Title Fat Removal Using a French Cryolipolysis Device

Cryolipolysis 27-01-2016
Title
Fat Removal Using a French Cryolipolysis Device: Retrospective Study of 418 Treatments
Author
Naouri M.1, 2
Author’s Affiliations
1
Centre Laser, 94130 Nogent sur Marne, France
2
Centre Laser International de la Peau, 75007 Paris, France
Corresponding Author
Michaël Naouri (email: [email protected])
Conflicts of Interest
Deleo: loaned the device and funded the study
Bioderma: paid conferences
Keywords
Cryolipolysis, fat removal, cellulitis, panniculitis, cold
Short Title
French Cryolipolysis
Statistics
•
Abstract: 256 Words
•
Text Core:
•
References: 19 References
1
Cryolipolysis 27-01-2016
•
Tables: 4 Tables
•
Figures: 5 Figures
•
Pictures: 6 Pictures
2
Acknowledgements
The author acknowledges the editing support of Laboratoire Bioderma, France and of its collaborator,
Sandrine Virassamynaik and the assistance of Arkesys, France for statistical analyses.
Cryolipolysis 27-01-2016
3
Abstract
Introduction: Cryolipolysis is a technique that induce selective apoptosis of the adipocytes using
controlled exposure to intense cold. The aim of our work was to study the tolerance and efficacy of a
new device.
Method: Retrospective observational post-commercial study in patients treated with a cryolipolysis
device designed and manufactured in France. The assessment was done after a single session, in patients
who paid for their treatment, by evaluating tolerance, measuring the perimeter of the treated area, and
measuring physician and patient satisfaction.
Results: A total of 418 areas were treated. The treatment was carried out without any significant side
effect in 92.5% of cases. Mean loss of circumference was 2.8 cm. The maximum circumference
differential was 10 cm. A loss of at least 1 cm was reported in 89.4% of cases. A total of 75.4% of
patients were satisfied or very satisfied. Out of all patients, 80.6% wanted to carry out a second session.
Treatment response was not correlated to initial perimeter or age.
Discussion: Most side effects had already been reported in the literature and no serious complications
were reported. This safety of use is important when faced with an influx of devices sold in Europe and
serious side effects are reported for some of them. We also show a high level efficacy of the French
cryolipolysis device based on objective and subjective evaluation in patients who paid for their
treatments. It was not possible to define the predictive factors for treatment response.
Conclusion: Cristal™ cryolipolysis is a safe and effective device for fat removal.
Cryolipolysis 27-01-2016
4
Introduction
Requests for dermatological aesthetic treatments are constantly increasing. Patients want less invasive
procedures with a minimum amount of sequelae and side effects. This explains why the demand for
surgical techniques is stagnating, while the demand for non-invasive medical techniques is booming.
Out of the most common requests, changing the body’s silhouette is the most difficult to carry out. This
generic term covers different problems: sagging skin (laxity), cellulite (orange peel effect), and excess
fat. There are several different techniques used for fat removal [1]: radio frequency and soft laser, whose
effects are relatively limited; high intensity focused ultrasound, that is cumbersome and/or a painful
procedure; and more recently cryolipolysis. The most important aspect of non-invasive aesthetic
treatments is selectivity. As with any treatment, selectivity helps act on the target area with minimal
impact on surrounding tissue: selective destruction of vessels, hair, or pigment using lasers, selective
treatment of a facial muscle with botulinum toxin, etc. The selective effect of cold on hypodermis is a
long-known medical phenomenon that can cause cold panniculitis when it reaches its peak, as described
in horsewomen [2] or children who eat popsicles [3]. Several authors have shown that it is possible to
induce selective apoptosis in the adipocytes of animals using controlled exposure to cold [4][5]. This
selective fat destruction technique using cold was named cryolipolysis [6] and was then applied to
humans using a device that can suck out fat pad into a hollow handpiece and then chill it with two
electrodes, cooled by the Pelletier effect. The first device’s safety and efficacy of use has been reported
in many studies [7][8][9][10][11][12][13]. This is a technique whose efficacy is particularly noteworthy
on localized fat pad. Many cryolipolysis devices have appeared on the market, especially in Europe, and
operate in the same way. Their reliability was recently challenged by an article on cold burns that
occurred after using some devices imported from Asia [14]. The goal of our study was to evaluate the
safety and efficacy of use of a new cryolipolysis device manufactured in France, and to try to determine
the predictive factors for treatment response.
Cryolipolysis 27-01-2016
5
Method
This retrospective observational monocentric post-commercial study was carried out in patients treated
at the Nogent sur Marne Laser Center in France for a period of one year. Inclusions were carried out
consecutively. All patients paid for their treatment. A single session was evaluated.
The inclusion criterion was: an adult patient treated for one or several localized fat pad that could be
pinched between the inch and index fingers.
Exclusion criteria included: a medical history of cold disorders (cryoglobulin, Raynaud’s disease, etc.),
visceral hernias, pregnancy, a recent Caesarean section (<6 months).
Treatment Procedure
The device used was the Cristal™ cryolipolysis (Deleo, Saint Raphael, France). It was designed and
built in France and benefits from medical CE marking. It has two slightly curved handpieces that can be
used simultaneously, and that exist in three different sizes. The device has been sold and used in France
for over a year for fat removal.
The largest sized handpiece was chosen based on the size of the area to be treated. A thick protective
epidermal membrane was laid over the treatment area after marking it with a dermographic pencil. The
markings were also used to measure circumference by noting the vertical distance between an
anatomical area (most often the umbilicus) and the area to be treated. The membrane was then soaked
with a cold-resistant gel in order to maximize the drop in temperature on the treatment area. The
treatment probe was placed on the area, then the suction was set between 3 and 4 according to the
thickness of the fat pad (the larger the bulge, the stronger the suction). The cooling temperature was
between -6°C and -10°C. In case of pain during the suction phase, suction was kept at 1 with a gradual
increase made possible by progressive anesthesia by the cold. Treatment duration was 60 minutes per
area. A 5-minute energetic massage was carried out immediately after the probe was removed. Cicabio
Arnica+ (Bioderma, Lyon, France) was prescribed and it was to be applied once daily for one week after
treatment.
Cryolipolysis 27-01-2016
6
Evaluation Criteria
Evaluation of Tolerance
Immediate tolerance was evaluated by examining the patient during the session and 15 minutes after
it. Patients were contacted, by telephone, the day after the session in order to screen out and take care
of any precocious complications. If needed, they were brought back in and treated until symptoms
were alleviated.
Patients were told to call in the case of unusual side effects, especially paradoxical adipose hyperplasia
on the treatment area within 3 months after cryolipolysis.
Evaluation of Efficacy
Efficacy was evaluated in patients at least 2 months after their session.
An objective evaluation was carried out by measuring the perimeter of the treated area before and 2
months after treatment according to anatomic markings with a deep breath. Three measurements were
taken for each evaluation and the mean of the three measurements was used as a reference if the
difference between them did not exceed 10%. Otherwise, three new measurements were taken.
A subjective evaluation was carried out by measuring physician and patient satisfaction (1:
dissatisfaction, 2: minimal satisfaction, 3: high satisfaction, 4: very high satisfaction). We also noted
patients who wanted to redo a cryolipolysis session.
Statistical analysis
All statistical analyses were performed using Minitab version 17 software with statistical significant set
at p<0.05. For quantitative variables, descriptive statistics were presented as mean ± standard deviation,
minimum, maximum and median. For qualitative variables, results were presented as a percentage.
Differences in quantitative outcomes were assessed using paired Student test (T-test). Pearson’s
correlation tests were performed when applicable.
Cryolipolysis 27-01-2016
7
Results
A total of 418 areas were treated on 147 patients. The treated areas were: the abdomen (144), the love
handles (156), the inner thighs (48), the back (26), the underside of buttocks (26), the inner knees (12),
and the breasts (6).
Tolerance (Table 1)
Out of all patients, 92.5% did not have any side effects except for minimal regressive ecchymosis.
Two treatments could not be carried out. The first one was supposed to be on the inner knees. The female
patient had a large lipedema. Suction while placing the probe led to high pain that was expected with
her condition [15]. This caused her to stop treatment. The second one was interrupted because a female
patient had a vasovagal attack with loss of consciousness 20 minutes after the probe was placed. The
patient fell off the examination table while she was alone. In spite of this incident, she wanted to start
the session again, but the physician refused. Two other vasovagal attacks without loss of consciousness
occurred. One at the beginning of the session and the other happened when the probe was taken out.
These were rapidly controlled after the patients were put in decubitus and treatment could be continued.
Eight patients had unusual sequelae:
-
painful induration after the session: 4 patients
-
skin anesthesia in the treatment area: 1 patient
-
large ecchymosis: 2 patients
-
erythema and phlyctena: 1 patient. The protective membrane slipped off the patient during
treatment.
No serious or irreversible complication was reported.
Efficacy
A total of 57 patients were reviewed for the efficacy evaluation (Table 2, Table 3, Figure 1 and Figure
2).
Cryolipolysis 27-01-2016
8
A mean statistically significant loss of circumference of 2.8 cm was observed after treatment including
the placement of 1-3 probes (Table 2). A loss of circumference of at least 1 cm was observed in 89.4%
of patients. The maximum loss observed was 10 cm. For the population studied, cryolipolysis induced
a significant loss of perimeter of the treated area that averaged between 2.218 and 3.361 cm; with a
confidence interval of 95% [CI at 95 % (2.218; 3.361)] (Table 2).
The mean loss according to localization and the number of areas treated appear in Table 3. The loss of
circumference was statistically higher after treating the abdomen and love handles (3 probes) than when
treating the abdomen only (1 probe) (Table 3 and Figure 2). On the other hand, the difference between
the loss of circumference between treating the abdomen (1 probe) and the love handles (2 probes) was
not statistically significant.
Mean patient and physician satisfaction was 3/4 (Figure 3). A total of 75.4% of patients and 75.4% of
physicians were satisfied or very satisfied.
Correlation Analyses
No correlation could be made between patient age or initial perimeter and the relative or absolute loss
of circumference (Table 4).
There was a positive correlation between the physician’s and patients’ subjective evaluation (Table 4)
and the objective decrease in circumference as well as between the physician’s and patients’ evaluation
(Figure 4 and Figure 5). The physician’s score was more precise and objective because it helped better
differentiate the mean level of loss.
Cryolipolysis 27-01-2016
9
Discussion
Our study showed the safety and efficacy of use for a new cryolipolysis device that was designed and
manufactured in France.
Because this was a device with aesthetic purposes used on healthy people, safety of use was the most
important factor. Most cryolipolysis devices operate according to the same principle, but may differ
based on how reliable their manufacturing and treatment protocols are. The objective retrospective
survey of all treatments carried out consecutively helped us objectively evaluate treatment safety on a
very large number of treated areas. To our knowledge, it is the largest series of safety studies on
cryolipolysis done after Diericx’s [7]. Our study showed that the device’s tolerance was excellent
without side effects or serious and/or irreversible complications in over 90% of cases. Most
complications described were already reported at higher frequencies with other cryolipolysis devices.
We had one case of cold burning. It was a superficial second-degree burn: erythema and phlyctena, and
healed rapidly without sequelae. It was due to the membrane slipping and not to the device
malfunctioning. This helped us to show the importance of epidermic protection with a suitable
membrane. Be as it may, the frequency of this complication was much lower than for other energybased devices commonly used for aesthetic purposes, such as hair removal lasers or radio frequency
[14]. On the other hand, there were no cases of paradoxical adipose hyperplasia reported. This is a much
more serious condition because it is irreversible and is the opposite result of the treatment goal [11][13].
We deliberately used strong sucking pressure in order to improve fat pad penetration in the probe and
to increase the cold effect by diminishing the treated area’s vascularization. Cicabio Arnica+ was
consistently prescribed in order to limit and treat secondary ecchymosis caused by suction from the
cryolipolysis probe. To our knowledge, vasovagal attacks have never been mentioned in the literature
with this type of technique. However, they are a common complication. This is why we always stayed
in the treatment room at the beginning and end of each session. In spite of this, we reported the
occurrence of a vasovagal attack after 20 minutes of treatment that could have led to traumatic and/or
neurological complications. After this incident, we put in place and recommend the consistent use of
alarms, as well as checking up on the patient every 15 minutes.
Cryolipolysis 27-01-2016
10
To our knowledge, this is the largest series of efficacy studies on cryolipolysis. Perimeter of the area
treated was the objective evaluation criterion of the efficacy that we chose for the clinical practice of
cryolipolysis. We tried to make it as objective as possible by using anatomical markings and taking
multiple measurements. In our experience, measuring perimeter is more easily reproducible that
measuring fat fold and it is especially matches with what the patient feels. So we believe it is the most
pertinent way to evaluate them.
Loss of circumference was around 3 cm with 1-3 probes. To evaluate the effect of the number of probes,
we focused on studying the abdominal area (abdomen +/- love handles) which was where most
treatments occurred. Even though we showed that treating these three areas is significantly more
efficacious than treating one area; it is interesting to note that the loss of circumference was not
significantly different between treating one (abdomen) and two areas (love handles). Treating the two
love handles was slightly less efficacious than only treating the abdomen. This shows the meshing of
different parameters in the loss of circumference: the number of treatment areas but also the quantity of
fatty tissue in the area and ease of suction. These two aspects were less important on the love handles
than in the abdominal area and may explain their lower response in spite of placing two probes there.
Patients’ objective response to treatment was very different from one patient to another. Some patients
were able to lose 10 cm of waist circumference after one session and others responded less well to
cryolipolysis. This is common with treatments using energy-based devices. In a previous study on CO2
fractional lasers, we showed that young patients with thin skin were the best responders [16][17]. In this
study, we tried to determine the predictive response criteria to cryolipolysis. Unfortunately, none of the
criteria tested (initial perimeter and patient age) were statistically correlated to treatment success. There
are probably individual susceptibilities of hypodermis to cold that cannot currently be predicted by
clinical exam or history taking.
In spite of this response variability, the level of patient satisfaction to the technique was very important.
According to some authors, the satisfaction criterion is the single most important consideration in the
choice of an energy-based device treatment [18]. Since all treated patients paid for their treatment, there
Cryolipolysis 27-01-2016
11
was no gratitude bias [19]. The fact that over 80% of patients were ready to pay for a new session also
showed how objective this high satisfaction is. This goes along with the 80% of patients who were ready
to recommend this treatment to a friend in Diericx’s study [7]. This last statistic encouraged us to suggest
several sessions to the patient during the preliminary consultation, even though we showed efficacy and
very high satisfaction after a single session. It has been shown that a second session could help improve
results [19]. We showed that patient satisfaction is more subjective than doctor satisfaction, which is
more correlated with the objective measurement of the perimeter. This impelled us to provide better
information to the patients on the reality of this technique in order to avoid excessive or unrealistic
expectations.
Our study showed that the Cristal™ cryolipolysis is a safe and effective device for fat removal with a
high level of patient satisfaction.
Cryolipolysis 27-01-2016
12
Tables
Table 1: Tolerance
N total
(patients)
Percentages
(%)
N total
(number of
areas treated)
Percentages
(%)
Vasovagal Attack
147
1.36
418
0.48
Painful Induration After the Session
147
2.72
418
0.96
Skin Anesthesia in the Treatment
Area
147
0.68
418
0.24
Large Ecchymosis
147
1.36
418
0.48
Erythema and Phlyctena
147
0.68
418
0.24
Total
147
6.80
418
2.39
Cryolipolysis 27-01-2016
13
Table 2: Efficacy
N
Mean
Standard
Deviation
Minimum
Maximum
Median
Age, Year
57
49.40
14.69
23
66
52
Perimeter
Before
Cryolipolysis,
cm
57
89.58*
17.28
38
116
95
Perimeter After
Cryolipolysis,
cm
57
86.79*
16.97
34
115
92
Loss of
Perimeter, cm
57
2.79
2.15
0
10
2
Wanted a
Second Session
57
*
Yes: 45 (80.36%)
No: 11 (19.64%)
Did Not Specify: 1 (1.75%)
: CI at 95 % for the difference in means: (2.218; 3.361)
T test of the difference in means = 0 (and ≠ 0) : T-value = 9.78; p-value = 0.000
Cryolipolysis 27-01-2016
14
Table 3: Loss of Perimeter Based on the Location and Number of Probes Placed
Loss of
Perimeter, cm
Location
N
(patients)
Number
of
Probes
Placed
Mean
Abdomen
24
1
2.38*
1.69
0
6
2
Thighs
3
1
1.33
0.58
1
2
1
Knees
3
1
2.00
1.73
1
4
1
2 Love
Handles
8
2
2.13**
2.36
0
7
2
Under the
Breasts
2
2
3.00
1.41
2
4
3
Buttocks
2
2
4.50
4.95
1
8
4.5
2 Love
Handles
+
Abdomen
15
3
4.00*, **
2.36
2
10
3
Standard
Minimum Maximum
Deviation
*
: CI at 95 % for the difference in means: (0.173; 3.077)
T test of the difference in means = 0 (and ≠): T-value = 2.32; p-value = 0.030 ; DL = 22
**
: CI at 95 % for the difference in means: (-0.34; 4.09)
T test of the difference in means = 0 (and ≠): T-value = 1.82 p-value = 0.091 DL = 14
Median
Cryolipolysis 27-01-2016
Table 4: Pearson’s Correlation Test
Pearson’s Correlation test
Pearson’s Correlation of Loss (cm) and Initial Perimeter (cm) = 0.206
p-value = 0.124
Pearson’s Correlation of Relative Loss (%) and Initial Perimeter (cm) = -0.114
p-value = 0.396
Pearson’s Correlation of Perimeter Loss (cm) and Age (year) = 0.036
p-value = 0.791
Pearson’s Correlation of Relative Loss (%) and Age (year) = -0.040
p-value = 0.768
Pearson’s Correlation of the Physician Satisfaction and Patient Satisfaction = 0.684
p-value = 0.000
15
Cryolipolysis 27-01-2016
16
Figures
Figure 1: Diagram of Individual Values (Before – After)
Figure 2: Diagram of Individual Values of Perimeter Loss Based on Location
10
Loss of perimeter (cm)
8
6
4
2
0
2
es
id
S
2
s
de
Si
+
m
do
Ab
en
m
do
Ab
en
hs
ig
Th
Location
ck
t to
u
B
s
es
ne
K
rt
de
n
U
he
st
ea
br
s
Cryolipolysis 27-01-2016
17
Figure 3: Histogram of Physician (up) and Patient (down) Satisfaction
25
20
Number
15
10
5
0
1
2
3
4
Physician satisfaction (score 1-4)
25
20
Number
15
10
5
0
1
2
3
Patient satisfaction (score 1-4)
4
Cryolipolysis 27-01-2016
18
Diagramme
des
indivduelles
Figure 4: Loss of Perimeter
(cm) Based
onvaleurs
Physician
Satisfaction
Perte de périmètre (cm) en fonction de l'évaluation du médecin
10
Loss of perimeter (cm)
8
6
4
2
0
1
2
3
4
Physician Satisfaction (score 1-4)
individual
values
Figure 5: Loss of Perimeter Diagramof
(cm) Based on
Patient Satisfaction
Loss of permeter (cm) based on patient satisfaction
10
Loss of perimeter (cm)
8
6
4
2
0
-2
-4
1
2
3
Patient satisfaction (score 1-4)
4
Cryolipolysis 27-01-2016
19
Pictures
Picture 1: red bump: appearance withdrawal of the probe, disappear in a few minutes during the
massage
Picture 2: phlyctena and erythema where the membrane slipped (to be noted, the absence of lesion
on the protected underlying pinkish skin)
Cryolipolysis 27-01-2016
Picture 3: Oedema and indurated appearance
Picture 4: Abdomen (Before/After)
20
Cryolipolysis 27-01-2016
Picture 5: Side (love handle) – Before/After
Picture 6: Abdomen + Love Handles (Before/After)
21
Cryolipolysis 27-01-2016
22
References
[1]
R. S. Mulholland, M. D. Paul, and C. Chalfoun, “Noninvasive body contouring with
radiofrequency, ultrasound, cryolipolysis, and low-level laser therapy.,” Clin. Plast. Surg., vol.
38, no. 3, pp. 503–20, vii–iii, Jul. 2011.
[2]
R. Vickers, “Equestrian cold panniculitis,” British Medical Journal (Clinical research ed.), vol.
282, no. 6261. BMJ Group, p. 405, 31-Jan-1981.
[3]
E. H. Epstein and M. E. Oren, “Popsicle panniculitis.,” N. Engl. J. Med., vol. 282, no. 17, pp.
966–7, Apr. 1970.
[4]
B. Zelickson, B. M. Egbert, J. Preciado, J. Allison, K. Springer, R. W. Rhoades, and D.
Manstein, “Cryolipolysis for noninvasive fat cell destruction: initial results from a pig model.,”
Dermatol. Surg., vol. 35, no. 10, pp. 1462–70, Oct. 2009.
[5]
T.-R. Kwon, K. H. Yoo, C. T. Oh, D. H. Shin, E. J. Choi, S. J. Jung, H. Hong, Y. S. Choi, and
B. J. Kim, “Improved methods for selective cryolipolysis results in subcutaneous fat layer
reduction in a porcine model.,” Skin Res. Technol., vol. 21, no. 2, pp. 192–200, May 2015.
[6]
M. M. Avram and R. S. Harry, “Cryolipolysis for subcutaneous fat layer reduction.,” Lasers
Surg. Med., vol. 41, no. 10, pp. 703–8, Dec. 2009.
[7]
C. C. Dierickx, J.-M. Mazer, M. Sand, S. Koenig, and V. Arigon, “Safety, tolerance, and
patient satisfaction with noninvasive cryolipolysis.,” Dermatol. Surg., vol. 39, no. 8, pp. 1209–
16, Aug. 2013.
[8]
S. R. Coleman, K. Sachdeva, B. M. Egbert, J. Preciado, and J. Allison, “Clinical efficacy of
noninvasive cryolipolysis and its effects on peripheral nerves.,” Aesthetic Plast. Surg., vol. 33,
no. 4, pp. 482–8, Jul. 2009.
[9]
L. Garibyan, L. Cornelissen, W. Sipprell, J. Pruessner, S. Elmariah, T. Luo, E. A. Lerner, Y.
Jung, C. Evans, D. Zurakowski, C. B. Berde, and R. Rox Anderson, “Transient Alterations of
Cutaneous Sensory Nerve Function by Noninvasive Cryolipolysis.,” J. Invest. Dermatol., vol.
135, no. 11, pp. 2623–31, Nov. 2015.
[10]
K. B. Klein, B. Zelickson, J. G. Riopelle, E. Okamoto, E. P. Bachelor, R. S. Harry, and J. A.
Preciado, “Non-invasive cryolipolysis for subcutaneous fat reduction does not affect serum
lipid levels or liver function tests.,” Lasers Surg. Med., vol. 41, no. 10, pp. 785–90, Dec. 2009.
[11]
H. R. Jalian, M. M. Avram, L. Garibyan, M. C. Mihm, and R. R. Anderson, “Paradoxical
adipose hyperplasia after cryolipolysis.,” JAMA dermatology, vol. 150, no. 3, pp. 317–9, Mar.
2014.
[12]
S. R. Keithi-Reddy, T. V Patel, a W. Armstrong, and a K. Singh, “Uremic pruritus,” Kidney
Int, vol. 72, no. 3, pp. 373–377, 2007.
[13]
S. M. Singh, E. R. C. Geddes, S. G. Boutrous, R. D. Galiano, and P. M. Friedman,
“Paradoxical adipose hyperplasia secondary to cryolipolysis: An underreported entity?,” Lasers
Surg. Med., vol. 47, no. 6, pp. 476–8, Aug. 2015.
[14]
A. M. Tremaine and M. M. Avram, “FDA MAUDE data on complications with lasers, light
sources, and energy-based devices.,” Lasers Surg. Med., vol. 47, no. 2, pp. 133–40, Feb. 2015.
[15]
M. Naouri, M. Samimi, M. Atlan, E. Perrodeau, C. Vallin, G. Zakine, L. Vaillant, and L.
Machet, “High-resolution cutaneous ultrasonography to differentiate lipoedema from
lymphoedema.,” Br. J. Dermatol., vol. 163, no. 2, pp. 296–301, Aug. 2010.
[16]
M. Naouri, M. Atlan, E. Perrodeau, G. Georgesco, R. Khallouf, L. Martin, and L. Machet,
“High-resolution ultrasound imaging to demonstrate and predict efficacy of carbon dioxide
fractional resurfacing laser treatment.,” Dermatol. Surg., vol. 37, no. 5, pp. 596–603, May
2011.
Cryolipolysis 27-01-2016
23
[17]
D. M. Ozog, “Predicting efficacy of fractional carbon dioxide laser treatment for resurfacing.,”
Dermatol. Surg., vol. 37, no. 5, p. 604, May 2011.
[18]
A. Faurschou, A. Olesen, J. Leonardi-Bee, and M. Haedersdal, “Lasers or light sources for
treating port-wine stains.,” Cochrane Database Syst Rev., vol. 11, no. 9, p. CD007152, 2011.
[19]
S. Y. Shek, N. P. Y. Chan, and H. H. Chan, “Non-invasive cryolipolysis for body contouring in
Chinese--a first commercial experience.,” Lasers Surg. Med., vol. 44, no. 2, pp. 125–30, Feb.
2012.