CFL-M Inca Protocol Inca measurement Author : Dinant Goorhorst, May 2014 Introduction This info sheet gives a brief description of a stepwise protocol of starting a Pressure Volume Measurement. Read also: ‘Quick start: How to perform a PV-study’ in the Inca user manual. Choice of Catheter Optimal pressure-volume recordings require that you use the right catheter. Read info sheet ‘Select the best catheter’ to read information on which catheter to use for different ventricular sizes and how to optimize catheter settings CD Leycom offers catheters with electrode spacings 6, 8, 10 or 12 mm. Study preparation Enter study settings During preparation of the patient, enter the study settings. Start with creating a new study. You must fill in a study name and an ID number. Protocol Inca – Monitoring Check the software settings 1. Choice of catheter Based on fluoroscopy or Echo/MRI values of EDV/ESV – EF Longaxis (mm) -> catheter type < < < > 55mm 70mm 90mm 90mm - 6 mm electrode spacing 8 mm electrode spacing 10 mm electrode spacing 12 mm electrode spacing Select “Scaling” and check the scaling values. Normally during start-up all channels are set to ‘Auto-scale’. Select “Catheter” in the settings menu. Enter the current catheter that will be used. Select “Channels” in the settings menu. Check and update the channel names and units for the external signals. Check also the channel-view assignment. The pressure will be used in the PV-loop and the ECG will be used to find the markers. 2. Study preparation a. Enter study settings b. Check software settings 3. Catheter placing a. Connect catheter to Pressure Module b. Optimal positioning -> pigtail in apex, Segment 7 around aortic valve (less than 1 cm under valves) c. Fixation catheter when optimal position reached. 4. Volume calibration a. SV calibration by Thermodilution – preferably 3 to 5 shots b. Alternated by EFcal 10ml of 6% saline when CO < 8L 10ml of 10% saline when CO> 8L -> to data 1… files + manual input SVcal or from SVcal files -> real-time absolute values The calibration of the conductance catheter is done automatically, however the user-supplied signals connected to the external inputs require initial calibration to enable the software to convert input signal voltages to user units. þ Note Select ‘Create new study’ during start-up, you will guide through these steps automatically. CD Leycom info 1423 CFL-M Inca Protocol Inca measurement Author : Dinant Goorhorst, May 2014 Catheter placing Measurement Prepare catheter Start your measurement always with a baseline. It is recommended to repeat a baseline (during and) after a measurement to verify if the situation is still the same (for instance: all the segments are still in the ventricle.) Flush the catheter first with a sterile saline solution. Then wet the sensor by immersing the distal portion of the catheter for approximately 10 seconds in a saline solution. Remove the catheter from saline, lay catheter on sterile table and cover sensor with sterile by saline wetted gauze. Connect catheter’s pressure arm Connect the catheter’s pressure arm to the pressure module of the Inca. On the screen you have to confirm that pressure calibration can be started. þ Important The catheter should not be moved or touched until the pressure calibration is ready (after approximately 15 seconds). Optimal positioning þ Optional During the insertion of the catheter, you may record the aorta pressure. During insertion of the catheter check the pressure curve to see if the pressure curve is changing from aorta pressure to left ventricular pressure. Connect the catheter’s volume arm to the volume module of the Inca. Select the screen with the segmental volumes. Check the segments who are moving simultaneously. þ Optional If 5 or less segments are in the ventricle, decrease the electric field by changing the current electrodes. Determination of slope factor SVcal Determine CO by a reference method following standard procedure. This could be performed by thermodilution to achieve a reliable average 5 measurements will be optimal. Each time that Cardiac Output is measured, using a reference method, data is acquired simultaneously with Conduct NT. Enter the value of the reference method in your comment and select the type ‘SVcal’. Now you can also enter the CO in the column ‘COref’. Determination of EFcal by measuring parallel conductance Acquire the EFcal-files during a saline injection. Use 10 ml of 6% with a CO less than 8L and use 10 ml of 10% with a CO above 8L. Wait for a steady state (2 – 3 seconds) as judges by the stability of the PV-loops. Start data-acquisition and inject the hypertonic saline through the distal port of the thermodilution catheter. The PV-loops move to the right because the volume is increasing slightly. Wait for the wash-out of the saline and PV loops starting to return to normal then stop acquisition. Fill in a comment and select the type ‘EFcal’. You can read the result of the saline injection directly in the displayed table. If you have three results close to each other, you know that you do not have to repeat the saline injection. If there is still some doubts regarding the results, you can repeat the saline once again. þ Note It is recommended to wait a while before you repeat a saline injection. In the mean time you can acquire an SVcal file after each EFcal-file or do a visual check on the acquired EFcal-files. Use 11 electrodes with 5 segments. Use 10 electrodes with 3 or 4 segments. þ Note It is recommended to display at least 1 segment that is outside the ventricle. þ Optional If MRI or Echo values available for EDV and ESV, you may prefer to use these values to calibrate volume. CD Leycom info 1423
© Copyright 2026 Paperzz