ARTICLE IN PRESS Ultramicroscopy 99 (2004) 73–85 Measuring the absolute position of EELS ionisation edges in a TEM P.L. Potapov*, D. Schryvers EMAT, University of Antwerp, RUCA, Groenenborgerlaan 171, Antwerp B-2020, Belgium Received 13 March 2003; received in revised form 10 June 2003; accepted 7 July 2003 Abstract Measurements of absolute positions of electron energy loss spectroscopy (EELS) core-loss edges in a transmission electron microscopy (TEM) are hampered by noticeable errors caused by instabilities of the primary energy of the incident electrons. These instabilities originate from a continuous drift and random ripple of the high tension and are unavoidable in the present generation of TEM and scanning TEM microscopes. However, more precise measurements are desired, for instance, to study the shift of the edge onset between atoms of different valency or chemical environment, the so-called chemical shift. A solution to this problem is presented by collecting a series of short low-loss acquisitions immediately followed by core-loss ones. To ensure a minimal time lapse between core-loss and low-loss acquisitions, all operations must be computer controlled. Accumulation of a number of acquisitions and their summation corrected for energy drift allows to cancel the energy instabilities and to relate the core-loss EELS spectra to the absolute energy scale. A practical algorithm is presented as well as the necessary calibrations for such a procedure. Also, examples of spectra collected using this principle and the resulting measured chemical shifts in several metaloxides are presented. r 2003 Elsevier B.V. All rights reserved. Keywords: Electron energy loss spectroscopy; Instrument control and alignment 1. Introduction Today, electron energy loss spectroscopy (EELS) has become a widespread attachment to transmission electron microscopes (TEM). The position and intensity of an EELS ionisation edge provide chemical information whereas its fine structure (energy loss near edge structure, ELNES) contains information on the bonding states in the *Corresponding author. Tel.: +32-3-218-0472; fax: +32-3218-0257. E-mail address: [email protected] (P.L. Potapov). material. It should be realised, however, that the majority of experimental data on EELS edges, as reported in the literature and obtained in current day microscopes, have a precision of 0.5 eV or worse. However, more accurate data are desired for several reasons. First, a given ionisation edge is known to appear at slightly different positions depending on the actual electronic structure of a given material [1–8]. This difference in onset position, also referred to as the chemical shift, in most cases does not exceed 1 eV and therefore cannot be properly reproduced by the existing experimental methods. Still, a chemical shift 0304-3991/$ - see front matter r 2003 Elsevier B.V. All rights reserved. doi:10.1016/S0304-3991(03)00185-2 ARTICLE IN PRESS 74 P.L. Potapov, D. Schryvers / Ultramicroscopy 99 (2004) 73–85 indicates changes in the relative energy levels of the inner electron shells with respect to the Fermi level and might provide important information complementary to the more widely used ELNES information. Second, the ongoing progress in filtered imaging methods asks for the application of smaller energy windows for valency mapping and spectrum-imaging. The use of windows less than 1 eV wide is now under discussion, but this small window size would require more precise measurements of the onset positions. Third, quantitative data of the absolute positions of the edge onset is important in view of theoretical understanding of the materials. Until recently, ab initio calculations could reproduce the real binding energies only approximately, but new advances in the calculation methods yield precisions comparable to those of conventional experiments [9,10]. Comparing different theoretical approaches would definitely require more precise experimental reference data. The main reason for the limited precision of the energy measurements in EELS spectra can be found in the instability of the energy of the primary electrons. The only available reference point for calibration of the origin of the energy axis is the position of the zero-loss peak, which consists of those electrons with primary energy. However, most ionisation edges appear in energy loss regions of a few to several hundred eV and cannot be observed in the same spectrum as the zero-loss peak unless a very rough energy dispersion is used. The common way to combine a high energy dispersion with an accurate energy calibration is by shifting the energy range between a lowloss region to a core-loss one by a small variation of the high tension of the microscope or by applying a voltage to the electrostatic drift of tube in the spectrometer. However, the precision of these shifts and the related energy measurements is limited by the energy instabilities of the currently available TEM and related scanning TEM (STEM) instruments. This limitation results in an unknown extra energy difference between any two consecutive acquisitions. At present there are only a few instruments in the world with a dedicated design in which these energy instabilities are minimised. For example, it is possible to decelerate the electrons before the spectral analysis and if the spectrometer is floating at the cathode potential, the acceleration and deceleration voltages cancel out exactly [11,12]. As a result, up to now, any reliable data about chemical shifts have almost exclusively been obtained from measurements on dedicated instruments [7,8]. However, those instruments are difficult in handling and are unlikely to become widespread in the near future. In the present work we suggest a new method enabling spectroscopists to measure precisely the absolute position of onsets using regular TEM instruments and standard spectrometers without the need for any additional hardware installations. The method is based on an computer script, which corrects for instabilities in the primary energy of the electrons. At this point we need to differentiate clearly between the precision and accuracy of a measurement. Better precision implies a smaller spread of the measured data, i.e., higher reproducibility of the measurements in a given instrument, while better accuracy means closer correspondence of the measured number to the real absolute value. In other words, precision deals with random errors while accuracy deals with systematic errors. Two other important terms in EEL spectroscopy are instrumental energy resolution and energy dispersion of the spectrometer. The former characteristic indicates how much an infinitely narrow peak (d-function) will be broadened by instrumental factors. The latter is simply the energy window covered by a single registration channel. In principle, the precision of measurements of the peak positions can be better than the energy resolution and energy dispersion, provided that the peak profile is fitted by the proper model and sufficient counts are collected. The script that has been developed yields an improved precision as well as accuracy for the measured position of the edge onset. 2. Instrumentation The present work is primarily performed on a Philips FEG CM30 ultratwin microscope with an external GIF200 detector. The primary energy of 300 kV is maintained by a Philips high-tension ARTICLE IN PRESS P.L. Potapov, D. Schryvers / Ultramicroscopy 99 (2004) 73–85 generator. The extractor voltage of the gun is reduced to improve the energy resolution, which is 0.75–0.8 eV in the present experiments. To optimise the precision of the measurements the finest energy dispersion available in the GIF200, namely 0.05 eV/channel, is chosen. The probe size is about 100 nm and the typical thickness of the investigated samples is 0.6–1.0 times the respective mean free path of the electron. All spectra are collected in image mode with the aperture in the objective lens plane providing a collection angle of 14 mrad. The second order aberrations in the GIF are carefully tuned, so that the energy resolution is almost the same when using a 2 mm or 0.6 mm GIF entrance aperture. This way, the 2 mm aperture can be used, increasing the intensity by a factor of 10 with respect to the 0.6 mm one. Tests of the energy stability are also conducted on a Jeol FEG JEM3000F microscope equipped with a GIF2000 and a Philips FEG Technai microscope with a monochromator and a high resolution-GIF detector (National Centre for HREM, Delft, Netherlands). The developed scripts are written in the digital micrograph (DM) code and can be implemented in any DM environment. 3. Factors affecting the precision and accuracy of absolute measurements Fig. 1a shows the variation with time of the primary electron energy measured through the GIF spectrometer in the CM30 and JEM3000F microscopes (lower section). The average energy of the electrons collected in a single acquisition is determined as the centre of mass of the zero-loss peak. This zero-loss peak is collected at a rate of 5 times per second with an exposure time of 0.02 s for each individual acquisition. In both CM30 and JEM3000F, a random energy ripple is detected having a dominant frequency between 1 and 3 Hz and a peak-to-peak magnitude up to 0.25 eV in the CM30 and 0.35 eV in the JEM3000F. The top part in Fig. 1a shows the variation of the high tension in the CM30 measured 5 times per second by a 8.5 digits voltmeter connected to the reference voltage in the high tension generator [13]. Taking into 75 account the actual precision of the voltmeter, which is 70.3 ppm (i.e., 0.09 V) at such a small integration time, the observed magnitude and frequency of the high-tension ripple correspond well to those detected via the zero-loss peak. Thus, the ripple of the primary energy originates from micro-instabilities of the high tension and seems to be attributed to the quality of the electronic feedback system in the high-tension generators. On top of this ripple, a continuous energy drift of 0.2 eV/min is observed in both CM30 and JEM3000F microscopes. Simultaneous measurements of the reference voltage and the zero-loss peak position reveal that also this drift mainly originates from the drift of the high tension. More details on this procedure can be found in a forthcoming paper [13]. It is interesting to compare these regular TEM instruments with other instruments specially designed for higher energy stability. Fig. 1a also shows stability tests performed on the dedicated Technai microscope at the Technical University Delft via the zero-loss position measurement. In this microscope, a monochromator is installed and the high-tension circuit is significantly improved in order to meet high-energy resolution requirements. As a result, the energy ripple is reduced to 0.08 eV and the drift to 0.04 eV/min. In practical terms, the apparent magnitude of the ripple can be reduced by optimising the exposure time. Fig. 1b shows the effect of a numerical filter averaging over 5 values from adjacent channels of the original ripple curve. This graph demonstrates that an increase of the exposure time from 0.02 to 1 s would result in a decrease of the apparent ripple peak-to-peak magnitude from 0.25 to 0.15 eV due to the averaging of fast changes within a single 1 s exposure. Still, random-like energy jumps up to 0.12 eV between any two subsequent acquisitions would remain. In practice, real experiments with 1 s exposure indicate that the latter also amounts to 0.15 eV. During conventional operation of an EELS experiment a pause between low-loss and coreloss acquisitions lasts more than 1 min because the operator should manually tune the illumination conditions and shift the energy region. Also, the ARTICLE IN PRESS 76 P.L. Potapov, D. Schryvers / Ultramicroscopy 99 (2004) 73–85 Fig. 1. (a, bottom) Variation of electron primary energy measured by EELS in different microscopes and (a, top) variation of high tension in CM30 measured by a 8.5 digits voltmeter; and (b) effect of a smoothing filter on the apparent magnitude of the energy ripple. The vertical spacing between the horizontal grid lines is 0.1 eV for (a) and 0.05 eV for (b). exposure time for core-loss acquisitions is usually chosen to be much larger than that for low-loss ones so that a new dark current reference image needs to be collected before each acquisition. Combining both the ripple and drift factors determined above, the typical random error in determining edge positions is expected to be about 0.5 eV or worse when standard microscopes and operation routines are used. Still, in the case of relative measurements, for instance when measuring a chemical shift of a given edge between two regions in one sample, the precision can be improved to 0.2–0.3 eV because the intensity does not have to be retuned and the ripple is partially averaged by long core-loss acquisitions. Furthermore, a systematic error of the same magnitude and related to the imperfect calibration of the high tension or the drift tube might be induced. This calibration is normally done by using the Ni L3 onset in nickel oxide and, taking ARTICLE IN PRESS P.L. Potapov, D. Schryvers / Ultramicroscopy 99 (2004) 73–85 77 Fig. 2. Calibration of the energy dispersion of the CCD camera for different settings and given in number of channels. the uncertainty in any energy measurement mentioned above into account, is only precise within a range of 70.5 eV. This calibration error is specific for each calibration procedure made on a given instrument. Additionally, in our opinion, the standard 854 eV value currently adopted by Gatan for the Ni L3 onset in NiO [14,15] is noticeably higher than the true value (see below). Thus, the majority of EELS spectra published in the literature are systematically shifted to higher energies. Note that the latter systematic error can easily be taken into account when comparing spectra from different sources provided that the reference point is properly described. Another factor which can potentially affect the precision and accuracy of the measurements is a non-linearity of the energy dispersion of the spectrometer. Ideally, the zero-loss peak and the examined core-loss feature should be put at the same channel by choosing an appropriate energy shift. In reality, however, even the sharpest features of ionisation edges are much broader than the zero-loss peak and are spread over 30–100 channels depending on the nominal energy dispersion used. Moreover, as will be demonstrated below, the zero-loss peak and core-loss onset should sometimes be put at different channels to avoid artefacts in the spectra. Thus, a range of at least a few hundred channels in the CCD camera needs to be properly calibrated. To calibrate the energy dispersion along the CCD camera the zeroloss peak is first precisely positioned at channel 100 (which was taken as a reference) and then shifted to the desired channel by the drift tube (for the calibration of the drift tube, see Section 6). The resulting position of the peak is then measured and compared with that expected from the nominal energy dispersion. This procedure was repeated several times over the same interval of channels to suppress the energy ripple effect and the smoothed average curve is shown in Fig. 2. The deviation from the expected position is maximal for an energy dispersion of 1 eV/channel and reduces with decreasing nominal dispersion. For the 0.05 eV/channel dispersion used in the present measurements, the deviation is negligible up to channel 500, the region of interest for the present experiments. 4. Acquisition script minimising the effects of energy instability In order to minimise the effects of the energy instability of the primary electrons a computer script was written performing several acquisitions in a row and in the mean time adjusting different artefacts and performing internal calibrations. To ARTICLE IN PRESS 78 P.L. Potapov, D. Schryvers / Ultramicroscopy 99 (2004) 73–85 equally average the energy ripple during all acquisitions a 1 s exposure time is adopted for both low-loss and core-loss acquisitions in the script. When switching from core-loss to low-loss acquisitions, the illumination is reduced by changing the current in the C2 condenser lens thus defocussing the spot. The energy region is shifted by applying the drift tube, a procedure which is much faster than changing the high tension. Both these operations are computer controlled minimising the pause between acquisitions. All acquired spectra are then gain normalised and corrected for dark current references, which are prepared at the start of the execution of the script. The schematic of the developed script is shown in Fig. 3. First a spectrum in a low-loss region is collected followed by the calculation of the position of the zero-loss peak. Based on this calculated value, a small drift tube voltage typically less than 1 V is applied to position the centre of the zero-loss peak exactly at channel 100. Then, a number of cycles each consisting of lowloss and core-loss acquisitions is executed. The pause between low-loss and core-loss acquisitions is only about 1 s, thus, according to Fig. 1b, the energy jump is expected to be smaller or equal to 0.15 eV, which yields a maximum shift of 3 channels when using the 0.05 eV/channel dispersion. After each cycle, the centre of the zeroloss peak is again calculated, and typically floats within the expected window of 3–4 channels around channel 100. This fluctuation cannot be avoided and results from the ripple of the high tension together with the continuous drift as described above. Then the data of the low-loss spectrum are shifted by several channels to realign the zero-loss peak with channel 100 and the same shift is applied to the subsequent core-loss spectrum. In those cases when the calculated deviation exceeds 3 channels, the drift tube correction is adjusted returning the zero-loss peak back to channel 100. With this combination of software and hardware corrections, respectively, the position of the zero-loss peak in each cycle can be maintained at channel 100. Thus, the continuous drift is cancelled allowing the script to execute as long as necessary. Still, the entire ripple of the primary energy is not completely suppressed by this operation since energy jumps of 0.15 eV between two subsequent low-loss and core-loss acquisitions within one cycle are still possible. However, this jump is random in nature and accumulation of a number of low-loss and coreloss spectra can average and minimise the effect of this feature. In practice, after about 15 cycles the position of a given peak converges to a fixed value. Fig. 3. Schematic of the script applied to cancel energy instabilities in a microscope. ARTICLE IN PRESS P.L. Potapov, D. Schryvers / Ultramicroscopy 99 (2004) 73–85 Finally the script yields a cumulative low-loss spectrum with the zero-loss peak placed at channel 100. The peak centre inside channel 100 is further calculated more precisely by measuring the centre of mass of the channels located within the halfwidth of the peak. Although the primary motivation for applying cycled acquisitions is improvement of the precision of measuring the peak positions, the suggested procedure exhibits another advantage. The instrumental resolution is usually measured in the lowloss region using short acquisitions. Long acquisitions, typical for core-loss regions, can result in additional energy broadening due to the energy instabilities in the course of acquisition. For instance, applying a 1 min acquisition might result in a further 0.2–0.3 eV deterioration of the effective energy resolution. In the developed script, 1 s acquisitions are applied and summed while taking the energy drift between them into account, resulting in a minimal deterioration of the original energy resolution. Regardless of the number of cycles, the energy broadening is determined by the very negligible energy drift within the 1 s time period of a single acquisition. Thus, the present procedure also combines a minimal energy broadening typical for short acquisitions with the good statistics achieved by long acquisitions.1 The disadvantage of the present version of the script is that the convergence angle changes with changing the C2 current. In particular, it implies that the low-loss spectrum cannot be used for deconvoluting the core-loss one from plural scattering. Further developments allowing us to register both low-loss and core-loss spectra under identical collection and convergence angles and applicable to both image and diffraction modes are in progress. 5. Artefacts in spectra Since two subsequent acquisitions are collected quite rapidly one after the other, the following 1 During preparation of this paper, other work has been published, which describes the similar script aimed to improve the energy resolution [16]. With minor changes, their script can be also employed for measuring the absolute onset positions. 79 artefact has been found. When executing the script without a sample in the electron beam path, a small spot is observed in the image of the second spectrum exactly at the place of the former zeroloss peak. This is shown in Fig. 4a where the dynamic range for the core-loss case (bottom image) is stretched by a factor 1000 with respect to that of the low-loss case (top image). A variation of the time delay between the low-loss and coreloss acquisitions demonstrates that this echo peak quickly decays with time and decreases to the noise level after approximately 5 s. This effect is similar to the incomplete discharge of the zero-loss peak observed after oversaturation of a CCD camera. In that case the charge cannot be read completely during the first following readout causing a partial memory of the previous acquisition. However, the incomplete readout is known to be recovered by applying a series of short acquisitions. In contrast, the echo peak observed in the present experiments decreased only with passing time and not with the number of readouts preceded the final acquisition. Fig. 4b shows the intensity of the echo peak after a series of readouts with different exposure times, as indicated. Regardless of the number of intermediate readouts, the decay curves practically coincide with one another and with the case when no readouts are applied. By fixing the delay time and varying the intensity of the initial zero-loss peak, an almost linear relation between the initial and echo peak intensities is found, as shown in Fig. 4c. Thus, the echo peak is not related with any oversaturation of the CCD camera but can possibly be attributed to the limited time response of the scintillator layer in front of the CCD array. Although the echo peak observed after a delay of 1 s is a thousand times lower in intensity than the original zero-loss peak, it is still noticeable in a low-intensity core-loss spectrum. The only way to cancel this artefact is by reducing the intensity of the zero-loss peak down to 10% of the saturation level. In this case, the statistics of a low-loss spectrum are still sufficient to calculate precisely the centre of the zero-loss peak while the intensity of the echo peak drops to the noise level as seen in Fig. 4c. Nevertheless, to avoid any potential influence of this artefact on the measurements, the edge onset of the core-loss spectrum was ARTICLE IN PRESS 80 P.L. Potapov, D. Schryvers / Ultramicroscopy 99 (2004) 73–85 Fig. 4. (a) Appearance of an echo peak in the image of a second spectrum, (b) dependency of the echo peak intensity on the pause between low-loss and core-loss acquisitions (number at each experimental point denotes the amount of intermediate readouts preceding the final acquisition) and (c) the echo peak intensity as a function of the original zero-loss peak. placed around channel 200 by the applied drift tube. Another disturbing feature is related with changing the beam intensity between low-loss and core-loss acquisitions. As described above, the intensity is tuned by changing the current in the C2 condenser lens. Unfortunately, it is noticed that the C2 current slightly affects the position of the zero-loss peak. This is shown by executing the script with a zero drift tube value yielding two lowloss spectra, one for defocussed and the other for focussed illumination. As for the latter C2 setting the intensity becomes very high, the exposure is reduced to 0.01 s, which results in a higher random energy jump between acquisitions as seen in Fig. 1b. This is then compensated by increasing the number of cycles up to 120. Fig. 5 shows the results of the measurements demonstrating that the apparent electron energy can be shifted by several tenths of eV when changing the C2 current with respect to a chosen reference value of 2500 mA. The energy shift is close to linear in the large range of C2 currents while the intensity entering the spectrometer, also shown in Fig. 5, ARTICLE IN PRESS P.L. Potapov, D. Schryvers / Ultramicroscopy 99 (2004) 73–85 81 distribution of a random variable, the average value Emeas measured in N cycles should occur with a 95% probability within the following confidence interval of precision sE near the true value Etrue s Eexp ¼ Emeas 71:96 pffiffiffiffiffi ¼ Emeas 7sE ; ð1Þ N where s is the standard deviation of a single measurement, which is estimated from the variation of the zero-loss peak position Ezl from cycle to cycle sffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi P ðEzl E% zl Þ2 s¼ : ð2Þ N Fig. 5. Effect of changing the current in the C2 condenser lens on the intensity coming into a GIF and the apparent position of the zero-loss peak. increases dramatically when the beam passes through the focus point at which its size becomes comparable with the spectrometer entrance. To correct for this C2 effect, the two standard C2 current settings were chosen at the beginning of every session, one for the low-loss and another for the core-loss region, aiming at a large intensity difference with a minimal energy shift. The remaining energy shift between the two C2 settings is measured and accounted for when calculating the reference point in the core-loss spectra. However, this shift can be calibrated only with the precision of 70.03 eV (see below), which introduces an additional uncertainty in the determination of the core-loss onsets. A possible solution for this could be to attenuate the intensity without any changing currents in the microscope lenses but rather by changes inside the GIF, for instance by wobbling the beam in the direction perpendicular to the energy dispersion axis. 6. Precision and accuracy of measuring the core-loss onsets As the energy jump between low-loss and coreloss acquisitions is random, the precision of the present measurements can be estimated by the standard statistics theory. Assuming a Gaussian When 1 s acquisitions are used, the standard deviation s of the zero-loss peak position is measured to be about 0.1 eV, which agrees well with the data displayed in Fig. 1b. This results in a 0.05 eV precision sE for a typical series of 20 cycles as used in the present work. The energy shift occurring with changing C2 current gives an additional error in measuring the onset. Calibration of this shift performed within the same script shows a higher standard deviation, namely 0.15 eV, because shorter exposure times are used. Since this calibration is needed only once a session, a higher number of cycles can be applied without affecting the efficiency of the procedure. Applying 120 cycles reduces the calibration error to 0.03 eV. As the effective duration of each cycle is only 3 s and provided that the continuous drift does not exceed 0.2 eV/min the contribution of the continuous energy drift within a given cycle should be less than 0.01 eV and can be neglected. Combining the sE with the calibration error of the C2 adjustment yields a total precision of 0.1 eV. The systematic error specific for a given instrument comes from the procedure of the drift tube calibration, which, following the above reasoning, can be performed with a precision of 0.1 eV. Since the drift tube can be considered as a linear feature, its calibration can be performed at large values, e.g. 1000 V, which yields a relative uncertainty of 0.01% which amounts to much smaller absolute values when decreasing the applied drift tube voltage, i.e., for edges with lower binding energies (e.g., 200 V70.02 V). ARTICLE IN PRESS 82 P.L. Potapov, D. Schryvers / Ultramicroscopy 99 (2004) 73–85 Table 1 Absolute position of the Ni L3 edge measured by dedicated EELS spectrometers [18,21] and XPS [22] Batson [18] Fink et al. [21] Fuggle et al. [22] Material Starting point (eV) Left half-height (eV) Maximum (eV) NiO Ni NiOa Ni NiOa — 851.9 — — — 852.0 852.5 852.9a 852.8 853.0a 852.75 853.0 853.2a 853.4a 853.6a The numbers for NiO oxide are extrapolated assuming a 0.2 eV chemical shift between Ni metal and NiO. a Extrapolated. A further uncertainty relates with the knowledge of the exact absolute position of the Ni L3 line in NiO, which is usually used for the drift tube calibration. Simple calibration of the voltage with a voltmeter is not entirely accurate in spectrometers in which an extra potential is applied to the drift tube, because the electron beam is likely not to be precisely centred in the drift tube. This typically creates an extra deflection of the beam as it is accelerated at the drift tube entrance and decelerated at the tube exit. The deflection results in a spurious shift of the energy at which a feature appears in the spectrum. In Gatan spectrometers, this effect typically amounts to 1–2 eV at an energy loss of 1000 eV [17]. At present, the value of 854.0 eV of the Ni L3 onset (defined as the left halfheight of the peak) in NiO is generally adopted for GATAN spectrometers [14,15]. However, data from dedicated EELS spectrometers in which energy instabilities are diminished support a lower value. Batson, using a dedicated STEM at the T.J. Watson IBM centre, obtained values of 852.0 eV for the half-height on the left shoulder of the EELS L3 peak and 852.75 eV for the maximum of the L3 peak in NiO [18].2 Fink et al. reported a value of 851.9 eV for the extrapolated starting point of the Ni L3 edge in pure nickel measured in a dedicated EELS spectrometer at . the Institut fur . Nukleare Festkorperphysik in Karlsruhe [21]. The latter number was derived by extrapolating the L3 edge to the background line 2 It should be noted that the number 853.2 eV cited by Egerton [19] unfortunately resulted from a misunderstanding when copying the respective data from the original source [18,20]. while the maximum of the peak stayed approximately 1.0 eV above this value [21]. As the instrumental energy resolution of both spectrometers is similar and as the chemical shift between nickel metal and nickel oxide is known these results can be properly compared. In this respect it is also worth to analyse the data from X-ray photoelectron emission spectroscopy (XPS), which is often used for calibration of XAS spectra. Fuggle et al. [22] reported a 852.8 eV binding energy for the Ni L3 shell in pure Ni. As this number is defined relative to the Fermi level it should be most closely compared with the halfheight of the EELS L3 peak. However, there is a slight difference in the ionisation event in XPS, in which an electron is ejected to the vacuum level, and in EELS, in which the final state locates just above the Fermi level. A created core-hole rises the effective attraction between a nuclei and electrons making all electronic shells deeper in energy and increasing the apparent binding energy. This effect is expected to be slightly stronger for XPS as an electron is ejected to a higher energy than in EELS. Thus, the XPS data could be considered as an upper limit for EELS binding energies. Table 1 lists the measured numbers and extrapolated half-heights and maxima of the L3 peak in NiO taking the natural lifetime broadening of edges and the chemical shift between Ni and NiO into account. As mentioned above, XPS data represent the upper limit for the EELS number indicating that the 854.0 eV value for the peak half-height currently used in the calibration of Gatan spectrometers is overestimated and should be revised. Also, the maximum of the L3 peak instead of the half-height is more relevant for ARTICLE IN PRESS P.L. Potapov, D. Schryvers / Ultramicroscopy 99 (2004) 73–85 calibration purposes as it is less dependent on the instrumental resolution. Finally, the value of 853.0 eV compiled from the results of Batson and Fink is adopted in the present work as a reference maximum for the Ni L3 peak in NiO. 7. Examples of collected spectra Fig. 6a shows oxygen K edges in several metaloxides, obtained using the above script and placed on an absolute energy scale. Except for a difference in ELNES as mentioned in the literature, the different positions of the onsets are now apparent. A chemical shift provides information complementary to ELNES and helps to differentiate between compounds of different valency. For instance, the TiO ELNES resembles the broadened ELNES of TiO2, so that TiO can be 83 confused with TiO2 when no sufficient instrumental energy resolution is available. However, the onset of TiO is clearly different from that in TiO2 leaving no possibility for wrong identification. In other cases like nickel and copper oxides, the oxygen edges differ significantly in both ELNES and onset position. The measured absolute positions of the first major peaks in NiO, CuO and TiO2 (rutile) oxides are in excellent agreement with numbers measured by XAS [23]. In all three cases, the peak positions in EELS spectra are only 0.1 eV lower than those in XAS, which is close to the estimated experimental error. Examples of a chemical shift in L3 edges are presented in Fig. 6b–d. Similar to the O K edges the Ti L edges of TiO2-rutile and TiO2-anatase are placed at almost the same position while the onset of TiO is shifted to a lower energy. Thus, rutile and anatase can be only differentiated by a small Fig. 6. (a) O K (b) Ti L3 (c) Ni L3 and (d) Cu L3 edges in different metal-oxides measured using the value of 853.0 eV for the Ni L3 peak maximum in NiO as a reference for the absolute energy. ARTICLE IN PRESS 84 P.L. Potapov, D. Schryvers / Ultramicroscopy 99 (2004) 73–85 difference in the shape of the second peak, which was already pointed out by Brydson et al. [5,6]. Comparison of the present data with data [5] reveals that the chemical shifts are almost identical although the absolute positions of the onsets in rutile and anatase measured by us are 0.4 eV lower than those in Ref. [5]. That can be related with different reference point for calibration of the drift tube mentioned in Chapter 6. A chemical shift of the Cu L3 edges in the series Cu–Cu2O–CuO is in good agreement with those measured in a dedicated STEM instrument [7]. However, our measurements for the Ni L3 edge contradict with data by Leapman et al. [2], who reported a 0.2 eV decrease of the peak position in nickel oxide compared with nickel metal. In contrast, we have measured a 0.25 eV increase for this position. On the other hand, Leapman’s number for a chemical shift between Cu and CuO [2] agrees with our data and data by Scheu et al. [7]. The reason for this discrepancy could relate with the fact that Leapman et al. used a regular TEM microscope with the energy instability problem discussed above. To improve the precision of measurements, they mounted two samples on the same grid and moved the probe rapidly from oxide to metal. Although this procedure improves the results, it is not always applicable and can still suffer from instabilities in the microscope. Strictly speaking, the measured values of the edge onsets are dependent on the energy resolution of the instrument. The position of the first major maximum in the ELNES structure is actually more reproducible when comparing spectra obtained in different instruments. However, the peak maximum has often no clear physical meaning while the position of the onset is related with the mutual location of the inner shell and the Fermi level. To ensure the correct determination of the onset positions, the obtained spectra are deconvoluted using the profile of the zero-loss peak collected in vacuum. Comparison between the original and deconvoluted spectra is presented in Fig. 7. In fact, the profile of ELNES changes only slightly with deconvolution indicating that the energy resolution of 0.75–0.8 eV is sufficient to track the onset positions located in the 400–1000 eV region. Table 2 lists the measured Fig. 7. Ti L3 edge in TiO2-rutile before and after deconvolution with the profile of the zero-loss peak. peak maxima and onset positions for the examined oxides and metals. 8. Conclusion A new method for measuring the absolute positions of EELS ionisation edges has been developed, which can be employed in regular TEM microscopes without additional hardware installation. The precision of the absolute measurements is improved by a factor of 5 compared with standard routines, thus yielding a precision of 0.1 eV. As an example several chemical shifts in metal-oxides have been measured and compared with literature data. Acknowledgements The authors like to thank P.E. Batson (IBM Th.J. Watson Research Centre) for discussions and providing the unpublished reference data. Extensive and helpful discussions with O.L. Krivanek (Nion Co.), R. Rodemeier (Gatan GmbH) and R.F. Egerton (University of Alberta) are gratefully acknowledged. Useful advice on DM scripting was kindly provided by M.Kundmann (Gatan R&D). J. Verbeeck (University of Antwerp) is acknowledged for the help with the high-tension measurements. The possibility to perform the test of energy stability in the Technai ARTICLE IN PRESS P.L. Potapov, D. Schryvers / Ultramicroscopy 99 (2004) 73–85 85 Table 2 Measured positions of the edge onsets and first major maxima in ELNES of different oxides Edge Compound Instrumental energy resolution (eV) Relative thickness Onset (deconvoluted spectra) Maximum of the first peak in ELNES (unprocessed spectra) (eV) Ni L3 edge NiO Ni 0.75 0.75 0.59 0.69 852.4 852.1 853.0a 852.8 Cu L3 edge Cu Cu2O CuO 0.75 0.75 0.80 0.85 0.91 0.98 932.6 933.0 930.8 933.7 933.7 931.4 Ti L3 edge TiO2 rutile TiO2 anatase TiO 0.80 0.75 0.75 0.97 0.94 0.90 457.2 457.3 456.0 457.8 457.9 457.6 O K edge NiO Cu2O CuO TiO2 rutile TiO2 anatase TiO 0.75 0.75 0.80 0.80 0.75 0.75 0.61 0.87 0.47 1.19 0.93 0.77 531.0 531.6 529.3 529.8 529.9 530.9 531.6 532.5 530.0 530.6 530.7 No sharp peak The drift tube value is calibrated assuming the maximum of the Ni L3 peak in NiO to be at 853.0 eV. a Calibration. microscope of Technical University of Delft is appreciated. This work is supported by a GOA project of the University of Antwerp entitled ‘‘Characterisation of nano-structures by means of advanced electron energy spectroscopy and filtering’’. References [1] J. Tafto, O.L. Krivanek, Phys. Rev. Lett. 48 (1982) 560. [2] R.D. Leapman, L.A. Grunes, P.L. Fejes, Phys. Rev. B 26 (1982) 614. [3] M.T. Otten, B. Miner, J.H. Rask, P.R. Buseck, Ultramicroscopy 18 (1985) 285. [4] G.J. Auchterlonie, D.R. McKenzie, D.J.H. Cockayne, Ultramicroscopy 31 (1989) 217. [5] R. Brydson, H. Sauer, W. Engel, J.M. Thomas, E. Zeiter, N. Kosugi, H. Kuroba, J. Phys.: Condens. Matter 1 (1989) 797. [6] R. Brydson, H. Sauer, W. Engel, TMS Annual Meeting, New Orleans, 1991, p. 131. [7] C. Scheu, G. Dehm, M. Ruhle, . R. Brydson, Phil. Mag. A 78 (1998) 439. [8] D.A. Muller, P.E. Batson, J. Silcox, Phys. Rev. B 58 (1998) 11970. [9] A.T. Paxton, M. van Schilfgaarde, M. MacKenzie, A.J. Graven, J. Phys.: Condens. Matter 12 (2000) 729. [10] D.R. Hamman, D.A. Muller, Phys. Rev. Lett. 89 (2002) 126404. [11] P.E. Batson, Rev. Sci. Instrum. 57 (1986) 43; J. Fink, Adv. Electron. Electron Phys. 75 (1989) 121. [12] J. Fink, Adv. Electron. Electron Phys. 75 (1989) 121. [13] J. Verbeeck, Ultramicroscopy, to be submitted. [14] C.C. Ahn, O.L. Krivanek, R.P. Burgner, M.M. Disco, P.R. Swann, EELS Atlas, Gatan Inc. and Solid State Science Department in Arizona State University, 1983. [15] Gatan Imaging Filter User’s Guide, Gatan Inc., 1996. [16] K. Kimoto, Y. Matsui, J. Microscopy 208 (2002) 224. [17] O.L. Krivanek, personal communication 2002. [18] P.E. Batson, personal communication 2002. [19] R.F. Egerton, in: Electron Energy-Loss Spectroscopy in the Electron Microscope, 2nd Edition, Plenum Press, New York, 1996, p. 436. [20] R.F. Egerton, personal communication 2002. [21] J. Fink, Th. Muller-Heinzerling, B. Sheerer, W. Speier, F.U. Hillebrecht, J.C. Fuggle, J. Zaanen, G.A. Zawatzky, Phys. Rev. B 32 (1985) 4899. [22] J.C. Fuggle, N. Martensson, J. Electron Spectrosc. Relat. Phenom. 21 (1980) 275. [23] F.M.F. de Groot, M. Grioni, J.C. Fuggle, J. Ghijsen, G.A. Zawatzky, H. Petersen, Phys. Rev. B 40 (1989) 5715.
© Copyright 2026 Paperzz