20/04/16 CRFS Workshop ANZSRS 2016 ASM The NUTS & BOLTS of Spirometry Mahesh Dharmakumara Respiratory Scientist The Alfred Hospital Melbourne, VIC. DISCLOSURE The presenter has advised that the following presentation will NOT include discussion on any specific commercial products or service and that there are NO financial interests or relationships with any of the Commercial Supporters of this years ASM. REFERENCES • Alfred Hospital Lung Function Laboratory, Principles and Practice of Spirometry, 2-Day Course, 2016. • ATS/ERS Task Force. Standardisation of Spirometry, 2005. • Borg, B. et al. Interpreting Lung Function Tests: A Step-by-Step Guide. Wiley-Blackwell Publishing, 2014. • Ruppel G. Manual of Pulmonary Function Testing (8th ed), St Louis: Mosby Year Book, 2003. 1 20/04/16 CRFS Workshop Practice Question Q. Why have you decided to attend this workshop? A. To understand fundamental concepts B. To refresh my knowledge C. To work through some practice questions D. All of the above OUTLINE • BTPS • Calibration • Spirogram • Sensors BTPS LUNG SPIROMETER 2.0 litres 1.8 litres BTPS ATPS 0 2.0 37o Celsius PH2O = 47 mm Hg 0 1.8 20o Celsius PH2O = 18 mm Hg When we blow into a “cold” spirometer, the volume recorded by the spirometer is less than that blown out of the lungs. Due to: 1. Gas shrinkage (Charles’ Law) 2. Condensation of water vapour (vapour pressure falls) Alfred 2-Day Spirometry Course, 2016 2 20/04/16 BTPS ATPS to BTPS Conversion Factors, given on page 1 of the CRFS Exam CALIBRATION Description • Calibration is the procedure in which a signal from a device is adjusted to produce a known output. • The adjustment assumes a profile for the relationship between an actual value (known) and a measured value (unknown). • This calibration profile (or curve) depends on the characteristics of the components used in the measurement and allows derivation of a calibration factor. • Calibration factor is stored and applied to all subsequent measurements. CALIBRATION Profile Actual Flow = Actual Volume (fixed, 3L syringe) Measured Time (varied) Calibration Factor = Actual Measured Measured Flow = Measured Volume (signal) Measured Time (varied) 3 20/04/16 CALIBRATION Derivation Volume Actual (3L Syringe) X Calibration Factor Measured (Device) Flow Slow Medium Fast Syringe Plunger Injection Measured Value x Calibration Factor = Actual Value CALIBRATION Procedure Example: Device: Flow Sensing Spirometer CALIBRATION • Standard: Certified 3.0 litre Calibration Syringe Procedure Volume (L) Vs. Time (sec) 4 20/04/16 CALIBRATION • Volume (L) Vs. Time (sec) CALIBRATION • Procedure Flow (L/sec) Vs. Volume (L) CALIBRATION • Procedure Procedure Flow (L/sec) Vs. Volume (L) 5 20/04/16 CALIBRATION Errors • Insufficient system warm-up time • Syringe not fully conditioned to testing environment • Insecure connection between syringe and sensor • Not entering ambient conditions when required • Failure to correct any signal drift by ‘zeroing’ • Not withdrawing and injecting the syringe plunger fully CALIBRATION Vs. Validation • Calibration procedure adjusts the signal whereas the validation procedure only checks the signal. • Validation is also known as a calibration check/test. Software may have a ‘syringe mode’ where BTPS conversion factors can be disabled for syringe validation purposes. • Some types of devices or sensors do not require/allow calibration. In this case, the way of checking if the device is within calibration limits is by performing a validation. • ATS/ERS 2005 states that volumes should meet the accuracy requirements within ±3.5% of true value CALIBRATION Practice Question Q. When calibrating a spirometer with a 3L syringe, the actual value divided by the measured value is known as the: A. Response factor No. This is a term used to describe a ratio in chromatography and spectroscopy. B. Calibration factor Yes. The measured value multiplied by the calibration factor equals the actual value. C. Correction factor No. While it is a type of correction factor, it is specifically not known as this. D. Profile factor No. The relationship between the actual values and a measured values generates a calibration profile. 6 20/04/16 CALIBRATION Practice Question Q55. A spirometer is tested by injecting 3 L volumes from a calibration syringe; the results are as follows: Trial 1 2 3 4 2.897 2.785 3.110 2.960 Which measurements DO NOT meet ATS/ERS (2005) recommendations? A. 1 and 2 B - 1, 2 and 3 C - 2 and 3 • • • • • • D - 2 only Syringe validation procedure, ±3.5% error limits 3.5% of 3.00L 3.5/100 x 3.00 = 0.105L Upper error limit (+3%) = 3.00 + 0.105 L = 3.105L Lower error limit (-3%) = 3.00 – 0.105 L = 2.895L Acceptable range = Lower limit to upper limit = 2.895L to 3.105L • So any value that falls outside the calculated range does not meet ARS/ERS recommendations. • This includes Trials 2 and 3, therefore the answer is C. 2014 Practice Exam Paper SPIROGRAM Types Two types of measured traces, each useful for identifying different sources of error during maneuver Complex Spirogram: Flow (L/sec) Vs. Volume (L) Flow (L/sec) Volume (L) Simple Spirogram: Volume (L) Vs. Time (sec) Volume (L) Time (sec) Alfred 2-Day Spirometry Course, 2016 SPIROGRAM Flow Flow Flow Sources of Error Volume Volume Acceptable effort Not completely full prior to blow Volume Premature termination or glottic closure Actual FVC 0 Time Volume Volume Volume Actual FVC 0 Time 0 Time Alfred 2-Day Spirometry Course, 2016 7 20/04/16 SPIROGRAM Flow Flow Flow Sources of Error Volume Volume Variable or submaximal effort Volume Cough or tongue occlusion 0 Volume Volume Volume Poor start or hesitation 0 Time Time 0 1 Time Alfred 2-Day Spirometry Course, 2016 SPIROGRAM Inspiration FEF Volume Vs. Time Volume (L) 25-75% FEF25%-75% Volume (L) FVC FEV1 FVC FEV1 Expiration 0 1 0 Time (sec) 1 Time (sec) Alfred 2-Day Spirometry Course, 2016 SPIROGRAM Volume Vs. Time Volume (L) • FVC: Forced Vital Capacity FEF25%-75% FVC • FEV1: Forced Expired Volume in the 1st second FEV1 • FER: Forced Expiratory Ratio • FEF25%-75%: Forced Expiratory Flow between 25% and 75% of the FVC 0 1 Time (sec) Alfred 2-Day Spirometry Course, 2016 8 20/04/16 Volume Vs. Time: Vitalograph Example SPIROGRAM Volume (L) Vitalograph Chart Paper • Vertical axis (left): - Volume (L, ATPS) - 1 small square = 1 increment = 0.05L or 50mL • Horizontal axis (bottom): - Time (sec) - 1 small square = 1 increment = 0.1sec or 100msec Time (sec) SPIROGRAM Volume Vs. Time FVC: 1) Identify plateau on the curve: • No change in volume (<0.025L) for ≥1sec, and • Exhalation for ≥6sec in subjects aged >10yrs and • ≥3sec in children aged <10yrs 2) Record the corresponding value from the Volume axis where the plateau occurs Volume (L) FVC 0 SPIROGRAM Volume Vs. Time Volume (L) FVC: 4L + 5 increments = 4L + (5 x 0.05L) = 4L + 0.25L = 4.25L (ATPS) Time (sec) FVC = 4.25L Time (sec) 9 20/04/16 SPIROGRAM Volume Vs. Time FEV1: 1) Identify where the 1 second from the start of expiration has occurred on the Time axis 2) Record the corresponding value from the Volume axis Volume (L) FEV1 0 1 SPIROGRAM FEV1: 3.5L – (1 increment) = 3.5L - (1 x 0.05L) = 3.5L - 0.05L = 3.45L (ATPS) Time (sec) Volume Vs. Time Volume (L) FEV1 = 3.45L Time (sec) SPIROGRAM Volume Vs. Time FER: 1) Determine highest acceptable FEV1 and FVC from the trace Volume (L) FVC 2) Express the ratio of this FEV1 and FVC as a percentage as follows FEV1 FER = FEV1 x 100 FVC 0 1 Time (sec) 10 20/04/16 SPIROGRAM Volume Vs. Time Volume (L) FER: • Highest acceptable FEV1 = 3.45L • Highest acceptable FVC = 4.25L = FEV1 FVC = 3.45L 4.25L FVC = 4.25L FEV1 = 3.45L = 0.81 x 100% = 81% Time (sec) SPIROGRAM Volume Vs. Time FEF25%-75%: 1) Calculate 25% and 75% of the FVC as follows: 25% FVC = 0.25 x FVC 75% FVC = 0.75 x FVC Volume (L) FEF25%-75% = Gradient FVC 75% FVC 2) Mark these values on the curve 3) Draw a straight line through both these points 25% FVC 0 Time (sec) SPIROGRAM Volume Vs. Time FEF25%-75%: 4) Flow is calculated as the change in volume over time which is the gradient of this drawn line Volume (L) FEF25%-75% = Gradient i.e. Flow = ∆V = Rise = Gradient ∆T Run 5) Gradient can be calculated by first using volume value (V2) from line for when time is 1sec 6) Then, use volume value (V1) from intersection of the line, that is, when time is 0sec 7) Finally, V2 V1 0 1 Time (sec) Gradient = V2 – V1 (L) = FEF25%-75% (L/sec) 1 (sec) 11 20/04/16 SPIROGRAM Volume Vs. Time Volume (L) FEF25%-75%: • FVC = 4.25L • 25% FVC = 0.25 x 4.25L = 1.06L FVC = 4.25L V2 = 4.0L • 75% FVC = 0.75 x 4.25L = 3.19L • Gradient = V2 – V1 (L) 1 (sec) = 4.0L – 0.7L = 3.30 L/sec 1sec 75% FVC V1 = 0.7L 25% FVC Time (sec) SPIROGRAM Final Reported Values: Volume Vs. Time Volume (L) • Need to correct all volumes from ATPS->BTPS. • How? Use conversion factor from look up table provided. • i.e. if room temperature is 21deg C, then factor is 1.096. FVC = 4.25L (ATPS) x 1.096 = 4.66L (BTPS) Time (sec) SPIROGRAM • FVC: Forced Vital Capacity Flow Vs. Volume Expiration ‘effort dependent’ PEF ‘effort independent’ • PEF: Peak Expiratory Flow FEF25% • PIF: Peak Inspiratory Flow • FEF75%: Forced Expiratory Flow at 75% of the FVC • Shape Analysis: Determination of acceptable effort and abnormal concavity Flow (L/sec) • FEF25%: Forced Expiratory Flow at 25% of the FVC 1 sec FEF75% FVC 0 Volume (L) PIF Inspiration Alfred 2-Day Spirometry Course, 2016 12 20/04/16 SPIROGRAM Flow Vs. Volume Shape Analysis: Normal 33yr Obstruction with reversibility Normal 60yr Restriction Obstruction Obstruction with reduced FVC Borg, B. et al. Interpreting Lung Function Tests: A Step-by-Step Guide. Wiley-Blackwell Publishing, 2014. SPIROGRAM Back Extrapolation Error Volume Flow Good Start Poor Start 0 Time Volume due to unsatisfactory start of expiration, characterised by slow start or poor take off or excessive hesitation Alfred 2-Day Spirometry Course, 2016 SPIROGRAM • • • Back Extrapolation Error The start of the trace is determined by the back extrapolation method. This will achieve an accurate ‘time zero’ and assure the FEV1 comes from an acceptable curve. Steps: 1) Ensure acceptable FVC achieved 2) Find steepest slope on the curve where PEF occurs from V-T graph 3) Draw back extrapolation line 4) Find new ‘time zero’ 5) Determine extrapolated volume 6) If extrapolated volume is <5% of acceptable FVC or 0.15L (150mL), whichever is greater, trace is acceptable 13 20/04/16 SPIROGRAM Back Extrapolation Error Volume (L) Time (sec) ATS/ERS 2005 states that extrapolated volume must be <5% of acceptable FVC or 0.15L (150mL), whichever is greater Alfred 2-Day Spirometry Course, 2016 SPIROGRAM Practice Question Q97. From the spirogram below performed at a room temperature of 21 degC, the FEV1 is: A. FEV1 = 3.60 L B. FEV1 = 3.65 L C. FEV1 = 4.00 L D. FEV1 = 4.60 L Volume Increments: 0.20L or 200mL FEV1 = 1.00L Time Increments: 0.20sec or 200msec 2014 Practice Exam Paper. SPIROGRAM Practice Question Q97. From the spirogram below performed at a room temperature of 21 degC, the FEV1 is: A. FEV1 = 3.60 L FEV1 = 4.60L B. FEV1 = 3.65 L FEV1 = 4.60L – 1.00L = 3.60L C. FEV1 = 4.00 L D. FEV1 = 4.60 L 1sec 2014 Practice Exam Paper. 14 20/04/16 SPIROGRAM Practice Question Q97. From the spirogram below performed at a room temperature of 21 degC, the FEV1 is: A. FEV1 = 3.60 L 4.65L B. FEV1 = 3.65 L FEV1 = 4.65L – 1.00L = 3.65L C. FEV1 = 4.00 L D. FEV1 = 4.60 L Determine start of test using back extrapolation method 1sec 2014 Practice Exam Paper. SPIROGRAM Practice Question Bonus Question: Based on the given information, does the extrapolated volume meet ATS/ERS recommendations? • Extrapolated Volume = half an increment = 100mL • FVC = 3.83L • 5% of FVC = 0.05 x 3.85 = 0.193 L = 193 mL Is Extrapolated Volume < 5% FVC or 150mL (whichever is greater)? 100mL < 193mL or 150mL Yes, acceptable amount of extrapolated volume. Therefore, calculation of FEV1 is acceptable. SPIROGRAM Practice Question Q68. When an FVC manoeuvre is being performed, the respiratory physiologist should emphasise the importance of a smooth continuous exhalation and which of the following: 1 - Maximal effort Yes. It is a forced manoeuvre therefore maximal effort is required to achieve acceptable (and repeatable) results. 2 - Unhesitating start Yes. This is part of the start-of-test criteria to prevent excessive extrapolated volumes. A. 1 and 2 B. 1 and 3 3 - Avoid leaning forward Yes. To prevent poor posture which may limit maximal performance but also to prevent subject from falling of chair should syncope occur. C. 2 and 3 D. 1, 2 and 3 2014 Practice Exam Paper 15 20/04/16 SPIROGRAM Practice Question Q36. The shape of the 'effort-dependent' portion of the expiratory flow volume curve is determined by which of the following? 2 - lung elastic recoil Yes. How quickly the lung ‘springs’ back effects the maximum flow. PEF Flow (L/sec) 1 - abdominal pressure during forced expiration Yes. The diaphragm being the main ‘pump’ drives abdominal region to generate maximum flow. FEF25% 1 sec FEF75% FVC 0 3 - flow resistance of the small airways No. This effects small airway measures such as FEF25%-75% which is on the ‘effort-independent’ portion of the curve. 4 - cross-sectional area of the small airways No. This effects the resistance of the small airways to flow, similar to 3. Volume (L) ‘effort dependent’ ‘effort independent’ A. 1 and 2 only B. 2 and 3 only C. 1, 3 and 4 D. 4 only 2014 Practice Exam Paper SENSORS Types Two types of measuring principles: • Volume-Displacement Direct measure of respired volume from the displacement of a component • Flow-Sensing Various physical principles used to produce a signal proportional to gas flow SENSORS Volume-Displacement ----------------------------- Wedge-Type (Bellows) ------------------------------- Water-Sealed (Bell) Dry Rolling-Seal (Piston) Alfred 2-Day Spirometry Course, 2016 16 20/04/16 SENSORS Turbine (Rotation) Fleisch (Pneumotachograph) Flow-Sensing Ultrasonic (Pulses) Heated/Hot-Wire (Heat-Loss) Pitot Tube (Pressure) Lily / Screen (Pneumotachograph) Alfred 2-Day Spirometry Course, 2016 SENSORS Volume-Displacement Water-Sealed: • Commonly found chain-compensated and Stead-Wells units • Consists of a bell that is sealed from the atmosphere by water • The bell falls and rises as the subject breathes in and out • A pulley system is attached to the bell and the marking pens record the subjects efforts on the rotating kymograph paper Alfred 2-Day Spirometry Course, 2016 SENSORS Volume-Displacement Wedge-Type (Bellows): • Vitalograph is the most common • Good graphical display of Spirogram (Volume Vs. Time) • Plastic bellows may have linearity problems • Back pressure may be a problem • Frequency response may be a problem • No flow-volume loop Alfred 2-Day Spirometry Course, 2016 17 20/04/16 SENSORS Volume-Displacement Dry-Rolling Seal: • Uses flexible silicone rubber or Teflon-coated rubber seal instead of water • Large-volume piston moves in and out as the patient breathes • Directly recorded by pen on paper • Rather large units Alfred 2-Day Spirometry Course, 2016 SENSORS Flow-Sensing Fleisch & Lily/Screen Pneumotachographs: • Have a resistive element (tubes or mesh screen) in the flow tube • Flow through pneumotach proportional to pressure difference • Tubing connect to differential-pressure transducer • Pneumotach must be linear over desired flow range • Measurements vary with gas viscosity, humidity and temperature • Good frequency response • Usually heated to prevent condensation of expired water vapour Alfred 2-Day Spirometry Course, 2016 SENSORS Flow-Sensing Turbine: • Flow signal obtained from rotation of the turbine, counted photoelectrically • Deposits of water droplets and dirt can change the mass of the turbine and rotational ability and thus affect the calibration • Fine hair/fluff can alter calibration • Not ideal for measuring PEF Alfred 2-Day Spirometry Course, 2016 18 20/04/16 SENSORS Flow-Sensing Heated/Hot-Wire: • Also known as heat-transfer or thermistor type • Type of mass airflow sensors • Heated thermistor that is cooled as gas flows past it • Temperature transducer automatically increases and measures the flow of electricity to the thermistor to keep it at the required temperature • Fragile Alfred 2-Day Spirometry Course, 2016 SENSORS Flow-Sensing Pitot Tube: • It is composed of a tube inserted horizontally along the axis of the gas flow line. • The pressure at the end of the tube is compared with the static pressure to determine the final gas flow rate within the flow line. Alfred 2-Day Spirometry Course, 2016 SENSORS Flow-Sensing Ultrasonic: • A pulse that travels with gas flow is sped up and against gas flow is slowed down • Gas flow measured from the transit times • No moving parts • Measurement independent of gas composition, pressure, temperature and humidity • Calibration not required, but validation possible Alfred 2-Day Spirometry Course, 2016 19 20/04/16 SENSORS Practice Question Q46. During repeated FVC manoeuvres using an unheated Fleisch pneumotach the recorded dynamic volumes are observed to progressively increase. The most likely explanation for this observation is: A - Condensation on the resistive elements of the pneumotach is increasing its resistance and therefore overestimating the measured flow Yes. Flow derivation is based on resistance. Increasing resistance on sensor is measured as increasing flow. B - Calibration error. No. If this was the case there would be consistent but inaccurate volumes recorded, not progressively increasing volumes. C - The pressure transducers are drifting. No. If this was the case, recorded volumes may be overestimated by no progressive increase. D - There is excessive noise in the flow signal. No. Recorded volumes may be inaccurate but a progressive increase would not be seen. 2014 Practice Exam Paper SENSORS Practice Question Q92. In order to measure flow using a pneumotachograph system which of the following electromechanical devices must also be used? A - pressure gauge No. This is a device for only measuring the pressure of a gas. B - spirometer No. This is a system as a whole which includes sensor and computing hardware. C - differential pressure transducer Yes. This device measures the mechanical pressure signal and converts it into an electrical signal. D – rotameter No. This is a type of flow meter. 2006 & 2009 Practice Exam Paper SENSORS Q57. Which of the following devices has the highest frequency response? Hint: Think about which sensor is good at measuring PEF accurately. A - Wedge spirometer B - Turbine peak flow meter C - Water-seal spirometer D - Differential pressure pneumotachograph 2006, 2009, 2014 Practice Exam Paper 20 20/04/16 CRFS Workshop ANZSRS 2016 ASM A final piece of advice… Think about why the correct answer is right, and why the incorrect answer is wrong. 21
© Copyright 2026 Paperzz