Protoplasma (1999) 209:126-131 PROTOPLASMA 9 Springer-Verlag 1999 Printed in Austria Dynamic studies of phloem and xylem flow in fully differentiated plants by fast nuclear-magnetic-resonance microimaging Rapid communication M. Rokitta 1'*, A. D. Peuke 2, U. Zimmermann 3, and A. Haase ~ Lehrstuhl f/jr Experimentelle Physik V, Physikalisches Institut, 2Julius-von-Sachs-Institut f/Jr Biowissenschaften, and 3Lehrstuhl f/jr Biotechnologie, Biozentrum, Universitfit W/jrzburg, W/Jrzburg Received February 9, 1999 Accepted May 5, 1999 Summary. A fast nuclear-magnetic-resonance imaging method was developed in order to measure simultaneously and quantitatively the water flow velocities in the xylem and the phloem of intact and transpiring plants. Due to technical improvements a temporal resolution of 7 rain could be reached and flow measurements could be performed over a time course of 12-30 h. The novel method was applied to the hypocotyl of 35- to 40-day-old, leafy plants of Ricinus comrnunis which were subjected to different light-dark regimes. The results showed that the xylem flow velocities and the xylem volume flow responded immediately to light on-off changes. Upon illumination the flow velocity and the volume flow increased as expected in respect to literature. In contrast, the phloem flow velocity did not change in response to the light-dark regimes. Interestingly, though, the volume flow in the phloem increased during darkness. These findings can be explained by assuming that the conducting area of the phloem becomes enlarged during the dark period due to opening of sieve pores. Keywords: Phloem flow; Xylem flow; Water transport; Nuclearmagnetic-resonance imaging; Ricinus communis L. Introduction N u c l e a r magnetic r e s o n a n c e ( N M R ) imaging is a noninvasive and nondestructive technique and provides a great variety of contrast m e c h a n i s m s such as chemical shift, relaxation times, diffusion or flow velocities. Recently, flow-weighted N M R imaging was used to obtain quantitative patterns of the flow velocity in the xylem of (intact) plants (Xia et al. 1993, K u c h e n b r o d et al. 1996). D u e to the small d i a m e t e r of the vessels (up to 50 ~tm) and flow velocities of less than 1 mm/s sophisticated e q u i p m e n t and pulse sequences are required to achieve a reasonable signal-to-noise ratio in the N M R images. P h l o e m flow is even m o r e difficult to m o n i t o r because the d i a m e t e r of the sieve tubes is significantly smaller than that of the xylem vessels and, thus, the total v o l u m e flow in the p h l o e m is m u c h less than that in the xylem. Despite these difficulties, K 6 c k e n b e r g e r et al. (1997) recently r e p o r t e d m e a s u r e m e n t s of the p h l o e m flow in R i c i n u s c o m m u n i s seedlings by N M R flow-imaging techniques. H o w e v e r , due to the dimensions of the m a g n e t bore, only 6-day-old (nontranspiring) seedlings could be investigated. M o r e o v e r , the total time for m e a s u r i n g a single flow data set was 4.5 h which m a d e it impossible to detect d y n a m i c flow changes in response to a light-dark regime. In this c o m m u n i c a t i o n , we will present m e a s u r e ments of p h l o e m flow t o g e t h e r with xylem flow in leafy, transpiring R. c o m m u n i s plants. D u e to technical i m p r o v e m e n t s the t e m p o r a l resolution was only 7 rain which was sufficient for d y n a m i c (functional) studies. This could be shown by m e a s u r e m e n t s of the response of the xylem and p h l o e m flow u p o n light-dark regimes. Material and methods *Correspondence and reprints: Physikalisches Institut, EP 5, Universitfit W/jrzburg, Am Hubland, D-97074 W/jrzburg, Federal Republic of Germany. The experiments were performed with 35- to 40-day-old R. communis L. plants, The cultivation of the plants was similar to the M. Rokitta et al.: Phloem and xylem flow in fully differentiated plants 127 f superconductingmagnet (Biospec70/20) halogen lamp 9 9 Q Q 9 9 9 9 9 9 9 9 9 9 9 9 9 j f air flow1 window O Q Q Q Q Q O 9 O O Q Q Q Q Q Q flesh air gas exchange system (CMS 400) | j timer 1. Schematic diagram of the setup used for the high-resolution NMR microimaging of the xylem and ]phloem flow velocities in intact and transpiring plants under controlled environmental conditions. For detailed explanations, see text Fig. procedure described by Peuke et al. (1994), except that the NMR equipment required some modifications. In order to fit the hypocotyl to the radio frequency (RF) coil, the seedlings were kept in darkness for a couple of days after germination. Under these conditions hypocotyls with a length of 15-20 cm were obtained. Then the seedlings were transferred to plastic pots filled with quartz sand. (The dimensions of the pots corresponded to the diameter and length of the magnetic bore.) The plants were watered daily with a nutrient solution containing 1.33 mol Ca(NO3)a and 1.33 mol K N O 3 per m 3 as N-source and were illuminated artificially [16 h light, 8 h dark regime; 300-500 gmol photon per (m 2. s)]. For the NMR measurements a leafy plant was installed in a home-built probe head with the transmitter/receiver RF coil about 8 cm above the roots (see Fig. 1).The probe head was sealed air tight in order to monitor the changes of the water vapour content and COz concentration in the air by a commercial gas exchange system (Walz, Effeltrich, Federal Republic of Germany). Illumination of the plant was performed with a halogen lamp through the front window of the probe head. The maximum light intensity (irradiance) was about 500 gmol photon per (m2. s). The temperature and the relative humidity were kept at about 25 ~ and 50%, respectively. The probe head was inserted into a Biospec 70/20 horizontal bore magnet (Bruker, Rheinstetten, Federal Republic of Germany). Magnetization-prepared NMR imaging (Bourgeois and Decorps 1991, Haase et al. 1993) was used to combine a very sensitive flowweighting sequence with a fast, high-resolution NMR microimaging technique. A phase contrast method employing stimulated echoes (Tanner 1970) combined with a suppression of the signals arising from stationary water (Lafirech et al. 1987, Bourgeois and Decorps 1991) was used for flow weighting. The imaging part of the sequence was a refocused FLASH (fast low angle shot) experiment (Haase et al. 1986). Thus, a relatively high spatial resolution (minimum 47 x 47 gm in plane, 3 mm slice thickness) as well as a high temporal resolution was obtained. A complete flow measurement took 7 min consisting of 16 averages. This resulted in a satisfactory signalto-noise ratio. Eight images were acquired with increasing flowencoding gradients. From this data set the flow velocities and signal amplitudes were calculated by a pixel-by-pixel fit of an appropriate function to the signal intensities of the images. In addition to the velocities, the upward or downward direction of the flows could be extracted from the above analysis as described in detail by Rokitta et al. (1999). The flow velocities given below represent average values of the conducting areas of the phloem and the xylem, respectively. However, for the evaluation of the xylem velocity only the areas of the primary vascular bundles were taken into account because they appeared uniformly in the NMR images of different plants in contrast to other structural conducting elements (see Fig. 2). It should also be noted that for the averaging of the data points in the selected phloem and xylem areas only data points were included which exhibited nonzero values for the flow velocity. Since the lowest detection limit for the flow velocity was about 0.2 mm/s (Rokitta et al. 1999), this procedure leads to a slight overestimation of the average flow velocity. For the estimation of the total volume flow from the average flow velocity values the total geometric conducting area of the xylem and phloem, respectively, must be known. However, these parameters are not available. The problem can be solved in the following way. An image pixel was generally larger than a single vessel as indicated in Fig. 2c.This means that for the calculation of the conducting area in this particular pixel the ratio of the regions with flowing and with nonflowing (stationary) water must be taken into account. A reasonable measure for this ratio is the signal amplitude derived from the fitting procedure mentioned above. Thus, the total volume flow can be estimated by multiplication of the average flow velocity value of each pixel with the corresponding signal amplitude and by subsequent summation. However, this procedure could lead to an error because regions of stationary water comprise both vessels with flow velocities below the detection limit as well as tissue cells. At present, it is not possible to discrimate between these two compartments and thus to quantify the volume flow in absolute units. Therefore, the values of the volume flows given below were normalized to the maximum value of each data set. For the morphological identification of the conducting elements in the FLASH images and flow maps a few spin echo images were made (for conditions, see legend to Fig. 2) as well as microscopical images (Axiophot; Zeiss, Oberkochen, Federal Republic of Germany). Results Figure 2 shows a typical FLASH i m a g e (a; 47 x 4 7 g m in-plane resolution, 3 mm slice thickness) and the corresponding flow velocity map (b; 78 x 313 gm in-plane resolution, 4 mm slice thickness) of a cross section of 128 M. Rokitta et al.: Phloem and xylem flow in fully differentiated plants Fig. 2. Cross sections of the shoot of 35- to 40-day-old intact and transpiring plant of R. communis made by NMR microimaging (a, b, and d) and light microscopy (c). a FLASH image (echo time, TE=4.3 ms; repetition time, TR= 11.5ms; field of view, FOV= 6 x 6 mm2; matrix, 128 x 128 zerofilled to 256 x 256; slice thickness, 3 ram; measuring time, 2 s; nominal in-plane resolution, 47 x 47 ~tm2).b Flow velocity map (matrix, 128 x 32 zerofilled to 128 x 128; slice thickness, 4 mm; FOV = 10 x 10 mm2reduced to 6 x 6 ram2;measuring time, 7 rain; nominal in-plane resolution, 78 x 313 gmZ). c Light microscopy; the size of one pixel in NMR flow imaging (b) is indicated in the upper right corner of the picture, d Spin echo image (TE = 15 ms; TR = 0.5 s; slice thickness, 6 mm; FOV = 8 x 8 mm2 reduced to 6 x 6 ram2; measuring time, 4.8 h; 64 averages; nominal in-plane resolution, 16 x 16 gm2). For further explanations, see text a R. c o m r n u n i s plant. The F L A S H image reflects the water distribution and thus the a r r a n g e m e n t of the structural elements in the cross section. Since b o t h images were m a d e at the same position of the shoot the flow image was projected on the F L A S H image in Fig. 2 b in o r d e r to facilitate the correlation b e t w e e n the flow p a t t e r n and the structural architecture. For unequivocal assignment of the structural and flow elements revealed by the N M R images to the m o r p h o logical elements seen usually in the light m i c r o s c o p e a typical microscopical cross section of the shoot of a plant is given additionally in Fig. 2c. C o m p a r i s o n of the F L A S H image with the light m i c r o s c o p y and the spin echo image shows that it is possible to identify the xylem and p h l o e m regions in the F L A S H image, but that the resolution of the F L A S H image was not as high as that of a spin echo image (Fig. 2 d). H o w e v e r , in contrast to the acquisition time of a F L A S H image ( a b o u t 2 s) a spin echo image of very high resolution (as shown in Fig. 2 d ) needs several hours. This period was not tolerable because this time would have b e e n missing for the m e a s u r e m e n t s of the kinetics of the xylem and p h l o e m flow velocities as a function of a light-dark regime. Therefore, flow images were m a d e in the following experiments only in combination with a F L A S H image. This was obviously also M. Rokitta et al.: Phloem and xylem flow in fully differentiated plants justified because the FLASH image apparently contained the necessary structural information. Both images show that the eight vascular bundles (initially separated from each other in younger plants) formed a closed ring due to secondary thickening. This ring appeared dark in the FLASH image because of the reduced free-water concentration per volume in the xylem area. However, the large vessels which can be seen very clearly in the spin echo image (Fig. 2 d) are also visible in the FLASH image (Fig. 2 a). The resolution of the FLASH image, particularly in combination with the flow map (Fig. 2 b), was also high enough to reveal the separation of the xylem from the phloem by the cambium (about 100-150 gm in thickness; see also Fig. 2 c). The flow image (Fig. 2b) shows that the xylem flow was directed upwardly. In order to demonstrate this, this flow direction is indicated by blue pixels in the image and the flow velocity values are provided with a positive sign. The phloem flow velocities were negative, i.e., opposite to the xylem flow velocities. This is indicated in Fig. 2b by the red pixels. Flow datasets as shown in Fig. 2b were performed regularly in 7 rain time intervals on five different plants subjected to various light-dark regimes. The corresponding average flow velocities and the calculated (normalized) volume flows in the xylem and in the phloem of two plants (A and B) are given in Fig. 3 a and b, respectively. Figure 3 c represents the transpiration and the assimilation rates measured simultaneously. As can be seen from Fig. 3, an increase in the transpiration and assimilation rate upon illumination is immediately reflected in an increase of the xylem flow velocity and volume flow. As also expected, when the light was switched off, the transpiration rate and the assimilation rate decreased (whereby the assimilation rate assumed negative values because of CO2 release from the plant due to respiration). Correspondingly, the flow velocity and the volume flow in the xylem decreased. In contrast to the xylem flow, the phloem velocity was completely independent of the light-dark regime, at least within the limits of accuracy. However, the volume flow apparently increased during the dark period in order to decrease again after the light was switched on. Discussion The objective of the present work was to demonstrate that the noninvasive method of NMR microimaging 129 allows the time-resolved observation of phloem and xylem flow simultaneously provided that appropriate modifications of the hardware and the pulse sequences were made. Up to now, this technique could only be applied to seedlings without transpiring leaves. The temporal resolution was with 4.5 h also much too long to resolve the kinetics both of phloem and xylem flow (K6ckenberger et al. 1997). In this study, the time resolution could be reduced to 7 min. This was fast enough to monitor the immediate effect of illumination on the kinetics of both flows which were opposed to each other as expected. The improved N M R imaging technique yielded values for the flow velocity in the xylem and the phloem which were in the ranges reported by Kuchenbrod et al. (1996) and by K6ckenberger et al. (1997), respectively. We could further demonstrate that the flow velocity and the volume flow in the xylem revealed a pattern in response to a light-dark regime which was consistent with the current belief of the effect of transpiration on xylem tension and flow (Schneider et al. 1997, Kuchenbrod et al. 1996). The most remarkable result was, however, that the flow velocity in the phloem was not changed by the lightdark regime despite an apparent increase in the phloem volume flow during darkness. These findings have two implications. First, the light independence of the phloem velocity is not necessarily in contrast to the pressure-flow hypothesis of Mtinch (1930) provided that a steady-state pressure gradient is maintained diurnally and nocturnally by osmoregulatory processes between the phloem and the accessory parenchyma cells or the xylem vessels (Zimmermann 1978; Van Bel 1990, 1993; Ziegler 1977). Second, in the light of the Hagen-Poiseuille law, the finding of an increase in volume flow during the dark period at constant velocity can only be explained by the assumption that the conducting area of phloem flow is increased during this period. This could mean that part of the sieve pores are continuously open, but that additional sieve pores which may be plugged by protein threads during the light period become temporarily open during darkness. This conclusion is in agreement with recent work of Knobtauch and Van Bel (1998). Van Bel and co-workers pointed out that the structure and mode of the sieve elements can only be elucidated when they are intact and actually translocating because of their extreme sensitivity to any kind of manipulation (cutting, chemical fixation, etc.). The noninvasive N M R flow-imaging technique presented 130 M. Rokitta et al.: Phloem and xylem flow in fully differentiated plants Plant , .)! ! ! ',~ ~ ~ Plant A ~ 0,8 " ; i : i : i ; i : : : i i i i .--~-- flow velcc~ty ---~-- normalized vomm~ ~o~ B 1,0 ~,0 0,8 0,8 < o 0,6 ,o. 0,4 ~ 0,2- fo ~ ~ z ~ : ' . : i ; : ; ; : : ; ,i : i i i i i i II i : : ~ ~ i . . . 0,0 l,O,,b), ' o,2 0,2 0,0 0,0 1,0 1,o i ! 0,8 o< i i ! 0,8 PHLOEM: i i i : ! i --'~'--" norml~llzed volume flow i : ! ! i i - o,6 ~<_ 11) 00 o,6 =E g= B ' 1 i" Ii : ~ ~ ~ ~ 1 00 ii!ii . . . . ' n . 3 0,4 i" 02- ; E- 0,2 i:i: iiili ' 0,0 0,0 10 . 9 ? ? . ! . : . ),.~ . ! ) ~ ?,r i ! ~ anspraton ate : 5 [ i /i : : : i "~0,6 ? ? ! i ~ ~0,4 i , ) i ? i i i 2 i i i ! ? 08 ~p (n ? ;2 ~-*~*~ ; : : : o I0,6 :; ': : -g_0,at- i i ~ i : ~~*"~' ! w; : ' i !:1 . . . . i; [i ii 0 ,4 m ;2 2=~. g o,a i ! ! i : : ! ~transplratlan 01 rate -2 i 0,,I! 0,0 18:00 00:00 06:00 : ~, i ; 12:00 ~ ' ",, ; ' :. i i : i ,, ! i i ! : ; i : ! ! : ; i 18:00 i : ; , , , 00:00 ~ ~ i ) ! i : . . . . . ~ . 06:00 -0 1-3 t i m e [hh:rnin] time [hh:min I 3. Xylem flow velocity and volume flow (a) and phloem flow velocity and volume flow (b) in dependence of a light-dark regime measured on two different plants (A and B). Simultaneous measurements of the transpiration and assimilation are given in e. The light and dark periods are indicated by white and black horizontal bars. Note that the assimilation rate assumes negative values during darkness because of respiration of the plants. For detailed discussion, see text Fig. here may, therefore, c o n t r i b u t e in f u t u r e to the elucid a t i o n of the local c o u p l i n g b e t w e e n x y l e m a n d p h l o e m flow and, in turn, to the l o n g - d i s t a n c e transport in the p h l o e m in r e s p o n s e to t i m e - v a r y i n g envir o n m e n t a l conditions. Acknowledgments The authors thank Ralf Deichmann for improvements in the software used for evaluation of the flow datasets. This work was sup- ported by two grants from the Deutsche Forschungsgemeinschaft: Schwerpunkt Apoplast Zi 99/9-2 to U.Z. and Graduiertenkolleg GRK 64/2 to A.H. and U.Z. References Bourgeois D, Decorps M (1991) Quantitative imaging of slow coherent motion by stimulated echoes with suppression of stationary water signal. J Magn Reson 94:20-33 Haase A, Frahm J, Matthei D, Merboldt K-D (1986) FLASH imaging: rapid NMR imaging using low flip angle pulses. J Magn Reson 67:258-266 M. Rokitta et al.: Phloem and xylem flow in fully differentiated plants Brandl M, Kuchenbrod E, Link A (1993) Magnetization-prepared NMR microscopy. J Magn Reson Set A 105:230-233 Knoblauch M, van Bel A l E (1998) Sieve tubes in action. Plant Cell 10:35-50 K6ckenberger W, Pope IM, X~a Y, Jeffrey KR, Komor E, Callaghan PT (1997) A non-invasive measurement of phloem and xylem water flow in castor bean seedlings by nuclear magnetic resonance microimaging. Planta 201:53-63 Kuchenbrod E, Landeck M, Ththmer F, Haase A, Zimmermann U (1996) Measurement of water flow in the xylem vessels of intact maize plants using flow-sensitive NMR imaging, Bot Acta 109: 184-186 Lahrech H, Briguet A, Graveron-Demilly D, Hiltbrand E, Moran PR (1987) Modified stimulated echo sequence for elimination of signals from stationary spins in MRI. Magn Reson Med 5: 196200 Mtinch E (1930) Die Stoffbewegung in der Pflanze. Fischer, Jena Peuke AD, Hartung W, Jeschke WD (1994) The uptake and flow of C, N and ions between roots and shoots in Ricinus cornmunis L. II: grown with low or high nitrate supply. J Exp Bot 45: 733740 - 131 Rokitta M, Zimmermann U, Haase A (1999) Fast NMR flow measurements in plants using FLASH imaging. J Magn Reson 137: 29-32 Schneider H, Zhu JJ, Zimmermana U (1997) Xytem and cell turgot pressure probe measurements in kttact roots of glycophytes: transpiration ilzduces a change in the radial and cellular reflection coef~cients. Plant Ceil Environ 20:221-229 Tanner JE (1970) Use of the stimulated echo in NMR diffusion studies. J Chem Phys 52:2523-2526 Van Bel AJE (1990) Xylem-phloem exchange via the rays: the undervalued route of transport. J Exp Bot 41:631-644 (1993) Strategies of phloem loading. Annu Rev Plant Physiol Plant Mol Biol 44:253-281 Xia Y, Sarafis V, Campbell EO, Callaghan PT (1993) Non-invasive imaging of water flow in plants by NMR microscopy. Protoplasma 173; 170-176 Ziegler H (1977) Flttssigkeitsstr0me in Pflanzen. In: Hoppe W, Lohmann W, Markl H, Ziegler H (eds) Biophysik. Springer, Berlin Heidelberg New York, pp 561-577 Zimmermann U (i978) Physics of turgor and osmoregulation. Annu Rev Plant Physiol 29:121-148 - Verleger: Springer-Verlag KG, Sachsenplatz 4-6, A-1201 Wien. - - Herausgeber: Prof. B. E. S. Gunning, Research School ol Biological Sciences,The Auslralian National University, Canberra Cily,ACT 2601,Australien. - - Redaktion: Sachsenplatz 4-6, A-1201Wien. - - Satz un d Umbruch. Best-set Typeseller Ltd., Hong Kong. - - Druck: Adolf Holzhausens Nachfolger, Kandlgasse 19-21, A-1070 Wien. - - Verlagsort: Wien. - - Herstellungsorl: Wien. -- Prinled in Austria
© Copyright 2026 Paperzz