T I I K A M K K I C A N .IOUUNAI, OF C L I N I C A L PATHOLOGY Vol. 39, No. 3, p p . 235-241 Mureli, 1%3 C o p y r i g h t © 1003 b y T h e Williams & Wilkins Co. 1'rinled in U.S.A. FALLING DROP METHOD FOR DETERMINATION OF SPECIFIC GRAVITY OF URINE M A R T H A W I N S T E A D , A.B., M . T . (ASCP), N . D . SISSON, M . D . , AND 11. L. M A R K E Y South Bend Medical Fovndalion, Inc., 531 North Main Street, South Bend 1, Indiana Many efforts have been made to establish a satisfactory method for determination of the specific gravity of urine on a small specimen. These began in 1926 when Kirkpatrick and Kling7 adjusted the relative proportions of xylene and carbon tetrachloride in a hydrometer cylinder while they simultaneously dropped the sample drop. The specific gravity of the exact mixture that would make the drop hover Avas then measured by means of a hydrometer, and this specific gravity was taken as representing that of the sample. Since Aaron's' article in 1945, the accuracy of measuring the specific gravity of urine by means of the usual hydrometer has been questioned, and numerous methods that circumvent the hydrometer have appeared in the literature. The gradient column, 4 ' 8 a series of tubes containing solutions with known specific gravities into which drops are allowed to fall,5, n and the specific gravity balance methods 9 seem to be cumbersome for use in routine measurement of the specific gravity of urine in the busy hospital laboratory. The falling drop method herein described, which has been in use in our laboratory since .1.956, is basically similar to that of Barbour and Hamilton. 2 It is simply a measure of the time required for a uniform drop of immiscible fluid to traverse a 12-in. distance in a solution of fixed specific gravity. The specific gravity of urine is found on a time-wersMS-specific gravity standard curve established with sucrose solutions of known specific gravity.* ' Received, March 15, 1902; revision received, September 21; accepted for publication, November 14. Miss Winstead is Chief of Quality Control, D r . Sisson is Pathologist., and Mr. Markey is Technical Advisor. * Sodium chloride solutions may also be used for standardization. 0 235 F I G . 1. Falling drop column 236 WINSTEAD BT The accuracy of falling drop methods (if corrections for temperature and proper standardization are considered) has been well established in relation to the gravimetric and refractometric methods. 6,12 Comparison is made between the values for the specific gravity when a simple American Optical Company refractometer (A-0 TS Meter)f and 4 urine hydrometers from 3 different manufacturers are used. METHOD 1. Draw the urine into a 1-ml. serologic pipet and immerse the tip approximately Ys in. into the fluid column. 2. Permit a drop (0.05 ml.) to form on the tip of the pipet, and remove the pipet from the fluid in order to release the drop from the tip. 3. With a stopwatch, time the fall of the drop from the upper to the lower line on the column. 4. Record the time and find the value for the specific gravity on the calibration chart. MATERIALS AND EQUIPMENT 1. Falling drop column. The column (Fig. 1) is made of heavy gage glass and has a uniform internal diameter of Yi hi. It is scribed with 2 lines, an upper and a lower, which are 12 in. apart. J The fall rate is the time needed for the drop to fall from the upper to the lower of these 2 lines. The bottom of the column is fitted with a Teflon stopcock that allows the urine drops that are collecting to be drawn off without disturbing the mixture in the column. In use, the column is held exactly vertical on a buret stand. (Note—This column should be siliconed with G. E. Drifilm #SC-87;§ aqueous silicones have proved to be not suitable.) Gentle rinsing with detergent cleans the column, but resiliconing must be performed occasionally. Old silicone is | Manufactured by American Optical Company, 10 S. Wabash Avenue, Silversmith Bldg., Box 804, Chicago 90, Illinois. J Available from Research Apparatus Co., Inc., Ill W. Mill St., Wauconda, Illinois. § Available from General Electric Chemical Dcpt., PiUsfield, Massachusetts. Vol. S& AL. TABLE 1 W E I G H T S FOR SUCROSE STANDARDS Specific Gravity* Sucrose Standards Cm./100 ml. 1.002 2.012 3.030 4.056 5.0S9 S.240 9.306 11.470 1.003 1.007 1.011 1.015 1.019 1.031 1.035 1.043 * At room temperature of 23 C. removed by means of allowing a dichromatesulfuric acid cleaning solution to stand in the column for 1 hr. Instructions for siliconing. Silicone isapplied in the following manner: A. Close the small end of the column with a rubber stopper and pour in a small amount of Drifilm. B. Plug the large end of the column and rotate so that the entire surface is covered. C. Pour out excess Drifilm, unstopper, and leave the column in an upright position overnight to dry. D. Rinse with tap water and dry by rinsing with acetone. 2. Thermometer hung beside the column. Although corrections are ordinarily unnecessary, they must be made for drastic changes in room temperature. The usual <j'f»e in scconas FIG. 2. Calibration curve for specific gravity March 1963 SPECIFIC GRAVITY OF 237 URINE TABLE 2 SPECIFIC GRAVITY OF SUCROSE STANDARDS Hydrometers* Standard Calculation of Sucrose by Weight Refractometer Falling Drop No. 1 No. 2 No. 3 No. 4 1.007 1.011 1.015 1.017 1.021 1.032 1.035 1.042 1.007 1.009 1.013 1.018 1.021 1.033 1.039 1.043 1.002 1.009 1.015 1.018 1.022 1.034 1.037 1.04G 1.008 1.012 1.014 1.010 1.021 1.031 1.039 1.045 sec. 1.003 1.007 1.011 1.015 1.019 1.031 1.035 1.043 1.003 1.0005 1.0105 1.0145 1.0185 1.0305 1.0345 1.043f 48, 23, 17, 13, 11, 8, 50 23 17 13 11 8 7M, 7K VA, <$A * Values are corrected to 1.000 with distilled water and are also corrected for temperature. f Solution was diluted 1:2 because the refractometer read only to 1.035. TABLE 3 SPECIFIC GRAVITY OF U R I N E BY M E A N S REFRACTOMETER, F A L L I N G D R O P , AND HYDROMETER AT 23 C. Urine No. Refractometer 1 2 3 4 5 0 7 S 9 10 11 12 13 14 15 10 17 18 19 20 1.008 1.008 1.0075 1.015 1.014 1.013 1.0145 1.010 1.019 1.019 1.023 1.025 1.024 1.023 1.0235 1.0215 1.025 1.0255 1.027 1.032 Hydrometers*t Falling Drop 1.008 1.008 1.010 1.013 1.0135 1.013 1.015 1.010 1.019 1.019 1.020 1.023 1.0225 1.021 1.023 1.020 1.025 1.027 1.025 1.031 OF No. 1 No. 2 No. 3 No. 4 1.012 1.011 1.012 1.015 1.018 1.019 1.020 1.021 1.024 1.023 1.025 1.030 1.029 1.020 1.029 1.024 1.029 1.011 1.011 1.012 1.015 1.018 1.020 1.021 1.023 1.024 1.026 1.030 1.028 1.026 1.031 1.025 1.030 1.014 1.010 1.011 1.014 1.017 1.017 1.018 1.018 1.023 1.024 1.025 1.029 1.028 1.020 1.029 1.023 1.028 1.011 1.010 1.012 1.015 1.018 1.018 1.021 1.022 1.024 1.024 1.027 1.030 1.028 1.028 1.029 1.025 1.029 — — — — — 1.030 1.031 1.030 1.030 1.035 1.037 1.030 1.037 * Values arc corrected to 1.000 with distilled water and arc also corrected for t e m p e r a t u r e . f —indicates t h a t there was not sufficient urine for determining specific gravity. correction of 0.001 specific gravity for each 3 C. variation is suitable for this method 2 ' 7 and should be applied for room temperature outside the 22 to 24 C. range. 3. Standard sucrose solutions. For use in selecting the appropriate points on the time-versMS-specific gravity graph, the required solutions of sucrose and their specific gravities are listed in Table 1. Although several dilutions of a concentrated solution could have been used, individual sucrose weighings were made in order to establish independent points on the graph. 4. Solution for the column. This material is prepared in the following manner: A. Add 28 ml. of bromobenzene* to 72 ml. of kerosene in a graduated cylinder. Mix well and fill the column with the mixture to approximately 2 in. from the top. B. Drop the 1.003 sucrose standard twice and time its fall. This should require 50 sec. ± 3 sec. A time longer than 50 sec. requires addition of kerosene; less time requires addition of bromobenzene. C. Adjust with the proper solution until the sucrose 1.003 standard gives the correct fall time. The adjusting solution is added dropwise, and the column (if the lower clamp is swung out of the way) can be inverted by means of rocking it in the upper clamp for easy mixing. (Note—Small droplets of standard suspended in the mixture affect the time. For the final check of time after addition of the adjusting mixture, allow the droplets to settle and retest the 1.003 sucrose standard.) * Available from E a s t m a n - K o d a k State St., Rochester 4, New York. Co., 343 238 WINSTEAD ET Vol. 39 AL. TABLE 4 C O N F I D E N C E L I M I T S (3 S.D.)* FOR THE V A R I O U S M E T H O D S OF D E T E R M I N A T I O N OF SPECIFIC GRAVITY Refractometer No. 1 No. 2 No. 3 Falling Drop Hydrometer No. 1 Avg. sp. gr. 3 S.D. Avg. sp. gr. 3 S.D. Avg. sp. gr. 3 S.D. 1.0050 1.0190 1.0310 No variation No variation No variation 1.0061 1.0199 1.0314 0.0006 0.0009 0.0005 1.0072 1.0217 1.0337 0.0025 0.0020 0.0020 S.D. = standard deviation. 5. Refractometer. An A-0 TS Meter is used. 6. Hydrometers. Four hydrometers are obtained from 3 different manufacturers. CALIBRATION Each of the standards should be dropped at least twice and their fall times recorded. A time-yersMS-specific gravity curve is constructed on a linear graph (Fig. 2), and a chart for routine use may be prepared from the graph. RESULTS The sucrose standards used for calibration of the falling drop method were checked with the refractometer and with 4 different urine hydrometers. Results are listed in Table 2. Twenty urines were tested for specific gravity by means of the falling drop refractometer and 4 urine hydrometers. Results are summarized in Table 3. In order to check reproducibility, 3 samples of urine with specific gravities of 1.005, 1.019, and 1.031 were tested 30 times each by means of each method. The results were then used to calculate 3 standard deviations by the replicate formula: I. S. y N -1 These comparisons are listed in Table 4. DISCUSSION The specific gravity of the bromobenzenekerosene mixture may be established at any point between and including 1.000 and 1.003, but once it has been established, it obviously must be kept the same, inasmuch as the time of fall is directly related to the specific gravity of the solution. The specific gravity is, in turn, directly related to the temperature of the solution, and continual correction of temperature is required unless room temperature is kept relatively constant. The difference in specific gravity resulting from a temperature change of 3 C. is 0.001; therefore, if 23 C. is selected in the standardization, the room temperature may range from 22 to 24 C. without correction. The statement made by the manufacturer and others3 that the A-0 TS refractometer was compensated for temperatures between 60 and 100 F. was viewed with some suspicion until the sucrose standards proved to yield values with this instrument that were essentially identical to the standard values. On the basis of this finding, in later portions of this work the refractometric value for specific gravity is accepted as the true value. The correction of the hydrometers used in this work to 1.000 for distilled water ranged from +0.004 to -0.001 (Aaron1 pointed out that the inherent calibration error of hydrometers can be as great as 0.009). All hydrometers had been calibrated for 60 F. (15.5 C ) ; therefore, all specific gravities they recorded required the addition of 0.003 for our room temperature of 23.5 C. If, as we suspect, the average laboratory seldom applies the required correction for temperature, and if, as Miles10 states, the temperature error summates with calibration error, it would seem reasonable that hydrometer determination of specific gravity frequently yields grossly inaccurate values. Reference to Table 2 indicates that even corrected, hydrometer values for specific gravity are significantly in error. March 1963 239 SPECIFIC GRAVITY OF URINE The size of the sample drop may be judged visually after one is accustomed to the amount regularly released by delivery of 0.05 ml. from a 1-ml. serologic pipet. In routine procedure, it is more rapid and practical to drop the urine with a disposable Pasteur pipet of uniform bore. The same pipet may be used for a whole run of urines if it is rinsed with sample between samples. As may be seen in Table 3, the specific gravity values of urine obtained by the falling drop method compare well with those obtained by the refractometric method. The hydrometer values, however, vary from the refractometric ones by a difference of 0.002 to 0.007. When the specific gravity of a single specimen of urine was measured 30 times with the refractometer, no variation in the reading was obtained. Comparison of the statistical data in Table 4 for 3 standard deviations (S.D.) clearly indicates that better reproducibility is obtained with the falling drop method than with use of the hydrometer. In order to compare the falling drop and hydrometer methods with regard to accuracy, the specific gravities for the 20 urines given in Table 3 were used to calculate " t " values. (This is a means of establishing whether or not differences between values by different methods are significant.) The refractometer value was regarded as the true value. For the falling drop method compared with the refractometer method, the " t " value of 1.9, calculated by means of the equation: d is not a significant difference at the 5 per cent level. Comparison of the hydrometer with the refractometer, however, resulted in a " t " value of 11.6—a highly significant difference at the 5 per cent level. In order to maintain this accuracy in routine use of the falling drop method, it is suggested that the fall time for 2 sucrose standards (1.003 and 1.035) be verified daily. As stated in the discussion of calibra- tion, the 1.003 standard should give 50 ± 3 sec. and the 1.035 standard, 7 ± Yi sec. (or whatever time was obtained for this standard calibration). Necessity for adjusting the column mixture may be detected at once. Standards should be kept in the refrigerator, but small aliquots for daily checking of solutions may be left at room temperature for periods as long as 2 weeks. Further precautions necessary are: 1. The column should be kept corked when not in use, in order to avoid evaporation of the fluid. 2. The column should be kept away from open flames and heat (Bunsen burners, sunlight, and so on), in order to avoid fires and alteration of the specific gravity of the solution. 3. Tor better accuracy, times shorter than 7 sec. should be rechecked on a 1:2 dilution of specimen with distilled water. 4. The drop of testing sample should be kept free of air bubbles. The time is entirely inaccurate if even a small bubble is present. CONCLUSIONS A falling drop method for determination of the specific gravity of urine has been described. It is accurate, extremely reproducible, and very simple for routine use. Calibration is made with sucrose standards and is entirely independent of the urine hydrometer. Specific gravity values are identical to those read when an A-0 TS Meter is used and thus furnish a means of establishing true specific gravity values for specimens of urine with which to compare values obtained by other methods. A simple check of the accuracy of daily maintenance calibration is described. There was no variation in the specific gravity replicates made with the use of the A-0 TS Meter. The 3 S.D. average value for reproducibility of the falling drop method was one-third that for the hydrometer (0.0007 to 0.0022). Judging from " t " test calculations, there is no significant difference at the 5 per cent level between the refractometer and the falling drop values. There is, however, a highly significant difference between the refractometer and the hydrometer values. 240 Vol. 39 WINSTEAD ET AL. SUMMARIO IN INTERLINGUA • Es describite un methodo a gutta cadente pro le determination del gravitate specific de urina. Le methodo es accurate, reproducibile, e multo simple a executar in usos routinari. Le calibration es facite con standards de sucrosa e es completemente independente ab le hydrometro de urina. Le valores del gravitate specific assi obtenite es identic con illos legite in le uso del refractometro A-0 TS e forni assi un medio pro establir ver valores de gravitate specific de specimens de urina, con le quales il es possibile comparar valores obtenite per altere methodos. Es describite un simple methodo pro le controlo del accuratia del calibration de mantenentia diurne. Esseva constatate nulle variation in le replicationes del determination de gravitates specific effectuate con le uso del refractometro A-0 TS. Le valor medie de 3 D.S. (deviation standard) pro le reproducibilitate del methodo a gutta cadente esseva un tertio de illo del methodo a hydrometro (0.0007 a 0.0022). A base de calculationes a test t il pote esser asserite que il existe nulle significative differentia al nivello de 5 pro cento inter le valores a refractometro e le valores a gutta cadente. Del altere latere, un differentia altemente significative se notava inter le valores a refractometro e le valores a hydrometro. REFERENCES 1. AARON, H . : The accuracy of clinical urinometers. Am. J . Clin. P a t h . , Tech. S e c , 9 : 70, 1945. 2. B A R B O U R , H. G., AND H A M I L T O N , W. F.: T h e falling drop method for determining specific gravity. J . Biol. Chem., 69: 625640, 1926. 3. B A R R Y , K . G., M C L A U R I N , A. W., P A R N B L L , B. L . : A practical temperature-compensated hand refractometer (the T S M e t e r ) : its clinical use and application in estimation of total serum proteins. J . L a b . & Clin. Med., 55: 803-808, 1960. 4. BROWN, M . E . : Measurements of t h e specific gravity of urine b y means of single gradient column. Am. J . Clin. P a t h . , 29: 188-190, 1958. 5. GOLDMAN, A. S.: A simple rapid method for determination of specific gravity of small samples of urine. Pediatrics, 24: 814-818, 1959. 6. KAGAN, B . M . : Simple method for estimation of total protein content of plasma a n d serum; falling drop method for the determination of specific gravity. J . Clin. Invest., 17: 369-372, 1938. 7. K I R K P A T R I C K , J . , AND K L T N G , D . H . : A sim- plified method for t h e determination of specific gravity of a few drops of urine. J. A. M. A., 87: 487-488, 1926. 8. L O W R Y , O. H . , AND H U N T E R , T . H . : T h e d e - termination of serum protein concentration with a gradient t u b e . J . Biol. Chem., 159: 465-474, 1945. 9. MAGATH, T . B . : Hydrometer for quick a n d accurate determination of specific gravity in urine. Am. J . Clin. P a t h . , 25: 1217-1219, 1955. 10. M I L E S , B . E . , P A T O N , A., AND DE W A R D E N E R , H. E . : Maximum urine concentration. Brit. M . J., 2: 901-905, 1954. 11. M O N R O E , L., AND H O P P E R , J . , J R . : A simple method of determining t h e specific gravity of small samples of urine. J . Lab. & Clin. Med., 31: 934-935, 1946. 12. M O O R E , N . S., AND V A N S L Y K E , D. D . : T h e relationships between plasma specific gravity, plasma protein content and edema in nephritis. J . Clin. Invest., 8: 337-355, 1930.
© Copyright 2026 Paperzz