VASCULAR & NONVASCULAR PROCEDURES EQUIPMENT AND ACCESSORIES USED IN THE VASCULAR LAB MAJOR EQUIPMENT X-RAY/FLUORO TUBE(S) • IMAGE INTENSIFIER/TV MONITOR • X-RAY TABLE • RAPID FILM CHANGERS • PRESSURE INJECTORS • DIGITAL SUBTRACTION • VCR • CATHETERS/GUIDEWIRES • ACCESSORIES OXYGEN SUPPLY • CONTRAST MEDIA/ WARMER • IV SUPPLIES, NEEDLES, SYRINGES • PHARMACOLOGICAL SUPPLIES • CRASH CART • EKG MONITOR • 1 CONTROL ROOM CONTROL PANEL (GENERATOR) • PROGRAM SELECTOR (1 PER FILM CHANGER) • DIGITAL COMPUTER • X-RAY TUBE MIN. 6 FILM/SEC x 5 SEC. CAPACITY • HIGH SPEED ROTOR (10K RPM) • 3-PHASE GENERATOR • 1,000 MA CAPACITY (MINIMUM) • FRACTIONAL FOCAL SPOT (0.3 mm) • ANODE ANGLE = 5° - 7° (Vs 10° - 15°) • SERIAL FILM CHANGER USED FOR STUDIES REQUIRING RAPID FILMING SEQUENCES USING 10X12 OR 14X14 INCH FILM. 2 FILM CHANGERS • ROLL FILM • CUT FILM ROLL FILM CHANGER USED A REEL OF 14” WIDE FILM. AFTER AN IMAGING SEQUENCE, THE TRAILING EDGE OF THE FILM IS CUT AND THE TAKE-UP ROLL IS PROCESSED. IMAGES ARE THEN CUT TO 14X14 “ RADIOGRAPHS. ROLL FILM CHANGER CAPABLE OF 4 TO 12 FILMS PER SECOND DEPENDING ON MANUFACTURER. THEY WERE SIMPLE TO OPERATE, BUT WASTED FILM. ROLL FILM WAS CUMBERSOME AND EASILY FOGGED. 3 CUT FILM CHANGERS MAGAZINE-LOADED FILM DEVICE DESIGNED TO USE 10X12 OR 14X14 INCH FILM. FILM RATE RANGES FROM 4 TO 6 FILMS PER SECOND. CUT FILM CHANGERS • LESS WASTE THAN ROLL CHANGERS • EASY FILM HANDLING • SLOWER FILM RATES • FREQUENT RELOADING FILM CHANGERS & BIPLANE FILMING FILM CHANGERS CAN BE POSITIONED ORTHOGONALLY TO PRODUCE RADIOGRAPHIC IMAGES IN TWO POSITIONS AT ONCE. TWO X-RAY TUBES ARE NEEDED, ONE FOR EACH FILM CHANGER. 4 BIPLANE FILMING XR A Y F I L M PATIENT X-RAY FILM PROGRAMMING COORDINATION OF THE FILMING SEQUENCE WITH THE CONTRAST INJECTOR, TABLE MOVEMENT AND XRAY EXPOSURE. 5 PROGRAM SELECTOR A DEVICE WHICH IS LOCATED IN THE CONTROL ROOM. MODERN PROGRAMMERS USE MICROPROCESSORS (COMPUTERBASED) WHICH EXECUTE THE INSTRUCTIONS PROGRAMMED INTO IT BY THE TECHNOLOGIST. PROGRAMMING VARIABLES • FILM CYCLE TIME •STATIONARY TIME (%) •TIME IN MOTION (%) • X-RAY GENERATOR •ZERO-TIME PHASING-IN TIME • MAXIMUM PERMISSIBLE EXPOSURE TIME FILM CYCLE TIME THIS IS THE TIME THAT THE FILM IS EITHER MOVING THROUGH THE FILM CHANGER OR REMAINING STATIONARY. EACH IS INDICATED IN PERCENTAGE TERMS. EXPOSURES ARE MADE WHEN THE FILM IS STATIONARY. 6 FILM CYCLE TIME NOT ALL OF THE STATIONARY TIME IS AVAILABLE FOR X-RAY EXPOSURE. THIS IS DUE TO DELAYS THAT OCCUR IN THE XRAY GENERATOR. ZERO TIME THE DELAY INHERENT IN OLDER SYSTEMS WHICH RELY ON MECHANICAL SWITCHES TO SIGNAL FOR AN EXPOSURE. MODERN SOLIDSTATE GENERATORS HAVE MADE ZERO TIME ALMOST NON-EXISTENT. PHASING-IN TIME WHEN THE ACTUAL EXPOSURE IS TRIGGERED, A DELAY CAN OCCUR DEPENDING ON THE POSITION OF THE SINE WAVE AT THAT MOMENT. ON 3-PHASE SYSTEMS, THE AVERAGE DELAY IS 2.78 ms. 7 MAXIMUM PERMISSIBLE EXPOSURE TIME IN ORDER TO PROPERLY PREVENT AN X-RAY EXPOSURE FROM OCCURRING WHILE THE FILM IS IN MOTION THROUGH THE FILM CHANGER, THE MAXIMUM EXPOSURE TIME MUST BE DETERMINED. MAXIMUM PERMISSIBLE EXPOSURE TIME -- STEP 1 EXPOSURE TIME 1000 _______________________________ = PER FILM # FILMS EXPOSED PER SECOND IN MSECONDS MAXIMUM PERMISSIBLE EXPOSURE TIME -- STEP 2 EXPOSURE TIME PER X FILM IN MSEC % THAT FILM IS = STATIONARY EXPOSURE TIME AVAILABLE PER FILM 8 MAXIMUM PERMISSIBLE EXPOSURE TIME -- STEP 3 EXPOSURE TIME - PHASING-IN TIME = AVAILABLE PER FILM IN MSECONDS MAXIMUM PERMISSIBLE EXPOSURE TIME PROBLEM: WHAT IS THE MAXIMUM PERMISSIBLE EXPOSURE TIME ON A 6 EXPOSURE/SEC FILMING SEQUENCE USING A 3-PHASE UNIT WITH A FILM STATIONARY TIME OF 40%? SOLUTION: • 1000/6 = 166.7 MSEC PER FILM • 166.7 X .40 = 66.7 MSEC AVAILABLE PER FILM • 66.7 - 2.78 = 63.9 MSEC MAXIMUM PERMISSIBLE PER FILM 9 PRESSURE INJECTORS ALSO CALLED ELECTROMECHANICAL INJECTORS ARE DEVICES WHICH CAN DELIVER IODINATED CONTRAST AT A PRECISE RATE AND PRESSURE FOR A SPECIFIED TIME. PRESSURE INJECTOR TYPES • CONSTANT FLOW RATE • CONSTANT PRESSURE 10 CONSTANT FLOW RATE INJECTORS PERMIT THE TECHNOLOGIST TO SET THE FLOW RATE (IN ML/SEC) AND THE INJECTION TIME (IN SECONDS). INJECTION PRESSURE WILL DEPEND ON VISCOSITY OF THE CONTRAST, DIAMETER, LENGTH AND TYPE OF THE CATHETER USED. CONSTANT PRESSURE INJECTORS REQUIRE SETTING THE PRESSURE (IN PSI). RESULTING FLOW RATE WILL DEPEND ON CATHETER LENGTH, CATHETER INSIDE DIAMETER, AND VISCOSITY OF CONTRAST PRESSURE INJECTORS -OPTIONAL EQUIPMENT • EKG MONITOR • DUAL SYRINGE SYSTEM 11 SAFETY CONSIDERATIONS • PRESSURE LIMIT CONTROL • RATE RISE CONTROL • MECHANICAL STOP • ELECTRICALLY INSULATED PLASTIC JACKET PROBLEM: 25 ML OF IODINATED CONTRAST IS INJECTED OVER THE COURSE OF 6 SECONDS. WHAT IS THE FLOW RATE? SOLUTION: • DOSE = FLOW RATE X TIME • 25 ML = FLOW RATE X 6 SEC • 25 ML/6 SEC = FLOW RATE • 4.17 ML/SEC = FLOW RATE 12 ANGIOGRAPHIC CATHETERS SLENDER FLEXIBLE TUBES OF VARYING LENGTHS, SHAPES AND DIAMETERS WHICH ARE INSERTED INTO BLOOD VESSELS IN ORDER TO DELIVER IODINATED CONTRAST. ANGIOGRAPHIC CATHETERS MADE FROM TEFLON, POLYETHYLENE OR POLYURETHANE. THEY COME WITH SINGLE OR MULTIPLE HOLES. ANGIOGRAPHIC CATHETERS TO GIVE ADDED STRENGTH AND THE ABILITY TO RESPOND TO TORQUE, SOME CATHETERS ARE BRAIDED. THAT IS, THE PLASTIC IS IMBEDDED WITH A WIRE MESH. 13 BRAIDED CATHETER ANGIOGRAPHIC CATHETERS THE SHAPE OF THE DISTAL END IS DESIGNED TO PERMIT INSERTION INTO THE APPROPRIATE VESSEL, ESPECIALLY DURING SELECTIVE STUDIES. ANGIOGRAPHIC CATHETERS THE NUMBER OF HOLES AT THE END OF THE CATHETER DEPENDS ON THE SIZE OF THE VESSEL TO BE CATHETERIZED. 14 ANGIOGRAPHIC CATHETERS SIZES ARE INDICATED USING THE FRENCH (F) SCALE. EACH FRENCH SIZE IS EQUAL TO 0.33 mm. THE OUTSIDE DIAMETER OF THE CATHETER DETERMINES THE FRENCH SIZE. ANGIOGRAPHIC CATHETERS THE AVERAGE ADULT REQUIRES THE USE OF A 4 OR 5 F CATHETER (WHICH IS 1.3 mm or 1.7 mm IN DIAMETER). GUIDEWIRES STAINLESS STEEL WIRE IS USED TO GUIDE THE ANGIOGRAPHIC CATHETER TO THE DESIRED LOCATION IN A BLOOD VESSEL. 15 GUIDEWIRES MAY BE OF SOLID STEEL CONSTRUCTION (A SINGLE WIRE) OR CONSIST OF A THIN STAINLESS STEEL CORE AROUND WHICH MORE STEEL IS WRAPPED. GUIDEWIRES GUIDEWIRES MAY HAVE A RIGID OR FLEXIBLE “J” TIP. FLEXIBLE TIPS HAVE MOVABLE CORES. 16 VESSEL DILATORS DILATORS ARE USED TO DILATE THE TISSUE TRACT AND LUMEN OF THE BLOOD VESSEL. IT IS SLIPPED OVER THE GUIDEWIRE. NEVER DILATE A VESSEL LARGER THAN THE CATHETER TO BE USED. 17
© Copyright 2026 Paperzz