thrombi stat Step by Step Guide to Monitoring thrombi stat is extremely simple to set up and use and offers you precise and accurate INR testing results. Here we take you through a series of simple steps demonstrating how to correctly set up and prepare thrombi stat MC1 and MC4 so that you are ready to start testing patients. How to maintain high levels of QC with thrombi stat How to set up thrombi stat MC1 Step 1 Step 2 • • • • • Allow reagent vials and reconstitution fluid to come to room temperature. Leave on bench overnight. Switch on thrombi stat MC1. Allow to warm up for 15 minutes. • • While the thrombi stat MC1 is warming up reconstitute the required vials of reagent. Allow to stand for a minimum of 15 minutes. Mix thoroughly before use. Do not allow to settle. Step 3 Step 4 • • • • While the thrombi stat MC1 is warming up reconstitute the QC plasma by adding 0.5ml of purified water to each vial (4 shots from the reagent pipette). Allow to stand for 10 minutes. Swirl to mix. While the thrombi stat MC1 is warming up place the required cuvettes in the rack and dispense a ball into each cuvette. Step 5 Step 6 • • Dispense 125 µl of well mixed reagent into each cuvette. • Step 7 • • • Test the QC plasma. If the QC plasma is in range you are ready to test patients. If results are out of range re-test and check for mistakes in preparation. Insert one cuvette into the measurement cell of the thrombi stat MC1. Allow to warm for 60 seconds. How to take a reading with thrombi stat MC1 Step 1 Step 2 Collection of a good quality capillary sample is critical for the accurate measurement of INR. A warm hand and good flow to the puncture site are essential to collect a good capillary sample. • • • To warm the hand you can,• Wash the hand with warm water. • Apply a warm compress for several minutes. • Gently massage the finger from base to tip several times to bring the blood to the finger tip. While the thrombi stat MC1 is warming up reconstitute the required vials of reagent. Allow to stand for a minimum of 15 minutes. Mix thoroughly before use. Do not allow to settle. Step 3 • • • While the thrombi stat MC1 is warming up reconstitute the QC plasma by adding 0.5ml of purified water to each vial (4 shots from the reagent pipette). Allow to stand for 10 minutes. Swirl to mix. Step 4 Step 5 • • Transfer the blood sample immediately to the test cuvette. Once the clot forms, the thrombi stat MC1 displays the measured INR on the screen.
© Copyright 2026 Paperzz