2010.02 DATA No.D-8 Page 1 / 1 Weekly Application Note Category Potentiometric Titrator COM series Sulfite ions in gold plating solution Plating by Redox titration method Referenced methods sulfite ions, gold plating solution, redox titration Key words; Outline The determination of sulfite ions contained in gold plating solution can be performed by the following procedure: Add a certain amount of iodine solution to the sample solution to react with sulfite ions in the sample as indicated formula (1). Æ After acidifying with hydrochloric acid, titrate the remaining iodine with sodium thiosulfate solution as shown in formula (2) by the redox back titration. SO32- + I2 2Na2S2O3 + I2 Æ SO42Æ + 2NaI 2H+ + 2I- ・・・ (1) Na2S2O6 ・・・ (2) + Reagents Titrant : 0.1 mol/L sodium thiosulfate standardized solution Additve : 0.05 mol/L iodine solution 1 mol/L hydrochloric acid (for acidification) Instruments & Electrodes Recommended automatic titrator More automated system! COM-1700S (High-end potentiometric titrator, stand-alone) COM-300A (Compact and simple titrator) COM-1700S + 2 additional burets B-1700-20 (*To automatically deliver iodine solution and hydrochloric acid) ・Platinum electrode PR-301 (P/N D231244-A) ・Reference electrode RE-201 (standard accessory) Additional burets ▲COM-1700S w/ additional burets Built-in Thermal Printer ▲COM-300A 2010.02 DATA No.D-8 Page 2 / 2 Condition parameters (example) Master file 1 1+2+3 *COM-1700 *automatically dispensing I2 and HCl3 Blank Master file 2 4+5+6 Measurement Condition 1 (Dispensing 10mL of I2) Condition 4 (Dispensing 10mL of I2) Method Disp Method Disp Buret No. 2 Buret No. 2 S-Timer 0 sec S-Timer 0 sec Disp Vol. 10 mL Disp Vol. 10 mL Condition 2 (Dispensing 5mL of HCl) Condition 5 (Dispensing 5mL of HCl) Method Disp Method Disp Buret No. 3 Buret No. 3 S-Timer 5 sec S-Timer 60 sec Disp Vol. 5 mL Disp Vol. 5 mL Condition 3 (Blank) Condition 6 (Dispensing 5mL of HCl) Method AUTO Method Auto Buret No. 1 Mode No. 8 Buret No. 1 Mode No. 8 Amp. No. 2 Pre Int 0 sec Amp No. 2 Pre Int 0 sec Meas Unit mV Del K 5 Meas Unit mV Del K 5 S-Timer 10 sec Del Sens 0 mV S-Timer 10 sec Del Sens 0 mV CP mL 0 mL Int Time 5 sec CP mL 0 mL Int Time 5 sec T-Timer 0 sec Int Sens 3 mV T-Timer 0 sec Int Sens 3 mV DP mL 0 mL Brt Speed 2 DP mL 0 mL Brt Speed 2 End Sens 300 Pulse 40 End Sens 100 Pulse 40 Over mL 0.3 mL Over mL 0 mL Max volume 20 mL Max Volume 20 mL Unit D Unit g/L Size 0 mL Size 1 mL Blank 0 Blank Result value of blank titration Molarity 0.1 Molarity 0.1 Factor Titre of the titrant Factor Titer of the titrant K 0 K 40.03 L 0 L 0 Formula D Formula (B-D) x F x M x K / S *When manually adding I2 and HCl, only condition 3 and 6 are used. For more information, please feel free to contact: Hiranuma Sangyo Co., Ltd. 1739 Motoyoshida-cho, Mito, Ibaraki 310-0836 Phone: +81-29-247-7343 / JAPAN Fax: +81-29-247-0381 URL http://www.hiranuma.com E-mail [email protected] ISO 9001 ISO 13485 Certified ISO 14001 Certified
© Copyright 2026 Paperzz