Compound analysis of limestone prepared as pressed powder pellets Epsilon 3X Introduction This data sheet investigates the capability of the Epsilon 3X, a bench top energy dispersive X-ray fluorescence spectrometer (EDXRF), as an analytical tool for a complete analysis of the quality and composition of limestone. Application background Limestone is the principal source of CaCO3, which is used as raw material in a wide range of applications. It is used as an additive in steel making, as main component for cement and, when purified, it even can be used as supplement for calciumrich animal feed or pharmaceutical fillers. Al2O3, SiO2, MgO and Fe2O3 are further considered as the main compounds of limestone, and are important in the use of limestone. The grade and application of the raw material is determined by the ratio between these main compounds, as well as the presence of other minor and trace compounds, like MnO, P2O5 and Pb. In this data sheet, the Epsilon 3X was used for a full and accurate analysis of limestone. Instrumentation Measurements were performed using a PANalytical Epsilon 3X EDXRF spectrometer, equipped with a 50 kV silver anode X-ray tube, 6 beam filters, a helium purge facility, a high-resolution silicon drift detector, a sample spinner and a 10-position removable sample changer. Sample preparation Pressed pellets were made from 11 commercially available limestone standards. The pellets were made by mixing 12 g of limestone powder with 3 g of Ultrawax binder. Measurement procedure Three measurement conditions were used to optimize the excitation of a group of elements (see Table 1). The total measurement time per sample was 10 minutes. Figure 1 shows an example of an XRF spectrum of a limestone standard. Table 1. Three measurement conditions Compounds kV μA Medium Filter Meas. time (s) Figure 1. A XRF spectrum of a limestone standard, obtained using the first measurement condition Calibration results The calibration was set up using 10 certified reference materials (CRM) and 1 standard kept aside for the precision measurements. Figure 2 shows a calibration graph acquired for CaO, using the conditions listed in Table 1. In Table 2 the calibration results are summarized for all compounds, with the corresponding correlation coefficients and RMS values (1 sigma). The calibrations show high correlations for all compounds of limestone. Also a calibration for traces of lead (Pb) was obtained but with limited quality because only three standards contain certified concentrations of Pb. Table 2.Calibration details (* RMS: The more accurate calibrations have the smaller RMS values.) Compounds Contration range (wt %) Correlation coefficient RMS* (wt %) CaO 29.95 -55.59 0.9998 0.226 Al2O3 0.021 - 3.200 0.9977 0.081 Fe2O3 0.018 - 1.250 0.9986 0.026 MgO 0.110 -17.500 0.9996 0.270 SiO2 0.120 - 9.500 0.9930 0.430 SrO 0.008 - 0.280 0.9999 0.002 K 2O 0.003 - 0.960 0.9998 0.008 MnO 0.0021 - 0.0420 0.9967 0.001 Na2O, MgO, Al2O3, SiO2, P2O5, SO3 4.5 1000 Helium None 420 Na2O 0.002 - 0.130 0.9002 0.023 P2O5 0.001 - 0.048 0.9405 0.007 CaO, K2O, MnO, TiO2 14 250 Helium Al-Thin 120 SO3 0.0170 - 0.1970 0.9854 0.009 TiO2 0.002 - 0.150 0.9991 0.002 SrO, Pb, Fe2O3 50 180 Air Ag 60 Pb (ppm) 0.9405 1.7 0.7 -15 CaO concentration (wt %) 27.0 30.0 36.0 33.0 39.0 CaO concentration (wt %) 51.5 51 50.5 Precision and accuracy To test the precision and accuracy of the method, the certified reference standard SX35-14 (Dillinger Hütte Laboratory) was measured 20 consecutive times. The measurement was repeated using 2 and 5 minutes as total measurements times, to analyze the influence of time on the precision. The total time for each measurement was divided over the three conditions according to Table 3, but all measurements were done using the calibration setup for 10 minutes. The results are summarized in Table 4, and a graphical representation of the 10 and 2 minute measurements for the CaO concentration is illustrated in Figure 3. The measured averages are close to the certified values, which show the accuracy of the method. The RMS values for shorter measurement times are slightly increased; however in all cases the repeatability is still acceptable. This is illustrated in Figure 3 with the dashed lines, showing 2 times RMS (2σ, 95% probability) for the different measurement times. 5 10 Table 3. Overview of the time used per condition, when the total measurement was shortened in the precision measurements. Total measurement time Compound 10 min 5 min 2 min Na2O, MgO, Al2O3, SiO2, P2O5, SO3 7 min 3 min 1 min CaO, K2O, MnO, TiO2 2 min 1 min 30 sec SrO, Pb, Fe2O3 1 min 1 min 30 sec 15 20 Measurement point Figure 3. The precision study of the CaO concentration in the SX35-14 standard performed with 10 and 2 min. measurement times. The dashed lines illustrate 2 times RMS (2σ, 95% probability) of each measurement time (10 min bleu, 2 min red). Table 4. Summary of the results of the repeatability measurements, obtained for 3 different measurement times. The certified values of the limestone standard SX35-14 are included. Certified Compound Conc. (wt %) RMS (wt %) 10 min Avg. (wt %) 51.29 RMS (wt %) 5 min 2 min RMS (wt %) RMS (wt %) CaO 51.49 0.09 0.02 0.04 0.06 Al2O3 0.483 0.006 0.443 0.002 0.002 0.005 Fe2O3 0.422 0.004 0.428 0.003 0.003 0.005 MgO 2.161 0.008 2.233 0.010 0.011 0.020 SiO2 2.533 0.011 2.339 0.006 0.008 0.014 SrO 0.030 0.001 0.028 0.0001 0.0001 0.0002 K2O 0.075 0.003 0.074 0.0015 0.0026 0.0029 MnO 0.020 <0.001 0.020 0.0004 0.0006 0.0010 NaO2 0.042 0.003 0.048 0.001 0.002 0.004 P2O5 0.034 0.001 0.037 0.001 0.001 0.001 SO3 0.197 0.005 0.207 0.001 0.002 0.004 TiO2 0.023 0.001 0.023 0.002 0.002 0.002 7.2 0.4 0.6 0.8 Pb (ppm) Global and near 0 42.0 Figure 2. Calibration graph for CaO in limestone standards prepared as pressed powder pellets 2 min - - Conclusions The results clearly demonstrate the excellent capability of the Epsilon 3X for the full analysis of the mineral content of limestone. The high performance of the silicon drift detector in the Epsilon 3X makes it possible to get accurate and fast results. The Epsilon 3X will give the user the opportunity to control the production process close to their production facilities, and analyze the grade of the excavated limestone. The required precision will determine the required measurement time of the application. PANalytical B.V. Lelyweg 1, 7602 EA Almelo P.O. Box 13, 7600 AA Almelo The Netherlands T+31 (0) 546 534 444 F+31 (0) 546 534 598 [email protected] www.panalytical.com Regional sales offices Americas T +1 508 647 1100 F +1 508 647 1115 Europe, Middle East, Africa T +31 (0) 546 834 444 F +31 (0) 546 834 499 Asia Pacific T +65 6741 2868 F +65 6741 2166 Although diligent care has been used to ensure that the information herein is accurate, nothing contained herein can be construed to imply any representation or warranty as to the accuracy, currency or completeness of this information. The content hereof is subject to change without further notice. Please contact us for the latest version of this document or further information. © PANalytical B.V. 2009. 9498 707 58211 PN10765 Corrected intensity 24.0 10 min 52
© Copyright 2024 Paperzz