AMU pH/Redox Operator’s Manual Version 3.50 and later SWAN sets the Standard A-96.250.441 / 060923 Customer Support SWAN and its representatives maintain a fully trained staff of technical specialists around the world. For any technical question, contact your nearest SWAN representative, or the manufacturer: SWAN ANALYTISCHE INSTRUMENTE AG Turicaphonstrasse 29 8616 Riedikon Switzerland Internet: www.swan.ch E-mail: [email protected] Document Status Title: AMU pH/Redox Operator’s Manual ID: A-96.250.441 Revision Issue 00 September 2006 © 2006, SWAN ANALYTISCHE INSTRUMENTE AG, Switzerland, all rights reserved Information contained in this document is subject to change without notice. AMU pH-Redox AMU pH-Redox AMU pH-Redox Table of Contents 1. Safety Instructions 1.1 1.2 Warning Notices. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . General Safety Regulations . . . . . . . . . . . . . . . . . . . . . . . 2. Product Description 2.1 2.2 2.3 2.4 2.5 2.6 2.7 2.7.1 2.7.2 2.7.3 Description of the System . . . . . . . . . . . . . . . . . . . . . . . . Technical Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Swansensor pH and Redox . . . . . . . . . . . . . . . . . . . . . . . Swansensor pH AY and Redox AY . . . . . . . . . . . . . . . . . Swansensor pH SI and Redox SI . . . . . . . . . . . . . . . . . . Swansensor pH FL and Redox FL. . . . . . . . . . . . . . . . . . Software. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Menu Navigation Keys . . . . . . . . . . . . . . . . . . . . . . . . . Software Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3. Installation 3.1 3.2 3.3 3.4 3.5 3.5.1 3.5.2 3.6 3.7 3.8 3.9 3.10 3.10.1 3.10.2 Installation Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . Mounting of the Instrument . . . . . . . . . . . . . . . . . . . . . . . Connection Schemes. . . . . . . . . . . . . . . . . . . . . . . . . . . . Power Supply. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Relay Contacts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Alarm Relay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Relay Contacts 1 and 2 . . . . . . . . . . . . . . . . . . . . . . . . Signal Outputs1 and 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Installation of Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . Interfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . RS232 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Profibus (optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-96.250.441 / 060923 3 4 5 6 8 8 8 8 9 9 10 11 12 13 14 16 16 16 16 16 16 17 17 19 19 19 1 AMU pH-Redox AMU pH-Redox 4. Instrument Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 5. Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 6. Maintenance (Quality assurance) 6.1 6.2 6.3 6.4 Maintenance Table. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Cleaning of Electrodes . . . . . . . . . . . . . . . . . . . . . . . . . . . Longer stop of operation . . . . . . . . . . . . . . . . . . . . . . . Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8. Trouble Shooting 8.1 8.2 Error Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Error List. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A. Appendix A.1 A.2 Program Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Program List and Explanations . . . . . . . . . . . . . . . . . . . . Default Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B. Material Safety Data Sheets B.1 B.2 B.3 B.4 2 Buffer pH 4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Buffer pH 7 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Buffer pH 9 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Buffer ORP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 23 24 25 27 28 30 33 50 53 56 59 62 C. RS232 Interface C.1 C.2 C.3 C.4 C.4.1 C.4.2 C.4.3 C.5 Connection of the Communication Cable. . . . . . . . . . . . . Configuration of the AMU Transmitter . . . . . . . . . . . . . . . HyperTerminal Configuration on the PC . . . . . . . . . . . . . Functions of the AMU RS232 Interface . . . . . . . . . . . . . . Logger Download . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Log download . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Clear Logger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Firmware Upload . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66 67 68 72 72 73 75 76 D. Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78 A-96.250.441 / 060923 AMU pH-Redox Safety Instructions 1. General Target audience Operator’s Manual Location 1.1 Safety Instructions The instructions included in this section explain the potential risks associated with instrument operation and provide important safety practices designed to minimize these risks. If you carefully follow the information contained in this section, you can protect yourself from hazards and create a safer work environment. More safety instructions are given throughout this manual, at the respective locations where observation is most important. Strictly follow all safety instructions in this publication. Operator: Qualified person who uses the equipment for its intended purpose. Instrument operation requires thorough knowledge of applications, instrument functions and software program as well as all applicable safety rules and regulations. The AMU Operator’s Manual shall be kept in proximity of the instrument. Warning Notices The symbols used for safety-related notices have the following significance: WARNING Symbols WARNING Generally, the triangular warning symbol indicates the possibility of personal injury or even loss of life if instructions are not followed, e.g. electrical shock hazard ATTENTION Symbols ATTENTION The attention symbol indicates the possibility of equipment damage, malfunctions or incorrect process results if instructions are not followed. A-96.250.441 / 060923 3 AMU pH-Redox Safety Instructions 1.2 General Safety Regulations Legal Requirements The user is responsible for proper system operation according to all local, state and federal laws that may apply. All precautions must be followed to ensure safe operation of the instrument. Spare Parts and Disposables Use only official SWAN spare parts and disposables. If other parts are used during the normal warranty period, the manufacturer’s warranty is voided. Modifications Modifications and instrument upgrades shall only be carried out by an authorized Service Technician. SWAN will not accept responsibility for any claim resulting from unauthorized modification or alteration. Electrical shock hazard WARNING 4 If proper operation is no longer possible, the instrument must be disconnected from all power lines, and measures must be taken to prevent inadvertent operation. To prevent from electrical shock, always make sure that the ground wire is connected. Service shall be performed by authorized personnel only. Whenever electronic service is required, disconnect instrument power and power of devices connected to • relay 1, • relay 2, • alarm relay. A-96.250.441 / 060923 AMU pH-Redox Product Description 2. Product Description 2.1 Description of the System This instrument is applicable for the measurement of the pH respectively ORP value in water. Application range Measuring Principle Temperature compensation pH and redox (ORP) is measured in many different applications: swimming pools waste water high purity water power plants (steam, condensate) industry Depending on the application, different electrode types and flow cells/fittings are needed. pH: The electrode consists of a pH sensitive glass electrode and a reference electrode = measuring chain. Depending on the pH value in the water, the mV value of the measuring chain changes. If both electrodes are integrated into one shaft it is a combined electrode. Reference electrodes are available with gel (Swansensor pH for potable water and swimming pools, Swansensor pH AY for waste water) or with liquid electrolyte (Swansensor pH SI for power plant cycle, Swansensor pH FL for high purity water). ORP: The electrode consists of a platinum pin and a reference electrode = measuring chain. Depending on the ORP value in the water, the mV value of the measuring chain changes. If both electrodes are integrated into one shaft it is a combined electrode. Reference electrodes are available with gel (Swansensor pH for potable water and swimming pools, Swansensor pH AY for waste water) or with liquid electrolyte (Swansensor ORP SI for power plant cycle, Swansensor ORP FL for high purity water). pH: The pH value depends on temperature. If the temperature of the sample is not stable, we recommend to install a temperature sensor. If no temperature sensor is chosen, please program the temperature of the sample in the instrument. For high purity water applications (power plant, high purity water): Additionally to the Nernst compensation a non-linear solution compensation or linear compensation with coefficient can be chosen. ORP: Temperature compensation is not necessary. A-96.250.441 / 060923 5 AMU pH-Redox Product Description 2.2 Technical Data The AMU measuring and control transmitter is used for panel installation. It has connections for a pH/ORP electrode and a Pt1000 temperature probe, and for a digital sample flow meter. Electronics case: Noryl® resin Protection degree: IP54 (front) Large backlit LC-Display, 75 x 45 mm Dimensions: 96 x 96 x 119 mm (DIN 43700) Weight: 0.45 kg Ambient temperature: -10 to +50°C Limit range of operation:25 to +65°C Storage and transport: -30 to +85°C Humidity: 10 to 90 % relative, non-condensing Voltage: 85 - 265 VAC, 47 - 63 Hz, or 24 VDC ±15% Power consumption: max. 7 VA Measuring range Resolution 0.00 to 14.00 pH Resolution: 0.01 pH -500 to 1500 mV Resolution: 1 mV Automatic range switching. Temperature Input Temperature compensation 6 for Pt1000 type sensor (DIN class A) Measuring range: -30 to +250 °C Resolution: 0.1 °C Nernst Non-linear solution compensation for high purity water Linear with coefficient for high purity water A-96.250.441 / 060923 AMU pH-Redox Product Description Temperature monitoring Safety features Alarm Relay Input Relays Signal Outputs Control Function Communication Interface Serial Interface Alarm if the transmitter temperature is higher than +65 °C or lower than 0 °C. No data loss after power failure, all data is saved in non-volatile memory. Over voltage protection of in- and outputs. Galvanic separation of measuring inputs and signal outputs. One potential free contact for summary alarm indication for programmable alarm values and instrument faults. Maximum load: 100 mA / 50 V One input for potential-free contact. Programmable ‘HOLD’ or ‘OFF’ function. Two potential-free contacts programmable as • limit switches for measuring values • controllers • timer for system cleaning with automatic hold function Maximum load: 100 mA / 50 V Two programmable signal outputs for measured values (freely scaleable, linear or bilinear) or as continuous control output (control parameters programmable). Current loop: 0/4 - 20 mA Maximum burden: 510 Ω Relays or current outputs programmable for 1 or 2 pulse dosing pumps, solenoid valves or for one motor valve. Programmable P, PI, PID or PD control parameters. Optional RS485 interface with Fieldbus protocol PROFIBUS DP or Modbus, galvanically separated. RS232 for data logger download to PC with Microsoft HyperTerminal and for Firmware updates. A-96.250.441 / 060923 7 AMU pH-Redox Product Description 2.3 Applications Swansensor pH and Redox Potable water Swimming pools Details see data sheet. Operating Conditions 0 to 50 °C Max. pressure: < 2 bar 2.4 Applications Swansensor pH AY and Redox AY Waste water Details see data sheet. Operating Conditions 0 to 50 °C Max. pressure: < 2 bar 2.5 Applications Swansensor pH SI and Redox SI Power plant cycle Details see data sheet. Operating Conditions 0 to 50 °C Pressure-free Flow speed: 5 to 10 l/h 2.6 Applications Swansensor pH FL and Redox FL High purity water Details see data sheet Operating Conditions 8 0 to 50 °C Pressure-free Flow speed: 5 to 10 l/h A-96.250.441 / 060923 AMU pH-Redox Product Description 2.7 Software 2.7.1 Display Menu navigation keys RUN HOLD normal operation The Input is activated. The Signal Outputs are frozen. OFF The Input is activated; the Signal Outputs are set to 0 (0/4 mA) error blinking: fatal error Status R1 R2 sample flow 8 l/h Time 7.43 pΗ 25 °C sample temperature Relay Status upper / lower limit reached upper / lower limit not yet reached control upw. / dnw.: no action control upw. / dnw.: active, dark bar indicates control intensity motor valve: valve closed motor valve: open, dark bar indicates approximate position timer timer: timing active (hand rotating) A-96.250.441 / 060923 9 AMU pH-Redox Product Description 2.7.2 Menu Navigation Keys Keys Exit to exit a menu or command (without accepting) to move back to the previous menu level to move DOWN in a menu list and to decrease digits to move UP in a menu list and to increase digits Enter Program Navigation RUN R1 R2 to open a selected sub-menu to confirm an entry 14:10:45 x pH 25.0°C Enter Exit Main Menu Messages Diagnostics Maintenance Operation Installation 1 For a detailed listing of the menu see Program Overview, p. 30 10 A-96.250.441 / 060923 AMU pH-Redox Product Description 2.7.3 Software Structure Main Menu Messages Diagnostics Maintenance Operation Installation 1 Messages Pending Errors Message List 1.1 Diagnostics Identification Sensors Sample I/O State Interface 2.1 3.1 Maintenance Calibration Simulation Set Time 02.04.05 16:49:30 A-96.250.441 / 060923 Operation Sensors Relay Contacts Logger 4.1 Installation Sensors Signal Outputs Relay Contacts Miscellaneous Interface 5.1 Menu Messages 1 Reveals pending errors as well as an event history (time and state of events that have occured at an earlier point in time). Menu Diagnostics 2 Provides user relevant instrument and sample data. Menu Maintenance 3 For instrument calibration, relay and signal output simulation and to set the instrument time. It is used by the service personnel. Menu Operation 4 Subset of menu Installation 5, but processrelated. User relevant parameters that might need to be modified during daily routine. Normally password protected and used by the processoperator. Menu Installation 5 For initial instrument set up by SWAN authorized person, to set all instrument parameters. Normally password protected. 11 AMU pH-Redox Installation 12 3. Installation 3.1 Installation Checklist Check Instrument’s specification must conform to your power ratings. Installation Mount the instrument in vertical position. Display should be at eye level. Electrical Wiring Connect all external devices like limit switches and current loops. Connect power cord; do NOT switch on power yet! Flow cell / fitting Install flow cell / fitting. Flow cell: Ensure outlet is pressure-free. Electrode Remove protective cap. Install in flow cell/fitting. Remove plug cap. Connect the sensor(s) to the AMU transmitter. Store sensor caps for interruption of operation. QV-flow: Install reference in the hole near the panel. Fitting: Ensure fitting is water-tight. Power up Flow cell: Turn on sample flow and wait until it is completely filled. Switch on power. Instrument setup Flow cell: Adjust sample flow. Program all parameters for sensor and external devices (interface, recorders, etc.). Program all parameters for instrument operation (limits, alarms). Run-in period Let the instrument run continuously for 1 h. pH electrode calibration Calibrate pH electrode. ORP in potable /waste water Calibrate ORP electrode. ORP in high purity water: Process calibration Let the electrode run in for at least 24 h. Make a valid manual measurement with a suitable ORP electrode. Compare this value to the value, indicated by the AMU. If necessary, correct. A-96.250.441 / 060923 AMU pH-Redox Installation 3.2 Mounting of the Instrument Standard housing according to DIN 43700: 96 x 96 x 119 mm 89.8 96 89.8 96 112 7 A-96.250.441 / 060923 13 AMU pH-Redox Installation 3.3 Connection Schemes Use only the terminals shown in this diagram, and only for the mentioned purpose. Use of any other terminals will cause short circuits with possible damage to material and personnel. Rear view 14 A-96.250.441 / 060923 AMU pH-Redox Installation SENSOR LINE RS232 RELAYS - INPUT - SIGNAL OUT OPTION PROFIBUS or MODBUS SHIELD The type plate for the product identification is on the back side of the AMU transmitter. A-96.250.441 / 060923 11.431.102 pH / Redox 15 AMU pH-Redox Installation 3.4 Power Supply Power Supply Connector Neutral Power 3.5 Relay Contacts 3.5.1 Alarm Relay Note: Max. load 100 mA / 50 V Terminal 5/6 3.5.2 Relay Contacts 1 and 2 Note: Max. load 100 mA / 50 V Relay 1: Terminal 1/2 Relay 2: Terminal 3/4 3.6 Signal Outputs1 and 2 Note: Max. burden 510 Ω If signals are sent to two different receivers, use signal isolator (loop isolator). Signal Output 1: Terminal 13 (+) and 12 (-) Signal Output 2: Terminal 14 (+) and 12 (-) Current loop: 0 - 20 mA resp. 4 - 20 mA 3.7 Input Use only potential-free (dry contacts). Terminal: 11/10 16 A-96.250.441 / 060923 AMU pH-Redox Installation 3.8 Sensor Connect the sensor as follows: pH / ORP Sensor: Terminal 21: shielded line Terminal 22: shield External reference sensor: Terminal 23: shielded line Terminal 24: shield Temperature sensor Pt1000: Terminal 20: shielded line Terminal 19: shield 3.9 Installation of Sensors Flow cell Cautiously pull off rubber cap from pH/ORP electrode tip. Fragile! M-Flow Handle with care! Insert pH/ORP electrode into one of the PG threads and fasten handtight. Take off plug cap. Connect to sensor cable. Store the plug and rubber cap for interruptions of operation. The temperature sensor is already mounted and connected to the cable. Flow cell QV- We recommend to mount the pH/ORP SI respectively Reference FL Flow electrode in the insert near the panel. Flow cell, view from above (simplified): pH SI / Reference FL sensor Temperature sensor pH FL sensor All sensors are fragile! Handle with care! Electrode Cautiously pull off rubber cap from pH/ORP electrode tip. Insert pH/ORP electrode into the opening (see above) and fasten hand-tight. Take off plug cap. Connect to sensor cable. Store the plug and rubber cap for interruptions of operation. A-96.250.441 / 060923 17 AMU pH-Redox Installation Temp. sensor The sensor is already mounted and connected to the cable. Reference The reference electrode (Swansensor pH SI, Swansensor ORP SI, electrode Swansensor FL) is delivered with an electrolyte tube, which is already mounted and filled. The sealed electrolyte bottle is also included in delivery. - Cautiusly take of the rubber cap from electrode tip. Make sure that the black cap at the tip of the electrode is fastened hand-tight. - The reference electrode has to be free of air bubbles: Hold the electrode, tip downwards, in one hand and hold the electrolyte tube with the other hand straight above the the electrode. Shake the electrode slightly in an angular position to release air, which might be inside. Release air bubbles in the electrolyte tube by knocking on it. - Insert the electrode into the flow cell and fasten hand-tight. - Cut off the tip of the electrolyte bottle. Take the plug out of the electrolyte tube. Put the electrolyte bottle firmly on the tube. - Mount the electrolyte bottle in the click holder (see 2.3 Instrument overview, p. 11). - Puncture the electrolyte bottle to allow pressure equilibration. - Take off plug cap. Connect to sensor cable. Store caps. UNIDIP fitting Please refer to the manual of the UNIDIP fitting how to install the pH/ ORP electrode and the temperature sensor in detail. Ensure that all surfaces and o-rings are clean. If mounted correctly, the UNIDIP fitting is tight. No water is allowed inside the UNIDIP fitting. Otherwise the cables corrode. If you use an UNIDIP Rinsematic, make sure the nozzles face the pH/ORP sensor. Temperature sensor and electrodes: Fragile! Handle with care! 18 A-96.250.441 / 060923 AMU pH-Redox Installation 3.10 Interfaces 3.10.1 RS232 The interface RS232 is on the backside of the AMU transmitter. For connecting the communication cable and the configuration (using RS232 for data logger download to PC and for firmware updates) see RS232 Interface, p. 66 3.10.2 Profibus (optional) The interface RS485, galvanically separated. To connect several instruments by means of a network, use a freecable end PROFIBUS cable. See picture below. A-96.250.441 / 060923 19 AMU pH-Redox Installation termination switch Please buy a free-cable end PROFIBUS cable from your local supplier. NOTE: The color of the wires might be different. Before connecting it to the transmitter, please check the corresponding data sheet. The PROFIBUS connections have to be short-circuited. The shield has the function to change or replace the transmitter without inactivating the bus connection. NOTE: Terminate bus by means of ON/OFF switch on the RS485 interface only. If only one instrument is installed, or on the last instrument in a bus, the termination switch must be ON. 20 A-96.250.441 / 060923 AMU pH-Redox Instrument Setup 4. Sensor Establish Sample Flow Instrument Setup Check if: Power is available Sensor is installed and connected correctly. For connecting the sensor to the transmitter, see Swansensor pH and Redox, p. 8 1 Check if sample is available. 2 Assure pressure-free sample outflow. 3 Open sample flow. (Wait until flow cell is completely filled.) 4 If applicable, check flow cell for leaks. 5 Connect the instrument to power. 6 Adjust sample flow. 7 Set all necessary parameters in Installation, Sensors 5.1. - Flow: Set if you use the Q-flow cell (with Swan flow meter) - Parameter: Set measuring parameter pH or ORP Sensor check: No. - Temperature: Set if a temperature sensor has been connected. If no, program the temperature of the sample (default temp.) If yes, you can set a temp. correction, if necessary. - Temperature compensation: Only for pH, not for ORP. - Standard solution(s): Program the buffer values (pH buffer table) or the ORP calibration solution. 8 Program all parameters for external devices (interface, recorders, etc.). Program all parameters for instrument operation (limits, alarms, etc.) 9 5. Let run in for 1 h. Correct if necessary, or calibrate. See Installation check list. Operation To have a general survey of the menu 4: Operation, see Program Overview, p. 30. For explanations see Program list and explanations. You can modify: The Filter Time Constant and the time for Hold after Calibration for the sensor The Settings for the Relay Contacts and the Input. A-96.250.441 / 060923 21 AMU pH-Redox Maintenance (Quality assurance) 6. Maintenance (Quality assurance) This section describes the activities intended to retain the instrument in, or to restore it to a state in which it maintains the required or specified performance. 6.1 Maintenance Table Swansensor pH / Swansensor Redox (ORP) Twice per month Calibrate electrode. Ensure buffers are not expired. If necessary, replace electrode. Yearly If necessary, clean electrode. Swansensor pH AY / Swansensor Redox (ORP) AY Twice per month Clean electrode. Monthly Calibrate electrode. Ensure buffers are not expired. If necessary, replace electrode. Swansensor pH SI or FL Weekly Check level in electrolyte bottle. If necessary, change bottle. Monthly Calibrate electrode. All 3 months Open cap of reference electrode slightly and let flow out 5 ml of electrolyte. Fasten cap hand-tight. Swansensor Redox (ORP) SI or FL 22 Weekly Check level in electrolyte bottle. If necessary, change bottle. Monthly If necessary, correct electrode. All 3 months Open cap of reference electrode slightly and let flow out 5 ml of electrolyte. Fasten cap hand-tight. A-96.250.441 / 060923 AMU pH-Redox Maintenance (Quality assurance) 6.2 Cleaning of Electrodes Cleaning Swansensor Take the electrode out of the flow cell. If necessary wipe the electrode pH and pH AY shaft and the green tip cautiously with a soft, clean, and damp paper tissue. Remove grease with a tissue moistured with alcohol (caution! Inflammable!). If the electrode is very dirty, put it into 1% diluted hydrochloric acid for roughly 1 min. Caution! Etching!! Afterwards rinse the electrode thoroughly with clean water. Remount in the flow cell again. Let the electrode run-in for 1 h before the first calibration. Swansensor Take the electrode out of the flow cell. If necessary, wipe off dirt ORP and ORP cautiously with a soft, clean, and damp paper tissue. AY Dull platinum surfaces indicate a contamination. Clean the electrode by putting it into the calibration solution or in 1% diluted hydrochloric acid for some minutes. Caution! Both solutions are etching! Afterwards rinse the electrode thoroughly with clean water. Remount in the flow cell again. Let the electrode run-in for 1 h before the first calibration. Swansensor If necessary, wipe the green tip of the electrode cautiously with a soft, pH SI and pH clean, and damp paper tissue. If necessary, remove grease with a FL tissue, moistured with alcolhol (Caution! Inflammable!). Do not insert into acids !!! Reference electrode: Open the cap (diaphragm) at the tip of the electrode slightly and let flow out some ml of electrolyte. Caution! KCL is corrosive! Fasten cap hand-tight again. Afterwards rinse the electrode thoroughly with high purity water. Remount in the flow cell again. Let the electrode run-in for 1 h before the first calibration. Swansensor If necessary, wipe the tip of the electrode cautiously with a soft, clean, ORP SI and and damp paper tissue. Do not insert into acids !!! ORP FL Reference electrode: Open the cap (diaphragm) at the tip of the electrode slightly and let flow out some ml of electrolyte. Caution! KCL is corrosive! Fasten cap hand-tight again. Afterwards rinse the electrode thoroughly with high purtiy water. Remount in the flow cell again. Let the electrode run-in for at least 24 h before the first correction. A-96.250.441 / 060923 23 AMU pH-Redox Maintenance (Quality assurance) 6.3 Longer stop of operation 24 Stop sample flow. Shut off power of the instrument. Take off the cable of the electrode(s) and put the plug cap on it. Take the electrode(s) out of the flow cell/fitting and rinse with clean water. Fill 3.5 molar KCl (if not available: water) into the rubber cap and place it on the tip of the electrode. Never store combined pH/ ORP electrodes, respectively reference electrode, dry! Store the electrode(s) with tip pointing downwards in a frostprotected room. A-96.250.441 / 060923 AMU pH-Redox Maintenance (Quality assurance) 6.4 Calibration Process pH / Use a suitable, high quality electrode and instrument to determine the ORP sample value. Ensure, the electrode is calibrated correctly! Enter this value under Process value in menu Process pH/ORP. 3.1 Maintenance Calibration Simulation Set Time 01.01.05 16:30:00 Enter Enter Enter Process pH Calibration Process pH Standard pH Enter 3.1.1 3.1.1.1 Current Value Offset 7.78 pH x mV Process Value Save 7.78 pH <Enter> Process pH Current Value Offset 7.78 pH x mV Process Value Save 7.70 pH <Enter> Enter correct value with arrow keys. 3.1.1.1 Process pH Current Value Offset 7.70 pH y mV Process Value Save 7.70 pH <Enter> Process pH Current Value Offset 3.1.1.1 press Enter 3.1.1.1 7.70 pH y mV Calibration successful Possible er- Offset error: ror message Last calibration wrong. Electrode old or dirty. Cable wet. A-96.250.441 / 060923 25 AMU pH-Redox Maintenance (Quality assurance) Standard pH 3.1 Maintenance Calibration Enter Simulation Set Time 01.06.04 16:30:00 Enter Enter Calibration Fill System Process pH Standard pH 3.1.3 Enter Remove the electrode (and, if applicable, the temp. sensor) from the flow cell. Now follow the instructions on the display. Standard 1 and 2 are set in 5.1.5 Buffer/Cal. solutions have to be clean. Do not use if expired. Always rinse electrode before dipping it into the solution. Flow cell QV- The flow cell has a removable calibration vessel. By a short left turning fow, high purity movement, this vessel can be unlocked and taken away. After rinsing, water you can fill the buffer directly into the vessel. To refasten the vessel again, turn it so that the groove is in front. Insert the vessel in the Bajonett fastening. Turn to the right till stop. cog groove Go to Menu 3.1.3. The instrument guides you through the complete calibration process. If you have completed the required action press <Enter> to proceed. • Rinse and dry electrode and put it into standard 1 • Standard 1, current value (Progress is shown). End for ORP. • Rinse and dry electrode and put it in standard 2 • Standard 2, current value (Progress is shown) • Rinse and dry electrode and put it into the flow cell Possible er- Offset error or Slope error: ror message Old / dirty / wrong buffer solutions. Electrode old or dirty. Cable wet. 26 A-96.250.441 / 060923 AMU pH-Redox Trouble Shooting 8. Trouble Shooting 8.1 Error Messages The menu 1 - Messages - reveals pending errors as well as the message list with text, time and state of errors that have occured before. Error Nonfatal error. The system continues to operate normally, except for the defective part. Fatal error (blinking symbol) If used for control and Relay 1, 2 as well as the Signal Outputs are deactivated, i. e. any control is interrupted. The indicated measured values are incorrect. Depending on the error (see Error List, p. 28) the system may stop operating. Pending Errors Error Code Sample Flow low <Enter> to Acknowledge 1.1.5 E010 Errors remain in the Pending Errors window until corrected, even if acknowledged. Acknowledging the error only deactivates the Alarm Relay, i.e. it switches off an alarm device if available. If an error is corrected, by intervention or by the system itself, it is listed in the the Message List (Menu 1.2). The symbol disappears. The last 64 messages are stored for viewing. They are deleted if SET Default Values (Menu 5.4.2) is carried out. A-96.250.441 / 060923 27 AMU pH-Redox Trouble Shooting 8.2 Error List Bold and red error code: Fatal error 28 Error Description Corrective action E001 Alarm high - check process - check electrode - check progr. value in 5.3.1.1 E002 Alarm low - check process - check electrode - check progr. value in 5.3.1.1 E007 Sample Temp. high - check sample temperature - check progr. value in 5.3.1.4 E008 Sample Temp. low - check sample temperature - check progr. value in 5.3.1.4 E009 Sample Flow high - check sample flow - check progr. value in 5.3.1.3 E010 Sample Flow low - establish sample flow - clean instrument - check progr. value in 5.3.1.3 E011 Temp. shorted check wiring of temperature sensor, check temp. sensor E012 Temp. disconnected check wiring of temperature sensor, check temp. sensor E013 Case Temp. high - check case/environment temperature - check progr. value in 5.3.1.5 E014 Case Temp. low - check case/environment temperature - check progr. value in 5.3.1.6 E017 Control time-out check control device or programming in Installation, Relay contact, Relay 1/2 5.3.2-3 A-96.250.441 / 060923 AMU pH-Redox Trouble Shooting Error Description Corrective action E024 Input active See Menu 5.3.4.4 (If Fault Yes is programmed in menu 5.3.4) E025 IC MK41T56 call service E026 IC LM75 call service E027 IC PCF8574 call service E028 EEProm Microcon call service E029 EEProm Motherboard call service E030 EEprom Front-End call service E031 Calibration Rec Out call service E032 Wrong Front-End call service E033 Power-on none, normal status E034 Power-down power on A-96.250.441 / 060923 29 AMU pH-Redox Appendix A. Appendix A.1 Program Overview Messages (Main menu 1) Pending errors, 1.1 Message list, 1.2 Identification, 2.1 Designation AMI pH/Redox, 2.1.1 Version 3.50, Instrument, 2.1.3.1 2.1.2 Mother Board, 2.1.3.2 Factory test, 2.1.3 Front-end, 2.1.3.3 Years, days, hours Operating time, 2.1.4 minutes, seconds, Electrode, 2.2.1 Current value, pH or mV, 2.2.1. Raw value, mV, 2.2.1.2 Cal. history, 2.2.1.5 Miscellaneous, 2.2.3 Case temp., °C, 2.2.3.1 Sensors, 2.2 Diagnostics (Main menu 2) Sample, 2.3 Sample ID, 2.3.1 Temperature °C, Pt1000, 2.3.2 Sample flow, l/h, Hz, 2.3.3 I/O State, 2.4 Alarm relay, 2.4.1 Relay 1/2, 2.4.2 Input Signal output 1/2/3 Interface, 2.5 Protocol, 2.5.1 Baud rate, 2.5.2 (3. SA option) (only with option RS485) Menu Messages and Diagnostics are always accessible for everybody. No password protection. No settings can be modified. 30 A-96.250.441 / 060923 AMU pH-Redox Appendix Calibration, 3.1 Maintenance Simulation, 3.2 (Main menu 3) Process pH or ORP, 3.1.1 Standard pH or ORP, 3.1.2 Alarm Relay, 3.2.1 Relay 1/2, 3.2.2-3 Signal output 1/2, 3.2.4-5 Set time, 3.3 Sensors, 4.1 Operation (Main menu 4) Relay contacts, 4.2 Logger, 4.3 Filter time constant, 4.1.1 Hold after cal., 4.1.2 Alarm relay, 4.3.1 Alarm , 4.3.1.1 Relay 1/2, 4.3.2-3 acc. Installation Input, 4.3.4 acc. Installation acc. Inst. Log Interval, 4.4.1 Clear logger, 4.4.2 Menu Maintenance is for service: Calibration, simulation of outputs and set time/ date. Please protect with password. Menu Operation is for the user, allowing to set limits, alarm values, etc. The presetting is done in the menu Installation (only for the System engineer). Please protect with password. A-96.250.441 / 060923 31 AMU pH-Redox Appendix Sensors, 5.1 Signal outputs, 5.2 Flow, 5.1.1 Type of sensor (pH/ORP), 5.1.2.1 Parameter, 5.1.2 Sensor check off/on, 5.1.2.2 Temp. sensor yes/no, 5.1.3.1 Temperature, 5.1.3 Temp. corr./Default Temp., 5.1.3.2 Temp. compensation, 5.1.40 (only for pH) Standards pH, 5.1.5 Standard 1 (buffer table), 5.1.5.1 Standard ORP, 5.1.5 Standard 2 (buffer table), 5.1.5.2 Signal output 1, 5.2.1 Signal output 2, 5.2.2 Signal output 3, 5.2.3 (only if ordered) pH or ORP Parameter, 5.2.1.1 Temp. Current loop, 5.2.1.2 flow Function, 5.2.1.3 depends on Funct. Scal./Contr.par. Alarm high Alarm low Hysteresis Alarm pH/ORP, Delay 5.3.1.1 Sample flow, 5.3.1.3 Installation Alarm relay, 5.3.1 (Main menu 5) Sample Temp, 5.3.1.40 Flow al. high Flow al. low Temp al. high Temp al. low (only with temp. sensor) Case Temp. high, 5.3.1.5 Case Temp. low, 5.3.1.6 Relay contacts, 5.3 Miscellaneous, 5.4 Interface, 5.5 Relay 1/2, 5.3.2-3 Function, 5.3.2.1 Parameter, 5.3.2.2 Limit/Set contr. depends on Funct. Input, 5.3.4 Active, 5.3.4.1 Signal output, 5.3.4.2 Outp./Contr., 5.3.4.3 Fault, 5.3.4.4 Delay, 5.3.4.5 Language, 5.4.1 Set Default, 5.4.2 Load firmware, 5.4.3 Password, 5.4.4 Sample ID, 5.4.5 Protocol, 5.5.1 Baud rate, 5.5.2 all menues (only with option RS485) Menu Installation: Defining assignment of all inputs and outputs, measuring parameters, interface, passwords, etc. Menu for the system engineer. Password strongly recommended 32 A-96.250.441 / 060923 AMU pH-Redox Appendix A.2 Program List and Explanations 1 Messages 1.1 Pending errors Provides list of active errors and status (active, acknowledged). If an active error is acknowledged with the enter key, the alarm relay opens again. Cleared errors are moved to the Message list. 1.2 Message Shows error history: Error code, date and time of issue, and status (active, acknowledged, cleared). Maximal 65 error codes are memorized. Then the oldest error is discharged to save the latest error (FIFO first in first out). 2 Diagnostics In diagnostics mode, the values can only be viewed, not modified. 2.1 Identification 2.1.1 View the Type of instrument: Always AMU pH/Redox (ORP) 2.1.2 Informs about the Version of firmware of instrument, e.g. V3.50 2.1.3 Read the date of the factory test of the instrument, mother board, front-end (sensor board) (can not be changed) 2.1.4 Reveals the Operating time in years, days, hours, minutes, seconds (can not be changed) 2.2 Sensors Review the diagnostic values of the photometer and pH electrode. 2.2.1 Sensor Current value: Indicates the actual disinfectant value in pH respectively (for ORP) mV Raw value: Read the actual electrode signal in mV. Calibration history: Informs about the diagnostic values of the electrode of last calibrations: pH: Offset in mV and slope in mV/pH. Typical offset of pH electrode: < +/- 30 mV. Max. tolerated offset <+/- 60 mV Typical slope of pH electrode: 55-65 mV/pH unit. A-96.250.441 / 060923 33 AMU pH-Redox Appendix Max. limits: 40-65 mV/pH. ORP: Offset in mV. <+/- 200 mV 2.2.3 Miscellaneous Case Temp.: Read the actual temperature of electronics housing. 2.3 Sample 2.3.1 Sample ID Sample ID: Review the programmed code. The code is defined by the user to identify the sample point in the plant. Temp.: Only if temp. sensor has been installed and programmed. Review the actual temperature of the sample in °C and in Ohm. Sample flow: Only with digital sample flow meter. Reveals the actual sample flow in l/h (litres per hour). Sample flow must be above 5 l/h. 2.4 I/O State 2.4.1- 2.4.2 Illustrates the actual status of all in and outputs: Alarm relay: Open or closed Relay 1 and Relay 2: Open or closed Input: Open or closed Signal output 1, 2, 3 (only if option is installed): Actual current in mA. 2.5 Interface Only available if option interface has been installed. Review programmed communication settings, set in Installation. 3 Maintenance 3.1 Calibration In this menu, you can correct the measuring value or calibrate (offset and slope) the electrode. 3.1.1 Process pH/ORP: Correction of electrode. See Calibration, p. 25, for details. 3.1.2 Standard pH/ORP: Calibration of electrode with the standard solution(s) programmed in Installation 5.1.5. See Calibration, p. 25, for details. 3.2 Simulation 3.2.1-3.2.6 To simulate in/outputs: Choose the required relay or signal output with the enter key and press the up/down keys to change the value. After confirming the setting with 34 A-96.250.441 / 060923 AMU pH-Redox Appendix the enter key, the value is simulated by the relay/signal output. As soon as you leave the simulation mode, the values will be reset. At the absence of any key activity the instrument will leave the simulation mode after 20 min. The simulated values will be reset. Alarm relay: Open or closed. Relay 1: Open or closed. Relay 2: Open or closed. Signal output 1: mA value. Signal output 2: mA value. Signal output 3: mA value (only if installed). 3.3 Set Time To adjust date and time. 4 Operation 4.1 Sensors 4.1.1 Filter time constant: Used to flatten noisy signals. The longer the filter time constant, the slower the system reacts to changes in the measuring value. Scalable values: 5-300 sec 4.1.2 Hold after calibration: Delay to enable the instrument to stabilize again after calibration, so no peaks show up on the signal output and no false limit warnings are issued. During calibration and hold after calibration, the signal outputs are frozen (held on last valid value), alarm values, limits are not active. Scalable values: 5 - 6000 sec 4.2 Relay contacts 4.2.1 Alarm relay: Here you reprogram alarm values high and low. Alarm: Alarm high: The measuring value must not rise above the alarm high value, set in this menu. Otherwise the alarm relay closes and E001 is issued. Scalable values: 0.00 - 14.00 pH, -500 to 1500 mV. Alarm low: If the measuring value falls below the alarm low value, set in this menu, the alarm relay closes and E002 is issued. Scalable values: 0.00 - 14.00 pH, -500 to 1500 mV. A-96.250.441 / 060923 35 AMU pH-Redox Appendix Hysteresis: Within the hysteresis area, the relay does not open if the alarm has been active. Used to prevent unnecessary relay switching (damage), which occurs if the measuring value fluctuates around the alarm value. Scalable values: 0.00 - 14.00 pH, -500 to 1500 mV. Delay: The time, the activation of the relay is retarded after the measuring value has risen above/fallen below the programmed alarm. Scalable values: 0-28800 sec. 4.2.2 and Relay 1 and 2: 4.2.3 Here you reprogram the limit value or set-point of control, or redefine the timer function. The available parameters depend on presetting in Installation. Limit upper/lower: Set-point: If the measuring value rises above (upper limit) respectively falls below (lower limit) the set-point, the relay closes. Hysteresis: Within the hysteresis area, (below limit upper respectively above limit lower), the relay does not open if the limit has been exceeded. Used to prevent unnecessary relay switching (damage), which occurs if the measuring value fluctuates around the limit value. Delay: The time, the activation of the relay is retarded after the measuring value has risen above/ fallen below the limit value. Control: To maintain a certain set-point in the system. Control upwards: Rises the measuring value to the set-point, assisted by a control device. E.g. dosing alcaline solution with a pump to rise pH value. Control downwards: Lowers the measuring value with the help of a control device, e.g. dosing acid with a pump into a swimming pool. Set-point: Program the required value which should be maintained in the system. P-Band: Area, above (downwards control) or below (upwards control) the set-point, where the dosing intensity is reduced from 100% to 0%, to equal set-point without overshooting. Timer: Relay closes as programmed, e.g. to issue the cleaning of the instrument. Mode: Set how often the relay should close. Choose between interval or daily or weekly If interval is chosen: Choose the time interval. Scalable values: 1 1440 min. If daily is chosen: The contact closes once per day. Set the starting time. If weekly is chosen: Program the starting time and the active days. The timer function starts on all days set to On. Run Time: For all modes. Time, how long the relay closes. Scalable values: 5 - 6000 s 36 A-96.250.441 / 060923 AMU pH-Redox Appendix Delay: For all modes. Signal outputs and control are on cont./hold/off (as programmed below) during run time + delay time. Scalable values: 0 - 6000 s Signal output: Choose between continuous or hold or off. Continuous: Signal outputs are operating continuously, the status display indicates RUN Hold: Signal outputs are frozen on the last valid value. Errors are not issued, except fatal errors, which are always issued. The status display indicates HOLD. Off: Signal outputs go to 0 respectively 4 mA. Errors are not issued, except fatal errors, which are always issued. The status display indicates OFF. Output (relay) / control (relay or signal output): Choose between continuous or hold or off. Continuous: Output/control are operating continuously. Hold: Limit/control are frozen on last valid value. The state of the limit/ control is not updated according to the actual measuring value, as long as the input is active (closed/opened) plus delay time. Reset time and derivative time of the controller are ignored. Off: Limits/control (signal output or relays 1+2) are interrupted. Errors are issued. Fieldbus Contact can be activated/deactivated by fieldbus. No parameter setting required. 4.2.4 Input: Define the status of signal outputs, relay contacts, and the delay if the input contact is activated. The setting depends on preprogramming in Installation. The status display shows the status of signal outputs. Active: Choose between open or close. Open: If the input is opened, the signal outputs and relays are on hold/ off/continuous as programmed below. Close: If the input is closed, the signal outputs and relays are on hold/ off/continuous as programmed below. Signal output: Choose between continuous or hold or off. Continuous: Signal outputs are operating continuously, the status display indicates RUN Hold: Signal outputs are frozen on the last valid value. The measurement is interrupted. Errors are not issued, except fatal errors, which are always issued. The status display indicates HOLD. Off: Signal outputs go to 0 respectively 4 mA. Errors are not issued, except fatal errors, which are always issued. The status display indicates OFF. A-96.250.441 / 060923 37 AMU pH-Redox Appendix Output (relay) / control (relay or signal output): Choose between continuous or hold or off. Continuous: Output/control are operating continuously. Hold: Limit/control are frozen on last valid value. The state of the limit/ control is not updated according to the actual measuring value, as long as the input is active (closed/opened) plus delay time. Reset time and derivative time of the controller are ignored. Off: Limits/control (signal output or relays 1+2) are interrupted. Errors are issued. Fault: Choose between yes or no. Fault no: No error message is issued when input is active (closed/ opened). Alarm relay does not close. Fault yes: Error E024 Input activated is issued. Alarm relay closes. Delay: The time the input deactivation (programmed above) is retarded after opening/closing the input. Scalable values: 0 - 6000 s 4.3 Logger The instrument is equipped with an internal logger which can be down loaded by interface, using the program Hyperterminal. The logger can save approximately 1500 data records. One data record consists of: Date, time, alarms, measuring value, temp., flow. 4.3.1 Log Interval: Choose the required logging interval. The table below gives the max. time interval, the logger can save. When the logger capacity is exhausted, the oldest data record is discharged to save the new data record (FIFO first in first out). Interval 1s 5s 1 min 5 min 10 min 30 min 1h Time 25 min 2h 25 h 5d 10 d 31 d 62 d 4.3.2 Clear logger: If confirmed with yes, the complete logger data is deleted. A new data series is started. 38 A-96.250.441 / 060923 AMU pH-Redox Appendix 5 Installation 5.1 Sensors 5.1.1 Flow: Define if the Swan flow meter is active or not. Choose between: None or Q-flow (with flow meter). 5.1.2 Parameter: Choose the type of sensor: pH or redox (ORP) Sensor check: not yet implemented. Set to off. 5.1.3 Temperature: Temp. sensor: Program if a temperature sensor Pt1000 has been connected. Choose between no or yes. If no, the Default Temp. is asked. Set the temp. of the sample. If yes, you can set a Temp. correction, to compensate errors of displayed value. 5.1.4 Temperature compensation: Only for pH. Available settings: Nernst/non-linear/coefficient. For potable water, waste water, swimming pools: Choose Nernst High purity water: Additionally to Nernst compensation you can choose between non-linear solution compensation or linear compensation with coefficient (pH/°C). 5.1.5 Standards: Here you program the standard solution(s) for the calibration of the electrode. pH: Enter the values of your buffer into the buffer table (exact buffer values for temperature 0° - 50°C) ORP: Enter the value of the standard solution. 5.2 Signal outputs 5.2.1 - 5.2.3 Signal output 1, Signal output 2, Signal output 3: Signal output 3 is only available if option 3rd signal output has been installed Here you assign a certain measuring value, current loop and function to the signal outputs. 5.2.1.1 - 5.2.3.1Parameter: Assign one of the measuring values to the signal output. Available settings: Measuring value/temperature/flow 5.2.1.2 - 5.2.3.2Current loop: Choose the current loop of the signal output. Make sure the connected device works with the same current loop. Available settings: 0-20 mA or 4-20 mA. 5.2.1.3 - 5.2.3.3Function: Define if the signal output should represent the actual measuring value or if it is used for control. A-96.250.441 / 060923 39 AMU pH-Redox Appendix If it should represent the actual measuring value you can choose between linear, bilinear or logarithmic scaling. See explanations below. Function lin/Shows one of the measuring values as current loop 0-20 mA/4-20 mA. bilinear:Linear: Program range low and range high. Range low: Choose the beginning of the scale. This value corresponds to 0 respectively 4 mA. If you have chosen 0-20 mA it is 0 mA, if you have chosen 4-20 mA, it is 4 mA. It is the lowest value represented by the signal output. Range high: Choose the end of the scale. This value corresponds to 20 mA. It is the highest value illustrated by the signal output. Bilinear: The beginning or end of the selected range can be scaled in a higher resolution, resulting in a semi-logarithmic presentation. Program range low, range high, range centre. Range low and high see above: Linear Range centre: Fixed on 10 respectively 12 mA. See also graphic below. mA 20 o res bilinear 0-20 mA: 10 4-20 mA: 12 0 ol. es 0 .00 6 .00 0 2 r 1 .00 l. 0 res ol. 0 linear 0/4 2 3 4 5 6 7 8 unit meas. value FunctionThe scaling is logarithmic. Program range high and range low. Log: Loga-Range low: Choose the beginning of the scale. This value corresponds rithmic scaleto 0 respectively 4 mA. If you have chosen 0-20 mA it is 0 mA, if you have chosen 4-20 mA, it is 4 mA. Range high: Choose the end of the scale. This value corresponds to 20 mA. ControlControl upwards or control downwards are available. For further explanations about control and control parameters (set-point/p-band) see 5.3.2.1 If the measuring value is out of the p-band, 20 mA are issued. Within 40 A-96.250.441 / 060923 AMU pH-Redox Appendix P-Band the mA value is reduced. If the measuring value equals setpoint, control is switched off, resulting in 0 mA (current loop 0-20 mA), or 4 mA (current loop 4-20 mA.) 5.2.1.4x x no. depends on chosen function of signal output: 5.2.3.4x x = 0=lin, 1=bilin, 2=log, 3=control 5.2.x.40 - Scaling: 5.2.x.42 Choose range low, range high, range centre (if applicable). For explanations see before: 5.2.1.3 Function. Available values: pH Range low/centre/high: -3.00 - 15.00 pH ORP Range low/centre/high: -500 - 1500 mV Temp. Range low/centre/high: -25.0 - 250.0 °C Flow Range low/centre/high: 0 - 50 l/h 5.2.1.43 Control: Program set-point and p-band. Please consult 5.3.2.1 for explanations and ranges. 5.3 Relay contacts 5.3.1 Alarm relay: Here you set the alarm values for the measuring value, temperature, sample flow, case temperature. 5.3.1.1 Alarm measuring value: Define the measuring value, which should issue an alarm high respectively low. Alarm high: The measuring value must not rise above the alarm high value, set in this menu. Otherwise, the alarm relay closes and E001 is issued. Scalable values: -3.00 - 15.00 ppm, -500 - 1500 mV Alarm low: If the measuring value falls below the set alarm value, the alarm relay closes and E002 is issued. Scalable values: -3.00 - 15.00 ppm, -500 - 1500 mV Hysteresis disinfectant: Within the hysteresis area, the relay does not open if the alarm has been active. Used to prevent unnecessary relay switching (damage), if the measuring value fluctuates around the alarm value. Scalable values: -3.00 - 15.00 ppm, -500 - 1500 mV Delay: The time, the closing of the alarm relay is retarded after the measuring value has risen above/fallen below the alarm value. Scalable values: 0-28800 sec See also graphic below. A-96.250.441 / 060923 41 AMU pH-Redox Appendix Delay 20 sec Hysteresis 0.05 0 Alarm high 0.40 red asu me relay open alarm off relay closed alarm on relay open alarm off 5.3.1.3 Alarm sample flow: Here you define at which sample flow a flow alarm should be issued and if the alarm relay should close. Alarm flow high: If measuring value rises above the programmed value, E009 flow alarm high will be issued. Scalable values: 10-50l/h Alarm flow low: If the measuring value falls below the programmed value, E010 flow alarm low will be issued. Scalable values: 0-9 l/h 5.3.1.4 Alarm Sample Temperature: Only if temp. senor has been connected. Alarm high: If the temperature rises above the programmed value, E007 is issued. Scalable values: -25 - 250 °C Alarm low: If the temperature falls below the programmed value, E008 is issued. Scalable values: -25 - 250 °C 5.3.1.5 Case temp. high: Set the alarm high value for temperature of electronics housing. If the temperature rises above the programmed value, E013 is issued. Scalable values: 30 to 75 °C 5.3.1.6 Case temp. low: Define the alarm low value for temperature of electronics housing. If the measuring value falls below the programmed value, E014 is issued. Scalable values: -10 to +20 °C 5.3.2 and Relay 1 and Relay 2 5.3.3 Program the function of the relays in detail. 5.3.2.1 and Function: Choose if the relay should 5.3.3.1 - close if the measuring value rises above/falls below the set limit, - if it should drive a control device (pump, motor valve, etc.) to maintain a certain set-point, - if it is used as timer (e.g. adding sensor cleaning solution), - if it should be activated by Profibus DP. Available settings: Limit upper/lower/control upwards/control downwards/timer/fieldbus 42 A-96.250.441 / 060923 AMU pH-Redox Appendix Limit upper/ See also graphic on page 63. lower: Set-point: If the measuring value rises above (limit upper) respectively falls below (limit lower) the set limit, the relay closes. Hysteresis: Within the hysteresis area (below limit upper respectively above limit lower), the relay does not open if the limit has been active. Used to prevent unnecessary relay switching (damage), which occurs if the measuring value fluctuates around the limit value. Delay: The time, the closing of the relay is retarded after the measuring value has risen above/fallen below the limit value. Control To maintain a certain set-point in the system. Control upwards: Rises the measuring value to the set-point, assisted by a control device. Control downwards: Lowers the measuring value to the set-point aided by a control device. Control parameters PID Set-point: Program the required value which should be maintained constant in the system. P-Band: Area, above (downwards control) or below (upwards control) the set-point, where the dosing intensity is reduced from 100% to 0%, to equal set-point without overshooting. See also graphic on page 53 PI controller: Reset time: The controller action increases the higher the deviation from the set-point is and the longer it lasts. PD controller: Derivative time: The controller action increases the higher the deviation from the set-point is and the faster the measuring value changes. Control If the controller action (dosing intensity) is constantly over 90% during time-out the control time-out (in min.), without the measuring value changing into the direction of set-point, dosing will be stopped for safety reasons, and an alarm is issued. Control Time The control device is driven by switching on/off its power supply. Define proportional a time interval (cycle time) as time base for control. To reduce the dosing intensity in the p-band, the pump runs only part-time of the cycle. Cycle time: Defines in which interval the pump is switched on. Response time: To avoid stressing the control device. Minimum time, the dosing unit has to run to avoid damage, or that elapses from the activation until the actual dosing starts. See also graphic below. Control by The pump is driven by adjusting the number of pulses. The relay frequency closes/opens in the pulse frequency. A-96.250.441 / 060923 43 AMU pH-Redox Appendix Frequency: Program the number of pulses/min. the pump can handle. Within the p-band the pulse number is reduced = dosing is reduced. Control Only one motor valve can be connected, because both relays are motor valve needed to work the valve. Relay 1 opens, and relay 2 closes the valve. Program the run time and neutral zone of your motor valve. Run time: Time, the motor valve needs to close/open completely. Neutral zone: Within the neutral zone the controller does not react to deviations. Unnecessary switching (damage) is avoided. Limit Delay 20 sec Hysteresis pH/mV m Limit upper pH/mV ue val red u s ea 0 relay open alarm off Time proportional control relay open alarm off relay closed alarm on unit set-point p-band 1. cycle time = 60s 2. cycle = 60s pump on p-band 60 s out of p-band 60 s 3. cycle = 60s 4. cycle = 60s 30 s 15 s 50% in p-band 75% in p-band time (s) Timer Relay closes as programmed, e.g. to issue cleaning of the instrument. Mode: Set how often the relay should close. Available settings: interval/daily/weekly 44 A-96.250.441 / 060923 AMU pH-Redox Appendix If interval is chosen: Program the time interval. Scalable values: 1 - 1440 min. If daily is chosen: The contact closes once per day. Choose the starting time. If weekly is chosen: Define the starting time and the active days. The timer function starts on all days set to On. Run Time: For all modes. Time how long the relay closes. Scalable values: 5 - 6000 s Delay: For all modes. Signal outputs and control are on cont/hold/off (as programmed below) during run time + delay time. Scalable values: 0 - 6000 s Signal output: Choose between continuous or hold or off. Continuous: Signal outputs are operating continuously, the status display indicates RUN Hold: Signal outputs are frozen on the last valid value. Errors are not issued, except fatal errors, which are always issued. The status display indicates HOLD. Off: Signal outputs go to 0 respectively 4 mA. Errors are not issued, except fatal errors, which are always issued. The status display indicates OFF. Outp.(relay)/contr.(relay or signal output): Choose between continuous or hold or off Continuous: Output/control are operating continuously. Hold: Frozen on last valid value. The state of the limit/control is not updated according to the measuring value, as long as the input is closed (+ during delay time). Reset time and derivative time of the controller are ignored. Off: Limits/control (signal output or relay) are interrupted. Errors are issued. Fieldbus Contact can be activated/deactivated by fieldbus. No parameter setting required. 5.3.2.2x and x no. depends on chosen function of the relay: 0=limit upper; 1=limit 5.3.3.2x lower; 2=control upwards; 3=control downwards; 4=timer A-96.250.441 / 060923 45 AMU pH-Redox Appendix Overview: 5.3.2.1 5.3.3.1 0/1 Limit upper/lower 2/3 Control upw./ downw. 4 Timer 5.3.2.2 5.3.3.2 Parameter Parameter Mode 5.3.2.3 5.3.3.3 Limit value Set-point Interval/Start time/ Calender 5.3.2.4 5.3.3.4 Hysteresis Control settings Run time 5.3.2.5 5.3.3.5 Delay Control parameters Delay 5.3.2.6 5.3.3.6 Signal outputs cont./hold/off 5.3.2.7 5.3.3.7 Output/control cont./hold/off 5.3.2.2(0-3) and Parameter: 5.3.3.2(0-3) Assign one measuring value to the relay (with chosen function). Available settings: Measuring value/temp./flow 5.3.2.24 and Mode of timer: 5.3.3.24 Choose how often the contact should close. Available settings: interval/ daily/weekly. For explanations see 5.3.2.1. 5.3.2.3xy and x no. depends on chosen function of relay: 0=limit upper; 1=limit lower; 5.3.3.3xy 2=control upwards; 3=control downwards; 4=timer y no. depends on chosen parameter OR mode of timer: For parameter: 0=measuring value; 1=Temp.; 2=sample flow For mode of timer: 0=interval; 1=daily, 2=weekly 5.3.2.3x(0-3) Limit value or Set-point: 5.3.3.3x(0-3) Choose the required limit/set-point. For explanations see 5.3.2.1. x=0,1,2,3 Scalable values depend on chosen parameter: pH: 0.00 -14.00 pH ORP: -500 - 1500 mV 5.3.2.34(0-2) Depends on chosen mode of timer: 0=interval; 1=daily; 2=weekly 5.3.3.34(0-2) 40: Interval in minutes 41: Start time: 00:00:00 46 A-96.250.441 / 060923 AMU pH-Redox Appendix 42: Calendar: Starting time and Mo-Su (Off/On) For explanations see 5.3.2.1 5.3.2.4xy and x no. depends on chosen function of relay = 0=limit upper; 1=limit 5.3.3.4xy and lower; 2=control upwards; 3=control downwards; 4=timer 5.3.2.5xy and y no. depends on chosen parameter. 5.3.3.5xy For parameter: 0=measuring value; 1=Temp.; 2=sample flow 5.3.2.40y or 41y Limit: Hysteresis: 5.3.3.40y or 41y Explanations see 5.3.2.1. Scalable values: pH: 0.00 -14.00 pH ORP: -500 - 1500 mV Limit Delay: Explanations see 5.3.2.1. Scalable values: 0-28800 s 5.3.2.42y or 43y Control: Control settings: 5.3.3.42y or 43y Explanations see 5.3.2.1. Actuator: Time prop.: Scalable values: Cycle time: 10 - 600 s, Response time: 0 - 240 s Actuator: Frequency: Scalable values: Frequency: 20 - 300/min. Actuator: Motor valve: Scalable values: Run time: 5 - 300 s, Neutral zone: 1 - 20% 5.3.2.52y or 53y Control: Control parameters: 5.3.3.52y / 53y: Explanations see 5.3.2.1. P-Band: Scalable values depend on setting in parameter: 0.00 - 14.00 pH or -500 - 1500 mV Reset time: If 0 s is chosen the function is switched off. Scalable values: 0 - 9000 s Derivative time: If 0 s is chosen the function is switched off. Scalable values: 0 - 9000 s Control time-out: If 0 min. is chosen, the function is switched off. Scalable values: 0 - 720 min. 5.3.2.44 or Timer: Run Time in seconds. Time the relay closes. 5.3.3.44 Scalable values: 5 - 6000 s 5.3.2.54 or Timer: Delay in sec. Time the signal outputs and output/control are on 5.3.3.54 cont./hold/off after deactivation of input. Available values: 0 - 6000 s 5.3.2.6 or Timer: Signal outputs: Define the status of the signal outputs 5.3.3.6 (programmed as current loop for measuring value) during run time (5.3.2.44) and delay (5.3.2.54): cont./hold/off. For explanations see 5.3.2.1. A-96.250.441 / 060923 47 AMU pH-Redox Appendix 5.3.2.7 andTimer: Output (relay)/control: Define the status of relays (and signal 5.3.3.7output, programmed as control) during run time and delay: cont./hold/ off. For explanations see 5.3.2.1. 5.3.4 Input Here you set the function of the input. The status display indicates the status of signal outputs 5.3.4.1 Active: Choose between open or close. Open: If the input is opened, the signal outputs, relays, and error relay are on hold/off/continuous as programmed below. Close: If the input is closed, the signal outputs, relays, and error relay are on hold/off/continuous as programmed below. 5.3.4.2 Signal output: Choose between continuous or hold or off. Continuous: Signal outputs are operating continuously, the status display indicates RUN Hold: Signal outputs are frozen on the last valid value. Errors are not issued, except fatal errors, which are always issued. If a non-fatal error is active when Input is closed, the error is frozen during Input closing time+delay. The status display indicates HOLD. Off: Signal outputs go to 0 respectively 4 mA. Errors are not issued, except fatal errors, which are always issued. If a non-fatal error is active when Input is closed, the error is frozen during Input closing time+delay. The status display indicates OFF. 5.3.4.3 Outp.(relay)/contr.(relay or signal output): Choose between continuous or hold or off Continuous: Output/control are operating continuously. Hold: Frozen on last valid value. The state of the limit/control is not updated according to the measuring value, as long as the input is closed (+ during delay time). Reset time and derivative time of the controller are ignored. Off: Limits/control (signal output or relay) are interrupted. Errors are issued. 5.3.4.4 Fault: Choose between yes or no. Fault no: No error message is issued when input is closed. Alarm relay does not close. Fault yes: Error E024 Input activated is issued. Alarm relay closes. 5.3.4.5 Delay: The time the input deactivation is retarded after opening/closing the input. Scalable values: 0 - 6000 s. 48 A-96.250.441 / 060923 AMU pH-Redox Appendix 5.4 Miscellaneous 5.4.1 Language: Set the required language. Available settings: Deutsch/english/français/español, default: English 5.4.2 Set default: If you set default, all programmed values will be set back to factory default values. See Default Values, p. 50. Available settings: No/calibration/in parts/completely. Calibration: Sets calibration values back to default. All other values are kept. In parts: Communication parameters are kept. All other values are set back to default values. Completely: Sets back all values including communication parameters. 5.4.3 Load firmware: Only for service personnel. For update of firmware. Not to be done by the user. Always keep No. 5.4.4 Password You can program passwords for all menus, except Diagnostics: Messages, Maintenance, Operation, Installation The passwords are 4-digit codes. If 0000 is programmed the function is switched off and no password is asked. If you program a different code, it will be asked, whenever you want to enter the menu. Every menu point may have a different password. Please make sure you do not forget the passwords. If you do, your dealer can help you. 5.4.5 Sample ID: Here you can set a code to identify the sample point in the plant. The code is defined by the user. 5.5 Interface 5.5.1 Protocol: Choose the required protocol: Available settings: Profibus, Modbus RTU, SwanBus, Hyperterminal, Web Server Please see separate manual how to perform a logger download by Hyperterminal. 5.5.2x x no. depends on chosen protocol in 5.5.1. 0=Profibus; 1=Modbus RTU; 2=SwanBus; 3=HyperTerminal; 4=WebServer 5.5.20 Profibus Device Address: Address of instrument A-96.250.441 / 060923 49 AMU pH-Redox Appendix 5.5.21 Modbus Device Address: Program address of instrument 5.5.22 SwanBus Device Address: Program address of instrument 5.5.23 Hyper Terminal Baud rate: Select transmission speed in baud. 5.2.24 Web Server Device address: Program address of instrument 5.5.3x x no. depends on chosen protocol in 5.5.1. 0=Profibus; 1=Modbus RTU; 2=SwanBus; 3=HyperTerminal; 4=WebServer 5.5.30 Profibus ID No.: Choose between Manufacturer/Analyzer/Multivariable 5.5.31 Modbus Baud rate: Select transmission speed in baud. 5.5.32 SwanBus Baud rate: Select transmission speed in baud. 5.5.4x x no. depends on chosen protocol in 5.5.1. 0=Profibus; 1=Modbus RTU; 2=SwanBus; 3=HyperTerminal; 4=WebServer 5.5.40 Profibus Local operation: Disabled/Enabled 5.5.41 Modbus Parity: None/Even/Odd Default Values Operation: Sensors:Filter time constant: 30 s Hold after calibration: 300 s Signal output same as in Installation 1-3: Relay 1, Relay same as in Installation 2, Input: Logger Interval: 30 min. Installation: Sensor: Flow: None Parameter: Type of sensor: pH Parameter Sensor check:off Temperature sensor: No Temperature sensor no: Default temp.: 25 °C Temperature sensor yes: Temp. Correction 0.00 °C 50 A-96.250.441 / 060923 AMU pH-Redox Appendix pH standard 1 / 2: Buffer table 0°C: 7.13 / 9.24 pH 5°C: 7.07 / 9.16 10°C: 7.05 / 9.11 15°C: 7.02 / 9.05 20°C: 7.00 / 9.00 25°C: 6.98 / 8.95 30°C: 6.97 / 8.91 35°C: 6.96 / 8.88 40°C: 6.95 / 8.85 50°C: 6.95 / 8.79 Standard Redox (ORP): 475 mV Signal outputs: Parameter: measuring value Current loop: 4-20 mA Function: lin Scaling: Range low: 0.00 pH Range high: 14.00 pH Salient point: 7.00 pH Salient point is always in the middle= switched off (lin scale). Temperature range low: 0.0 °C Temperature range high: 50.0 °C Flow range low: 0 l/h Flow range high: 50 l/h Alarm relay: pH alarm high: 15.00 pH pH alarm low: -3.00 pH Hysteresis: 0.10 pH Delay: 5 s Sample flow alarm high: 20 l/h Sample flow alarm low: 5 l/h Sample temp. alarm high: 55 °C Sample temp. alarm low: 5 °C Case temp. high: 65 °C Case temp. low: 0 °C Relay1/2 Function: limit upper Parameter: Relay 1: measuring value, Relay 2: measuring value Limit: Relay 1: 14.00 pH, Relay 2: 14.00 pH Limit Hysteresis: 0.1 pH Limit Delay: 30 s Control: Actuator: Frequency Control: Frequency: Frequency: 120/min. Control: Time prop.: Cycle time: 60s, Response time: 10 s Control: Motor valve: Run time: 60s, Neutral zone: 5% A-96.250.441 / 060923 51 AMU pH-Redox Appendix Control: Control parameter P-band: 0.10 pH Control: Control parameter Reset time: 0 s Control: Control parameter Derivative time: 0 s Control: Control parameter Control time-out: 0 min. Timer: Mode: Interval: 1 min. Timer: Mode: daily/weekly: Starting time: 00.00.00 Timer: Run time: 10s Timer: Delay: 5s Timer: Signal output: cont. Timer: Output/Control: cont. Input Active: when closed Signal output: hold Output/control: off Fault: no Delay: 10 s Miscellaneous Language: English Set default: no Load firmware: no Password: for all modes 0000 Sample ID: -Interface 52 Protocol: Hyperterminal A-96.250.441 / 060923 AMU pH-Redox Material Safety Data Sheets B. Material Safety Data Sheets B.1 Buffer pH 4 According to EC Directive 91/155/EWG 1 Identification of the substance/preparation and of the company/undertaking Identification of the product: Catalogue No.: A-85.112.300, A-85.112.500, A-85.112.700 Product name: Calibration Solution pH 4 Manufacturer/supplier identification: Company: SWAN Analytische Instrumente AG, CH - 8616 Riedikon, Switzerland Tel.: +41 1 943 63 00, Fax: +41 1 943 63 01 2 Composition/information on ingredients Synonyms: CAS-No.: Molar mass: Molecular formula: EINECS-No.: Potassium Hydrogen Phthalate, means of preservation and color 877-24-7 (Potassium Hydrogen Phthalate, not diluted) 204.23 g/mol (Potassium Hydrogen Phthalate, not diluted) C8H5KO4 (Potassium Hydrogen Phthalate, not diluted) 2128894 (Potassium Hydrogen Phthalate, not diluted) Hazardous ingredients: Name according to EC Directives: None 3 Hazards identification No hazardous product according to the directive 67/548/EWG. 4 First aid measures After inhalation: After skin contact: After eye contact: After ingestion: 5 -wash off with plenty of water. rinse out with plenty of water with the eyelid held wide open. make victim drink plenty of water, summon doctor if feeling unwell. Fire-fighting measures Suitable extinguishing media: 6 -- Special risks: -- Other information: non-combustible Accidental release measures: Procedures for cleaning / absorption: Dilute with water and take-up with liquid-absorbent material. Environmental-protection measures: None A-96.250.441 / 060923 53 AMU pH-Redox Material Safety Data Sheets 7 8 Handling and storage: Handling: no further requirements Storage: tightly closed. Dry. In a well-ventilated place. At room temperature . No public product. Keep out of children's reach. Exposure controls/personal protection Personal protective equipment: Respiratory protection: not required. Eye protection: not required. Hand protection: not required Industrial hygiene: Contact with eye or skin: Wash with plenty of water. 9 10 11 Physical and chemical properties Form: Color: Odor: liquid reddish odorless pH value: (20 oC) 4 Melting temperature: Boiling temperature: not available near 100 oC Ignition temperature: Flash point: not available not available Explosion limits: lower upper not available not available Density: (20 oC) 1.1 g/cm3 Solubility in water: (20 oC) very good and completely soluble Stability and reactivity Conditions to be avoided: -- Substances to be avoided: -- Hazardous decomposition products: none Toxicological information Acute toxicity: No toxicological data available Further toxicological information: After skin contact: -After eye contact: -After ingestion: -- 54 A-96.250.441 / 060923 AMU pH-Redox Material Safety Data Sheets 12 Ecological information Ecotoxic effects: -Further ecologic data: None 13 Disposal considerations Product: There are no uniform EC regulations for the disposal of chemicals or residues. Chemical residues generally count as special waste. The disposal of the latter is regulated in the EC member countries through corresponding laws and regulations, and in Switzerland also by the cantons. We recommend that you contact either the authorities in charge or approved waste disposal companies which will advise you on how to dispose of special waste. Packaging: Disposal in compliance with official regulations. Handle contaminated packaging in the same way as the substance itself. If not officially specified differently, non-contaminated packaging may be treated like household waste or recycled. 14 Transport information Not subject to transport regulations. 15 Regulatory information Labeling according to EG Directives Symbol: -Name: -R-phrases: -- S-phrases: -- Other national regulations: Swiss toxic class: None 16 Other information Date of issue: 04.03.2005 The information contained herein is based on the present state of our knowledge. It characterizes the product with regard to the appropriate safety precautions. It does not represent a guarantee of the properties of the product. A-96.250.441 / 060923 55 AMU pH-Redox Material Safety Data Sheets B.2 Buffer pH 7 1 Identification of the substance/preparation and of the company/undertaking Identification of the product: Catalogue No.: 85-113300 or 85-113350 Product name: Calibration solution pH 7 Use of substance/preparation: Buffer solution, no public product, only for laboratory use Manufacturer/supplier identification: Company: Swan Analytical Instruments AG, CH - 8616 Riedikon Tel.: 01/943 63 00, Fax: 01/943 63 01 2 Composition/information on ingredients Synonyms: CAS-No.: Molar mass: Molecular formula: EINECS-No.: Phosphate buffer solution ----- Hazardous ingredients: Name according to EC Directives: None CH: not toxic class 3 Hazards identification No hazardous product according to the directive 67/548/EWG. 4 First aid measures After inhalation: After skin contact: After eye contact: After ingestion: 5 6 -wash off with plenty of water. rinse out with plenty of water with the eyelid held wide open. Summon eye specialist. make victim drink plenty of water, summon doctor if feeling unwell. Fire-fighting measures Suitable extinguishing media: -- Special risks: -- Other information: non-combustible Accidental release measures: Procedures for cleaning / absorption: Dilute with water and take-up with liquid-absorbent material. Environmental-protection measures: None 56 A-96.250.441 / 060923 AMU pH-Redox Material Safety Data Sheets 7 Handling and storage: Handling: Storage: public 8 no further requirements tightly closed. Dry. In a well-ventilated place. At room temperature . No product. Keep out of children's reach. Exposure controls/personal protection Personal protective equipment: Respiratory protection: not required. Eye protection: not required. Hand protection: not required Industrial hygiene: Contact with eye or skin: Wash with plenty of water. 9 Physical and chemical properties Form: Colour: Odour: liquid yellow odourless pH value: (20 oC) not available near 100 oC Ignition temperature: Flash point: not available not available Explosion limits: 10 11 7 Melting temperature: Boiling temperature: lower upper not available not available o Density: (20 C) 1.1 g/cm3 Solubility in water: (20 oC) very good and completely soluble Stability and reactivity Conditions to be avoided: -- Substances to be avoided: -- Hazardous decomposition products: none Toxicological information Acute toxicity: No toxicological data available Further toxicological information: After skin contact: -After eye contact: -After ingestion: -- A-96.250.441 / 060923 57 AMU pH-Redox Material Safety Data Sheets 12 Ecological information Ecotoxic effects: -Further ecologic data: None 13 Disposal considerations Product: There are no uniform EC regulations for the disposal of chemicals or residues. Chemical residues generally count as special waste. The disposal of the latter is regulated in the EC member countries through corresponding laws and regulations, and in Switzerland also by the cantons. We recommend that you contact either the authorities in charge or approved waste disposal companies which will advise you on how to dispose of special waste. Packaging: Disposal in compliance with official regulations. Handle contaminated packaging in the same way as the substance itself. If not officially specified differently, non-contaminated packaging may be treated like household waste or recycled. 14 Transport information Not subject to transport regulations. 15 Regulatory information Labelling according to EG Directives Symbol: -Name: -R-phrases: -- S-phrases: -- Other national regulations: Swiss toxic class: None 16 Other information Date of issue: 19.05.2003 The information contained herein is based on the present state of our knowledge. It characterizes the product with regard to the appropriate safety precautions. It does not represent a guarantee of the properties of the product. 58 A-96.250.441 / 060923 AMU pH-Redox Material Safety Data Sheets B.3 Buffer pH 9 1 Identification of the substance/preparation and of the company/undertaking Identification of the product: Catalogue No.: 85-114300 or 85-114350 Product name: Calibration solution pH 9 Use of substance/preparation: Buffer solution, no public product, only for laboratory use Manufacturer/supplier identification: Company: Swan Analytical Instruments AG, CH - 8616 Riedikon Tel.: 01/943 63 00, Fax: 01/943 63 01 2 Composition/information on ingredients Synonyms: CAS-No.: Molar mass: Molecular formula: EINECS-No.: Phosphate buffer solution ----- Hazardous ingredients: Name according to EC Directives: None 3 Hazards identification No hazardous product according to the directive 67/548/EWG. 4 First aid measures After inhalation: After skin contact: After eye contact: After ingestion: 5 6 -wash off with plenty of water. rinse out with plenty of water with the eyelid held wide open. Summon eye specialist. make victim drink plenty of water, summon doctor if feeling unwell. Fire-fighting measures Suitable extinguishing media: -- Special risks: -- Other information: non-combustible Accidental release measures: Procedures for cleaning / absorption: Dilute with water and take-up with liquid-absorbent material. Environmental-protection measures: None A-96.250.441 / 060923 59 AMU pH-Redox Material Safety Data Sheets 7 Handling and storage: Handling: Storage: public 8 no further requirements tightly closed. Dry. In a well-ventilated place. At room temperature . No product. Keep out of children's reach. Exposure controls/personal protection Personal protective equipment: Respiratory protection: not required. Eye protection: not required. Hand protection: not required Industrial hygiene: Contact with eye or skin: Wash with plenty of water. 9 10 11 Physical and chemical properties Form: Colour: Odour: liquid blue odourless pH value: (20 oC) 9 Melting temperature: Boiling temperature: not available near 100 oC Ignition temperature: Flash point: not available not available Explosion limits: lower upper not available not available Density: (20 oC) 1.1 g/cm3 Solubility in water: (20 oC) very good and completely soluble Stability and reactivity Conditions to be avoided: -- Substances to be avoided: -- Hazardous decomposition products: none Toxicological information Acute toxicity: No toxicological data available Further toxicological information: After skin contact: -After eye contact: -After ingestion: -- 60 A-96.250.441 / 060923 AMU pH-Redox Material Safety Data Sheets 12 Ecological information Ecotoxic effects: -Further ecologic data: None 13 Disposal considerations Product: There are no uniform EC regulations for the disposal of chemicals or residues. Chemical residues generally count as special waste. The disposal of the latter is regulated in the EC member countries through corresponding laws and regulations, and in Switzerland also by the cantons. We recommend that you contact either the authorities in charge or approved waste disposal companies which will advise you on how to dispose of special waste. Packaging: Disposal in compliance with official regulations. Handle contaminated packaging in the same way as the substance itself. If not officially specified differently, non-contaminated packaging may be treated like household waste or recycled. 14 Transport information Not subject to transport regulations. 15 Regulatory information Labelling according to EG Directives Symbol: -Name: -R-phrases: -- S-phrases: -- Other national regulations: Swiss toxic class: None 16 Other information Date of issue: 19.05.2003 The information contained herein is based on the present state of our knowledge. It characterizes the product with regard to the appropriate safety precautions. It does not represent a guarantee of the properties of the product. A-96.250.441 / 060923 61 AMU pH-Redox Material Safety Data Sheets B.4 Buffer ORP 1 Identification of the substance/preparation and of the company/undertaking Identification of the product: Catalogue No.: 85-121300 / 85-121500 / 85-121700 Product name: Redox Calibration Solution for redox electrodes, 430 mV Pt/Calomel, 475 mV Pt/AgCl Manufacturer/supplier identification: Company: SWAN Analytische Instrumente AG, CH-8616 Riedikon, Switzerland Tel.: +41 1 943 63 00, Fax: +41 1 943 63 01 2 Composition/information on ingredients Synonyms: CAS-No.: Molar mass: Molecular formula: EINECS-No.: Diluted sulphuric acid / iron sulphate solution 7664-93-9 98.08 H2SO4 231-639-5 Hazardous ingredients: Name according to EC Directives: Sulphuric acid Hazard symbol: C R-phrases: 35 EC-Index-No.: 016-020-00-8 Causes severe burns CAS-No.: 7664-93-9 Content: <10% 3 Hazards identification Causes severe burns. 4 First aid measures After inhalation: After skin contact: After eye contact: After ingestion: 5 fresh air. Summon doctor. wash off with plenty of water. Immediately remove contaminated clothing. rinse out with plenty of water with the eyelid held wide open. Immediately summon eye specialist. make victim drink plenty of water, avoid vomiting, immediately summon doctor. Do not attempt to neutralize. Fire-fighting measures Suitable extinguishing media: in adaptation to materials stored in the immediate neighborhood. 62 Special risks: Development of hazardous vapors possible in the event of fire. Hydrogen may form upon contact with metals (danger of explosion!). The following may develop in the event of fire: sulphur oxides. Other information: Non combustible. A-96.250.441 / 060923 AMU pH-Redox Material Safety Data Sheets 6 Accidental release measures: Person-related precautionary measures Do not inhale vapors / aerosols. Procedures for cleaning / absorption: Dilute with lots of water. Neutralize with sodium carbonate. Take up dry. Forward for disposal. Clean up affected area. Environmental-protection measures: Undiluted solution should not be allowed to enter the sewage system in bigger quantities. 7 8 Handling and storage: Handling: No further requirements Storage: Tightly closed. Dry. In a well-ventilated place. At room temperature. No public product. Keep out of children's reach. Exposure controls/personal protection Specific control parameter: MAK Germany (max. workplace conc.) sulphuric acid: 1 mg/m3 total dust Personal protective equipment: Respiratory protection: required when vapors / aerosols are generated. Eye protection: required. Hand protection: required Industrial hygiene: Change contaminated clothing and immerse in water. Protect skin. Wash hands and face after working with substance. 9 Physical and chemical properties Form: Color: Odor: liquid yellow odorless pH value: at 50 g/l H2O: (20 oC) |1 Melting temperature: Boiling temperature: not available ca. 100 oC Ignition temperature: Flash point: not available not available Explosion limits: lower upper not available not available Density: Bulk density: (20 oC) not available 1.1 g/cm3 (20 oC) very good soluble Solubility in water: A-96.250.441 / 060923 63 AMU pH-Redox Material Safety Data Sheets 10 Stability and reactivity Conditions to be avoided: strong heating Substances to be avoided: alkali metals, alkali compounds, ammonia, alkaline earth metals, alkaline earth compounds, acids, bases, metals, metal alloys, combustible substances, organic solvents, halogenate, permanganate. 11 Hazardous decomposition products: toxic gases Further information: corrosive, incompatible with metals Toxicological information Acute toxicity: Toxicological values are not available due to other dangerous properties of the substance. Further toxicological information: Contact with skin causes severe burns under formation of sloughs. Eye contact leads to corneal damage. Ingestion is followed by severe pains (risk of perforation!), nausea, vomiting, and diarrhea, and after a latency period of some weeks possibly pyloric stenosis. Inhalation of aerosols resulting in damage of the mucous membranes. 12 Ecological information Ecotoxic effects: Biological effects: Toxic for aquatic organisms. Caustic even in diluted form. Endangers drinking water supplies if allowed to enter soil and/or water in large quantities. Harmful due to pH shift. Does not cause biological oxygen deficit. Further ecologic data: The following applies to sulphuric acid: biological effects: toxic for aquatic organisms; harmful effect due to pH shift. LD 50 96h: 10 mg/l; fish: lethal as from 1.2 mg/l; as from 6.3 mg/l lethal in 24 h. Do not allow larger quantities to enter drinking water supplies, waste water, or soil! 13 Disposal considerations Product: There are no uniform EC regulations for the disposal of chemicals or residues. Chemical residues generally count as special waste. The disposal of the latter is regulated in the EC member countries through corresponding laws and regulations, and in Switzerland also by the cantons. We recommend that you contact either the authorities in charge or approved waste disposal companies which will advise you on how to dispose of special waste. Packaging: Disposal in compliance with official regulations. Handle contaminated packaging in the same way as the substance itself. If not officially specified differently, non-contaminated packaging may be treated like household waste or recycled. 64 A-96.250.441 / 060923 AMU pH-Redox Material Safety Data Sheets 14 Transport information Land transport ADR/RID and GGVS/GGVE (Germany) GGVS/GGVE-class: 8 Number and letter: ADR/RID-class: 8 Number and letter: Name of material: 2796 SCHWEFELSAEURE 1B 1B River transport ADN/ADNR: not examined Sea transport IMDG: IMDG-class: 8 EmS: 8-06 Correct technical name: UN-No: 2796 MFAG: 700 SULPHURIC ACID <10% Air transport ICAO-TI and IATA-DGR: ICAO/IATA-class: 8 UN-/ID-No: 2796 Correct technical name: SULPHURIC ACID <10% Packaging group: II Packaging group: II The transport regulations are cited according to international regulations and in the form applicable in Germany (GGVS/GGVE). Possible national deviations in other countries are not considered. 15 Regulatory information Labeling according to EG Directives Symbol: C Name: Corrosive R-phrases: 35 Causes severe burns. S-phrases: 2-26-30 Keep out of reach of children. After eye contact rinse with plenty of water and immediately summon doctor. Never add water. German regulations Water pollution class: 1 (weakly polluting substance) (KBwS classification) Other national regulations: Swiss toxic class: 3 16 Other information Date of issue: 03.02.2005 The information contained herein is based on the present state of our knowledge. It characterizes the product with regard to the appropriate safety precautions. It does not represent a guarantee of the properties of the product. A-96.250.441 / 060923 65 AMU pH-Redox RS232 Interface C. RS232 Interface C.1 Connection of the Communication Cable 1 Disconnect the instrument from the mains! 2 Connect the communication cable for AMU transmitters to the transmitter. RS232 with connected communication cable 66 3 Connect the other end of the cable to the PC. 4 Connect the instrument to power. A-96.250.441 / 060923 AMU pH-Redox RS232 Interface C.2 Configuration of the AMU Transmitter 1 Enter the menu 5.5.1 on the AMU transmitter and choose HyperTerminal for the sub-menu Protocol. Interface Protocol Baud Rate 5.5.1 HyperTerminal 115200 Baud Protocol Profibus Modbus RTU SWAN Bus HyperTerminal Webserver Confirm with <Enter> 2 Enter menu 5.5.2 and choose 115200 Baud for the Baud rate. Interface Protocol Baud Rate 5.5.2 HyperTerminal 115200 Baud Baud Rate 2400 Baud 4800 Baud 9600 Baud 19200 Baud 38400 Baud 57600 Baud 115200 Baud Confirm with <Enter> A-96.250.441 / 060923 67 AMU pH-Redox RS232 Interface C.3 HyperTerminal Configuration on the PC * 68 A-96.250.441 / 060923 AMU pH-Redox RS232 Interface * Note: If your computer has no COM port, you have to use an USBCOM-Converter to connect the cable to the PC. After having it connected find out via the Device Manager which port you are using (Settings/Control Panel/System/Hardware/ Device Manager). Clicking on the Device Manager leads you to the following scheme: Clicking on Ports (COM & LPT) shows you which port is used. This address of the port should be used for configuring the HyperTerminal. A-96.250.441 / 060923 69 AMU pH-Redox RS232 Interface 70 A-96.250.441 / 060923 AMU pH-Redox RS232 Interface A-96.250.441 / 060923 71 AMU pH-Redox RS232 Interface C.4 Functions of the AMU RS232 Interface The AMU RS232 Interface has two functions: It allows the transfer of log data onto a PC and the upgrade of AMU instruments with a new firmware. C.4.1 Logger Download Log interval The internal logger can save approx. 1500 data records, stored at a selectable interval. The following table shows the period of time covered with the corresponding interval. Operation Sensors Signal Outputs Relay Contacts Logger 4.4 4.4.1 Logger Log Interval x Sec/Min/Hour Clear Logger no/yes Log Interval 1 Second 5 Seconds 1 Minute 5 Minutes 10 Minutes 30 Minutes 1 Hour Interval 1s 5s 1 min 5 min 10 min 30 min 1h Time 25 min 2h 25 h 5d 10 d 31 d 62 d Default value for the the interval is set on 30 min. 72 A-96.250.441 / 060923 AMU pH-Redox RS232 Interface C.4.2 Log download Download Procedure Choose the configured connection. On the keyboard, press Esc. Browse to the folder you have created to archive printouts. A text file AMU*.TXT is created. Existing files are renamed AMU*0.TXT, AMU*1.TXT, AMU*2.TXT, etc. * = Abbr. for the corresponding instrument, e.g. AMUCON.TXT is the file for the instrument AMU pH-Redox. A-96.250.441 / 060923 73 AMU pH-Redox RS232 Interface Example of a logger file: AMUCON.txt Date 08.05.06 31.05.06 01.06.06 Time 15:00:00 09:30:00 17:30:00 Logger download of an AMU pH-Redox. Alarms 0536871936 0536871424 0536871424 pH [pΗ] 0.0 0.0 0.0 [] Temp. [°C] FLOW [l/h] 25.0 0.0 0.0 7.0 0.0 7.0 You can also download the Message List and the Calibration History. Message List of an AMU pH-Redox: EVTCON.txt Date 02.03.06 02.03.06 22.06.06 Time 13:34:00 14:49:37 17:12:41 Code E033 E012 E011 Desc Power-on Temp. disconnected Temp. shorted State Active Acknowledged Active Calibration History of an AMU pH-Redox: CALCON.txt pH Date 22.06.06 Time 17:12:26 Changing the folder for download data 74 Offset [mV] Slope [mV] -3 58.21 To change the folder where the files are saved, 1 2 3 press the “Receive”-button. choose the directory and the folder and the receiving protocol must be Zmodem. A-96.250.441 / 060923 AMU pH-Redox RS232 Interface C.4.3 Clear Logger To delete all log data, proceed as follows: Operation Sensors Signal Outputs Relay Contacts Logger Logger Log Interval Clear Logger 4.4 4.4.2 30 Minutes no Clear Logger no yes 4.4.2 Logger Log Interval Save? 30 Minutes Clear Logger yes Yes No A-96.250.441 / 060923 Select yes and press to accept. Press Exit Enter . yes highlighted, press Enter to clear logger. The system reboots, all log data is deleted. 75 AMU pH-Redox RS232 Interface C.5 Firmware Upload Only SWAN trained and authorized personnel shall perform the tasks described in this section. Requisites Firmware upload The HyperTerminal connection must be configured. See HyperTerminal Configuration on the PC, p. 68 1 Choose the configured connection. AMU pH-Redox 3 Press <Enter> and confirm it by pressing “y”. 76 A-96.250.441 / 060923 AMU pH-Redox RS232 Interface 4 If upload fails (instrument does not restart), repeat steps 1 and 4. Perform Set defaults 5.4.2 > completely and setup the instrument according to user specifications. A-96.250.441 / 060923 77 AMU pH-Redox Index D. Index A Alarm Relay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7, 16 5 C Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22, 23 Communication Cable . . . . . . . . . . . . . . . . . . . . . . . . . . 66 Control Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 D Default Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50, 78 Description of the System . . . . . . . . . . . . . . . . . . . . . . . . 5 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 78 E Error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Establish Sample Flow . . . . . . . . . . . . . . . . . . . . . . . . . Explanations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27 21 33 F Firmware Upload . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76 I Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7, 16 7, 19 L Logger Download . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72 M Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Material Safety Data Sheets . . . . . . . . . . . . . . . . . . . . . . Measuring Principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 53 5 13 P Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 A-96.250.441 / 060923 AMU pH-Redox Index R Relay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Relay Contacts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . RS232 1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 16 9, 66 S Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Sample Flow, Establish . . . . . . . . . . . . . . . . . . . . . . . . . Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Sensor, Installation of . . . . . . . . . . . . . . . . . . . . . . . . . . . Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Signal Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Software Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Symbol, Attention . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Symbol, Warning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . System, Description of . . . . . . . . . . . . . . . . . . . . . . . . . . 3 21 8, 17 17 21 7, 16 9 11 3 3 5 T Technical Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Trouble Shooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 27 A-96.250.441 / 060923 79 AMU pH-Redox SWAN is represented worldwide by subsidiary companies and distributors. cooperates with independent representatives all over the world. SWAN Products Analytical Instruments for: High Purity Water Feedwater, Steam and Condensate Potable Water Pool and Sanitary Water Cooling Water Waster Water and Effluents Made in Switzerland. 80 A-96.250.441 / 060923
© Copyright 2024 Paperzz