Cuvette-Test LCK 348 Phosphate

PA 348 - 06/1
Cuvette-Test
LCK 348 Page 1
Programming instructions
CADAS 100 LPG 185
Phosphate-Phosphorus
Edition 01/06
ortho and total
Attention! The edition date of this programming instruction must be identical to the edition date of the instruction
attached to the reagents. This is the only way to ensure that your instrument will be programmed with the correct data.
Please note!
If you make an error during programming, you can cancel the last entry by pressing the key ➞
on the keyboard.
and L
indicate the right and left
➞
➞
➞
The input symbols R
keys respectively.
Deletion of the old LCK 348 Phosphate-Phosphorus program
Select the program CPRG (deletion of tests).
Press function key below symbol »348 E«.
Press function key below »CLEAR«.
Press function key below » > « resp. » «. Return to main program.
➞
●
●
●
●
Programming of LCK 348 Phosphate-Phosphorus
Select the program PRG (programming of a test, see page 1, 2 and 3).
Select the program VAR (variation of a test, see page 4 and 5).
Phosphate-Phosphorus (PO4-P) (Program PRG)
Input via keyboard
Program number:
for example 1
Name:
PO L
1
λ = … nm:
850
C1 = A
1
C1 = A 1 x F
1
C 1 = A 1 x F1 + A
0
C 1 = A 1 x F1 + F
2
$ 348 L
➞
L
Number of measurement points:
➞
Symbol:
▼ 4- L
➞
➞
Display
P
E
C 1 = A 1 x F1 + F 2 + A
C2 = A
Factor:
Factor:
=
2
=
-0.054
/R
➞
mg L
➞
Dimension 1 :
6.47
l
Extension 1 :
All data entries are documented on the printer; their transfer to storage is confirmed by the printer message »stored«.
You should check the printed data to ensure that all entries are correct.
PA 348 - 06/1
Programming instructions
CADAS 100 LPG 185
Phosphate-Phosphorus
Cuvette-Test
LCK 348 Page 2
Edition 01/06
ortho and total
Attention! The edition date of this programming instruction must be identical to the edition date of the instruction
attached to the reagents. This is the only way to ensure that your instrument will be programmed with the correct data.
Please note!
If you make an error during programming, you can cancel the last entry by pressing the key ➞
on the keyboard.
and L
indicate the right and left
➞
➞
➞
The input symbols R
keys respectively.
Deletion of the old LCK 348 Phosphate program
Select the program CPRG (deletion of tests).
Press function key below symbol »348«.
Press function key below »CLEAR«.
Press function key below » > « resp. » «. Return to main program.
➞
●
●
●
●
Programming of LCK 348 Phosphate
Select the program PRG (programming of a test, see page 1, 2 and 3).
Select the program VAR (variation of a test, see page 4 and 5).
Phosphate (PO4 ) (Program PRG)
Input via keyboard
Program number:
for example 1
Name:
PO L
➞
Display
L
Number of measurement points:
1
λ = … nm:
850
C1 = A
1
➞
Symbol:
C1 = A 1 x F
1
C 1 = A 1 x F1 + A
0
C 1 = A 1 x F1 + F
2
▼ 4
$ 348
C 1 = A 1 x F1 + F 2 + A
C2 = A
Factor:
Factor:
=
2
=
-0.146
/R
➞
mg L
➞
Dimension 1 :
19.84
l
Extension 1 :
All data entries are documented on the printer; their transfer to storage is confirmed by the printer message »stored«.
You should check the printed data to ensure that all entries are correct.
PA 348 - 06/1
Cuvette-Test
LCK 348 Page 3
Programming instructions
CADAS 100 LPG 185
Phosphate-Phosphorus
Edition 01/06
ortho and total
Attention! The edition date of this programming instruction must be identical to the edition date of the instruction
attached to the reagents. This is the only way to ensure that your instrument will be programmed with the correct data.
Please note!
If you make an error during programming, you can cancel the last entry by pressing the key ➞
on the keyboard.
and L
indicate the right and left
➞
➞
➞
The input symbols R
keys respectively.
Deletion of the old LCK 348 Phosphorus-Pentoxide program
Select the program CPRG (deletion of tests).
Press function key below symbol »348 P«.
Press function key below »CLEAR«.
Press function key below » > « resp. » «. Return to main program.
➞
●
●
●
●
Programming of LCK 348 Phosphorus-Pentoxide
Select the program PRG (programming of a test, see page 1, 2 and 3).
Select the program VAR (variation of a test, see page 4 and 5).
Phosphorus-Pentoxide (P2 O5 ) (Program PRG)
Program number:
for example 1
Name:
P L
λ = … nm:
850
C1 = A
1
C1 = A 1 x F
1
C 1 = A 1 x F1 + A
0
C 1 = A 1 x F1 + F
2
➞
1
$ 348 L
➞
L
Number of measurement points:
➞
Symbol:
▼ 2 L
➞
Input via keyboard
➞
Display
O L
▼ 5
P
C 1 = A 1 x F1 + F 2 + A
C2 = A
Factor:
Factor:
=
2
=
-0.107
/R
➞
mg L
➞
Dimension 1 :
14.83
l
Extension 1 :
All data entries are documented on the printer; their transfer to storage is confirmed by the printer message »stored«.
You should check the printed data to ensure that all entries are correct.
PA 348 - 06/1
Programming instructions
CADAS 100 LPG 185
Phosphate-Phosphorus
Cuvette-Test
LCK 348 Page 4
ortho and total
Edition 01/06
Attention! The edition date of this programming instruction must be identical to the edition date of the instruction
attached to the reagents. This is the only way to ensure that your instrument will be programmed with the correct data.
Please note!
If you make an error during programming, you can cancel the last entry by pressing the key ➞
on the keyboard.
Programming of LCK 348 Phosphate-Phosphorus
Select the program VAR.
Select function key below symbol »$ 348 E«.
Phosphate-Phosphorus (PO4-P) (Program VAR)
Display
Input via keyboard
Name:
PO4-P
Symbol:
$ 348 E
Number of measurement points: 1
λ =
850 nm
C1 = A1 x F1 + F2 + A
C2 = A
Factor 1 =
6.47
Factor 2 =
-0.054
Dimension 1 =
mg/l
Extension 1 =
Programming of LCK 348 Phosphate
Select the program VAR.
Select function key below symbol »$ 348«.
Phosphate (PO4 ) (Program VAR)
Display
Input via keyboard
Name:
PO4
Symbol:
$ 348
Number of measurement points: 1
λ =
850 nm
C1 = A1 x F1 + F2 + A
C2 = A
Factor 1 =
19.84
Factor 2 =
-0.146
Dimension 1 =
mg/l
Extension 1 =
All data entries are documented on the printer; their transfer to storage is confirmed by the printer message »stored«.
You should check the printed data to ensure that all entries are correct.
PA 348 - 06/1
Programming instructions
CADAS 100 LPG 185
Phosphate-Phosphorus
Cuvette-Test
LCK 348 Page 5
ortho and total
Edition 01/06
Attention! The edition date of this programming instruction must be identical to the edition date of the instruction
attached to the reagents. This is the only way to ensure that your instrument will be programmed with the correct data.
Please note!
If you make an error during programming, you can cancel the last entry by pressing the key ➞
on the keyboard.
Programming of LCK 348 Phosphorus-Pentoxide
Select the program VAR.
Select function key below symbol »$ 348 P«.
Phosphorus-Pentoxide (P2 O5 ) (Program VAR)
Display
Input via keyboard
Name:
P2O5
Symbol:
$ 348 P
Number of measurement points: 1
λ =
850 nm
C1 = A1 x F1 + F2 + A
C2 = A
Factor 1 =
14.83
Factor 2 =
- 0.107
Dimension 1 =
mg/l
Extension 1 =
All data entries are documented on the printer; their transfer to storage is confirmed by the printer message »stored«.
You should check the printed data to ensure that all entries are correct.