ab65349 NADP/NADPH Assay Kit Instructions for Use For the rapid, sensitive and accurate measurement of NADP/NADPH in various samples. This product is for research use only and is not intended for diagnostic use. 1 Table of Contents 1. Overview 3 2. Protocol Summary 3 3. Components and Storage 4 4. Assay Protocol 6 5. Data Analysis 9 6. Troubleshooting 10 2 1. Overview Assays of nicotinamide nucleotides are of continual interest in the studies of energy transforming and redox state of cells or tissue. Abcam’s NADP/NADPH Assay Kit provides a convenient tool for sensitive detection of the intracellular nucleotides: NADP, NADPH and their ratio. The enzymes in the system specifically recognize NADP/NADPH in an enzyme cycling reaction (It does not recognize NAD+/NADH). There is no need to purify NADP/NADPH from sample mix. The enzyme cycling reaction significantly increases detection sensitivity. Results can be quantified using plate reader at OD450nm. 2. Protocol Summary Sample Preparation Standard Curve Preparation Add NADP Cycling Mix Measure Optical Density 3 3. Components and Storage A. Kit Components Item Quantity NADP/NADPH Extraction Buffer 50 mL NADP Cycling Buffer 15 mL NADP Cycling Enzyme Mix 0.2 mL NADPH Developer Stop Solution NADPH Standard (MW:833.36) 1 vial 1.2 mL 166.7 μg Store kit at -80°C. Ensure that the NADP Cycling Buffer is at room temperature before use. The optimal temperature is 22°C. Keep other enzymes on ice during the assay and protect from light as much as possible. Reconstitute NADPH standard with 200 μl pure DMSO to generate 1 nmol/μl NADPH standard stock solution. 4 Reconstitute NADPH Developer with 1.2 ml of ddH2O. Pipette up and down several times to completely dissolve the pellet into solution. Aliquot enough NADP Cycling Enzyme mix (2 μl per assay) for the number of assays to be performed in each experiment and freeze the stock solution immediately at 80°C for future use. The reconstituted enzymes are stable for up to 2 months at -80°C. B. Additional Materials Required Microcentrifuge Pipettes and pipette tips Colorimetric microplate reader 96 well plate Orbital shaker 5 4. Assay Protocol 1. Sample Preparation: a. For cell samples: Wash cells with cold PBS. Pellet 105 cells for each assay in a microcentrifuge tube (2000 rpm for 5 min). Extract the cells with 200 μl of NADP/NADPH Extraction Buffer by freeze/thaw two cycles (20 min on dry-ice, then 10 min at room temperature), or homogenization. Vortex the extraction for 10 sec. Spin the sample at 14000 rpm for 5 min. Transfer the extracted NADP/NADPH solution into a new labeled tube. b. For tissue samples: Weigh ~20 mg tissue for each assay, wash with cold PBS, homogenize with 400 μl of NADP/NADPH Extraction Buffer in a microcentrifuge tube. Spin the sample at 14000 rpm for 5 min. Transfer the extracted NADP/NADPH solution into a new labeled tube. Note: Cell or tissue lysates may contain enzymes that consume NADPH rapidly. We suggest removal of these enzymes from the sample by filtering the samples through 10 kDa molecular weight cut off filters (ab93349) before performing the assays. 6 2. Standard Curve Preparation: Dilute 10 μl of the 1 nmol/μl NADPH standard with 990 μl NADP/NADPH Extraction Buffer to generate 10 pmol/μl standard NADPH (Note: diluted NADPH solution is unstable, must be used within 4 hours). Add 0, 2, 4, 6, 8, 10 μl of the diluted NADPH standard into labeled 96-well plate in duplicate to generate 0, 20, 40, 60, 80, 100 pmol/well standard. Aliquot remaining standard and store at -20oC. Make the final volume to 50 μl with NADP/NADPH extraction buffer. a. To detect total NADP/NADPH: (NADPt):Transfer 50 μl of extracted samples into labeled 96-well plate in duplicates. b. To detect NADPH only: Aliquot 200 μl samples into eppendorf tubes. Heat samples to 60°C for 30 min in a water bath or a heating block. Under the conditions, all NADP will be decomposed while NADPH will still be intact. Cool samples on ice. Quick spin samples if precipitates occur. Transfer 50 μl of NADPH samples into labeled 96-well plate in duplicates Several sample dilutions should be performed to ensure the reading can be within the standard curve range. 7 4. NADP Cycling Mix: Mix enough reagent for the number of samples and standards to be performed. For each well, prepare a total 100 μl NADP Cycling Mix: NADP Cycling Buffer Mix: NADP Cycling Enzyme Mix: 98 μl 2 μl Mix well and add 100 μl Reaction Mix into each well; mix well. 5. Incubate the plate at room temperature for 5 min to convert NADP to NADPH. 6. Development: Add 10 μl NADPH Developer into each well. Let the reaction develop for 1 to 4 hours. Read the plate at OD450nm. Note: The signal increases with the reaction time. The plate can be read multiple times while the color is developing. The reaction can be stopped by addition of 10 μl Stop Solution each well and mix well. The color should be stable within 48 hours in a sealed plate, after the reactions are stopped. 8 5. Data Analysis Apply the sample OD450nm reading to standard curve. The background reading is very significant in the enzyme cycling reaction, therefore should be subtracted from standard and samples. The amount of NADPt or NADPH can be expressed in pmol/106 cells or ng/mg protein (NADPH molecular weight 745.4). NADP/NADPH Ratio is calculated as: NADPt - NADPH NADPH 9 6. Troubleshooting Problem Reason Solution Assay not working Assay buffer at wrong temperature Assay buffer must not be chilled - needs to be at RT Protocol step missed Plate read at incorrect wavelength Unsuitable microtiter plate for assay Unexpected results Re-read and follow the protocol exactly Ensure you are using appropriate reader and filter settings (refer to datasheet) Fluorescence: Black plates (clear bottoms); Luminescence: White plates; Colorimetry: Clear plates. If critical, datasheet will indicate whether to use flat- or U-shaped wells Measured at wrong wavelength Use appropriate reader and filter settings described in datasheet Samples contain impeding substances Unsuitable sample type Sample readings are outside linear range Troubleshoot and also consider deproteinizing samples Use recommended samples types as listed on the datasheet Concentrate/ dilute samples to be in linear range 10 Samples with inconsistent readings Unsuitable sample type Samples prepared in the wrong buffer Samples not deproteinized (if indicated on datasheet) Cell/ tissue samples not sufficiently homogenized Too many freezethaw cycles Samples contain impeding substances Samples are too old or incorrectly stored Lower/ Higher readings in samples and standards Not fully thawed kit components Out-of-date kit or incorrectly stored reagents Reagents sitting for extended periods on ice Incorrect incubation time/ temperature Incorrect amounts used Refer to datasheet for details about incompatible samples Use the assay buffer provided (or refer to datasheet for instructions) Use the 10kDa spin column (ab93349) Increase sonication time/ number of strokes with the Dounce homogenizer Aliquot samples to reduce the number of freeze-thaw cycles Troubleshoot and also consider deproteinizing samples Use freshly made samples and store at recommended temperature until use Wait for components to thaw completely and gently mix prior use Always check expiry date and store kit components as recommended on the datasheet Try to prepare a fresh reaction mix prior to each use Refer to datasheet for recommended incubation time and/ or temperature Check pipette is calibrated correctly (always use smallest volume pipette that can pipette entire volume) 11 Problem Reason Solution Standard curve is not linear Not fully thawed kit components Wait for components to thaw completely and gently mix prior use Pipetting errors when setting up the standard curve Incorrect pipetting when preparing the reaction mix Air bubbles in wells Concentration of standard stock incorrect Errors in standard curve calculations Use of other reagents than those provided with the kit Try not to pipette too small volumes Always prepare a master mix Air bubbles will interfere with readings; try to avoid producing air bubbles and always remove bubbles prior to reading plates Recheck datasheet for recommended concentrations of standard stocks Refer to datasheet and re-check the calculations Use fresh components from the same kit For further technical questions please do not hesitate to contact us by email ([email protected]) or phone (select “contact us” on www.abcam.com for the phone number for your region). 12 13 14 UK, EU and ROW Email: [email protected] Tel: +44 (0)1223 696000 www.abcam.com US, Canada and Latin America Email: [email protected] Tel: 888-77-ABCAM (22226) www.abcam.com China and Asia Pacific Email: [email protected] Tel: 108008523689 (中國聯通) www.abcam.cn Japan Email: [email protected] Tel: +81-(0)3-6231-0940 www.abcam.co.jp 15 Copyright © 2012 Abcam, All Rights Reserved. The Abcam logo is a registered trademark. All information / detail is correct at time of going to print.
© Copyright 2026 Paperzz