2048B - Princeton Instruments

X-RAY GROUP
PI-MTE: 2048B
2048 x 2048 imaging array | 13.5 x 13.5 µm pixels
The PI-MTE: 2048B from Princeton Instruments is an innovative, high-sensitivity imaging system designed to deliver scientific
performance inside high-vacuum chambers over long operation times. The PI-MTE: 2048B incorporates a back-illuminated
CCD without AR coating for ultra-low-energy X-ray detection. With 13.5 x 13.5 µm pixels and 100% fill factor, this
system offers a large imaging area with very high spatial resolution and dynamic range. The thermoelectrically cooled
design features PCBs, thermally linked to the circulating coolant to provide reliable operation inside vacuum chambers. The
system’s compact size and flexible tubing facilitate the positioning of the detector in limited space or on a movable arm.
FEATURES
BENEFITS
Back-illuminated CCD (without AR coating)*
Best response for soft x-ray
2048 x 2048 imaging array
13.5 x 13.5 mm pixels
Largest available area without AR coating for imaging
Unique thermal design
Efficient heat removal in high-vacuum environment
Flexible binning and readout
Increases frame rate and signal-to-noise ratio (SNR)
Compact design
For installation in limited space or on a movable arm in vacuum
Thermoelectric cooling
Minimizes dark current
Up to 2 MHz digitization
Delivers high frame rates without compromising system performance
Dual-digitizer option
Multiple-speed digitization allows complete freedom to select between “slow
operation” for low noise and highest SNR or “fast operation” for rapid image
acquisition
Seamless, plug-and-play connection to PC notebooks and desktops
Easy OEM integration
Flexible software packages for data acquisition, display and analysis;
LightField offers intuitive, cutting edge user interface, IntelliCal® and more.
Compatible with Windows 8/7/XP, and Linux;
Universal programming interfaces for easy custom programming.
USB 2.0 interface configuration
Optional: LightField® (for Windows 8/7, 64-bit)
Or WinView/Spec (for Windows 8/7/XP, 32-bit)
PICAM (64-bit) / PVCAM (32-bit)
software development kits (SDKs)
*Contact your local sales specialist for other CCD options.
Applications:
Soft X-ray Imaging , X-ray Microscopy,
EUV Lithography, X-ray Plasma Imaging
Page 1 of 4
PI-MTE: 2048B Rev. N2.1
X-RAY GROUP
SPECIFICATIONS
CCD image sensor
Scientific-grade, back-illuminated, MPP device without AR coating
CCD format
2048 x 2048 imaging pixels
13.5 x 13.5 mm pixels
100% fill factor
27.6 x 27.6 mm imaging area (optically centered)
MINIMUM
TYPICAL
Linear full well
80 ke-
100 ke-
Output amplifier
130 ke-
200 ke-
System read noise
@ 50 kHz digitization
@ 100 kHz digitization
@ 1 MHz digitization
@ 2 MHz digitization
Dark current @ -50° C operation
Deepest cooling temperature
TE (chilled liquid)
Vacuum Compatibility
-45° C
Nonlinearity @ 100 kHz
2%
Readout bits / speed
16 bits @ 50 kHz to 1 MHz
Parallel shift rate
96.2 µsec/row
NOTE: All specifications subject to change
MAXIMUM
3 e- rms
4 e- rms
10 e- rms
12 e-rms
0.02 e-/p/s
4 e- rms
7 e- rms
15 e- rms
18 e-rms
0.1 e-/p/s
-50° C
-55° C
10-4 Torr
10-6 Torr*
* Requires dynamic vacuum
Page 2 of 4
PI-MTE: 2048B Rev. N2.1
X-RAY GROUP
QE (%)
Quantum Efficiency Curve
100
100
90
90
80
80
70
70
60
60
50
50
40
40
30
30
20
20
10
10
0
1
10
100
1000
10000
0
100000
Photon Energy (eV)
BI basic Process, NoAR
BI, enhanced process, NoAR
BI DD NoAR
BI DD AR Coated
0.50 mm Be/FI_DD
0.50 mm Be/BI_DD
Page 3 of 4
PI-MTE: 2048B Rev. N2.1
X-RAY GROUP
2.00 50.80
4.90 124.33
SEE NOTE 1.
4.57 115.95
3.15 80.01
2.35 59.69
1.75 44.45
MTE Drawing
1.43 36.20
1.08 27.43
SEE NOTE 1.
1.04 26.47
2.00 50.80
WATER OUT
WATER IN
(4X) 4-40 UNC - 2B
.25 6.35
0.42 10.62
ARRAY FOCAL PLANE
(2X) THESE SURFACES ONLY
#6-32 UNC - 2B 0.30[07.6]
STANDARD 1/4 INCH COPPER TUBING
NOTES:
1.25 31.75
STANDARD FEMALE VCR FITTING
0.55 13.84 MIN.
SEE NOTE 1.
0.76 19.30 MIN.
SEE NOTE 1.
1. WATER TUBE LENGTH AND POSITION MAY BE CUSTOMIZED,
SUBMIT DIMENSIONAL REQUIREMENTS AT TIME OF ORDER.
Vacuum Interface
2.75” Conflat® flange with feedthrough tubes
Swagelok fittings
outside vacuum
VCR fittings
inside vacuum
4.5” Conflat® flange with vacumm-compatible
electrical feedthrough connectors
NOTE: Includes high vacuum compatible 36” Kapton® cable and 36” flexible stainless steel tubing
Readout Rates
Binning
1x1
2x2
4x4
@ 2 MHz
2.39 sec
1.185 sec
0.61 sec
@ 1 MHz
4.59 sec
1.765 sec
0.77 sec
@ 100 kHz
40.25 sec
11.254 sec
3.456 sec
@ 50 kHz
78.42 sec
20.80 sec
5.845 sec
www.princetoninstruments.com [email protected]
TOLL-FREE +1.877.474.2286 | PHONE +1 609.587.9797
Visit the Princeton Instruments website to find a representative in your area.
Page 4 of 4