TECHNICAL XX NOTES JOURNAL OF APPLIED BIOMECHANICS, 1998, 14, 93-104 Measurement © 1998 by Humanof Kinetics Pressure Publishers, and Inc. Shear 93 A Device for Simultaneous Measurement of Pressure and Shear Force Distribution on the Plantar Surface of the Foot Brian L. Davis, Julie E. Perry, Donald C. Neth, and Kevin C. Waters A device has been designed to simultaneously measure the vertical pressure and the anterior–posterior and medial–lateral distributed shearing forces under the plantar surface of the foot. The device uses strain gauge technology and consists of 16 individual transducers (each with a surface area measuring 2.5 3 2.5 cm) arranged in a 4 3 4 array. The sampling frequency is 37 Hz and data may be collected for 2 s. The device was calibrated under both static and dynamic conditions and revealed excellent linearity (±5%), minimal hysteresis (±7.5%), and very good agreement between applied and measured loads (±5%). Vector addition of the distributed loads gave resultant forces that were qualitatively very similar to those obtained from a standard force plate. Data are presented for measurements from the forefoot of 4 diabetic subjects during the initiation of gait, demonstrating that distributed shear and pressure on the sole of the foot can be measured simultaneously. Key Words: diabetic foot, plantar pressure, shear measurement, instrumentation Ulceration of the plantar tissue of the foot represents a potentially serious problem for individuals who have sensory neuropathy secondary to diabetes. The principal mechanical factors leading to ulceration are ill-defined. A number of studies (Boulton et al., 1983; Ctercteko, Dhanendran, Hutton, & LeQuesne, 1981; Stokes, Faris, & Hutton, 1975) have shown that sites of ulceration correspond to areas of elevated plantar loading, while others (Pollard & LeQuesne, 1983) have shown that ulceration occurs at sites of maximum shear stress. While high vertical or tangential forces alone could damage skin tissue, there is quite possibly some combination of these forces that is more detrimental than others. Davis (1993) proposed three scenarios in which the interaction of pressure and shear forces could cause skin loading conditions that might contribute to ulcer formation. Although shear is believed to be a significant factor contributing to ulceration (Bauman, Girling, & Brand, 1963; Brand, 1988; Delbridge, Ctercteko, Fowler, Reeve, & LaQuesne, 1985), few researchers have investigated frictional forces due to the difficulty in constructing devices capable of measuring shear (Cavanagh & Ulbrecht, 1991; Masson The authors are with the Department of Biomedical Engineering, The Lerner Research Institute, The Cleveland Clinic Foundation, 9500 Euclid Ave., Cleveland, OH 44195. 93 94 Davis, Perry, Neth, and Waters & Boulton, 1991; Thompson 1983). Most devices have been limited to measuring shear forces in one direction only (Laing et al., 1992; Pollard & LeQuesne, 1983; Pollard, LeQuesne, & Tappin, 1983; Tappin, Pollard, & Beckett, 1980; Tappin & Robertson, 1991) and, thus, likely underestimate the maximum shear stresses. A transducer capable of measuring both shear components on the plantar surface of the foot was developed by Lord, Hosein, and Williams (1992). The transducer measured 16 mm in diameter and 4 mm thick and was placed in an inlay to permit in-shoe measurement. One limitation of discrete transducers is that prior knowledge is required about the areas of interest in order to determine placement of individual transducers. As a result, areas experiencing high stresses may remain undetected if they do not coincide with the areas of interest determined a priori. The transducer used by Lord and colleagues is also capable of measuring pressure if an additional component is added. However, all three stresses were not measured simultaneously because the pressure-measuring component made the transducer too large for in-shoe measurements. More recently, Huo and Nicol (1995) reported on the development of a 3-D force distribution measurement system using strain gauge technology. The device consists of an array of 336 sensors. Although no pressure or shear data were described, the authors reported that the device was suitable for investigating foot loading conditions during the takeoff phase of the high jump. In our laboratory a device has been developed, also based on strain gauge technology, which is capable of simultaneously measuring the distribution of all three force components (vertical, mediolateral, and anteroposterior) on the plantar surface of the foot. Methods Mechanical Design The device has an overall surface area measuring 10.5 cm 3 10.5 cm and is composed of 16 individual transducers. Each transducer measures 2.5 cm 3 2.5 cm and has two components (Figure 1). The hollow cylindrical aluminum column has two sets of T–strain gauge rosettes. Each set is mounted at a different level, and within each set four rosettes are equidistantly spaced around the column. This configuration permits measurement of both mediolateral and anteroposterior shear forces even in the presence of superimposed vertical forces. Each rosette consists of a 90° arrangement of two 350V standard constantan strain gauges (Measurements Group, Inc., Raleigh, NC). Compressive forces are measured with an S-shaped cantilever that fits into the end of each vertical column. The cantilever is instrumented with four rosettes, one centered on each of the four vertical faces (two interior and two exterior faces). Since surface material affects the coefficient of friction between the foot and the contact surface, the device is designed to accommodate different surface materials. Specifically, 2.5 cm squares of material are mounted to thin (3 mm) aluminum base plates that have two small posts on the inferior surface. The top of each transducer is designed with two small holes at opposite corners to accommodate the “caps” of surface material. Sixteen individual transducers are arranged in a 4 3 4 array and anchored in a 2 cm thick aluminum base plate (Figure 2). Spacing between transducers is 1.5 mm. This provides an overall surface area of approximately 10.5 cm 3 10.5 cm. Although this size is not large enough to permit force measurements under the entire plantar surface, it is large enough to examine the forefoot area, which is of primary interest since most diabetic foot ulcers occur in the regions of the metatarsal heads and toes. Measurement of Pressure and Shear 95 Figure 1 — Schematic representation of the transducer. The transducer consists of a hollow column with an S-shaped cantilever, both machined from aluminum (Young’s modulus = 70 GPa). All measurements are in centimeters. An enlarged view of the rosette depicts the orthogonal arrangement of the strain gauges. Figure 2 — The fully assembled shear and pressure device consisting of 16 transducers in a 4 3 4 array. The total surface area measures 10.5 3 10.5 cm with a 1.5 mm space between adjacent transducers. 96 Davis, Perry, Neth, and Waters Electronic Design The electronic interface board (Figure 3) consists of six Wheatstone bridges per post. Four bridges detect the shear forces and two detect the vertical forces. To handle the complexity of sampling from 96 bridges (16 posts 3 6 bridges per post), 32 multiplexors are used and a straight data acquisition method is employed. A 1 V precision regulator is Figure 3 — Diagram of the electronic interface board showing the representative arrangement of the six bridges per post. Bridges A and B measure compression, C and D anterior–posterior shear, and E and F medial–lateral shear. When pressure is applied to the S element, gauges on either side of the vertical sections of the S change resistance in opposite ways (the resistance on the outside gauges increases and the resistance on the inside gauges decreases). The same happens to the gauges on the vertical tube when shear is applied—the resistance increases or decreases depending on whether the gauge is on the left or right of the tube. Therefore, the same Wheatstone bridge configuration is used for both the shear and pressure channels. Measurement of Pressure and Shear 97 used to maintain a constant excitation voltage to all 96 bridges. The positive and negative outputs from each bridge are fed to one of six channels of the analog multiplexors that are used differentially in pairs. Four digital output lines from the computer select the bridge from which data will be acquired. The acquisition sequence first reads Bridge 1 on each of the 16 transducers, then Bridge 2, and so on until Bridge 6 is read. This permits data to be sampled at a frequency of 37 Hz for 2 s. The output from each pair of multiplexors is fed into an amplifier with a gain of 500. The output from each amplifier is then sent to one of 16 analog-to-digital converters. Each multiplexor is controlled via software that has been written to ensure reliable acquisition and storage of data. Device Calibration Each individual transducer was calibrated under static conditions. For both loading and unloading conditions, calibration was carried out under compressive loads (0, 31.15, 75.65, 120.15, 164.65 N) and both anteroposterior and mediolateral loads (0, 22.25, 44.50, and 66.75 N). Two trials of loading and unloading were conducted. During compressive tests, the weights were placed directly on top of the transducer. Shear loads were applied by hanging weights from a cord attached at one end to the transducer and passed over a simple pulley system. More in-depth testing and calibration were then carried out on one representative transducer in the array. A series of seven different tests was conducted: 1. A compressive load was applied through point loading to the center of the transducer and to each of the corners at separate times. A load of 111.25 N was clamped to a board through which three nails were driven. The load was then positioned over the transducer such that one nail was positioned on the transducer in the region of interest and the other two nails rested on a surrounding support surface. The compressive load measured by the transducer was then recorded under these five different loading conditions (center and the four corners). 2. Compressive loads from 31.15 to 164.65 N were applied to the transducer, and the cross-talk present on the shear force channels was measured. 3. Shear loads ranging from 22.25 to 66.75 N were applied to the transducer, and the cross-talk present on the compression channels was recorded. 4. Shear forces were applied 1 cm off-center to create a torsional force, and the effect on both compression and shear was examined. Torsional forces ranged from 0 to 77.88 N. 5. Compressive (0–164.65 N) and shear (0–66.75 N) forces were applied simultaneously. 6. A force that was the vector addition of mediolateral and anteroposterior forces was applied. The load, ranging from 0 to 77.88 N, was applied along an axis midway between the mediolateral and anteroposterior axes. 7. A load cell (Sensotek, Columbus, OH; Model 13/2443-08, range 0–225 N) capable of measuring pressure dynamically was used to apply cyclical loads to the transducer. The range of loading was 0 to 60 N at a rate of approximately 3 Hz and 0 to 100 N at approximately 1 Hz. Data were collected simultaneously from the Sensotek load cell (370 Hz) and the shear and pressure device (37 Hz). As a secondary check of the quality of the transducer array, the force profiles obtained from the device were compared to those from an AMTI force plate (AMTI Inc., Newton, MA) for one individual. These were two separate trials: one stepping off the force plate and one stepping off the shear and pressure device. The manner of testing was 98 Davis, Perry, Neth, and Waters the same as for the subjects (see below). Although calibration of the device was the primary means of validation, this comparison with the force plate profile provided a secondary check of the quality of the output from the shear and pressure device under dynamic conditions. Preliminary Subject Trials Data were collected on the right foot of 4 individuals with diabetes. Two of the subjects had sensory neuropathy (means: age = 73, height = 1.8 m, weight = 69 kg, vibration score [Vibratron II] = 16.43, could not feel 6.10 Semmes-Weinstein monofilament), and 2 had protective sensation (means: age = 65, height = 1.6 m, weight = 91 kg, vibration score = 6.20, could feel 5.07 monofilament). The array of transducers was set flush in a 92 cm 3 122 cm platform, and caps of Minorplast (a material commonly used in therapeutic shoes) were placed on each of the transducers. A 5 mm thick pad of rubber matting was placed on the wooden platform surrounding the transducer array. The subjects were tested in their bare feet and stood with their right forefoot on the transducers. Walking was initiated with the left leg, and the individuals took two to three steps. In this manner the forces under the right forefoot were recorded during the initiation phase of walking for three trials. The purpose of these trials was not to make group comparisons but to show the utility of the shear and pressure device and provide representative subject data in addition to calibration data. Results Calibration Trials For each of the 16 transducers, a regression equation was determined for applied versus measured loads for compression, anteroposterior shear, and mediolateral shear under static loading conditions. A 99% confidence interval was determined for the slopes, and the ranges were as follows: compression (0.981–1.03), anteroposterior (0.996–1.00), and mediolateral (0.997–1.00). Compressive results had slightly more variation than shear, but the widths of the confidence intervals were small and in close proximity to the line of identity (a value of 1.0). Calibration plots (Figure 4) are presented for the four conditions that introduced noise into the system. Even with the possible presence of cross-talk, the data showed very good linearity and excellent agreement between applied and measured loads. The quality of the data was not degraded by the introduction of torsion, the combination of mediolateral/ anteroposterior forces, or combined compressive/shear forces. With the present design, forces in all three directions can be measured with accuracies that are within 5% of the true value. For each of the plots in Figure 4, linearity values were determined as a percentage of the full-scale measurable output. These values, corresponding to Figure 4a–4d, respectively, were ±4.5%, ±5.1%, ±2.4%, and ±4.2%. The values for the first three plots represent the maximum error present, whereas in the fourth condition the value is the average error. This fourth condition measured the effect of torsion on compression, and we do not feel that the maximum error calculated as ±12% provides a realistic representation of the limitations of the device. (Torsion was created by applying a shear load at the extreme corner of the transducer, and it is highly unlikely that loading conditions under the foot would replicate this unusual loading scenario.) For the point-loading test, in which vertical point loads were applied at the extreme corners of the sensor, the highest reading was 4.8% (1.3 N) greater than the load Measurement of Pressure and Shear 99 Figure 4— Calibration plots for the conditions that introduce cross-talk into the system: (a) vector addition of medial–lateral and anterior–posterior forces, (b) effect of compression on shear measurements, and the effect of torsion on (c) shear and (d) compression. Data correspond to three trials of loading (solid lines) and unloading (dashed lines). Details of loads are given in the Device Calibration section of the manuscript. measured in the center of the sensor. The lowest reading was 14% (3.8 N) lower than the central reading of 27.2 N. When vertical loads were applied to the transducer, the average cross-talk on the shear channels was 5.2% (3.2 N) of the applied load; with anteroposterior loading, the average cross-talk on the vertical channels was 13.3% (4.8 N); and during mediolateral loading, the average cross-talk was 4.3% (2.3 N) of the applied load. Dynamic calibration of a single transducer with the Sensotek load cell (Figure 5) resulted in an average difference of 3 N between the load cell and the pressure transducer. This difference corresponded to 3% and 5% of the maximum applied loads under 1 Hz and 3 Hz nominal loading rates, respectively. From the 3 Hz data, the hysteresis of the transducer was determined to be ±7.4%. Because technical limitations did not permit dynamic calibration at greater than 3 Hz, a Fourier transform was performed on the data to determine the frequency content of the applied load. A signal-to-noise ratio of 20:1 was used as the cutoff for meaningful signal content. With this criterion, the maximum frequency contained in the signal was 14 Hz. The secondary check of the device, as a whole, under dynamic conditions revealed that the force profiles obtained from the device qualitatively resembled very closely the data from the AMTI force plate (Figure 6). 100 Davis, Perry, Neth, and Waters Figure 5 — Dynamic performance of a transducer in the shear and pressure device when loaded at rates of approximately (a) 1 Hz and (b) 3 Hz using a Sensotek load cell. The average difference between the load cell and the transducer was 3 N. Subject Trials The sites of maximum pressure and shear were the same for some of the subjects; however, for the entire group the sites of maximum shear and pressure did not correspond (Figure 7). The highest pressure occurred in the region corresponding to the medial metatarsal heads (Area 3, S2), whereas the site of highest shear force was in the lateral metatarsal heads (Area 3, S3). The greatest difference between shear and pressure values occurred in Region 3, S2. The average magnitude of the shear forces was approximately 25% of the compressive loads. In terms of the direction of shear forces at adjacent sites (not shown), the peak value of the difference between forces directed away from each other was about twofold greater than the difference between forces directed toward each other, indicating a greater tendency for the tissue to be stretched than bunched (Perry, Davis, & Hall, 1995). Discussion Identifying sites of maximum shear and pressure is important in working with diabetic individuals because ulceration of the foot is most likely to occur at these sites (Boulton et Measurement of Pressure and Shear 101 Figure 6 — Comparison of force profiles in all three planes for the shear and pressure device and an AMTI force plate. Qualitatively, the profiles are nearly identical. Data represent one trial on each device. al., 1983; Cavanagh & Ulbrecht, 1991). In contrast to shear devices that have been used in previous investigations, the current device is capable of measuring true shear (the resultant of mediolateral and anteroposterior forces) rather than being limited to measuring only one component of the shear force. This permits a more accurate and representative analysis of the frictional forces acting on the plantar surface of the foot. Furthermore, with the current device, it is now possible to examine the combined effects of distributed vertical and shear forces. As mentioned previously, some authors have shown that ulceration 102 Davis, Perry, Neth, and Waters Figure 7 — Magnitude of peak pressure and shear results obtained from 4 individuals with diabetes. The vertical axes refer to force in newtons. The other axes define the array of transducers. Block (1, S1) represents the hallux; (4, S1) the 1st metatarsal head, and (4, S4) the fifth metatarsal head. Note that the scales are different for the pressure and shear plots, and the direction of the shear forces is not depicted. occurs at sites of maximum shear, while others have shown that it corresponds to sites of maximum pressure. Since all three force components can be measured simultaneously, this issue can now be investigated in greater detail. One major limitation of the current design is the small overall size of the device combined with the large individual transducer area of 6.25 cm2. The large transducer area likely results in underestimations of true peak pressures and limits the ability to discriminate between adjacent areas of high stresses. The small overall size of the transducer array restricts the area of analysis to only a portion of the plantar surface of the foot and limits the activities under which forces may be investigated. Furthermore, the device is not suitable for in-shoe measurements. The frequency response of a system is also an important aspect of system design. Frequency response can be approximated by considering each individual transducer element as a simple cantilever with a mass at one end (i.e., the aluminum column with the Sshaped mass at one end). The resonant frequency of the unloaded device is calculated to be 350 Hz. This suggests that the device can be loaded under activities similar to those typically used with force plates (the resonant frequency of the AMTI force plate is about 500 Hz). However, it is necessary to consider that the sampling frequency of the device is currently limited to 37 H, which limits the loading responses that can be investigated. Thermal stability of the device is maintained through the Wheatstone bridge arrangement. The four gauges (per bridge) are mounted on the same piece of aluminum so changes in temperature should affect all gauges similarly. Two gauges are mounted in parallel with the axis of strain application, while two gauges are mounted orthogonal to the axis. This orthogonal arrangement compensates for changes in temperature. Additionally, it has been demonstrated that the device behaves very linearly, there is good agreement between applied and measured loads, and there is minimal hysteresis. These Measurement of Pressure and Shear 103 characteristics are typical of transducers that are (a) constructed from aluminum or steel, which behave in a linear manner up to the yield point and (b) strain-gauged according to the guidelines prescribed by the gauge manufacturer (Measurements Group, Inc., Raleigh, NC). Conclusions With the development of the current device, we have demonstrated that loading under the feet can be simultaneously measured in three dimensions (vertical, mediolateral, and anteroposterior) with accuracies to within 5% of the true value. The ability to quantify loading under the foot in this manner has the potential to (a) fill a void in the clinical literature by allowing researchers to investigate complex loading conditions at the foot/ground interface and (b) address limitations in the biomechanical literature by better defining boundary conditions that are important in finite element models. References Bauman, J.H., Girling, J.P., & Brand, P.W. (1963). Plantar pressures and trophic ulceration: An evaluation of footwear. Journal of Bone and Joint Surgery, 45B, 652-673. Boulton, A.J.M., Hardisty, C.A., Betts, R.P., Franks, C.I., Worth, R.C., Ward, J.D., & Duckworth, T. (1983). Dynamic foot pressure and other studies as diagnostic and management aids in diabetic neuropathy. Diabetes Care, 6, 26-33. Brand, P.W. (1988). Repetitive stress in the development of diabetic foot ulcers. In M.E. Levin & L.W. O’Neal (Eds.), The diabetic foot (4th ed., pp. 83-90). St. Louis, MO: Mosby. Cavanagh, P.R., & Ulbrecht, J.S. (1991). Biomechanics of the diabetic foot: A quantitative approach to the assessment of neuropathy, deformity and plantar pressure. In M.H. Jahss (Ed.), Disorders of the foot and ankle (2nd ed., pp. 1864-1907). Philadelphia: Saunders. Ctercteko, G.C., Dhanendran, M., Hutton, W.C., & LeQuesne, L.P. (1981). Vertical forces acting on the feet of diabetic patients with neuropathic ulceration. British Journal of Surgery, 68, 608614. Davis, B.L. (1993). Foot ulceration: Hypotheses concerning shear and vertical forces acting on adjacent regions of skin. Medical Hypotheses, 40, 44-47. Delbridge, L., Ctercteko, G., Fowler, C., Reeve, T.S., & LeQuesne, L.P. (1985). The aetiology of diabetic neuropathic ulceration of the foot. British Journal of Surgery, 72, 1-6. Huo, M., & Nicol, K. (1995). 3-D force distribution measuring system. In K. Hakkinen, K.L. Keskinen, P.V. Komi, & A. Mero (Eds.), XVth Congress of the International Society of Biomechanics: Book of Abstracts (pp. 410-411). Jyväskylä, Finland: Gummerus. Laing, P., Deogan, H., Cogley, D., Crerand, S., Hammond, P., & Klenerman, L. (1992). The development of the low profile Liverpool shear transducer. Clinical Physics and Physiological Measurement, 13, 115-124. Lord, M., Hosein, R., & Williams, R.B. (1992). Method for in-shoe shear stress measurement. Journal of Biomedical Engineering, 14, 181-186. Masson, E.A., & Boulton, A.J.M. (1991). Pressure assessment methods in the foot. In R.G. Frykberg (Ed.), The high risk foot in diabetes mellitus (pp. 139-149). New York: Churchill Livingstone. Perry, J.E., Davis, B.L., & Hall, J.O. (1995). Profiles of shear loading in the diabetic foot. In K. Hakkinen, K.L. Keskinen, P.V. Komi, & A. Mero (Eds.), XVth Congress of the International Society of Biomechanics: Book of Abstracts (pp. 722-723). Jyväskylä, Finland: Gummerus. Pollard, J.P., & LeQuesne, L.P. (1983). Method of healing diabetic forefoot ulcers. British Medical Journal, 286, 436-437. Pollard, J.P., LeQuesne, L.P., & Tappin, J.W. (1983). Forces under the foot. Journal of Biomedical Engineering, 5, 37-40. Stokes, I.A.F., Faris, I.B., & Hutton, W.C. (1975). The neuropathic ulcer and loads on the foot in diabetic patients. Acta Orthopaedica Scandinavica, 46, 839-847. 104 Davis, Perry, Neth, and Waters Tappin, J.W., Pollard, J., & Beckett, E.A. (1980). Method of measuring ‘shearing’ forces on the sole of the foot. Clinical Physics and Physiological Measurement, 1, 83-85. Tappin, J.W., & Robertson, K.P. (1991). Study of the relative timing of shear forces on the sole of the forefoot during walking. Journal of Biomedical Engineering, 13, 39-42. Thompson, D.E. (1983). Pathomechanics of soft tissue damage. In M.E. Levin & L.W. O’Neal (Eds.), The diabetic foot (3rd ed., pp. 148-161). St. Louis, MO: Mosby. Acknowledgments Construction of the shear and pressure device was financially supported by an NIH-NIAMS39750 Center Grant and by Thor-Lo, Inc., Statesville, NC.
© Copyright 2025 Paperzz