INTRODUCTION
P.2
PRECAUTIONS FOR SAFE OPERATION
P.3
ENVIRONMENTAL CONDITIONS
P.6
SHAKER COMPONENTS
P.7
Control panel and keypad
P.8
INSTALLATION SITE
P.9
START-UP OF UNIT
P.10
FIXING SAMPLES ONTO PLATFORM
P.10
OPTIONAL COMPONENTS
P.11
LOCATION OF ERLENMEYER FLASK
Location of Erlenmeyer flask (MIR-38PFT)
P.12
Location of Erlenmeyer flask (MIR-38PFN)
P.13
NUMBER OF FLASKS AND REVOLUTION
P.14
CAUTION FOR CULTURE
P.15
OPERATION OF CONTROL PANEL
P.16
SETTING THE ROTATIONAL ACCELERATION
P.17
KEY LOCK FUNCTION
P.17
AUTO LIMIT FUNCTION
P.18
ALARMS & SAFETY FUNCTIONS
P.19
ROUTINE MAINTENANCE
Cleaning
P.20
Removal of platform
P.20
Replacement of consumable
P.20
TROUBLESHOOTING
P.21
DISPOSAL OF UNIT
P.22
COMBINATION OF MIR-S100
P.27
INTEGRATION INSIDE INCUBATOR
P.28
SPECIFICATIONS
P.29
PERFORMANCE
P.29
SAFETY CHECK SHEET
P.30
1
INTRODUCTION
■ Read this operating instruction carefully before using the Product and follow the instructions for safety
operation.
■ Our company disavows any responsibility for safety if the Product is used for other than the intended
use or used with any procedures other than those given in this operating instruction.
■ Keep this operating instruction in a suitable place so that it can be referred to as necessary.
■ The contents of this operating instruction are subject to change without notice for improvement of
performance or functions.
■ Contact our sales representative or agent if any page of the operating instruction is lost or the page
order is incorrect.
■ Contact our sales representative or agent if any point in this operating instruction is unclear or if there
are any inaccuracies.
■ No part of this operating instruction may be reproduced in any form without the expressed written
permission of our company.
CAUTION
Our company guarantees the product under certain warranty conditions. Our company in no
way shall be responsible for any loss of content or damage of content.
2
PRECAUTIONS FOR SAFE OPERATION
It is imperative that the user complies with this operating instruction
as it contains important safety advice.
Items and procedures are described so that you can use this unit correctly and safely.
If the precautions advised are followed, this will prevent possible injury to the user and
any other person.
Precautions are illustrated in the following way:
WARNING
Failure to observe WARNING signs could result in a hazard to personnel
possibly resulting in serious injury or death.
CAUTION
Failure to observe CAUTION signs could result in injury to personnel and
damage to the unit and associated property.
Symbol shows;
this symbol means caution or warning.
this symbol means an action is prohibited.
this symbol means an instruction must be followed.
Be sure to keep this operating instruction in a place accessible to users of this unit.
3
PRECAUTIONS FOR SAFE OPERATION
WARNING
Do not use the unit outdoors. Current leakage or electric shock may result if the unit is exposed to
rain water.
Only qualified engineers or service personnel should install the unit.
unqualified personnel may cause electric shock or fire.
The installation by
Install the unit on a sturdy floor and take an adequate precaution to prevent the unit from
turning over. If the floor is not strong enough or the installation site is not adequate, this may result
in injury from the unit falling or tipping over.
Never install the unit in a humid place or a place where it is likely to be splashed by water.
Deterioration of the insulation may result which could cause current leakage or electric shock.
Never install the unit in a flammable or volatile location. This may cause explosion or fire.
Never install the unit where acid or corrosive gases are present as current leakage or electric
shock may result due to corrosion.
Always ground (earth) the unit to prevent electric shock. If the power supply outlet is not
grounded, it will be necessary to install a ground by qualified engineers.
Never ground the unit through a gas pipe, water main, telephone line or lightning rod. Such
grounding may cause electric shock in the case of an incomplete circuit.
Connect the unit to a power source as indicated on the rating label attached to the unit. Use
of any other voltage or frequency other than that on the rating label may cause fire or electric shock.
Do not insert metal objects such as a pin or a wire into any vent, gap or any outlet on the unit.
This may cause electric shock or injury by accidental contact with moving parts.
Use this unit in safe area when treating the poison, harmful or radiate articles.
may cause bad effect on your health or environment.
Improper use
Turn off the power switch (if provided) and disconnect the power supply to the unit prior to
any repair or maintenance of the unit in order to prevent electric shock or injury.
Do not touch any electrical parts (such as power supply plug) or operate switches with a wet
hand. This may cause electric shock.
4
WARNING
Ensure you do not inhale or consume medication or aerosols from around the unit at the time of
maintenance. These may be harmful to your health.
Never splash water directly onto the unit as this may cause electric shock or short circuit.
Never put containers with liquid on the unit as this may cause electric shock or short circuit when
the liquid is spilled.
Never bind, process, or step on the power supply cord, or never damage or break the power
supply plug. A broken supply cord or plug may cause fire or electric shock.
Do not use the supply cord if its plug is loose.
Such supply cord may cause fire or electric shock.
Never disassemble, repair, or modify the unit yourself. Any such work carried out by an
unauthorized person may result in fire, or electric shock or injury due to a malfunction.
Disconnect the power supply plug if there is something wrong with the unit.
abnormal operation may cause electric shock or fire.
Continued
When removing the plug from the power supply outlet, grip the power supply plug, not the cord.
Pulling the cord may result in electric shock or fire by short circuit.
Disconnect the power supply plug before moving the unit.
cord. A damaged cord may cause electric shock or fire.
Take care not to damage the power
Disconnect the power plug when the unit is not used for long periods. Keeping the connection
may cause electric shock, current leakage, or fire due to the deterioration of insulation.
The disposal of the unit should be accomplished by appropriate personnel. Remove doors to
prevent accidents such as suffocation.
Do not put the packing plastic bag within reach of children as suffocation may result.
5
PRECAUTIONS FOR SAFE OPERATION
CAUTION
This unit must be plugged into a dedicated circuit protected by branch circuit breaker.
Use a dedicated power source as indicated on the rating label attached to the unit.
may cause fire resulting from abnormal heating.
A multiple-tap
Check the setting when starting up of operation after power failure or turning off of power
switch. The stored items may be damaged due to the change of setting.
Prepare a safety check sheet when you request any repair or maintenance for the safety of service
personnel.
ENVIRONMENTAL CONDITIONS
This equipment is designed to be safe at least under the following conditions (based on the IEC
61010-1):
■ Indoor use;
■ Altitude up to 2000 m;
■ Ambient temperature 5 oC to 40 oC
■ Maximum relative humidity 80 % for temperature up to 31 oC decreasing linearly to 50 % relative
humidity at 40 oC;
■ Mains supply voltage fluctuations not to exceed ±10 % of the nominal voltage;
■ Other supply voltage fluctuations as stated by the manufacturer;
■ Transient overvoltages according to Installation Categories (Overvoltage Categories) II; For mains
supply the minimum and normal category is II;
■ Pollution degree 2 in accordance with IEC 60664.
6
SHAKER COMPONENTS
2
5
7
6
3
4
1. Power switch:
1
Main switch of the shaker.
2. Platform (MIR-38PFT/MIR-38PFN):
can be cleaned by removing.
3. Control panel:
The clamps and test tube rack can be fixed on the platform.
It
The revolution speed and operating time can be set through the control panel.
4. Platform fixing screw:
5. Top plate:
removed.
Also functions as an over-current breaker.
This screw is used to fix the platform to the top plate.
The platform can be fixed on this plate.
6. Power cable port:
The surface can be cleaned after the platform is
Use the enclosed power supply cord.
7. DAQ (option) mounting location
7
SHAKER COMPONENTS
Control panel and keypad
2
5
1
4
3
6
7
8
9
10
1. Display of revolution speed and time
In operation, this indicator shows the revolution speed or time. (In timer running, the indicator shows
expiration time. In continuous running, it shows integrated time up to 99.9 hours.*1) In setting mode,
the indicator shows setting value of a revolution speed or time. The display can change over by
pressing select key. And this indicator also shows the error code when detects abnormal by
self-diagnosis function.
2. RPM lamp (RPM): This lamp lights when the display shows the revolution speeds, and flashes
during the revolution speed setting mode.
3. TIME lamp (TIME): This lamp lights when the display shows the remaining time or elapsed time, and
flashes during setting mode of time.
4. Auto limit lamp (AUTO LIMIT): This lamp lights when the auto limit function is on, and flashes when
the auto limit function is activated. Details of auto limit function should be referred to page 18.
5. Start/Stop key (START/STOP): Pressing this key gets the shaker started. Pressing this key again
stops the shaker. To operate with auto limit function off, press this key for 3 seconds.
6. Select key (SELECT):
and time in the display.
Pressing this key changes over setting or present value of revolution speed
7. Set key (SET): Pressing this key enable the revolution speed and time to be set.
display will blink during the setting mode.
Indication in the
8. Digit shift key (
): Pressing this key in the setting mode causes the changeable digit to shift.
Press this key more than 5 seconds to enter the key lock mode.
9. Numerical value shift key (
): Pressing this key in the setting mode causes the numerical value
to shift. Press this key more than 5 seconds to enter function mode.
10. Enter key (ENT):
Pressing this key ensures and memorizes the set value.
Note:
*1. The time display is indicated 0.1 hour (6 minutes) in more than 6 minutes indication and by a second
in less than 6 minutes indication.
*2. Power failure alarm (PF display, buzzer < intermittent tone >) is released by pushing any key.
8
INSTALLATION SITE
To operate this unit properly and to obtain maximum performance, install the unit in a location with the
following conditions:
■ Sturdy and level floor
Always install the unit on a sturdy and level floor. The uneven floor or tilted installation may cause
failure or injury. Install the unit in stable condition to avoid the vibration or noise. Unstable condition
may cause vibration or noise.
WARNING
Install the unit on a sturdy floor. If the floor is not strong enough or the installation site is not
adequate, this may result in injury from the unit falling or tipping over.
Select a level and sturdy floor for installation. This precaution will prevent the unit from tipping.
Improper installation may result in water spillage or injury from the unit tipping over.
■ Low humidity
Install the unit in the ambient of 80 %R.H. or less humidity.
current leakage or electric shock.
Installation under high humidity may cause
WARNING
Do not use the unit outdoors. Current leakage or electric shock may result if the unit is exposed to
rain water.
Never install the unit in a humid place or a place where it is likely to be splashed by water.
Deterioration of the insulation may result which could cause current leakage or electric shock.
■ No flammable or corrosive gas
Never install the unit in a flammable or corrosive location. This may cause explosion or fire or may
result in the current leakage or electric shock by the corrosion of the electrical components.
9
START-UP OF UNIT
Starting trial operation or operation of this shaker, shall be followed by the next procedure.
1. Remove the packaging materials and tapes.
Note:
Remove the cable tie banding the power supply cord. Prolonged banding may cause the corrosion of
the cord coating.
2. If the product is dirty, clean it with a diluted neutral dishwashing detergent. (Undiluted detergent can
damage the plastic components. For the dilution, refer to the instruction of the detergent.) After the
cleaning with the diluted detergent, always wipe it off with a wet cloth. Then wipe off it with a dry cloth.
3. Plug the power supply cord into the port on the left of the rear side of the shaker.
4. Set up the unit correctly according to “INSTALLATION SITE” of page 9.
FIXING SAMPLES ONTO PLATFORM
●Fix the platform to top plate
Set a platform along with two stopping holes in rear of the top plate, and secure the platform by the two
fixing screws.
●Fix an Erlenmeyer flask to platform
Screw the optional clamp to the platform and set an Erlenmeyer flask in it. (Refer to page 12 and 13)
Do not get the Erlenmeyer flask out of platform widely. In case of setting in an incubator, there is the
possibility of touching the wall inside.
CAUTION
Fix a clamp to the platform securely. Improper fixing is dangerous. To prevent unbalance, set an
Erlenmeyer flask symmetrically at the center of the platform, and do not get the flasks out of platform
widely.
Reduce the revolution speed when the total weight of flasks exceed the allowable limit. (To
prevent abnormal vibration.)
10
OPTIONAL COMPONENTS
[Optional components]
The following table lists for the MIR-S100.
Variety
Model
Applicable flask
Maximum Mounting Quantity
for each platform
MIR-38PFT
MIR-38PFN
MIR-51FC
50 ml Erlenmeyer Flask
18 pcs.
20 pcs.
MIR-101FC
100 ml Erlenmeyer Flask
15 pcs.
11 pcs.
MIR-201FC
200 ml Erlenmeyer Flask
9 pcs.
6 pcs.
MIR-301FC
300 ml Erlenmeyer Flask
9 pcs.
6 pcs.
MIR-501FC
500 ml Erlenmeyer Flask
5 pcs.
6 pcs.
MIR-1001FC
1000 ml Erlenmeyer Flask
3 pcs.
3 pcs.
MIR-2001FC
2000 ml Erlenmeyer Flask
2 pcs.
2 pcs.
Test tube rack
MIR-204SC
Various test tube
1 pc.
1 pc.
Stabilizer sheet
MIR-38SS
Petri dish and Flask
1 sheet
1 sheet
Spring rack
MIR-38SP
Various test tube
Clamp
All in one rack
(Max clamp)
[Clamp example]
500 ml
100 ml
Test tube rack
Clamps for Erlenmeyer flasks
●Fix a selected clamp to the platform firmly with M4 screw (MIR-38PFT) or M5 screw (MIR-38PFN) <A
flask of 50 ml-300 ml is fixed by one screw and that of 500 ml-2000 ml is fixed by four screws>. Set the
clamp symmetrically with respect to the center point of platform. (Refer to page 12 and 13)
●The spring of clamp may lose holding power after long-term usage. Replace the spring of clamp as
necessary.
●About optional components, contact a qualified service personnel.
●Do not use other competitor’s pressure-sensitive adhesive sheet to fix a clamp.
11
LOCATION OF ERLENMEYER FLASK
Location of Erlenmeyer flask (MIR-38PFT)
50 ml Erlenmeyer flask × 18
100 ml Erlenmeyer flask × 15
200 ml Erlenmeyer flask × 9
300 ml Erlenmeyer flask × 9
1000 ml Erlenmeyer flask × 3
500 ml Erlenmeyer flask × 5
Screw holes
2000 ml Erlenmeyer flask × 2
Screw holes location of clamps for 500 ml-2000 ml flasks
12
Location of Erlenmeyer flask (MIR-38PFN)
50 ml Erlenmeyer flask × 20
100 ml Erlenmeyer flask × 11
200 ml Erlenmeyer flask × 6
300 ml Erlenmeyer flask × 6
500 ml Erlenmeyer flask × 6
1000 ml Erlenmeyer flask × 3
Screw holes
2000 ml Erlenmeyer flask × 2
Screw holes location of clamps for 500 ml-2000 ml flasks
13
NUMBER OF FLASKS AND REVOLUTION
This figure is verification data when the shaker is set on a stable testing bench. Depending on the
location where MIR-S100 is set on, the revolution speed shown in the graph is not always attainable.
Adequate maximum revolution (rpm)
2 L flask (Medium 1000 ml)
2 L flask (Medium 400 ml)
300
300
250
250
200
200
150
150
1
2
1
1 L flask (Medium 500 ml)
1 L flask (Medium 200 ml)
500
500
400
400
300
300
200
200
1
2
2
3
1
500 ml flask (Medium 250 ml)
500
2
3
500 ml flask (Medium 100 ml)
500
400
400
380
300
300
200
200
1
2
3
4
5
1
Flask capacity 50 %
2
3
4
5
Flask capacity 20 %
A number of flasks
*50 ml-300 ml of flasks are operational in 500 rpm with maximum number.
14
CAUTION FOR CULTURE
● Be careful of the rotation part (platform).
Be careful not to catch a finger between body and platform in operation. Do not insert metal objects
such as a pin or a wire into gap between platform and top plate. This may cause electric shock or injury
by accidental contact with moving parts.
● Use the power supply cord enclosed with the unit.
Use of an inferior power supply cord may cause electric shock or fire.
● Be careful of frostbite when operating in low temperature.
Do not touch the body and platform with a wet hand when the product is used in 0 oC ambient. This may
cause frostbite.
15
OPERATION OF CONTROL PANEL
The table below shows the basic procedure for setting the control panel. The example in the table is
based on the assumption that the revolution speed is 200 rpm and the culture time is 1 hour.
Note: The unit is set at the factory that the revolution speed is 50 rpm and culture time is continuous
running (00.0).
Table1
Basic operation sequence (Example:revolution speed 200 rpm, culture time 1 hour)
Description of operation
Key operated
Indication after operation
1
Turn the power switch ON
-----
The figure of 0 is shown in indicator
and RPM lamp is lighted.*1
2
Press set key.
SET
The left digit and RPM lamp
are flashed.
3
By pressing digit shift key
and numerical value shift key,
set the figure to 200.
4
Press enter key.
5
Press select key.*2
6
Press set key.
7
By pressing digit shift key and
numerical value shift key, set the
figure to 01.0.
8
Press enter key.
9
Press start/stop key.
When pressed,
the changeable digit is shifted.
When pressed,
the figure of settable digit changes.
ENT
The revolution speed is memorized,
and back to the initial display.
SELECT
Timer mode is shown in indicator
and TIME lamp is lighted.
SET
The left digit and TIME lamp are
flashed.
When pressed,
the changeable digit is shifted.
When pressed,
the figure of settable digit changes.
ENT
START/STOP
Timer setting is memorized,
and back to the initial display.
Start operation in 200 rpm
and the remaining time is indicated.
*1. When TIME lamp is lighted, turn on RPM lamp by pressing select key (SELECT).
*2. Culture time is 00.0 (continuous running) in default.
*3. Settable range of culture time is from 0 to 99 hours 54 minutes (00.0-99.9). Setting value of 00.0 is a
continuous running.
*4. Count down display is shown by 0.1 hour until remaining times is 6 minutes, and shown by 1 second
when remaining time is less than 5 minutes 59 seconds.
*5. In case of not using timer mode (culture time setting is 00.0), timer indicator displays total running
time.
CAUTION
Handle the reagents and samples used to react with care. In case of accidental entry into the eyes or
ingestion, this may be harmful to health. In terms of toxicity of regents, inquire at a regent manufacturer.
Handle the samples and reaction products. Contamination between samples may cause unexpected
harmful products.
CAUTION
Do not get up the body in operation. This may cause injury by catching a finger between the body and
platform.
16
SETTING THE ROTATIONAL ACCELERATION
The table below shows the basic procedure for setting a rotational acceleration speed at the start up. The
acceleration speeds has 3 speeds (1, 2, 3). Speed 3 is the fastest.
Note: The unit is set at the factory that the rotational acceleration is 2.
Table 2
3.)
Procedure for changing a rotational acceleration speed. (Example: changing from 1 to
Description of operation
Key operated
Indication after operation
1
Press numerical value shift key
for 5 seconds.
A function mode is shown
in the indicator.
When pressed, the changeable digit is shifted.
2
By pressing digit shift key
and numerical value shift key,
set the figure to F23.
3
Press enter key.
4
Press numerical value shift key
and scroll the figure from 1 to 3.
5
Press enter key.
When pressed,
the figure of settable digit changes.
ENT
When pressed, the current
acceleration speed is flashed.
When pressed, the acceleration
speed value changes.
ENT
Set and back to the initial display.
*Do not change other functional code except F23.
KEY LOCK FUNCTION
This shaker is provided with a key lock function
which disable START/STOP and change of setting
value by the key operation.
Note: The unit is set at the factory that key lock is
OFF (L0)
Table 3
MODE
FUNCTION
L0
Key lock OFF
Enable to change
L1
Key lock ON
Disable to change
Procedure for key lock setting (change from key lock OFF to key lock ON)
Description of operation
Key operated
Indication after operation
Displays current RPM or the time.
1
Press digit shift key for 5 seconds.
Displays L0 in indicator.
2
Set 1 in initial place by pressing
numerical value shift key.
When pressed,
the figure of settable digit changes.
3
Press enter key.
ENT
Key lock on and current RPM
or the time is dispayed.
●A key lock function is valid for start/stop, and the setting of revolution / culture time function only.
(Entering setting mode and check of setting value is possible, but change of setting value is impossible.)
●For release of key lock, select key lock OFF (L0).
17
AUTO LIMIT FUNCTION
Auto limit function is ON after power-ON, and it activates when it detects abnormal vibration during
shaker running. When the auto limit function is activated, the buzzer sounds auto limit lamp flashes and
RPM is reduced 20 rpm. If the abnormal vibration still continues, RPM is reduced 20 rpm again with
buzzerand lamp. In the case of 50 rpm, the shaker stops running because the speed cannot be reduced
anymore. The flashing of auto limit lamp continues after the shaker stopping. The flashing stops when
new rotational speed is set or the running again.
When auto limit function is not used, press a start/stop key (START/STOP) for 3 seconds, and release
start/stop key (STRAT/STOP) after checking the auto limit lamp is turn off. The shaker starts operating
without auto limit function. (Refer to *1-3 below)
CAUTION
In the case of not using auto limit function, be careful of abnormal vibration.
vibration may cause fall down or crash of the shaker, or machine failure.
Table 4
Procedure to cancel auto limit function.
Description of operation
1
2
Too much abnormal
Key operated
Press start/stop key
for 3 seconds.
START
STOP
Release start/stop key
START
STOP
Indication after operation
ON
AUTO
LIMIT
OFF
AUTO
LIMIT
Auto limit lamp turns off.
Start operatings with auto limit function OFF.
*1. Auto limit function is ON after power-ON.
*2. After power failure in operation, the shaker resumes running with the setting before power failure.
*3. When continue pressing start/stop key (START/STOP), ON and OFF of auto limit function is
alternated every 3 seconds.
18
ALARMS & SAFETY FUNCTIONS
This shaker has the alarms and safety functions as shown in table 5, and also self diagnostic functions.
Table 5
Alarm &
Safety
Auto limit
function
Alarms and safety functions
Situation
When occurring abnormal
vibration.
Indication
LED lamp is
flashed
-----
Buzzer
Safety operation
Intermittent Reduce revolution
tone
speed by 20 rpm
-----
Unable of setting
change and
start/stop operation.
Releasing
method
Refer to
page 18
Key lock
When the key lock is “ON”
Auto start
after power
failure.* 1
Auto start after power
failure.
Auto return
When there is no key
pressing in each setting
mode for 90 seconds.
Temperature
abnormality
When the shaker temp. is
higher than over-heat limit
temp.
Revolution
abnormality
Revolution speed do not
arrive setting speed or over
setting speed.
“E07” is flashed
Or deviating 10 rpm from
setting speed even after
being stabilized
Intermittent
tone
shutdown
Motor
overlord
protection
circuit
When the motor is
overloaded and the
overcurrent is detected.
“E06” is flashed
Intermittent
tone
Maintenance
notice
function
*2
Total running time is
10,000 hours or more.
“CH1” is flashed
twice after
power-ON.
-----
-----
-----
Failure
predicting
function*3
When the arrival time to
“CH2” is flashed
100 rpm (or to setting RPM
twice just after
if it is less than 100 rpm)
power-ON.
takes 90 seconds or more.
-----
-----
-----
Alternative
Intermittent
indication of PF/real
tone
revolution
-----
“E08” is flashed
-----
----PV indication after
90 seconds ( return
RPM or time )
-----
Press any
key.
-----
Intermittent
tone
Turns off
the power
switch, and
turn on
again.
*PF(Power Failure), PV(Present Value)
*1. Auto start after PF : The shaker restarts operation after power failure when shaker stops operation
by power failure. The timer is not counted during power failure and the count is resumed after power
failure when the timer mode selected.
*2. Maintenance notice function : This function is to notice the replacement of consumable. (Refer to
page 20)
*3. Failure predicting function : This function is to notify the abnormality of MIR-S100 when it dose not
achieve average revolution. ( When “CH2” is flashed with maximum allowable load weight, it is
recommend to call for service.)
19
ROUTINE MAINTENANCE
Cleaning
Perform the following procedure to clean the MIR-S100 when use it for the first time or contaminate under
the platform.
●Clean removing platform with a diluted neutral dishing detergent with distilled water.
●Wipe the interior parts and inside platform with alcohol for sterilization and then wipe off with a dry
gauze so as not to remain alcohol for the sterilization.
●Do not use a solution of disodium chlorate or other halogen-based solution because this may cause
rust.
●In case of heavy dirt, consult sales representative or agents.
Removal of platform
1. Turn off the power switch and unplug.
4. Pull the platform forward.
2. Rotate platform fixing screws to
counterclockwise direction.
5. Below is the shaker removed the platform.
3. Pick up the platform at the front.
6. To replace the platform, follow the procedure
in reverse.
I
Replacement of consumable
The belt and bearings used in MIR-S100 are consumable parts. MIR-S100 has a maintenance notice
function.
When “CH1” is indicated in the display after power ON, it recommends replacement of those parts. For
the replacement, consult sales representative or agents.
20
TROUBLESHOOTING
If the unit malfunctions, check out the following before calling for service.
Malfunction
The shaker does not
operate at all.
Check/Remedy
• The shaker is not plugged correctly into a power outlet.
• The circuit breaker at the power source is active or a power failure
has occurred.
• The power supply cord is not connected to the port on the rear of the
shaker.
The key operation is
disabled.
• The key lock function is set in ON mode. (Refer to page 17)
Under the setting mode.
The alarm function
operates.
• The shaker is not running within a correct temperature range.
• The relation between number of flasks and shaker speeds is not
correct. (Refer to page 14)
Shaker speed goes down
to lower than the set value.
The shaker stops during
operation.
“CH2” flashes in display
after power-ON.
(The auto limit function operates continuously in the case of conditions
below.)
• The setting table is microvibrating by being installed in the product
with a compressor etc.
• The installation site is not stable or leveled.
• Flask is not fixed securely.
• Location of clamp is not right.
• Amount of culture medium is not right.
• The timer function is set to ON mode.
• The sample weight on platform is too much. The allowable load
weight is 5 kg.
Note:
If the malfunction is not eliminated after checking the above items, or the malfunction is not shown in the
above table, contact our sales representative or agent.
21
DISPOSAL OF UNIT
WARNING
The disposal of the unit should be undertaken by appropriate personnel.
22
Note:
This symbol mark and recycle system are applied only to EU countries
and not applied to the countries in the other area of the world.
Waste Electrical and Electronic Equipment (WEEE) Directive-2002/96/EC
(English)
Your Panasonic product is designed and manufactured with high quality materials and components which
can be recycled and reused.
This symbol means that electrical and electronic equipment, at their end-of-life, should be disposed of
separately from your household waste.
Please dispose of this equipment at your local community waste collection/recycling centre.
In the European Union there are separate collection systems for used electrical and electronic products.
Please help us to conserve the environment we live in!
(German)
Ihr Panasonic Produkt wurde entworfen und hergestellt mit qualitativ hochwertigen Materialien und
Komponenten, die recycelt und wiederverwendet werden können.
Dieses Symbol bedeutet, daß elektrische und elektronische Geräte am Ende ihrer Nutzungsdauer von
Hausmüll getrennt entsorgt werden sollen.
Bitte entsorgen Sie dieses Gerät bei Ihrer örtlichen kommunalen Sammelstelle oder im Recycling Centre.
In der Europäischen Union gibt es unterschiedliche Sammelsysteme für Elektrik- und Elektronikgeräte.
Helfen Sie uns bitte, die Umwelt zu erhalten, in der wir leben!
23
DISPOSAL OF UNIT
(French)
Votre produit Panasonic est conçu et fabriqué avec des matèriels et des composants de qualité
supérieure qui peuvent être recyclés et réutilisés.
Ce symbole signifie que les équipements électriques et électroniques en fin de vie doivent être éliminés
séparément des ordures ménagères.
Nous vous prions donc de confier cet équipement à votre centre local de collecte/recyclage.
Dans l’Union Européenne, il existe des systèmes sélectifs de collecte pour les produits électriques et
électroniques usagés.
Aidez-nous à conserver l’environnement dans lequel nous vivons !
Les machines ou appareils électriques et électroniques contiennent fréquemment des matières qui, si
elles sont traitées ou éliminées de manière inappropriée, peuvent s’avérer potentiellement dangereuses
pour la santé humaine et pour l’environnement.
Cependant, ces matières sont nécessaires au bon fonctionnement de votre appareil ou de votre machine.
Pour cette raison, il vous est demandé de ne pas vous débarrasser de votre appareil ou machine usagé
avec vos ordures ménagères.
(Spanish)
Los productos Panasonic están diseñados y fabricados con materiales y componentes de alta calidad,
que pueden ser reciclados y reutilizados.
Este símbolo significa que el equipo eléctrico y electrónico, al final de su ciclo de vida, no se debe
desechar con el resto de residuos domésticos.
Por favor, deposite su viejo “televisor” en el punto de recogida de residuos o contacte con su
administración local.
En la Unión Europea existen sistemas de recogida específicos para residuos de aparatos eléctricos y
electrónicos.
Por favor, ayúdenos a conservar el medio ambiente!
24
(Portuguese)
O seu produto Panasonic foi concebido e produzido com materiais e componentes de alta qualidade que
podem ser reciclados e reutilizados.
Este símbolo significa que o equipamento eléctrico e electrónico no final da sua vida útil deverá ser
descartado separadamente do seu lixo doméstico.
Por favor, entregue este equipamento no seu ponto local de recolha/reciclagem.
Na União Europeia existem sistemas de recolha separados para produtos eléctricos e electrónicos
usados.
Por favor, ajude-nos a conservar o ambiente em que vivemos!
(Italian)
Il vostro prodotto Panasonic è stato costruito da materiali e componenti di alta qualità, che sono
riutilizzabili o riciclabili.
Prodotti elettrici ed elettronici portando questo simbolo alla fine dell’uso devono essere smaltiti
separatamente dai rifiuti casalinghi.
Vi preghiamo di smaltire questo apparecchio al deposito comunale.
Nell’Unione Europea esistono sistemi di raccolta differenziata per prodotti elettrici ed elettronici.
Aiutateci a conservare l’ambiente in cui viviamo!
25
DISPOSAL OF UNIT
(Dutch)
Panasonic producten zijn ontwikkeld en gefabriceerd uit eerste kwaliteit materialen, de onderdelen kunnen
worden gerecycled en weer worden gebruikt.
Het symbool betekent dat de elektrische en elektronische onderdelen wanneer deze vernietigd gaan
worden , dit separaat gebeurt van het normale huisafval.
Zorg ervoor dat het verwijderen van de apparatuur bij de lokaal erkende instanties gaat gebeuren.
In de Europese Unie wordt de gebruikte elektrische en elektronische apparatuur bij de daarvoor wettelijke
instanties aangeboden.
Alstublieft help allen mee om het milieu te beschermen.
(Swedish)
Din Panasonic produkt är designad och tillverkad av material och komponenter med hög kvalitet som kan
återvinnas och återanvändas.
Denna symbol betyder att elektriska och elektroniska produkter, efter slutanvändande, skall sorteras och
lämnas separat från Ditt hushållsavfall.
Vänligen, lämna denna produkt hos Din lokala mottagningstation för avfall/återvinningsstation.
Inom den Europeiska Unionen finns det separata återvinningssystem för begagnade elektriska och
elektroniska produkter.
Vänligen, hjälp oss att bevara miljön vi lever i!
26
COMBINATION OF MIR-S100
Combination of MIR-S100 with cooled incubator, and versatile environmental test chamber.
● Cooled incubator: MIR-153, MIR-253, MIR-553
● Versatile environmental test chamber : MLR-350, MLR-351 (Can not be used in MLR-350(H),
MLR-351(H))
*It is recommended to install MIR-S100 on the lower shelf in the incubator to prevent abnormal vibration.
Caution
・Select sturdy floor when combine MIR-S100 with our incubators and set the incubator in place with
leveling foot.
In the case of combining MIR-S100 with cooled incubator or versatile environmental test chamber, use
fixing kit MIR-15FP provided as optional component.
Fixing procedure of MIR-15FP to cooled incubator
1. Remove a shelf of incubator to be set the MIR-S100. (The lower shelf is recommended)
2. Fix the plate to the shelf with 6 screws, by using two plate fixing supports. (See figure below)
3. Replace the shelf with plate into the incubator. Fix the shelf to incubator by using the shelf fixing metal
right, left, and shelf fixing support. (See figure below)
4. Remove the cap on the access port of the incubator and pass the power code of MIR-S100 through
the port.
5. Put the cap enclosed with the kit on the access port.
6. Connect the power cord into MIR-S100, and set MIR-S100 on the plate.
Plate
Shelf fixing metal left
Shelf
Shelf fixing metal right
Shelf fixing supports
Plate fixing supports
27
INTEGRATION INSIDE INCUBATOR
Recommended maximum RPM in our incubator.
Unit:RPM
MIR-153
MIR-253
MIR-553
MLR-350
MLR-351
100 ml Flask
×15
500 ml Flask
×6
Lower shelf
400
300
Upper shelf
300
200
Lower shelf
300
200
Upper shelf
300
300
Lower shelf
300
300
Upper shelf
200
200
Lower shelf
300
200
●Measured by flask with 50 % water of capacity.
●As MIR-153 is lower type cooled incubator, installation of MIR-S100 is applicable to lower shelf only.
28
SPECIFICATIONS
Product name
Shaker MIR-S100
External dimensions
W410 mm x D305 mm x H120 mm
Body: Painted steel
Cover: Stainless
Exterior
Platform dimensions
W380 mm x D250 mm
Platform
Copper alloyed stainless steel
Shaking motion
Circling
Amplitude
(Circling diameter)
20 mm
Revolution
50 rpm - 500 rpm
Allowable load weight
5 kg
Flask intake capacity
100 ml Erlenmeyer Flask: x 15 500 ml Erlenmeyer Flask: x 5 ( MIR-38PFT )
100 ml Erlenmeyer Flask: x 11 500 ml Erlenmeyer Flask: x 6 ( MIR-38PFN )
Display
3-digit LED (speeds/time changeable)
Motor
DC brushless motor
Accessories
Power supply cord
Power source
100 V to 240 V, 50 Hz/60 Hz
Weight
Optional component
15 kg
Platform (MIR-38PFT, MIR-38PFN)
Clamp (MIR-51FC, MIR-101FC, MIR-201FC, MIR-301FC, MIR-501FC, MIR-1001FC,
MIR-2001FC), Test tube rack (MIR-204SC), Spring rack (MIR-38SP)
Stabilizer sheet (MIR-38SS), Fixing kit (MIR-15FP), Interface board (MTR-480)
Note: Refer to the updated catalog when ordering an optional component.
Designs and specifications are subject to change without notice.
PERFORMANCE
Product name
Model No.
Shaker MIR-S100
MIR-S100-PT
MIR-S100-PC
Maximum power consumption
27 W
Maximum heat emission
97 kJ/h
Rated voltage, frequency
AC 100 V to 240 V, 50 Hz/60 Hz
Amperage
Usable environment condition
MIR-S100-PE
Max. 0.27 A
o
o
Temperature; 0 C - 60 C, Humidity; equal or less than 80 %R.H.
*These data based on our measuring method.
*Design or specifications will be subject to change without notice.
29
CAUTION
Please fill in this form before servicing.
Hand over this form to the service engineer to keep for his and your safety.
Safety check sheet
1. Shaker contents :
Risk of infection:
Risk of toxicity:
Risk from radioactive sources:
□Yes
□Yes
□Yes
□No
□No
□No
(List all potentially hazardous materials that have been stored in this unit.)
Notes :
2. Contamination of the unit
Unit interior
No contamination
Decontaminated
Contaminated
Others:
□Yes
□Yes
□No
□No
□Yes
□No
3. Instructions for safe repair/maintenance of the unit
□Yes
□No
a) The unit is safe to work on
□Yes
□No
b) There is some danger (see below)
Procedure to be adhered to in order to reduce safety risk indicated in b) below.
Date :
Signature :
Address, Division :
Telephone :
Product name :
Shaker
Model No.
MIR-
Serial number :
Please decontaminate the unit yourself before calling the service engineer.
30
Date of Installation :
1-1-1, Sakata Oizumi-Machi Ora-Gun, Gunma 370-0596, Japan
Printed in Japan
7FB6P15137400
© Panasonic Healthcare Co., Ltd. 2012