FEMA ELECTRÓNICA DOUBLE COUNTER RATEMETER Model DAN985I Double COUNTER / RATEMETER for 4/20 mA signals IDEAL SOLUTION for measuring flow applications, visualizing both the current flow and the accumulated flow, and with the capability to scale the visualized signal in different units (m3, liters, ...) USER’S MANUAL (HT0205-r050705) FEMA ELECTRÓNICA - Page 1 User’s Manual - Instrument DAN985I Model DAN985I Double COUNTER / RATEMETER for 4/20 mA Signals The DAN985I is a particular version of the double unit DAN984 wiht function Counter / Ratemeter. The input signal from 4 to 20 mA reaching the DAN985I is integrated and a new frequency signal is generated from 0 to 2500 impulses/second proportional to the 0/ 10 Vdc signal. With the new impulses/second signal, the instrument works as a standard ratemeter / counter , for applications in flow measuring, at instantaneous and accumulated flow measuring. The DAN985I is a 6 digit panel meter, with 2 internal instruments, named «A» and «B». Only one of the internal instruments («A» or «B») can be visualized on the display. The internal instrument which is not being visualized, is still working normally, and its value can be visualized by acting on the instrument controls in order to switch between instruments «A» and «B». Instruments «A» and «B» are by default configured for «A» acting as a ratemeter (display proportional to the impulses/second arriving to the instrument) and «B» acting as a counter (display proportional to the number of impulses received since the last reset). Additionaly, instrument «A» can be configured as a counter instead of ratemeter, being the whole DAN984 instrument a COUNTER/COUNTER unit. The DAN985I has 2 signal inputs for impulses, named «INPUT1» and «INPUT2». By default, instruments «A» and «B» receive impulses from «INPUT1», and «INPUT2» is by default configured as an inhibit control. «INPUT2» can be also configured to act as a second input for applications with bidirectional encoders, or to control the add/substract function, for example. The DAN985I instrument is configured with codes. These codes activate specific functions, and the codes and values are listed in this manual, together with the procedure to activate them from the frontpanel of the instrument. The RATEMETER instrument (internal instrument «A») displays INSTANTANEOUS FLOW, and the COUNTER instrument (internal instrument «B») is the display for ACCUMULATED flow. Technical Data DISPLAYS SIMBOLS DIGIT HEIGHT POLARITY INDICATION DECIMAL POINT Led 7 segments, red color 9.9.9.9.9.9 14,2 mm. (0,56") negative counts indicated by LED from 0 to 999999 6 positions, programmable SIGNAL INPUT 4/20 mA RESET At rearside by contact terminal At front by push button +10Vdc (maximum 50 mA) 0.012% E2 Prom non volatile 100.000 minimum 10 years minimum Vexc OUTPUT ACCURACY MEMORY CYCLES ON/OFF DATA RETENTION RELAY OUTPUTS 2 Relay Contacts (NC,NO,Common) 9A-250Vac NON INDUCTIVE load POWER CONSUMPTION EXTERNAL FUSE 230 Vac (50 Hz) 3VA 50 mA at 230 Vac (not incluided) CONNECTIONS WIRE SECTION WORKING TEMP. STORAGE TEMP RELATIVE HUMIDITY HOUSING DIMENSIONS PANEL CUT-OUT HOUSING MATERIAL WEIGHT Plug-in screw clamps max. 2.5 mm2 (recommended 1mm2) from 0 to +50 ºC from -40 to +70 ºC 90 % at 40°C DIN 43700 96 x 48 x 117mm (3,78 x 1,89 x 117‘’) 92,5 x 44,5 mm (3,64 x 1,75’’) ABS black color 310 gr. Technical Data for Integrator Linearity Accuracy Thermal Drift Stabilization Time FEMA ELECTRÓNICA - Page 2 0.1% 0.1% <150 ppm 10 minutes User’s Manual - Instrument DAN985I 1.- Front side Description and Connections Button PS .- Move through the programming codes Button PE .- Apply the programming code selected Button PR .- Reset Button PS (Blue) Programming Codes Button PE (Green) AL1 AL2 PDIS ResetA 6 5 4 3 2 1 E ± 1 to 6 AL1 AL2 E Programming codes Status for Alarm1 Status for Alarm2 Validate the code visualized on display A B ± instrument «A» active on display instrument «B» active on display changes the active instrument A B Sign Led Led for negative count indication Programming Blocked (Indicates Jumper J1 Closed) ResetA (No function) Button PR (Red) Sign PDIS Alarm AL2 4/20 mAc + Alarm AL1 2.- Connections and Signal Type nc no c c no nc JUMPERS at rear side JUMPERS internal Close ADF Open J Control of «INPUT2» with MECHANICAL CONTACT terminals 6 and 3 Internal jumper K closed FEDCBA 7 6 5 4 3 Reset Input2 NPN SIGNAL terminal 6 (Signal), 3 (Common) and 4 (Vexc. 10Vdc) Control of RESET with 21 + - Power Common MECHANICAL CONTACT terminal 7 (Reset) and 3 (Common) 3.- Internal Jumpers (Frequency Filters) Jumpers J,K .- Frequency Filters Jumpers J,K Closed Filter at < 100 Hz G H J K Jumpers J,K, Open Filter at <10 KHz Note .- Use filter at <100Hz only to prevent rebounds on input from Mechanical Contact Jumper L .- Enable front RESET Jumper L Closed Front RESET Enabled Jumper L Open Front RESET Disabled Jumper L Jumper J1 .- Programing Blocked Jumper J1 Closed Programming Blocked Jumper J1 Open Programming Normal Jumper J2 CLosed for 4/20 mA Note .- activated led PDIS (Programming Disabled) at the front of the instrument OFFSET 4.- Mechanical Dimensions (mm / inches) 48 (1,89) Jumper J1 5 max (0,20) 96 (3,78) 13 102 15 (0,51) (4,02) (0,59) 123456789012345678 123456789012345678 123456789012345678 123456789012345678 Panel Cut-out 123456789012345678 123456789012345678 92,5 x 44,5 123456789012345678 123456789012345678 (3,64 x 1.75) 123456789012345678 123456789012345678 FEMA ELECTRÓNICA - Page 3 SPAN User’s Manual - Instrument DAN985I 5.- Introducing DAN985I configuration The programming codes for the DAN985I are composed of 2 digits and a value asigned to the code. For example, code «46» controls the decimal point position for instrument «B», and its default value is «1» (no decimal points). This programming code is named as «46 1E». The «46 3E» code activates decimal point on instrument «B» at position «XXXX.XX». The «E» at the end of the code means that code «46 3E» needs to be validated («E») in order to activate. A description follows on how to activate code «46 3E» : Press PS until number «4» is selected Press PE to validate Number «4» displays 4 Press PS until number «6» is selected Press PE to validate Number «6» displays 46 1 Note .- together with code «46» appears the value «1» (total «46 1») indicating that code «46» has a value of «1» Press PS until number «3» is selected Press PE to validate. Number «3» substitutes «1» 46 3 Press PS until number «E» is selected Press PE to validate Instrument activates decimal point XXXX . XX IMPORTANT .- After 6 seconds without programming, the instrument rolls-back and cancels the ongoing programmation Some of the programming codes for the DAN985I are made of only 1 digit, which is the case for codes «1», «2» and «3», controlling setpoints for alarms «AL1» and «AL2», and the «Multiplication Factor» of the instruments «A» and «B». When introducing code «1», the display shows a 6 digit value, which is the current value for «AL1». To modify each one of the 6 digits, select one of the numbers «1» to «6» and press PE until you get on display the desired alarm value. To validate, select «E» and press PE. MULTIPLICATION FACTORS .- By default, the instrument is made of several multiplication factors, all with a value of x1, which multiply the received signal before display. The counter and the ratemeter display by default the actual impulses/second and the total impulses since the last reset. These values can be multiplied in order to modify (scale) the value displayed. The multiplication factors are activated with the programming codes. There are multiplicators common to both instruments («A» and «B») and mutliplicators that only affect one of the instruments. Multiplicators Common to instruments «A» and «B» «44» Multiplication by Flanges x1 , x2 Does not affect to instrument «A» when functions as Ratemeter Multiplicators for instrument «A» when functions as Ratemeter «3» «62» «64» Scale Factor Time Factor Multiplicator from 0.0000 to 5.9999 x1, x60, x3600 x1, x10, x100x 1000, x0.1, x0.01 Multiplicators for instrument «A» when functions as Counter «3» «61» Scale Factor Multiplicator from 0.0000 to 5.9999 x1, x0.1, x0.01, x0.001 Multiplicators for instrument «B» when functions as Counter «3» «45» Scale Factor Multiplicator from 0.0000 to 5.9999 x1, x0.1, x0.01, x0.001 FEMA ELECTRÓNICA - Page 4 User’s Manual - Instrument DAN985I 6.- Basic configuration of the instrument «41 1E» Instrument «A» acting as RATEMETER (instrument «B» functions always as COUNTER) «41 2E» Instrument «A» acting as COUNTER (instrument «B» functions always as COUNTER) 7.- Modes of Operation «43 1E» «INPUT1» functions as impulses input for instruments «A» and «B» «INPUT2» functions as inhibit control when is driven to logical state «0» «43 2E» «INPUT1» functions as impulses input for instruments «A» and «B» «INPUT2» funcions as ADD/SUBSTRAT control for the counters. Substracts when driven to state «0» «43 3E» Not available in this unit «43 4E» Not available in this unit «43 5E» Not available in this unit «43 6E» Not available in this unit NOTE .- The ratemeter is not affected by functions RESET, INHIBIT and SUBSTRACT 8.- Display instrument «A» or «B» ± Select «±» and press «PE» to change the instrument on display «A» or «B» The leds at the bottom front part, labeled «A» and «B», indicate the active instrument 9.- Reset of the instrument configuration «41 -1E»Reset of the instrument configuration. Activates the default parameters (see section 29) In order to place the negative sign in the code, introduce codes «4», «1», «1» and «±» and «PE» 10.- Operating mode of the RESET (Front Reset and Rearside Reset) «42 1E» funtions as Reset for instrument «A» and deactivates relays assigned to instrument «A» «42 2E» funtions as Reset for instrument «B» and deactivates relays assigned to instrument «B» «42 3E» funtions as Reset for both instruments and deactivates both relays 11.- Assignation of «AL1» and «AL2» to instruments «A» and «B» «51 1E» «AL1» assigned to Instrument «A» «AL2» assigned to Instrument «B» «51 2E» «AL1» and «AL2» assigned to Instrument «B» «51 3E» «AL1» and «AL2» assigned to Instrument «A» (Activates code «56 1E» Reset Normal) 12.- Programming the value for «AL1» and «AL2» «1» «2» displays the value of «AL1» displays the value of «AL2» To modify the value of the alarm, use leds 1,2,3,4,5,6 to modify the value of the digits 1,2,3,4,5,6.When the led of the digit position we want to change, is selected, press PE to mofify the value (each press will add +1 to the digit selected).Once the desired value of the alarm is displayed, select «E» and press PE to validate. FEMA ELECTRÓNICA - Page 5 User’s Manual - Instrument DAN985I 13.- Scale Factor «3» The «Scale Factor» is a multiplicator between 0.0000 and 5.9999. Value by default is 1.0000 Instruments «A» and «B» have its own independent «Scale Factor» The value of instruments «A» and «B» is multiplied by the value of its «Scale Factor», and displayed Introducing «3» displays the value of the «Scale Factor» for the active instrument («A» or «B») 14.- Flange Multiplication «44 1E» «44 2E» Multiplication x1 (Counts at down flanges) Multiplication x2 (Counts both up and down flanges) Note .- Code «44» can multiply x1 or x2 the display, and affects only to counters, both in «A» and in «B» Note .- Not compatible with code «43 6E» 15.- Configuration of Instrument «A» when functions as Ratemeter «61 «61 «61 «61 1E» 2E» 3E» 4E» Digit Fixed to Zero Digit Fixed to Zero Digit Fixed to Zero No Zeros XXXXX0 XXXX00 XXX000 XXXXXX This option moves the dislay to the left while placing zeros to the right «62 1E» «62 2E» «62 3E» Display 'PER SECONDS' (x1) Display 'PER MINUTE' (x60) Display 'PER HOUR' (x3600) Display = Frequency Input x1 Display = Frequency Input x60 Display = Frequency Input x3600 «63 «63 «63 «63 «63 «63 1E» 2E» 3E» 4E» 5E» 6E» Time Base between 0,5 and 1 seconds Time Base between 1 and 2 seconds Time Base between 2 and 4 seconds Time Base between 4 and 8 seconds Time Base between 8 and 16 seconds(Frequency Maximum 7500 Hz) Time Base between 16 and 32 seconds(Frequency Maximum 3250Hz) «64 «64 «64 «64 «64 «64 1E» 2E» 3E» 4E» 5E» 6E» Multiplicator x1000 Multiplicator x100 Multiplicator x10 Multiplicator x1 Multiplicator x0.1 Multiplicator x0.01 16.- Configuration of instrument «A» when functions as Counter «64 1E» «64 2E» «64 3E» «64 4E» Multiplication Factor x1 Multiplication Factor x0,1 Multiplication Factor x0,01 Multiplication Factor x0,001 Acts with the counter impulses x0.1 and displays the impulses counted /10 17.- Configuration of instrument «B» when functions as Counter «45 «45 «45 «45 1E» 2E» 3E» 4E» Multiplication Factor x1 Multiplication Factor x0,1 Multiplication Factor x0,01 Multiplication Factor x0,001 Acts with the counter impulses x0.1 and displays the impulses counted /10 18.- Position for Decimal Point Inst. «A» «65 1E» «65 2E» «65 3E» «65 4E» «65 5E» Inst. «B» «46 1E» «46 2E» «46 3E» «46 4E» «46 5E» 0 0.0 0.00 0.000 0.0000 Note .- The same codes, but with negative, allow display of zeros to the left. For example, code «46 -2E» will indicate 00000.0 FEMA ELECTRÓNICA - Page 6 User’s Manual - Instrument DAN985I 19.- Configuration of «AL1» operating as «ALARM OF MAXIMUM» Functionality .- «RELAY1» activates when display is higher than the value of «AL1» and is deactivated when: «52 1E» «RELAY1» is deactivated when «RELAY2» is activated (Needs «RELAY1» and «RELAY2» assigned to counter «B») «52 2E» «RELAY1» is deactivated when «RELAY2» is deactivated (Needs «RELAY1» and «RELAY2» assigned to counter «B») «52 3E» «52 4E» «52 5E» «52 6E» «RELAY1» is deactivated when pressing front «RESET» «RELAY1» is deactivated when releasing front «RESET» «RELAY1» is deactivated when finishing the delay time (See code «53») «RELAY1» is deactivated when display is lower than the value of «AL1» 20.- Configuration of «AL2» operating as «ALARM OF MAXIMUM» Functionality .- «RELAY2» activates when display is higher than the value of «AL2» and is deactivated when «54 1E» «RELAY2» is deactivated when «RELAY1» is activated (Needs «RELAY1» and «RELAY2» assigned to counter «B») «54 2E» «RELAY2» se desactiva cuando «RELAY1» se desactiva (Needs «RELAY1» and «RELAY2» assigned to counter «B») «54 3E» «54 4E» «54 5E» «54 6E» «RELAY2» is deactivated when pressing front «RESET» «RELAY2» is deactivated when releasing front «RESET» «RELAY2» is deactivated when finishing the delay time (See code «55») «RELAY2» is deactivated when display is lower than the value of «AL2» (Activates «56 1E») 21.- Configuration of «AL1» operating as «ALARM OF MINIMUM» Functionality .- «RELAY1» is deactivated when display is higher than the value of «AL1» and is activated when «52 -1» «RELAY1» is activated when «RELAY2» is activated (Needs «RELAY1» and «RELAY2» assigned to counter «B») «52 -2» «RELAY1» is activated when «RELAY2» is deactivated (Needs «RELAY1» and «RELAY2» assigned to counter «B») «52 «52 «52 «52 -3» -4» -5» -6» «RELAY1» is activated when pressing front «RESET» «RELAY1» is activated when releasing front «RESET» «RELAY1» is activated when finishing the delay time (See code «53») «RELAY1» is activated when display is lower than the value of «AL1» 22.- Configuration of «AL2» operating as «ALARM OF MINIMUM» Functionality .- «RELAY2» is deactivated when display is higher than the avlue of «AL2» and is activated when «54 -1E» «RELAY2» is activated when «RELAY1» is activated (Needs «RELAY1» and «RELAY2» assigned to counter «B») «54 -2E» «RELAY2» is activated when «RELAY1» is deactivated (Needs «RELAY1» and «RELAY2» assigned to counter «B») «54 -3E» «54 -4E» «54 -5E» «54 -6E» «RELAY2» is activated when pressing front «RESET» «RELAY2» is activated when releasing front «RESET» «RELAY2» is activated when finishing the delay time (See code «53») «RELAY2» is activated when display is lower than the value of «AL2» (Activates «56 1E») 23.- Delay on activation of «RELAY1» and «RELAY2» «53» Delay of «RELAY1» in seconds (from 0,01 to 599,99 seconds) Needs code «52 5E» «55» Delay of «RELE2» in seconds (from 0,01 to 599,99 seconds) Needs code «54 5E» FEMA ELECTRÓNICA - Page 7 User’s Manual - Instrument DAN985I 24.- Special Functions on RESET «56 1E» Loads on display 000000 «56 2E» Loads on display the value of «AL2» and the impulses substract from the display value When reaching 000000 «RELAY2» is activated and continues substracting in negative «56 3E» Loads on display 000000 and when reaching the value of «AL2», the display backs to 000000 (Value for delay «55» will be taken in consideration, and needs code «54 5E») «56 4E» Loads on display the value of «AL2» and the impulses substract from the display value When reaching a 000000 value of «AL2» is reloaded on display and impulses substract from display «56 5E» Loads on display 000000 and when reaching the value of «AL2» display is reloaded to 000000 «56 6E» Loads on display the value of «AL2» and the impulses substract from the display value When reaching 000000 value of «AL2» is reloaded on display and impulses substract from display 25.- Programming Example Working with a flowmeter offering a 4/20 mA signal with a maximum flow of 100 liters/hour. We want to connect the output of the flowmeter to the DAN985I instrument in order to display the instantaneous flow in liters/hour and the accumulated flow in liters. 1 .- General Configuration «A» = Ratemer (Instantaneous flow) «B» = Counter (Accumulated Flow) code «41 1E» 2.- Ratemeter Configuration 4 mA = 0 impulses / second = 0 liters / hour 20 mA = 2500 impulses / second = 100 liters / hour we need to configure an input from 0 to 2500 imp/sec. with a display of 0 to 100 liters/hour 2500 x MULTIPLICATOR = 100 MULTIPLICATOR = 100 / 2500 = 0.0400 The programming codes for the Ratemeter will be : Scale Factor = 0.0400 code «3» with a value of 0.0400 3.- Counter Configuration Look for the addition to the display for a single impulse received 2500 imp./sec. = 100 liters/hour x 1 hour/3600 sec. = 0.027777 liters/sec. 2500 imp. = 0.027777 liters 1 imp. = 0.000011111 liters = 0.0111 X 0.001 we need to programm the Counter wit the following codes : Scale Factor = 0.0111 Multiplication Factor = x 0.001 code «3» with a value of 0.0111 code «45 4E» NOTES .- Note that once the overall multiplication factor has been calculated (both in the ratemeter and in the counter) we split the value in 2 parts, in order to use the decimals of the Scale Facto r«3», which can have value between 0.0000 and 5.9999. FEMA ELECTRÓNICA - Page 8 User’s Manual - Instrument DAN985I 26.- Blocking the Programming Jumper J1 Open.Jumper J1 Closed.«66 1E» «66 2E» «66 3E» «66 4E» «66 5E» «66 6E» allows all programming actions blocks the programming of the instrument, depending on value of code «66» Allows RESET Allows RESET and modify values of «AL1» and «AL2» Allows RESET and «Scale Factor» Allows RESET, «Scale Factor», «AL1», «AL2» Allows RESET, «AL1», «AL2» and load a value in the counter Allows RESET, «AL1», «AL2», «Scale Factor» and load a value in the counter Note .- the same codes but in negative, allow the same functions but without RESET. For example, code «66 -2E» allows to modify «AL1» and «AL2» but does not allow RESET 27.- Error Messages and Display Tests Error P Error E «6±» Error at start-up - Press ENTER. Check the programmation Error at the micro - Press ENTER. Check the programming generates a test display sequence to check the lightning of the digits 28.- Limitations on the internal Counter The DAN985I instrument is made of an internal counter with 23 bits, plus 1 bit for the sign. In total there are 24 bits for the impulse codification, with a maximum internal counting of 8.388.608 impulses (more than 8 million impulses). Once this value has been achieved, the effect on display is the activation of the negative sign and the counting in SUBSTRACT mode for the COUNTERS. In case of finding this situation, press the RESET and the counters will recover functionality Also «AL2» can be programmed in order to release a RESET automatically when reaching the value programmed, without need of a manual RESET (code «56 5E»). FEMA ELECTRÓNICA - Page 9 User’s Manual - Instrument DAN985I 29.- Default Programming Codes «1» «2» «3» «3» 500 1000 1.0000 1.0000 Value for «AL1» Value for «AL2» «Scale Factor» for Instrument «A» «Scale Factor» for Instrument «B» «41 1E» «42 3E» «43 1E» «44 1E» «45 1E» «46 1E» «61 4E» «62 1E» «63 1E» «64 4E» «65 1E» «51 2E» «52 3E» «53» «54 3E» «55» «56 1E» «66 1E» Instrument «A» acting as Ratemeter Reset for instruments «A» and «B» and deactivation of «RELAY1» y «RELAY2» «INPUT1» impulse input, «INPUT2» inhibit control Flanges Multiplication x1 instrument «B» Counter with «Multiplication Factor» x1 Instrument «B» without Decimal Point Instrument «A» Ratemeter without fixed zero Instrument «A» Ratemter with «Time Multiplicator» x1 Time Base for instrument «A» Ratementer between 0,5 and 1 seconds Instrument «A» Ratemeter with Multiplicator x1 Instrument «A» without Decimal Point «AL1» and «AL2» assigned to instrument «B» «RELAY1» as maximum alarm, and deactivates when pressing fron reset 00000.1 Delay on activation for «RELAY1» at 0.1 seconds «RELAY2» as maximum alarm, and deactivates when pressing fron reset 00000.1 Delay on activation for «RELAY2» at 0.1 seconds Special Functions. Reset normal Blocking exceptions, allows RESET NOTES - INSTRUMENT DAN985I FEMA ELECTRÓNICA - Page 10 User’s Manual - Instrument DAN985I WARRANTY FEMA ELECTRÓNICA, S.A. warrants this unit to be free of manufacturing defects for a perior of 12 MONTHS from shipment date. This warranty covers both materials and manufacturing processes. This warranty is VOID if the unit shows evidence of damages as a result of misapplication, accident, misuse or if the product had been tampered or repaired by personnel or companies without the official authorization of FEMA ELECTRÓNICA, S.A. This warranty is also VOID for damages caused by defective or inappropriate applications. In case of malfunction, the unit should be sent to the manufacturer for its evaluation. Within the warranty period, and always previous examination from FEMA ELECTRÓNICA, S.A., the unit will be repaired or replaced to the discretion of the manufacturer. Limitation of liability : FEMA ELECTRÓNICA, S.A. shall not be responsible for any damage or loss to other equipment however caused, which may be experienced as a result of the installation or use of this product. FEMA ELECTRÓNICA, S.A. liability shall not exceed the purchase price paid of the product upon which liability is based. In no event shall FEMA ELECTRÓNICA, S.A. be liable for consequential, inicidental or special damages RETURNS FOR REPAIR Ship free of charges and properly packed to : FEMA ELECTRÓNICA, S.A. Pol.Ind.Santiga (Altimira 14, T-14, N-2) E-08210 BARBERÀ (Barcelona) España - Spain INSTALLATION PRECAUTIONS.- The installation and use of this unit must be done by qualified operators. The unit has not power switch, and the unit has not internal protection fuse, and will start operation as soon as power is connected. The installation must incorporate an external main switch with a protection fuse of : fuse 50 mA (at 230 Vac power) Also the necessary devices to protect the operator and the process when using the unit to control a machine or process where injury to personnel or damage to equipment or process, may occur as a result of failure of the unit. SAFETY PRESCRIPTIONS.- The unit has been designed and tested under UNE 20553 rules and is delivered in good condition. This manual contains information for electrical connections. Do not make wiring signal changes or connections when power is applied to the unit. Make signal connections before power is applied and, is reconnection is required, disconnect the AC (mains) power before such wiring is attempted. Install the unit in places with a good ventilation to avoid the excessive heating. And far from electrical noise source or magnetic field generators such as power relays, electrical motors, speed controls etc... The unit cannot be installed in open places. Do not use until the installation is finished. POWER SUPPLY.- The power supply must be connected to the adequate terminals (see the connection instructions). The characteristics of the power supply are written on the attached label. Please make sure that the unit is correctly connected to a power supply of the correct voltage and frequency. Do not use other power supply otherwise permanent damage may be caused to the unit. Do not connect the unit to power sources heavily loaded or to circuits which power loads in cycle ON-OFF or to circuits which power inductive loads. SIGNAL WIRING.- Certain considerations must be given when install the signal input and control wires. Long wires can act like antennas and introduce electrical noise to the unit, therefore : A.- Do not install the signal input or control wires in the same conduit with power lines, heaters, solenoids, SCR controls etc....and always far from these elements. B.- When shielded wires are used, connect the shield to the common terminal and leave unconnected the other end of the shield and do not connect to the machine ground. EXCITATION VOLTAGE.- The unit supplies excitation voltage for sensors, at the terminals 3 and 4. Do not connect these terminals to other external power supply, permanent damages may result to the unit. SAFETY CONSIDERATIONS .- Before starting any operation of adjustment, replacement, maintenance or repair, the unit must be disconnected from the power supply. Keep the unit clean , to assure good functioning and performance. To prevent electrical or fire hazard, do not expose the unit to excessive moisture. Do not operate the unit in the presence of flammable gases or fumes, such as environment constitutes a definite safety hazard. The unit is designed to be mounted in a metal panel. If the unit shows signs of damage, or is not able to show the expected measures, or has been stored in a bad conditions or a protection failure can occur, then do not attempt to operate and keep the unit out of service. IN CASE OF FIRE 1.- Disconnect the unit from the power supply. 2.- Give the alarm according to the local rules. 3.- Switch off all the air conditioning devices. 4.- Attack the fire with carbonic snow, do not use water in any case. WARNING : In closed areas do not use systems with vaporized liquids. CONNECTIONS All wiring connections are made using push-in cable connectors. There is a separate connector block for power supply, input signals and each relay output. Please make sure that each connector block is connected on the adequate place. PANEL MOUNTING.- The instrument size case is 1/8 DIN. The internal electronic circuit can be inserted or removed by the front part and is held on to the case by the two lateral lugs, which must be put in their corresponding holes placed on each case side. Verify that the panel cut-out is correctly according to the dimensions indicated with a minimum depth of 135 mm. (5.31"). Install the fixation clips of blue colour in the lateral guides of the unit by its rear part and then slide and press them firmly against the panel, until the unit is totally hold on. FEMA ELECTRÓNICA - Page 11 more products Programmable Panel Meters Signal Converters & Isolated Transmitters Large Displays Standard Panel Meters Miniature Panel Meters Large Displays for TIME www.fema.es ELECTRONIC INSTRUMENTATION FOR INDUSTRY FEMA ELECTRÓNICA, S.A. Pol. Ind. Santiga - Altimira 14 (T14 - N2) E-08210 Barberà - BARCELONA ESPAÑA - SPAIN Tel. (+34) 93.729.6004 - www.fema.es Fax (+34) 93.729.6003 - [email protected] FEMA ELECTRÓNICA - Page 12
© Copyright 2026 Paperzz