Data sheet - nke INSTRUMENTATION

Reference: STPS-UK.doc
Revision: 2
The STPS measures and records the temperature, the conductivity (the salinity is calculated) and the pressure.
There are 3 versions in various depth ranges.
STPS xxx-SI: Station version (for stationary measurements), Inductive communication (via an Inductive Data Pencil).
STPS xxx-PI: Profil version (reduced temperature sensor response time, for dynamic measurements), Inductive
communication.
STPS xxx-PR: Profil version, Radio communication (via a Radio Data Pencil or a Concentrator).
This version has been developed in cooperation with Ifremer.
xxx: depth range 10m, 30m, 100m or 300m.
The STPS can be equiped with a local chlorination system to protect the sensors against biofouling.
Easy to use
-its transmission data interface (inductive or radio) allows dialog with a PC in hardness environment,
using an inductive or radio (around 10m range) Data Pencil interface,
-a friendly software is run on PC to program, collect and display data.
Strongly built
-no external connector,
-sensors easy to maintain thanks to the protection unscrewing,
-dialog with STPS is possible under water (inductive version),
-lithium battery offering several years life.
Examples of STPS using:
-water quality monitoring in oyster zones,
-salinity measurements for authentification of hydrodynamical models...
DESIGNATION
ORDER CODE
Depth
Temperature
Conductivity
Salinity
Autonomy
SI version
PI version
PR version
range
accuracy
resolution
range
accuracy
resolution
response time
range
accuracy
resolution
range
accuracy
resolution
Time
Automatic starting
Sample rate
memory
energy
STPS 10
STPS 30
STPS 100
STPS 300
60-07-221
------------------------------10 m
3 cm
0.24 cm
60-07-232
60-07-238
----------------30 m
9 cm
0.72 cm
60-07-233
60-07-239
60-07-234
100 m
30 cm
2.4 cm
------------60-07-267
60-07-222
300 m
90 cm
7.2 cm
-5 +35°C
<0.05°C
0.001°C at 10°C
Station: <3s at 63% - Profiler: <0.5s at 63%
0 to 70 mS/cm
<0.05 mS/cm in the range 0 to 60mS/cm
0.0012mS/cm
0 to 2000µS/cm
5µS/cm
0.04µS/cm
2 to 42°/ °° at 20°C
+/-0.1°/ °°
0.0011°/ °°
0 to 1.137°/ °° at 20°C
<0.002°/ °°
0.0001°/ °°
Internal clock with calendar (+/-1mn/month)
Programmable level trigger pressure
Programmable from 1 second to 99 hours
1 Mb, 4 years recording at 10mn
>4 years recording at 10mn (replaceable lithium batteries)
Mechanical characteristics
size 217 mm x 30 mm
weight in the air 180 gr
DISTRIBUTOR
Rue Gutenberg Z.I. Kérandré
56700 HENNEBONT FRANCE
Tel: (33) 02 97 36 10 12
Fax : (33) 02 97 36 55 17
[email protected]
www.nke-instrumentation.com