ARTICLE IN PRESS Radiation Physics and Chemistry 71 (2004) 713–715 X-ray radiation source for total dose radiation studies D. Biselloa,*, A. Candeloria, A. Kaminskia, A. Litovchenkoa, E. Noahb, L. Stefanuttia a Dipartimento di Fisica, INFN Sezione di Padova, Universita di Padova, Via Marzolo 8, Padova I-35131, Italy b Imperial College, London SW7 2AZ, UK 1. Introduction 3. The X-ray emission field Low-energy (p60 keV) X-ray tube sources are currently used for investigating the tolerance to ionizing radiation in total dose tests of unpackaged semiconductor devices and circuits, due to their higher dose rates (100–1000 rad(Si)/s) and lower costs than 60Co g-ray sources (Palkuti and LePage, 1982; ASTM), and to the radioprotection advantage that the radiation can be switched off. In this study we report on the X-ray field emission characteristics, measured by silicon diodes, of the irradiation facility installed at the INFN National Laboratory of Legnaro (Padova, Italy) for total dose tests of electronics devices and circuits. X-ray emissions from tungsten anodes have been widely investigated in the literature. For operating voltages between 20 and 60 kV, they consists of the bremsstrahlung component with an energy spectrum up to Y keV, where Y is the tube operating voltage in kV, plus the multi-line L characteristic radiation of tungsten, which contains 12 lines for the La ; Lb ; and Lg series between 8.3 and 11.7 keV (Birch et al., 1982). The X-ray spectra emitted from the tube is additionally filtered for total dose tests by 0.15 mm Al in order to attenuate the low-energy (o8 keV) component (ASTM). An example of X-ray spectra calculated by XCOMP (Nowotny and Hofer, 1995) for a typical tube operating condition in total dose tests is shown in Fig. 1, where both the bremsstrahlung and characteristic radiation components are taken into account. The X-ray emission field from the tube has been measured by a square silicon diode calibrated in accordance to the ASTM recommendations, with active area A¼ 5 5 mm2 and thickness d ¼ 300 mm, positioned on the chuck of a commercial semiautomatic probe station under the X-ray tube. Silicon diodes are considered for X-ray dose rate measurements in total dose tests because silicon is the reference material in microelectronics devices. During the X-ray field measurements, the diode was reverse biased at full depletion and the active area was grounded, as well as the guardring contact to prevent border effects on the current induced by X-rays in the diode active volume V=A d. The dose rate from the X-ray tube at the diode surface referred to silicon (Dr ) is determined by Dr ¼ CIx =ðqEeh VrÞ; where C is a calibration constant to report the average dose rate in the diode active volume to the surface value (Noah, 2000), which 2. The X-ray irradiation system The total dose test facility is the RP-149 Semiconductor Irradiation System from Seifert (Ahrensburg, Germany) equipped with a standard tube for X-ray diffraction analysis: maximum power 3000 W, maximum voltage 60 kV, tungsten anode, 0.3 mm Be inherent filtration, take off angle 6 , and focus dimensions 0.4 12 mm2. The tube voltage and current can be set with kV and mA resolution, respectively. The tube is located inside a shielding cabinet and it can be moved with an accuracy of 30 mm along the x-, y- (motorized) and z- (manual) axis of 15, 20 and 450 cm, respectively, for accurate positioning on the tested device. *Corresponding author. Tel.: +39-049-8277216; Fax: +39049-8277237. E-mail address: [email protected] (D. Bisello). 0969-806X/$ - see front matter r 2004 Elsevier Ltd. All rights reserved. doi:10.1016/j.radphyschem.2004.04.071 ARTICLE IN PRESS D. Bisello et al. / Radiation Physics and Chemistry 71 (2004) 713–715 714 Photon spectra normalized to the maximum value 1.0 0.8 0.6 0.4 0.2 0.0 0 10 20 30 40 50 Energy (keV) Fig. 1. X-ray spectrum normalized to the maximum value calculated by XCOMP (Nowotny and Hofer, 1995), anode material W, take-off angle 6 , tube voltage 50 kV, inherent filtration 0.3 mm Be, additional filtration 0.15 mm Al, distance from the tube aperture 15 cm. 140 Dose rate (rad(Si)/s) 120 100 13.9 mm 80 60 15.5 mm 40 16.1 mm 20 20.6 0 -20 -15 -10 (a) -5 0 5 X position (mm) 10 15 20 10 15 20 140 Dose rate (rad(Si)/s) 120 7.2 mm 100 80 60 9.2 mm 40 11.4 mm 20 19.4 mm 0 -20 (b) -15 -10 -5 0 5 Y position (mm) Fig. 2. Dose rate from the X-ray tube as a function of the position along the x-direction perpendicular (a) and the y-direction parallel (b) to the anode-cathode line at different distances from the tube aperture: 10 cm (squares), 15 cm (diamonds), 20 cm (triangles), and 40 cm (circles). Experimental conditions: tube voltage 50 kV, tube current 10 mA, filtration 0.15 mm Al, take off angle 6 . The horizontal lines delimited by two open circles correspond to the intervals for the 10% dose rate variation with respect to the maximum value. ARTICLE IN PRESS D. Bisello et al. / Radiation Physics and Chemistry 71 (2004) 713–715 accounts for the dose profile in the silicon diode as a function of the depth; Ix is the diode current induced by X-rays obtained by subtracting the diode current with and without the X-ray radiation; q is the electron charge; Eeh ¼ 3:6 eV is the mean energy to create and electron hole pair in silicon; r is the silicon density. The X-ray dose rate fields from the tube operating at 50 kV and 10 mA, measured by moving the probe station chuck on which the silicon diode is positioned, along the x- (perpendicular to the anode–cathode line) and y- (parallel to the anode–cathode line) axes, are shown in Fig. 2. Measurements have been repeated at different distances between the tube aperture (E11 cm down to the tube focus) and the diode surface. As expected, the X-ray emission is symmetric along the direction perpendicular to the anode–cathode line, but unsymmetrical along the anode–cathode direction due to the standard X-ray tube geometry: electrons impact perpendicularly on the anode and induce a cylindrical (not spherical) symmetric photon emission. The effect of the take-off angle (6 ) is seen on the position of the maximum emission along the AC directions, which slightly decreases (from 1 to 4 mm) by increasing the distance from the tube. Data in Fig. 2 shows that uniform irradiation with 10% accuracy can be obtained on 13.9 7.2 mm2 (20.6 19.4 mm2) area at 10 cm (40 cm) from the tube aperture, even if the dose rate 715 is reduced by a factor 8.6 in this distance range. Depending on the DUT dimension, the dose rate can be further increased (decreased) by decreasing (increasing) the distance from the source. Finally, we remark that the 10% accuracy estimation is a worst case condition, because the diode averages the dose rate on its 5 5 mm2 active area when positioned at (x,y). References American Society for Testing and Materials (ASTM). Standard guide use of an X-ray tester (E10 KeV photons) in ionizing radiation effects testing of semiconductor devices and microcircuits. ASTM designation F 1457–99. Birch, R., Marshall, M., Peaple, L.H.J., 1982. Theoretical and measured L X-rays from a solid tungsten target. Phys. Med. Biol. 27, 1119–1129. Noah, E., 2000. X-ray dosimetry: determination of ionization dose delivered by the Seifert Rp-149 X-ray machine. Imperial College, London, UK. Nowotny, R., Hofer, A., 1995. Ein program fur die berechung von diagnostiscen Rontgenspekten. Fortschr. Rontgenstr., 142, 685-689. Available on line: ftp.bmtp.akh-wien.ac.at/ BMTP/xray/. Palkuti, L.J., LePage, J.J., 1982. X-ray wafer probe for total dose testing. IEEE Trans. Nucl. Sci. 29, 1832–1837.
© Copyright 2026 Paperzz