Cuvette-Test LCK 348 Phosphate

PA 348 - 06/1
Cuvette-Test
LCK 348 Page 1
Programming instructions
CADAS 100 LPG ≥ 210
Phosphate-Phosphorus
Edition 01/06
ortho and total
Attention! The edition date of this programming instruction must be identical to the edition date of the instruction
attached to the reagents. This is the only way to ensure that your instrument will be programmed with the correct data.
Please note!
If you make an error during programming, you can cancel the last entry by pressing the key ➞
on the keyboard.
and L
indicate the right and left
➞
➞
➞
The input symbols R
keys respectively.
Programming of LCK 348 Phosphate-Phosphorus
Select the program PRG (programming of a test, see page 1 and 2) or
Select the program VAR (variation of a test, see page 3 and 4).
Phosphate-Phosphorus (PO4-P) (Program PRG)
Program number:
for example 1
Symbol:
L
348 L
1
W:
λ
=
Factor:
1
=
2
3
STA
FPRG
... nm
➞
Pro:
E
▼ 4- L
PO L
➞
Name:
➞
Input via keyboard
➞
Display
P
Press the function key below 1
850
6.47
mg L
Measuring limit:
yes
no
/
X
➞
-0.054
Unit:
➞
Constant:
R
l
Press function key below yes
Lower limit:
0.5
Upper limit:
5.0
Programming of LCK 348 Phosphate
Select the program PRG (programming of a test, see page 1 and 2) or
Select the program VAR (variation of a test, see page 3 and 4).
Phosphate (PO4) (Program PRG)
Display
Input via keyboard
Program number:
for example 1
➞
L
Name:
PO L
348
➞
Symbol:
Pro:
1
W:
λ
=
Factor:
1
=
2
... nm
3
STA
FPRG
Press the function key below 1
850
19.84
Unit:
mg L
yes
no
/
X
➞
-0.146
➞
Constant:
Measuring limit:
▼ 4
R
l
Press function key below yes
Lower limit:
1.5
Upper limit:
15.0
All data entries are documented on the printer; their transfer to storage is confirmed by the printer message »stored«.
You should check the printed data to ensure that all entries are correct.
PA 348 - 06/1
Cuvette-Test
LCK 348 Page 2
Programming instructions
CADAS 100 LPG ≥ 210
Phosphate-Phosphorus
Edition 01/06
ortho and total
Attention! The edition date of this programming instruction must be identical to the edition date of the instruction
attached to the reagents. This is the only way to ensure that your instrument will be programmed with the correct data.
Please note!
If you make an error during programming, you can cancel the last entry by pressing the key ➞
on the keyboard.
and L
indicate the right and left
➞
➞
➞
The input symbols R
keys respectively.
Programming of LCK 348 Phosphorus-pentoxide
Select the program PRG (programming of a test, see page 1 and 2) or
Select the program VAR (variation of a test, see page 3 and 4).
Phosphorus-Pentoxide (P2 O5 ) (Program PRG)
Program number:
for example 1
Symbol:
L
348 L
1
W:
λ
=
Factor:
1
=
2
... nm
3
STA
FPRG
O L
▼ 5
Press the function key below 1
850
14.83
mg L
yes
no
/
X
➞
-0.107
Unit:
➞
Constant:
Measuring limit:
▼ 2 L
➞
Pro:
P
➞
P L
➞
Name:
➞
Input via keyboard
➞
Display
R
l
Press function key below yes
Lower limit:
1.2
Upper limit:
11.5
All data entries are documented on the printer; their transfer to storage is confirmed by the printer message »stored«.
You should check the printed data to ensure that all entries are correct.
PA 348 - 06/1
Programming instructions
CADAS 100 LPG ≥ 210
Phosphate-Phosphorus
Cuvette-Test
LCK 348 Page 3
ortho and total
Edition 01/06
Attention! The edition date of this programming instruction must be identical to the edition date of the instruction
attached to the reagents. This is the only way to ensure that your instrument will be programmed with the correct data.
Please note!
If you make an error during programming, you can cancel the last entry by pressing the key ➞
on the keyboard.
Programming of LCK 348 Phosphate-Phosphorus
Select the program VAR.
Select function key below symbol »348 E«.
Phosphate-Phosphorus (PO4-P) (Program VAR)
Display
Input via keyboard
Program number:
for example 1
Symbol:
348 E
Name:
PO4 - P LCK 348 E
Pro:
1
3
STA
FPRG
Press function key below 1
λ = 850 nm
W:
Factor:
2
1
=
6.47
Constant:
-0.054
Unit:
mg/l
Measuring limit:
yes
Lower limit:
0.5 mg/l
Upper limit:
5.0 mg/l
no
Press function key below yes
Programming of LCK 348 Phosphate
Select the program VAR.
Select function key below symbol »348«.
Phosphate (PO4 ) (Program VAR)
Display
Input via keyboard
Program number:
for example 1
Symbol:
348
Name:
PO4 LCK 348
Pro:
1
3
STA
FPRG
Press function key below 1
λ = 850 nm
W:
Factor:
2
1
=
19.84
Constant:
-0.146
Unit:
mg/l
Measuring limit:
yes
Lower limit:
1.5 mg/l
Upper limit:
15.0 mg/l
no
Press function key below yes
All data entries are documented on the printer; their transfer to storage is confirmed by the printer message »stored«.
You should check the printed data to ensure that all entries are correct.
PA 348 - 06/1
Programming instructions
CADAS 100 LPG ≥ 210
Phosphate-Phosphorus
Cuvette-Test
LCK 348 Page 4
ortho and total
Edition 01/06
Attention! The edition date of this programming instruction must be identical to the edition date of the instruction
attached to the reagents. This is the only way to ensure that your instrument will be programmed with the correct data.
Please note!
If you make an error during programming, you can cancel the last entry by pressing the key ➞
on the keyboard.
Programming of LCK 348 Phosphorus-Pentoxide
Select the program VAR.
Select function key below symbol »348 P«.
Phosphorus-Pentoxide (P2 O5 ) (Program VAR)
Display
Input via keyboard
Program number:
for example 1
Symbol:
348 P
Name:
P2 O5 LCK 348 P
Pro:
1
3
STA
FPRG
Press function key below 1
λ = 850 nm
W:
Factor:
2
1
=
14.83
Constant:
-0.107
Unit:
mg/l
Measuring limit:
yes
Lower limit:
1.2 mg/l
Upper limit:
11.5 mg/l
no
Press function key below yes
All data entries are documented on the printer; their transfer to storage is confirmed by the printer message »stored«.