An Introduction to Artifact Recognition Part 1 By Wayne Peacock, RPSGT Artifact recognition is an essential part of quality study is captured for the reviewing the sleep technologist’s job. Timely, accurate technologist and physician. recognition and correction of several types There are many types of artifact, some of of artifact lead to the acquisition of a highwhich may be beneficial. For example, snoring quality and artifact-free recording. In order artifact seen in the chin electromyogram to achieve this, the recording technologist (EMG) may actually add value to the study by must pay close attention to proper electrode confirming the presence of snoring; muscle and sensor application, ensure accurate signal artifact during movement helps confirm processing and maintain the integrity of the patient movement.3 The task of the sleep recording. Attention to these three things technologist includes recognizing the artifact, will determine the quality of the recording determining its origin, and then deciding the wayne peacock, obtained. most appropriate way to address the problem. rpsgt Artifact can affect every channel and To determine the origin, begin with the patient. biopotential recorded. In order to correctly handle Check all electrodes and wires and work toward the various types of interference, the recording technologist amplifiers and eventually back to the computer.1 To put must possess a working knowledge of what is being it simply, follow the signal pathway. recorded and how to best respond to each type of Figure 1 shows a typical patient circuit. artifact that may affect that channel. For the purpose of this article it is assumed that the recording Patient Sensor Headbox Amplifier AD Converter Computer technologist used proper 10-20 electrode placement for Figure 1. A typical patient circuit1 the electroencephalogram (EEG), performed careful site preparation, and secured to the skin quality electrodes with relatively low impedances (<5kOhms). Impedances Most artifacts may be recognized by an exaggerated for both EEG and electrooculogram (EOG) should range appearance, although some may be difficult to from one to five kOhms.1 Even with detailed attention to distinguish from physiological signals. An example all of the aspects described above, a sleep recording may of this is the appearance of sweat artifact in the EEG still encounter intermittent artifact. Impedances should channels. This artifact is generally identified by its be checked periodically throughout the recording, excessively slow frequency; however, the intruding waves especially when signs of artifact are noted.2 Knowing may be similar in their appearance to high-amplitude how to deal with these occurrences will ensure that a delta waves.3 ECG Artifact ECG artifact is common and is usually seen in the EEG, EOG or EMG channels. 3 See Figure 2. Strong ECG voltages may be detected from almost any location on the body. ECG artifact may be increased due to impedance imbalance between the exploring and referencing electrode. This should always be considered. In some cases, as with obese patients, some amount of ECG artifact will be present in spite of low impedances. To reduce the probability of ECG artifact, care should be taken to avoid placing referencing electrodes (M1 and M 2) on soft, fatty tissue. The technologist may find it advantageous to place these electrodes slightly higher on the mastoid process, just above the ear on the bony portion; they also may be placed on the earlobes.3 If artifact remains, then the technologist may link or “double reference” the two referencing electrodes. This can be accomplished either with jumpers or by using a derivation that allows the M 1 to be linked with 20 A2Zzz 2008 • volume 17 • number 4 M 2 (M 1+M 2). 3 This method is called double referencing, which was accomplished via the electrode selector panel with analog systems. Many of today’s digital recording systems have provisions allowing double referencing, which may be as simple as the click of a mouse. Figure 2. ECG Artifact Movement Artifact Oximetry Artifacts Figure 3. Movement artifact If it does not disappear, then there may have been a dislodging of electrodes during the repositioning. Movement artifact may be minimized with proper electrode application. Figure 4. Oximeter probe displaced Thick or painted nails also may present problems when recording oximetry tracings. It is imperative for the recording technologist to recognize poor or improper tracings from the oximeter since this signal is significant in scoring and diagnosing sleep related breathing disorders. Often times this artifact may be corrected by simply re-attaching the probe. When this does not correct the problem it may be necessary to replace the probe or simply trace the connections to ensure all cables are securely in place. This type of artifact is usually a generalized artifact in all channels caused by the patient repositioning. This artifact usually dissipates once the patient has settled. See Figure 3. Improper probe attachment or a faulty probe may be the cause of oximetry signal artifact.3 Occasionally an oximetry signal also may be lost temporarily due to large movements. See Figure 4. This article will continue in the next issue of A2Zzz with examples of other common artifacts. References 1. American Association of Sleep Technologists. American Association of Sleep Technologists technical guideline for standard polysomnography: update 3-2008. Available to members online at www.aastweb. org. 2. American Academy of Sleep Medicine. The AASM manual for the scoring of sleep and associated events: rules, terminology and technical specifications. Westchester, Ill: American Academy of Sleep Medicine; 2007. p. 19. 3. Butkov N. Recording quality and artifacts. In: Butkov N, Lee-Chiong T, editors. Fundamentals of sleep technology. Philadelphia: Lippincot Williams & Wilkins; 2007. p. 351-365. Wayne Peacock, RPSGT, has been in the sleep field for 10 years and is the manager of the Sleep Disorders Center and Neurodiagnostic Department at Baptist Hospital in Pensacola, Fla. A2Zzz 2008 • volume 17 • number 4 21
© Copyright 2026 Paperzz