Presented at the XX Congress of The International Society on Thrombosis and Haemostasis, August 6-12, 2005, Sydney, Australia Characterization and Performance of the HemosIL™ Silica Clotting Time (SCT) Test for Detection of Lupus Anticoagulants (LA). R. Bottenus, C. Cueto, M. Triscott Instrumentation Laboratory, Orangeburg, NY, USA. Introduction The Silica Clotting Time (SCT) is an APTT-based assay first described by Chantarangkul, et.al., Thrombosis Research 67: 355-365, 1992. The current version utilizes synthetic phospholipid for the cephalin component of the reagents, instead of the original bovine cephalin. The kit contains a low concentration phospholipid screening reagent and a high concentration phospholipid confirmatory reagent. A calcium chloride activator is also provided that contains polybrene to minimize heparin interference. It is recommended that the results of the screening and confirmatory assays are calculated using a normalized ratio, to help increase the specificity of the assay. These studies evaluated the performance of the HemosIL SCT assay, relative to the HemosIL LAC DRVVT assay. Materials • HemosIL Silica Clotting Time (SCT) kit contains vials of three liquid reagents: - SCT Screen Reagent - SCT Confirm Reagent - SCT Calcium Chloride • HemosIL LAC Screen and LAC Confirm kits • HemosIL Normal Calibrator • Staclot LA (Diagnostica Stago) Methods and Results SCT Ratio - Procedures and Calculations The following procedure should be used to calculate the SCT Screen and SCT Confirm ratios: 1. For each new lot of SCT Screen and SCT Confirm kit, determine new Normal Range according to NCCLS Document H21-A4, Vol. 23, No. 35. 2. Determine the Mean of each Normal Range in seconds. 3. Use the mean of each normal range as a constant denominator in the calculations of ratios. SCT Screen Ratio = Patient Screen results (in seconds) Mean of Screen Normal Range (in seconds) SCT Confirm Ratio = Patient Confirm results (in seconds) Mean of Confirm Normal Range (in seconds) Normalized SCT Ratio = Screen Ratio Confirm Ratio Interpretation The final result should be expressed as the Normalized SCT Ratio. If ratio is > 1.24 (ACL Futura/ACL Advance) or > 1.20 (ACL 8000/9000/10000), then LA is present If ratio ≤ 1.24 (ACL Futura/ACL Advance) or ≤ 1.20 (ACL 8000/9000/10000) and SCT Screen and SCT Confirm clotting times are prolonged, then mixing studies should be performed to investigate factor deficiencies or inhibitors. If the mixing test is still prolonged, some anticoagulant other than LA may be present in the test plasma. Sample Type SCT Screen SCT Confirm LAC Screen LAC Confirm Normalized Ratio SCT LAC Sec Sec Sec Sec Normal 35.9 33.4 31.6 28.2 1.06 0.98 OAT 64.1 63.2 67.3 52.1 1.03 1.12 DIC 33.4 40.9 34.9 35.7 0.83 0.85 UFH 32.7 34.8 35.2 35 0.95 0.87 LMWH 32.6 28.6 30.5 26.5 1.16 1.00 Factor VIII Def. 46.9 49.1 28.1 28.2 0.97 0.86 Factor IX Def. 73.6 88.1 37.9 31.7 0.85 1.04 Factor XI Def. 54.5 56.3 39.9 36.6 0.98 0.94 Factor XII Def. 43.6 49.1 45.4 39.5 0.9 1.00 LA-ACA 101.3 38.8 100.2 39.1 2.65 2.22 LA-APA 84.2 50.1 124.6 44.3 1.7 2.44 A-ß2GPI 57.6 37.3 55.6 29.6 1.57 1.63 Table 2 Normal Range Study One hundred twenty fresh, citrated plasma samples (60 female and 60 male) were obtained from apparently healthy donors and tested on an ACL Advance using a single lot of HemosIL Silica Clotting Time (SCT Screen and SCT Confirm) reagents. The distribution of the normalized ratios is parametric as shown in the frequency distribution plot below. Abnormal Samples OAT DIC LA-APA A-ß2GPI LA-ACA UFH LMWH Factor VIII Def. Factor IX Def. Factor XI Def. Factor XII Def. Quanity 20 10 10 10 10 7 7 4 4 4 4 Normalized Ratios Method Comparison N Normal Range Study Figure 1 The parametric reference interval (normal range) was calculated according to the guidelines established by NCCLS C28-A2, i.e. 95% reference interval with 90% confidence intervals (2 SD) around the reference limits. These data support the normal range for the normalized ratio in the HemosIL Silica Clotting Time product insert of the ACL Advance, n=120, Normalized SCT Ratio: 0.70 to 1.24. Due to many variables that may affect results, each laboratory should establish its own normal range. Method Comparison Using SCT and LAC Reagents with Patient Samples A comparative performance study was conducted using 210 patient citrated plasma samples (120 normals and 90 abnormals) analyzed in singlet with HemosIL Silica Clotting Time (SCT Screen and SCT Confirm) versus the predicate HemosIL LAC Screen/LAC Confirm on an ACL Advance. SCT Normalized Ratio 4.00 3.00 y = 1.0989x - 0.086 r = 0.8738 n = 210 2.00 1.00 0.00 0.00 1.00 2.00 3.00 4.00 LAC Normalized Ratio Figure 2 Sensitivity and Specificity The sensitivity and specificity of HemosIL Silica Clotting Time (SCT Screen/SCT Confirm) as compared directly to the predicate HemosIL LAC Screen/LAC Confirm were also calculated from this study: Table 3 LAC Screen/Confirm Normalized Ratio SCT Screen/Confirm Normalized Ratio • Relative Sensitivity (31/33) + - + 31 3 - 2 174 = 93.9% (95% C.I. = 77.0 - 96.9) • Relative Specificity (174/177) = 98.3% (95% C.I. = 95.9 - 98.7) • Overall Agreement (205/210) = 97.6% Final Interpretation Calculation Mixing studies and an additional qualitative test, Staclot LA from Diagnostica Stago was used to resolve the five discrepant results. The table below shows the final interpretation of results. other interferences. The ease of use of SCT on automated instruments for screening and confirmation of Lupus inhibitors makes this test a convenient alternative for LA Screening. Table 4 Final Interpretation SCT Screen/Confirm Normalized Ratio + - + 31 3 - 0 176 • Relative Sensitivity (31/31) = 100% (95% C.I. = 89.0 - 100.0) • Relative Specificity (176/179) = 98.3% (95% C.I. = 95.9 - 98.7) • Overall Agreement (207/210) = 98.6% Table 5 Sample Type SCT LAC Staclot Mixing Study Final Cut-off > 1.24 Cut-off > 1.2 Cut-off > 8 sec N. Ratio > 1.24 Interpretation Normal 1.34(+) 1.00(-) 2.6(-) UF Heparin OAT Factor VIII Factor XII 1.13(-) Negative 1.28(+) 1.12(-) 1.01(-) 1.22(+) 4.5(-) 1.10(-) Negative 2.5(-) Not Done Negative 1.44(+) 0.82(-) 0.98(-) -1.4(-) 1.02(-) Negative 1.21(+) 2.4(-) 1.04(-) Negative Conclusions The HemosIL Silica Clotting Time assay is a convenient automated test for detection of LA, that runs on ACL instrument platforms. The SCT assay can be used with a second screening test (e.g. DRVVT- based) to satisfy current guidelines on the use of at least two different LA assay principles for detection of LA. The sensitivity and specificity of the SCT assay is comparable to the DRVVT-based HemosIL LAC assay. Analysis of 210 patient citrated plasma samples (120 normals and 90 abnormals) with SCT and LAC assays gave a relative Sensitivity of 93.9% (versus LAC), a relative Specificity of 98.3% and an overall agreement of 97.6%. Both tests identified all the Lupus-positive samples correctly. Five samples were found to be discrepant between the two assays, but upon follow-up mixing studies, the samples were correctly identified as negative samples. Using a normalized ratio to calculate the Screen and Confirm assay results allows the SCT assay to be insensitive to OAT patients, heparin patients, as well SCTBOT.PS 08/05
© Copyright 2026 Paperzz