Modified Method: Digestion and distillation of total cyanide by MICRO DIST and FIA Appendix A / 4.5.2: Modified Method Compilation [30 November 2000], Revision 2.1 Digestion and distillation of total cyanide in drinking and wastewaters using MICRO DIST and determination of cyanide by flow injection analysis Lachat Instruments QuikChem Method 10-204-00-1-X 1.0 SCOPE AND APPLICATION 1.1 This method is for the determination of total cyanide which is the sum of strong acid dissociable (SAD), weak acid dissociable (WAD), and free cyanide, in samples of drinking water and domestic and industrial wastewaters. The method is for use in EPA’s data gathering and monitoring programs associated with the Clean Water Act, Resource Conservation and Recovery Act, Comprehensive Environmental Response, Compensation and Liability Act, and Safe Drinking Water Act. 1.2 The applicable range is 0.005 (minimum level, ML) to 0.500 mg CN-/L. Higher concentrations can be determined by dilution of the original sample or by reducing the volume of the sample loop. In a single laboratory, the method detection limit (MDL) is 0.001 mg CN-/L. The MDL obtained from a pooled interlaboratory study, is 0.002 mg CN-/L. Sample throughput is 80 injections / h, or 45 s / sample. 1.3 Modifications to the distillation procedure beyond those expressly stated in this method are strictly prohibited. The individual laboratory is permitted to make modifications to the colorimetric determinative procedure described in this method provided that all quality control requirements specified in Sections 9 and 17.5 of the method are met. The laboratory must fully document any such changes for review, as specified in Section 9. 2.0 SUMMARY OF METHOD By means of a passive miniature distillation device, MICRO DIST, the cyanide in the samples is released by digesting and acidifying cyanide complexes, converting them to hydrocyanic acid (HCN). The cyanide ion is trapped in a 1.0 M sodium hydroxide absorbing solution which is diluted to 0.25 M solution during the distillation. By means of flow injection analysis, the 0.25 M NaOH distillate is converted to cyanogen chloride by reaction with chloramine-T, pyridine and barbituric acid to give a red-colored complex. The Modified Method: Digestion and distillation of total cyanide by MICRO DIST and FIA absorbance of this complex is measured at 570 nm by measuring the peak area resulting from the sample. The peak area is proportional to the concentration of the cyanide in the sample. The determinative step of this method is described here only so that a complete method is provided. Any approved continuous-flow method may be used for the determination of cyanide in the 0.25 M NaOH MICRO DIST distillates. NOTE: Other approved distillation procedures may use different concentrations of NaOH in the absorber or trapping solution. Because the determinative method’s sensitivity is pH dependent, it is important to ensure that the pH of the absorber solution from any distillation procedure is adjusted to match the pH of calibration standards specified in the approved determinative method being used. USEPA Reference Method “Methods for the Chemical Analysis of Water and Wastes”, EPA-600/4-79020, Revised March 1983 and 1979, Method 335.2 specifies that absorbance be measured at 578 nm using a batch spectrophotometer. USEPA Reference Method “Methods for the Determination of Inorganic Substances in Environmental Samples”, EPA-600/R-93-100, August 1993, Method 335.4 specifies that absorbance be measured at 570 nm. The latter wavelength is specified in this Modified Method because Method 335.4 uses an automated continuous flow-through detector similar to that used in the following flow injection method. 3.0 DEFINITIONS 3.1 CALIBRATION BLANK (CB) -- A volume of reagent water in the same matrix as the calibration standards, but without the analyte. 3.2 CALIBRATION STANDARD (CAL) -- A solution prepared from the primary dilution standard solution or stock standard solutions. The CAL solutions are used to calibrate the instrument response with respect to analyte concentration. 3.3 INSTRUMENT PERFORMANCE CHECK SOLUTION (IPC) -- A solution of one or more method analytes used to evaluate the performance of the instrument system with respect to a defined set of criteria. 3.4 LABORATORY SPIKED BLANK (LSB) or LABORATORY FORTIFIED BLANK (LFB) -- an aliquot of reagent water or other blank matrices to which known quantities of the method analytes are added in the laboratory. The LSB or LFB is analyzed exactly like a sample, and its purpose is to determine whether the methodology is in control and whether the laboratory is capable of making accurate and precise measurements. The LSB or LFB can be used as the INITIAL PRECISION AND RECOVERY (IPR) and ONGOING PRECISION AND RECOVERY (OPR) samples as needed. 3.5 LABORATORY REAGENT BLANK (LRB) -- An aliquot of reagent water or other blank matrices that is digested exactly as a sample including exposure to all glassware, equipment, and reagents that are used with other samples. The LRB is used to determine if method analytes or other interferences are present in the laboratory environment, the reagents, or the apparatus. 3.6 LINEAR CALIBRATION RANGE (LCR) -- The concentration range over which the instrument response is linear. 3.7 MATRIX SPIKE/MATRIX SPIKE DUPLICATE (MS/MSD) -- An aliquot of an environmental sample to which known quantities of the method analytes are added in the laboratory (MS) – the fortified sample can then be split to provide a duplicate (MSD). The MS and MSD are analyzed exactly like a sample, and their purpose is to determine whether the sample matrix contributes bias or imprecision to the analytical results. The background concentrations of the analytes in the sample matrix must be determined in a separate aliquot and the measured values or the MS and MSD corrected for background concentrations. Modified Method: Digestion and distillation of total cyanide by MICRO DIST and FIA 3.8 MATERIAL SAFETY DATA SHEET (MSDS) -- Written information provided by vendors concerning a chemical's toxicity, health hazards, physical properties, fire, and reactivity data including storage, spill, and handling precautions. 3.9 METHOD DETECTION LIMIT (MDL) -- The minimum concentration of an analyte that can be identified, measured and reported with 99% confidence that the analyte concentration is greater than zero. 3.10 QUALITY CONTROL SAMPLE (QCS)/ONGOING PRECISION AND RECOVERY SAMPLE (OPR) -- A solution of method analytes of known concentrations that is used to spike an aliquot of LRB. The QCS is obtained from a source external to the laboratory and different from the source of calibration standards. The OPR sample may be prepared from the standard source or external source (if the latter, the OPR and QCS are equivalent). 4.0 INTERFERENCES 4.1 Most non-volatile interferences are eliminated or minimized by the distillation procedure. Some of the known interferences are aldehydes, nitrate-nitrite, and oxidizing agents, such as chlorine, thiosulfate, and sulfide. Multiple interferences may require the analysis of a series of matrix spikes and duplicates (MS/MSDs) to verify the suitability of the chosen treatment. 4.2 4.3 Thiocyanate will interfere if present. Sulfides adversely affect the procedure by producing hydrogen sulfide during distillation. If a drop of the sample on lead acetate test paper indicates the presence of sulfide, treat 25 mL more than the stabilized sample (pH ≥ 12) than that required for the cyanide determination with powdered cadmium carbonate. Yellow cadmium sulfide precipitates if the sample contains sulfide. Repeat this operation until a drop of the treated sample solution does not darken the lead acetate test paper. Filter the solution through a dry filter paper into a dry beaker, and from the filtrate, measure the sample to be used for analysis. Avoid a large excess of cadmium and a long contact time in order to minimize a loss by complexation or occlusion of cyanide on the precipitated material. NOTE: If samples contain particulate that would be removed upon filtration, the samples must be filtered prior to treatment with cadmium carbonate. The collected particulate must be saved, and the filtrate then treated using the sulfide removal procedure described above. The collected particulate and treated filtrate must be recombined, homogenized, and included in the total cyanide distillation. 4.4 Studies have shown that sulfide concentrations of up to to 10 mg S2-/L in the distillate can be tolerated. However, when is it expected that hydrogen sulfide will be generated from the distilled sample during the distillation, every effort should be made to analyze distillates within 2 hours from the start of distillation. 4.5 If exceptionally high concentrations of sulfide (>>10 mg S2-/L) are generated by the decomposition of sulfur-containing compounds during distillation, they could significantly bias results, violating the QC criteria in Section 9 of this method. Should this situation occur, a MICRO DIST tube containg the lead cation, as described in other approved distillation methods, should be used. When using a sulfide removal procedure, make sure to initially analyze QC samples to ensure criteria are being met with the modified procedure. 4.6 High results may be obtained for samples that contain nitrate and/or nitrite. During the distillation nitrate and nitrite will form nitrous acid that will react with some organic compounds to form Modified Method: Digestion and distillation of total cyanide by MICRO DIST and FIA oximes. These oximes will decompose under test conditions to generate HCN. The interference of nitrate and nitrite is eliminated by pretreatment with sulfamic acid. 4.7 Oxidizing agents, such as residual chlorine, decompose most of the cyanides. Test a drop of the sample with potassium iodide (KI)-starch paper at time of collection; a blue color indicates the need for treatment. Add ascorbic acid, a few crystals (about 0.6 g each) at a time, until a drop of sample produces no color in the indicator paper; then add an additional 0.6 g of ascorbic acid for each liter of sample volume. Sodium arsenite has also been employed to remove oxidizing agents. 4.8 Method interferences may be caused by contaminants in the reagent water, reagents, and sample processing apparatus that bias response. 5.0 SAFETY 5.1 The toxicity or carcinogenicity of each reagent used in this method has not been fully established. Each chemical should be regarded as a potential health hazard and exposure should be as low as reasonably achievable. Cautions are included for known extremely hazardous materials. 5.2 Each laboratory is responsible for maintaining a current awareness file of OSHA regulations regarding the safe handling of the chemicals specified in this method. A reference file of Material Safety Data Sheets (MSDS) should be made available to all personnel involved in the chemical analysis. The preparation of a formal safety plan is also advisable. 5.3 All waste materials should be disposed of in a responsible manner, in accordance with federal, state, territory, local, and tribal regulations. 5.4 The following chemicals have the potential to be highly toxic or hazardous, for detailed explanations consult the MSDS. 5.4.1 5.4.2 5.4.3 5.4.4 Cyanide Hydrochloric acid Pyridine Sulfuric acid 6.0 EQUIPMENT AND SUPPLIES 6.1 6.2 Balance -- analytical, with a 0.0001 g resolution 6.3 Flow injection analysis equipment designed to deliver and react sample and reagents in the required order and ratios. Glassware -- Class A volumetric flasks and pipettes or polyethylene containers as required. Samples may be stored in plastic or glass. 6.3.1 6.3.2 6.3.3 6.3.4 6.3.5 6.4 Autosampler Multichannel proportioning pump Reaction unit or manifold Colorimetric detector Data system Special Apparatus 6.4.1 In-line Heating Unit Modified Method: Digestion and distillation of total cyanide by MICRO DIST and FIA 6.4.2 MICRO DIST Block and Cyanide-1 tubes. The MICRO DIST block is available from Zellweger Analytics, Lachat Instrument Division, Milwaukee, WI. as Lachat part number A17100. The Cyanide-1 tubes are available as Lachat part number A17001. 6.4.3 Filter paper, No. 2 7.0 REAGENTS AND STANDARDS 7.1 Preparation of Reagents Use reagent water, distilled or deionized, free of the analyte of interest, ASTM-Type II or equivalent, for all solutions. Degassing with helium: To prevent bubble formation, degas all solutions except the standards with helium. Use He at 140 kPa (20 lb/in2) through a helium degassing tube. Bubble He vigorously through the solution for one minute. Reagent 1. Carrier, 0.25 M Sodium Hydroxide In a 1 L plastic container dissolve 10.0 g NaOH [CASRN-1310-73-2] in 1.00 L or 1.00 kg water. Reagent 2. Phosphate Buffer, 0.71 M By Volume: In a 1 L volumetric flask, dissolve 97 g potassium phosphate, monobasic, anhydrous, (KH2PO4) [CASRN-7778-77-0] in approximately 800 mL water. Dilute to the mark and invert to mix. Prepare fresh monthly. By Weight: To a 1 L tared container add 97 g potassium phosphate, monobasic, anhydrous, (KH2PO4) and 975 g water. Stir or shake until dissolved. Prepare fresh monthly. Reagent 3. Chloramine-T Dissolve 2.0 g chloramine-T hydrate [CASRN-127-65-1] in 500 mL or 500 g water. Prepare fresh daily. Reagent 4. Pyridine-Barbituric Acid Reagent By Volume: In the fume hood, place 15.0 g barbituric acid [CASRN-67-52-7]in a 1 L beaker and add 100 mL water, rinsing down the sides of the beaker to wet the barbituric acid. Add 75 mL pyridine (C5H5N) [CASRN-110-86-1] while stirring and mix until the barbituric acid dissolves. Add 15 mL concentrated hydrochloric acid (12 M HCl) [CASRN-7647-01-0] and mix. Transfer to a 1 L volumetric flask, dilute to the mark, and invert to mix. Prepare fresh weekly. By Weight: In the fume hood, place 15.0 g barbituric acid in a tared 1 L container and add 100 g water, rinsing down the sides of the beaker to wet the barbituric acid. Add 73 g pyridine (C5H5N) with stirring and mix until the barbituric acid dissolves. Add 18 g concentrated hydrochloric acid (12 M HCl), then add an additional 825 g water and mix. Prepare fresh weekly. Reagent 5. MICRO DIST Distillation Releasing Agent This will make 200 mL of 7.11 M sulfuric acid / 0.79 M magnesium chloride. It is best to do this in the hood as HCl fumes will be released. Place a 500 mL beaker on a top-loading balance with a 400 g capacity or greater. Tare the balance. Into the tared 500 mL beaker place 110.8 g water. Then add and dissolve completely 32.2 g magnesium chloride hexahydrate (MgCl2.6H2O) [CASRN-7786-30-3] in this water. Use 99% A.C.S. reagent as any trace cyanide in this reagent will contribute to a distillation blank. Slowly add 139 g concentrated sulfuric acid in increments of 40 g at a time, swirling and allowing to cool. HCl fumes will be released. Use the purest grade you can find as any trace cyanide in this reagent will contribute to a distillation blank. CAUTION! Modified Method: Digestion and distillation of total cyanide by MICRO DIST and FIA When adding sulfuric acid to water, the solution will become very hot. NEVER ADD WATER TO CONCENTRATED ACID! Transfer the solution to the MICRO DIST automatic pipet container. Place the assembled and calibrated pipet cap on loosely and allow the solution to cool to room temperature in the hood. When the solution is at room temperature, screw the cap on tightly. Prime the pipet and the solution is ready to use. 7.2 Preparation of Standards Pre-calibration with non-distilled standards (see Section 7.2 – STANDARDS NOT TO BE DISTILLED WITH SAMPLES) is recommended to test instrument operation. However, final calibration must be performed with distilled standards (See Section 7.2 – STANDARDS TO BE DISTILLED WITH SAMPLES). Standard 1. Stock Standard 1250 mg CN-/L In a 1 L volumetric flask dissolve 2.0 g potassium hydroxide (KOH) [CASRN-1310-58-3] in approximately 500 mL water. Add 3.129 g potassium cyanide (KCN) [CASRN-151-50-8] . CAUTION: KCN IS HIGHLY TOXIC. AVOID INHALATION OF DUST OR CONTACT WITH THE SOLID OR SOLUTIONS. Dilute to the mark with water and invert three times. Prepare fresh weekly. STANDARDS TO BE DISTILLED WITH SAMPLES: Standard 2. Stock Standard 12.5 mg CN-/L in water In a 100 mL volumetric flask add 1.00 mL Standard 1 (1250 mg CN-/L). Dilute to the mark with water and invert to mix. Prepare fresh weekly. Table 1 These standards will be distilled Working Standards (Prepare Daily) Concentration mg CN-/L A 0.500 B 0.100 C 0.05 D 0.01 E 0.005 F 0.00 10.0 2.0 1.0 --- --- --- --- --- --- 5.0 2.5 --- Weight (g) of stock standard 2 diluted to final weight (~250 g) divided by factor below with water 10.0 2.0 1.00 --- --- --- Weight (g) of standard A diluted to final weight (~250 g) divided by factor below with water --- --- --- 5.0 2.5 --- Division Factor Divide exact weight of the standard by this factor to give final weight 0.04 0.008 0.004 0.02 0.01 --- By Volume Volume (mL) of stock standard 2 diluted to 250 mL with water Volume(mL) of standard A diluted to 250 mL with DI water By Weight STANDARDS NOT TO BE DISTILLED WITH SAMPLES: Blank in Distillation Matrix (0.25 M NaOH) By Volume: In a 2 L volumetric flask containing approximately 1200 mL water, add 20 g sodium hydroxide (NaOH). Dilute to the mark and invert to mix. Prepare the blank in a large single batch versus several smaller batches as slight changes in the 0.25 M diluent matrix can Modified Method: Digestion and distillation of total cyanide by MICRO DIST and FIA adversely affect the determined concentrations. The non distilled standards, blank, and carrier should all be prepared from this single diluent. By Weight: To a tared 2 L container, add 2 kg water and 20 g sodium hydroxide (NaOH). Invert to mix. Prepare the blank in a large single batch versus several smaller batches as slight changes in the 0.25 M diluent matrix can adversely affect the determined concentrations. The non distilled standards, blank, and carrier should all be prepared from this single diluent. Standard 2. Stock Standard 12.5 mg CN-/L in Distillation Matrix In a 100 mL volumetric flask add 1 mL of Standard 1 (1250 mg CN-/L). Dilute to the mark with 0.25 M NaOH and invert to mix. Prepare fresh weekly. These standards will not be distilled Working Standards (Prepare Daily) Concentration mg CN-/L A 0.500 B 0.100 C 0.05 D 0.01 E 0.005 F 0.00 Volume (mL) of standard 2 diluted to 250 mL with 0.25 M NaOH 10.0 2.0 1.0 --- --- --- Volume(mL) of standard A diluted to 250 mL with 0.25 M NaOH --- --- --- 5.0 2.5 --- Weight (g) of standard 2 diluted to final weight (~250 g) divide by factor below with 0.25 M NaOH 10.0 2.0 1.0 --- --- --- Weight (g) of standard A diluted to final weight (250 g) divide by factor below with 0.25 M NaOH --- --- --- 5.0 2.5 --- Division Factor Divide exact weight of the standard by this factor to give final weight 0.04 0.008 0.004 0.02 0.01 --- By Volume By Weight Modified Method: Digestion and distillation of total cyanide by MICRO DIST and FIA 8.0 SAMPLE COLLECTION, PRESERVATION AND STORAGE The following procedure is consistent with 40 CFR 136.3(c) Table II. When this method is used for EPA compliance monitoring, preservation techniques and holding times may not be altered by the laboratory. 8.1 Samples may be collected in polyethylene or glass bottles. All bottles must be thoroughly cleaned and rinsed with reagent water. Volume collected should be sufficient to insure a representative sample, allow for replicate analysis (if required), and minimize waste disposal. 8.2 Sulfides (S2-) may adversely affect the procedure by producing hydrogen sulfide during distillation. Sulfide also rapidly converts CN- to SCN-, especially at high pH, and the sample must be analyzed within 24 h of collection unless the sample is treated as follows. 8.2.1 If a drop of the sample on lead acetate test paper indicates the presence of sulfide, treat 25 mL more of the stabilized sample (pH ≥ 12) than that required for the cyanide determination with powdered cadmium carbonate. Yellow cadmium sulfide precipitates if the sample contains sulfide. Repeat this operation until a drop of the treated sample solution does not darken the lead acetate test paper. Filter the solution through a dry filter paper into a dry beaker, and from the filtrate, measure the sample to be used for analysis. Avoid a large excess of cadmium and a long contact time in order to minimize a loss by complexation or occlusion of cyanide on the precipitated material. The resulting filtrate must have a concentration of 10 mg S2-/L or less. NOTE: If samples contain particulate that would be removed upon filtration, the samples must be filtered prior to treatment with cadmium carbonate. The collected particulate must be saved, and the filtrate then treated using the sulfide removal procedure described above. The collected particulate and treated filtrate must be recombined, homogenized, and included in the total cyanide distillation. 8.3 Samples must be preserved with sodium hydroxide at a pH ≥ 12 and cooled to 4oC at the time of collection. Samples should be analyzed as soon as possible after collection. If storage is required, preserved samples should be maintained at 4oC and may be held for up to 14 days. 8.4 Oxidizing agents, such as residual chlorine, decompose most of the cyanides. Test a drop of the sample with potassium iodide (KI)-starch paper at time of collection; a blue color indicates the need for treatment. Add ascorbic acid, a few crystals (about 0.6 g each) at a time, until a drop of sample produces no color in the indicator paper. Sodium arsenite has also been employed to remove oxidizing agents. 9.0 Quality Control 9.1 Each laboratory that uses this method is required to operate a formal quality assurance program. The minimum requirements of this program consist of an initial demonstration of laboratory capability, ongoing analyses of standards and blanks as a test of continued performance, and analyses of matrix spike (MS) and matrix spike duplicate (MSD) samples to assess accuracy and precision. Laboratory performance is compared to established performance criteria to determine if the results of analyses meet the performance characteristics of the method. 9.1.1 The analyst shall make an initial demonstration of the ability to generate acceptable accuracy and precision with this method. This ability is established as described in Section 9.2. The laboratory must maintain performance records that define the quality of the data that are generated. See Section 17.5. Modified Method: Digestion and distillation of total cyanide by MICRO DIST and FIA 9.1.2 9.1.1.1 Each time a modification is made to the method, the analyst is required to repeat the procedure in Section 9.2. If the change will affect the detection limit of the method, the laboratory is required to demonstrate that the MDL (40 CFR Part 136, Appendix B) is lower than the MDL for that analyte in this method, or one-third the regulatory compliance level, whichever is higher. If the change will affect calibration, the analyst must recalibrate the equipment according to Section 10. 9.1.1.2 Alternative determinative techniques and changes that degrade method performance are not allowed. If an analytical technique other than the techniques specified in this method is used, that technique must have specificity equal to or better than the specificity of the techniques in this method for the analytes of interest. 9.1.1.3 The laboratory is required to maintain records of modifications made to this method. These records include the following, at a minimum: 9.1.1.3.1 The names, titles, addresses and telephone numbers of the analyst(s) who performed the analyses and modification, and of the quality control officer who witnessed and will verify the analyses and modification. 9.1.1.3.2 A listing of pollutant measured (total cyanide) 9.1.1.3.3 A narrative stating reason(s) for the modification. 9.1.1.3.4 Results from all quality control (QC) tests comparing the modified method to this method, including: (a) Calibration (Section 10) (b) Calibration verification (Section 9.3.3 and 9.7) (c) Initial precision and recovery (Section 9.2.2) (d) Analysis of blanks (Section 9.3.4) (e) Accuracy assessment (Section 9.3 and 9.4) (f) Ongoing precision and recovery (Section 9.6) 9.1.1.3.5 Data that will allow an independent reviewer to validate each determination by tracing the instrument output (weight or other signal) to the final result. This data will include: (a) Sample numbers and other identifiers. (b) Extraction dates. (c) Analysis dates and times. (d) Analysis sequence/run chronology. (e) Sample weight or volume. (f) Extract volume. (g) Make and model of analytical balance and weights traceable to NIST. (h) Copies of logbooks, printer tapes and other recordings of raw data. (i) Data system outputs, and other data to link the raw data to the results reported. Analyses of matrix spike and matrix spike duplicate samples are required to demonstrate method accuracy and precision and to monitor matrix interferences (interferences caused Modified Method: Digestion and distillation of total cyanide by MICRO DIST and FIA by the sample matrix). The procedure and QC criteria for spiking are described in Section 9.4. 9.2 9.1.3 Analyses of laboratory blanks are required to demonstrate freedom from contamination. The procedure and criteria for analysis of a blank are described in Section 9.3.4. 9.1.4 The laboratory shall, on an ongoing basis, demonstrate through calibration verification and analysis of the ongoing precision and recovery sample that the analysis system is in control. These procedures are described in Sections 9.6 and 9.7, respectively. 9.1.5 The laboratory should maintain records to define the quality of data that is generated. Development of accuracy statements is described in Sections 9.3.2. Initial demonstration of laboratory performance. 9.2.1 Method Detection Limit (MDL)— To determine MDL values, take seven replicate aliquots of the fortified reagent water and process through the entire analytical method. The fortifying concentration should be 1 to 5 times the MDL (see 40 CFR Part 136, Appendix B for more information). Perform all calculations defined in the method and report the concentration values in the appropriate units. Calculate the MDL as follows: MDL = tS where t is the Student's t value for a 99% confidence level and 7-1 degrees of freedom; t = 3.14 for seven replicates, and S is the standard deviation of the replicates. MDLs should be determined every 6 months, when a new operator begins work, or whenever there is a significant change in the background or instrument response. 9.2.2 Initial precision and recovery (IPR)—To establish the ability to generate acceptable precision and accuracy, the analyst shall perform the following operations: 9.2.2.1 Measure four samples of prepared at 0.2 mg-CN/L according to the procedure beginning in Section 11. 9.2.2.2 Using the results of the set of four analyses, compute the average percent recovery (x) and the standard deviation of the percent recovery (s) for cyanide. Use the following equation for calculation of the standard deviation of the percent recovery: Equation 1 s = Σx2 – (Σx)2 n n–1 where: n = Number of samples x = Concentration in each sample Modified Method: Digestion and distillation of total cyanide by MICRO DIST and FIA 9.2.2.3 9.3 Compare (s) and (x) with the corresponding limits for initial precision and recovery in Table 2. If (s) and (x) meet the acceptance criteria, system performance is acceptable and analysis of samples may begin. If, however, (s) exceeds the precision limit or (x) falls outside the range for recovery, system performance is unacceptable. In this event, correct the problem and repeat the test. Assessing laboratory performance: 9.3.1 Ongoing Precision and Recovery Standard (OPR). The laboratory must analyze at least one LFB spiked at a level of 0.2 mg-CN/L with each batch of samples. Calculate accuracy as percent recovery (Sect. 9.4.3). If the recovery of any analyte falls outside the required control limits [see Section 17.5] the method is judged out of control, and the source of the problem must be identified and resolved before continuing analyses. 9.3.2 When sufficient internal performance data become available (usually a minimum of 2030 analyses), optional control limits can be developed from the percent mean recovery (x) and the standard deviation (S) of the mean recovery. These data can be used to establish the upper and lower control limits as follows: UPPER CONTROL LIMIT = x+3S LOWER CONTROL LIMIT = x-3S The optional control limits must be equal to or better than the required control limits specified in Section 17.5. After each five to 10 new recovery measurements, new control limits can be calculated using only the most recent 20-30 data points. Also, the standard deviation (S) data should be used to establish an ongoing precision statement for the level of concentrations included in the LFB. These data must be kept on file and be available for review. 9.3.3 Instrument Performance Check Solution (IPC). For all determinations, the laboratory must analyze the IPC (a midrange check standard) and a calibration blank immediately following daily calibration, after every 10th sample (or more frequently, if required), and at the end of the sample run. Analysis of the IPC solution and calibration blank immediately following calibration must verify that the instrument is still within ± 10 percent. If the calibration cannot be verified with the specified limits, reanalyze the IPC solution. If the second analysis of the IPC solution confirms calibration to be outside the limits, sample analysis must be discontinued and the cause determined. In the case of drift, the instrument should be recalibrated. All samples following the last acceptable IPC solution must be reanalyzed. The analysis data of the calibration blank and IPC solution must be kept on file with the sample analyses data. 9.3.4 Laboratory reagent blanks (LRB)—laboratory reagent water blanks are analyzed to demonstrate freedom from contamination. 9.3.4.1 Measure a laboratory reagent water blank initially (i.e., with the tests in Section 9.2) and with each analytical batch. The blank must be subject to the same procedural steps as a sample. 9.3.4.2 If material is detected in the blank at a concentration greater than the Minimum Level (Section 1.2), analysis of samples is stopped until the source of contamination is eliminated and a blank shows no evidence of contamination. All samples must be associated with an uncontaminated Modified Method: Digestion and distillation of total cyanide by MICRO DIST and FIA method blank before the results may be reported for regulatory compliance purposes. 9.4 Matrix spikes/Matrix Spike Duplicates (MS/MSD)—The laboratory must spike, in duplicate, a minimum of 10 percent of all samples (one sample in each batch of 10 samples) from a given sampling site or, if for compliance monitoring, from a given discharge. The two sample aliquots shall be spiked with a cyanide standards 9.4.1 The concentration of the spike in the sample shall be determined as follows: 9.4.1.1 For compliance monitoring, the concentration of cyanide in the sample being checked against a regulatory concentration limit, and the spiking level shall be at that limit or one to five times higher than the background concentration of the sample (determined in Section 9.3.1), whichever concentration is higher. 9.4.1.2 If the concentration of cyanide in a sample is not being checked against a limit, the spike shall be at the concentration of the precision and recovery standard (Section 9.6) or at one to five times higher than the background concentration, whichever concentration is higher. 9.4.2 Analyze one sample aliquot out of each set of 10 samples from each site or discharge according to the procedure beginning in Section 11 to determine the background concentration (B) of cyanide. 9.4.3 Calculate the percent recovery (P) of cyanide in each aliquot using the following equation: Equation 2 P = 100 (A – B) T where: A = Measured concentration of analyte after spiking B = Measured background concentration of cyanide T = True concentration of the spike 9.4.4 Compare the percent recovery of cyanide with the corresponding QC acceptance criteria in Section 17.5. 9.4.4.1 If the results of the spike fail the acceptance criteria, and the recovery of the QC standard in the ongoing precision and recovery test (Section 9.6) for the analytical batch is within the acceptance criteria in Section 17.5, an interference is present. In this case, the result may not be reported for regulatory compliance purposes and the analyst must assess the potential cause for the interference. If the interference is attributable to sampling, the site or discharge should be resampled. If the interference is attributable to a method deficiency, the analyst must modify the method, repeat the tests required in Section 9.1.2, and repeat the analysis of the sample and the MS/MSD. Modified Method: Digestion and distillation of total cyanide by MICRO DIST and FIA 9.4.4.2 9.4.5 If the results of both the spike and the ongoing precision and recovery test fail the acceptance criteria, the analytical system is judged to be out of control, and the problem shall be identified and corrected, and the sample reanalyzed. Compute the relative percent difference (RPD) between the two results (not between the two recoveries) using the following equation: Equation 3 D1 – D2 RPD = x 100 (D1 + D2)/2 where: D1 = concentration of cyanide in the sample D2 = concentration of cyanide in the second (duplicate) sample 9.4.6 The relative percent difference for duplicates shall meet the acceptance criteria in Section 17.5. If the criteria are not met, the analytical system is judged to be out of control, and the problem must be immediately identified and corrected, and the analytical batch reanalyzed. As part of the QC program for the laboratory, method precision and accuracy for samples should be assessed and records should be maintained. After the analysis of five spiked samples in which the recovery passes the test in Section 9.3.4, compute the average percent recovery (Pa) and the standard deviation of the percent recovery (sp). Express the accuracy assessment as a percent recovery interval from Pa – 2 sp to Pa + 2 sp. For example, if Pa = 90 percent and sp = 10 percent for five analyses of cyanide, the accuracy interval is expressed as 70 – 110 percent. Update the accuracy assessment on a regular basis (e.g., after each five to 10 new accuracy measurements). 9.5 Depending upon specific program requirements, field replicates and field spikes of the analytes of interest into samples may be required to assess the precision and accuracy of the sampling and sample transporting techniques. 9.6 Ongoing precision and recovery (OPR)—To demonstrate that the analysis system is in control, and acceptable precision and accuracy is being maintained with each analytical batch, the analyst shall perform the following operations: 9.6.1 Prepare a precision and recovery standard with each analytical batch according to the procedure beginning in Section 11. 9.6.2 Compare the concentration with the limits for ongoing precision and recovery (OPR) in Section 17.5. If the concentration is in the range specified, the process is in control and analysis of blanks and samples may proceed. If, however, the concentration is not in the specified range, the analytical process is not in control. In this event, correct the problem, retest the analytical batch, and repeat the ongoing precision and recovery test. Modified Method: Digestion and distillation of total cyanide by MICRO DIST and FIA The laboratory should add results that pass the specification in Section 9.6.2 to IPR and previous OPR data and update QC charts to form a graphic representation of continued laboratory performance. The laboratory should also develop a statement of laboratory data quality for each analyte by calculating the average percent recovery and the standard deviation of the percent recovery (sr). Express the accuracy as a recovery interval from R – 2sr to R + 2sr. For example, if R = 95 percent and sr = 5 percent, the accuracy is 85 percent to 105 percent. 9.6.3 It is suggested that the laboratory obtain a quality control sample (QCS) from a source different from the source for the cyanide standard used routinely in this method, and that the QCS be used for verification of the concentrations of cyanide. 9.7 The standards used for initial precision and recovery (Section 9.2.2), matrix spikes (Section 9.4), and ongoing precision and recovery (Section 9.6) should be identical so that the most precise results will be obtained. 9.8 Depending upon specific program requirements, field replicates and field spikes of the analytes of interest into samples may be required to assess the precision and accuracy of the sampling and sample transporting techniques. 10.0 CALIBRATION AND STANDARDIZATION 10.1 Prepare a series of at least 3 standards, covering the desired range, and a blank by diluting suitable volumes of standard solution. 10.2 10.3 Calibrate the instrument, as described in section 11. 10.4 After the calibration has established, it must be verified by the analysis of a suitable quality control sample (QCS). If measurements exceed +/-10% of the established QCS value, the analysis should be terminated and the instrument recalibrated. The new calibration must be verified before continuing analysis. Periodic reanalysis of the QCS is recommended as a continuing calibration check. Prepare standard curve by plotting instrument response against concentration values. A calibration curve may be fitted to the calibration solution concentration/response data using the computer. Acceptance or control limits should be established using the difference between the measured value of the calibration solution and the "true value" concentration. 11.0 PROCEDURE 11.1 Sample Pretreatment 11.1.1 Sample Preparation at collection or prior to distillation by MICRO DIST 11.1.1.1 When this method is used for EPA compliance monitoring, preservation techniques and holding times may not be altered by the laboratory without approval of a variance by EPA. Modified Method: Digestion and distillation of total cyanide by MICRO DIST and FIA 11.1.1.2 11.1.1.3 11.1.1.4 11.1.1.5 Sulfides may adversely affect the procedure by producing hydrogen sulfide during distillation. If a drop of the sample on lead acetate test paper indicates the presence of sulfide, treat 25 mL more of the stabilized sample (pH ≥ 12) than that required for the cyanide determination with powdered cadmium carbonate. Yellow cadmium sulfide precipitates if the sample contains sulfide. Repeat this operation until a drop of the treated sample solution does not darken the lead acetate test paper. Filter the solution through a dry filter paper into a dry beaker, and from the filtrate, measure the sample to be used for analysis. Avoid a large excess of cadmium and a long contact time in order to minimize a loss by complexation or occlusion of cyanide on the precipitated material. NOTE: If samples contain particulate that would be removed upon filtration, the samples must be filtered prior to treatment with cadmium carbonate. The collected particulate must be saved, and the filtrate then treated using the sulfide removal procedure above. The collected particulate and treated filtrate must be recombined, homogenized, and included in the total cyanide distillation. High results may be obtained for samples that contain nitrate and/or nitrite. During the distillation nitrate and nitrite will form nitrous acid that will react with some organic compounds to form oximes. These oximes will decompose under test conditions to generate HCN. The interference of nitrate and nitrite is eliminated by pretreatment with sulfamic acid. Oxidizing agents, such as residual chlorine, decompose most of the cyanides. Test a drop of the sample with potassium iodide (KI)-starch paper at time of collection; a blue color indicates the need for treatment. Add ascorbic acid, a few crystals (about 0.6 g each) at a time, until a drop of sample produces no color in the indicator paper. Sodium arsenite has also been employed to remove oxidizing agents. If the sample volume originally collected is much greater than the 6.0 mL sample distilled, and the original sample contains suspended solids, homogenize the sample (after preservation with NaOH to pH > 12) so that the 6.0 mL sample distilled contains the same concentration of suspended solids as in the original sample. 11.1.2 Sample and Standard Distillation by MICRO DIST (MICRO DIST Method Cyanide-1) A summary of the procedure for the digestion and distillation of total cyanide is given below. The user should read and be familiar with the MICRO DIST Reference and Methods Manual supplied with the MICRO DIST system. The figure references in the following procedure refer to figures in this manual. In the following procedure, D and M refer to the marks on the ends of the collector tube. D means ‘distillation’, or ‘discarded’ end and M means ‘measuring’ end as shown below: Modified Method: Digestion and distillation of total cyanide by MICRO DIST and FIA Procedure: 1. 2. 3. 4. 5. 6. 7. 8. 9. 10. Set the MICRO DIST block temperature controller to 120°C. Allow the block to warm up. This will take about 40 min. With the M end up, put as many collector tubes as you have samples into the collector tube rack. Put as many sample tubes into the sample tube rack as you have samples to distill; up to 21 for one block. Place 6.0 mL of sample or standard into each sample tube with an automatic pipet . In this method, the standards is distilled with the samples. In order to both release the free cyanide as HCN(g) and to digest complexed cyanides during the digestion/distillation, first prime the re-pipettor several times into a waste container. Then, add 0.75 mL of 7.11 M sulfuric acid / 0.79 M magnesium chloride solution to the sample tube using the automatic pipet. Immediately push the D end of a Cyanide-1 collector tube over the open end of each sample tube to start the seal. Place the assembly in the press, putting the sample tube through the hole in the white base. Before pressing the user should grip the collector tube firmly at the breakaway point to keep the tube from shifting during the pressing procedure. The pressing motion should be a smooth constant pressure which is just enough to slide the sample tube inside the collector tube. A jerky, forced motion may cause added strain to the tube and could potentially crack it. Press down on the handle until the stop ring on the sample tube hits the D end of the collector tube. Put on the heat-resistant gloves. Push the sample tube and D end of each tube all the way into the preheated block so that the collector tube stop ring touches the block. Placing 21 tubes should take less than one minute. Set the timer for 30 min. When 30 min is up, put on the heat-resistant gloves. Remove the first tube from the block and immediately pull off its sample tube using a downward, twisting motion as opposed to a sideways motion. You must pull off the sample tube within 4 s of removing it from the block or suck-back of the sample will occur. Drop the sample tube and the hot solution left in it into a waste bucket reserved for this purpose. Modified Method: Digestion and distillation of total cyanide by MICRO DIST and FIA 11. 12. 13. 14. 15. 16. 17. 11.2 Invert each collector tube and place it into the collector tube rack, now with the D end up. It should take less than two minutes to pull and separate all 21 tubes. Allow tubes to cool for at least 10 min. For each collector tube, hold the tube horizontally and rinse its walls with the distillate in order to homogenize it. Slowly roll the distillate around in the tube to gather all droplets clinging to the tube walls into the bulk of the distillate. Then, slowly return the collector tube to an upright position so that the D end is up. Stubborn drops will often fall into the M end when the tube is flicked with your finger. With the D end still up, break the collector tube in half by pulling the D end hard towards yourself to break it, then twisting and tearing off the D end. Discard the D end. In the remaining M end of the collector tube, dilute to the 6.0 mL mark with reagent water. This results in the original sample volume, but now in 0.25 M NaOH. Push the second cap onto the open end of the tube (Figure 11) as far as it will go. Shake the tube with a gentle whipping motion to mix in diluent water (Figure 12). Do not invert the sample. With the M end down, place the tube into the collector tube rack (Figure 5). Determine the solution in the tube with the flow injection method described in this method or other approved continuous flow method. Using the following flow injection analysis method, there is enough distillate for two 25 s sample periods. Samples should be determined as quickly after analysis as possible. If a batch of samples is being analyzed (so that immediate analysis every sample is not possible), seal both ends of the tube with Parafilm until analysis can be performed. FLOW INJECTION ANALYZER START-UP PROCEDURE 11.2.1 Prepare reagent and standards as described in section 7. 11.2.2 Set up manifold as shown in section 17. 11.2.3 Input data system parameters as in section 17. 11.2.4 Pump reagents and allow the system to equilibrate. Allow 15 minutes for heating unit to warm up to 60°C. 11.2.5 Place samples and/or standards in the autosampler. Input the information required by the data system, such as concentration, replicates and QC scheme. 11.2.6 Calibrate the instrument by injecting the standards. 11.3 System Notes 11.3.1 If sample concentrations are greater than the high standard, the distilled sample should be diluted with 0.25 M sodium hydroxide (NaOH) diluent. When an automated diluter is used, 0.25 M NaOH should be used as the diluent. Do not dilute distilled samples or standards with water as this will cause a large positive matrix effect. 11.3.2 For analyzing distilled samples, prepare a standard curve by plotting the peak area of distilled standards against concentration values. Compute concentrations by comparing sample peak area with the standard curve. 12.0 DATA ANALYSIS AND CALCULATIONS 12.1 The data system will then prepare a calibration curve by plotting instrument response against standard concentration. Sample concentration is calculated from the linear regression equation: Modified Method: Digestion and distillation of total cyanide by MICRO DIST and FIA Concentration = (Slope) Instrument Response + Intercept. 12.2 Report only those values that fall between the lowest and the highest calibration standards. Samples exceeding the highest standard should be diluted and reanalyzed. 12.3 Report results in mg /L. 13.0 METHOD PERFORMANCE 13.1 Typical method performance data are presented as support data in section 17. 14.0 POLLUTION PREVENTION 14.1 Pollution prevention encompasses any technique that reduces or eliminates the quantity or toxicity of waste at the point of generation. Numerous opportunities for pollution prevention exist in laboratory operation. Whenever feasible, laboratory personnel should use pollution prevention techniques to address their waste generation. When wastes cannot be feasibly reduced at the source, the USEPA recommends recycling as the next best option. 14.2 The quantity of chemicals purchased should be based on expected usage during its shelf life and disposal cost of unused material. Actual reagent preparation volumes should reflect anticipated usage and reagent stability. 14.3 All waste materials should be disposed of in a responsible manner, in accordance with federal, state, territory, local, and tribal regulations. 14.4 For information about pollution prevention that may be applicable to laboratories and research institutions, consult "Less is Better: Laboratory Chemical Management for Waste Reduction," available from the American Chemical Society's Department of Government Regulations and Science Policy, 115 16th Street N. W., Washington D. C. 20036, (202) 872-4477. 15.0 WASTE MANAGEMENT 15.1 The Environmental Protection Agency (USEPA) requires that laboratory waste management practice be conducted consistent with all applicable rules and regulations. Excess reagents, samples and method process wastes should be characterized and disposed of in an acceptable manner. The agency urges laboratories to protect the air, water, and land by minimizing and controlling all releases from hoods, and bench operations, complying with the letter and spirit of any waster discharge permit and regulations, and by complying with all solid and hazardous waste regulations, and by complying with all solid and hazardous waste regulations, particularly the hazardous waste identification rules and land disposal restrictions. For further information on waste management consult the "Waste Management Manual for Laboratory Personnel", available from the American Chemical Society at the address listed in Sect. 14.3. 16.0 REFERENCES 16.1 U.S. Environmental Protection Agency, Methods for Chemical Analysis of Water and Wastes, EPA-600/4-79-020, Revised March 1983, Method 335.4 16.2 Guideline and Format for EMSL-Cincinnati Methods. EPA-600/8-83-020, August 1983. Modified Method: Digestion and distillation of total cyanide by MICRO DIST and FIA 16.3 Standard Methods For the Examination of Water and Wastewater, 20th ed., AWWA/WEF/APHA (1999), Method 4500-CN--N, p4-50. 17.0 TABLE, DIAGRAMS, FLOWCHARTS, AND VALIDATION DATA 17.1 Flow Injection Analysis Manifold diagram CARRIER is 0.25 M sodium hydroxide solution (Reagent 1). All manifold tubing is 0.8 mm (0.030 in) i.d. This is 5.2 uL/cm. 4.5 is 70 cm of tubing on a 4.5 cm coil support APPARATUS: An injection valve, a 10 mm path length flow cell , and a colorimetric detector module is shows 650 cm of tubing wrapped around the heater block at the specified required. The temperature. Note 1: 2 meter back pressure loop, 0.52 mm i.d. 17.2 Typical parameters for Lachat Instruments QuikChem AE Flow Injection Analyzer Sample throughput: 80 samples/h, 45 s/sample Pump speed: 35 Cycle Period: 45 s Inject to Start of Peak Period: 42 s Modified Method: Digestion and distillation of total cyanide by MICRO DIST and FIA 17.3 Typical parameters parameters for Lachat Instruments QC 8000 Flow Injection Analyzer Sample throughput: 80 samples/h, 45 s/sample Pump speed: 35 Cycle Period: 45 s Analyte Data: Concentration Units: mg CN-/L Peak Base Width: 39 s % Width Tolerance: 100 % Threshold: 25000 Inject to Peak Start: 42 s Chemistry: Direct Calibration Data: Levels Concentration mg/L 1 2 3 4 5 6 0.500 0.100 0.050 0.010 0.005 0.000 Calibration Fit Type: 1st Order Polynomial Calibration Rep. Handling: Average Weighting Method: 1/X Sampler Timing: Min. Probe in Wash Period: 14 s Probe in Sample Period: 20 s Valve Timing: Load Time: 0.0 s Load Period: 20 s Inject Period: 25 s Modified Method: Digestion and distillation of total cyanide by MICRO DIST and FIA 17.4 Typical data for Lachat Instruments QC 8000 Flow Injection Analyzer Calibration Graph and Statistics Level Area mg CN-/L Determined Rep 1 Rep 2 Replica STD Replica %RSD % residual 1 31230856 0.5 0.503 3120856 31241620 6214.6 0.0 -0.6 2 615006 0.1 0.099 6150060 6158613 4937.8 0.1 1.0 3 295031 0.05 0.0475 2950321 2972790 12972.8 0.4 5.0 4 636723 0.01 0.0103 636723 659544 13175.8 2.0 -2.5 5 333965 0.005 0.0054 333965 356110 12784.4 3.7 -7.5 6 0 0 0 0 0.0 0.0 0.0 --- Method Detection Limit using Distilled Standards Modified Method: Digestion and distillation of total cyanide by MICRO DIST and FIA 17.5 Method Performance The method is required to perform according to the following QC Acceptance Criteria: Acceptance Criterion Section NPDES Limit (%) NPDWR Limit (%) 32 65-129 26 66-118 62-132 35 63-121 29 62-132 63-121 9.2.2 Initial precision and recovery Precision (s) Recovery (X) Matrix spike/matrix spike duplicate Recovery RPD 9.2.2.2 9.2.2.2 9.3 9.3.4 9.3.5 9.6 Ongoing precision and recovery Recovery 9.2.3
© Copyright 2026 Paperzz