i-STAT Operations and Testing Procedure

POCT Program
Massachusetts General Hospital - Pathology Service
55 Fruit Street, Boston, MA 02114
Title: i-STAT Operations and Testing Procedure
Cross References: i-STAT Operator Training Checklist
i-STAT INR Fingerstick Training and Assessment Record
i-STAT Competency Assessment 6-month and Annual
i-STAT Instrument Competency-Written Test
i-STAT Cartridge Temperature Storage Log
i-STAT Instrument Corrective Action Log
i-STAT Instrument Validation Log
i-STAT Record of Receipt-QC Documentation Log
i-STAT Training Guide
MGH POCT Inter-Laboratory Correlation Procedure
MGH POCT QC Storage Ordering and Documentation Guide
POCT Proficiency Testing Procedure
POCT Proficiency Testing Evaluation Worksheet
i-STAT CLEW update form
Contents:
Purpose................................................................................................................................................................... 2
Scope ...................................................................................................................................................................... 2
Policy and Procedure Statement............................................................................................................................. 2
Proficiency Testing (For G3+, CG8+, CG4+, ACT, PT/INR and Troponin cartridges)............................................ 2
Regulatory Requirements ....................................................................................................................................... 3
Competency Assessment ....................................................................................................................................... 3
Limitations/Interferences ......................................................................................................................................... 3
Test Kit/Supplies/Equipment ................................................................................................................................... 6
Calibration ............................................................................................................................................................... 7
Calibration Verification ............................................................................................................................................ 7
Electronic Quality Control........................................................................................................................................ 7
Liquid Quality Control .............................................................................................................................................. 8
Barcode Scanning................................................................................................................................................... 8
Calculations............................................................................................................................................................. 9
Reporting results ..................................................................................................................................................... 9
Preventive Maintenance.......................................................................................................................................... 9
Instrument Replacement Policy ............................................................................................................................ 10
Connectivity and LIS / HIS Down Time Procedure - OR....................................................................................... 10
Connectivity and LIS / HIS Down Time Procedure - Imaging ............................................................................... 12
Troubleshooting .................................................................................................................................................... 13
Alternative method ................................................................................................................................................ 13
References............................................................................................................................................................ 13
Reference Ranges ................................................................................................................................................ 14
ACT (Kaolin) Cartridge Procedure ........................................................................................................................ 16
CG8+, CG4+, and G3+ Cartridges Procedure ...................................................................................................... 20
Creatinine Cartridge Procedure ............................................................................................................................ 24
PT/INR Cartridge Procedure ................................................................................................................................. 27
Troponin I Procedure ............................................................................................................................................ 31
Date: 7/08
Written By:
Kim Gregory MT(ASCP)
Nancy Toscano MT (ASCP)
Date: 12/09
Title (with LTR): i-STAT Operations and Testing Procedure (LTR19694)
Last Approved: Gregory, Kimberly (Electronic Signature Timestamp: 8/11/2014 3:04:44 PM)
1
Purpose
This document outlines policies and procedures pertaining to testing on the i-STAT. In an effort to be concise, some
information from the manufacturer’s recommended procedure may be excluded. It is recommended that operators
familiarize themselves with the manufacturer’s product information that accompanies each package and their
manual, if one exists.
Scope
Level of Personnel: All RN’s, Catheterization lab techs, Perfusionists, Respiratory Therapists, Laboratory
Technologists, and MD’s who have successfully completed initial training and fulfilled the specific competency
requirements for the complexity level of each cartridge.
Testing Site: Clinical sites approved and on file with the Pathology service’s POCT Division.
Approved Cartridges: G3+, CG8+, CG4+, Creatinine, ACT (Kaolin), PT/INR, Troponin I
Policy and Procedure Statement
The i-STAT is a portable clinical analyzer used in conjunction with i-STAT cartridges to obtain definitive quantitative
measurements of ACT, blood gas parameters, lactate, creatinine, PT/INR and Troponin I utilizing i-STAT cartridges.
The disposable cartridge contains all of the system’s components (reagents, sensors, and waste container).
Proficiency Testing (For G3+, CG8+, CG4+, ACT, PT/INR and Troponin cartridges)
The College of American Pathologists (CAP) sends unknown samples to the laboratory for analysis several times
per year. Results are submitted to the CAP within 10 days of survey receipt. If a site fails 2 out of 3 events or two
consecutive events according to federal law, it may be required to discontinue testing.

All Survey results are to be handled and reported in the same manner as clinical results following the
directions on the CAP Survey package insert. The samples are not to be analyzed in duplicate unless
clinical specimens are analyzed in duplicate. Actions or decisions must be documented.

Participation must be random and not assigned to specific individuals. Successful participation may be used
as demonstrating successful competency for that year.

Upon receiving the survey:

The POCT program will contact the participating departments regarding the survey and the timeline
of the survey to be performed.

The departments must be available within the period identified by POCT.

The Key operators must make sure of the following:
o Instruments are in good working order.
o Randomly select staff to participate, but ensure that subsequent surveys are rotated among
different staff (e.g. document in a log)
o Maintain original CAP survey form with the results documented.
o Maintain the signed Attestation form.
o In addition, retain copies of above in the files of testing personnel.

Once results are obtained, they should be given to the POCT Coordinators who will send them to CAP via
mail, fax or electronic entry on the CAP website.

Site Director and CLIA certificate Director or designees shall review survey results to assess performance
and ensure compliance with the standard and comment.

Scores of 100% minimally requires documentation of review by the Director or designee.

Scores between 100% and 80% requires a comprehensive investigation and remedial action documented of
unsuccessful challenges.

Scores less than 80% requires a comprehensive investigation and documentation of remedial action of
unsuccessful challenges. Scores of less than 80 percent may jeopardize a sites ability to continue to
perform testing.

Should a site fail proficiency, they will be required to immediately perform a comprehensive investigation and
document remedial action. Operator re-training may be required.

In order to avoid cessation of testing, a site failing a challenge will be expected to develop and implement a
more aggressive plan for performance improvement
Title (with LTR): i-STAT Operations and Testing Procedure (LTR19694)
Last Approved: Gregory, Kimberly (Electronic Signature Timestamp: 8/11/2014 3:04:44 PM)
2

Each site is responsible for completing survey challenges when they arrive.
Anticipated Survey Periods:
AQ3 (Blood Gases), & PCARM
(Cardiac Markers)
CT-1/CT5
(Activated Clotting Time)
WP3
(PT/INR)
Product Receipt
Evaluation receipt
Product Receipt
Evaluation receipt
Product Receipt
Evaluation receipt
March
June
October
April
July
November
February
May
September
March
June
October
January
July
February
August
Regulatory Requirements
I. Each testing site must have a documented quality control program, which is developed in collaboration with or has
been approved by the MGH Pathology Service.
II. All test results must be maintained in patient records with all required information for four years
Required information:
1. Patient’s name
2. Medical Record Number
3. Patient’s gender
4. Patient’s age or date of birth
5. Date & time test collected, performed and reported
6. Ordering Physician
7. Responsible physician (if not 6)
8. Reference or Target Range
9. Test Performed
10. Test units
11. Lab name
III. Additional information that must be retained for four years:
1. Testing personnel records
2. Quality control results
3. Product information (i.e. serial number, lot numbers, expiration dates, etc.), information on quality control
and any remedial action
4. QC charts, maintenance sheets, reference and critical ranges
IV. Other
1. Universal precautions must be observed when handling any patient specimen.
2. A physician’s order or standing order is required prior to performing test.
3. The Hospital Hand Hygiene policy must be adhered to at all times.
V. Linearity/Calibration Verification
The POCT program will perform and document linearity/calibration verification checks every six months for non waived cartridges if applicable.
Competency Assessment
Testing personnel must demonstrate competency prior to direct patient testing. Refer to the individual cartridge
procedures for specific competency requirements. All operators must read the procedure manual and complete the
Operator Training Checklist after initial training and Competency Assessment Checklist after completing the
competency assessment.
Expired Operators:
Operators that fail to meet competency requirements within 365 days will be locked out of the system. They will be
required to undergo retraining and competency assessment according to above.
Limitations/Interferences
Interferences
Title (with LTR): i-STAT Operations and Testing Procedure (LTR19694)
Last Approved: Gregory, Kimberly (Electronic Signature Timestamp: 8/11/2014 3:04:44 PM)
3
An interferent is a substance, which, if present at significant levels in the blood specimen being analyzed, will
produce an error in the result of the analyte being measured. For example, in the table below, β-hydroxybutyrate at
sample concentration level of 16 mmol/L would decrease the measured sodium by 5 mmol/L.
Analyte
Interferent
ACT
Incorrect handling
Exogenously
added heparin,
citrate, oxalate or
EDTA
Creatinine
Acetaminophen
Ascorbate
Bromide
CO2
Details/Interferent
Concentration/
Analyzer not maintained
on a level surface during
testing
Heparinized syringe
1 mmol/L
0.227 mmol/L
100 mg/dL
When Creat < 2mg/dL:
pCO2 >40 mmHg
pCO2 < 40 mmHg
When Creat >2mg/dL:
pCO2 >40 mmHg
pCO2 < 40 mmHg
Creatine
N-acetylcysteine
5 mg/dL
16.6mmol/L
.
Hydroxyurea
Glucose
(Cartridge)
Bromide
PH
37.5 mmol/L (300 mg/dL)
pH: per 0.1 pH units below
7.4 @ 37°C
pH: per 0.1 pH units above
7.4 @ 37°C
Oxygen
PO2 less than 20 mmHg
@ 37°C
Hydroxyurea
100 μmol/L
Thiocyanate
24 mmol/L (140 mg/dL)
Effect on Analyte Result
Results may be affected by >10%
Will interfere with results
Increase (↑)Creat by 0.25 mg/dL
Increase (↑)Creat by 0.7 mg/dL
Increase (↑)Creat by 0.8 mg/dL
Increase (↑)Creat by 6.9%/10 mmHg
Decrease (↓)Creat by 6.9%/10mmHg
Decrease (↓)Creat by 3.7%/10mmHg
Increase (↑) Creat by 3.7%/10mmHg
Increase(↑) Creat by 0.20 mg/dL
Increase(↑) Creat by 0.4 mg/dL
Use an alternative method to
measure creatinine when patients
have been administered
hydroxyurea.
Decrease (↓) glucose by 30 mg/dL
Decrease (↓) glucose by 0.9 mg/dL
(0.05
mmol/L)
Increase (↑) glucose by 0.8 mg/dL (0.04
mmol/L)
May decrease (↓) glucose
Increase (↑) glucose 8 mg/dL (0.44
mmol/L)
Decrease (↓) glucose by approx. 23%
Analyte
Interferent
Magnesium
β-hydroxybutyrate
Lactate
Salicylate
Details/Interferent
Concentration
1.0 mmol/L above normal
20 mmol/L
20 mmol/L
4.34 mmol/L
Ionized
Calcium
Lactate
Effect on Analyte Result
Increase (↑) iCa by 0.04 mmol/L
Decrease (↓) iCa by 0.1 mmol/L
Decrease (↓) iCa by 0.05 mmol/L
Decrease (↓) iCa by 0.1 mmol/L
Bromide
25 mmol/L (200 mg/dL)
Decrease (↓) lactate by 40%
Cysteine
6.4 mmol/L (101 mg/dL)
Decrease (↓) lactate by 11%
Hydroxyurea
100 μmol/L
Increase(↑) Use alternative method to
measure lactate
Glycolic Acid
Increase (↑) lactate
Title (with LTR): i-STAT Operations and Testing Procedure (LTR19694)
Last Approved: Gregory, Kimberly (Electronic Signature Timestamp: 8/11/2014 3:04:44 PM)
4
PCO2
Propofol
(Diprovan®)
For patients administered propofol or
thiopental sodium, i-STAT recommends
the use of CG4+, CG8+, EG6+, and
EG7+ cartridges, which are free from
clinically significant interference at all
relevant therapeutic doses. i-STAT does
not recommend the use of EC8+
cartridges for patients receiving propofol
or thiopental sodium.
This antibiotic has been found to cause
a concentration-dependent false
prolongation of prothrombin time (PT)
and elevation of INR when using the iSTAT PT/INR test. It’s recommended
that an alternate testing method be
used.
Thiopental
Sodium
PT/INR
Cubicin
(daptomycin for
injection)
Chlorhexidine
Gluconate
Sodium
Troponin I
Lupus
Anticoagulant
Antibodies
The i-STAT PT/INR test may report
false prolongation of the
prothrombin time (PT) and an
elevation of the INR on samples
contaminated with Chlorhexidine
Gluconate.
β-hydroxybutyrate
Lactate
Bromide
If the presence of lupus anticoagulant
antibodies is known or suspected, use
an alternate testing method.
Decrease (↓) Na by 5 mmol/L
Decrease (↓) Na by 5 mmol/L
Increase (↑) Na by 5 mmol/L
Heparin
Gross Hemolysis
Rare Antibodies
to troponin or its
circulating
complexes
16 mmol/L (166 mg/dL)
20 mmol/L
37.5 mmol/L
90 U/mL
From traumatic draw or
residual alcohol from
drawing site
Decrease (↓) troponin by 20%
Decrease (↓) troponin
Samples from patients who have been
exposed to animals or who have
received therapeutic or diagnostic
procedures employing immunoglobulins
or reagents derived from
immunoglobulins may contain
antibodies, i.e. HAMA or other
heterophile antibodies, which may
interfere with immunoassays and
produce erroneous results. While this
product contains reagents that minimize
the effect of these interferents, and QC
algorithms designed to detect their
effects, the possibility of interference
causing erroneous results should be
evaluated carefully in cases where there
are inconsistencies in the clinical
information. Results from the i-STAT
cTnI assay should be considered in the
context of the entirety of the available
clinical information.
Title (with LTR): i-STAT Operations and Testing Procedure (LTR19694)
Last Approved: Gregory, Kimberly (Electronic Signature Timestamp: 8/11/2014 3:04:44 PM)
5
Test Kit/Supplies/Equipment
Product
i-STAT Liquid Controls:
Level 1
Level 2
Level 3
Calibration Verification
(4 ampoules each levels 1 thru 5)
Vendor
i-STAT
i-STAT
Manufacturer
#
PeopleSoft
#
06F12-01
06F13-01
06F14-01
062741
163167
218509
06F15-01
142135
Tests
Storage
requirement
Na, K, iCa, Glucose, pH,
PCO2, PO2, lactate,
Creatinine
See individual
cartridge
procedure
Na, K, iCa, Glucose, pH,
PCO2, PO2, Lactate
Store at 2-8ºC
Na, K, iCa, Glucose, pH,
PCO2, PO2,
calculated parameters:
TCO2, HCO3, BE, sO2
Lactate, pH, PCO2, PO2,
calculated parameters:
TCO2, HCO3, BE, sO2
pH, PCO2, PO2,
calculated parameters:
TCO2, HCO3, BE, sO2
See individual
cartridge
procedure
CG8+ Cartridge
i-STAT
03P88-25
101554
CG4+ Cartridge
i-STAT
03P85-25
162960
G3+ Cartridge
i-STAT
03P78-25
218460
Creatinine Cartridge
i-STAT
03P84-25
163684
Creatinine
See individual
cartridge
procedure
ACT Liquid Controls
Level 1
Level 2
i-STAT
06P17-15
06P17-16
141622
141623
ACT (Kaolin)
See individual
cartridge
procedure
ACT (Kaolin) Cartridge
i-STAT
03P87-25
329023
ACT (Kaolin)
PT/INR Cartridge
i-STAT
03P89-24
118473
PT/INR
INR Liquid Controls
Level 1
Level 2
i-STAT
06P17-13
06P17-14
137158
137159
PT/INR
cTnI Cartridge
i-STAT
03P90-25
329022
Troponin I
i-STAT
06F12-02
06F13-02
06F14-02
218512
218513
218514
Troponin I
i-STAT
06F15-02
218515
Troponin I
Electronic simulator
i-STAT
06F11-01
218532
N/A
Syringes 1 ml
B&D
309602
08653
Syringes 3 ml
Portex
4041
102044
Luerlock syringe
Portex
4042-2
28720
I-STAT1 9V NiMH
Battery
ISTAT
06F23-55
128162
cTnI Liquid Controls
Level 1
Level 2
Level 3
Cardiac Markers cTnI
Calibration Verification
Kit (3 levels x 2 each
level)
See individual
cartridge
procedure
See individual
cartridge
procedure
See individual
cartridge
procedure
See individual
cartridge
procedure
See individual
cartridge
procedure
See individual
cartridge
procedure
See individual
cartridge
procedure
< -18º C (-1º F),
cTnI Liquid
Controls
Room
Temperature
Room
Temperature
Room
Temperature
Room
Temperature
Room
Temperature
Title (with LTR): i-STAT Operations and Testing Procedure (LTR19694)
Last Approved: Gregory, Kimberly (Electronic Signature Timestamp: 8/11/2014 3:04:44 PM)
6
Calibration
Calibration is performed automatically as part of the test cycle on each cartridge type. Operator intervention is not
necessary.
Calibration Verification
Calibration Verification (CAL VER) using the “CAL VER” solutions (low, medium, and high levels) on all non-waived
analytes is required. (i.e. G3+, CG4+, CG8+, and Troponin I)
CAL VER is performed:
1. On any new, or replacement device before clinically used
2. Every 6 months
3. As part of the troubleshooting process when controls reflect an unusual trend or are consistently out of
range.
 Run each level once.
 i-STAT analyzer must be docked after running the CAL VER solutions.
 POCT program will print out the reports from Precision web.
 Signed and approved by Medical director or designee.
To perform CAL VER on i-STAT:
 Press Menu. Press 3 to select Quality Tests.
 Press 3 to select CAL VER.
If the Calibration Verification fails, the analyzer must be removed from use.
Electronic Quality Control
Electronic Simulators:
The Electronic Simulator (both external and internal) is a quality control device for the analyzer. It simulates two
levels of electrical signals that stress the analyzer’s signal detection function both below and above measurement
ranges. It provides an independent check on the ability of the analyzer to take accurate and sensitive
measurements- of voltage, current and resistance from the cartridge. An analyzer will pass or fail this test
depending on whether or not it measures these signals within specified limits. The Electronic Simulator will fail if
high relative humidity interferes with the measurements, making separate recording of humidity unnecessary.
A. Internal Electronic Simulator:
The Internal Electronic Simulator runs every 8 hours for blood gases and for all the analytes every 24 hours.
Inserting a cartridge triggers the internal simulator. When 8 hours has elapsed since the last Electronic
Simulator test (internal or external), it will automatically perform the internal test before the sample is tested,
adding about 15-20 seconds to the testing cycle.
1. If it fails, the testing cycle stops and FAIL displays on the screen.



Immediately rerun the cartridge in the same analyzer to confirm the FAIL.
If the cartridge fails again rerun the cartridge in a different analyzer (if less than three minutes has
lapsed from the time of cartridge filling.)
If the cartridge fails in more than one analyzer, use a different cartridge.
2. If it passes, the cartridge cycle continues to completion and the user receives the test results.
B. External Electronic Simulator:
The External Electronic Simulator is a stable electronic device, which is inserted into the cartridge port of the
analyzer as described above. It should be run when:
1. The analyzer has been dropped
2. Every six months
3. After major malfunction
4. Replacement analyzer received from i-STAT
To run external Simulator:
 Press the Menu key to access the Administration menu.
 Press the #3 for Quality test
 Press the #4 for Simulator
Title (with LTR): i-STAT Operations and Testing Procedure (LTR19694)
Last Approved: Gregory, Kimberly (Electronic Signature Timestamp: 8/11/2014 3:04:44 PM)
7
1. Insert the simulator into the analyzer with the “I” facing up. (Do not touch the contact pads).
Cartridge Locked is displayed indicating that the simulator should not be removed. DO NOT
ATTEMPT TO REMOVE THE SIMULATOR WHILE THE Cartridge Locked MESSAGE IS
DISPLAYED. AN ATTEMPT TO REMOVE FORCIBLY THE SIMULATOR DURING THIS CYCLE
MAY DAMAGE THE ANALYZER.
2. If PASS is displayed on the analyzer screen, remove the simulator after the Cartridge, Locked
message disappears from the display screen.
3. If FAIL is displayed on the analyzer screen:
 Repeat the procedure with the same Electronic Simulator. If PASS is displayed, use the
analyzer as required.
 If FAIL is displayed, repeat the procedure with a different Electronic simulator, if
available.
a) If PASS is displayed with the second simulator, use the analyzer as
required.
b) If FAIL is displayed with the second Electronic simulator:
(1) Do not analyze patient samples with the analyzer.
(2) Transmit the result to Precision Web.
(3) Record the failure in the Instrument Corrective Action Log, along
with action taken.
C. Thermal Probe Check
i-STAT analyzers contain a thermal control subsystem consisting of two thermal probes with thermistors and heating
contact wires. When measurements are performed at a controlled temperature, the thermal probes in the analyzer
contact the metalized area under the chips in the cartridge and maintain the temperature of the sensors and the
fluids that come into contact with these sensors at the required temperature of 0.10C.
The thermal probe check must be verified twice each year, after i-STAT software update. Check the thermal probes
as follows:
1. If the analyzer and simulator have been stored separately in areas where the ambient
temperature differs by more than 3C (5F), allow the simulator and analyzer to stand in the
same place, out of drafts, for 30 minutes before inserting the simulator into the analyzer.
Handle the simulator as little as possible to maintain its thermal uniformity and stability.
2. Insert the simulator into the analyzer.
3. When the results are displayed, press the period button and the difference between the thermal
probes can be viewed on the analyzer’s screen. (Acceptable: a value equal to or less that 0.1)
If a FAIL message with “t” Quality Check Code, or a value greater than 0.1 displays repeat the
procedure. Repeat the procedure if “--.--“ is displayed. Call Technical Support for repeated
failures.
4. Documentation of results appears electronically in Precision Web under the test detail of the
electronic simulator.
5. The results of the Thermal Probe Check for each analyzer are documented on the Thermal
Probe Check Log.
Liquid Quality Control
Liquid Quality Control must be performed:
1.
2.
3.
4.
5.
On every analyzer prior to initial placement for patient testing
On every new lot number and every new shipment of cartridges
When there’s a suspected cartridge or storage problem
On any back up analyzer placed in use before patient testing
On all cartridge/reagent types as well as all analytes. If the same analyte occurs on multiple
cartridge/reagent types, it must be tested on every cartridge/reagent type
6. Monthly to check storage conditions of cartridges
Note: Implementing a new or replacement device that uses Non-Waived Cartridges: i-Stat is considered an
option #1 device under CLIA. Point of care testing staff must run liquid controls daily for the first 10 days
concurrently with EQC, both internal and external. If no QC failures occur then the analyzer is ready to use. Any
failure will require further evaluation. Document results on i-STAT Instrument Validation Log.
Barcode Scanning
To scan a barcode
 Hold the analyzer approximately 3-12 inches and at a 10 degree angle from the barcode to be scanned.
 Press and hold down the Scan key to start the barcode scanner.
Title (with LTR): i-STAT Operations and Testing Procedure (LTR19694)
Last Approved: Gregory, Kimberly (Electronic Signature Timestamp: 8/11/2014 3:04:44 PM)
8



When the analyzer accepts the barcode, it will beep in acknowledgment and automatically turn off the beam.
The beam will also turn off after 3 seconds. You may view the data that scanned on the analyzer
After viewing the data, release the Scan key.
The meter contains a laser. DO NOT stare into the laser light or point it toward anyone.
Calculations
The i-STAT analyzer contains a microprocessor that performs all calculations required for reporting results.
Reporting results
The handheld analyzer will display test results once the cartridge has been unlocked and the Cartridge Locked
prompt disappears. Results display for 2 minutes. Recall results to the display screen by pressing the Menu key.
For dark units where it is difficult to view the display, pressing and holding the 0 (zero) key will turn the backlight on.
A. Suppressed Results
The i-STAT will not display results under three conditions:
1. Results outside of the system’s reportable ranges are flagged with a “<“(less than) or “>“(greater than)
indicating that the result is below the lower limit or above the upper limit of the reportable range
respectively. See Analytical Range under Reference Ranges for these values. If a flag of “<“(less
than) or “>“(greater than) occurs, SEND A SPECIMEN TO THE CLINICAL LABORATORY FOR
ANALYSIS.
Action: Follow unit protocol for critical results.
2. A result that is flagged with *** means that:
a. There is an interfering substance in the sample.
b. An individual sensor for that test has been compromised.
c. This particular test is a calculated result that depends on the results of another test that has ***.
d. The first time *** appears when a sample is tested, retest the sample using a new cartridge
(certain tests such as ACTk and PT/INR require fresh samples).
e. If *** appear again when the sample is tested a second time on the i-STAT SEND A SPECIMEN
TO THE CLINICAL LABORATORY FOR ANALYSIS.
f. If *** reappears on multiple patient samples, call Point of Care Testing for assistance.
3. Results will not be reported if a test cycle has a problem with the sample, calibrant solution, sensors,
mechanical, or electrical functions of the analyzer.
Action: Take the action displayed with the message that identifies the problem. Refer to the i-STAT
system manual’s troubleshooting section if necessary. Use a backup analyzer if necessary.
B. Transmitting Results to the Electronic Medical Record
1. Place the analyzer in the Downloader. The Power light must be green and the blue proximity light will
turn on automatically. When properly aligned, all unsent results will be automatically transmitted. The
analyzer does NOT need to be turned on. The message “Communication in progress” is displayed.
Verify that the display is turned off before removing from Downloader.
2. If unable to resolve data transmission issues, notify the Point of Care Testing Program.
Email: mghpoctcoordinators, pager: 35058
Preventive Maintenance
A. Rechargeable battery:
A nickel-metal-hydride (NiMH) rechargeable battery powers the analyzer. The battery recharges when the
analyzer is placed in a Downloader/Recharger. The battery pack can also be removed from the analyzer and
placed in the separate recharging compartment on the Downloader/Recharger. Full recharge from a
discharged state takes approximately 40 hours. The analyzer will display “Low Battery” when battery recharge
is needed. Spare batteries may be acquired from the POCT program staff.
B.
Disinfect exterior of analyzer between each patient with hospital-approved disinfectant, e.g. Sani wipe cloths or
Virex 256.
C.
CLEW/JAMS software updates:
Software updates performed twice per year re-establishes standardization values and incorporates refinements
to the internal quality monitoring system. External Quality Control and the Thermal Probe Check are performed
along with the CLEW update. The POCT Program will ensure CLEW updates are implemented prior to
software expiration.
Title (with LTR): i-STAT Operations and Testing Procedure (LTR19694)
Last Approved: Gregory, Kimberly (Electronic Signature Timestamp: 8/11/2014 3:04:44 PM)
9
Instrument Replacement Policy
A.
The replacement instrument policy applies to the following situations:
1. Any malfunction of the instrument
2. Any new instrument
3. Any major maintenance
B.
The POCT program will perform the following before the replacement instrument is installed:
1.
2.
3.
4.
5.
6.
7.
8.
External Simulator (Refer to the procedure for external Simulator)
Thermal probe check (Refer to the procedure for Thermal probe check)
CAL VER (Refer to the procedure for performing CAL VER)
Liquid Quality control (Refer to the procedure for performing liquid controls)
Run liquid QC, internal QC and external simulator for 10 days (on analyzers used for non-waived testing)
Document all the results on the Instrument Validation Log
Inform the POCT sites regarding the replacement of the instrument.
All the reports must be approved and signed by the Medical Director or designee.
Connectivity and LIS / HIS Down Time Procedure - OR
Title (with LTR): i-STAT Operations and Testing Procedure (LTR19694)
Last Approved: Gregory, Kimberly (Electronic Signature Timestamp: 8/11/2014 3:04:44 PM)
10
Title (with LTR): i-STAT Operations and Testing Procedure (LTR19694)
Last Approved: Gregory, Kimberly (Electronic Signature Timestamp: 8/11/2014 3:04:44 PM)
11
Connectivity and LIS / HIS Down Time Procedure - Imaging
Title (with LTR): i-STAT Operations and Testing Procedure (LTR19694)
Last Approved: Gregory, Kimberly (Electronic Signature Timestamp: 8/11/2014 3:04:44 PM)
12
Troubleshooting
A. Analyzer:
1. No display: Either if the display screen remains blank, after a cartridge has been properly inserted or after
the On/Off key has been pressed, the batteries should be replaced.
2. Cartridge Locked: Cartridge Locked appears on the screen during the testing cycle to indicate that the
cartridge or simulator is locked in the analyzer and should not be removed. A cartridge or simulator must
be removed only after the Cartridge Locked prompt disappears from the screen. Normally the analyzer
will reset and release the cartridge after the testing cycle is completed. If the analyzer cannot reset, the
Cartridge Locked prompt will remain on the screen. If this occurs, wait until the analyzer deactivates (display
screen blank) and press the On/Off key. The analyzer will try to reset. If the Cartridge Locked prompt
does not disappear, do not attempt to remove the cartridge. Contact the Point of Care Testing program:
Email: mghpoctcoordinators, pager: 35058
3. Low Battery: Recharge battery in Downloader/Recharge or swap with extra rechargeable battery.
4. Electronic simulator fail will appear if the analyzer has not successfully completed the electronic simulator
test. Refer to ELECTRONIC QUALITY CONTROL section.
5. CLEW expiring, Upgrade required, Point of Care Testing would need to perform a software update.
6. Messages and Quality Check Codes: If a problem is detected during a testing cycle, the cycle will be
stopped and a message box will appear on the screen. The messages will identify the code number. Refer
to i-STAT’s technical bulletin(s) to determine the meaning of the code number and the suggested action. If
the analyzer deactivates before the detected problem is addressed, the message box will reappear the next
time the On/Off key is pressed. Document all problems on the Instrument Corrective Action Log and notify
Point of Care Testing.
 Temperature out of Range error message: The analyzer must be moved to an area where the
temperature is between 18-30° C. Equilibration can take up to 30 minutes.
 Other error messages: Refer to the i-STAT System Manual’s Troubleshooting section for more
information.
7. i-STAT Dropped: Run simulator prior to patient testing.
8. Power Outage or Disaster scenario: Adaptors are available from the POCT program to allow i-STAT to run
on regular 9V batteries. Contact the POCT program for assistance.
B. Cartridge: *** Instead of Results:
1. Test the patient specimen with a new cartridge. If *** reappears, send the sample to the Clinical Laboratory
for analysis. The sample may contain a substance that interferes with the test.
2. If *** are obtained for the same test performed on an additional (different) patient, contact Point of Care
Testing for assistance: Email: mghpoctcoordinators, pager: 35058. (This problem may have been caused
by improper cartridge storage).
Technical support may be contacted 24 hours x 7 days at 1-800-366-8020.
Alternative method
Should the i-STAT System become inoperable for any reason, specimens should be collected and submitted to the
laboratory in accordance with the Laboratory Procedure Manual.
References
A. i-STAT1 System Manual, i-STAT Corporation, Princeton, N.J., 06/01/01.
B. Gornall, A.G.: Applied biochemistry of clinical disorders, Chapter 6, Respiratory Disorders, p. 94, 1980.
C. Gradwohl’s Blood gas analysis and acid-base balance: Principles & Techniques, Chapter 17, p. 351, 1980.
Fleisher & Schwartz.
D. Pesce, A. and Kaplan, Methods in Clinical Chemistry, C.V. Mosby Co., St. Louis, MO, 1987.
E. Tietz, Norbert W., Editor, Fundamentals of Clinical Chemistry, W.B. Saunders Co., Philadelphia, PA, 1986.
Title (with LTR): i-STAT Operations and Testing Procedure (LTR19694)
Last Approved: Gregory, Kimberly (Electronic Signature Timestamp: 8/11/2014 3:04:44 PM)
13
Reference Ranges
The tested ranges are those ranges that were tested during functional sensitivity and linearity testing.
Test
Calcium, ionized
Reference Range
1.14 - 1.30 mmol/L
Critical Value Range
(Approved by Med Poll
Committee)*
Analytical
Range
<0.8 or >1.54 mmol/L
0.25 – 2.50
PCO2 (arterial, venous, mix,
capillary) mmHg
Art: 14D, 35 - 42
Art: 0 30 - 35
Ven: 38 - 50
Mix: 35 - 50
Cap: 35 - 42
<20 or >75 mmHg, first
only (within past 24 hours)
pH (arterial, venous, mix,
capillary)
Art:
Ven:
Mix:
Cap:
<7.10 or >7.59 units, first
only (within past 24 hours)
PO2 (arterial, venous, mix, capillary)
mmHg
7.35 - 7.45
7.30 - 7.40
7.32 – 7.45
7.35 – 7.45
5 - 130
6.5 - 8.2
Art: 0-14 D: 60 - 80
Art: >14 D: 80 - 100
Mix: 40 - 90
Ven: 35 - 50
Cap: 45 - 60
5 - 800
<40 mmHg
Glucose (Adult)
70-100 mg/dL
<50 or >400
20 - 700
Glucose, plasma (newborn)
60-100 mg/dL
<40 or >500 mg/dL
20 - 700
Adult: 3.4 - 4.8 mmol/L
<1 mo: 4.0 - 5.6 mmol/L
<2.8 or >6.0 mmol/L
Sodium
135 - 145 mmol/L
<120 or >160 mmol/L
100 - 180
Lactate
0.5- 2.2 mmol/L
None
0.30 - 20.0
TCO2 (calculated)
Adult: 23.0 – 31.9 mmol/L
14 days-1 yr: 22.0 – 27.0 mmol/L
<14 days: 19.0 – 22.0 mmol/L
<11.0 or >40.0
HCO3 (calculated)
> 1 year: 24 - 30 mmol/L
14 days-1 yr: 22 - 27 mmol/L
0 - 14 days: 19 - 22 mmol/L
Potassium
2-9
5 - 50
None
1.0 - 85.0
BE (calculated)
None
None
(-30) - (+30)
SO2 (calculated)
None
None
0 - 100
PT 11.0-13.7 sec (age 6 months to
adult); PT-INR values for Warfarin
Anti-Coag Therapy: Standard
Intensity 2.0-3.0; High Intensity 2.53.5
EP Lab: Cardio aversion INR >2.0
All other EP procedures <3.5
Cath Lab: <1.5 Internal jugular
approach
<1.7 Femoral approach
<2.0 Radial approach
INR >5 POCT
INR >5 Core Lab
0.9-8.0
Troponin I
Negative
Borderline or positive
samples reflex to the core
laboratory for additional
testing
0.00-50.00
Creatinine
0.6-1.5 mg/dL
<10 yr: 0.3-1.0 mg/dL
None
0.5-14.0
PT/INR
Title (with LTR): i-STAT Operations and Testing Procedure (LTR19694)
Last Approved: Gregory, Kimberly (Electronic Signature Timestamp: 8/11/2014 3:04:44 PM)
14
*Clinical policy and Procedure Manual Laboratory Results: Guidelines for Retrieving and Reporting.
Patient
Addressograph
Activated Clotting Time Target Ranges
The Institution's Medical Staff believe Reference Ranges are not appropriate for Activated Clotting Time
(ACT), instead Target Values or Ranges that vary with procedures, protocols. In some cases, physician
clinical preferences may be appropriate. The following table provides guidelines that have been
established by site Medical Directors and staff and is intended to take the place of Reference Ranges. ACT
ranges are for heparin unless otherwise specified.
Location
Neuro-Radiology
Typical Target Range
Used at MGH
Qualifier or Description
Target Range is patient-specific and
1.5-2.0 times the baseline value.
subject to physician’s discretion
during the case.
250-330 seconds
EP laboratory
Respiratory:
Catheterization Lab
The Target Range is determined for
200-220 seconds with an upper limit of
each patient in the doctor’s order.
240 seconds
The Target Range is procedure and
patient-specific.
Interventional Catheterizations:
170-200 (upper limit 230) for patients on
platelet G2b-3a inhibitors and 200-220
(upper limit 240) for patients off platelet
G2b-3a inhibitors
Cardiac Surgery:
Patients undergoing Diagnostic
Patient will receive 0-4000 units heparin,
Catheterizations
regardless of ACT level.
Target ranges may vary with
procedure.
Baseline:
90 - 130 seconds
Therapeutic heparin drip:
160 - 180 seconds
(or 2 - 2.5 control aPTT + 100 seconds to
convert to an ACT value)
On Cardiopulmonary bypass
regardless of the surgical procedure:
>450 seconds
Off Cardiopulmonary bypass CABG:
>400 seconds
Therapeutic bivalirudin for CPB:
Initial ACT: >400 seconds
To remain in the Medical Record
Reviewed by: K Lewandrowski, MD
Date: 2/2/10
Title (with LTR): i-STAT Operations and Testing Procedure (LTR19694)
Last Approved: Gregory, Kimberly (Electronic Signature Timestamp: 8/11/2014 3:04:44 PM)
15
ACT (Kaolin) Cartridge Procedure
The Activated Clotting Time (ACT) is a measure of the time required for complete activation of the coagulation
cascade. The endpoint is indicated by the conversion of a thrombin substrate other than fibrinogen and an
electrochemical sensor is used to indicate that conversion. The ACT is primarily used to monitor a patient’s state
of anticoagulation due to heparin administered during a medical or surgical procedure (e.g. cardiac
catheterization, renal dialysis, Percutaneous Transluminal Coronary Angioplasty).
Specimen Collection
Suitable Specimens for ACT:
 Fresh whole blood (venous or arterial) without anticoagulant collected in a plastic syringe properly
labeled with the patient name, date of birth, or medical record number. If from an in-dwelling line, flush
the line with 5 mL saline and discard the first 5 mL of blood or three to six dead space volumes of the
catheter.
 Fresh whole blood (venous or arterial) collected in a plastic tube properly labeled with the patient name,
date of birth, or medical record number without anticoagulant, clot activators, or serum separators.
Device used to transfer sample to cartridge must be plastic.
 If a repeat measurement is needed, a fresh sample must be obtained
Criteria for Specimen Rejection:
 Evidence of clotting
 Other sample types such as urine, CSF and pleural fluid
 Specimen collected in a glass syringe or tube or with anticoagulant of any kind
NOTE: Avoid the Following Circumstances
 Drawing a specimen from an arm with an I.V
 Stasis (tourniquet left on longer than one minute before venipuncture)
 Extra muscle activity (fist pumping)
 Hemolysis (alcohol left over puncture site or a traumatic draw)
 Icing before filling cartridge
 Time delays before filling cartridge
Reagent Storage Requirements
Verification of Cartridge Storage Conditions during Shipment




A.
On receipt of new cartridges, verify that the transit temperatures were satisfactory using the fourwindow temperature indicator strip included with the cartridge boxes.
Read the temperature strip immediately since it will change with exposure to room temperature.
Record observations on the Record of Receipt that comes with each box and attach the receipt with
strip in the space shown on the Record of Receipt /QC Documentation Log.
Perform liquid QC on the cartridges and document results on the Record of Receipt /QC
Documentation Log.
Refrigerated Cartridges








Store at 2 - 8 C. Do not allow cartridges to freeze.
Verify that the cartridges stored in the refrigerator are all within the expiration date printed on the
boxes. Do not use after expiration date.
Verify that the refrigerator did not exceed the limits of 2 to 8°C (35 to 46°F).
Record the temperature of refrigerator on temperature log sheet.
Cartridges must remain in pouches until time of use. Do not use cartridge if pouch has been
punctured.
Individual cartridges may be used after standing five minutes at room temperature. An entire box of
cartridges should stand at room temperature for one hour.
Remedial Action:
If the temperature is outside the range of 2 to 8°C (35 to 46°F), quarantine the cartridges in the
storage refrigerator.
Notify the Site Coordinator of the temperature failure.
Title (with LTR): i-STAT Operations and Testing Procedure (LTR19694)
Last Approved: Gregory, Kimberly (Electronic Signature Timestamp: 8/11/2014 3:04:44 PM)
16

B.
DO NOT USE the cartridges from this refrigerator.
Room Temperature Cartridges:




C.
Cartridges at room temperature 18 to 30°C (64 to 86 °F) are good for 14 days. Mark the
room temperature expiration date on each cartridge.
Do not return cartridges to the refrigerator once they have been at room temperature.
Record the daily room temperature on temperature log sheet
Return any cartridges that are beyond the 14-day room temperature stability to the site coordinator
for tracking purposes.
i-STAT ACT Liquid Controls, Levels 1 and 2



Store at 2 to 8 C.
Do not use after expiration date on box and ampoules.
Record date and time removed from refrigeration. Control solutions may be stored at room
temperature for up to 4 hours. If left out longer than 4 hours at room temperature, they should be
discarded.
Competency Assessment
All operators must read the procedure manual and complete the “Operator Training Checklist” after initial training.
The competency assessment process is done following initial training, after six months during the first year, then
annually thereafter.
Competency is assessed using six methods, examples of which are below:
1.
2.
3.
4.
5.
6.
Successful performance of routine patient testing, verified by direct observation
Supervisor monitoring of the recording and reporting of test results
Supervisor review of intermediate test results, QC, proficiency tests, and preventative
maintenance performance
Successful performance of instrument maintenance function checks and calibration, verified by direct
observation
Testing previously analyzed samples, proficiency testing samples, internal blind testing samples
Assessment of problem-solving skills
Quality Control
Prior to performing quality control, determine if additional preparation is required.
Liquid Control preparation:
ACT Coagulation Control Level 1 and Level 2
1. i-Stat ACT controls are contained in 6-ml vials. Separate 6 ml vials contain 103 ml of calcium chloride
solution for reconstitution.
2. Prior to testing, vials containing the lyophilized plasma and CaCl2 reconstituting fluid should stand at room
temperature for a minimum of 45 minutes.
3. Reconstitute only one level of control plasma at a time. CONTROL SOLUTIONS MUST BE USED
IMMEDIATELY (less than 30 seconds) AFTER COMPLETING THE RECONSTITUTION AND MIXING
STEPS. Pour the entire contents of the calcium chloride vial into the lyophilized human plasma control vial.
4. Allow the vial to sit at room temperature for 1 minute
5. Mix the contents by swirling gently for 1 minute, then inverting slowly for 30 seconds.
6. Visually inspect the control vial to ensure that the sample is fully reconstituted. If not, discard the
reconstituted fluid and start over with fresh vials.
7. Using a plastic transfer pipette, plastic syringe, or plastic capillary tube with no anticoagulant, immediately
transfer the solution from the vial into the ACT cartridge. NOTE: additional ACT cartridges may be tested
with the remaining fluid if used within 30 seconds of complete reconstitution of the sample.
Running the controls:
1.
2.
3.
4.
Put on gloves.
Press the On/Off key
Press the Menu key
Press 3 to select Quality Tests.
NOTE: Always remember to analyze control materials in the Control pathway under the Quality
Tests option of the i-STAT 1 Analyzer Administration Menu. Do not analyze controls under the
Title (with LTR): i-STAT Operations and Testing Procedure (LTR19694)
Last Approved: Gregory, Kimberly (Electronic Signature Timestamp: 8/11/2014 3:04:44 PM)
17
Patient Test pathway. This must be done before inserting the test cartridge into the analyzer. The
analyzer allows 15 minutes to insert the cartridge after the last keystroke.
5.
6.
7.
8.
9.
10.
11.
12.
Press 1 to select Control.
Press “Scan” to scan Operator ID
Press “Scan” to scan Control Lot number
Press “Scan” to scan the cartridge lot number from the box or manually enter the number using
the keypad and press Enter.
Fill cartridge with the control (see below). Insert cartridge.
Enter 1 for i-STAT Level 1. Enter 2 for i-STAT Level 2.
Results will appear. At the bottom of one of the result pages, you will see 1- Test Options
Select 1.
You will now have the following options:
1- Next level
2- Repeat level
3- History
Using these options will allow you to continue your Quality Control testing without having to
start again at the top of the QC menu.
Control Evaluation:
1. Document results on the Record of Receipt /QC Documentation Log and compare results to the ranges to
determine acceptability.
2. The operator performing the quality controls is responsible for evaluating the controls and performing any
necessary corrective actions. If the results are out of range, corrective action must be documented on the
Record of Receipt /QC Documentation Log.
3. If all results are within expected ranges, use the cartridges as needed. Transmit the results when finished.
4. The result statistics will be reviewed monthly by the Medical director or Designee.
Quality Control Corrective Action:
1. If any liquid control result is unacceptable, patient testing can not be performed until the problem has
been resolved. Use another analyzer or send the specimen to the Clinical Laboratory.
2. Repeat any out of range control, verifying that the control procedure above has been followed carefully.
3. Document all the results and corrective actions on Record of Receipt /QC Documentation Log.
4. If new control solutions fail a second time, analyzer, control solutions and cartridges should be pulled
from use and the POCT program notified Email: mghpoctcoordinators, pager: 35058. An investigation
will be done to determine the cause of failure.
Testing Procedure
1. Follow hand hygiene protocol and put on gloves.
2. Turn the analyzer on and press 2 for i-STAT Cartridge.
3. Scan or enter the operator ID and patient CSN. Repeat if prompted.
4. Remove the cartridge from its pouch. Avoid touching the contact pads or exerting pressure over the
calibrant pack in the center of the cartridge.
5. Following thorough mixing of the sample, direct the dispensing tip or capillary tube containing the
blood into the sample well and dispense the sample until it reaches the fill mark on the cartridge
and the well is about half full.
6. Close the cover over the sample well until it snaps into place. (Do not press over the sample well.)
7. Insert the cartridge into the cartridge port until it clicks into place. When using an ACT cartridge, the analyzer
must remain horizontal during the testing cycle.
8. Select tests to be reported.
9. Enter additional parameters on the Chart page if required:


Choose the number corresponding to the type of sample used when prompted at the Sample Type
field.
Press the → key to return to the results page.
10. View results shown on the analyzer’s display screen.
Title (with LTR): i-STAT Operations and Testing Procedure (LTR19694)
Last Approved: Gregory, Kimberly (Electronic Signature Timestamp: 8/11/2014 3:04:44 PM)
18
11. Remove the cartridge after “Cartridge Locked” message disappears. The analyzer is ready for the next test.
12. Remove gloves and perform hand hygiene.
Documentation
Dock meter after each test performed to transmit results. During network downtime, results are held in analyzer until
they can be transmitted to patients’ electronic medical record.
Title (with LTR): i-STAT Operations and Testing Procedure (LTR19694)
Last Approved: Gregory, Kimberly (Electronic Signature Timestamp: 8/11/2014 3:04:44 PM)
19
CG8+, CG4+, and G3+ Cartridges Procedure
The CG8+ cartridge is designed to directly measure glucose, sodium, potassium, ionized calcium, Hematocrit, pH,
pCO2, and pO2, while calculating values for hemoglobin, TCO2, HCO3, BE and sO2. The CG4+ cartridge directly
measures pH, pCO2, and pO2, and lactate and calculates values for TCO2, HCO3, BE and sO2. The G3+ is similar to
the CG4+, except that lactate is not measured.
Sodium, Potassium, Ionized Calcium, pH, and pCO2
These analytes are measured potentiometrically by ion-selective electrode. Concentrations are calculated from the
measured potential through the Nernst equation.
1.
2.
3.
4.
5.
6.
7.
Sodium and potassium test results are important in the diagnosis and treatment of patients suffering from
hypertension, renal failure or impairment, cardiac distress, disorientation, dehydration, nausea and diarrhea.
Ionized calcium is the biologically active fraction of calcium. Through its role in a number of enzymatic
reactions and in membrane transport mechanisms, ionized calcium is vitally important in blood coagulation,
nerve conduction, and neuromuscular transmission and in muscle contraction. Measurements of ionized
calcium may be particularly useful under these clinical conditions: transfusion of citrated blood, liver
transplantation, open-heart surgery, neonatal hypocalcaemia, renal disease, hyperparathyroidism, malignancy,
hypertension and pancreatitis.
pH is an index of the acidity or alkalinity of the blood.
pCO2, along with pH, is used to assess acid-base balance. pCO2 is the respiratory component of acid-base
balance. pCO2 represents the balance between cellular production of CO2 and ventilatory removal of CO2.
Causes of primary respiratory acidosis (increase in pCO2) are airway obstruction, sedatives and anesthetics,
respiratory distress syndrome, and chronic obstructive pulmonary disease. Causes of primary metabolic
acidosis are ketoacidosis, lactate acidosis and diarrhea. Causes of primary respiratory alkalosis (decrease in
pCO2) are hypoxia due to chronic heart failure, edema and neurological disorders, and mechanical
hyperventilation. Causes of primary metabolic alkalosis are vomiting and antacid treatment.
pO2, a measurement of the pressure of oxygen dissolved in the blood, is measured
amperometrically. The oxygen permeates through a gas permeable membrane from the sample
into an internal electrolyte solution where it is reduced at the cathode. The oxygen reduction
current is proportional to the dissolved oxygen concentration.
Decreased pO2 may be caused by ventilation problems or by alteration in the flow of blood within the heart or
lungs.
Glucose is measured amperometrically. Hydrogen peroxide, produced by the oxidation of glucose, is oxidized
at an electrode to produce an electric current proportional to the glucose concentration. Glucose is the primary
source of energy for the body and the only source of energy for brain tissue. Measurements of blood glucose
levels are important for identification of both hypo- and hyperglycemia, as well as for managing patient
treatment to maintain glucose within designated limits.
Lactate is measured amperometrically. The enzyme lactate oxidase, immobilized in the
lactate biosensor, selectively converts lactate to pyruvate and hydrogen peroxide. The liberated
hydrogen peroxide is oxidized by the platinum electrode to produce a current that is proportional
to the sample concentration.
Elevated levels of lactate are mainly found in conditions of hypoxia such as shock and
hypovolumia; in conditions associated with diseases such as diabetes mellitus, neoplasia, and
liver disease.
(The i-STAT will not be used for Hematocrit and/or hemoglobin determinations)
Specimen Collection
Suitable Specimens for Cartridges for Blood Gases, Electrolytes, and Chemistries:

Fresh whole blood collected in a plain capillary collection tube or capillary collection tube balanced with
heparin
 Fresh whole blood collected in a collection tube with lithium or sodium heparin. Fill tube to capacity
 Fresh whole blood collected in a plain plastic syringe or in a blood gas syringe.
Criteria for Specimen Rejection:
 Evidence of clotting
 Specimens collected in vacuum tubes with anticoagulant other than lithium or sodium heparin
 Syringe for pH, PCO2 and PO2 with air bubbles in sample
 Incompletely filled vacuum tube for the measurement of ionized calcium
Title (with LTR): i-STAT Operations and Testing Procedure (LTR19694)
Last Approved: Gregory, Kimberly (Electronic Signature Timestamp: 8/11/2014 3:04:44 PM)
20

Other sample types such as urine, CSF and pleural fluid
NOTE: Avoid the Following Circumstances
 Drawing a specimen from an arm with an I.V
 Stasis (tourniquet left on longer than one minute before venipuncture)
 Extra muscle activity (fist pumping)
 Hemolysis (alcohol left over puncture site or a traumatic draw)
 Icing before filling cartridge
 Time delays before filling cartridge
 Exposing the sample to air when measuring pH, PCO2 and PO2
Reagent Storage Requirements
Verification of Cartridge Storage Conditions during Shipment

On receipt of new cartridges, verify that the transit temperatures were satisfactory using the fourwindow temperature indicator strip included with the cartridge boxes.

Read the temperature strip immediately since it will change with exposure to room temperature.

Record observations on the Record of Receipt /QC Documentation Log that comes with each box
and attach the receipt with strip, in the space shown on the log.

Perform liquid QC on the cartridges and document results on the Record of Receipt /QC
Documentation Log.
A.
Refrigerated cartridges






Store at 2 - 8 C .Do not allow cartridges to freeze.
Verify that the cartridges stored in the refrigerator are all within the expiration date printed on the
boxes. Do not use after expiration date.
Verify that the refrigerator did not exceed the limits of 2 to 8°C (35 to 46°F).
Record the temperature of refrigerator on temperature log sheet.
Cartridges must remain in pouches until time of use. Do not use cartridge if pouch has been
punctured.
Individual cartridges may be used after standing five minutes at room temperature. An entire box of
cartridges should stand at room temperature for one
Remedial Action:

If the temperature is outside the range of 2 to 8°C (35 to 46°F), quarantine the cartridges in the
storage refrigerator.

Notify the Site Coordinator of the temperature failure.

DO NOT USE the cartridges from this refrigerator.
B.
Room temperature cartridges:

Cartridges at room temperature 18 to 30°C (64 to 86 °F) are good for 2 months. Mark the room
temperature expiration date on each cartridge.

Do not return cartridges to the refrigerator once they have been at room temperature.

Record the daily room temperature on temperature log sheet

Return any cartridges that are beyond the 2 month room temperature stability to the site coordinator
for tracking purposes.
C.
i-STAT Aqueous Liquid Controls, Levels 1, 2 and 3

Store at 2 to 8 C.

Do not use after expiration date on box and ampoules.

Aqueous liquid controls used for oxygen measurement must stand at room temperature a minimum
of 4 hours before use. Controls used for all other analytes must equilibrate for 30 minutes at room
temperature.

Record date and time removed from refrigeration. Aqueous liquid controls expire 5 days after
removal from refrigeration.
Competency Assessment
All operators must read the procedure manual and complete the “Operator Training Checklist” after initial
training. For ACT, Blood Gases, and Lactate, the competency assessment process is done following initial
training, after 6 months during the first year, then annually thereafter.
Title (with LTR): i-STAT Operations and Testing Procedure (LTR19694)
Last Approved: Gregory, Kimberly (Electronic Signature Timestamp: 8/11/2014 3:04:44 PM)
21
Competency is assessed using six methods, examples of which are below:
1. Successful performance of routine patient testing, verified by direct observation
2. Supervisor monitoring of the recording and reporting of test results
3. Supervisor review of intermediate test results, QC, proficiency tests, and preventative maintenance
performance
4. Successful performance of instrument maintenance function checks and calibration, verified by direct
observation
5. Testing previously analyzed samples, proficiency testing samples, or internal blind testing samples.
6. Assessment of problem-solving skills
Quality Control
Prior to performing quality control, determine if additional preparation is required.
i-STAT Aqueous Liquid Controls: (For G3+, CG8+, CG4+)
1. Aqueous liquid controls require different temperature stabilization times depending on whether or not oxygen
is to be measured. If oxygen is to be measured, equilibrate the ampoule for 4 hours. If not, equilibrate the
ampoule for approximately 30 minutes at room temperature.
2. Hold the ampoule at the top, bottom (with forefinger and thumb), and shake 15-20 times (about 10 seconds)
to mix the solution. Tap the ampoule to restore the liquid to the bottom of the ampoule.
3. Open the ampoule by snapping off the tip. Protect fingers with gauze or use an ampoule breaker.
4. Immediately transfer control to syringe as follows: Use a clean 3 mL syringe. Replace attached needle with
a 19-gauge blunt tip needle.
5. Aspirate the control from the ampoule into the syringe. Be careful that air is not drawn in with the liquid.
Expel one or two drops of liquid before filling the cartridge. If desired, you may detach the blunt tip needle
before filling the cartridge.
6. Immediately transfer the solution into a cartridge.
7. Immediately seal the cartridge and insert it into an analyzer.
8. Do not use the solution left in a syringe or ampoule for additional testing of cartridges that contain sensors
for ionized calcium, pH, PCO2, or PO2. Open a new ampoule.
Running the controls:
1.
2.
3.
4.
Put on gloves.
Press the On/Off key
Press the Menu key
Press 3 to select Quality Tests.
NOTE: Always remember to analyze control materials in the Control pathway under the Quality
Tests option of the i-STAT 1 Analyzer Administration Menu. Do not analyze controls under the
Patient Test pathway. This must be done before inserting the test cartridge into the analyzer. The
analyzer allows 15 minutes to insert the cartridge after the last keystroke.
5.
6.
7.
8.
9.
10.
11.
12.
Press 1 to select Control.
Press “Scan” to scan Operator ID
Press “Scan” to scan Control Lot number
Press “Scan” to scan the cartridge lot number from the box or manually enter the number using
the keypad and press Enter.
Fill cartridge with the control (see below). Insert cartridge.
Enter 1 for i-STAT Level 1. Enter 2 for i-STAT Level 2.
Results will appear. At the bottom of one of the result pages, you will see 1- Test Options
Select 1.
You will now have the following options:
1- Next level
2- Repeat level
3- History
Using these options will allow you to continue your Quality Control testing without having to
start again at the top of the QC menu.
Control Evaluation:
Title (with LTR): i-STAT Operations and Testing Procedure (LTR19694)
Last Approved: Gregory, Kimberly (Electronic Signature Timestamp: 8/11/2014 3:04:44 PM)
22
1. Document results on the Record of Receipt /QC Documentation Log and compare results to the ranges to
determine acceptability.
2. The operator performing the quality controls is responsible for evaluating the controls and
performing any necessary corrective actions. If the results are out of range, corrective action
must be documented on the Record of Receipt /QC Documentation Log.
3. If all results are within expected ranges, use the cartridges as needed. Transmit the results when
finished.
4. The result statistics will be reviewed monthly by the Medical director or Designee.
Quality Control Corrective Action:
1. If any liquid control result is unacceptable, patient testing can not be performed until the problem has been
resolved. Use another analyzer or send the specimen to the Clinical Laboratory.
2. Repeat any out of range control, verifying that the control procedure above has been followed carefully.
3. Document all the results and corrective actions on the Record of Receipt /QC Documentation Log.
4. If new control solutions fail a second time, analyzer, control solutions and cartridges should be pulled from
use and the POCT program notified - Email: mghpoctcoordinators, pager: 35058. An investigation will be
done to determine the cause of failure.
Testing Procedure
1. Follow hand hygiene protocol and put on gloves.
2. Turn the analyzer on and press 2 for i-STAT Cartridge.
3. Scan or enter the operator ID and patient’s CSN. Repeat if prompted.
4. Remove the cartridge from its pouch. Avoid touching the contact pads or exerting pressure over the
calibrant pack in the center of the cartridge.
5. Following thorough mixing of the sample, direct the dispensing tip or capillary tube containing the
blood into the sample well and dispense the sample until it reaches the fill mark on the cartridge
and the well is about half full.
6. Close the cover over the sample well until it snaps into place. (Do not press over the sample well.)
7. Insert the cartridge into the cartridge port until it clicks into place. When using an ACT cartridge, the analyzer
must remain horizontal during the testing cycle.
8. Select tests to be reported.
9. Enter additional parameters on the Chart page if required:

Choose the number corresponding to the type of sample used when prompted at the Sample
Type field.
 Patient temperature can be entered as degrees Centigrade or Fahrenheit
 FIO2 can be entered as the numbers of liters or as a percentage of oxygen a patient is receiving
 Press the → key to return to the results page
10. View results shown on the analyzer’s display screen.
11. Remove the cartridge after “Cartridge Locked” message disappears. The analyzer is ready for
the next test.
12. Remove gloves and perform hand hygiene.
Documentation
Dock meter after each test performed to transmit results. During network downtime, results are held in analyzer until
they can be transmitted to patients’ electronic medical record.
Title (with LTR): i-STAT Operations and Testing Procedure (LTR19694)
Last Approved: Gregory, Kimberly (Electronic Signature Timestamp: 8/11/2014 3:04:44 PM)
23
Creatinine Cartridge Procedure
Creatinine is measured amperometrically. Hydrogen peroxide, which is produced through various enzymatic
reactions within the cartridge, is oxidized at the platinum electrode to produce a current which is proportional to the
sample creatinine concentration. Elevated levels of creatinine are mainly associated with abnormal renal function
and occur whenever there is a significant reduction in glomerular filtration rate or when urine elimination is
obstructed.
Specimen Collection
Suitable Specimens for measuring creatinine:


Venous whole blood collected in a collection tube with lithium or sodium heparin. Fill tube to capacity
Venous whole blood collected in a plain plastic syringe or in a blood gas syringe.
Criteria for Specimen Rejection:



Evidence of clotting
Specimens collected in vacuum tubes with anticoagulant other than lithium or sodium heparin
Other sample types such as urine, CSF and pleural fluid
NOTE: Avoid the Following Circumstances
 Drawing a specimen from an arm with an I.V
 Stasis (tourniquet left on longer than one minute before venipuncture)
 Extra muscle activity (fist pumping)
 Hemolysis (alcohol left over puncture site or a traumatic draw)
 Icing before filling cartridge
 Time delays before filling cartridge
Reagent Storage Requirements
Verification of Cartridge Storage Conditions during Shipment

On receipt of new cartridges, verify that the transit temperatures were satisfactory using the fourwindow temperature indicator strip included with the cartridge boxes.

Read the temperature strip immediately since it will change with exposure to room temperature.

Record observations on the Record of Receipt that comes with each box and attach the receipt, with
strip, in the space shown on the Record of Receipt /QC Documentation Log.

Perform liquid QC on the cartridges and document results on the Record of Receipt /QC
Documentation Log.
A.
Refrigerated cartridges

Store at 2 - 8 C. Do not allow cartridges to freeze.

Verify that the cartridges stored in the refrigerator are all within the expiration date printed on the
boxes. Do not use after expiration date.

Verify that the refrigerator did not exceed the limits of 2° to 8°C (35 to 46°F).

Record the temperature of refrigerator on temperature log sheet.

Cartridges must remain in pouches until time of use. Do not use cartridge if pouch has been
punctured.

Individual cartridges may be used after standing five minutes at room temperature. An entire box of
cartridges should stand at room temperature for one hour.
Remedial Action:

If the temperature is outside the range of 2° to 8°C (35 to 46°F), quarantine the cartridges in the
storage refrigerator.

Notify the Site Coordinator of the temperature failure.

DO NOT USE the cartridges from this refrigerator.
Title (with LTR): i-STAT Operations and Testing Procedure (LTR19694)
Last Approved: Gregory, Kimberly (Electronic Signature Timestamp: 8/11/2014 3:04:44 PM)
24
B.
C.
Room temperature cartridges:

Cartridges at room temperature 18° to 30°C (64 to 86 °F) are good for 14 days. Mark the roomtemperature expiration date on each cartridge.

Do not return cartridges to the refrigerator once they have been at room temperature.

Record the daily room temperature on temperature log sheet

Return any cartridges that are beyond the 14-day room temperature stability to the site coordinator
for tracking purposes.
i-STAT Aqueous Liquid Controls, Levels 1 and 3

Store at 2° to 8 C.

Do not use after expiration date on box and ampules.

Aqueous liquid controls used for oxygen measurement must stand at room
temperature a minimum of 4 hours before use. Controls used for all other analytes
must equilibrate for 30 minutes at room temperature.

Record date and time removed from refrigeration. Aqueous liquid controls expire 5 days
after removal from refrigeration.
Competency Assessment
All operators must read the procedure manual and complete the “Operator Training Checklist” after initial
training. Competency is assessed at orientation and annually using at least two of the following methods:
1.
2.
3.
4.
Performing a test on a blind specimen.
Supervisor observes performance of routine work.
Each user’s quality control performance is monitored.
Written testing specific to the method.
Quality Control
Prior to performing quality control, determine if additional control material preparation is required.
i-STAT Aqueous Liquid Controls: (For Creat)
1. Aqueous liquid controls require different temperature stabilization times depending on whether or not oxygen
is to be measured. If oxygen is to be measured, equilibrate the ampoule for 4 hours. If not, equilibrate the
ampoule for approximately 30 minutes at room temperature.
2. Hold the ampoule at the top, bottom (with forefinger and thumb), and shake 15-20 times (about 10 seconds)
to mix the solution. Tap the ampoule to restore the liquid to the bottom of the ampoule.
3. Open the ampoule by snapping off the tip. Protect fingers with gauze or use an ampoule breaker.
4. Aspirate the control from the ampoule into a clean 3 ml syringe, until it is half full.
5. Expel three drops of liquid before filling the cartridge
6. Immediately transfer the solution into a cartridge.
7. Immediately seal the cartridge and insert it into an analyzer.
Running the controls:
1.
2.
3.
4.
Put on gloves.
Press the On/Off key
Press the Menu key
Press 3 to select Quality Tests.
NOTE: Always remember to analyze control materials in the Control pathway under the Quality
Tests option of the i-STAT 1 Analyzer Administration Menu. Do not analyze controls under the
Patient Test pathway. This must be done before inserting the test cartridge into the analyzer. The
analyzer allows 15 minutes to insert the cartridge after the last keystroke.
5.
6.
7.
8.
Press 1 to select Control.
Press “Scan” to scan Operator ID
Press “Scan” to scan Control Lot number
Press “Scan” to scan the cartridge lot number from the box or manually enter the number using
the keypad and press Enter.
9. Fill cartridge with the control (see below). Insert cartridge.
10. Enter 1 for i-STAT Level 1. Enter 3 for i-STAT Level 3.
11. Results will appear. At the bottom of one of the result pages, you will see 1- Test Options
Select 1.
Title (with LTR): i-STAT Operations and Testing Procedure (LTR19694)
Last Approved: Gregory, Kimberly (Electronic Signature Timestamp: 8/11/2014 3:04:44 PM)
25
12. You will now have the following options:
1- Next level
2- Repeat level
3- History
Using these options will allow you to continue your Quality Control testing without having to start
again at the top of the QC menu.
Control Evaluation:
1. Document results on the Record of Receipt /QC Documentation Log and compare results to the ranges to
determine acceptability.
2. The operator performing the quality controls is responsible for evaluating the controls and
performing any necessary corrective actions. If the results are out of range, corrective action
must be documented on the Record of Receipt /QC Documentation Log.
3. If all results are within expected ranges, use the cartridges as needed. Transmit the results when
finished.
4. The result statistics will be reviewed monthly by the Medical director or Designee.
Quality Control Corrective Action:
1. If any liquid control result is unacceptable, patient testing can not be performed until the problem has been
resolved. Use another analyzer or send the specimen to the Clinical Laboratory.
2. Repeat any out of range control, verifying that the control procedure above has been followed carefully.
3. Document all the results and corrective actions on the Record of Receipt /QC Documentation Log
4. If new control solutions fail a second time, analyzer, control solutions and cartridges should be pulled from
use and the POCT program notified Email: mghpoctcoordinators, pager: 35058. An investigation will be
done to determine the cause of failure.
Testing Procedure
Note: Use only VENOUS blood for Creatinine testing!!
1. Follow hand hygiene protocol and put on gloves.
2. Turn the analyzer on and press 2 for i-STAT Cartridge.
3. Scan or enter the operator ID and patient CSN. Repeat if prompted.
4. Remove the cartridge from its pouch. Avoid touching the contact pads or exerting pressure over the
calibrant pack in the center of the cartridge.
5. Following thorough mixing of the sample, direct the dispensing tip or capillary tube containing the blood into
the sample well and dispense the sample until it reaches the fill mark on the cartridge and the well is about
half full.
6. Close the cover over the sample well until it snaps into place. (Do not press over the sample well.)
7. Insert the cartridge into the cartridge port until it clicks into place.
8. Select tests to be reported.
9. Enter additional parameters on the Chart page if required:
 Choose the number corresponding to the type of sample used when prompted at the Sample Type field.
 Patient temperature can be entered as degrees Centigrade or Fahrenheit
 Press the → key to return to the results page.
10. View results shown on the analyzer’s display screen.
11. Remove the cartridge after “Cartridge Locked” message disappears. The analyzer is ready for the next test.
12. Remove gloves and perform hand hygiene.
Documentation
Dock meter after each test performed to transmit results. During network downtime, results are held in analyzer until
they can be transmitted to patients’ electronic medical record.
Title (with LTR): i-STAT Operations and Testing Procedure (LTR19694)
Last Approved: Gregory, Kimberly (Electronic Signature Timestamp: 8/11/2014 3:04:44 PM)
26
PT/INR Cartridge Procedure
The i-STAT® PT/INR test is a whole blood determination of the prothrombin time used for monitoring oral
anticoagulant (warfarin) therapy. The test determines the time required for complete activation of the
extrinsic pathway of the coagulation cascade when initiated (activated) with a thromboplastin.
The PT/INR test result is reported as an International Normalized Ratio (INR) and, optionally, in seconds. The
INR is the recommended method of result reporting for monitoring of oral anticoagulant therapy.
Specimen Collection
Suitable Specimens for PT/INR
 Fresh whole blood without anticoagulant collected in a plastic syringe or plastic evacuated tube without
clot activators or serum separators. Device used to transfer sample to cartridge must be plastic.
 Fresh capillary whole blood dispensed directly into the cartridge from the finger
Criteria for Specimen Rejection
 Evidence of clotting
 Specimens collected in glass syringes or tubes or with anticoagulant of any kind
 Other sample types such as urine, CSF, and pleural fluid
NOTE: Avoid the Following Circumstances
 Drawing a specimen from an arm with an I.V.
 Stasis (tourniquet left on longer than one minute before venipuncture)
 Extra muscle activity (fist pumping)
 Hemolysis (alcohol left over puncture site, or a traumatic draw)
 Icing before filling cartridge
 Time delays before filling cartridge
Reagent Storage Requirements
Verification of Cartridge Storage Conditions during Shipment




A.
On receipt of new cartridges, verify that the transit temperatures were satisfactory using the fourwindow temperature indicator strip included with the cartridge boxes.
Read the temperature strip immediately since it will change with exposure to room temperature.
Record observations on the Record of Receipt that comes with each box and attach the receipt, with
strip, in the space shown on the Record of Receipt /QC Documentation Log.
Perform liquid QC on the cartridges and document results on the Record of Receipt /QC
Documentation Log.
Refrigerated cartridges






Store at 2 - 8 C. Do not allow cartridges to freeze.
Verify that the cartridges stored in the refrigerator are all within the expiration date printed on the
boxes. Do not use after expiration date.
Verify that the refrigerator did not exceed the limits of 2° to 8°C (35 to 46°F).
Record the temperature of refrigerator on temperature log sheet.
Cartridges must remain in pouches until time of use. Do not use cartridge if pouch has been
punctured.
Individual cartridges may be used after standing five minutes at room temperature. An entire box of
cartridges should stand at room temperature for one hour.
Remedial Action:



If the temperature is outside the range of 2° to 8°C (35 to 46°F), quarantine the cartridges in the
storage refrigerator.
Notify the Site Coordinator of the temperature failure.
DO NOT USE the cartridges from this refrigerator.
Title (with LTR): i-STAT Operations and Testing Procedure (LTR19694)
Last Approved: Gregory, Kimberly (Electronic Signature Timestamp: 8/11/2014 3:04:44 PM)
27
B.
Room temperature cartridges:




C.
Cartridges at room temperature 18° to 30°C (64 to 86 °F) are good for 14 days. Mark the roomtemperature expiration date on each cartridge.
Do not return cartridges to the refrigerator once they have been at room temperature.
Record the daily room temperature on temperature log sheet
Return any cartridges that are beyond the 14-day room temperature stability to the site coordinator
for tracking purposes.
i-STAT PT/INR Liquid Controls, Levels 1 and 2



Store at 2 to 8 C.
Do not use after expiration date on box and ampoules.
Record date and time removed from refrigeration. Control solutions may be stored at room
temperature for up to 4 hours. If left out longer than 4 hours at room temperature, they should be
discarded.
Competency Assessment
All operators must read the procedure manual and complete the “Operator Training Checklist” after initial
training. For PT/INR, the competency assessment process is done following initial training, after 6 months during
the first year, then annually thereafter.
Competency is assessed using six methods, examples of which are below:
1. Successful performance of routine patient testing, verified by direct observation
2. Supervisor monitoring of the recording and reporting of test results
3. Supervisor review of intermediate test results, QC, proficiency tests, and preventative maintenance
performance
4. Successful performance of instrument maintenance function checks and calibration, verified by direct
observation
5. Testing previously analyzed samples, proficiency testing samples, or internal blind testing samples.
6. Assessment of problem-solving skills
Quality Control
Prior to performing quality control, determine if additional preparation is required.
Liquid Control preparation:
PT/INR Control Level 1 and Level 2
1. i-Stat PT/INR controls are contained in 6-ml vials. Separate 6 ml vials contain 103 ml of calcium chloride
solution for reconstitution.
2. Prior to testing, vials containing the lyophilized plasma and CaCl2 reconstituting fluid should stand at room
temperature for a minimum of 45 minutes.
3. Reconstitute only one level of control plasma at a time. CONTROL SOLUTIONS MUST BE USED
IMMEDIATELY (less than 30 seconds) AFTER COMPLETING THE RECONSTITUTION AND MIXING
STEPS. Pour the entire contents of the calcium chloride vial into the lyophilized human plasma control vial.
4. Allow the vial to sit at room temperature for 1 minute
5. Mix the contents by swirling gently for 1 minute, then inverting slowly for 30 seconds.
6. Visually inspect the control vial to ensure that the sample is fully reconstituted. If not, discard the
reconstituted fluid and start over with fresh vials.
7. Using a plastic transfer pipette, plastic syringe, or plastic capillary tube with no anticoagulant, immediately
transfer the solution from the vial into the PT/INR cartridge. NOTE: additional PT/INR cartridges may be
tested with the remaining fluid if used within 30 seconds of complete reconstitution of the sample.
Running the controls:
1.
2.
3.
4.
Put on gloves.
Press the On/Off key
Press the Menu key
Press 3 to select Quality Tests.
NOTE: Always remember to analyze control materials in the Control pathway under the Quality
Tests option of the i-STAT 1 Analyzer Administration Menu. Do not analyze controls under the
Title (with LTR): i-STAT Operations and Testing Procedure (LTR19694)
Last Approved: Gregory, Kimberly (Electronic Signature Timestamp: 8/11/2014 3:04:44 PM)
28
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
Patient Test pathway. This must be done before inserting the test cartridge into the analyzer. The
analyzer allows 15 minutes to insert the cartridge after the last keystroke.
Press 1 to select Control.
Press “Scan” to scan Operator ID
Press “Scan” to scan Control Lot number
Press “Scan” to scan the cartridge lot number from the box or manually enter the number using
the keypad and press Enter.
Fill cartridge with the control (see below). Insert cartridge.
Enter 1 for i-STAT Level 1. Enter 2 for i-STAT Level 2.
Results will appear. At the bottom of one of the result pages, you will see 1- Test Options
Select 1.
You will now have the following options:
1- Next level
2- Repeat level
3- History
Using these options will allow you to continue your Quality Control testing without having to start again
at the top of the QC menu.
Control Evaluation:
1. Document results on the Record of Receipt /QC Documentation Log and compare results to the ranges to
determine acceptability.
2. The operator performing the quality controls is responsible for evaluating the controls and performing any
necessary corrective actions. If the results are out of range, corrective action must be documented on the
Record of Receipt /QC Documentation Log.
3. If all results are within expected ranges, use the cartridges as needed. Transmit the results when finished.
4. The result statistics will be reviewed monthly by the Medical director or Designee.
Quality Control Corrective Action:
1. If any liquid control result is unacceptable, patient testing can not be performed until the problem has been
resolved. Use another analyzer or send the specimen to the Clinical Laboratory.
2. Repeat any out of range control, verifying that the control procedure above has been followed carefully.
3. Document all the results and corrective actions on Record of Receipt /QC Documentation Log
4. If new control solutions fail a second time, analyzer, control solutions and cartridges should be pulled from
use and the POCT program notified Email: mghpoctcoordinators, pager: 35058. An investigation will be
done to determine the cause of failure.
Testing Procedure
Caution: The i-STAT PT/INR cartridge is designed to accept a sample between 20 and 45 micro liters.
A single drop of blood from either a finger puncture or as formed at the tip of a syringe will typically be within this
range. If a larger volume is delivered to the sample well, use caution when closing the cartridge as excess blood
may be expelled from the cartridge. The i-STAT PT/INR test can be performed using capillary or venous samples.
Finger Stick:
1. Follow hand hygiene protocol and put on gloves.
2. Turn the analyzer on and press 2 for i-STAT Cartridge.
3. Scan or enter the operator ID and patient CSN. Repeat if prompted.
4. Remove cartridge from foil pouch and place the cartridge on a flat surface Avoid touching the contact pads
or exerting pressure over the calibrant pack in the center of the cartridge.
5. Prepare lancet device and set aside until needed.
6. Clean and prepare the finger to be sampled. Allow finger to dry thoroughly before sampling.
7. Prick the bottom side of the fingertip with the lancet device.
8. Gently squeeze the finger, developing a hanging drop of blood and perform the test with the first drop of
blood. Avoid strong repetitive pressure (“milking”) as it may cause hemolysis or tissue fluid contamination of
the specimen.
9. Touch the drop of blood against the bottom of the sample well. Once in contact with the sample well, the
blood will be drawn into the cartridge.
Title (with LTR): i-STAT Operations and Testing Procedure (LTR19694)
Last Approved: Gregory, Kimberly (Electronic Signature Timestamp: 8/11/2014 3:04:44 PM)
29
10. Apply sample until it reaches the fill mark indicated on the cartridge.
11. Fold the sample closure over the sample well.
12. Press the rounded end of the closure until it snaps into place.
13. Insert the cartridge into the cartridge port until it clicks into place. When using a PT/INR cartridge, the
analyzer must remain on a level, vibration-free surface with the display facing up during the testing cycle.
14. Select tests to be reported.
15. Enter additional parameters on the Chart page if required:

Choose the number corresponding to the type of sample used when prompted at the Sample Type
field.

Patient temperature can be entered as degrees Centigrade or Fahrenheit

FIO2 can be entered as the numbers of liters or as a percentage of oxygen a patient is receiving

Press the → key to return to the results page
16. View results shown on the analyzer’s display screen.
17. Remove the cartridge after “Cartridge Locked” message disappears. The analyzer is ready for the next test.
18. Remove gloves and perform hand hygiene.
Venipuncture:






Collection technique resulting in good blood flow must be used.
The sample for testing should be drawn into a plastic collection device (either a plastic syringe or
plastic evacuated tube).
The collection device cannot contain anticoagulants such as heparin, EDTA, oxalate, or citrate.
The collection device cannot contain clot activators or serum separators.
If a second measurement is required, a fresh sample should be obtained.
Note: Some experts recommend drawing and discarding a (venous) sample of at least 1.0 mL prior to
drawing sample for coagulation testing.
1. Follow hand hygiene protocol and put on gloves.
2. Turn the analyzer on and press 2 for i-STAT Cartridge.
3. Scan or enter the operator ID and patient CSN. Repeat if prompted.
4. Remove cartridge from foil pouch and place the cartridge on a flat surface Avoid touching the contact pads
or exerting pressure over the calibrant pack in the center of the cartridge.
5. Apply sample until it reaches the fill mark indicated on the cartridge.
6. Fold the sample closure over the sample well.
7. Press the rounded end of the closure until it snaps into place.
8. Insert the cartridge into the cartridge port until it clicks into place. When using a PT/INR cartridge, the
analyzer must remain on a level, vibration-free surface with the display facing up during the testing cycle.
9. Select tests to be reported.
10. Enter additional parameters on the Chart page if required:

Choose the number corresponding to the type of sample used when prompted at the Sample Type
field.

Patient temperature can be entered as degrees Centigrade or Fahrenheit

FIO2 can be entered as the numbers of liters or as a percentage of oxygen a patient is receiving

Press the → key to return to the results page
11. View results shown on the analyzer’s display screen.
12. Remove the cartridge after “Cartridge Locked” message disappears. The analyzer is ready for the next test.
13. Remove gloves and perform hand hygiene.
Documentation
Title (with LTR): i-STAT Operations and Testing Procedure (LTR19694)
Last Approved: Gregory, Kimberly (Electronic Signature Timestamp: 8/11/2014 3:04:44 PM)
30
Dock meter after each test performed to transmit results. During network downtime, results are held in analyzer until
they can be transmitted to patients’ electronic medical record.
Troponin I Procedure
The i-STAT cTnI test is an in vitro diagnostic test for the quantitative measurement of cardiac troponin I (cTnI)
in whole blood or plasma samples. Cardiac troponin I measurements can be used as an aid in the diagnosis
and treatment of myocardial infarction and in the risk stratification of patients with acute coronary syndromes
with respect to their relative risk of mortality. Elevated levels of cardiac-specific troponins convey prognostic
information beyond that supplied by the patient’s clinical signs and symptoms, the ECG at presentation, and the predischarge exercise test. In addition, the ability of cTnI to be measured at the low-end of the concentration range
allows therapeutic intervention to be considered at any elevation above the normal range
Whole blood comes into contact with a sensor containing troponin-specific antibody and an enzyme conjugate
specific to a separate portion of the cTnI molecule. The cTnI within the patient’s sample becomes labeled with the
conjugate and captured onto the eletrochemical sensor. A wash phase results in an enzymatic cleaving of the
antibody/antigen/antibody sandwich, releasing a detectable electrochemical product.
Specimen Collection
Title (with LTR): i-STAT Operations and Testing Procedure (LTR19694)
Last Approved: Gregory, Kimberly (Electronic Signature Timestamp: 8/11/2014 3:04:44 PM)
31
Suitable Specimens for Troponin I/cTnI
 Fresh heparinized whole blood or plasma samples collected in syringes or evacuated tubes containing
lithium or sodium heparin. Collection tubes must be filled at least half full.
 Non heparinized whole blood samples tested within one minute of patient draw collected into a plastic
syringe or plastic evacuated tube containing no additives.
Criteria for Specimen Rejection
 Evidence of clotting
 The use of whole blood or plasma samples containing other anticoagulants such as EDTA, oxalate, and
citrate will cause a in decreased cTnI result.
 Other sample types such as urine, CSF, and pleural fluid
 Capillary tubes and direct skin punctures (e.g. fingersticks) should not be used with the cTnI cartridge
 Samples should not be used unless the blood collection tube is filled at least half full
NOTE: Avoid the Following Circumstances
 Drawing a specimen from an arm with an I.V
 Stasis (tourniquet left on longer than one minute before venipuncture)
 Extra muscle activity (fist pumping)
 Hemolysis (alcohol left over puncture site or a traumatic draw)
 Icing before filling cartridge
 Time delays before filling cartridge
Reagent Storage Requirements
Verification of Cartridge Storage Conditions during Shipment

On receipt of new cartridges, verify that the transit temperatures were satisfactory using the fourwindow temperature indicator strip included with the cartridge boxes.

Read the temperature strip immediately since it will change with exposure to room temperature.

Record observations on the Record of Receipt that comes with each box and attach the receipt, with
strip, in the space shown on the Record of Receipt /QC Documentation Log.

Perform liquid QC on the cartridges and document results on the Record of Receipt /QC
Documentation Log.
A.
Refrigerated cartridges

Store at 2 - 8 C. Do not allow cartridges to freeze.

Verify that the cartridges stored in the refrigerator are all within the expiration date printed on the
boxes. Do not use after expiration date.

Verify that the refrigerator did not exceed the limits of 2° to 8°C (35 to 46°F).

Record the temperature of refrigerator on temperature log sheet.

Cartridges must remain in pouches until time of use. Do not use cartridge if pouch has been
punctured.

Individual cartridges may be used after standing five minutes at room temperature. An entire box of
cartridges should stand at room temperature for one hour.
Remedial Action:
 If the temperature is outside the range of 2° to 8°C (35 to 46°F), quarantine the cartridges in the
storage refrigerator.
 Notify the Site Coordinator of the temperature failure.
 DO NOT USE the cartridges from this refrigerator.
B.
Room temperature cartridges:

Cartridges at room temperature 18° to 30°C (64 to 86 °F) are good for 14 days. Mark the roomtemperature expiration date on each cartridge.

Do not return cartridges to the refrigerator once they have been at room temperature.

Record the daily room temperature on temperature log sheet

Return any cartridges that are beyond the 14-day room temperature stability to the site coordinator
for tracking purposes.
C.
i-STAT Cardiac Marker Controls Levels 1, 2, and 3:

Store at ≤ –18°C (-1 °F) in a non-defrosting freezer. After thaw, the opened or
Title (with LTR): i-STAT Operations and Testing Procedure (LTR19694)
Last Approved: Gregory, Kimberly (Electronic Signature Timestamp: 8/11/2014 3:04:44 PM)
32
unopened vial is stable for 4 hours when capped and stored at 2-8°C (35-46°F).
Do not re-freeze
Competency Assessment
All operators must read the procedure manual and complete the “Operator Training Checklist” after initial training.
For cTnI, the competency assessment process is done following initial training, after 6 months during the first year,
then annually thereafter.
Competency is assessed using six methods, examples of which are below:
1. Successful performance of routine patient testing, verified by direct observation
2. Supervisor monitoring of the recording and reporting of test results
3. Supervisor review of intermediate test results, QC, proficiency tests, and preventative maintenance
performance
4. Successful performance of instrument maintenance function checks and calibration, verified by direct
observation
5. Testing previously analyzed samples, proficiency testing samples, or internal blind testing samples.
6. Assessment of problem-solving skills
Quality Control
Prior to performing quality control, determine if additional preparation is required.
Liquid Control preparation:
cTnI Control Level 1, 2, or Level 3
1. Remove vial from freezer and thaw at room temperature (18-30°C) for 15 minutes.
2. Thoroughly mix by gently swirling the bottle. Avoid foaming of the sample.
Running the controls:
1.
2.
3.
4.
Put on gloves.
Press the On/Off key
Press the Menu key
Press 3 to select Quality Tests.
NOTE: Always remember to analyze control materials in the Control pathway under the Quality
Tests option of the i-STAT 1 Analyzer Administration Menu. Do not analyze controls under the
Patient Test pathway. This must be done before inserting the test cartridge into the analyzer. The
analyzer allows 15 minutes to insert the cartridge after the last keystroke.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
Press 1 to select Control.
Press “Scan” to scan Operator ID
Press “Scan” to scan Control Lot number
Press “Scan” to scan the cartridge lot number from the box or manually enter the number using
the keypad and press Enter.
Dispense a drop of sample directly from the vial into the i-STAT cTnI cartridge and seal the cartridge.
Insert cartridge.
Enter 1 for i-STAT Level 1. Enter 2 for i-STAT Level 2, etc.
Results will appear. At the bottom of one of the result pages, you will see 1- Test Options
Select 1.
You will now have the following options:
4- Next level
5- Repeat level
6- History
Using these options will allow you to continue your Quality Control testing without having to start again
at the top of the QC menu.
Control Evaluation:
1. Document results on the Record of Receipt /QC Documentation Log and compare results to the ranges to
determine acceptability.
Title (with LTR): i-STAT Operations and Testing Procedure (LTR19694)
Last Approved: Gregory, Kimberly (Electronic Signature Timestamp: 8/11/2014 3:04:44 PM)
33
2. The operator performing the quality controls is responsible for evaluating the controls and performing any
necessary corrective actions. If the results are out of range, corrective action must be documented on the
Record of Receipt /QC Documentation Log.
3. If all results are within expected ranges, use the cartridges as needed. Transmit the results when finished.
4. The result statistics will be reviewed monthly by the Medical director or Designee.
Quality Control Corrective Action:
1. If any liquid control result is unacceptable, patient testing can not be performed until the problem has been
resolved. Use another analyzer or send the specimen to the Clinical Laboratory.
2. Repeat any out of range control, verifying that the control procedure above has been followed carefully.
3. Document all the results and corrective actions on Record of Receipt /QC Documentation Log
4. If new control solutions fail a second time, analyzer, control solutions and cartridges should be pulled from
use and the POCT program notified Email: mghpoctcoordinators, pager: 35058. An investigation will be
done to determine the cause of failure
Testing Procedure
1. Follow hand hygiene and put on gloves.
2. Turn the analyzer on and press 2 for i-STAT Cartridge.
3. Scan or enter the operator ID and patient CSN. Repeat if prompted.
4. Remove the cartridge from its pouch. Avoid touching the contact pads or exerting pressure over the
calibrant pack in the center of the cartridge.
5. Following thorough mixing of the sample, direct the dispensing tip or capillary tube containing the blood into
the sample well and dispense the sample until it reaches the fill mark on the cartridge and the well is about
half full.
6. Close the cover over the sample well until it snaps into place. (Do not press over the sample well.)
7. Insert the cartridge into the cartridge port until it clicks into place.
8. Select tests to be reported.
9. Enter additional parameters on the Chart page if required:

Choose the number corresponding to the type of sample used when prompted at the Sample Type
field.

Patient temperature can be entered as degrees Centigrade or Fahrenheit
11. Press the → key to return to the results page.
12. View results shown on the analyzer’s display screen.
13. Remove the cartridge after “Cartridge Locked” message disappears. The analyzer is ready for
the next test.
14. Remove gloves and perform hand hygiene.
Documentation
Dock meter after each test performed to transmit results. During network downtime, results are held in analyzer until
they can be transmitted to patients’ electronic medical record
Title (with LTR): i-STAT Operations and Testing Procedure (LTR19694)
Last Approved: Gregory, Kimberly (Electronic Signature Timestamp: 8/11/2014 3:04:44 PM)
34