R E S E A R C H Internal contamination of air-driven low-speed handpieces and attached prophy angles Judith R. Chin, DDS, MS; Chris H. Miller, MS, PhD; Charles John Palenik, MS, PhD, MBA A J ✷ D A ✷ I A T IO N Background. In an in vitro crossover study, the authors investigated whether the interior of lowN C speed handpiece/prophy-angle systems becomes conU A ING EDU 3 taminated during operation and submersion into RT ICLE Geobacillus stearothermophilus. Methods. This study involved two types of handpieces attached to eight brands of prophy angles. The researchers operated angles attached to sterile handpieces for 60 seconds. They then analyzed the inside surfaces of the angle, nosecone and motor. They tested each prophy angle and handpiece 10 times. Results. In the 160 tests of handpieces contaminated at the prophy cup end, the spores traveled into the motor gears 32 times (20 percent). In the other 160 tests in which the motor gears were contaminated, the test bacterium traveled through the prophy cup in 75 instances (47 percent). Conclusions. The in vitro data suggest that low-speed handpiece motors can become contaminated internally during use with prophy angles. Also, internal contaminants appear to have been released from the handpiece. Clinical Implications. The results suggest that low-speed handpieces can become contaminated internally during use. Unless low-speed handpieces are sterilized properly between patients, they may become cross-contaminated. Key Words. Infection control; handpiece; prophy angle; Geobacillus stearothermophilus; equipment contamination; sterilization. JADA 2006;137(9):1275-80. T T ABSTRACT CON he Centers for Disease Control and Prevention recommends that dentists “clean and heatsterilize handpieces and other intraoral instruments that can be removed from the air and waterlines of dental units between patients.”1 Several studies have recommended heat sterilization of high-speed handpieces because of the potential for internal contamination during use.2-9 The justification for the heat sterilization of the low-speed handpiece system is less clear. Pressurized air is needed to operate the air-driven low-speed handpiece. This air must escape or be reduced to eliminate excessive heat buildup. All disposable and reusable types of prophy angles have a vent or opening to reduce or eliminate excessive heat buildup. This vent may allow internal contamination of a low-speed handpiece system because it is not a sealed system. This could lead to subsequent cross-contamination unless the handpiece is heat-sterilized between uses. However, there is only preliminary information concerning the internal contamination of low-speed handpiece systems during use.10 Thus, we conducted a study to determine the potential for internal Dr. Chin is an associate professor, Oral Health Research Institute, School of Dentistry, Indiana University, 415 Lansing St., Indianapolis, Ind. 46202, e-mail “[email protected]”. Address reprint requests to Dr. Chin. Dr. Miller is the executive associate dean and a professor of oral microbiology, School of Dentistry, Indiana University, Indianapolis. Dr. Palenik is the sponsored research manager, School of Dentistry, Indiana University, Indianapolis. JADA, Vol. 137 http://jada.ada.org Copyright ©2006 American Dental Association. All rights reserved. September 2006 1275 R E S E A R C H contamination of low-speed handpiece motors and prophy angles and their potential for patient-topatient transmission of organisms. We tested two hypotheses. The first was that microbes can enter the prophy-angle/handpiece system at the prophy angle end and travel to the gears of the air-driven handpiece motor. The second hypothesis was that microbes on the gears of the low-speed handpiece air-driven motor can travel to and be expelled from the prophy angle. MATERIALS AND METHODS Study design. This study was conducted at a single center, with an eight-by-two crossover design using two types of low-speed handpieces attached to seven different brands of disposable and one brand of reusable prophy angle. In a crossover design, an object receives all treatments in sequence. By contrast, in a parallel-groups design, some objects get the first treatment and different objects get the second treatment. The crossover design represents a special situation in which there is not a separate comparison group. In effect, each object serves as its own control. Also, since the same object receives both treatments, there is no possibility of covariate imbalance. Low-speed handpieces. Two different styles of air-driven low-speed handpieces were used for this study. One was the Prophy Star single piece low-speed handpiece (Star Dental, Lancaster, Pa.), and the other was the Titan 3 two-piece lowspeed handpiece (Star Dental). Both were connected to the dental unit using a four-line swivel attachment (360 Quick Connect Swivel, Star Dental). We cleaned and sterilized the handpieces in a steam autoclave at 1,210 C for 30 minutes before use and after testing with each type of prophy angle. Prophy angles. We used eight types of prophy angles, seven disposable and one reusable. The seven disposable angles we used were dAcclean firm cup (Henry Schein, Melville, N.Y.); dClassic Traditional Web (Young Dental, Earth City, Mo.); dClassic Twin Rib GUM (Sunstar Butler, Chicago); dClinpro Firm (3M ESPE, St. Paul, Minn.); dDENSCO Firm Yellow (Waterpik Technologies, Newport Beach, Calif.); dNupro revolv (Dentsply, York, Pa.); dOriginal Green regular (Denticator, Earth City, Mo.). 1276 JADA, Vol. 137 The reusable metal prophy angle we used was the TS2 Prophy Angle (Young Dental), with a disposable screw-type prophy cup (Classic Traditional Web, Young Dental). Experiment 1: testing for the spread of microbes from the prophy angle to the handpiece motor. We covered a sterile handpiece with a disposable plastic sleeve (Lowspeed Handpiece Sleeve, Denticator) and attached a prophy-angle with a cup. We then partially wrapped the prophy angle/handpiece unit with plastic wrap to further prevent contamination from the outside. The prophy angle’s vent was not covered. We submerged the prophy angle head aseptically in 12.5 milliliters of a suspension of sterile phosphate-buffered saline (PBS, Sigma Chemical, St. Louis) containing 2.0 × 106 colonyforming units per mL of Geobacillus stearothermophilus spores (ATCC 7953, Ethox International, STS Life Sciences Division, Rush, N.Y.) and 10 percent by volume of sterile defibrinated sheep blood (Colorado Serum, Denver). The spore/blood suspension was contained in a sterile 60-mL Pyrex centrifuge tube (No. 8540, Corning, Corning, N.Y.) (Figure 1). We covered the tube with a sterilized piece of aluminum foil to prevent external aerosol contamination. We turned on the handpiece and pressed the prophy cup against the inside walls of the tube and ran it 30 times within 60 seconds to simulate use. We then blotted the prophy angle/handpiece with a paper towel to remove excess external liquid and removed the plastic wraps aseptically. We donned fresh sterile gloves, separated the handpiece components, and carefully sampled and cultured the gears of the handpiece motor as described in the following section (Microbiological Analysis). We repeated each of the 16 combinations (eight types of prophy angles and two types of handpieces) 10 times with new sterilized handpieces and prophy angles, for a total of 160 units tested. As a negative control, we tested and sampled 20 uncontaminated handpieces to detect any microbial contamination from the environment. The negative controls tested consisted of 10 single-unit handpieces and 10 two-unit handpieces. We used only one brand of prophy angle (Classic Twin Rib GUM disposable) for all negative controls. Experiment 2: testing for the spread of microbes from the handpiece motor to the prophy angle. We placed 0.1 mL of 2.0 × 106 CFU/mL G. stearothermophilus in PBS con- http://jada.ada.org September 2006 Copyright ©2006 American Dental Association. All rights reserved. R E S E A R C H Figure 2. Inoculation of the single-unit handpiece motor. Figure 1. Handpiece system wrapped with plastic barriers and being tested for internal spread of contamination. taining 10 percent sheep blood on the motor gears of a sterilized handpiece (Figure 2). We covered the handpiece with a plastic sleeve and attached a prophy-angle with cup. We partially wrapped the prophy angle/handpiece unit with plastic wrap to further prevent outside contamination. The prophy angle’s vent was not covered. We submerged the prophy angle head in 12.5 mL of sterile PBS contained in a sterile 60-mL Pyrex glass centrifuge tube. We covered the tube with a sterilized piece of aluminum foil to prevent external aerosol contamination. We turned on the handpiece and pressed the prophy cup against the inside walls of the tube and ran it 30 times within 60 seconds. We removed the prophy angle/handpiece from the tube and cultured the PBS as described in the section below (Microbiological Analysis). We tested each of the 16 combinations (eight types of prophy angles and two types of handpieces) 10 times with new sterilized handpieces and prophy angles, for a total of 160 units Figure 3. Sampling the handpiece motor. tested. As a negative control, we tested 20 uncon taminated handpieces and sampled them to detect any microbial contamination from the environment. The negative controls tested consisted of 10 single-unit handpieces and 10 two-unit handpieces. Again, we used only one brand of prophy angle (Classic Twin Rib GUM disposable) for all negative controls. MICROBIOLOGICAL ANALYSIS Experiment 1. The goal of this analysis was to determine if the test spores had spread internally JADA, Vol. 137 http://jada.ada.org Copyright ©2006 American Dental Association. All rights reserved. September 2006 1277 R E S E A R C H TABLE 1 Number of prophy angle/handpiece units showing spread of bacteria from the prophy angle head to the handpiece motor gears. TYPE OF PROPHY ANGLE SINGLE-UNIT HANDPIECE† TWO-UNIT HANDPIECE† Acclean Disposable 1/10 3/10 Classic Traditional Web Disposable 0/10 8/10 Classic Twin Rib GUM Disposable 2/10 2/10 PROPHY ANGLE* PBS and incubated it at 56 C for seven days. Turbidity indicated positive growth, which we analyzed on TSA plates for G. stearothermophilus as described above. RESULTS Experiment 1: Testing for the spread of microbes from the Disposable 0/10 2/10 Clinpro prophy angle to the Disposable 1/10 2/10 DENSCO Firm Yellow handpiece motor. SamDisposable 0/10 8/10 Nupro revolv pling of the 20 uncontaminated negative control Disposable 2/10 0/10 Original Green handpieces did not reveal Metal 1/10 0/10 TS2 Prophy Angle bacterial growth. Table 1 Not applicable 7/80 25/80 TOTALS shows the results from the test handpiece units. Of the * The angles’ manufacturers are as follows: Acclean firm cup (Henry Schein, Melville, N.Y.); Classic Traditional Web (Young Dental, Earth City, Mo.); Classic Twin Rib GUM (Sunstar Butler, Chicago); 80 prophy-angle/single-unit Clinpro Firm (3M ESPE, St. Paul, Minn.); DENSCO Firm Yellow (Waterpik Technologies, Newport Beach, handpiece systems tested, Calif.); Nupro revolv (Dentsply, York, Pa.); Original Green regular (Denticator, Earth City, Mo.); TS2 Prophy Angle (Young Dental). seven exhibited test spores † Number with positive motor gears over total number of units tested. on the motor gears. Of the 80 prophy-angle/two-unit handpiece systems tested, 25 exhibited test from the prophy angle to the handpiece motor spores on the motor gears. In total, in the 160 after the simulated use. We disassembled the tests of handpieces contaminated at the prophy prophy-angle/handpiece units aseptically, taking cup end, the spores traveled into the motor gears great care not to contaminate the internal compo32 times (20 percent). nents. We sampled the gears of the handpiece Experiment 2: Testing for the spread of motor with a cotton-tipped applicator (Figure 3). microbes from the handpiece motor to the We then streaked the applicator onto trypticase prophy angle. Sampling of the uncontaminated soy agar (TSA) plates (Becton, Dickinson, Sparks, control handpieces did not reveal bacterial Md.), which we placed into trypticase soy broth growth. Table 2 shows the results from the test (TSB) supplemented with 0.25 percent glucose handpiece units. Of the 80 prophy angle/single(Becton, Dickinson). We incubated the TSA and unit handpiece systems tested, 30 released test TSB media for seven days at 56 C. After incubation, we analyzed both the TSA and TSB media spores into the PBS in which the prophy-angle for the presence of test bacteria. The appearance head was submerged. Of the 80 prophy angle/twoof turbidity in the TSB broth indicated microbial unit handpiece systems tested, 45 released the growth from the sampled site. We confirmed positest spores into the PBS in which the prophy tive growth as G. stearothermophilus by analysis angle head was submerged. In total, in the 160 of distinctive white colonies of gram-positive rods tests in which the motor gears were contamion the 56 C TSA plates. nated, the test bacterium traveled through the Experiment 2. The goal of this analysis was prophy cup in 75 instances (47 percent). to determine if the test spores had spread interDISCUSSION nally from the handpiece motor to the prophy angle and out into the environment after the simThe results from this in vitro study substantiate ulated use. We analyzed the PBS in which the those of an earlier preliminary study using simprophy head was submerged for the presence of ilar methodology with Serratia marcescens as the test spores. To do so, we streaked an aliquot of test microbe.9 In our study, operation of prophythe PBS onto a TSA plate, then added an equal angle/handpiece systems in a microbial environvolume of double-strength TSB to the remaining ment resulted in movement of those microbes 1278 JADA, Vol. 137 http://jada.ada.org September 2006 Copyright ©2006 American Dental Association. All rights reserved. R E S E A R C H internally from the prophy TABLE 2 angle to the gears of the Number of prophy angle/ handpiece units showing handpiece motor in 9 perspread of bacteria from the handpiece motor out cent of the single-unit through the prophy angle. handpiece systems studied and in 31 percent of the TWO-UNIT SINGLE-UNIT PROPHY ANGLE* TYPE OF PROPHY ANGLE HANDPIECE HANDPIECE two-unit handpiece systems studied. These results sup- Acclean 10/10 3/10 Disposable port the first hypothesis for 4/10 4/10 Disposable Classic Traditional Web this study, that microbes 5/10 3/10 Disposable Classic Twin Rib GUM can enter the prophyangle/handpiece system at 7/10 8/10 Disposable Clinpro the prophy angle end and 2/10 3/10 Disposable DENSCO Firm Yellow travel to the gears of the 8/10 5/10 Disposable air-driven handpiece motor. Nupro revolv They also support a recom- Original Green 7/10 2/10 Disposable mendation to clean and 1/10 2/10 Metal TS2 Prophy Angle heat-sterilize low-speed 45/80 30/80 Not applicable TOTALS handpiece motors, nose cones and reusable prophy- * The angles’ manufacturers are as follows: Acclean firm cup (Henry Schein, Melville, N.Y.); Classic Traditional Web (Young Dental, Earth City, Mo.); Classic Twin Rib GUM (Sunstar Butler, Chicago); angles between patients. Clinpro Firm (3M ESPE, St. Paul, Minn.); DENSCO Firm Yellow (Waterpik Technologies, Newport Beach, Calif.); Nupro revolv (Dentsply, York, Pa.); Original Green regular (Denticator, Earth City, Mo.); TS2 There were differences in Prophy Angle (Young Dental). levels of contamination † Number with positive motor gears over total number of units tested. between single- and twounit handpiece systems, as well as between discontribute to cross-contamination between posable plastic and reusable metal prophy angles. patients unless those handpiece motors, nose Differences between the types of handpieces may cones and reusable angles are heat-sterilized be attributed to the areas tested. We were able to between uses. So decontaminating (for example, test four areas of the two-unit handpiece system wiping off) just the outside of these items between versus three areas of the single-unit handpiece patients would not prevent potential for crosssystem. The single-piece handpiece system is contamination. factory-sealed between the motor of the handpiece Conditions of actual use of low-speed handpieces in the mouth differ from the simulated use and the gears of the motor, so we could not anainvolved in this study. Such differences would lyze this area. If we could have done so, it could include the number and types of microbes have shown higher levels of contamination than involved, the degree of stress placed on the in the two-unit handpiece system. The plastic disprophy cups, the types of prophy angles and posable prophy angles had higher levels of contahandpieces used, and the length of time the handmination than did the reusable metal prophy piece system is operated. A follow-up study anaangle. This may be attributed to the fact that we lyzing the internal surfaces of low-speed handtested only one brand of reusable metal angle piece systems actually used on patients would be versus seven disposable plastic angles. This study valuable. Although there could be a small chance analyzed the more common prophy angles availfor outside contamination of the internal portions able in the United States, and we did not test all of the handpiece systems tested in this study, we available prophy angles. took great care to prevent this by using plastic The results also supported the second hypothbarriers and other aseptic techniques. Also, we esis in this study—that microbes on the gears of detected no environmental contamination in the the low-speed handpiece air-driven motor can control handpieces. travel to and be expelled from the prophy angle. This release was demonstrated in 38 percent of CONCLUSIONS the single-unit handpiece systems tested and 56 The results of this in vitro study suggest that the percent of the two-unit handpiece systems tested. internal portions of some low-speed handpiece Therefore, this suggests that the internally conmotors have the potential to become contamitaminated handpiece systems have a potential to † † JADA, Vol. 137 http://jada.ada.org Copyright ©2006 American Dental Association. All rights reserved. September 2006 1279 R E S E A R C H nated during use with prophy angles. They also show that there is a potential for internal contaminants to be released through the prophy angle into the mouth of a patient. This could contribute to transmission of organisms between patients if low-speed handpiece motors, components and prophy angles are not heat-sterilized between uses. Future clinical trials analyzing the internal surfaces of low-speed handpiece systems actually used on patients are recommended. ■ The authors would like to thank Kenneth Burgess, Huong Vu, Richard Le and Sharon Gwinn for their expertise and assistance with this study. 1. Kohn WG, Collins AS, Cleveland JL, et al. Guidelines for infection control in dental health-care settings: 2003. Centers for Disease Control and Prevention. MMWR Recomm Rep 2003(52[RR-17]):1-61. 1280 JADA, Vol. 137 2. Epstein JB, Rea G, Sibau L, Sherlock CH. Rotary dental instruments and the potential risk of transmission of infection: herpes simplex virus. JADA 1993;124(12):55-9. 3. Epstein JB, Rea G, Sibau L, Sherlock CH, Le ND. Assessing viral retention and elimination in rotary dental instruments. JADA 1995;126(1):87-92. 4. Cottone JA, Molinari JA. State-of-the-art infection control in dentistry. JADA 1991;122(8):33-41. 5. Molinari JA. Dental handpiece sterilization: historical and technological advances. Compend Contin Educ Dent 1998;19(7):724-7. 6. Sanchez E, Macdonald G. Decontaminating dental instruments: testing the effectiveness of selected methods. JADA 1995;126(3):359-62, 364, 366, passim. 7. Saunders MJ, Cottone JA. What is good enough? Sterilization and its discontents. Tex Dent J 1987;104(9):20-9. 8. Underhill TE, Terezhalmy GT, Cottone JA. Prevention of crossinfections in the dental environment. Compend Contin Educ Dent 1986;7(4):260, 262, 264, passim. 9. Molinari JA, Palenik CJ. Environmental surface infection control, 2003. Compend Contin Educ Dent 2004;25(1 supplement):30, 32-4, 36-7. 10. Waskow JR, CH., Palenik, CJ. Justification for sterilizing slowspeed handpiece motors between patients (abstract 3181). J Dent Res 1996;75(special issue):415. http://jada.ada.org September 2006 Copyright ©2006 American Dental Association. All rights reserved.
© Copyright 2026 Paperzz