A59440 AA July 2008 Beckman Coulter Dissolved Oxygen Probe A57199 How the Dissolved Oxygen Probe Works The Beckman Coulter Dissolved Oxygen (DO) Probe is a galvanic electrochemistry device and does not require power from the meter or controller to generate its signal. The DO probe will provide a millivolt signal proportional to the concentration of oxygen in the water. A DO probe does not measure oxygen directly. It measures the partial pressure of oxygen in the water, which is directly proportional to the percentage of oxygen saturation in the water. The concentration (mg/L or ppm) can be calculated based upon the oxygen solubility, temperature, salinity, and total atmospheric pressure. The DO probe consists of a cathode, anode, and electrolyte separated from your process fluid by an oxygen-permeable membrane. The oxygen passing through the membrane reacts with the cathode, giving up electrons, which produces an electrical current. Reconditioning the Probe 1. Remove the pre-membraned cap from the tip of the probe by turning it counterclockwise and discard, as shown in Figure 1a. 2. Open the tip on the bottle of DO electrolyte solution. Place the needle on the plastic bottle, as shown in Figure 1b. 3. Insert the needle into one of the four holes surrounding the silver cathode. Squeeze the bottle to inject the solution until it leaks out of a fill hole, as shown in Figure 1c. 4. Replace the cap on the probe. Figure 1. Reconditioning the Probe b a c Inject until solution leaks out of fill hole (approx. 5 mL) Calibrating the Probe The simplest method to calibrate the Dissolved Oxygen probe is in saturated air. If the water is air saturated, the partial pressure of air in the water will be the same as it is in air. Therefore, the millivolt output of the DO probe in air corresponds to 100% saturation in water. Be sure to remove water drops from membrane when calibrating in air. To calibrate the DO probe in line, bubble oxygen or air into your process fluid to oxygen saturate the fluid. The millivolt output of the DO probe under these conditions will correspond to 100% saturation. If a two-point calibration is needed, a saturated solution of sodium sulfite will provide a zero oxygen environment. The DO probe will take several minutes to reach zero millivolts after it is submerged into the saturated sodium sulfite. Please see the calibration instructions in the meter user’s guide for additional details. ©2008 Beckman Coulter, Inc. 4300 North Harbor Boulevard, Fullerton, CA92834-3100 Printed in USA. Turn counterclockwise to remove cap Storing the Probe If long-term storage of the DO probe is required, remove the pre-membraned cap and probe, leaving only a little solution on the membrane to keep it wet. Leaving a full volume of electrolyte in the DO probe long-term without use will deplete the anode. See the Reconditioning the Probe section for instructions on how to open and close the probe housing. Probe Specifications Output at 100% saturation: 36-54 mV Output at 0% saturation: <1 mV Temperature Range: Maximum 50°C Minimum 0°C Accuracy: ± 2% when measuring temp. equal to cal temp. Response time: 5 minutes to reach 95% of final reading Water flow rate: Wetted Materials: Body Membrane Min. 2 in./sec. across membrane WARNING Always follow laboratory safety procedures, including wearing protective gloves and eye wear. Avoid consuming, inhaling, or coming into direct contact with chemicals or solutions. 警告 始终按照实验室安全程序操作,包括佩戴防护手套和 眼罩。避免吞食、吸入或直接接触化学物质或溶液。 NOTE Epoxy HDPE Dissolved Oxygen Probe and Accessories A57199 Dissolved Oxygen Probe A58726 Dissolved Oxygen Refill Solution A58728 Replacement Cap with Membrane If you experience problems with your Beckman Coulter electrode that these procedures cannot resolve, call your local Beckman dealer. In the U.S., please call the toll-free Product Support line at (800) 742-2345. NOTE For MSDS information, please visit www.beckmancoulter.com. Click on customer support, then MSDS. NOTE Please visit www.beckmancoulter.com for a complete list of electrochemistry meters, electrodes, and accessories.
© Copyright 2026 Paperzz