Microfluid Nanofluid (2015) 18:357–366 DOI 10.1007/s10404-014-1454-3 RESEARCH PAPER Serpentine and leading-edge capillary pumps for microfluidic capillary systems Roozbeh Safavieh • Ali Tamayol • David Juncker Received: 4 March 2014 / Accepted: 26 June 2014 / Published online: 17 July 2014 Ó Springer-Verlag Berlin Heidelberg 2014 Abstract Microfluidic capillary systems operate using capillary forces only and require capillary pumps (CPs) to transport, regulate, and meter the flow of small quantity of liquids. Common CP architectures include arrays of posts or tree-like branched conduits that offer many parallel flow paths. However, both designs are susceptible to bubble entrapment and consequently variation in the metered volume. Here, we present two novel CP architectures that deterministically guide the filling front along rows while preventing the entrapment of bubbles using variable gap sizes between posts. The first CP (serpentine pump) guides the filling front following a serpentine path, and the second one (leading-edge pump) directs the liquid along one edge of the CP from where liquid fills the pump row-by-row. We Electronic supplementary material The online version of this article (doi:10.1007/s10404-014-1454-3) contains supplementary material, which is available to authorized users. R. Safavieh A. Tamayol D. Juncker McGill University and Genome Quebec Innovation Centre, McGill University, 740 Dr. Penfield Avenue, Montreal, QC H3A 0G1, Canada R. Safavieh A. Tamayol D. Juncker Biomedical Engineering Department, McGill University, 3775 University Avenue, Montreal, QC H3A 2B4, Canada A. Tamayol Harvard-MIT Division of Health Sciences and Technology, Cambridge, MA, USA A. Tamayol Harvard Medical School, Boston, MA, USA D. Juncker (&) Department of Neurology and Neurosurgery, McGill University, 3801 University Street, Montreal, QC H3A 2B4, Canada e-mail: [email protected] varied the angle of the rows with respect to the pump perimeter and observed filling with minimal variation in pumping pressure and volumetric flow rate, confirming the robustness of this design. In the case of serpentine CP, the flow resistance for filling the first row is high and then decreases as subsequent rows are filled and many parallel flow paths are formed. In the leading-edge pump, parallel flow paths are formed almost immediately, and hence, no spike in flow resistance is observed; however, there is a higher tendency of the liquid for skipping a row, requiring more stringent fabrication tolerances. The flow rate within a single CP was adjusted by tuning the gap size between the microposts within the CP so as to create a gradient in pressure and flow rate. In summary, CP with deterministic guidance of the filling front enables precise and robust control of flow rate and volume. Keywords Capillary flow Passive pumping Microfluidics Pore network modeling Point-of-care 1 Introduction Capillary systems are widely used for manipulating liquids in heat pipes (Groll et al. 1998), regulating fluid flow in low-gravity environment in space (Weislogel 2003), patterning biomolecules on surfaces (Delamarche et al. 1997, 2005; Lovchik et al. 2010), immunoassays (Cesaro-Tadic et al. 2004), and diagnostic applications (Juncker et al. 2002; Gervais and Delamarche 2009; Gervais et al. 2011a; Safavieh and Juncker 2013). Capillary systems are selfpowered and self-regulated by capillary effects and are thus self-contained as there is no need for external pumps and tubes. Capillary pumps (CPs) are essential components of many capillary systems and provide a negative (suction) 123 358 pressure for manipulating liquids. CPs are notably used in assays to flow accurate volumes of sample (i.e., 1–10 lL) over the assay surface to reach a high sensitivity (Gervais et al. 2011b; Safavieh and Juncker 2013). The volume of the CP serves as a metering device—the pump’s capacity sets the volume of the liquid drawn in—and as a waste reservoir for reagents (Juncker et al. 2002; Safavieh and Juncker 2013). Indeed, since capillary systems operate by applying a negative pressure, CPs are placed at the end of the flow path and thus hold the liquid after it passes through the reaction zones. For immunoassays, it is often desired to have CPs that can provide a desired flow rate for a considerable amount of time (i.e., 1–30 min) during the incubation step. CPs that generate a particular capillary pressure and that do not significantly increase the overall flow resistance as they are being filled are thus needed. Successful designs include CPs with parallel microchannels arranged in an arborescent pattern (Juncker et al. 2002) or arrays of posts (Gervais et al. 2011b), or meshes (Mirzaei et al. 2010). Postarrays and meshes are more robust than branching channels, because they are less prone to clogging or flow stopping as the liquid can progress along many parallel, connected paths. In practice, pumps are often covered to avoid evaporation and contamination of the sample. As a consequence, bubbles can easily get trapped, if the filling front of the liquid progresses faster along one side of the CP and reaches the outlet prior to filling the CP entirely. Trapped bubbles reduce the liquid drawn into the CP and thus result in unreliable metering and inaccurate incubation time during immunoassays. An occurrence of bubble trapping is illustrated in the Movie S1 in the electronic supplementary information (ESI). The issue of bubble trapping becomes even more pronounced when multiple pumps are connected in series (Zimmermann et al. 2007). A strategy to avoid bubble trapping in microfluidic systems is to control the progression of the filling front. This has been widely implemented in hydrophobic systems where external pumps, pressure, vacuum, or centrifugal forces are used to drive liquids, and where the progression is controlled and preprogramed by adding hydrophobic obstacles to the flow path (McNeely et al. 1999; Vulto et al. 2011). The control of the liquid front is more difficult in self-filling, hydrophilic capillary systems because of the constraints of maintaining self-filling. Zimmerman et al. (2007, 2008) designed CPs with posts to enable filling and imposing specific angles on the filling front-indented sidewalls that slow down the progression along the edges. However, their CP did not provide deterministic flow control, and the filling front was shown to jump rows, and in angled pump, the front could progress more rapidly on one side than the other. Hence, trapping of bubbles could not be excluded, and as a result, they introduced outlets 123 Microfluid Nanofluid (2015) 18:357–366 made by a series of arch-shaped, hierarchical ‘‘bridges’’ with increasing size that fill in predetermined order. The outlets further help minimize bubble trapping, but at the expense of a weaker capillary pressure that reduces further as the cross section of the bridging channels increases, thus making the system susceptible to contaminations and stopping of the flow. Moreover, this structure was supplemented with another structure in the form of a small serpentine side channel, which acted as a delay valve (Zimmermann et al. 2008). This channel however comes at a cost of reduced metering accuracy, and ultimately, each new external element added to the CP affects the functionality of the circuit and may itself become a source of failure. Here, we introduce CPs that deterministically guide the progression of the filling front via the geometry of the microstructures of the CPs. These CPs comprise arrays of microposts that are spaced differently along the X (parallel to the bulk flow) and Y (perpendicular to the bulk flow) axes so as to form rows along the Y axis. The liquid entering the pump flows into the first row and fills the narrow gaps between the posts on the downstream side of the row, which acts as a temporary stop valve; the valving effect is robust because of the negative pressure of the filling front in the row (Zimmermann et al. 2007). The liquid front is thus guided and fills the row entirely, and only at the edge of the CP does it progress into the next row. Two CP designs and strategies were developed that allow a precise control over the advancement of the filling front: first, a serpentine CP, where the front is guided in an alternate direction in each row in a serpentine pattern; second, a leading-edge CP, where the liquid initially is guided to fill one of the edges of the CP and then fills each row starting from this ‘‘leading-edge’’. The filling front advances perpendicularly to the bulk flow along each row, and as a result, both follow different paths. We analyze and compare the two CPs and demonstrate the robustness of both designs by changing the angle of the rows. Next, we develop an electrical network analog for the serpentine pump and use it to calculate the flow resistance of the CP as the liquid fills it. Finally, a serpentine CP with continuously varying gap between the rows was designed, and the theoretical and experimental filling speeds were compared. 2 Materials and methods 2.1 Chemicals and materials Polydimethylsiloxane (PDMS) was purchased from Dow corning (SylgardÒ 184, MI, USA). Curing agent and polymer base were manually mixed at a ratio of 1:10 and Microfluid Nanofluid (2015) 18:357–366 cured for 8 h in an oven (Lindberg Blue M, Fisher Scientific) at 60 °C. A 1 mg/mL solution of Fluorescein sodium salt (C20H10Na2O5) (fluorescein dye, Sigma-Aldrich, USA) in water (Milli-Q purified water, Millipore, USA) was used for characterizing the pumps; 600 Si wafers with the thickness of 650 ± 10 lm were purchased from Silicon Quest International (San Jose, USA). Shipley-1813, a positive photoresist, was purchased from Microchem Inc. (Massachusetts, USA). 2.2 CPs fabrication We designed CPs using a layout editor software (CleWin 5, WieWeb software, Netherlands) and obtained 700 9 700 chrome mask with 65,000 dots per inch resolution from Thin Metal Parts Inc. (Colorado Springs, USA). The CPs were fabricated by first coating a Si wafer with a 2.8 lm thick photoresist layer followed by photolithography and deep reactive ion etching (DRIE, Tegal SDE 110, USA). Next, we diced the wafers into individual chips using a diamond scriber (Ted Pella Inc., USA). The Si chips were rendered hydrophilic by flaming them for *45 s until they became glowing red using a microtorch (Butane fuel microtorch MT-51, Master Appliance, USA). A flat 2-mm-thick PDMS was cut to size, and vents were punched with a biopsy puncher (1 mm diameter, Ted Pella Inc., USA). The PDMS was sealed onto the Si chip to serve as a cover. 359 3 Results and discussion 3.1 Design of the capillary pumps The CPs comprise an array of 40 9 80 lm2 posts with rounded edges in a staggered arrangement. The gap between adjacent posts in two separate rows, W1, is larger than the distance between two columns, W2, Fig. 1a. By making the surface hydrophilic, the liquid spontaneously fills the CP owing to the capillary pressure generated inside the CP, and first enters the large space forming the first row (t0), but then immediately fills the narrow gaps between adjacent posts (t1). However, the liquid does not flow into the next row, because the geometry of the posts acts as a temporary capillary stop valve (Zimmermann et al. 2008). The liquid front thus progresses along the next post and 2.3 Characterization of the capillary pumps CPs were imaged using optical (LV150 industrial microscope, Nikon, Japan) and scanning electron microscopy (S3000N variable pressure SEM, Hitachi, Japan). The progression of the filling front in the CPs was visualized by capturing the flow of fluorescein in deionized water. For each experiment, 2.5 lL of solution was delivered to the loading port and using a stereomicroscope (Leica MZ10f, Leica Microsystems, Germany). Fluorescence videos were recorded at a rate of 30 frames per second (infinity 3 CCD camera, USA). Videos were converted into images using VirtualDubMod Surround 1.6.0 (Lee 2010) and the velocity of the filling front was established by measuring the change in position between individual frames. At least three independent experiments were performed, and standard deviations were calculated. The flow resistance of the CP as a function of the position of the filling front was calculated by modeling the pump as a nodal network and using an in house code programed in Matlab (v7.13, MathWorks Inc., USA). The advancing and static contact angles between DI water and a flat Si surface, which was rendered hydrophilic, were measured using a contact angle analysis equipment (VCA optima, AST products Inc., USA). Fig. 1 Schematics of serpentine and leading-edge CPs with arrays of posts. a Schematic illustrating the progression of capillary flow between the posts from time t0 to t2. The liquid fills the rows by capillary pressure, as well as the gap between the posts, but the liquid cannot jump into the next row as each pair of posts forms a capillary stop valve, and thus, the liquid progresses along the row until it reaches the edge of the pump and proceeds to fill the next row. b Serpentine CPs comprise guiding structures at the end of each alternating row to guide the liquid filling front in a serpentine pattern. c Leading-edge CPs draw the liquid into the narrow edge with a capillary pressure (DP2), and then sequentially into each of the rows, which have the highest capillary pressure (DP1), starting with the first one. Finally, the wide gap (DP3) is filled up to the exit. The path of the filling front is indicated by the red arrows and the bulk flow by the white hashed arrows; note the difference in the paths. d A capillary system with a loading port, a CRV, a flow resistor, and a vent is used to test the CPs (color figure online) 123 360 further along the row (t2), until it fills the next narrow gap, and so on. Oval shaped posts were chosen to both perform as a temporary stop valve, preventing the liquid to leak from one row to the other, and avoid trapping bubbles while filling the posts one after the other in the same row. It may, however, also be possible to use rectangular posts, but the flow resistance for the same gap would be higher and the fabrication becomes more difficult. The experimental data confirmed the effectiveness of the oval shaped posts in temporarily stopping of liquid. The filling front thus fills the entire row, and only after reaching the edge of the CP, advances to the next row. The CP is thus filled row-by-row. Two distinct CPs were designed that guide the filling front in two different ways. The first is called a serpentine pump, because it guides the liquid front along a serpentine path, Fig. 1b. As it reaches the edge of the CP, the liquid front is guided to the next row while guiding structures are included to prevent premature leakage into the subsequent row. The second CP is called leading-edge pump, because the liquid front is led along one edge of the CP and starts sequentially filling each row from this edge, Fig. 1c. The liquid front is guided by reducing the gap between the ‘‘leading’’ edge and the posts compared with the other edges—hence creating a higher capillary pressure at the leading edge—and thus forcing the liquid to fill each row from that edge. In both CPs, the angle of the rows is varied and tilted between 08 to 458 relative to the edge of the CP. Manipulation of the filling front along the required directions allows filling of CPs with various geometries and thus could save the footprint of the microfluidic circuits. Also, by varying the filling front direction, the number of posts per unit area of the CP and its capacity can be adjusted. To test the two CP designs, simple capillary systems consisting of a loading port, a capillary retention valve (CRV), an external flow resistor, and the respective CP were built, Fig. 1d (Juncker et al. 2002; Hitzbleck et al. 2011). PDMS is widely used in the fabrication of microfluidic systems but it is hydrophobic natively, and although it can be hydrophilized by plasma treatment, it reverts to a hydrophobic state within hours. To avoid complications owing the changes in surface wettability of PDMS, the chips were made out of Si. Si is capped by a native SiO2 layer, which can be hydrophilized readily, for example, by plasma treatment, and remains hydrophilic over extended periods of time. This enables us to preserve the chips over for few weeks and test them, when they are needed. The system was then sealed with a hydrophobic PDMS cover that contained a loading port to fill samples, and one opening for venting. The capillary pressure, DP, in the CPs was estimated using the pressure equation for rectangular cross sections (Delamarche et al. 1998): 123 Microfluid Nanofluid (2015) 18:357–366 DP ¼ c cos þ cos cos þ cos b t l r þ w h ð1Þ where c is the surface tension of the liquid, i = b, t, l, r are the contact angles of the liquid–air meniscus on the bottom, top, left, and right walls, respectively, and w and h are the width and height of the rows formed between the pillars. The actual pressure is expected to be lower owing to the fact that the channel is not continuous, but lined by posts with narrow gaps. The narrow gaps, W2, between the pillars were constant and equal to 15 lm for all CPs except the one with the variable capillary pressure, which was equal to 25 lm, Fig. S1A in ESI. The width of the rows, W1, was varied from 34 to 130 lm, depending on the CP. The depth of the pumps was fixed during the DRIE fabrication step and was typically *100 lm. The spiral shaped external resistor had a width of 30 lm and an overall length of 3,400 lm, Fig. S1C in ESI. The footprints of the CPs are from 4.2 9 4 to 4.2 9 4.5 mm2 with a porosity ranging from 47 to 73 % and thus can accumulate between 1.02 and 1.65 lL. To have CPs with higher volumes, one could either increase the height of the structures or instead expand the footprint Fig. 2 Micrographs of CPs and the capillary system. a A serpentine CP with a 15° angle with ridges at the edge. b A leading-edge CP with 15° showing the leading-edge gap and the trailing-edge gap. Note that the posts are cut at the edges here. c Overview of the capillary system used to test the CPs. The PDMS cover with a vent was aligned to the chip manually using an stereomicroscope Microfluid Nanofluid (2015) 18:357–366 361 of the CPs. The deeper the structure, the higher the aspect ratio of the posts, which is difficult to extend beyond 10 in Si. Increasing the area of the CPs leads to an increase in the overall footprint of each device, and an increased cost per unit. For the leading-edge CPs, Fig. 1c, we adjusted the gaps between rows that generate DP1, the gap of the leading edge (left and top) that results in DP2, and the gap of the trailing edge (bottom and right) that creates DP3, so that the absolute value of the capillary pressures was jDP1 j [ jDP2 j [ jDP3 j ð2Þ and the PDMS cover. The gap of W2 acts as a temporary stop valve and prevents the filling front from jumping from one row to the next, until the entire row is filled. The robustness of these temporary stop valves can be increased by adjusting 3 different parameters: (i) by decreasing the ratio of W2/W1; (ii) by making the surface of the chip less hydrophilic; and (iii) by increasing the capillary pressure within a row to create a strong negative pressure. 3.2 CPs with rows at different angles Figure 2 shows the fabricated serpentine and leadingedge CPs as well as a capillary system used in the experiment, Fig. 2a–c. The functionality of the CP depends on the balance of capillary pressures, and the wettability of the Si structures The deterministic control over the filling path allows designing serpentine and leading-edge CPs with angled rows, Fig. 3. We studied the effect of the angle on the flow rate of the serpentine CPs. The three serpentine CPs with row angles relative to the edges of a = 0°, 15°, and 45° are Fig. 3 Time lapse imaging showing the progression of the filling front. Three serpentine CPs with a 0°, b15°, c 45° angle are shown and d a leading-edge CP with a 15° angle. Dilute solution of fluorescein in distilled water was used to perform all the tests. The red arrows show the direction of the liquid filling front, whereas the white arrows demonstrate the direction of the flow. Real-time movies of the filling as shown in the ESI. The scale bar is 1 mm (color figure online) 123 362 Microfluid Nanofluid (2015) 18:357–366 shown in Fig. 3a–c. In general, no significant difference was observed between the flow rates through the pumps. We also fabricated a leading-edge CP with an angle of a = 15° shown in Fig. 3d. Real-time movies S2-6 in (ESI) illustrate the filling of different CPs. Angled rows afford greater flexibility in directing the filling front, and a 45° angle further minimizes the risk of bubble trapping at the outlet as the final rows are very short and row jumping less likely to occur. 3.3 Flow model of the CPs The flow rates of the system used here are few tens of nanoliters per second only; thus, the flow is laminar and corresponds to what is known as creeping flow. Under these conditions, the relationship between capillary pressure, DP, the flow resistance, R, and volumetric flow rate, Q, of the system becomes (Juncker et al. 2002; White 2005; Safavieh et al. 2011): R¼ DP Q ð3Þ This is similar to the electrical relationship between voltage, resistance, and current; however, in this case, the flow resistance of the CP continuously changes as the filling front progresses, while flow front and bulk flow follow different paths. Both of these factors need to be considered when calculating the flow resistance and flow rate. The flow resistance of rectangular microchannels is defined as (White 2005; Akbari et al. 2009): 3 pw1 wh 192h Rc ¼ Pa s m3 1 5 tanh pw 2h 12lL ð4Þ where L, w, and h are the length, width, and height of the microchannel, respectively, and l is the viscosity of the liquid. The CP can be considered as an array of unit cells that allow for flow along the row and across the row through the small gaps, Fig. 4a. To simplify the calculation, the gap between the posts can be treated as straight conduits. The resistance of each unit cell in a row can be lumped into a single resistance Rh, and the resistance along the narrow gap between posts as Rv, Fig. 4b. Tamayol et al. (2013) have shown that the suitable length scale for determining the pressure drop between adjacent cylinders is the minimum gap size. Thus, to simplify the analysis, we neglected the effects of the rounded edges in the post and assumed that the posts are completely rectangular shape with 40 9 80 lm2 and that the effective gaps to calculate Rh and Rv are W1 = 35 lm and W2 = 15 lm, respectively. The values of Rh and Rv can then be obtained using Eq. 4. More details are provided in ESI. 123 The regular structure of the CP lends itself to a pore network modeling approach, common in the analysis of flow through porous media, which is an equivalent to network approach as used in electronics (Ho et al. 1975; Chapuis and Prat 2007; Prat 2011). In this approach, the pump can be modeled as a network of nodes interconnected by Rh and Rv resistors, Fig. S2 and S3. For a typical serpentine CP, Rh = 3.99 9 1011 Pa s m-3 and Rv = 9.90 9 1011 Pa s m-3, ESI. The staggered architecture of the CP with a = 0°, Fig. 4b, was modeled as an in-line array of pillars, so that all the four resistances connect to a single node, thus simplifying the calculations. This approximation is justified by the fact that the shift in the structure of the resistance R2h is one-fifth of that of Rv and thus the contribution of the lateral flow resistance between two rows is relatively negligible to the resistance of the narrow gap. To calculate the equivalent resistance of the network model, each node was assigned a coordinate from 1 to (j 9 m), where j and m are the number of rows and columns, respectively, as shown in Fig. 4b. Using Kirchhoff’s circuit laws (Smith 1998), a conductance matrix of the 1 form C P ¼ Q was constructed, where C ¼ R is the conductance matrix and has the dimension of (j 9 m) 9 (j 9 m), and once is multiplied by the matrix of nodal pressures, P, the summation of the flow rates flowing in and out of each node is obtained with P and Q having the dimensions of (j 9 m). Assigning an arbitrary value for the Qj,m, and the same negative value for Q1,1, we calculated the associated values for Pj,m, and P1,1 by solving the matrix. Finally, the resistance between the node 1 and j 9 m of the pump can be calculated as: R¼ Pð1; 1Þ Pðj; mÞ Qj;m ð5Þ The detailed algorithm is provided in the ESI. The total flow resistance of the circuit is the summation of the resistance associated with the CP and the connecting channel. Figure 4c illustrates the resistance of the CP as the function or position of the filling front. For a serpentine CP with a = 0°, the resistance increases rapidly, as the filling front fills the first row, because here, the bulk flow and filling front follow the same path, and the resistance reaches a local maximum when the liquid reaches the edge. Indeed, as the liquid travels back along the second row, the two rows become connected by an increasing number of narrow gaps, which all serve as parallel flow paths, and hence, the flow resistance diminishes as the liquid front advances. In fact, the flow resistance of the CP shown here diminishes between the first and seventh row, from R = 6.98 9 1012 to R = 1.87 9 1012 Pa s m-3, but then starts to increase continuously until it is filled entirely as the vertical Microfluid Nanofluid (2015) 18:357–366 363 Fig. 4 Flow resistance of a serpentine CP. a Schematic showing the liquid flow of a CP. b Lumped resistance model of the CP. c Average flow resistance for the filling front at each row based on nodal analysis of the electrical circuit equivalent, see ESI. The calculated flow resistance decreases sharply as the liquid fills the first few rows of the CP, and starts to increase linearly after the seventh row. d Comparison of theoretical and experimental flow rates using a capillary pressure 1,210 Pa for fitting. The error bars in the experimental data are the RMS values of three independent experiments resistance of the narrow gaps dominate the overall flow resistance. Our analysis showed that for after a spike in resistance in the first row that has no parallel flow path, the resistance diminishes in subsequent rows as many parallel flow paths exist, and then gradually increases in a linear manner up to R = 7.53 9 1012 Pa s m-3, Fig. 4c. The overall flow resistance of the CP remains small compared with microchannels, and in a full capillary circuit, the fluctuations would be further minimized owing to by upstream resistances that may be much larger than the one of the CP. Alternatively, using a CP at a 45° angle, or an inlet placed closer to an edge, the large jump in flow resistance can be avoided as the initial resistance increase is due to the filling of the first, long row in the design shown in Fig. 4. The normalized calculated flow rates of the CP obtained from the computed flow resistance using the capillary pressure as a free parameter set to P = 1,210 Pa matches and reproduces the flow rates observed experimentally, Fig. 4d. The average capillary pressure for each unit cell of the typical serpentine and leading-edge CPs (shown in Fig. 3) is constant and can be calculated by dividing the change in the Gibbs free energy of the unit cell (before and after filling by the liquid) over the volume of the liquid, which is transported into the cell. The equation can be formulated as (Leverett 1941; Hassanizadeh and Gray 1993): DP ¼ cLV ðcos ht DASVt þ cos hb DASVb þ 2 cos hr DASVr Þ Vl ð6Þ where cLV is the surface tension of the liquid (i.e., water), hi=t,b,r are the contact angle of the liquid–air meniscus in the top, bottom, and right surfaces, DASVi¼t;b;r are the respective changes of interfacial areas between solid, and air before and after filling of the cell for the top, bottom, and right surfaces, and Vl is the volume of the liquid transferred in the unit cell (see ESI for more details). From experiments shown in Figs. S6 and S7, we observe that the advancing contact angle of a freshly treated flat Si (associated with the right and bottom surfaces) is 26.4° ± 3.3°, and the advancing contact angle of a layer of PDMS (associated with the top surface of the CP) is 118.7° ± 1.7°; thus, the average capillary pressure generated calculated from Eq. (6) equals to DP = 3,375 Pa (see the ESI for the details of the calculation). Equation 6, however, only constitutes an approximation as it does neither include corner flow effects nor the change in contact angles upon rapid filling (Weislogel and Lichter 1998). 123 364 We measured the flow rates for the pump and repeated the experiments at least three times. We observed the variation in the flow rates in the order of 20 % as shown in the error bars of the experiment, Fig. 4d. Our experiments allowed determining the relative change of the flow rate over time, and the capillary pressure was used as a free parameter to correlate flow rate with calculated flow resistance and was DP = 1,210 Pa. This value corresponds to an equivalent of a capillary pressure generated from a glass tube with the diameter of 80 lm (which can be calculated analytically), which is commensurate with the dimensions of the CP used here with a depth of *100 lm and a distance of *75 lm between the rows. The flow rate of the CP is governed by the capillary pressure and the flow resistance, which are both dominated by the geometry of the CP. The capillary pressure is dominated by the cross section of the rows, and for deep pumps, mainly by W1, and is independent on the overall dimension. The flow resistance is mostly dominated by W2 and the number of posts and gaps between them. The flow resistance of CPs with nonrectangular geometries can be calculated using the electrical network analogy by changing the number of nodes and resistors on each line to match the width of the pump. But as mentioned, in most cases, the external resistance will dominate the overall flow resistance, and the contribution of the CP may be negligible. 3.4 Preprogramed serpentine CPs with varying capillary pressure In many applications, for example, microfluidic immunoassays (Parsa et al. 2008), it is desirable to adjust the flow rate for various steps of the assay. The control over the filling front enables to design CPs with variable capillary pressure. We designed CPs with a gradual decrease in the capillary pressure by increasing the width of the rows from 34 to 130 lm with incremental steps of 3 lm, Fig. 5a and Movie S7 in ESI. Figure 5b illustrates the variation in the average flow rates from one row of the CP to the other. Although the capillary pressure decreases continuously, initially, the speed of the filling front increases rapidly from 3 mm s-1 in the 1st row to 5 mm s-1 in the 8th row of the CP. This value corresponds to a growth from 33 to 62 nl s-1 in the flow rate of the CP. This increase is due to the decrease in the flow resistance, as explained in the section describing on the flow model of the CP. Subsequently, the speed of the filling front diminishes continuously to 2 mm s-1 in the last row, which corresponds to a flow rate 23 nl s-1. It is known that the contact angle of a liquid with a surface increases for increasing flow velocity (Hoffman 1982; Seebergh and Berg 1992; Weislogel and Lichter 1998). Bracke et al. (1989) empirically established a 123 Microfluid Nanofluid (2015) 18:357–366 Fig. 5 Preprogramed serpentine pumps with variable capillary pressure. a Fluorescence image of a CP filled with a green fluorescent solution revealing the increasing width of the rows that lead to a capillary pressure gradient. b Flow rates in each of the rows measured by video microscopy. The error bars are standard deviations of seven experiments. The scale bar is 1 mm (color figure online) relationship between contact angle and capillary number Ca = Ul c , with U being the velocity of the filling front as: cosha coshd ¼ 2 Ca0:5 cosha þ 1 ð7Þ With ha being the static advancing contact angle and hd the dynamic advancing contact angle. The static contact angles of Si and PDMS surfaces are ha = 26.4° and ha = 118.7°, respectively, and thus, the dynamic contact angles of these surfaces are expected to increase in the 8th and last row to 27.3° and 29° for the Si, and to 118.8° and 119.1° for PDMS, respectively, Table S1. Thus, when calculating the capillary pressure at different rows, one needs not only to consider the change in geometry, the increase in flow resistance, but also the change in contact angle. If we neglect the increasing flow resistance of the capillary pump and other dissipative energies, in theory, the average ratio of the generated capillary pressure between the 34th row and the 8th row of the pump is expected to be: DP8 ¼ 2:0 DP34 ð8Þ where DP8 and DP34 are the average capillary pressure generated in the 8th and 34th row, respectively. This ratio is orderly consistent with our experimental data QQ348 ¼ 2:7, where Q8 and Q34 are the average flow rates in the 8th and 34th rows, respectively, and the difference may be in part due to the approximations that were made. The detail of the calculation for the ratio of the capillary pressures can be found in ESI. 4 Conclusions We introduced novel designs for CPs that deterministically guides the filling front along a predetermined path along rows and at the edges. We implemented two novel Microfluid Nanofluid (2015) 18:357–366 architectures, namely a serpentine and leading-edge filling path, respectively. The flexibility and robustness of the design was highlighted using CPs with different filling angles and gradients of pressures. A pore network analysis model was developed to calculate the flow resistance and was used to predict the pressure and flow properties of the different CPs. In comparison with previous CPs, our serpentine and leading-edge CPs can control the progression of the filling front row-by-row deterministically; thus, they can (i) robustly avoid trapping bubbles and be used as capillary driven metering elements, and (ii) they can be used to accurately change the capillary pressure of the CPs at each row. Moreover, it is expected that with this technology, the shape of the CP may be adjusted as well, and various geometry could be made, for example, to modulate the flow resistance as the CP is being filled. The developed CPs are robust and can further expand the functionality of capillary systems. For example, by tailoring the capillary pressure of the CP and the flow resistance of the circuit, the filling time may be tuned from a few seconds to tens of minutes, while accurately metering the liquid. CPs with gradients of capillary pressure may allow varying the liquid flow rate during the incubation steps of bioassays to minimize the time to saturation (Parsa et al. 2008). Also, it will be possible to make CPs that produce a constant flow rate, by compensating the change in resistance with an increase in capillary pressure. We believe programmable CPs will expand the possibilities of passive, capillary driven microfluidics for carrying out advanced fluidic operations autonomously, and this may open new opportunities for the use of capillary systems in immunoassays, point-of-care diagnostics and other areas. Acknowledgments We would like to acknowledge funding from NSERC, CIHR and CFI, and the assistance of the McGill Nanotools and Microfab Laboratory (funded by CFI, NSERC and Nanoquebec). We also acknowledge Prof. Elizabeth Jones for allowing us to use the facilities in her lab. We also acknowledge Arash Kashi, Mohammad Ameen Qasaimeh, and Mohammadali Safavieh for their helpful discussions. DJ acknowledges the support from Canada Research Chair. References Akbari M, Sinton D, Bahrami M (2009) Pressure drop in rectangular microchannels as compared with theory based on arbitrary cross section. J Fluids Eng 131:041202 Bracke M, Devoeght F, Joos P (1989) The kinetics of wetting—the dynamic contact angle. Trends Colloid Interface Sci 79:142–149 Cesaro-Tadic S, Dernick G, Juncker D, Buurman G, Kropshofer H, Michel B et al (2004) High-sensitivity miniaturized immunoassays for tumor necrosis factor a using microfluidic systems. Lab Chip 4:563–569 Chapuis O, Prat M (2007) Influence of wettability conditions on slow evaporation in two-dimensional porous media. Phys Rev E 75:046311-1–046311-5 365 Delamarche E, Bernard A, Schmid H, Michel B, Biebuyck H (1997) Patterned delivery of immunoglobulins to surfaces using microfluidic networks. Science 276:779–781 Delamarche E, Bernard A, Schmid H, Bietsch A, Michel B, Biebuyck H (1998) Microfluidic networks for chemical patterning of substrate: design and application to bioassays. JACS 120:500–508 Delamarche E, Juncker D, Schmid H (2005) Microfluidics for processing surfaces and miniaturizing biological assays. Adv Mater 17:2911–2933 Gervais L, Delamarche E (2009) Toward one-step point-of-care immunodiagnostics using capillary-driven microfluidics and PDMS substrates. Lab Chip 9:3330–3337 Gervais L, de Rooij N, Delamarche E (2011a) Microfluidic chips for point-of-care immunodiagnostics. Adv Mater 23:H151–H176 Gervais L, Hitzbleck M, Delamarche E (2011b) Capillary-driven multiparametric microfluidic chips for one-step immunoassays. Biosens Bioelectron 27:64–70 Groll M, Schneider M, Sartre V, Zaghdoudi MC, Lallemand M (1998) Thermal control of electronic equipment by heat pipes. Rev Gen Therm 37:323–352 Hassanizadeh SM, Gray WG (1993) Thermodynamic basis of capillary pressure in porous media. Water Resour Res 29:3389–3405 Hitzbleck M, Gervais L, Delamarche E (2011) Controlled release of reagents in capillary-driven microfluidics using reagent integrators. Lab Chip 11:2680–2685 Ho C, Ruehli A, Brennan P (1975) The modified nodal approach to network analysis. IEEE Trans Circuits Syst 22:504–509 Hoffman RL (1982) A study of the advancing interface.2. The theoretical prediction of the dynamic contact angle in liquid gas systems. J Rheol 26:606 Juncker D, Schmid H, Drechsler U, Wolf H, Wolf M, Michel B et al (2002) Autonomous microfluidic capillary system. Anal Chem 74:6139–6144 Lee A (2010) VirtualDub. In: VirtualDubMod Surround 1.6.0.0 (ed) Leverett MC (1941) Capillary behavior in porous solids. Trans Am Inst Min Metall Eng 142:152–169 Lovchik RD, Bianco F, Tonna N, Ruiz A, Matteoli M, Delamarche E (2010) Overflow microfluidic networks for open and closed cell cultures on chip. Anal Chem 82:3936–3942 McNeely MR, Spute MK, Tusneem NA, Oliphant AR (1999) Hydrophobic microfluidics. In: Ahn CH, Frazier AB (eds) Microfluidic devices and systems, vol 3877. Spie-Int Soc Optical Engineering, Bellingham, pp 210–220 Mirzaei M, Pla-Roca M, Safavieh R, Nazarova E, Safavieh M, Li H et al (2010) Microfluidic perfusion system for culturing and imaging yeast cell microarrays and rapidly exchanging media. Lab Chip 10:2449–2457 Parsa H, Chin CD, Mongkolwisetwara P, Lee BW, Wang JJ, Sia SK (2008) Effect of volume- and time-based constraints on capture of analytes in microfluidic heterogeneous immunoassays. Lab Chip 8:2062–2070 Prat M (2011) Pore network models of drying, contact angle, and film flows. Chem Eng Technol 34:1029–1038 Safavieh R, Juncker D (2013) Capillarics: pre-programmed, selfpowered microfluidic circuits built from capillary elements. Lab Chip 13:4180–4189 Safavieh R, Zhou GZ, Juncker D (2011) Microfluidics made of yarns and knots: from fundamental properties to simple networks and operations. Lab Chip 11:2618–2624 Seebergh JE, Berg JC (1992) Dynamic wetting in the low capillary number regime. Chem Eng Sci 47:4455–4464 Smith KC (1998) Introduction to electronics. In: KC’s problems and solutions for microelectronic circuits, chap 1, 4th edn, vol 1. Oxford University Press, USA, pp 1–9 123 366 Tamayol A, Yeom J, Akbari M, Bahrami M (2013) Low Reynolds number flows across ordered arrays of micro-cylinders embedded in a rectangular micro/minichannel. Int J Heat Mass Transf 58:420–426 Vulto P, Podszun S, Meyer P, Hermann C, Manz A, Urban GA (2011) Phaseguides: a paradigm shift in microfluidic priming and emptying. Lab Chip 11:1596–1602 Weislogel MM (2003) Some analytical tools for fluids management in space: Isothermal capillary flows along interior corners. In: Narayanan R (ed) Gravitational effects in physico-chemical processes, vol 32, pp 163–70 123 Microfluid Nanofluid (2015) 18:357–366 Weislogel MM, Lichter S (1998) Capillary flow in an interior corner. J Fluid Mech 373:349–378 White FM (2005) Solutions of the Newtonian viscous-flow equations. In: Viscous fluid flow, 3rd edn. McGraw-Hill Science/Engineering/Math, pp 116–125 Zimmermann M, Schmid H, Hunziker P, Delamarche E (2007) Capillary pumps for autonomous capillary systems. Lab Chip 7:119–125 Zimmermann M, Hunziker P, Delamarche E (2008) Valves for autonomous capillary systems. Microfluid Nanofluid 5:395–402
© Copyright 2025 Paperzz