Endoscopic Multiple Clip Applier (EL5ML)

LIGAMAX5™
Endoscopic Multiple Clip Applier
(EL5ML)
Optimal Device Performance
Optimal Device Performance
The LIGAMAX5™ 5mm Endoscopic Multiple Clip Applier is intended for use on tubular structures or vessels
wherever a metal ligating clip is indicated. The following information is intended to highlight key steps to provide
optimal device performance. Before using this or any other medical device, please refer to full Instructions for Use
found in the package insert.
Do not insert the clip applier through a trocar if a clip is
present in the jaws. This could result in clip malformation,
dislodged clips or damage to the instrument.
Prior to loading a clip in the jaws and firing the
instrument: ensure that the jaws are fully open by
verifying that the line of demarcation between the
jaws and the instrument shaft is past the distal end
of the trocar cannula.
Avoid firing the instrument over another clip or
instrument. Firing it in this manner may distort or
yield the instrument jaws, which could cause the
instrument to “spit” clips.
Do not excessively apply a side load to the jaws that
would cause them to partially collapse and potentially
result in a clip malformation.
Do not excessively twist or torque the instrument
jaws when positioning the instrument on a vessel and
firing. This could result in clip malformation.
Relax downward force on the proximal end of the
instrument jaws when firing. Excessive force on
the proximal end of the jaws could result in clip
malformation.
Customer support
For product use assistance, clinical guidelines, service and repair, emergency assistance, copy of a 510(k) clearance letter or
complaints, please contact our Customer Support Center at [email protected] or by calling 1-877-ETHICON (3844266). Our Support Center is staffed 24 hours a day, 7 days a week by qualified nurses to answer your product-related questions.
For complete product information, including full steps for use, indications, contraindications, warnings and precautions, please see
the Instructions for Use.
Visit ethicon.com/ligamax5 for more information.