the unfractionated PC1 and PC2 polycarbonate samples were prepared in THF at a concentration of 2mg/mL, whereas the PC1 and PC2 SEC fractions were dissolved in 80 mL of THF after complete evaporation of the eluent. Equal volumes of polycarbonate sample solutions and matrix solution were mixed in order to obtain a 1:1 or 1:3 ratio (v/v, sample/matrix). One microliter of each sample/matrix solution was spotted on the MALDI sample holder and allowed to dry slowly in order to improve matrix crystallization. Molar mass calculations The molar mass of the two polydisperse PC samples were calculated from the SEC curves by the Polymer Lab Caliber software using the absolute calibration curves obtained by plotting the log Mw of each SEC selected fraction (calculated from the MALDI-TOF spectra) as a function of the corresponding elution volume. The average molar masses (Mn and Mw) of all SEC fractions were calculated from MALDI-TOF analysis by the appropriate Perseptive-Voyager-Grams software, using Eqns ((1)) and ((2)): Mn Ni Mi =Ni 1 Mw Ni Mi2 =Ni Mi 2 where Ni is the relative abundance of the polymer chain with molecular mass Mi. The Ni values were measured as the relative intensity of the molecular ion peak with a molecular mass Mi in the MALDI spectrum. Rapid Commun. Mass Spectrom. 13, 2260–2267 (1999) 2264 ANALYSIS OF POLYCARBONATES BY SEC/MALDI 1 Figure 4. MALDI-TOF spectra of the SEC fractions from sample PC1 collected at: 26.65 mL (a), 27.8 mL (b) and 28.54 mL (c). Viscometry Inherent viscosity [inh = (ln r)/C;C = 0.5 g/dL] measurements were performed at 30 0.1 °C in CHCl3 as solvent using a Desreux-Bishoff suspended-level viscometer, yielding the values of 0.42 and 0.44 for PC1 and PC2, respectively. RESULTS AND DISCUSSION End group determination The MALDI spectra of the two unfractionated polycarbonate samples (PC1 and PC2) are reported in Figs 1 and 2, respectively, showing well-resolved peaks up to 16 000 Da. Rapid Commun. Mass Spectrom. 13, 2260–2267 (1999) These spectra were recorded in reflectron mode with a mass resolution of about 3000, allowing the accurate structural characterization of the two polymer samples. The inset expansions in Figs 1 and 2 show the presence of peaks belonging to six mass series, assigned in Table 1, which are desorbed as [M Na] adducts accompanied by less intense potassiated adducts. Cyclic oligomers (species of type D, Table 1) appear as the most abundant species at low masses in both samples and their peak intensity decreases rapidly with size. Different end groups are attached to the linear chains of the two polycarbonates. Oligomers of type A, bearing phenylcarbonate groups at both chain ends (Table 1), predominate in sample PC1 at masses higher than 2500 Da (Fig. 1). Copyright # 1999 John Wiley & Sons, Ltd. ANALYSIS OF POLYCARBONATES BY SEC/MALDI 1 2265 Figure 5. MALDI-TOF spectra of the SEC fractions from sample PC1 collected at: 30.2 mL (a), 35.3 mL (b) and 40.0 mL (c). In the MALDI spectrum of sample PC2 the peaks corresponding to the sodiated oligomers of type A' (Fig. 2), containing p-tert-butylphenylcarbonate at both chain ends (Table 1), become the most intense starting from 4000 Da and are accompanied by satellite peaks (plus 16 mass units) due to the potassiated species. Oligomers of type C (Fig. 1, Table 1) can be assigned to polycarbonate chains containing phenol groups at both ends. In the inset expansion in Fig. 1, species C appearat 14 Da higherthan the most intense peak series (due to species A, Table 1). Oligomers of type C can also be identified in the inset expansion of the MALDI spectrum in Fig. 2, Copyright # 1999 John Wiley & Sons, Ltd. although they appear with low intensity. In Figs 1 and 2 the oligomer series B and B' (Table 1), containing phenol/ phenylcarbonate and phenol/p-tert-butylphenylcarbonate chain ends, respectively, can also be detected. Our results are in agreement with recent reports on end group analysis of some industrial BPA-polycarbonates.8,16,17 A minor ion fragmentation process due to the use of HABA as a matrix (see experimental section) has been claimed to occur in the MALDI spectra of PC samples,8 but the latter claim was not confirmed in the present work. Rapid Commun. Mass Spectrom. 13, 2260–2267 (1999) 2266 ANALYSIS OF POLYCARBONATES BY SEC/MALDI 1 Figure 6. SEC calibration lines of linear (&) and cyclic (*) PC chains, calculated by separate summation of molar masses of mass spectral peaks corresponding to linear and cyclic oligomers, respectively, observed in the MALDI-TOF spectra of the SEC fractions of sample PC1. SEC/MALDI measurements The average molar masses (Mw and Mn) of the PC1 and PC2 samples were determined by the off-line SEC/MALDI method6,7 (see experimental section), which consists of the analytical SEC fractionation of polydisperse polymers to collect numerous fractions (about 100) containing nearly monodisperse samples. The amount of sample present in each fraction (about 2.5 mg on average) exceeds many times the quantity needed for a MALDI spectrum. The weight average molar mass values (Mw) of these fractions are calculated from the MALDI spectra and are plotted against the corresponding elution volumes (Ve), producing an absolute MM calibration line which enables the computation of the average MM and MMD of the unfractionated sample directly from its SEC trace. Figure 3 reports the SEC traces of samples PC1 and PC2, together with the calibration line constructed after the fractionation of these two samples and the MALDI analysis of the narrow distribution fractions collected. The mass range covered by the MALDI spectra of the PC fractions covered the mass range 1000 to 100 000 Da (see below). The average MM values computed by using the calibration line in Fig. 3 are as follows: PC1: Mw = 22 200, Mn = 10 300; PC2: Mw = 28 000, Mn = 13 200. These values are lower by a factor of two with respect to those calculated by using the calibration line obtained from polystyrene standards (PC1: Mw = 55 800, Mn = 23 600; PC2: Mw = 59 300, Mn = 27 300). (The universal calibration curve was obtained using ten PS, ten PMMA, and five PEG samples. These samples were primary SEC standards obtained by anionic polymerization, and their MMs were all accurately measured by conventional techniques.20) The strong discrepancy between the two sets of data is most likely ascribed to the higher hydrodynamic volume of Rapid Commun. Mass Spectrom. 13, 2260–2267 (1999) polycarbonate with respect to the random coiled polystyrene standards. These results are in agreement with data reported in the literature, where the SEC was performed both with polystyrene SEC standards and with monodisperse polycarbonate samples.23 Figures 4(a)–(c) show the MALDI spectra of the SEC fractions collected from sample PC1 at the lower elution volumes (26.7, 27.8, 28.5mL, respectively). These spectra display well-resolved oligomer peaks up to 70 000Da (Fig. 4(b)), allowing the identification of the polymer structure. In fact, the MALDI spectra in Figs 4(b) and 4(c) are mainly constituted of linear chains terminated by phenylcarbonate groups at both ends (oligomers of type A, Table 1). Figures 5(a)–(c) display instead the MALDI spectra of the SEC fractions collected from sample PC1 at higher elution volumes (30.2, 35.3, 40.0mL, respectively). The MALDI spectra in Figs 5(a) and 5(b) still show oligomers of type A as major components, but in the inset of Fig. 5(b) other, less abundant, species (oligomers B, C and D, Table 1) are identified. In the MALDI spectrum in Fig. 5(c), obtained from the PC1 fraction taken at 40.0 mL, the most abundant peaks are those corresponding to cyclic oligomers (species D, Table 1). Similar results were obtained in the MALDI analysis of PC2 SEC fractions. It should be remarked that even those fractions collected (Fig. 3) at high elution volumes (>37 mL), where the SEC traces are flat suggesting that the PC had been totally eluted, in fact produced excellent MALDI spectra. This is due to the high sensitivity of MALDI, and also to the fact that, although the weight fraction of these low-mass eluted polycarbonate oligomers is quite low, the number fraction is still sizeable. Owing to the presence of both linear and cyclic chains, the SEC calibration line reported in Fig. 3 shows a deviation from linearity at elution volumes higher than 34.0 mL. Copyright # 1999 John Wiley & Sons, Ltd. ANALYSIS OF POLYCARBONATES BY SEC/MALDI 1 Although such deviations are not unusual and might appear to not merit further consideration, in the present case it has a very specific origin. It depends on the fact that the molar masses of the SEC fractions were calculated from the MALDI spectra as summations of all the peaks appearing in the spectra, irrespective of whether they belonged to linear or cyclic oligomers. It is well known19–21 that cyclic species are eluted at later times with respect to linear chains having the same backbone length. As a consequence, the calibration lines for cyclic and linear PDMS chains are different19–21 and, according to theory, the ratio (Mcycle/Mlinear)Ve should be 1.24.19 However, the MALDI spectra of the low mass fractions (Fig. 5) allow the detection of linear and cyclic oligomers contained in these PC samples, and separate average molar masses for linear and for cyclic oligomers can be calculated. When this procedure was applied, two slightly differing SEC calibration plots were obtained (Fig. 6), due to the smaller hydrodynamic volume of the polycarbonate cyclic chains with respect to the linear ones. In agreement with the theory, the ratio (Mcycle/Mlinear)Ve was found to be 1.22. CONCLUSIONS Because of its unprecedented sensitivity, MALDI-TOF constitutes an excellent SEC detector and the combination of SEC with MALDI brings new possibilities to polymer analysis. Our results show that the MALDI spectra of the SEC fractions allow not only the detection of linear and cyclic oligomers contained in these samples, but also the simultaneous determination of the average molar masses of linear and cyclic oligomers. Two parallel SEC calibration plots were obtained, and the ratio (Mcycle/Mlinear) was found to be in agreement with theory. Acknowledgements Partial financial support from the Italian Ministry for University and for Scientific and Technological Research (MURST), from Progetto Finalizzato Materiali per Tecnologie Avanzate II (CNR, Rome), and Copyright # 1999 John Wiley & Sons, Ltd. 2267 from the National Council of Research (CNR, Rome) is gratefully acknowledged. REFERENCES 1. Bahr U, Deppe A, Karas M, Hillenkamp F, Giessman U. Anal. Chem. 1992; 64:2866. 2. Montaudo G, Montaudo MS, Puglisi C, Samperi F. Rapid Commun. Mass Spectrom. 1995; 9:453. 3. Larsen BS, Michelle Byrd HC, McEwen CN. Symposium on Synthetic Polymer Analysis by Mass Spectrometry, Pittcon, March 12, 1999, Orlando, FL; 1365. 4. Scamporrino E, Maravigna P, Vitalini D, Mineo P. Rapid Commun. Mass Spectrom. 1998; 12:646. 5. Servaty S, Kohler W, Meyer WH, Rosenauer C, Spickermann J, Rader HJ, Wegner G. Macromolecules 1998; 31:2468. 6. Montaudo G, Garozzo D, Montaudo MS, Puglisi C, Samperi F. Macromolecules 1995; 28:7983. 7. Montaudo G, Montaudo MS, Puglisi C, Samperi F. Rapid Commun. Mass Spectrom 1995; 9:1158. 8. Nielen MWF, Malucha S. Rapid Commun. Mass Spectrom. 1997; 11:1194; Nielen MWF. Anal. Chem. 1998; 70:1563. 9. Dwyer J, Botten D. Internat. Lab. January, 1997; 13A. 10. Fei X, Murray KK. Anal. Chem. 1996; 68:3555. 11. Montaudo G. Polym. Prep. (A.C.S., Div. Polym. Chem.) 1996; 36:290. 12. Danis PO, Saucy DA, Huby FJ. Polym. Prep. (A.C.S., Div. Polym. Chem.) 1996; 36:311. 13. Wiley RH. Macromolecules 1971; 4:264. 14. Foti S, Giuffrida M, Maravigna P, Montaudo G. J. Polym. Sci., Polym. Chem. Ed. 1983; 21:1567. 15. Puglisi C, Sturialc L, Montaudo G. Macromolecules 1999; 32:2194. 16. Montaudo G, Montaudo MS, Puglisi C, Samperi F. Anal. Chem. 1994; 66:4366. 17. Przybilla L, Rader HJ, Mullen K. Eur. Mass Spectrom. 1999; 5:133. 18. Maravigna P, Montaudo G. in Comprehensive Polymer Science Vol. 5, Pergamon Press Oxford, 1989; 63; Montaudo G. Macromolecules 1991; 24:5289. 19. Wright PV, Beevers MS. In Cyclic Polymers, Semlyen JA (ed). Elsevier, 1986; 109. 20. Montaudo G, Montaudo MS, Puglisi C, Samperi F. Rapid Commun. Mass Spectrom. 1998; 12:519. 21. Horbach A, Vernaleken H, Weirauch K. Makromol. Chem. 1980; 181:111. 22. Puglisi C, Samperi F, Montaudo G, Carroccio S. Rapid Commun. Mass Spectrom. 1999; 13:000. 23. Bruessau RJ. Macromol. Symp. 1996; 110:15. Rapid Commun. Mass Spectrom. 13, 2260–2267 (1999)
© Copyright 2026 Paperzz