CLINICAL RESEARCH Europace (2015) 17, 1526–1532 doi:10.1093/europace/euu410 Ablation for atrial fibrillation In vivo contact force measurements and correlation with left atrial anatomy during catheter ablation of atrial fibrillation Fabienne Schluermann 1, Tobias Krauss 2, Juergen Biermann 1, Maximilian Hartmann 1, Luca Trolese 1, Gregor Pache 2, Christoph Bode 1, and Stefan Asbach1* 1 Cardiology and Angiology I, Heart Center, Freiburg University, Freiburg, Germany; and 2Department of Radiology, University Hospital, Freiburg, Germany Received 6 October 2014; accepted after revision 24 December 2014; online publish-ahead-of-print 5 March 2015 Aims Lesion formation during catheter ablation crucially depends on catheter-tissue contact. We sought to evaluate the impact of anatomical characteristics of the left atrium (LA) and the pulmonary veins (PVs) on contact force (CF) measurements. ..................................................................................................................................................................................... Methods and An anatomical map of the LA was obtained in 25 patients prior to catheter ablation of atrial fibrillation. Contact force results (operator blinded) and local bipolar electrogram amplitudes (EGM) were measured in eight pre-defined segments around the PVs. After unblinding, points with low CF (≤5 g) were corrected to CF .5 g, and the distance between points was measured. In a pre-procedural computed tomography of the heart, LA volume as well as sizes and circumferences of the PV ostia were measured and correlated to CF measurements. Four hundred and twenty-six points in eight pre-defined LA locations were assessed. Low CF (,5 g) was found in 25.0% (43.5%) of points superior, 33.3% (66.7%) anterior, 32.1% (44.4%) inferior, and 15.5% (15.9%) posterior to the right (left) PVs. The mean distance after correction was 5.8 + 3.4 mm. Local bipolar electrogram amplitudes between low- and high-CF points did not differ (1.21 + 1.54 vs. 1.13 + 1.3 mV, P ¼ ns). The mean CF at the left PVs was significantly lower than at the right PVs (7.91 + 3.74 vs. 13.95 + 6.34 g, P , 0.001), with the lowest CF anterior to the left PVs (5.2 + 3.6 g). Contact force measurements did not correlate to LA volume, size, and circumference of the PVs. ..................................................................................................................................................................................... Conclusion Contact force during LA mapping significantly differs according to the location within the LA. These differences are independent of LA volume and anatomy of the PV ostia. ----------------------------------------------------------------------------------------------------------------------------------------------------------Keywords Atrial fibrillation † Contact force † Left atrial volume † Area of pulmonary veins † Circumference of pulmonary veins Introduction Electrical isolation of the pulmonary veins (PVI) is an accepted treatment option for patients with symptomatic atrial fibrillation (AF).1 A standard approach is to create point-by-point lesions around the pulmonary veins (PVs), usually guided by an electroanatomical mapping system. Studies have shown a high rate of conduction recurrence,2 which also is a major determinant of clinical arrhythmia recurrence. Optimal lesion formation therefore is a major goal during ablation, and this crucially depends on adequate catheter-tissue contact. While this usually is estimated by tactile feedback, interpretation of local electrograms, and impedance drop during ablation, new technologies allow for direct measurement of amount, and direction of contact force (CF) applied with the catheter. In experimental studies, it has been shown that CF correlates with lesion size and depth, and that excessive CF predisposes to thrombus formation and cardiac perforation.3,4 Clinical trials demonstrated the safety and beneficial effects regarding procedural parameters.5 – 7 It has been shown that reconnection of the PVs occurs at sites with poor CF,8 – 11 and consistent with this, evidence for a correlation of CF and arrhythmia recurrence is emerging.12 – 15 Beyond CF, anatomical conditions have also been related to outcomes in AF ablation. A major predictor has been shown to be the left atrial (LA) volume, with increasing volumes the chances of sustained rhythm control are decreasing.16,17 Data on the role of pulmonary vein anatomy and its impact on AF recurrence after ablation are * Corresponding author. Cardiology and Angiology I, Heart Center, Freiburg University, Hugstetter Str. 55, 79106 Freiburg, Germany. Tel: +49 761 270 34010; fax: +49 761 270 33882. E-mail address: [email protected] Published on behalf of the European Society of Cardiology. All rights reserved. & The Author 2015. For permissions please email: [email protected]. 1527 In vivo contact force measurements and correlation with left atrial anatomy What’s new? † Despite detailed mapping of the left atrium, low local contact force (CF) is a frequent finding during left atrial (LA) mapping, and CF cannot be estimated by local electrogram amplitude. † Low CFs cluster in the anterior region of the left pulmonary veins. † The distribution of low CFs is independent of LA volume. † The distribution of low CFs is independent of pulmonary vein anatomy. conflicting.17 – 19 It may be argued that larger LA volumes provide a substrate for AF beyond PV triggers, thus resulting in higher recurrence rates, however, larger LA volumes also could impose a greater challenge for the operator to achieve adequate cathetertissue contact and create optimal lesions. We, therefore, conducted the present study to evaluate the relationship of CF and anatomical LA and PV characteristics. Methods Twenty-five consecutive patients referred for ablation of paroxysmal (n ¼ 16) or persistent (n ¼ 9) AF (15 male, 63.3 + 15.5 years) were included at a single centre from April to December 2012. Ablation procedure After written informed consent was obtained, the ablation procedure was performed under analgosedation and spontaneous breathing in the catheter laboratory. Guiding by intracardiac ultrasound (AcuNavTM , Siemens) was used in all procedures as previously described.20 A decapolar electrophysiology catheter was placed into the coronary sinus allowing pacing manoeuvres to verify PV isolation at the end of the procedure. Double transseptal access was obtained to deploy the circular, decapolar mapping catheter (LassoNavw, Biosense Webster, Inc.) and the CF measuring ablation catheter (Thermocoolw SmarttouchTM , bidirectional navigation catheter, D/F curve, Biosense Webster, Inc.). A steerable transseptal sheath (AgilisTM , St Jude Medical) was used for the CF catheter in all cases. Saline-irrigation rate during LA mapping was 2 mL/min. Using the CARTOw3-System (Biosense Webster, Inc.), a fast anatomical map (FAM) of the LA was performed using the LassoTM Navw catheter at highest possible resolution. The AccuResp Module was used to reduce the effect of respiration on FAM acquisition. Thereafter, the CF measuring catheter was positioned in the eight pre-defined areas around both PVs in order to delineate the planned ablation circle. These points were taken with the operator blinded to CF measurements. Orientation was supported by tactile feedback, the pre-acquired FAM, and local bipolar signal amplitude. Signal amplitude was recorded for each acquired point. After unblinding, points with CF ≤5 g were re-assessed and points in vicinity as close as possible with CF .5 g were taken. Distances between points were measured with the CARTOintegrated software algorithm. During subsequent PV isolation, the operator remained unblinded for CF measurements. A single LSPV LAA LIPV Figure 1 Example of a FAM with points taken blinded to CF measurements (white) and after correction (grey), with an example of distance measurement between points. LSPV, left superior pulmonary vein; LIPV, left inferior pulmonary vein; LAA, left atrial appendage. operator performed all procedures. Figure 1 shows an example of a FAM with points, uncorrected and corrected, and the distance measurement. Computed tomography data acquisition and post-processing All examinations were performed using a 320-detector-row CT (MDCT) scanner (Aquillion One; Toshiba, Otawara, Japan). For contrast-enhanced data acquisition, 80– 90 mL of iodinated contrast agent (Imeron 350, Bracco Imaging Konstanz, Germany) was injected (flow rate 4 mL/s. followed by 50 mL chaser bolus of saline solution). Patients were scanned in a supine position during breath hold, with the scan extending from the carina to the diaphragm. The scan was started with a delay of 6 s after automatic detection of contrast in the LA with a threshold of +180 Hounsfield Units. The centre of the data acquisition window was set at 40% of the R–R interval. Volume data were reconstructed into axial images with a slice thickness of 1.0 mm and a medium smooth-tissue convolution kernel. Three-dimensional assessment of LA anatomy Multiplanar reconstructions of the MDCT dataset were used to calculate the LA volume and to assess ostial dimensions using dedicated software (Aquarius 3D Workstation, TeraRecon, San Mateo, CA, USA). LA volumetry was performed by tracing around the margins of LA from the level of the mitral annulus to the roof. The LA appendage was excluded. Multiplanar reformations were manually oriented to display the orthogonal view of the cross-section of each PV ostium. Ostial diameter and cross-sectional area of each PV were calculated automatically. Figure 2A and B show examples of CT measurements of areas and circumferences of PV ostia, and a LA volume, respectively. Statistical analysis Continuous variables are reported as mean + standard deviation, discrete variable as counts (percentages). The x2 and Fisher’s exact tests 1528 F. Schluermann et al. A B Figure 2 (A) An example of left atrial volumetry, (B) an example of a measurement of a pulmonary vein ostium. 1529 In vivo contact force measurements and correlation with left atrial anatomy 50 40 CF (g) were used to test for differences in discrete variables. Means of intra-individual CF measurements were calculated overall and for each of the eight pre-defined regions within the LA. These values were averaged over all patients. Two-tailed analysis of variance was used to test for overall differences in means, and Student’s paired t-test was used to test for differences in pairwise comparisons. For correlation, Pearson’s correlation coefficient r was calculated. All tests were two-sided with a significance level a ¼ 5%. Analyses were per& formed using GraphPad Prism, version 5.0c, GraphPad Software, Inc. 30 20 10 Results n Figure 3 Local distribution of CF within the left atrium in different locations around the right (RPV) and left (LPV) pulmonary veins. Anterior Analysis of CF We analysed CF and local bipolar signal amplitude of 426 points. The overall mean CF was 11.0 + 3.7 g. CF ≤5 g was measured in 152 (35.7%), 6– 10 g in 111 (26.1%) and .10 g in 163 (38.3%) points. Significant differences were found with respect to the LA location. The mean CF at the left PVs was lower than at the right PVs (7.9 + 3.7 vs. 14.0 + 6.3 g, P ¼ 0.0003). Lowest CF was found anterior to the left PVs, at the ridge between the left PVs and the atrial appendage (5.2 + 3.6 g). Contact force here was lower than superior (7.5 + 5.7 g, P ¼ 0.09), posterior (11.3 + 6.4 g, P ¼ 0.0001), and inferior (7.6 + 6.1 g, P ¼ 0.09) to the left PVs. Highest CF was found posterior to the right PVs (15.7 + 8.9 g). Please refer to Figure 3 for distribution of CF according to the LA region. Consistent with the distribution of CF within the LA, points with very low (,5 g) CF also clustered around the anterior parts of the left and right PVs, with significant differences in the distribution (P , 0.0001). Figure 4 shows the distribution of very low CF. Signal amplitude of the local bipolar electrogram (EGM) did not differ between points with low vs. high CF (1.21 + 1.54 vs. 1.13 + 1.3 mV, P ¼ ns). No correlation between EGM and CF was found (r 2 ¼ 0.002, P ¼ 0.391). Distance measurements The overall mean distance of corrected points to their counterpart with low CF (≤5 g) was 5.8 + 3.4 mm. While no regional differences of the distances could be noted, we found a significant correlation of low CF and distance after correction, i.e. points with lower CF also needed a larger distance to correct these points to achieve better contact (r ¼ 20.30, P , 0.001, Figure 5). R ea M The procedure was performed in all 25 patients (63.3 + 15.5 years) without complications. The echocardiographically determined LA diameter was 43.7 + 8.2 mm, left ventricular ejection fraction was 52.0 + 7.1%. Ejection fraction did not differ between patients with paroxysmal and persistent AF (52.2 + 7.7 vs. 51.7 + 6.1%, P ¼ 0.9), whereas the LA diameter was larger in patients with persistent AF (49.2 + 8.7 vs. 40.6 + 6.2 mm, P ¼ 0.008). The total procedure time (including ablation) was 3.1 + 0.6 h, fluoroscopy time 12.3 + 4.8 min. ov er PV all p R ost PV R ant PV s R up PV m e a in n f R LP PV V po LP st V LP ant V su LP p M Vi ea nf n LP V 0 Clinical characteristics 0g 1g 2g 3g 4g 5g Posterior Anterior LSPV RSPV LIPV RIPV L A A Figure 4 Distribution of points acquired with low CF (,5 g) around the pulmonary veins. LAA, left atrial appendage; LSPV, left superior pulmonary vein; LIPV, left inferior pulmonary vein; RSPV, right superior pulmonary vein; RIPV, right inferior pulmonary vein. the PV ostia. The ostial area of the right superior PV (2.9 + 0.9 cm2) was significantly larger than the ostial areas of the left superior PV (2.1 + 0.7 cm2) and the left inferior PV (2.2 + 0.9 cm2) (P , 0.05). In four patients with a left common PV (LCPV), the areas and circumferences were measured at the level of separation of the veins. Patients with an LCPV did not differ according to LA volume (123.2 + 36.9 vs. 124.5 + 32.5 cm3, P ¼ 0.9) and area or circumference of the PVs. Analysis of computed tomography scans The mean LA volume was 124.3 + 32.4 cm3. A good correlation was found with LA diameter as determined by echocardiography (r ¼ 0.69, P ¼ 0.0001). Please refer to Table 1 for dimensions of Correlation of CF and LA anatomy The mean CF was not different in patients with smaller (,124 cm3) LA volumes (10.8 + 3.3 g) vs. larger (.124 cm3) LA volumes 1530 F. Schluermann et al. (11.1 + 4.2 g, P ¼ 0.83). Correlation was found neither for CF and LA volume nor for CF and size or circumference of the right of left PV ostia (Figures 6 and 7A and B). Patients with an LCPV were neither different with respect to mean CF (12.7 + 1.0 vs. 10.7 + 3.9 g, P ¼ 0.33) nor were regional differences in CF noted (P ¼ ns). Discussion The main findings of the presented study are (i) CF is heterogeneously distributed around the PVs, with lowest values found anterior to the left PVs at the ridge to the LA appendage. (ii) Areas with very low CF have a similar distribution pattern and here, (iii) CF and distances to adjacent points with CF .5 g inversely correlate. (iv) Despite an elaborate anatomical reconstruction of the LA, still a significant number of points exhibit low CF. (v) LA volume and PV anatomy do not influence CF. After the introduction of the catheter technology that allows online measurement of CF, a number of studies have revealed that CF is a major determinant of lesion formation.3,4 It has also become quantifiable what has been clinically suspected; CF sensing technology reduces ablation and procedure times,6,13 and adequate catheter-tissue contact is harder to achieve at specific anatomical locations, especially at the ridge between the left PVs and the LA appendage.8 – 10,12 The results of the presented study in this regard confirm this observation, even in a setting with much effort for a precise anatomical map, including the FAM with highest resolution, thorough interpretation of local signal amplitude, intracardiac ultrasound, and the use of a steerable sheath. Without CF, a large number of points were taken with inadequate low CF (36% ,5 g) and would presumably not have produced transmural lesions. This further emphasizes the additional benefit of CF measurements during mapping and ablation. Data from the EFFICAS I trial suggest that a much higher CF (20 g) should be targeted to achieve a transmural lesion.9 Even if the distance measurements are afflicted with some inaccuracy (such as that the measured distance not necessarily represents the shortest possible distance to the LA surface), still the unexpected long average distance of .5 mm and the correlation of distance with CF clearly support the notion that no effective lesion can be achieved in such a position. While a relationship of CF and lesion formation is firmly established, CF is not the only predictor of AF recurrence. Hof et al. have shown that for each 10 mL increase in LA volume, the risk of AF recurrence after catheter ablation increases by 14%, and Abecasis et al. 16,17 found that patients with an LA volume of .145 mL have higher recurrence rates after catheter ablation. Both studies, similar to our study, used LA volumes derived from CT scans. Similar results could be obtained when LA volume is assessed by echocardiography or electroanatomical mapping,21,22 and volumetry has been shown to have more predictive value than LA diameter derived from echocardiography.23 However, these studies did not provide CF values from the ablation procedures. Therefore, it still remained unclear whether the reported relation of AF recurrence and LA dimensions is due to a difference in the substrate that underlies AF, or due to technical challenges imposed on the operator when ablating in a larger LA. 25 200 LA volume (mm3) LA diameter (mm) Distance (mm) 20 15 10 LA diameter LA volume 150 100 50 5 0 0 0 0 2 4 6 5 10 15 20 25 CF (g) CF (g) Figure 5 Linear regression of CF and distance to adjacent points after correction to adequate catheter contact. Figure 6 No correlation was found between LA diameter (as measured by echocardiography) or volume (as measured by CT) and CF. Table 1 Anatomical characteristics of the LA and PVs RSPV RIPV LSPV LIPV P-value ............................................................................................................................................................................... 2 Area (cm ) Circumference (mm) 2.9 + 0.9 2.5 + 0.9 2.1 + 0.7 2.2 + 0.9 0.0087 62.4 + 9.0 57.5 + 11.5 55.0 + 7.7 55.2 + 10.6 0.0299 1531 In vivo contact force measurements and correlation with left atrial anatomy Limitations The overall number of studied patients is limited. Procedures were performed by a single operator in order to reduce inter-operator variations in CF, which have previously been demonstrated.5 This limits generalization of our observations to a broader operator population. All procedures were performed during spontaneous breathing without a systematic assessment of the effect of respiration. As it has been shown that the mode of ventilation influences CF,24 timing within the breathing cycle during spontaneous breathing also may have affected CF measurements. LPV area (cm2) LPV circumference (cm) A 80 70 60 50 40 6 3 Conclusion 0 0 10 20 Mean CF LPVs (g) 30 RPV area (cm2) RPV circumference (cm) B Contact force during LA mapping significantly differs according to the location within the LA. The lower the CF, the larger is the distance to the adjacent point with adequate CF. Local differences within the LA are independent of LA volume and anatomy of the PV ostia. 80 References 70 60 50 40 6 3 0 0 10 20 Mean CF RPVs (g) 30 Figure 7 (A and B) No correlation was found for mean CF around the left (LPV, A) and right (RPV, B) pulmonary veins and area or circumference of the pulmonary veins. The results of our study are in favour of the former assumption, since CF was equally distributed without relationship to LA volume, area or circumference of the PVs. With this observation, it seems plausible to accept that CF is linked more to anatomical features within the LA (e.g. low CF at the ridge anterior to the left PVs), and possibly the angle of the transseptal access site to the specific mapping/ablation site, more than to the overall volume of the LA. It has recently been shown that the presence of an LCPV is a predictor for freedom of AF after AF ablation.19 The authors speculated that the presence of an LCPV facilitates greater CF especially in the anterior parts, and that the LCPV may more frequently harbour triggers that initiate AF. While the authors recognized the limitation in their study that they could not provide CF data, we did not find differences in CF in patients with LCPV when compared with those that had two discrete left PVs. This would favour the latter speculation, that an LCPV more often harbours triggers for AF, though we are unaware of further supportive evidence. 1. Camm AJ, Lip GY, De Catarina R, Savelieva I, Atar D, Hohnloser SH et al. 2012 focused update of the ESC Guidelines for the management of atrial fibrillation. Europace 2012;14:1385 –413. 2. Cappato R, Negroni S, Pecora D, Bentivegna S, Lupo PP, Carolei A et al. Prospective assessment of late conduction recurrence across radiofrequency lesions producing electrical disconnection at the pulmonary vein ostium in patients with atrial fibrillation. Circulation 2003;108:1599 –604. 3. Yokoyama K, Nakagawa H, Shah D, Lambert H, Leo G, Aeby N et al. Novel contact force sensor incorporated in irrigated radiofrequency ablation catheter predicts lesion size and incidence of steam pop an thrombus. Circ Arrhythm Electrophysiol 2008;1:354 –62. 4. Thiagalingam A, D’Avila A, Forley L, Guerero JL, Lambert H, Leo G et al. Importance of catheter contact force during irrigated radiofrequency ablation: evaluation in a porcine ex vivo model using a force-sensing catheter. J Cardiovasc Electrophysiol 2010;21:806 –11. 5. Kuck KH, Reddy V, Schmidt B, Natale A, Neuzil P, Saoudi N et al. A novel radiofrequency ablation catheter using contact force sensing: Toccata study. Heart Rhythm 2012;9:18 –23. 6. Martinek M, Lemes C, Sigmund E, Derndorfer M, Aichinger J, Winter S et al. Clinical impact of an open-irrigated radiofrequency catheter with direct force measurement on atrial fibrillation ablation. Pacing Clin Electrophysiol 2012;35:1312 –8. 7. Stabile G, Solimene F, Calò L, Anselmino M, Castro A, Pratola C et al. Catheter-tissue contact force for pulmonary veins isolation: a pilot multicentre study on effect on procedure and fluoroscopy time. Europace 2014;16:335 –40. 8. Haldar S, Jarman JWE, Panikker S, Jones DG, Salukhe T, Gupta D et al. Contact force sensing technology identifies sites of inadequate contact and reduces acute pulmonary vein reconnection: a prospective case control study. Int J Cardiol 2013;168: 1160 –6. 9. Neuzil P, Reddy V, Kautzner J, Petru J, Wichterle D, Shah D et al. Electrical reconnection after pulmonary vein isolation is contingent on contact force during initial treatment: results from the EFFICAS I study. Circ Arrhythm Electrophysiol 2013;6: 327 – 33. 10. Park C, Lehrmann H, Keyl C, Weber R, Schiebeling J, Allgeier J et al. Mechanisms of pulmonary vein reconnection after radiofrequency Ablation of atrial fibrillation: the deterministic role of contact force and interlesion distance. J Cardiovasc Electrophysiol 2014;25:701 –8. 11. le Polain de Waroux JB, Weerasooriya R, Anvardeen K, Barbraud C, Marchandise S, De Meester C et al. Low contact force and force-time integral predict early recovery and dormant conduction revealed by adenosine after pulmonary vein isolation. Europace 2015;17:877 –83. 12. Reddy V, Shah D, Kautzner J, Schmidt B, Saoudi N, Herrera C et al. The relationship between contact force and clinical outcome during radiofrequency catheter ablation of atrial fibrillation in the TOCCATA study. Heart Rhythm 2012;9: 789 – 95. 13. Kimura M, Sasaki S, Owada S, Horiuchi D, Sasaki K, Itoh T et al. Comparison of lesion formation between contact force-guided and non-guided circumferential pulmonary vein isolation: a prospective, randomized study. Heart Rhythm 2014;11:984 – 91. 14. Marijon E, Fazaa S, Narayanan K, Guy-Moyat B, Bouzeman A, Providencia R et al. Real-time contact force sensing for pulmonary vein isolation in the setting of 1532 15. 16. 17. 18. 19. paroxysmal atrial fibrillation: procedural and 1-year results. J Cardiovasc Electrophysiol 2013;25:130 –7. Providência R, Marijon E, Combes S, Bouzeman A, Jourda F, Khoueiry Z et al. Higher contact-force values associated with better mid-term outcome of paroxysmal atrial fibrillation ablation using the SmartTouch(tm) catheter. Europace 2015;17:56 –63. Abecasis J, Dourado R, Ferreira A, Saraiva C, Cavaco D, Santos KR et al. Left atrial volume calculated by multi-detector computed tomography may predict successful pulmonary vein isolation in catheter ablation of atrial fibrillation. Europace 2009;11: 1289 –94. Hof I, Chilukuri K, Arbab-Zadeh A, Scherr D, Dalal D, Nazarian S et al. Does left atrial volume and pulmonary vein anatomy predict the outcome of catheter ablation of atrial fibrillation? J Cardiovasc Electrophysiol 2009;20:1005 – 10. Den Uijl DW, Tops LF, Delgado V, Schuijf JD, Kroft LJM, de Roos A et al. Effect of pulmonary vein anatomy and left atrial dimensions on outcome of circumferential radiofrequency catheter ablation for atrial fibrillation. Am J Cardiol 2011;107:243 –9. McLellan AJA, Ling LH, Ruggiero D, Wong MCG, Walters TE, Nisbet A et al. Pulmonary vein isolation: The impact of pulmonary venous anatomy on long-term F. Schluermann et al. 20. 21. 22. 23. 24. outcome of catheter ablation for paroxysmal atrial fibrillation. Heart Rhythm 2014;11:549 – 56. Biermann J, Bode C, Asbach S. Intracardiac Echocardiography during catheter-based ablation of atrial fibrillation. Cardiol Res Pract 2012;2012, doi:10.1155/2012/921746. Shin SH, Park MY, Oh WJ, Hong SJ, Pak HN, Song WH et al. Left atrial volume is a predictor of atrial fibrillation recurrence after catheter ablation. J Am Soc Echocardiogr 2008;21:697 –702. Maciel W, Andrea E, Araujo N, Carvalho H, Belo LG, Siqueira L et al. Prognotic criteria of success and recurrence in circumferential ablation for the treatment of atrial fibrillation. Arq Bras Cardiol 2007;88:134 –43. Sohns C, Sohns JM, Vollmann D, Luethje L, Bergau L, Dorenkamp M et al. Left atrial volumetry from routine diagnostic work up prior to pulmonary vein ablation is a good predictor of freedom from atrial fibrillation. Eur Heart J Cardiovasc Imaging 2013;14:684 –91. Kumar S, Morton JB, Halloran K, Spence SJ, Lee G, Wong MCG et al. Effect of respiration on catheter-tissue contact force during ablation of atrial arrhythmias. Heart Rhythm 2012;9:1041 –47. EP CASE EXPRESS doi:10.1093/europace/euv199 Online publish-ahead-of-print 2 September 2015 ............................................................................................................................................................................. Clipping to sinus rhythm: cardioversion of atrial fibrillation by a thoracoscopic left atrial appendage occlusion Vedran Velagić*, Gian-Battista Chierchia, and Mark La Meir Centre for Cardiovascular Diseases, Universitair Ziekenhuis Brussel, Vrije Universiteit Brussel, 101 Laarbeeklaan, 1090 Brussels, Belgium * Corresponding author. Tel: +385 91 7929284; fax: +322 477 6851, E-mail address: [email protected] A 52-year-old male patient underwent a hybrid procedure for the treatment of a long-standing persistent atrial fibrillation (second ablation). A thoracoscopic video-assisted approach was used for the epicardial, radiofrequent pulmonary vein isolation, and the ‘box lesion’ set creation. Epicardial electrograms at the different sites in the left atrium were recorded, before and after the completion of ablation. The cycle length of the left atrium appendage activation decreased after the ablation, and electrograms became fractionated (Figure A). Afterwards, the epicardial appendage clipping was performed during which patient converted to sinus rhythm. There was no more electrical activity in the appendage, entry and exit blocks were confirmed (Figure B). Endocardially, the pulmonary vein isolation was confirmed and no further ablation was performed. We hypothesize that, by isolating pulmonary veins and posterior wall of the left atrium, the appendage became the main atrial fibrillation driver, marked by a new fractionation of the local electrograms. Consequently, the effective electric isolation of the appendage stopped the arrhythmia and patient returned to sinus rhythm. It is important to note that the epicardial appendage clipping might provide not only the functional occlusion but also the electrical isolation, unlike the percutaneous devices that leave the arrhythmogenic potential of the appendage untouched. The full-length version of this report can be viewed at: http://www.escardio.org/Guidelines-&-Education/E%E2%80%93learning/Clinicalcases/Electrophysiology/EP-Case-Reports. Published on behalf of the European Society of Cardiology. All rights reserved. & The Author 2015. For permissions please email: [email protected].
© Copyright 2026 Paperzz