AMU pH/Redox Operator`s Manual

AMU pH/Redox
Operator’s Manual
Version 3.50 and later
SWAN sets the Standard
A-96.250.441 / 060923
Customer Support
SWAN and its representatives maintain a fully trained staff of technical specialists
around the world. For any technical question, contact your nearest
SWAN representative, or the manufacturer:
SWAN ANALYTISCHE INSTRUMENTE AG
Turicaphonstrasse 29
8616 Riedikon
Switzerland
Internet: www.swan.ch
E-mail: [email protected]
Document Status
Title:
AMU pH/Redox Operator’s Manual
ID:
A-96.250.441
Revision
Issue
00
September 2006
© 2006, SWAN ANALYTISCHE INSTRUMENTE AG, Switzerland, all rights reserved
Information contained in this document is subject to change without notice.
AMU pH-Redox
AMU pH-Redox
AMU pH-Redox
Table of Contents
1.
Safety Instructions
1.1
1.2
Warning Notices. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
General Safety Regulations . . . . . . . . . . . . . . . . . . . . . . .
2.
Product Description
2.1
2.2
2.3
2.4
2.5
2.6
2.7
2.7.1
2.7.2
2.7.3
Description of the System . . . . . . . . . . . . . . . . . . . . . . . .
Technical Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Swansensor pH and Redox . . . . . . . . . . . . . . . . . . . . . . .
Swansensor pH AY and Redox AY . . . . . . . . . . . . . . . . .
Swansensor pH SI and Redox SI . . . . . . . . . . . . . . . . . .
Swansensor pH FL and Redox FL. . . . . . . . . . . . . . . . . .
Software. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Menu Navigation Keys . . . . . . . . . . . . . . . . . . . . . . . . .
Software Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.
Installation
3.1
3.2
3.3
3.4
3.5
3.5.1
3.5.2
3.6
3.7
3.8
3.9
3.10
3.10.1
3.10.2
Installation Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Mounting of the Instrument . . . . . . . . . . . . . . . . . . . . . . .
Connection Schemes. . . . . . . . . . . . . . . . . . . . . . . . . . . .
Power Supply. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Relay Contacts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Alarm Relay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Relay Contacts 1 and 2 . . . . . . . . . . . . . . . . . . . . . . . .
Signal Outputs1 and 2 . . . . . . . . . . . . . . . . . . . . . . . . . . .
Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Installation of Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . .
Interfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
RS232 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Profibus (optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . .
A-96.250.441 / 060923
3
4
5
6
8
8
8
8
9
9
10
11
12
13
14
16
16
16
16
16
16
17
17
19
19
19
1
AMU pH-Redox
AMU pH-Redox
4.
Instrument Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . .
21
5.
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
21
6.
Maintenance (Quality assurance)
6.1
6.2
6.3
6.4
Maintenance Table. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Cleaning of Electrodes . . . . . . . . . . . . . . . . . . . . . . . . . . .
Longer stop of operation . . . . . . . . . . . . . . . . . . . . . . .
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8.
Trouble Shooting
8.1
8.2
Error Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Error List. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
A.
Appendix
A.1
A.2
Program Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Program List and Explanations . . . . . . . . . . . . . . . . . . . .
Default Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
B.
Material Safety Data Sheets
B.1
B.2
B.3
B.4
2
Buffer pH 4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Buffer pH 7 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Buffer pH 9 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Buffer ORP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
22
23
24
25
27
28
30
33
50
53
56
59
62
C.
RS232 Interface
C.1
C.2
C.3
C.4
C.4.1
C.4.2
C.4.3
C.5
Connection of the Communication Cable. . . . . . . . . . . . .
Configuration of the AMU Transmitter . . . . . . . . . . . . . . .
HyperTerminal Configuration on the PC . . . . . . . . . . . . .
Functions of the AMU RS232 Interface . . . . . . . . . . . . . .
Logger Download . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Log download . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Clear Logger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Firmware Upload . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
66
67
68
72
72
73
75
76
D.
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
78
A-96.250.441 / 060923
AMU pH-Redox
Safety Instructions
1.
General
Target
audience
Operator’s
Manual Location
1.1
Safety Instructions
The instructions included in this section explain the potential risks
associated with instrument operation and provide important safety
practices designed to minimize these risks.
If you carefully follow the information contained in this section, you
can protect yourself from hazards and create a safer work
environment.
More safety instructions are given throughout this manual, at the
respective locations where observation is most important.
Strictly follow all safety instructions in this publication.
Operator:
Qualified person who uses the equipment
for its intended purpose.
Instrument operation requires thorough knowledge of applications,
instrument functions and software program as well as all applicable
safety rules and regulations.
The AMU Operator’s Manual shall be kept in proximity of the
instrument.
Warning Notices
The symbols used for safety-related notices have the following
significance:
WARNING
Symbols
WARNING
Generally, the triangular warning symbol indicates the possibility of
personal injury or even loss of life if instructions are not followed,
e.g.
electrical shock hazard
ATTENTION
Symbols
ATTENTION
The attention symbol indicates the possibility of equipment
damage, malfunctions or incorrect process results if instructions
are not followed.
A-96.250.441 / 060923
3
AMU pH-Redox
Safety Instructions
1.2
General Safety Regulations
Legal
Requirements
The user is responsible for proper system operation according to all
local, state and federal laws that may apply. All precautions must
be followed to ensure safe operation of the instrument.
Spare Parts
and
Disposables
Use only official SWAN spare parts and disposables. If other parts
are used during the normal warranty period, the manufacturer’s
warranty is voided.
Modifications
Modifications and instrument upgrades shall only be carried out by
an authorized Service Technician. SWAN will not accept
responsibility for any claim resulting from unauthorized modification
or alteration.
Electrical
shock hazard
WARNING
4
If proper operation is no longer possible, the instrument must be
disconnected from all power lines, and measures must be taken to
prevent inadvertent operation.
Š
To prevent from electrical shock, always make sure that
the ground wire is connected.
Š
Service shall be performed by authorized personnel
only.
Š
Whenever electronic service is required, disconnect
instrument power and power of devices connected to
• relay 1,
• relay 2,
• alarm relay.
A-96.250.441 / 060923
AMU pH-Redox
Product Description
2.
Product Description
2.1
Description of the System
This instrument is applicable for the measurement of the pH
respectively ORP value in water.
Application
range
Measuring
Principle
Temperature
compensation
pH and redox (ORP) is measured in many different applications:
Š swimming pools
Š waste water
Š high purity water
Š power plants (steam, condensate)
Š industry
Depending on the application, different electrode types and flow
cells/fittings are needed.
pH: The electrode consists of a pH sensitive glass electrode and a
reference electrode = measuring chain. Depending on the pH value
in the water, the mV value of the measuring chain changes. If both
electrodes are integrated into one shaft it is a combined electrode.
Reference electrodes are available with gel (Swansensor pH for
potable water and swimming pools, Swansensor pH AY for waste
water) or with liquid electrolyte (Swansensor pH SI for power plant
cycle, Swansensor pH FL for high purity water).
ORP: The electrode consists of a platinum pin and a reference
electrode = measuring chain. Depending on the ORP value in the
water, the mV value of the measuring chain changes. If both
electrodes are integrated into one shaft it is a combined electrode.
Reference electrodes are available with gel (Swansensor pH for
potable water and swimming pools, Swansensor pH AY for waste
water) or with liquid electrolyte (Swansensor ORP SI for power
plant cycle, Swansensor ORP FL for high purity water).
pH: The pH value depends on temperature. If the temperature of
the sample is not stable, we recommend to install a temperature
sensor. If no temperature sensor is chosen, please program the
temperature of the sample in the instrument.
For high purity water applications (power plant, high purity water):
Additionally to the Nernst compensation a non-linear solution
compensation or linear compensation with coefficient can be
chosen.
ORP: Temperature compensation is not necessary.
A-96.250.441 / 060923
5
AMU pH-Redox
Product Description
2.2
Technical Data
The AMU measuring and control transmitter is used for panel
installation. It has connections for a pH/ORP electrode and a
Pt1000 temperature probe, and for a digital sample flow meter.
Electronics case:
Noryl® resin
Protection degree:
IP54 (front)
Large backlit LC-Display,
75 x 45 mm
Dimensions:
96 x 96 x 119 mm (DIN 43700)
Weight:
0.45 kg
Ambient temperature: -10 to +50°C
Limit range of operation:25 to +65°C
Storage and transport: -30 to +85°C
Humidity:
10 to 90 % relative, non-condensing
Voltage:
85 - 265 VAC, 47 - 63 Hz, or 24 VDC ±15%
Power consumption:
max. 7 VA
Measuring range
Resolution
0.00 to 14.00 pH Resolution: 0.01 pH
-500 to 1500 mV Resolution: 1 mV
Automatic range switching.
Temperature
Input
Temperature
compensation
6
for Pt1000 type sensor (DIN class A)
Measuring range: -30 to +250 °C
Resolution: 0.1 °C
Š Nernst
Š Non-linear solution compensation for high purity water
Š Linear with coefficient for high purity water
A-96.250.441 / 060923
AMU pH-Redox
Product Description
Temperature
monitoring
Safety
features
Alarm Relay
Input
Relays
Signal
Outputs
Control
Function
Communication Interface
Serial
Interface
Alarm if the transmitter temperature is higher than +65 °C
or lower than 0 °C.
No data loss after power failure, all data is saved in non-volatile
memory. Over voltage protection of in- and outputs.
Galvanic separation of measuring inputs and signal outputs.
One potential free contact for summary alarm indication for
programmable alarm values and instrument faults.
Maximum load: 100 mA / 50 V
One input for potential-free contact. Programmable ‘HOLD’ or
‘OFF’ function.
Two potential-free contacts programmable as
• limit switches for measuring values
• controllers
• timer for system cleaning with automatic hold function
Maximum load: 100 mA / 50 V
Two programmable signal outputs for measured values
(freely scaleable, linear or bilinear) or as continuous control output
(control parameters programmable).
Current loop: 0/4 - 20 mA
Maximum burden: 510 Ω
Relays or current outputs programmable for 1 or 2 pulse dosing
pumps, solenoid valves or for one motor valve.
Programmable P, PI, PID or PD control parameters.
Optional RS485 interface with Fieldbus protocol PROFIBUS DP or
Modbus, galvanically separated.
RS232 for data logger download to PC with Microsoft
HyperTerminal and for Firmware updates.
A-96.250.441 / 060923
7
AMU pH-Redox
Product Description
2.3
Applications
Swansensor pH and Redox
Š
Š
Potable water
Swimming pools
Details see data sheet.
Operating
Conditions
0 to 50 °C
Max. pressure: < 2 bar
2.4
Applications
Swansensor pH AY and Redox AY
Š
Waste water
Details see data sheet.
Operating
Conditions
0 to 50 °C
Max. pressure: < 2 bar
2.5
Applications
Swansensor pH SI and Redox SI
Š
Power plant cycle
Details see data sheet.
Operating
Conditions
0 to 50 °C
Pressure-free
Flow speed: 5 to 10 l/h
2.6
Applications
Swansensor pH FL and Redox FL
Š
High purity water
Details see data sheet
Operating
Conditions
8
0 to 50 °C
Pressure-free
Flow speed: 5 to 10 l/h
A-96.250.441 / 060923
AMU pH-Redox
Product Description
2.7
Software
2.7.1 Display
Menu navigation keys
RUN
HOLD
normal operation
The Input is activated. The Signal Outputs are frozen.
OFF
The Input is activated; the Signal Outputs are set to 0
(0/4 mA)
error
blinking: fatal error
Status
R1
R2
sample
flow
8 l/h
Time
7.43 pΗ
25 °C
sample
temperature
Relay Status
upper / lower limit reached
upper / lower limit not yet reached
control upw. / dnw.: no action
control upw. / dnw.: active, dark bar indicates control intensity
motor valve: valve closed
motor valve: open, dark bar indicates approximate position
timer
timer: timing active (hand rotating)
A-96.250.441 / 060923
9
AMU pH-Redox
Product Description
2.7.2 Menu Navigation Keys
Keys
Exit
to exit a menu or command (without accepting)
to move back to the previous menu level
to move DOWN in a menu list and
to decrease digits
to move UP in a menu list and
to increase digits
Enter
Program
Navigation
RUN
R1
R2
to open a selected sub-menu
to confirm an entry
14:10:45
x pH
25.0°C
Enter
Exit
Main Menu
Messages
Diagnostics
Maintenance
Operation
Installation
1
For a detailed listing of the menu see Program Overview, p. 30
10
A-96.250.441 / 060923
AMU pH-Redox
Product Description
2.7.3 Software Structure
Main Menu
Messages
Diagnostics
Maintenance
Operation
Installation
1
Messages
Pending Errors
Message List
1.1
Diagnostics
Identification
Sensors
Sample
I/O State
Interface
2.1
3.1
Maintenance
Calibration
Simulation
Set Time 02.04.05 16:49:30
A-96.250.441 / 060923
Operation
Sensors
Relay Contacts
Logger
4.1
Installation
Sensors
Signal Outputs
Relay Contacts
Miscellaneous
Interface
5.1
Menu Messages 1
Reveals pending errors as
well as an event history
(time and state of events
that have occured at an
earlier point in time).
Menu Diagnostics 2
Provides user relevant
instrument and sample
data.
Menu Maintenance 3
For instrument calibration,
relay and signal output
simulation and to set the
instrument time. It is used
by the service personnel.
Menu Operation 4
Subset of menu
Installation 5, but processrelated. User relevant
parameters that might
need to be modified during
daily routine. Normally
password protected and
used by the processoperator.
Menu Installation 5
For initial instrument set up
by SWAN authorized
person, to set all
instrument parameters.
Normally password
protected.
11
AMU pH-Redox
Installation
12
3.
Installation
3.1
Installation Checklist
Check
Instrument’s specification must conform to your
power ratings.
Installation
Mount the instrument in vertical position. Display
should be at eye level.
Electrical
Wiring
Connect all external devices like limit switches and
current loops.
Connect power cord; do NOT switch on power yet!
Flow cell /
fitting
Install flow cell / fitting.
Flow cell: Ensure outlet is pressure-free.
Electrode
Remove protective cap. Install in flow cell/fitting.
Remove plug cap. Connect the sensor(s) to the
AMU transmitter. Store sensor caps for interruption of operation.
QV-flow: Install reference in the hole near the
panel.
Fitting: Ensure fitting is water-tight.
Power up
Flow cell: Turn on sample flow and wait until it is
completely filled. Switch on power.
Instrument
setup
Flow cell: Adjust sample flow.
Program all parameters for sensor and external
devices
(interface, recorders, etc.). Program all parameters for instrument operation (limits, alarms).
Run-in period
Let the instrument run continuously for 1 h.
pH electrode
calibration
Calibrate pH electrode.
ORP in potable
/waste water
Calibrate ORP electrode.
ORP in high
purity water:
Process
calibration
Let the electrode run in for at least 24 h.
Make a valid manual measurement with a suitable
ORP electrode. Compare this value to the value,
indicated by the AMU. If necessary, correct.
A-96.250.441 / 060923
AMU pH-Redox
Installation
3.2
Mounting of the Instrument
Standard housing according to DIN 43700: 96 x 96 x 119 mm
89.8
96
89.8
96
112
7
A-96.250.441 / 060923
13
AMU pH-Redox
Installation
3.3
Connection Schemes
Use only the terminals shown in this diagram, and only for the
mentioned purpose. Use of any other terminals will cause short
circuits with possible damage to material and personnel.
Rear view
14
A-96.250.441 / 060923
AMU pH-Redox
Installation
SENSOR
LINE
RS232
RELAYS - INPUT - SIGNAL OUT
OPTION
PROFIBUS or MODBUS
SHIELD
The type plate for the
product identification is
on the back side of the
AMU transmitter.
A-96.250.441 / 060923
11.431.102
pH / Redox
15
AMU pH-Redox
Installation
3.4
Power Supply
Power Supply Connector
Neutral
Power
3.5
Relay Contacts
3.5.1 Alarm Relay
Note: Max. load 100 mA / 50 V
Terminal 5/6
3.5.2 Relay Contacts 1 and 2
Note: Max. load 100 mA / 50 V
Relay 1: Terminal 1/2
Relay 2: Terminal 3/4
3.6
Signal Outputs1 and 2
Note:
Max. burden 510 Ω
If signals are sent to two different receivers,
use signal isolator (loop isolator).
Signal Output 1: Terminal 13 (+) and 12 (-)
Signal Output 2: Terminal 14 (+) and 12 (-)
Current loop: 0 - 20 mA resp. 4 - 20 mA
3.7
Input
Use only potential-free (dry contacts).
Terminal: 11/10
16
A-96.250.441 / 060923
AMU pH-Redox
Installation
3.8
Sensor
Connect the sensor as follows:
pH / ORP Sensor:
Terminal 21: shielded line
Terminal 22: shield
External reference sensor:
Terminal 23: shielded line
Terminal 24: shield
Temperature sensor Pt1000:
Terminal 20: shielded line
Terminal 19: shield
3.9
Installation of Sensors
Flow cell Cautiously pull off rubber cap from pH/ORP electrode tip. Fragile!
M-Flow Handle with care!
Insert pH/ORP electrode into one of the PG threads and fasten handtight.
Take off plug cap. Connect to sensor cable.
Store the plug and rubber cap for interruptions of operation.
The temperature sensor is already mounted and connected to the
cable.
Flow cell QV- We recommend to mount the pH/ORP SI respectively Reference FL
Flow electrode in the insert near the panel. Flow cell, view from above
(simplified):
pH SI / Reference FL sensor
Temperature
sensor
pH FL sensor
All sensors are fragile! Handle with care!
Electrode Cautiously pull off rubber cap from pH/ORP electrode tip.
Insert pH/ORP electrode into the opening (see above) and fasten
hand-tight.
Take off plug cap. Connect to sensor cable.
Store the plug and rubber cap for interruptions of operation.
A-96.250.441 / 060923
17
AMU pH-Redox
Installation
Temp. sensor The sensor is already mounted and connected to the cable.
Reference The reference electrode (Swansensor pH SI, Swansensor ORP SI,
electrode Swansensor FL) is delivered with an electrolyte tube, which is already
mounted and filled. The sealed electrolyte bottle is also included in
delivery.
- Cautiusly take of the rubber cap from electrode tip. Make sure that the
black cap at the tip of the electrode is fastened hand-tight.
- The reference electrode has to be free of air bubbles: Hold the
electrode, tip downwards, in one hand and hold the electrolyte tube
with the other hand straight above the the electrode. Shake the
electrode slightly in an angular position to release air, which might be
inside. Release air bubbles in the electrolyte tube by knocking on it.
- Insert the electrode into the flow cell and fasten hand-tight.
- Cut off the tip of the electrolyte bottle. Take the plug out of the
electrolyte tube. Put the electrolyte bottle firmly on the tube.
- Mount the electrolyte bottle in the click holder (see 2.3 Instrument
overview, p. 11).
- Puncture the electrolyte bottle to allow pressure equilibration.
- Take off plug cap. Connect to sensor cable. Store caps.
UNIDIP fitting Please refer to the manual of the UNIDIP fitting how to install the pH/
ORP electrode and the temperature sensor in detail. Ensure that all surfaces and o-rings are clean. If mounted correctly, the UNIDIP fitting is
tight. No water is allowed inside the UNIDIP fitting. Otherwise the cables
corrode. If you use an UNIDIP Rinsematic, make sure the nozzles face
the pH/ORP sensor.
Temperature sensor and electrodes: Fragile! Handle with care!
18
A-96.250.441 / 060923
AMU pH-Redox
Installation
3.10
Interfaces
3.10.1 RS232
The interface RS232 is on the backside of
the AMU transmitter.
For connecting the communication cable and the configuration
(using RS232 for data logger download to PC and for firmware
updates) see RS232 Interface, p. 66
3.10.2 Profibus (optional)
The interface RS485,
galvanically separated.
To connect several instruments by means of a network, use a freecable end PROFIBUS cable. See picture below.
A-96.250.441 / 060923
19
AMU pH-Redox
Installation
termination
switch
Please buy a free-cable end PROFIBUS cable from your
local supplier.
NOTE: The color of the wires might be different. Before connecting
it to the transmitter, please check the corresponding data sheet.
The PROFIBUS connections have to be short-circuited.
The shield has the function to change or replace the transmitter
without inactivating the bus connection.
NOTE: Terminate bus by means of ON/OFF switch on the RS485
interface only. If only one instrument is installed, or on the last
instrument in a bus, the termination switch must be ON.
20
A-96.250.441 / 060923
AMU pH-Redox
Instrument Setup
4.
Sensor
Establish
Sample Flow
Instrument Setup
Check if:
Š Power is available
Š Sensor is installed and connected correctly.
For connecting the sensor to the transmitter, see Swansensor pH
and Redox, p. 8
1
Check if sample is available.
2
Assure pressure-free sample outflow.
3
Open sample flow. (Wait until flow cell is completely filled.)
4
If applicable, check flow cell for leaks.
5
Connect the instrument to power.
6
Adjust sample flow.
7
Set all necessary parameters in Installation, Sensors 5.1.
- Flow: Set if you use the Q-flow cell (with Swan flow meter)
- Parameter: Set measuring parameter pH or ORP
Sensor check: No.
- Temperature: Set if a temperature sensor has been connected.
If no, program the temperature of the sample (default temp.)
If yes, you can set a temp. correction, if necessary.
- Temperature compensation: Only for pH, not for ORP.
- Standard solution(s): Program the buffer values (pH buffer table)
or the ORP calibration solution.
8 Program all parameters for external devices (interface,
recorders, etc.). Program all parameters for instrument
operation (limits, alarms, etc.)
9
5.
Let run in for 1 h. Correct if necessary, or calibrate. See
Installation check list.
Operation
To have a general survey of the menu 4: Operation, see Program
Overview, p. 30. For explanations see Program list and
explanations.
You can modify:
Š The Filter Time Constant and the time for Hold after Calibration
for the sensor
Š The Settings for the Relay Contacts and the Input.
A-96.250.441 / 060923
21
AMU pH-Redox
Maintenance (Quality assurance)
6.
Maintenance (Quality assurance)
This section describes the activities intended to retain the
instrument in, or to restore it to a state in which it maintains the
required or specified performance.
6.1
Maintenance Table
Swansensor pH / Swansensor Redox (ORP)
Twice per month
Calibrate electrode. Ensure buffers are not expired. If
necessary, replace electrode.
Yearly
If necessary, clean electrode.
Swansensor pH AY / Swansensor Redox (ORP) AY
Twice per month
Clean electrode.
Monthly
Calibrate electrode. Ensure buffers are not expired. If
necessary, replace electrode.
Swansensor pH SI or FL
Weekly
Check level in electrolyte bottle. If necessary, change bottle.
Monthly
Calibrate electrode.
All 3 months
Open cap of reference electrode slightly and let flow out
5 ml of electrolyte. Fasten cap hand-tight.
Swansensor Redox (ORP) SI or FL
22
Weekly
Check level in electrolyte bottle. If necessary, change bottle.
Monthly
If necessary, correct electrode.
All 3 months
Open cap of reference electrode slightly and let flow out
5 ml of electrolyte. Fasten cap hand-tight.
A-96.250.441 / 060923
AMU pH-Redox
Maintenance (Quality assurance)
6.2
Cleaning of Electrodes
Cleaning
Swansensor Take the electrode out of the flow cell. If necessary wipe the electrode
pH and pH AY shaft and the green tip cautiously with a soft, clean, and damp paper
tissue.
Remove grease with a tissue moistured with alcohol (caution!
Inflammable!).
If the electrode is very dirty, put it into 1% diluted hydrochloric acid for
roughly 1 min. Caution! Etching!!
Afterwards rinse the electrode thoroughly with clean water.
Remount in the flow cell again.
Let the electrode run-in for 1 h before the first calibration.
Swansensor Take the electrode out of the flow cell. If necessary, wipe off dirt
ORP and ORP cautiously with a soft, clean, and damp paper tissue.
AY Dull platinum surfaces indicate a contamination. Clean the electrode by
putting it into the calibration solution or in 1% diluted hydrochloric acid
for some minutes. Caution! Both solutions are etching!
Afterwards rinse the electrode thoroughly with clean water.
Remount in the flow cell again.
Let the electrode run-in for 1 h before the first calibration.
Swansensor If necessary, wipe the green tip of the electrode cautiously with a soft,
pH SI and pH clean, and damp paper tissue. If necessary, remove grease with a
FL tissue, moistured with alcolhol (Caution! Inflammable!). Do not insert
into acids !!!
Reference electrode: Open the cap (diaphragm) at the tip of the
electrode slightly and let flow out some ml of electrolyte. Caution! KCL
is corrosive! Fasten cap hand-tight again.
Afterwards rinse the electrode thoroughly with high purity water.
Remount in the flow cell again.
Let the electrode run-in for 1 h before the first calibration.
Swansensor If necessary, wipe the tip of the electrode cautiously with a soft, clean,
ORP SI and and damp paper tissue. Do not insert into acids !!!
ORP FL Reference electrode: Open the cap (diaphragm) at the tip of the
electrode slightly and let flow out some ml of electrolyte. Caution! KCL
is corrosive! Fasten cap hand-tight again.
Afterwards rinse the electrode thoroughly with high purtiy water.
Remount in the flow cell again.
Let the electrode run-in for at least 24 h before the first correction.
A-96.250.441 / 060923
23
AMU pH-Redox
Maintenance (Quality assurance)
6.3 Longer stop of operation
Š
Š
Š
24
Stop sample flow.
Shut off power of the instrument.
Take off the cable of the electrode(s) and put the plug cap on it.
Take the electrode(s) out of the flow cell/fitting and rinse with clean
water. Fill 3.5 molar KCl (if not available: water) into the rubber cap
and place it on the tip of the electrode. Never store combined pH/
ORP electrodes, respectively reference electrode, dry!
Store the electrode(s) with tip pointing downwards in a frostprotected room.
A-96.250.441 / 060923
AMU pH-Redox
Maintenance (Quality assurance)
6.4
Calibration
Process pH / Use a suitable, high quality electrode and instrument to determine the
ORP sample value. Ensure, the electrode is calibrated correctly! Enter
this value under Process value in menu Process pH/ORP.
3.1
Maintenance
Calibration
Simulation
Set Time 01.01.05 16:30:00
Enter
Enter
Enter
Process pH
Calibration
Process pH
Standard pH
Enter
3.1.1
3.1.1.1
Current Value
Offset
7.78 pH
x mV
Process Value
Save
7.78 pH
<Enter>
Process pH
Current Value
Offset
7.78 pH
x mV
Process Value
Save
7.70 pH
<Enter>
Enter correct value with arrow keys.
3.1.1.1
Process pH
Current Value
Offset
7.70 pH
y mV
Process Value
Save
7.70 pH
<Enter>
Process pH
Current Value
Offset
3.1.1.1
press
Enter
3.1.1.1
7.70 pH
y mV
Calibration successful
Possible er- Offset error:
ror message Last calibration wrong.
Electrode old or dirty.
Cable wet.
A-96.250.441 / 060923
25
AMU pH-Redox
Maintenance (Quality assurance)
Standard pH
3.1
Maintenance
Calibration
Enter
Simulation
Set
Time 01.06.04 16:30:00
Enter
Enter
Calibration
Fill System
Process pH
Standard pH
3.1.3
Enter
Remove the electrode
(and, if applicable, the
temp. sensor) from the
flow cell.
Now follow the
instructions on the
display.
Standard 1 and 2 are
set in 5.1.5
Buffer/Cal. solutions have to be clean. Do not use if expired. Always
rinse electrode before dipping it into the solution.
Flow cell QV- The flow cell has a removable calibration vessel. By a short left turning
fow, high purity movement, this vessel can be unlocked and taken away. After rinsing,
water you can fill the buffer directly into the vessel.
To refasten the vessel again, turn it so that the groove is in front. Insert
the vessel in the Bajonett fastening. Turn to the right till stop.
cog
groove
Go to Menu 3.1.3. The instrument guides you through the complete
calibration process. If you have completed the required action press
<Enter> to proceed.
• Rinse and dry electrode and put it into standard 1
• Standard 1, current value (Progress is shown). End for ORP.
• Rinse and dry electrode and put it in standard 2
• Standard 2, current value (Progress is shown)
• Rinse and dry electrode and put it into the flow cell
Possible er- Offset error or Slope error:
ror message Old / dirty / wrong buffer solutions.
Electrode old or dirty.
Cable wet.
26
A-96.250.441 / 060923
AMU pH-Redox
Trouble Shooting
8.
Trouble Shooting
8.1
Error Messages
The menu 1 - Messages - reveals pending errors as well as the
message list with text, time and state of errors that have occured
before.
Error
Nonfatal error. The system continues to operate normally, except
for the defective part.
Fatal error
(blinking symbol)
If used for control and Relay 1, 2 as well as the Signal Outputs are
deactivated, i. e. any control is interrupted. The indicated measured
values are incorrect. Depending on the error (see Error List, p. 28)
the system may stop operating.
Pending Errors
Error Code
Sample Flow low
<Enter> to Acknowledge
1.1.5
E010
Errors remain in the Pending
Errors window until corrected,
even if acknowledged.
Acknowledging the error only
deactivates the Alarm Relay, i.e.
it switches off an alarm device if
available.
If an error is corrected, by intervention or by the system itself,
it is listed in the the Message List (Menu 1.2). The symbol
disappears.
The last 64 messages are stored for viewing. They are deleted if
SET Default Values (Menu 5.4.2) is carried out.
A-96.250.441 / 060923
27
AMU pH-Redox
Trouble Shooting
8.2
Error List
Bold and red error code: Fatal error
28
Error
Description
Corrective action
E001
Alarm high
- check process
- check electrode
- check progr. value in 5.3.1.1
E002
Alarm low
- check process
- check electrode
- check progr. value in 5.3.1.1
E007
Sample Temp.
high
- check sample temperature
- check progr. value in 5.3.1.4
E008
Sample Temp. low
- check sample temperature
- check progr. value in 5.3.1.4
E009
Sample Flow high
- check sample flow
- check progr. value in 5.3.1.3
E010
Sample Flow low
- establish sample flow
- clean instrument
- check progr. value in 5.3.1.3
E011
Temp. shorted
check wiring of temperature
sensor, check temp. sensor
E012
Temp. disconnected
check wiring of temperature
sensor, check temp. sensor
E013
Case Temp. high
- check case/environment
temperature
- check progr. value in 5.3.1.5
E014
Case Temp. low
- check case/environment
temperature
- check progr. value in 5.3.1.6
E017
Control time-out
check control device or programming in Installation, Relay
contact, Relay 1/2 5.3.2-3
A-96.250.441 / 060923
AMU pH-Redox
Trouble Shooting
Error
Description
Corrective action
E024
Input active
See Menu 5.3.4.4
(If Fault Yes is programmed in
menu 5.3.4)
E025
IC MK41T56
call service
E026
IC LM75
call service
E027
IC PCF8574
call service
E028
EEProm Microcon
call service
E029
EEProm Motherboard
call service
E030
EEprom Front-End
call service
E031
Calibration Rec Out
call service
E032
Wrong Front-End
call service
E033
Power-on
none, normal status
E034
Power-down
power on
A-96.250.441 / 060923
29
AMU pH-Redox
Appendix
A.
Appendix
A.1
Program Overview
Messages
(Main menu 1)
Pending errors, 1.1
Message list, 1.2
Identification, 2.1
Designation AMI pH/Redox, 2.1.1
Version 3.50,
Instrument, 2.1.3.1
2.1.2
Mother Board, 2.1.3.2
Factory test, 2.1.3
Front-end, 2.1.3.3
Years, days, hours
Operating time, 2.1.4 minutes, seconds,
Electrode, 2.2.1
Current value, pH or mV, 2.2.1.
Raw value, mV, 2.2.1.2
Cal. history, 2.2.1.5
Miscellaneous, 2.2.3
Case temp., °C, 2.2.3.1
Sensors, 2.2
Diagnostics
(Main menu 2)
Sample, 2.3
Sample ID, 2.3.1
Temperature °C, Pt1000, 2.3.2
Sample flow, l/h, Hz, 2.3.3
I/O State, 2.4
Alarm relay, 2.4.1
Relay 1/2, 2.4.2
Input
Signal output 1/2/3
Interface, 2.5
Protocol, 2.5.1
Baud rate, 2.5.2
(3. SA option)
(only with option RS485)
Menu Messages and Diagnostics are always accessible for everybody. No password protection. No settings can be modified.
30
A-96.250.441 / 060923
AMU pH-Redox
Appendix
Calibration, 3.1
Maintenance
Simulation, 3.2
(Main menu 3)
Process pH or ORP, 3.1.1
Standard pH or ORP, 3.1.2
Alarm Relay, 3.2.1
Relay 1/2, 3.2.2-3
Signal output 1/2, 3.2.4-5
Set time, 3.3
Sensors, 4.1
Operation
(Main menu 4)
Relay contacts, 4.2
Logger, 4.3
Filter time constant, 4.1.1
Hold after cal., 4.1.2
Alarm relay, 4.3.1
Alarm , 4.3.1.1
Relay 1/2, 4.3.2-3
acc. Installation
Input, 4.3.4
acc. Installation
acc. Inst.
Log Interval, 4.4.1
Clear logger, 4.4.2
Menu Maintenance is for service: Calibration, simulation of outputs and set time/
date. Please protect with password.
Menu Operation is for the user, allowing to set limits, alarm values, etc. The
presetting is done in the menu Installation (only for the System engineer). Please
protect with password.
A-96.250.441 / 060923
31
AMU pH-Redox
Appendix
Sensors, 5.1
Signal outputs, 5.2
Flow, 5.1.1
Type of sensor (pH/ORP), 5.1.2.1
Parameter, 5.1.2
Sensor check off/on, 5.1.2.2
Temp. sensor yes/no, 5.1.3.1
Temperature, 5.1.3
Temp. corr./Default Temp., 5.1.3.2
Temp. compensation, 5.1.40 (only for pH)
Standards pH, 5.1.5
Standard 1 (buffer table), 5.1.5.1
Standard ORP, 5.1.5 Standard 2 (buffer table), 5.1.5.2
Signal output 1, 5.2.1
Signal output 2, 5.2.2
Signal output 3, 5.2.3
(only if ordered)
pH or ORP
Parameter, 5.2.1.1
Temp.
Current loop, 5.2.1.2 flow
Function, 5.2.1.3
depends on Funct.
Scal./Contr.par.
Alarm high
Alarm low
Hysteresis
Alarm pH/ORP,
Delay
5.3.1.1
Sample flow, 5.3.1.3
Installation
Alarm relay, 5.3.1
(Main menu 5)
Sample Temp,
5.3.1.40
Flow al. high
Flow al. low
Temp al. high
Temp al. low
(only with temp. sensor)
Case Temp. high, 5.3.1.5
Case Temp. low, 5.3.1.6
Relay contacts, 5.3
Miscellaneous, 5.4
Interface, 5.5
Relay 1/2, 5.3.2-3
Function, 5.3.2.1
Parameter, 5.3.2.2
Limit/Set contr.
depends on Funct.
Input, 5.3.4
Active, 5.3.4.1
Signal output, 5.3.4.2
Outp./Contr., 5.3.4.3
Fault, 5.3.4.4
Delay, 5.3.4.5
Language, 5.4.1
Set Default, 5.4.2
Load firmware, 5.4.3
Password, 5.4.4
Sample ID, 5.4.5
Protocol, 5.5.1
Baud rate, 5.5.2
all menues
(only with option RS485)
Menu Installation: Defining assignment of all inputs and outputs, measuring
parameters, interface, passwords, etc. Menu for the system engineer. Password strongly recommended
32
A-96.250.441 / 060923
AMU pH-Redox
Appendix
A.2
Program List and Explanations
1 Messages
1.1 Pending errors
Provides list of active errors and status (active, acknowledged). If an
active error is acknowledged with the enter key, the alarm relay opens
again. Cleared errors are moved to the Message list.
1.2 Message
Shows error history: Error code, date and time of issue, and status
(active, acknowledged, cleared).
Maximal 65 error codes are memorized. Then the oldest error is
discharged to save the latest error (FIFO first in first out).
2 Diagnostics
In diagnostics mode, the values can only be viewed, not modified.
2.1 Identification
2.1.1 View the Type of instrument: Always AMU pH/Redox (ORP)
2.1.2 Informs about the Version of firmware of instrument, e.g. V3.50
2.1.3 Read the date of the factory test of the instrument, mother board,
front-end (sensor board) (can not be changed)
2.1.4 Reveals the Operating time in years, days, hours, minutes, seconds
(can not be changed)
2.2 Sensors
Review the diagnostic values of the photometer and pH electrode.
2.2.1 Sensor
Current value: Indicates the actual disinfectant value in pH respectively
(for ORP) mV
Raw value: Read the actual electrode signal in mV.
Calibration history: Informs about the diagnostic values of the electrode
of last calibrations:
pH: Offset in mV and slope in mV/pH.
Typical offset of pH electrode: < +/- 30 mV. Max. tolerated offset
<+/- 60 mV
Typical slope of pH electrode: 55-65 mV/pH unit.
A-96.250.441 / 060923
33
AMU pH-Redox
Appendix
Max. limits: 40-65 mV/pH.
ORP: Offset in mV. <+/- 200 mV
2.2.3 Miscellaneous
Case Temp.: Read the actual temperature of electronics housing.
2.3 Sample
2.3.1 Sample ID
Sample ID: Review the programmed code. The code is defined by the
user to identify the sample point in the plant.
Temp.: Only if temp. sensor has been installed and programmed.
Review the actual temperature of the sample in °C and in Ohm.
Sample flow: Only with digital sample flow meter. Reveals the actual
sample flow in l/h (litres per hour). Sample flow must be above 5 l/h.
2.4 I/O State
2.4.1- 2.4.2 Illustrates the actual status of all in and outputs:
Alarm relay: Open or closed
Relay 1 and Relay 2: Open or closed
Input: Open or closed
Signal output 1, 2, 3 (only if option is installed): Actual current in mA.
2.5 Interface
Only available if option interface has been installed.
Review programmed communication settings, set in Installation.
3 Maintenance
3.1 Calibration
In this menu, you can correct the measuring value or calibrate (offset
and slope) the electrode.
3.1.1 Process pH/ORP:
Correction of electrode. See Calibration, p. 25, for details.
3.1.2 Standard pH/ORP:
Calibration of electrode with the standard solution(s) programmed in
Installation 5.1.5. See Calibration, p. 25, for details.
3.2 Simulation
3.2.1-3.2.6 To simulate in/outputs:
Choose the required relay or signal output with the enter key and press
the up/down keys to change the value. After confirming the setting with
34
A-96.250.441 / 060923
AMU pH-Redox
Appendix
the enter key, the value is simulated by the relay/signal output.
As soon as you leave the simulation mode, the values will be reset.
At the absence of any key activity the instrument will leave the
simulation mode after 20 min. The simulated values will be reset.
Alarm relay: Open or closed.
Relay 1: Open or closed.
Relay 2: Open or closed.
Signal output 1: mA value.
Signal output 2: mA value.
Signal output 3: mA value (only if installed).
3.3 Set Time
To adjust date and time.
4 Operation
4.1 Sensors
4.1.1 Filter time constant:
Used to flatten noisy signals. The longer the filter time constant, the
slower the system reacts to changes in the measuring value.
Scalable values: 5-300 sec
4.1.2 Hold after calibration:
Delay to enable the instrument to stabilize again after calibration, so no
peaks show up on the signal output and no false limit warnings are
issued. During calibration and hold after calibration, the signal outputs
are frozen (held on last valid value), alarm values, limits are not active.
Scalable values: 5 - 6000 sec
4.2 Relay contacts
4.2.1 Alarm relay:
Here you reprogram alarm values high and low.
Alarm:
Alarm high: The measuring value must not rise above the alarm high
value, set in this menu. Otherwise the alarm relay closes and E001 is
issued. Scalable values: 0.00 - 14.00 pH, -500 to 1500 mV.
Alarm low: If the measuring value falls below the alarm low value, set in
this menu, the alarm relay closes and E002 is issued. Scalable values:
0.00 - 14.00 pH, -500 to 1500 mV.
A-96.250.441 / 060923
35
AMU pH-Redox
Appendix
Hysteresis: Within the hysteresis area, the relay does not open if the
alarm has been active. Used to prevent unnecessary relay switching
(damage), which occurs if the measuring value fluctuates around the
alarm value. Scalable values: 0.00 - 14.00 pH, -500 to 1500 mV.
Delay: The time, the activation of the relay is retarded after the
measuring value has risen above/fallen below the programmed alarm.
Scalable values: 0-28800 sec.
4.2.2 and Relay 1 and 2:
4.2.3 Here you reprogram the limit value or set-point of control, or redefine
the timer function. The available parameters depend on presetting in
Installation.
Limit upper/lower:
Set-point: If the measuring value rises above (upper limit) respectively
falls below (lower limit) the set-point, the relay closes.
Hysteresis: Within the hysteresis area, (below limit upper respectively
above limit lower), the relay does not open if the limit has been
exceeded. Used to prevent unnecessary relay switching (damage),
which occurs if the measuring value fluctuates around the limit value.
Delay: The time, the activation of the relay is retarded after the
measuring value has risen above/ fallen below the limit value.
Control: To maintain a certain set-point in the system.
Control upwards: Rises the measuring value to the set-point, assisted
by a control device. E.g. dosing alcaline solution with a pump to rise pH
value.
Control downwards: Lowers the measuring value with the help of a
control device, e.g. dosing acid with a pump into a swimming pool.
Set-point: Program the required value which should be maintained in
the system.
P-Band: Area, above (downwards control) or below (upwards control)
the set-point, where the dosing intensity is reduced from 100% to 0%,
to equal set-point without overshooting.
Timer: Relay closes as programmed, e.g. to issue the cleaning of the
instrument.
Mode: Set how often the relay should close. Choose between interval
or daily or weekly
If interval is chosen: Choose the time interval. Scalable values: 1 1440 min.
If daily is chosen: The contact closes once per day. Set the starting
time.
If weekly is chosen: Program the starting time and the active days. The
timer function starts on all days set to On.
Run Time: For all modes. Time, how long the relay closes. Scalable
values: 5 - 6000 s
36
A-96.250.441 / 060923
AMU pH-Redox
Appendix
Delay: For all modes. Signal outputs and control are on cont./hold/off
(as programmed below) during run time + delay time.
Scalable values: 0 - 6000 s
Signal output: Choose between continuous or hold or off.
Continuous: Signal outputs are operating continuously, the status
display indicates RUN
Hold: Signal outputs are frozen on the last valid value. Errors are not
issued, except fatal errors, which are always issued. The status display
indicates HOLD.
Off: Signal outputs go to 0 respectively 4 mA. Errors are not issued,
except fatal errors, which are always issued. The status display
indicates OFF.
Output (relay) / control (relay or signal output):
Choose between continuous or hold or off.
Continuous: Output/control are operating continuously.
Hold: Limit/control are frozen on last valid value. The state of the limit/
control is not updated according to the actual measuring value, as long
as the input is active (closed/opened) plus delay time. Reset time and
derivative time of the controller are ignored.
Off: Limits/control (signal output or relays 1+2) are interrupted. Errors
are issued.
Fieldbus
Contact can be activated/deactivated by fieldbus. No parameter setting
required.
4.2.4 Input:
Define the status of signal outputs, relay contacts, and the delay if the
input contact is activated. The setting depends on preprogramming in
Installation.
The status display shows the status of signal outputs.
Active: Choose between open or close.
Open: If the input is opened, the signal outputs and relays are on hold/
off/continuous as programmed below.
Close: If the input is closed, the signal outputs and relays are on hold/
off/continuous as programmed below.
Signal output: Choose between continuous or hold or off.
Continuous: Signal outputs are operating continuously, the status
display indicates RUN
Hold: Signal outputs are frozen on the last valid value. The
measurement is interrupted. Errors are not issued, except fatal errors,
which are always issued. The status display indicates HOLD.
Off: Signal outputs go to 0 respectively 4 mA. Errors are not issued,
except fatal errors, which are always issued. The status display
indicates OFF.
A-96.250.441 / 060923
37
AMU pH-Redox
Appendix
Output (relay) / control (relay or signal output):
Choose between continuous or hold or off.
Continuous: Output/control are operating continuously.
Hold: Limit/control are frozen on last valid value. The state of the limit/
control is not updated according to the actual measuring value, as long
as the input is active (closed/opened) plus delay time. Reset time and
derivative time of the controller are ignored.
Off: Limits/control (signal output or relays 1+2) are interrupted. Errors
are issued.
Fault: Choose between yes or no.
Fault no: No error message is issued when input is active (closed/
opened). Alarm relay does not close.
Fault yes: Error E024 Input activated is issued. Alarm relay closes.
Delay: The time the input deactivation (programmed above) is retarded
after opening/closing the input.
Scalable values: 0 - 6000 s
4.3 Logger
The instrument is equipped with an internal logger which can be down
loaded by interface, using the program Hyperterminal.
The logger can save approximately 1500 data records. One data
record consists of: Date, time, alarms, measuring value, temp., flow.
4.3.1 Log Interval: Choose the required logging interval.
The table below gives the max. time interval, the logger can save.
When the logger capacity is exhausted, the oldest data record is
discharged to save the new data record (FIFO first in first out).
Interval
1s
5s
1 min
5 min
10 min
30 min
1h
Time
25 min
2h
25 h
5d
10 d
31 d
62 d
4.3.2 Clear logger:
If confirmed with yes, the complete logger data is deleted. A new data
series is started.
38
A-96.250.441 / 060923
AMU pH-Redox
Appendix
5 Installation
5.1 Sensors
5.1.1 Flow: Define if the Swan flow meter is active or not. Choose between:
None or Q-flow (with flow meter).
5.1.2 Parameter:
Choose the type of sensor: pH or redox (ORP)
Sensor check: not yet implemented. Set to off.
5.1.3 Temperature:
Temp. sensor: Program if a temperature sensor Pt1000 has been
connected. Choose between no or yes.
If no, the Default Temp. is asked. Set the temp. of the sample.
If yes, you can set a Temp. correction, to compensate errors of
displayed value.
5.1.4 Temperature compensation: Only for pH.
Available settings: Nernst/non-linear/coefficient.
For potable water, waste water, swimming pools: Choose Nernst
High purity water: Additionally to Nernst compensation you can choose
between non-linear solution compensation or linear compensation with
coefficient (pH/°C).
5.1.5 Standards: Here you program the standard solution(s) for the
calibration of the electrode.
pH: Enter the values of your buffer into the buffer table (exact buffer
values for temperature 0° - 50°C)
ORP: Enter the value of the standard solution.
5.2 Signal outputs
5.2.1 - 5.2.3 Signal output 1, Signal output 2, Signal output 3: Signal output 3 is
only available if option 3rd signal output has been installed
Here you assign a certain measuring value, current loop and function
to the signal outputs.
5.2.1.1 - 5.2.3.1Parameter: Assign one of the measuring values to the signal output.
Available settings: Measuring value/temperature/flow
5.2.1.2 - 5.2.3.2Current loop: Choose the current loop of the signal output. Make sure
the connected device works with the same current loop.
Available settings: 0-20 mA or 4-20 mA.
5.2.1.3 - 5.2.3.3Function: Define if the signal output should represent the actual
measuring value or if it is used for control.
A-96.250.441 / 060923
39
AMU pH-Redox
Appendix
If it should represent the actual measuring value you can choose
between linear, bilinear or logarithmic scaling.
See explanations below.
Function lin/Shows one of the measuring values as current loop 0-20 mA/4-20 mA.
bilinear:Linear: Program range low and range high.
Range low: Choose the beginning of the scale. This value corresponds
to 0 respectively 4 mA. If you have chosen 0-20 mA it is 0 mA, if you
have chosen 4-20 mA, it is 4 mA. It is the lowest value represented by
the signal output.
Range high: Choose the end of the scale. This value corresponds to
20 mA. It is the highest value illustrated by the signal output.
Bilinear: The beginning or end of the selected range can be scaled in a
higher resolution, resulting in a semi-logarithmic presentation. Program
range low, range high, range centre.
Range low and high see above: Linear
Range centre: Fixed on 10 respectively 12 mA.
See also graphic below.
mA
20
o
res
bilinear
0-20 mA: 10
4-20 mA: 12
0
ol.
es
0
.00
6
.00
0
2
r
1
.00
l. 0
res
ol.
0
linear
0/4
2
3
4
5
6
7
8
unit
meas. value
FunctionThe scaling is logarithmic. Program range high and range low.
Log: Loga-Range low: Choose the beginning of the scale. This value corresponds
rithmic scaleto 0 respectively 4 mA. If you have chosen 0-20 mA it is 0 mA, if you
have chosen 4-20 mA, it is 4 mA.
Range high: Choose the end of the scale. This value corresponds to
20 mA.
ControlControl upwards or control downwards are available. For further
explanations about control and control parameters (set-point/p-band)
see 5.3.2.1
If the measuring value is out of the p-band, 20 mA are issued. Within
40
A-96.250.441 / 060923
AMU pH-Redox
Appendix
P-Band the mA value is reduced. If the measuring value equals setpoint, control is switched off, resulting in 0 mA (current loop 0-20 mA),
or 4 mA (current loop 4-20 mA.)
5.2.1.4x x no. depends on chosen function of signal output:
5.2.3.4x x = 0=lin, 1=bilin, 2=log, 3=control
5.2.x.40 - Scaling:
5.2.x.42 Choose range low, range high, range centre (if applicable). For
explanations see before: 5.2.1.3 Function.
Available values:
pH Range low/centre/high: -3.00 - 15.00 pH
ORP Range low/centre/high: -500 - 1500 mV
Temp. Range low/centre/high: -25.0 - 250.0 °C
Flow Range low/centre/high: 0 - 50 l/h
5.2.1.43 Control:
Program set-point and p-band. Please consult 5.3.2.1 for explanations
and ranges.
5.3 Relay contacts
5.3.1 Alarm relay:
Here you set the alarm values for the measuring value, temperature,
sample flow, case temperature.
5.3.1.1 Alarm measuring value: Define the measuring value, which should
issue an alarm high respectively low.
Alarm high: The measuring value must not rise above the alarm high
value, set in this menu. Otherwise, the alarm relay closes and E001 is
issued. Scalable values: -3.00 - 15.00 ppm, -500 - 1500 mV
Alarm low: If the measuring value falls below the set alarm value, the
alarm relay closes and E002 is issued.
Scalable values: -3.00 - 15.00 ppm, -500 - 1500 mV
Hysteresis disinfectant: Within the hysteresis area, the relay does not
open if the alarm has been active. Used to prevent unnecessary relay
switching (damage), if the measuring value fluctuates around the alarm
value. Scalable values: -3.00 - 15.00 ppm, -500 - 1500 mV
Delay: The time, the closing of the alarm relay is retarded after the
measuring value has risen above/fallen below the alarm value.
Scalable values: 0-28800 sec
See also graphic below.
A-96.250.441 / 060923
41
AMU pH-Redox
Appendix
Delay 20 sec
Hysteresis
0.05
0
Alarm high
0.40
red
asu
me
relay open
alarm off
relay closed
alarm on
relay open
alarm off
5.3.1.3 Alarm sample flow: Here you define at which sample flow a flow alarm
should be issued and if the alarm relay should close.
Alarm flow high: If measuring value rises above the programmed
value, E009 flow alarm high will be issued. Scalable values: 10-50l/h
Alarm flow low: If the measuring value falls below the programmed
value, E010 flow alarm low will be issued. Scalable values: 0-9 l/h
5.3.1.4 Alarm Sample Temperature: Only if temp. senor has been connected.
Alarm high: If the temperature rises above the programmed value,
E007 is issued. Scalable values: -25 - 250 °C
Alarm low: If the temperature falls below the programmed value, E008
is issued. Scalable values: -25 - 250 °C
5.3.1.5 Case temp. high: Set the alarm high value for temperature of
electronics housing. If the temperature rises above the programmed
value, E013 is issued. Scalable values: 30 to 75 °C
5.3.1.6 Case temp. low: Define the alarm low value for temperature of
electronics housing. If the measuring value falls below the
programmed value, E014 is issued. Scalable values: -10 to +20 °C
5.3.2 and Relay 1 and Relay 2
5.3.3 Program the function of the relays in detail.
5.3.2.1 and Function: Choose if the relay should
5.3.3.1 - close if the measuring value rises above/falls below the set limit,
- if it should drive a control device (pump, motor valve, etc.) to maintain
a certain set-point,
- if it is used as timer (e.g. adding sensor cleaning solution),
- if it should be activated by Profibus DP.
Available settings: Limit upper/lower/control upwards/control
downwards/timer/fieldbus
42
A-96.250.441 / 060923
AMU pH-Redox
Appendix
Limit upper/ See also graphic on page 63.
lower: Set-point: If the measuring value rises above (limit upper) respectively
falls below (limit lower) the set limit, the relay closes.
Hysteresis: Within the hysteresis area (below limit upper respectively
above limit lower), the relay does not open if the limit has been active.
Used to prevent unnecessary relay switching (damage), which occurs
if the measuring value fluctuates around the limit value.
Delay: The time, the closing of the relay is retarded after the measuring
value has risen above/fallen below the limit value.
Control To maintain a certain set-point in the system.
Control upwards: Rises the measuring value to the set-point, assisted
by a control device.
Control downwards: Lowers the measuring value to the set-point aided
by a control device.
Control parameters PID
Set-point: Program the required value which should be maintained
constant in the system.
P-Band: Area, above (downwards control) or below (upwards control)
the set-point, where the dosing intensity is reduced from 100% to 0%,
to equal set-point without overshooting.
See also graphic on page 53
PI controller: Reset time: The controller action increases the higher the
deviation from the set-point is and the longer it lasts.
PD controller: Derivative time: The controller action increases the
higher the deviation from the set-point is and the faster the measuring
value changes.
Control If the controller action (dosing intensity) is constantly over 90% during
time-out the control time-out (in min.), without the measuring value changing
into the direction of set-point, dosing will be stopped for safety reasons,
and an alarm is issued.
Control Time The control device is driven by switching on/off its power supply. Define
proportional a time interval (cycle time) as time base for control. To reduce the
dosing intensity in the p-band, the pump runs only part-time of the
cycle.
Cycle time: Defines in which interval the pump is switched on.
Response time: To avoid stressing the control device. Minimum time,
the dosing unit has to run to avoid damage, or that elapses from the
activation until the actual dosing starts.
See also graphic below.
Control by The pump is driven by adjusting the number of pulses. The relay
frequency closes/opens in the pulse frequency.
A-96.250.441 / 060923
43
AMU pH-Redox
Appendix
Frequency: Program the number of pulses/min. the pump can handle.
Within the p-band the pulse number is reduced = dosing is reduced.
Control Only one motor valve can be connected, because both relays are
motor valve needed to work the valve. Relay 1 opens, and relay 2 closes the valve.
Program the run time and neutral zone of your motor valve.
Run time: Time, the motor valve needs to close/open completely.
Neutral zone: Within the neutral zone the controller does not react to
deviations. Unnecessary switching (damage) is avoided.
Limit
Delay 20 sec
Hysteresis
pH/mV
m
Limit upper
pH/mV
ue
val
red
u
s
ea
0
relay open
alarm off
Time proportional control
relay open
alarm off
relay closed
alarm on
unit
set-point
p-band
1. cycle time = 60s 2. cycle = 60s
pump on
p-band
60 s
out of p-band
60 s
3. cycle = 60s
4. cycle = 60s
30 s
15 s
50% in p-band
75% in p-band
time (s)
Timer Relay closes as programmed, e.g. to issue cleaning of the instrument.
Mode: Set how often the relay should close.
Available settings: interval/daily/weekly
44
A-96.250.441 / 060923
AMU pH-Redox
Appendix
If interval is chosen: Program the time interval.
Scalable values: 1 - 1440 min.
If daily is chosen: The contact closes once per day. Choose the starting
time.
If weekly is chosen: Define the starting time and the active days. The
timer function starts on all days set to On.
Run Time: For all modes. Time how long the relay closes. Scalable
values: 5 - 6000 s
Delay: For all modes. Signal outputs and control are on cont/hold/off
(as programmed below) during run time + delay time.
Scalable values: 0 - 6000 s
Signal output: Choose between continuous or hold or off.
Continuous: Signal outputs are operating continuously, the status
display indicates RUN
Hold: Signal outputs are frozen on the last valid value. Errors are not
issued, except fatal errors, which are always issued. The status display
indicates HOLD.
Off: Signal outputs go to 0 respectively 4 mA. Errors are not issued,
except fatal errors, which are always issued. The status display
indicates OFF.
Outp.(relay)/contr.(relay or signal output):
Choose between continuous or hold or off
Continuous: Output/control are operating continuously.
Hold: Frozen on last valid value. The state of the limit/control is not
updated according to the measuring value, as long as the input is
closed (+ during delay time). Reset time and derivative time of the
controller are ignored.
Off: Limits/control (signal output or relay) are interrupted. Errors are
issued.
Fieldbus Contact can be activated/deactivated by fieldbus. No parameter setting
required.
5.3.2.2x and x no. depends on chosen function of the relay: 0=limit upper; 1=limit
5.3.3.2x lower; 2=control upwards; 3=control downwards; 4=timer
A-96.250.441 / 060923
45
AMU pH-Redox
Appendix
Overview:
5.3.2.1
5.3.3.1
0/1
Limit upper/lower
2/3
Control upw./
downw.
4
Timer
5.3.2.2
5.3.3.2
Parameter
Parameter
Mode
5.3.2.3
5.3.3.3
Limit value
Set-point
Interval/Start time/
Calender
5.3.2.4
5.3.3.4
Hysteresis
Control settings
Run time
5.3.2.5
5.3.3.5
Delay
Control parameters
Delay
5.3.2.6
5.3.3.6
Signal outputs
cont./hold/off
5.3.2.7
5.3.3.7
Output/control
cont./hold/off
5.3.2.2(0-3) and Parameter:
5.3.3.2(0-3) Assign one measuring value to the relay (with chosen function).
Available settings: Measuring value/temp./flow
5.3.2.24 and Mode of timer:
5.3.3.24 Choose how often the contact should close. Available settings: interval/
daily/weekly. For explanations see 5.3.2.1.
5.3.2.3xy and x no. depends on chosen function of relay: 0=limit upper; 1=limit lower;
5.3.3.3xy 2=control upwards; 3=control downwards; 4=timer
y no. depends on chosen parameter OR mode of timer:
For parameter: 0=measuring value; 1=Temp.; 2=sample flow
For mode of timer: 0=interval; 1=daily, 2=weekly
5.3.2.3x(0-3) Limit value or Set-point:
5.3.3.3x(0-3) Choose the required limit/set-point. For explanations see 5.3.2.1.
x=0,1,2,3 Scalable values depend on chosen parameter:
pH: 0.00 -14.00 pH
ORP: -500 - 1500 mV
5.3.2.34(0-2) Depends on chosen mode of timer: 0=interval; 1=daily; 2=weekly
5.3.3.34(0-2) 40: Interval in minutes
41: Start time: 00:00:00
46
A-96.250.441 / 060923
AMU pH-Redox
Appendix
42: Calendar: Starting time and Mo-Su (Off/On)
For explanations see 5.3.2.1
5.3.2.4xy and x no. depends on chosen function of relay = 0=limit upper; 1=limit
5.3.3.4xy and lower; 2=control upwards; 3=control downwards; 4=timer
5.3.2.5xy and y no. depends on chosen parameter.
5.3.3.5xy
For parameter: 0=measuring value; 1=Temp.; 2=sample flow
5.3.2.40y or 41y Limit: Hysteresis:
5.3.3.40y or 41y Explanations see 5.3.2.1.
Scalable values: pH: 0.00 -14.00 pH
ORP: -500 - 1500 mV
Limit Delay:
Explanations see 5.3.2.1. Scalable values: 0-28800 s
5.3.2.42y or 43y Control: Control settings:
5.3.3.42y or 43y Explanations see 5.3.2.1.
Actuator: Time prop.:
Scalable values: Cycle time: 10 - 600 s, Response time: 0 - 240 s
Actuator: Frequency:
Scalable values: Frequency: 20 - 300/min.
Actuator: Motor valve:
Scalable values: Run time: 5 - 300 s, Neutral zone: 1 - 20%
5.3.2.52y or 53y Control: Control parameters:
5.3.3.52y / 53y: Explanations see 5.3.2.1.
P-Band: Scalable values depend on setting in parameter:
0.00 - 14.00 pH or -500 - 1500 mV
Reset time: If 0 s is chosen the function is switched off.
Scalable values: 0 - 9000 s
Derivative time: If 0 s is chosen the function is switched off.
Scalable values: 0 - 9000 s
Control time-out: If 0 min. is chosen, the function is switched off.
Scalable values: 0 - 720 min.
5.3.2.44 or Timer: Run Time in seconds. Time the relay closes.
5.3.3.44 Scalable values: 5 - 6000 s
5.3.2.54 or Timer: Delay in sec. Time the signal outputs and output/control are on
5.3.3.54 cont./hold/off after deactivation of input.
Available values: 0 - 6000 s
5.3.2.6 or Timer: Signal outputs: Define the status of the signal outputs
5.3.3.6 (programmed as current loop for measuring value) during run time
(5.3.2.44) and delay (5.3.2.54): cont./hold/off.
For explanations see 5.3.2.1.
A-96.250.441 / 060923
47
AMU pH-Redox
Appendix
5.3.2.7 andTimer: Output (relay)/control: Define the status of relays (and signal
5.3.3.7output, programmed as control) during run time and delay: cont./hold/
off. For explanations see 5.3.2.1.
5.3.4 Input
Here you set the function of the input. The status display indicates the
status of signal outputs
5.3.4.1 Active: Choose between open or close.
Open: If the input is opened, the signal outputs, relays, and error relay
are on hold/off/continuous as programmed below.
Close: If the input is closed, the signal outputs, relays, and error relay
are on hold/off/continuous as programmed below.
5.3.4.2 Signal output: Choose between continuous or hold or off.
Continuous: Signal outputs are operating continuously, the status
display indicates RUN
Hold: Signal outputs are frozen on the last valid value. Errors are not
issued, except fatal errors, which are always issued. If a non-fatal error
is active when Input is closed, the error is frozen during Input closing
time+delay. The status display indicates HOLD.
Off: Signal outputs go to 0 respectively 4 mA. Errors are not issued,
except fatal errors, which are always issued. If a non-fatal error is
active when Input is closed, the error is frozen during Input closing
time+delay. The status display indicates OFF.
5.3.4.3 Outp.(relay)/contr.(relay or signal output):
Choose between continuous or hold or off
Continuous: Output/control are operating continuously.
Hold: Frozen on last valid value. The state of the limit/control is not
updated according to the measuring value, as long as the input is
closed (+ during delay time). Reset time and derivative time of the
controller are ignored.
Off: Limits/control (signal output or relay) are interrupted. Errors are
issued.
5.3.4.4 Fault: Choose between yes or no.
Fault no: No error message is issued when input is closed. Alarm relay
does not close.
Fault yes: Error E024 Input activated is issued. Alarm relay closes.
5.3.4.5 Delay: The time the input deactivation is retarded after opening/closing
the input.
Scalable values: 0 - 6000 s.
48
A-96.250.441 / 060923
AMU pH-Redox
Appendix
5.4 Miscellaneous
5.4.1 Language:
Set the required language.
Available settings: Deutsch/english/français/español, default: English
5.4.2 Set default:
If you set default, all programmed values will be set back to factory
default values. See Default Values, p. 50.
Available settings: No/calibration/in parts/completely.
Calibration: Sets calibration values back to default. All other values are
kept.
In parts: Communication parameters are kept. All other values are set
back to default values.
Completely: Sets back all values including communication parameters.
5.4.3 Load firmware:
Only for service personnel. For update of firmware. Not to be done by
the user. Always keep No.
5.4.4 Password
You can program passwords for all menus, except Diagnostics:
Messages, Maintenance, Operation, Installation
The passwords are 4-digit codes. If 0000 is programmed the function is
switched off and no password is asked. If you program a different code,
it will be asked, whenever you want to enter the menu.
Every menu point may have a different password.
Please make sure you do not forget the passwords. If you do, your
dealer can help you.
5.4.5 Sample ID:
Here you can set a code to identify the sample point in the plant. The
code is defined by the user.
5.5 Interface
5.5.1 Protocol:
Choose the required protocol:
Available settings: Profibus, Modbus RTU, SwanBus, Hyperterminal,
Web Server
Please see separate manual how to perform a logger download by
Hyperterminal.
5.5.2x x no. depends on chosen protocol in 5.5.1.
0=Profibus; 1=Modbus RTU; 2=SwanBus; 3=HyperTerminal;
4=WebServer
5.5.20 Profibus Device Address: Address of instrument
A-96.250.441 / 060923
49
AMU pH-Redox
Appendix
5.5.21 Modbus Device Address: Program address of instrument
5.5.22 SwanBus Device Address: Program address of instrument
5.5.23 Hyper Terminal Baud rate: Select transmission speed in baud.
5.2.24 Web Server Device address: Program address of instrument
5.5.3x x no. depends on chosen protocol in 5.5.1.
0=Profibus; 1=Modbus RTU; 2=SwanBus; 3=HyperTerminal;
4=WebServer
5.5.30 Profibus ID No.: Choose between Manufacturer/Analyzer/Multivariable
5.5.31 Modbus Baud rate: Select transmission speed in baud.
5.5.32 SwanBus Baud rate: Select transmission speed in baud.
5.5.4x x no. depends on chosen protocol in 5.5.1.
0=Profibus; 1=Modbus RTU; 2=SwanBus; 3=HyperTerminal;
4=WebServer
5.5.40 Profibus Local operation: Disabled/Enabled
5.5.41 Modbus Parity: None/Even/Odd
Default Values
Operation:
Sensors:Filter time constant: 30 s
Hold after calibration: 300 s
Signal output same as in Installation
1-3:
Relay 1, Relay same as in Installation
2, Input:
Logger Interval: 30 min.
Installation:
Sensor: Flow: None
Parameter: Type of sensor: pH
Parameter Sensor check:off
Temperature sensor: No
Temperature sensor no: Default temp.: 25 °C
Temperature sensor yes: Temp. Correction 0.00 °C
50
A-96.250.441 / 060923
AMU pH-Redox
Appendix
pH standard 1 / 2: Buffer table
0°C: 7.13 / 9.24 pH
5°C: 7.07 / 9.16
10°C: 7.05 / 9.11
15°C: 7.02 / 9.05
20°C: 7.00 / 9.00
25°C: 6.98 / 8.95
30°C: 6.97 / 8.91
35°C: 6.96 / 8.88
40°C: 6.95 / 8.85
50°C: 6.95 / 8.79
Standard Redox (ORP): 475 mV
Signal outputs: Parameter: measuring value
Current loop: 4-20 mA
Function: lin
Scaling:
Range low: 0.00 pH
Range high: 14.00 pH
Salient point: 7.00 pH
Salient point is always in the middle= switched off (lin scale).
Temperature range low: 0.0 °C
Temperature range high: 50.0 °C
Flow range low: 0 l/h
Flow range high: 50 l/h
Alarm relay:
pH alarm high: 15.00 pH
pH alarm low: -3.00 pH
Hysteresis: 0.10 pH
Delay: 5 s
Sample flow alarm high: 20 l/h
Sample flow alarm low: 5 l/h
Sample temp. alarm high: 55 °C
Sample temp. alarm low: 5 °C
Case temp. high: 65 °C
Case temp. low: 0 °C
Relay1/2 Function: limit upper
Parameter: Relay 1: measuring value, Relay 2: measuring value
Limit: Relay 1: 14.00 pH, Relay 2: 14.00 pH
Limit Hysteresis: 0.1 pH
Limit Delay: 30 s
Control: Actuator: Frequency
Control: Frequency: Frequency: 120/min.
Control: Time prop.: Cycle time: 60s, Response time: 10 s
Control: Motor valve: Run time: 60s, Neutral zone: 5%
A-96.250.441 / 060923
51
AMU pH-Redox
Appendix
Control: Control parameter P-band: 0.10 pH
Control: Control parameter Reset time: 0 s
Control: Control parameter Derivative time: 0 s
Control: Control parameter Control time-out: 0 min.
Timer: Mode: Interval: 1 min.
Timer: Mode: daily/weekly: Starting time: 00.00.00
Timer: Run time: 10s
Timer: Delay: 5s
Timer: Signal output: cont.
Timer: Output/Control: cont.
Input Active: when closed
Signal output: hold
Output/control: off
Fault: no
Delay: 10 s
Miscellaneous Language: English
Set default: no
Load firmware: no
Password: for all modes 0000
Sample ID: -Interface
52
Protocol: Hyperterminal
A-96.250.441 / 060923
AMU pH-Redox
Material Safety Data Sheets
B.
Material Safety Data Sheets
B.1
Buffer pH 4
According to EC Directive 91/155/EWG
1
Identification of the substance/preparation and of the company/undertaking
Identification of the product:
Catalogue No.:
A-85.112.300, A-85.112.500, A-85.112.700
Product name:
Calibration Solution pH 4
Manufacturer/supplier identification:
Company:
SWAN Analytische Instrumente AG, CH - 8616 Riedikon, Switzerland
Tel.: +41 1 943 63 00, Fax: +41 1 943 63 01
2
Composition/information on ingredients
Synonyms:
CAS-No.:
Molar mass:
Molecular formula:
EINECS-No.:
Potassium Hydrogen Phthalate, means of preservation and color
877-24-7 (Potassium Hydrogen Phthalate, not diluted)
204.23 g/mol (Potassium Hydrogen Phthalate, not diluted)
C8H5KO4 (Potassium Hydrogen Phthalate, not diluted)
2128894 (Potassium Hydrogen Phthalate, not diluted)
Hazardous ingredients:
Name according to EC Directives: None
3
Hazards identification
No hazardous product according to the directive 67/548/EWG.
4
First aid measures
After inhalation:
After skin contact:
After eye contact:
After ingestion:
5
-wash off with plenty of water.
rinse out with plenty of water with the eyelid held wide open.
make victim drink plenty of water, summon doctor if feeling unwell.
Fire-fighting measures
Suitable extinguishing media:
6
--
Special risks:
--
Other information:
non-combustible
Accidental release measures:
Procedures for cleaning / absorption:
Dilute with water and take-up with liquid-absorbent material.
Environmental-protection measures:
None
A-96.250.441 / 060923
53
AMU pH-Redox
Material Safety Data Sheets
7
8
Handling and storage:
Handling:
no further requirements
Storage:
tightly closed. Dry. In a well-ventilated place. At room temperature . No
public product. Keep out of children's reach.
Exposure controls/personal protection
Personal protective equipment:
Respiratory protection:
not required.
Eye protection:
not required.
Hand protection:
not required
Industrial hygiene:
Contact with eye or skin: Wash with plenty of water.
9
10
11
Physical and chemical properties
Form:
Color:
Odor:
liquid
reddish
odorless
pH value:
(20 oC)
4
Melting temperature:
Boiling temperature:
not available
near 100 oC
Ignition temperature:
Flash point:
not available
not available
Explosion limits:
lower
upper
not available
not available
Density:
(20 oC)
1.1 g/cm3
Solubility
in water:
(20 oC)
very good and completely soluble
Stability and reactivity
Conditions to be avoided:
--
Substances to be avoided:
--
Hazardous decomposition products:
none
Toxicological information
Acute toxicity:
No toxicological data available
Further toxicological information:
After skin contact:
-After eye contact:
-After ingestion:
--
54
A-96.250.441 / 060923
AMU pH-Redox
Material Safety Data Sheets
12
Ecological information
Ecotoxic effects:
-Further ecologic data:
None
13
Disposal considerations
Product:
There are no uniform EC regulations for the disposal of chemicals or residues. Chemical residues
generally count as special waste. The disposal of the latter is regulated in the EC member countries
through corresponding laws and regulations, and in Switzerland also by the cantons. We recommend
that you contact either the authorities in charge or approved waste disposal companies which will
advise you on how to dispose of special waste.
Packaging:
Disposal in compliance with official regulations. Handle contaminated packaging in the same way as the
substance itself. If not officially specified differently, non-contaminated packaging may be treated like
household waste or recycled.
14
Transport information
Not subject to transport regulations.
15
Regulatory information
Labeling according to EG Directives
Symbol:
-Name:
-R-phrases:
--
S-phrases:
--
Other national regulations:
Swiss toxic class:
None
16
Other information
Date of issue:
04.03.2005
The information contained herein is based on the present state of our knowledge. It characterizes the
product with regard to the appropriate safety precautions. It does not represent a guarantee of the
properties of the product.
A-96.250.441 / 060923
55
AMU pH-Redox
Material Safety Data Sheets
B.2 Buffer pH 7
1
Identification of the substance/preparation and of the company/undertaking
Identification of the product:
Catalogue No.:
85-113300 or 85-113350
Product name:
Calibration solution pH 7
Use of substance/preparation:
Buffer solution, no public product, only for laboratory use
Manufacturer/supplier identification:
Company:
Swan Analytical Instruments AG, CH - 8616 Riedikon
Tel.: 01/943 63 00, Fax: 01/943 63 01
2
Composition/information on ingredients
Synonyms:
CAS-No.:
Molar mass:
Molecular formula:
EINECS-No.:
Phosphate buffer solution
-----
Hazardous ingredients:
Name according to EC Directives: None
CH: not toxic class
3
Hazards identification
No hazardous product according to the directive 67/548/EWG.
4
First aid measures
After inhalation:
After skin contact:
After eye contact:
After ingestion:
5
6
-wash off with plenty of water.
rinse out with plenty of water with the eyelid held wide open. Summon eye
specialist.
make victim drink plenty of water, summon doctor if feeling unwell.
Fire-fighting measures
Suitable extinguishing media:
--
Special risks:
--
Other information:
non-combustible
Accidental release measures:
Procedures for cleaning / absorption:
Dilute with water and take-up with liquid-absorbent material.
Environmental-protection measures:
None
56
A-96.250.441 / 060923
AMU pH-Redox
Material Safety Data Sheets
7
Handling and storage:
Handling:
Storage:
public
8
no further requirements
tightly closed. Dry. In a well-ventilated place. At room temperature . No
product. Keep out of children's reach.
Exposure controls/personal protection
Personal protective equipment:
Respiratory protection:
not required.
Eye protection:
not required.
Hand protection:
not required
Industrial hygiene:
Contact with eye or skin: Wash with plenty of water.
9
Physical and chemical properties
Form:
Colour:
Odour:
liquid
yellow
odourless
pH value:
(20 oC)
not available
near 100 oC
Ignition temperature:
Flash point:
not available
not available
Explosion limits:
10
11
7
Melting temperature:
Boiling temperature:
lower
upper
not available
not available
o
Density:
(20 C)
1.1 g/cm3
Solubility
in water:
(20 oC)
very good and completely soluble
Stability and reactivity
Conditions to be avoided:
--
Substances to be avoided:
--
Hazardous decomposition products:
none
Toxicological information
Acute toxicity:
No toxicological data available
Further toxicological information:
After skin contact:
-After eye contact:
-After ingestion:
--
A-96.250.441 / 060923
57
AMU pH-Redox
Material Safety Data Sheets
12
Ecological information
Ecotoxic effects:
-Further ecologic data:
None
13
Disposal considerations
Product:
There are no uniform EC regulations for the disposal of chemicals or residues. Chemical residues
generally count as special waste. The disposal of the latter is regulated in the EC member countries
through corresponding laws and regulations, and in Switzerland also by the cantons. We recommend
that you contact either the authorities in charge or approved waste disposal companies which will
advise you on how to dispose of special waste.
Packaging:
Disposal in compliance with official regulations. Handle contaminated packaging in the same way as the
substance itself. If not officially specified differently, non-contaminated packaging may be treated like
household waste or recycled.
14
Transport information
Not subject to transport regulations.
15
Regulatory information
Labelling according to EG Directives
Symbol:
-Name:
-R-phrases:
--
S-phrases:
--
Other national regulations:
Swiss toxic class:
None
16
Other information
Date of issue:
19.05.2003
The information contained herein is based on the present state of our knowledge. It characterizes the
product with regard to the appropriate safety precautions. It does not represent a guarantee of the
properties of the product.
58
A-96.250.441 / 060923
AMU pH-Redox
Material Safety Data Sheets
B.3 Buffer pH 9
1
Identification of the substance/preparation and of the company/undertaking
Identification of the product:
Catalogue No.:
85-114300 or 85-114350
Product name:
Calibration solution pH 9
Use of substance/preparation:
Buffer solution, no public product, only for laboratory use
Manufacturer/supplier identification:
Company:
Swan Analytical Instruments AG, CH - 8616 Riedikon
Tel.: 01/943 63 00, Fax: 01/943 63 01
2
Composition/information on ingredients
Synonyms:
CAS-No.:
Molar mass:
Molecular formula:
EINECS-No.:
Phosphate buffer solution
-----
Hazardous ingredients:
Name according to EC Directives: None
3
Hazards identification
No hazardous product according to the directive 67/548/EWG.
4
First aid measures
After inhalation:
After skin contact:
After eye contact:
After ingestion:
5
6
-wash off with plenty of water.
rinse out with plenty of water with the eyelid held wide open. Summon eye
specialist.
make victim drink plenty of water, summon doctor if feeling unwell.
Fire-fighting measures
Suitable extinguishing media:
--
Special risks:
--
Other information:
non-combustible
Accidental release measures:
Procedures for cleaning / absorption:
Dilute with water and take-up with liquid-absorbent material.
Environmental-protection measures:
None
A-96.250.441 / 060923
59
AMU pH-Redox
Material Safety Data Sheets
7
Handling and storage:
Handling:
Storage:
public
8
no further requirements
tightly closed. Dry. In a well-ventilated place. At room temperature . No
product. Keep out of children's reach.
Exposure controls/personal protection
Personal protective equipment:
Respiratory protection:
not required.
Eye protection:
not required.
Hand protection:
not required
Industrial hygiene:
Contact with eye or skin: Wash with plenty of water.
9
10
11
Physical and chemical properties
Form:
Colour:
Odour:
liquid
blue
odourless
pH value:
(20 oC)
9
Melting temperature:
Boiling temperature:
not available
near 100 oC
Ignition temperature:
Flash point:
not available
not available
Explosion limits:
lower
upper
not available
not available
Density:
(20 oC)
1.1 g/cm3
Solubility
in water:
(20 oC)
very good and completely soluble
Stability and reactivity
Conditions to be avoided:
--
Substances to be avoided:
--
Hazardous decomposition products:
none
Toxicological information
Acute toxicity:
No toxicological data available
Further toxicological information:
After skin contact:
-After eye contact:
-After ingestion:
--
60
A-96.250.441 / 060923
AMU pH-Redox
Material Safety Data Sheets
12
Ecological information
Ecotoxic effects:
-Further ecologic data:
None
13
Disposal considerations
Product:
There are no uniform EC regulations for the disposal of chemicals or residues. Chemical residues
generally count as special waste. The disposal of the latter is regulated in the EC member countries
through corresponding laws and regulations, and in Switzerland also by the cantons. We recommend
that you contact either the authorities in charge or approved waste disposal companies which will
advise you on how to dispose of special waste.
Packaging:
Disposal in compliance with official regulations. Handle contaminated packaging in the same way as the
substance itself. If not officially specified differently, non-contaminated packaging may be treated like
household waste or recycled.
14
Transport information
Not subject to transport regulations.
15
Regulatory information
Labelling according to EG Directives
Symbol:
-Name:
-R-phrases:
--
S-phrases:
--
Other national regulations:
Swiss toxic class:
None
16
Other information
Date of issue:
19.05.2003
The information contained herein is based on the present state of our knowledge. It characterizes the
product with regard to the appropriate safety precautions. It does not represent a guarantee of the
properties of the product.
A-96.250.441 / 060923
61
AMU pH-Redox
Material Safety Data Sheets
B.4 Buffer ORP
1
Identification of the substance/preparation and of the company/undertaking
Identification of the product:
Catalogue No.:
85-121300 / 85-121500 / 85-121700
Product name:
Redox Calibration Solution
for redox electrodes, 430 mV Pt/Calomel, 475 mV Pt/AgCl
Manufacturer/supplier identification:
Company:
SWAN Analytische Instrumente AG, CH-8616 Riedikon, Switzerland
Tel.: +41 1 943 63 00, Fax: +41 1 943 63 01
2
Composition/information on ingredients
Synonyms:
CAS-No.:
Molar mass:
Molecular formula:
EINECS-No.:
Diluted sulphuric acid / iron sulphate solution
7664-93-9
98.08
H2SO4
231-639-5
Hazardous ingredients:
Name according to EC Directives: Sulphuric acid
Hazard symbol: C
R-phrases: 35
EC-Index-No.: 016-020-00-8 Causes severe burns
CAS-No.:
7664-93-9
Content:
<10%
3
Hazards identification
Causes severe burns.
4
First aid measures
After inhalation:
After skin contact:
After eye contact:
After ingestion:
5
fresh air. Summon doctor.
wash off with plenty of water. Immediately remove contaminated clothing.
rinse out with plenty of water with the eyelid held wide open. Immediately
summon eye specialist.
make victim drink plenty of water, avoid vomiting, immediately summon
doctor. Do not attempt to neutralize.
Fire-fighting measures
Suitable extinguishing media: in adaptation to materials stored in the immediate neighborhood.
62
Special risks:
Development of hazardous vapors possible in the event of fire.
Hydrogen may form upon contact with metals (danger of explosion!). The
following may develop in the event of fire: sulphur oxides.
Other information:
Non combustible.
A-96.250.441 / 060923
AMU pH-Redox
Material Safety Data Sheets
6
Accidental release measures:
Person-related precautionary measures
Do not inhale vapors / aerosols.
Procedures for cleaning / absorption:
Dilute with lots of water. Neutralize with sodium carbonate. Take up dry. Forward for disposal. Clean up
affected area.
Environmental-protection measures:
Undiluted solution should not be allowed to enter the sewage system in bigger quantities.
7
8
Handling and storage:
Handling:
No further requirements
Storage:
Tightly closed. Dry. In a well-ventilated place. At room temperature. No
public product. Keep out of children's reach.
Exposure controls/personal protection
Specific control parameter:
MAK Germany (max. workplace conc.) sulphuric acid: 1 mg/m3 total dust
Personal protective equipment:
Respiratory protection:
required when vapors / aerosols are generated.
Eye protection:
required.
Hand protection:
required
Industrial hygiene:
Change contaminated clothing and immerse in water. Protect skin. Wash hands and face after working
with substance.
9
Physical and chemical properties
Form:
Color:
Odor:
liquid
yellow
odorless
pH value:
at 50 g/l H2O:
(20 oC)
|1
Melting temperature:
Boiling temperature:
not available
ca. 100 oC
Ignition temperature:
Flash point:
not available
not available
Explosion limits:
lower
upper
not available
not available
Density:
Bulk density:
(20 oC)
not available
1.1 g/cm3
(20 oC)
very good soluble
Solubility
in water:
A-96.250.441 / 060923
63
AMU pH-Redox
Material Safety Data Sheets
10
Stability and reactivity
Conditions to be avoided:
strong heating
Substances to be avoided:
alkali metals, alkali compounds, ammonia, alkaline earth metals, alkaline earth compounds, acids,
bases, metals, metal alloys, combustible substances, organic solvents, halogenate, permanganate.
11
Hazardous decomposition products:
toxic gases
Further information:
corrosive, incompatible with metals
Toxicological information
Acute toxicity:
Toxicological values are not available due to other dangerous properties of the substance.
Further toxicological information:
Contact with skin causes severe burns under formation of sloughs. Eye contact leads to corneal
damage. Ingestion is followed by severe pains (risk of perforation!), nausea, vomiting, and diarrhea,
and after a latency period of some weeks possibly pyloric stenosis. Inhalation of aerosols resulting in
damage of the mucous membranes.
12
Ecological information
Ecotoxic effects:
Biological effects: Toxic for aquatic organisms. Caustic even in diluted form. Endangers drinking water
supplies if allowed to enter soil and/or water in large quantities.
Harmful due to pH shift. Does not cause biological oxygen deficit.
Further ecologic data:
The following applies to sulphuric acid: biological effects: toxic for aquatic organisms; harmful effect due
to pH shift. LD 50 96h: 10 mg/l; fish: lethal as from 1.2 mg/l; as from 6.3 mg/l lethal in 24 h.
Do not allow larger quantities to enter drinking water supplies, waste water, or soil!
13
Disposal considerations
Product:
There are no uniform EC regulations for the disposal of chemicals or residues. Chemical residues
generally count as special waste. The disposal of the latter is regulated in the EC member countries
through corresponding laws and regulations, and in Switzerland also by the cantons. We recommend
that you contact either the authorities in charge or approved waste disposal companies which will
advise you on how to dispose of special waste.
Packaging:
Disposal in compliance with official regulations. Handle contaminated packaging in the same way as the
substance itself. If not officially specified differently, non-contaminated packaging may be treated like
household waste or recycled.
64
A-96.250.441 / 060923
AMU pH-Redox
Material Safety Data Sheets
14
Transport information
Land transport ADR/RID and GGVS/GGVE (Germany)
GGVS/GGVE-class:
8
Number and letter:
ADR/RID-class:
8
Number and letter:
Name of material:
2796 SCHWEFELSAEURE
1B
1B
River transport ADN/ADNR:
not examined
Sea transport IMDG:
IMDG-class:
8
EmS:
8-06
Correct technical name:
UN-No:
2796
MFAG:
700
SULPHURIC ACID <10%
Air transport ICAO-TI and IATA-DGR:
ICAO/IATA-class: 8
UN-/ID-No: 2796
Correct technical name: SULPHURIC ACID <10%
Packaging group:
II
Packaging group:
II
The transport regulations are cited according to international regulations and in the form applicable in
Germany (GGVS/GGVE). Possible national deviations in other countries are not considered.
15
Regulatory information
Labeling according to EG Directives
Symbol:
C
Name:
Corrosive
R-phrases:
35
Causes severe burns.
S-phrases:
2-26-30
Keep out of reach of children. After eye contact rinse with plenty
of water and immediately summon doctor. Never add water.
German regulations
Water pollution class:
1 (weakly polluting substance) (KBwS classification)
Other national regulations:
Swiss toxic class:
3
16
Other information
Date of issue:
03.02.2005
The information contained herein is based on the present state of our knowledge. It characterizes the
product with regard to the appropriate safety precautions. It does not represent a guarantee of the
properties of the product.
A-96.250.441 / 060923
65
AMU pH-Redox
RS232 Interface
C.
RS232 Interface
C.1
Connection of the Communication Cable
1
Disconnect the instrument from the mains!
2
Connect the communication cable for AMU transmitters to the
transmitter.
RS232 with connected communication
cable
66
3
Connect the other end of the cable to the PC.
4
Connect the instrument to power.
A-96.250.441 / 060923
AMU pH-Redox
RS232 Interface
C.2
Configuration of the AMU Transmitter
1
Enter the menu 5.5.1 on the AMU transmitter and choose
HyperTerminal for the sub-menu Protocol.
Interface
Protocol
Baud Rate
5.5.1
HyperTerminal
115200 Baud
Protocol
Profibus
Modbus RTU
SWAN Bus
HyperTerminal
Webserver
Confirm with <Enter>
2
Enter menu 5.5.2 and choose 115200 Baud for the Baud
rate.
Interface
Protocol
Baud Rate
5.5.2
HyperTerminal
115200 Baud
Baud Rate
2400 Baud
4800 Baud
9600 Baud
19200 Baud
38400 Baud
57600 Baud
115200 Baud
Confirm with <Enter>
A-96.250.441 / 060923
67
AMU pH-Redox
RS232 Interface
C.3
HyperTerminal Configuration on the PC
*
68
A-96.250.441 / 060923
AMU pH-Redox
RS232 Interface
* Note:
If your computer has no COM port, you have to use an USBCOM-Converter to connect the cable to the PC.
After having it connected find out via the Device Manager which
port you are using (Settings/Control Panel/System/Hardware/
Device Manager).
Clicking on the Device
Manager leads you to
the following scheme:
Clicking on Ports
(COM & LPT)
shows you which
port is used.
This address of the port should be used for configuring the
HyperTerminal.
A-96.250.441 / 060923
69
AMU pH-Redox
RS232 Interface
70
A-96.250.441 / 060923
AMU pH-Redox
RS232 Interface
A-96.250.441 / 060923
71
AMU pH-Redox
RS232 Interface
C.4
Functions of the AMU RS232 Interface
The AMU RS232 Interface has two functions: It allows the transfer
of log data onto a PC and the upgrade of AMU instruments with a
new firmware.
C.4.1 Logger Download
Log interval
The internal logger can save approx. 1500 data records, stored at a
selectable interval. The following table shows the period of time
covered with the corresponding interval.
Operation
Sensors
Signal Outputs
Relay Contacts
Logger
4.4
4.4.1
Logger
Log Interval x Sec/Min/Hour
Clear Logger
no/yes
Log Interval
1 Second
5 Seconds
1 Minute
5 Minutes
10 Minutes
30 Minutes
1 Hour
Interval
1s
5s
1 min
5 min
10 min
30 min
1h
Time
25 min
2h
25 h
5d
10 d
31 d
62 d
Default value for the the interval is set on 30 min.
72
A-96.250.441 / 060923
AMU pH-Redox
RS232 Interface
C.4.2 Log download
Download
Procedure
Choose the configured
connection.
Š On the keyboard,
press Esc.
Š Browse to the
folder you have
created to archive
printouts.
A text file AMU*.TXT is created. Existing files are renamed
AMU*0.TXT, AMU*1.TXT, AMU*2.TXT, etc.
* = Abbr. for the corresponding instrument, e.g. AMUCON.TXT is the
file for the instrument AMU pH-Redox.
A-96.250.441 / 060923
73
AMU pH-Redox
RS232 Interface
Example of a
logger file:
AMUCON.txt
Date
08.05.06
31.05.06
01.06.06
Time
15:00:00
09:30:00
17:30:00
Logger download of an AMU pH-Redox.
Alarms
0536871936
0536871424
0536871424
pH [pΗ]
0.0
0.0
0.0
[]
Temp. [°C] FLOW [l/h]
25.0
0.0
0.0
7.0
0.0
7.0
You can also download the Message List and the Calibration
History.
Message List of an AMU pH-Redox: EVTCON.txt
Date
02.03.06
02.03.06
22.06.06
Time
13:34:00
14:49:37
17:12:41
Code
E033
E012
E011
Desc
Power-on
Temp. disconnected
Temp. shorted
State
Active
Acknowledged
Active
Calibration History of an AMU pH-Redox: CALCON.txt
pH
Date
22.06.06
Time
17:12:26
Changing the
folder for
download
data
74
Offset [mV] Slope [mV]
-3
58.21
To change the folder where the files are saved,
1
2
3
press the “Receive”-button.
choose the directory and the folder
and the receiving protocol must be Zmodem.
A-96.250.441 / 060923
AMU pH-Redox
RS232 Interface
C.4.3 Clear Logger
To delete all log data, proceed as follows:
Operation
Sensors
Signal Outputs
Relay Contacts
Logger
Logger
Log Interval
Clear Logger
4.4
4.4.2
30 Minutes
no
Clear Logger
no
yes
4.4.2
Logger
Log Interval
Save? 30 Minutes
Clear Logger
yes
Yes
No
A-96.250.441 / 060923
Select yes and press
to accept.
Press
Exit
Enter
.
yes highlighted, press Enter
to clear logger. The system
reboots, all log data is
deleted.
75
AMU pH-Redox
RS232 Interface
C.5
Firmware Upload
Only SWAN trained and authorized personnel shall perform the
tasks described in this section.
Requisites
Firmware
upload
The HyperTerminal connection must be configured. See
HyperTerminal Configuration on the PC, p. 68
1
Choose the
configured
connection.
AMU pH-Redox
3 Press <Enter> and
confirm it by
pressing “y”.
76
A-96.250.441 / 060923
AMU pH-Redox
RS232 Interface
4
If upload fails (instrument
does not restart), repeat
steps 1 and 4.
Perform Set defaults 5.4.2 > completely and setup the
instrument according to user specifications.
A-96.250.441 / 060923
77
AMU pH-Redox
Index
D.
Index
A
Alarm Relay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7, 16
5
C
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
25
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22, 23
Communication Cable . . . . . . . . . . . . . . . . . . . . . . . . . .
66
Control Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7
D
Default Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50, 78
Description of the System . . . . . . . . . . . . . . . . . . . . . . . .
5
Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
9
78
E
Error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Establish Sample Flow . . . . . . . . . . . . . . . . . . . . . . . . .
Explanations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
27
21
33
F
Firmware Upload . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
76
I
Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7, 16
7, 19
L
Logger Download . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
72
M
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Material Safety Data Sheets . . . . . . . . . . . . . . . . . . . . . .
Measuring Principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
22
53
5
13
P
Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
16
A-96.250.441 / 060923
AMU pH-Redox
Index
R
Relay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Relay Contacts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
RS232 1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7
16
9, 66
S
Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Sample Flow, Establish . . . . . . . . . . . . . . . . . . . . . . . . .
Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Sensor, Installation of . . . . . . . . . . . . . . . . . . . . . . . . . . .
Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Signal Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Software Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Symbol, Attention . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Symbol, Warning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
System, Description of . . . . . . . . . . . . . . . . . . . . . . . . . .
3
21
8, 17
17
21
7, 16
9
11
3
3
5
T
Technical Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Trouble Shooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6
27
A-96.250.441 / 060923
79
AMU pH-Redox
SWAN
is represented worldwide by subsidiary companies
and distributors.
cooperates with independent representatives
all over the world.
SWAN Products
Analytical Instruments for:
High Purity Water
Feedwater, Steam and Condensate
Potable Water
Pool and Sanitary Water
Cooling Water
Waster Water and Effluents
Made in Switzerland.
80
A-96.250.441 / 060923