A Comparison of Differential White Blood Cell Counts Using Manual Technic and the Coulter® S-Plus IV ROBERT M. GREENDYKE, M.D., DONALD R. KANTER, PH.D., LOUISE D E B O O V E R , B.S., LINDA SAVAGE, B.S., AND SUSAN VANGELDER, B.S. The results of 100-cell and 500-cell manual differential white blood cell counts were compared with those obtained using a Coulter® S-Plus IV electronic particle counter. Significant (P < 0.001) correlations were observed between the manual and instrument-derived data for percentages of granulocytes and lymphocytes. (Key words: Differential white blood cell counts; Coulter® S-Plus IV) Am J Clin Pathol 1985; 84: 348-350 TO VALIDATE the routine use of quantifying three white blood cell (WBC) subpopulations using a Coulter® S-Plus IV electronic blood cell counter as a replacement for manual performance of differential WBC counts, a study has been conducted to compare data produced by the instrument and by conventional technic. Little has been published to document the acceptability of the substitution. Methods One hundred consecutive single Wright's-stained peripheral blood smears were prepared from routine specimens submitted from patients hospitalized in this institution. The blood samples came from patients with psychiatric and/or generally chronic medical problems representative of the hospital population as a whole, and included no cases of primary hematologic disorder. One hundred-cell and 500-cell manual differential WBC counts were independently performed on the same smears by each of three of the authors (LDB, LS and SVG) who all are certified MT (ASCP) and have several years experience as laboratory technologists. Differential counts were performed in duplicate on the same blood specimens on the Coulter S-Plus IV cell counter according to the manufacturer's directions.1 The instrument distinguishes three WBC subpopulations (segmented and band granulocytes, small lymphocytes, and large cells including monocytes, large lymphocytes, and myeloid cells younger than bands). To permit comparisons, values derived from the manual counts on eosinophils, basophils, band neutrophils, and segmented neutrophils Laboratory Service, U.S. Veterans Administration Medical Center, Canandaigua, New York were combined for comparison with instrument-generated granulocyte data. Manually determined monocyte percentages were compared to instrument generated mononuclear (large cell) values. The data were then subjected to statistical study using the Pearson ProductMoment Correlation.2 Results Granulocytes 100-Cell Count Determinations. The results of the Pearson Product-Moment Correlation on granulocyte data derived from 100-count manual determinations and the Coulter Counter instrument are indicated in Table 1. Each of the three manual counts showed highly significant correlations to each other and to the Coulter determination. 500-Cell Count Determinations. As indicated in Table 2, correlational analyses of 500-count manual determinations also revealed highly significant correlations between each of the manual counts, as well as between manual counts and Coulter counts. In every case, the 500-cell count manual determinations showed higher correlations than 100-cell count values. Received November 1, 1984; received revised manuscript and accepted for publication January 16, 1985. Address reprint requests to Dr. Greendyke: Chief Laboratory Service, VA Medical Center, Canandaigua, New York 14424. 348 Lymphocytes 100-Cell Count Determinations. Correlational analyses of manual 100-cell count lymphocyte determinations and the Coulter instrument count resulted in highly statistically significant correlations, which are summarized in Table 3. 500-Cell Count Determinations. Again, 500-count determinations of lymphocyte percentages produced higher correlations than did 100-count determinations. These analyses are summarized in Table 4. Mononuclear Cells 100-Cell Count Determinations. Analyses of manual counts indicated significant correlations between the BRIEF SCIENTIFIC REPORTS Vol. 84 • No. 3 Table 3. 100-Cell Count Manual Determinations and Coulter Instrument Count Percentages for Lymphocytes Table 1. 100-Cell Count Manual Determinations and Coulter Instrument Count Percentages for Granulocytes Tech 1 Tech 1 versus Tech 2 Tech 1 versus Tech 3 Tech 1 versus Coulter Tech 2 Tech 2 versus Tech 3 Tech 2 versus Coulter Tech 3 Tech 3 versus Coulter Coulter Mean SD 67.35 11.77 r 0.87* 0.88* 0.81* 68.54 11.38 0.88* 0.85* 69.48 11.83 0.86* 73.19 349 9.43 • /'<o.obi. Tech 1 Tech 1 versus Tech 2 Tech 1 versus Tech 3 Tech 1 versus Coulter Tech 2 Tech 2 versus Tech 3 Tech 2 versus Coulter Tech 3 Tech 3 versus Coulter Coulter Mean SD 27.83 10.97 r 0.87* 0.85* 0.83* 27.38 10.73 0.86* 0.88* 25.96 11.14 21.90 8.73 0.86* * P< 0.001. three technologists (see Table 5); however, only one of the three technologists' determinations showed a significant correlation with the Coulter instrument count. 500-Cell Count Determinations. All correlations between the technologists and between technologists and machine counts were significant for 500-cell count determinations. However, the technologists demonstrated a significantly higher level of agreement with each other than with the instrument count (see Table 6). Discussion The difficulty in attempting to evaluate the validity of differential WBC counts produced by an electronic particle counter by comparing them with manually generated results is readily apparent. The two methods use different criteria for leukocyte identification, and the Coulter instrument categorizes the cells into only three groups. The problems associated with manual WBC differential counts are well known, and include nonrandom distribution of cell types on the smear, possibly Table 2. 500-Cell Count Manual Determinations and Coulter Instrument Count Percentages for Granulocytes Mean SD Tech 1 Tech 1 versus Tech 2 Tech I versus Tech 3 Tech 1 versus Coulter 68.25 10.71 Tech 2 Tech 2 versus Tech 3 Tech 2 versus Coulter 68.65 Tech 3 Tech 3 versus Coulter 68.56 10.50 10.81 0.97* 0 97* 0.89* 0 97* 0.91* 0.90* inadequate sample size counted, and generally minor difficulties with cell identification.3 The Coulter S-Plus IV counts 20,000 cells, thereby minimizing error variance associated with sampling limitations, but substitutes a potentially larger problem in cell identification by virtue of using only cell size as a criterion. The present study indicated highly significant (P < 0.001) correlations among the data produced manually by three technologists and also between manual and instrument-generated granulocyte and lymphocyte percentages. Five hundred-cell differential counts yielded higher correlation coefficients between manual and instrument data for granulocyte and lymphocyte percentages than did 100-cell counts, suggesting that sample size may have contributed to the differences. Not unexpectedly, in light of the different cell identification criteria, the manual monocyte vs. instrument mononuclear cell percentages correlated poorly. There are obvious limitations to the use of three- Table 4. 500-Cell Count Manual Determinations and Coulter Instrument Count Percentages for Lymphocytes Tech 1 Tech 1 versus Tech 2 Tech 1 versus Tech 3 Tech 1 versus Coulter Tech 2 Tech 2 versus Tech 3 Tech 2 versus Coulter Tech 3 Tech 3 versus Coulter Coulter */>< 0.001. Mean SD 27.19 9.95 27.31 9.85 27.78 10.34 0.96* 0.96* 0.90* 0.95* 0.91* 0.89* 21.90 8.73 350 GREENDYKE ET AL. Table 7. Criteria Based upon Instrument Printout for Manual Repetition of Automated WBC Differential Counts and Frequency of Occurrence of Each in 500 Consecutive Tests* Table 5. 100-Cell Count Manual Determinations and Coulter Instrument Count Percentages for Mononuclear Cells Tech 1 Tech 1 versus Tech 2 Tech 1 versus Tech 3 Tech 1 versus Coulter Tech 2 Tech 2 versus Tech 3 Tech 2 versus Coulter Tech 3 Tech 3 versus Coulter Coulter Mean SD 4.8 2.49 r 2.3 0.38t 0.16 NS 4.61 3.0 4.82 1.59 0.17 NS • /' < 0.05. t/"< 0.001. population differential WBC counts as produced by the Coulter S-Plus IV instrument, especially for hospital inpatients. It may be argued that stained peripheral blood smears should be manually scanned in all cases, and this currently is routine in the authors' laboratory. Such a practice, however, vitiates some of the advantage of automated blood cell counting. It is beyond the scope of this paper to discuss the relative merits of data produced by the Coulter instrument relative to red blood cell indices and relative distribution width vs. traditional estimates of hemoglobinization and anisocytosis, or of platelet count, mean platelet volume, and platelet distribution width vs. estimates from blood smears. That there are WBC distribution abnormalities that are missed without manual examination is readily apparent, two of Table 6. 500-Cell Count Manual Determinations and Coulter Instrument Count Percentages for Mononuclear Cells Tech 1 Tech 1 versus Tech 2 Tech 1 versus Tech 3 Tech 1 versus Coulter Tech 2 Tech 2 versus Tech 3 Tech 2 versus Coulter Tech 3 Tech 3 versus Coulter Coulter • P < 0.001. t/'<0.0l. t P < 0.05. Mean SD 4.44 2.06 r 0.69* 0.72* 0.26t 3.99 Percentage Frequency of Occurrence Criteria 0.42t 0.34f 0:25* 4.08 AJ.C.P. .September 1985 Increased total WBC Decreased total WBC Increased granulocyte percentage Decreased granulocyte percentage Increased lymphocyte percentage Increased mononuclear cell percentage "Valley failure" 43 6 44 6 16 13 14 • Thirty-two percent of counts requiring manual repetition met two or more criteria. For discussion of "valley failure" see text. In addition, a very small number of additional differential counts were repeated as a result of eosinophilia, basophilia, or band neutrophilia as isolated findings on routine scanning of peripheral blood smears from samples not displaying any of the other abnormalities listed above. the most significant being the presence of band neutrophilia and eosinophilia, although it is acknowledged that the former usually is associated with granulocytosis. The cost effectiveness of routinely screening for such abnormalities has not been assessed. The criteria established in this institution for manual repetition of the automated WBC differential count are listed in Table 7, together with the frequency of occurrence of each in 500 consecutive blood samples. The phenomenon of "valley failure" with the use of the Coulter instrument requires comment. In 14% of the present cases, the instrument was unable to discriminate between the mononuclear and lymphocyte populations because of a continuous cell size gradient resulting from the presence of numerous large lymphocytes. The absence of a "valley" between the cell size distribution peaks (whence the term "valley failure") precluded enumeration of the two types of cells and required manual counting. A sampling of 1,591 random differential WBC counts performed in the 12 months during which the above criteria have been used indicates that 58.6% have required manual repetition. Clinical acceptance of data so generated has been excellent. Acknowledgment. Roy Johnston, Ph.D., and B. W. Lundy, Ph.D., performed the initial statistical analysis of the data presented. 1.75 0.68* 0.27t 3.85 2.37 4.82 1.59 0.21J References 1. Coulter Counter Model S-Plus IV with three-population differential product reference manual. Hialeah, FL, Coulter Electronics, October 1983 2. Hays W: Statistics for the Social Sciences. Second edition. New York, Holt, Rhinehart and Winston, 1973, p 616 3. Nelson DA, Morris MW: Basic methodology, Clinical Diagnosis and Management by Laboratory Methods. Edited by JB Henry. Edition 17. Philadelphia, Saunders, 1984, pp 611-612
© Copyright 2025 Paperzz