UNIVERSITY OF MALTA DEPARTMENT OF BIOLOGY BIO1010 Cellular and Biochemical Basis of Life PRACTICAL WORK IN CELL BIOLOGY Contents: I. Setting up a microscope For Critical illumination For Kohler illumination Use of Immersion Objectives Selecting an area for observation Light and illumination in practice Observations Care of the stand and cleaning of optical components II. Microscopic Preparations Notes on the various methodologies Professor V. Axiak MICROSCOPIC OBSERVATIONS Setting up a Microscope For Critical Illumination 1. Adjust seat height (and inclination of the tube, when possible) so that the eye can be brought easily before the eyepiece; the draw-tube, if fitted, should be adjusted to the correct tube-length. 2. Turn on a low-power objective, e.g. l0x, with the revolving nosepiece; bring the condenser in its highest position. 3. Switch on the lamp and manipulate the (flat!) mirror until the front lens of the condenser lights up. 4. Move a stained object in the light beam from the condenser, using the mechanical stage (object lights up). 5. Look in eyepiece and bring the object into focus with the coarse adjustment so that an image appears. 6. Remove the eyepiece from the tube and manipulate the mirror and the height adjustment on the condenser so that a disc with even illumination is seen when looking down the tube. Close the substage diaphragm so that the outer border (about ¼-¾ of the diameter) of the disc, actually the entrance pupil of the microscope, is screened off. This is no more than a rule-of-thumb to obtain a tentative adaptation of the numerical aperture of the condenser to that of the objective, so that with an average stained specimen not too much glare is generated and no appreciable loss in aperture occurs. 7. Re-insert the eyepiece and adjust the condenser so that maximal intensity of illumination is obtained. When the image of the surface of the opal bulb is found disturbing, the condenser may be moved slightly up or down. For Kohler-Illumination In setting up a microscope with a built-in Kohler-illumination, the steps 1-5 are the same; the subsequent steps are: 6. Close the field diaphragm and focus the border of the illuminated disc sharply in the specimen. 7. Bring the illuminated disc to the centre of the field using the centring screws of the condenser, which should be present with any microscope with Kohler-illumination. 8. Open the field diaphragm until its image fills the field of view, so that the border is no longer visible. 9. Adjust the aperture of the illumination cone with the substage diaphragm, as described in Crit. III. 6). As the Kohler illumination in itself already reduces glare, an excessive aperture of the illumination becomes less easily manifest than with a critical illumination; this adaptation is required nevertheless in order to obtain an optimal result. 10. In changing an objective, both field diaphragm and aperture diaphragm should be adjusted, while the condenser has not infrequently to be realigned with the objective. Use of Immersion Objectives The main practical point to consider here is not so much the introduction of an oil film between the cover glass and the front lens, but the combination of a very short working distance with a very thin depth of field. Missing the image plan in focusing is a far from imaginary event; in the assumption that the image still has to appear, the objective can be forced without noticeable resistance against the object with the advancement of the fine adjustment. This can have serious consequences, as the front lens of the objective might get damaged or become loose in its mounting because of the pressure (which often leads to a crushing of the cover-glass too): due to the high costs of individual manufacture the reparation of a damaged front lens is often more expensive than a complete new objective of the same type. A few technical devices exist for this very old problem. The most universally applied guarding device is the so-called spring-mount; this consists of a telescoping system in which the objective proper can be pushed against the pressure of a spring in an outer tube-mount. This type of mount is not only applied with oil immersion-objectives - for which it has become more or less the standard practice - but also for some high-power dry objectives. When the correct image plane has been missed, the telescoping of the spring-mount, when the objective is forced down onto the slide with the fine adjustment, becomes evident only when the lens is inspected from the side, for there is no image! When the spring- mount has reached the end of its movement, it has no more effect and thus the system is far from fool-proof. The same is the case with other guarding devices, such as a slipping of the fine adjustment when a certain pressure is exceeded, or an adjustable mechanical stop to the fine adjustment. These latter devices can even give a false feeling security; with any appreciable variation of the thickness of slide + mounting medium + coverslip they fail completely. The following general rules can be recommended for maximal safety, although caution should always be exercised. I First select the area to be studied with a low-power dry objective. II Apply one, and not more than one, medium drop of oil on the cover glass exactly where the specimen is illuminated by the condenser, e.g. with a plastic bottle with a long spout. It is important to let any air with escape first from the spout, as air bubbles in oil on the cover glass are very difficult to remove. III In the case of parafocally adjusted objectives the immersion objective can be rotated with the nosepiece into position in the oil without changing the height adjustment of coarse or fine adjustment. IV When the immersion objective has snapped into position, further adjustment should be made with the micrometer until the image comes into focus. When the same low-power objective is always used first, the necessary correction will be identical each time (e.g. half a turn forward with the fine adjustment). With objectives which are not parafocally adjusted such as may be the case with an older microscope type, or a set of objectives of various makes, the focusing of an immersion objective cannot be performed in the way just described. When a suitable area has been selected with a low-power lens, the oil immersion objective is rotated into position and racked up 8-10 mm above the object. When the oil has been applied as described previously, the eye level should be placed near the top of the object slide, so that the objective can be inspected from the side. The oil-immersion objective should be lowered carefully with the coarse adjustment until it touches the oil. At the moment of contact, the oil spreads with a clearly visible jerking movement in the slit-like space between the lower surface of the objective mount and the coverslip. The objective is now lowered towards the object with the fine adjustment, while looking in the eyepiece until the image appears. This is a very important point, which may also apply when an oil-immersion lens can be set up in the easier way discussed at the beginning of this section, with a parafocally adjusted low-power objective. Selecting area for observation In selecting an area to be focused with oil-immersion objective, one has to make sure that under any circumstances an image can be expected to appear. Often the rather small object field of a high-power objective is overestimated, so that one may try to focus on an emptyfield. Moreover, the centring of the objectives is seldom perfect. It is important, therefore, to select a homogeneous area with contrast-rich details for setting up the high-power objective, if necessary even outside of the area to be studied. When the correct level of focus has been found, it is relatively easy to pass a few “empty” areas in the specimen with the mechanical stage. Uncertainty whether one is under or over the level of the focus is a situation which should be avoided; when this state of affairs has been reached (and movements with the mechanical stage do not yield anything like an image either) it is best to stage anew with a low-power objective. In some situations, e.g. a smear of scattered cells of low contrast, it can be difficult to find the correct level of focus with the oil immersion lens. In some circumstances it is even recommended to mount a very small piece of cigarette piece or thin metal foil which is visible to the naked eye with the specimen under the coverslip, so than an approximate level of focus can be found at once at the border of this object in the specimen. With parafocally adjusted objectives, it is easy to turn back eventually to a low-power objective without removing the oil. In contrast to what is often thought, the film of oil does not interfere very much with the formation of an image at low power, as long as the front lens does not touch the oil (which will not easily occur with e.g. a l0x objective with its long free working distance). The oil film on the cover glass will cause, of course, an exaggeration of the cover-glass effect. Even in using a non-resinifying oil both in oil immersion objective and the slide should always be cleaned carefully after use. When a condenser immersion has been applied, the condenser should be cleaned, moreover, and the slide on both sides. As the use of a condenser immersion is seldom justified, even from a theoretical point of view, the rather tedious immersion of the condenser and the cleaning it entails (often also of the stage) is kept for a few selected situations with a highly corrected condenser and a contrast object, so that the full condenser aperture can effectively be used. Light and illumination in practice In describing the standard-procedure for setting up a microscope, a position of the condenser diaphragm has been indicated which, in the case of a stained specimen with average contrast, leads to a good adaptation of the N.A. of objective and condenser. In many cases the illumination cannot be considered as optimal under these conditions, however: the aperture of the condenser may need further adjusting, or the intensity of illumination may be too high or too low. These two factors have to be dealt with independently; the use of the condenser diaphragm for tempering the brightness of the image is one of the most common mistakes of beginners in microscopy. It is true that in closing the substage aperture diaphragm the amount of light passed is reduced, but this is accompanied by a change in the aperture cone of the illumination, which quickly brings about a loss in resolving power and an increase in diffraction phenomena in the specimen. The best way to change the brightness of the image is a variable resistance on the lamp tension, as can be easily installed with a low voltage lamp. When such a control is not feasible (e.g. with a simple high voltage illumination) or the lamp current should be kept constant for certain purposes (e.g. in photometry or photomicrography), an eventual change in image brightness should be brought about with filters. In view of the low luminance of the high voltage bulb poverty of image brightness at high power observation will be a more probable event than a too intense illumination, although this may occur with low-power work. With high-power observation (e.g. using a l00x, N.A. 1.25-1.35 objective) the high-voltage bulb will certainly fall short when a binocular tube is used. The uninitiated observer generally has a tendency to use a too low level of image brightness with a stained specimen, underestimating the consequences of this. A dimly illuminated image reduces the actual resolution (not the resolving power) considerably, however, certain details can be made clear merely by increasing the image brightness. Local differences in light absorption which determine the contrasts in the image of a stained specimen, come out only with a certain degree of lighting intensity. With monocular observation, a too low brightness of the image will increase moreover the habit of tightly closing the eye which is not in use. The tense attitude which this brings about will promote fatigue and headache, which phenomena are often ascribed too easily to monocular observation. A somewhat more dim lighting of the image, on the other hand, is sometimes preferable with unstained objects which are poor in absorbing details and in which the contrasts are determined mainly by diffraction at boundaries. The condenser diaphragm should be in a much more closed position than with a stained specimen, in view of the fact that a higher degree of coherence in the illumination beam, in combination with a minimum of glare in the specimen is desirable in these situations. Observations The study of a microscopic specimen starts - after an inspection with the unaided eye - by focusing a low-power objective on it as already described. With some experience this can be performed with a few quick actions. When the microscope has been set up and the object focused with e.g. a l0x objective, the observation proper can start. This will differ so much from case to case, that it is virtually impossible to make any sensible general remark in this respect. Problems which will often recur, however, will be those of a systematical search for certain details and marking and re-locating of a particular field, once such a detail has been found. In searching a microscopic specimen, a comparatively arbitrary way can be followed when a structure to be studied with high-power can be found easily. The situation is quite different, however, when a specimen has to be examined systematically, e.g. in making a differential count of various components or in finding (or excluding the occurrence of) a certain rarely occurring detail. It is often the best in these situations to search the specimen meander- like, in which use is made of the fact that it is possible to pursue straight tracks through the specimen with the mechanical stage. The track is followed in a given direction up to the border of the preparation after which the position of the object is shifted with the other control until a detail which could just be observed at one side still remains visible at the other side. When the track is followed back after this shift, the strips of object-field just overlap, so that a complete inventory of the specimen is possible if the same shift (or slightly less than the diameter of the object field) is always made when the border of the preparation is reached. In surveying quickly large series of preparations in the search for certain details, it is sometimes easier to move the specimen by hand (depending on the magnification, entirely free on the stage or with one side under a spring clip) than with a mechanical stage. In many cases the mechanical stage is detachable; a built-in mechanical stage can be moved often to an extreme position so that the object can be put on the stage outside of its object holder. When quick screening of larger specimens is often required, the mounting of a gliding stage can be considered. Care of the stand and the cleaning of optical components A microscope stand of good make hardly needs any maintenance for long periods; the moving parts, such as the rack-work, only need cleaning and new grease at very long intervals. Care should be taken never to use thin oil on rack-work, as this may cause the spontaneous sinking of a tube or condenser. A microscope should be kept as much as possible free from dust in a case or under a plastic cover; dust do not only adhere to the stand and its moving parts; but accumulate especially on lenses and other glass surfaces (cover plates of built-in illumination, mirrors). The stage should be cleaned regularly; if anything is spilled on the stage it should be wiped off. Especially mounting material (Canada balsam and the like) should be removed at once (when necessary with some xylol) as it can form hard cakes and be difficult to remove afterwards. The following general rules can be given for the cleaning of optical surfaces, they hold true especially for objectives. It should be kept in mind that external surfaces of optical parts, treated with anti-reflection coating are relatively hard, but that those coverings are very thin. Some of the internal coatings are soft and easily damaged. 1. Non-adhering dust can be removed best with a dry soft brush (if shortly warmed against the surface of a bulb it picks up dust particles more easily) or lens paper. Dust should never be removed by fingers from optical surfaces; the skin which is moist and greasy will usually take up the dust, but leaves a greasy substance instead. Fingermarks or other adhering grease or dirt should be removed with soft cloth or lens paper barely moistened with xylol, or petrol if available. Never soak or immerse a lens in xylol; avoid under any circumstance the use of methylated spirits or other alcohols, ether or acetone in view of the cementing material. When water is used aqua dest. should always be taken instead of tap-water, which leaves a deposit of salts on evaporation. 2. Always take lens paper, linen, etc. double over the finger to prevent infiltration by moisture or grease from the skin. When the border of the mount is protruding over the lens and/or the front lens is concave (as is often the case with plan-objectives) it can be difficult to reach the entire lens surface with lens paper or a cloth. The cleaning should be performed then with a piece of cotton wool impregnated with some xylene around a match stick. 3. Oil-immersion objectives should always be cleaned after use, even with use of nonhardening oils, as these also alter slowly with time. Preferably first use dry lens paper, which is mostly sufficient, finish when necessary with a piece of linen slightly moistened in xylene. When too much of a solvent is on the cloth, touch it with a dry part; the lens should never be soaked in any cleaning fluid. Some very old objectives have lens cements which can be weakened by xylene; cleaning of older objectives should preferably be done entirely dry. 4. Never attempt to take an objective lens to pieces for cleaning; even when everything has been brought back into place correctly, mutual distances can be changed. This does not hold true for eye-pieces with their more simple construction; they can be unscrewed for cleaning without any danger. With most modern objectives unscrewing has become impossible because the mount is made in one piece. Condensers, on the other hand often have a front lens which can be unscrewed to lower aperture of the illumination apparatus; even when this device is not used, dirt can accumulate between both lens systems and should be removed from time to time. Condensers with a swing-put front lens quickly become dusty on the uncovered top side of the large lens. When they cannot be cleaned sufficiently with an air current from a blower brush (a soft brush at the end of the spout of a rubber blower, also useful for cleaning the top cover of a built-in illumination) it should be taken out and cleaned in the manner already described. Recording observations from the microscope: Well annotated drawings are extremely useful in recording microscopical studies. Please follow the following rules in making drawings which are to be presented to the tutor for assessment: • • Draw only what you see, not what is indicated in the textbook or suggested by your imagination. Remember that you are examining a three-dimensional object. Therefore in some cases, a more accurate impression is obtained by racking the focus back and forth, to bring different levels into focus. If possible serial section in T.S. and in L.S. should be examined. • • • • • Use the microscope with BOTH eyes open and have your notebook on the right of the instrument. Drawings under L.P. should show the outlines and the main features drawn to the same scale as seen under the microscope. Magnification should be noted and recorded. Drawings under H.P. should show only representative areas of the field of view, never the whole of it, and different cells should be properly labelled. Drawings of cells should not represent holes or brackets or fish scales! They should represent individual cells as seen under YOUR microscope. Careful observation will indicate that not only have the individual cells their individual boundaries but that adjacent boundaries are interconnected in some quite specific manner according to the nature of the cells. Drawings should be in pencil, while shading or other markings should be kept to the essential minimum and then in such a way that the shading is clearly differentiated from the detail on the drawing - each dot or line on a drawing represents a dot or line on the original and vice versa. Each drawing is to be properly titled: i.e. name of specimen in full, magnification, objective used, and when possible the stains utilised, together with the orientation (L.S. or T.S.). MICROSCOPIC PREPARATIONS Microscopic preparations may either be temporary or permanent. Both involve a series of micro technical processes to which a specimen or tissue must be subjected. These processes include fixing the tissue to kill it, hardening, sectioning, staining, dehydration clearing and mounting. These processes all have some effects on the cellular structure of the specimen. This may result in artifacts or distortions of the actual structure of the tissue. The presence of these artifacts must be borne in mind when interpreting a microscopic preparation. TEMPORARY PREPARATIONS Temporary preparations are well suited for the examination of living or fresh material. One main advantage is that the specimen is subjected to as little preparation as possible, so that it is more or less in a natural state. However the microscopic study of living material presents several problems. The isolated cells of larger organisms, eggs embryo or unicellular organisms must be kept alive, or at least the changes occurring after death delayed. The cells are mostly colourless and transparent, so that their examination presents some difficulties. The latter have been overcome by the use of special techniques for illuminating and observing the specimens, for example phase contrast microscopy, interference microscopy, etc. Mounting in water This is the familiar method of studying small organisms in a drop of water. It is often necessary to support the coverslip using a glass ring or else use a cavity slide. Marine microorganisms should be examined in sea water. Mounting in isotonic saline solutions This technique is used to examine cells from small pieces of tissue. The fluids used are solutions whose composition resembles the plasma normally surrounding the tissue. They are usually dilute solutions of sodium chloride and other salts. Vital staining Certain dyes, known as vital dyes, are taken up by living cells where they colour elements within the cell, e.g. mitochondria. Dye particles can also be taken into the cell by phagocytosis. Intravital staining refers to differentiation of cellular constituents by injecting dyes into a living animal. Supravital staining indicates staining of isolated cells, small pieces of tissues, or small organisms by placing them in isotonic saline solutions containing dyes. PERMANENT PREPARATIONS The techniques mentioned in the previous section will only last a few hours at most: to keep a permanent record they must be photographed or accurately drawn. Moreover certain cellular components are only poorly defined in such preparations. The following sections illustrate the processes involved in making permanent preparations. FIXATION The first step is to fix the tissue or specimen. Fixation is intended to terminate life processes quickly, preventing decay by bacteria and autolysis by the enzymes present in the cells themselves. Fixation, therefore, preserves the cellular components as faithfully as possible and stabilizes their condition during the steps to follow. Fixatives (or fixing agents) also harden the tissue so that it can be embedded and sectioned, and also improves the effect of certain stains. The following are a few of the fixatives in use: Desiccation This often brings about distortion in most kinds of cells, but is used for smear preparation of blood, lymph, pus, spermatozoa etc. Fixation is often completed by heat or by treatment with alcohol. Heat Heat brings about denaturation of enzymes, but also causes extreme shrinkage and distortion. It is now used in conjunction with other methods. Formaldehyde This is a gas which is soluble in water, to give formalin. It is normally employed as a 10% formalin solution. It is a strong fixing agent, rendering tissues tough and elastic. It can be used on its own or in combination with other chemicals (compound fixatives). Alcohol 70% ethanol precipitates cell proteins such as albumen and globulin in an insoluble and denatured form while it renders glycogen and nucleic acids water soluble. It tends to shrink the tissues. Compound Fixatives Choosing a fixative depends on what parts of a tissue are to be investigated. If the organisation of the tissue is important, then a micro-anatomical fixing agent is to be used. If intracellular components are to be demonstrated, then use a cytological fixing agent. Zenker’s Fluid This is a microanatomical fixative used for routine fixation. It penetrates rapidly and evenly. Fixation is usually complete after 12 hours. Zenker’s fluid consists of a solution of mercuric chloride (5g), potassium dichromate (2.5g) and sodium sulphate (1g) in distilled water (l00ml). 5mls of glacial acetic acid are added just before use. Formal-saline This is the most popular and useful of all fixatives. Tissues left in it for long periods suffer very little from its effect. Fixation time is about 24 hours. It consists of l0mls of formalin and 0.9g of sodium chloride in l00ml of distilled water. Carnoy’s Fluid Intended as a cytological fixative, it penetrates rapidly and gives excellent nuclear fixation. It can be made mixing absolute alcohol (60ml) chloroform (30ml) and glacial acetic acid (l0ml). APPLYING FIXATIVES It is not only necessary to use a suitable fixative, but also to apply it in the most effective method, so that it would be able to penetrate as rapidly as possible, without excessive distortion. As with other reagents, fixatives are to be treated carefully as many contain poisonous chemicals. Fixing by Immersion This is the usual method of fixation. With larger specimens it may be necessary to inject the fixative as well. Always use about 20 times the volume of fixative as that of the tissue to be fixed. The penetrating power determines the size of the tissue to be fixed. For example use only small pieces of tissue with low penetrating fixatives. Hot fixatives usually penetrate faster; if heat is to be used, the fixative is to be warmed before the specimen is added. Fixing by injection (Perfusion) A fixative penetrates dense or bulky tissues rapidly when it is injected into the blood cavities of body cavities. The fixative is introduced using a cannula. Washing It is often necessary to wash the fixative out of the tissues before proceeding with the preparation. For example, mercury components, unless removed, tend to form precipitates which are very difficult to remove. Washing is carried out with water or with alcohol, depending on the fixative. Small specimens, must be washed using repeated changes of water or alcohol. It is best to use running water with larger specimens. Bleaching Bleaching is carried out on specimens which contain dark pigments, in order to render the rest of the material more visible. Mayer’s Chlorine bleach is suitable for bleaching sections of skin. Decalcification It is impossible to section material which contains calcium salts, unless these are first removed. An acid is always incorporated in the decalcifying solution, together with other substances to prevent distortion. Some fixatives also contain sufficient acid to decalcify small objects during the course of fixation. Recently, chelating agents have been used to decalcify tissues. EDTA (ethylene-diamine-tetraacetic acid) is useful for this purpose. EMBEDDING AND SECTIONING The cutting of sections is the most important way in which organisms or tissues are separated into thin portions for microscopic study. Sections must be thin enough for light to pass through them. Generally, sections are one or two cells thick (about 8- 12µm). Embedding Methods Using Pith or Cork This was the earliest aid to hand sectioning, using pieces of pith or cork to support the tissues. These methods are no longer in much use. Paraffin Wax Method The tissue is impregnated with wax and so it can be cut into extremely thin sections. This method consists of the following steps: 1. fixation (and washing). 2. dehydration - because water is immiscible with paraffin wax, the fixed tissue must be dehydrated before impregnating with wax. This is carried out by transferring the tissue from 30% to absolute alcohol, passing through 50%, 70% and 90% grades. The final bath in absolute alcohol should have at least two changes. 3. “Clearing” - because alcohol itself is immiscible with paraffin wax, the dehydrated tissue must next be transferred to a “clearing agent” which are solvents of paraffin wax. They are so called because they render the tissue transparent. Chloroform, xylene and cedarwood oil are the commonest clearing agents. 4. Impregnation with wax - this is carried out in molten wax at a temperature some 2-3°C higher than the melting point of the wax (usually about 54°C). All the wax must be filtered before use, to keep the crystals as small as possible. The tissue is first transferred from the clearing agent to a solution of wax in the cleaning agent, before being actually transferred to successive baths of molten wax. 5. Block making - “blocking” is carried out by transferring the impregnated tissue to a mould containing molten wax (this may be a plastic container, paper boxes two L-shaped pieces of metal resting on a glass). The tissue is oriented in the mould and a slip of paper bearing the necessary details inserted in one corner. The wax is then cooled rapidly by blowing gently and then by playing a jet of water it. When cold, the wax block is removed from the mould. 6. Trimming the block - before sectioning, the block must be trimmed. This should be done with a safety razor blade. Trim each side straight and make the opposite sides parallel; leaving about 2 mm from the widest part of the specimen. The face of the block should be a perfect rectangle. The trimmed block should now be attached to the chuck of the microtome. The chuck should have a layer of wax on its surface. Melt this wax with a hot scalpel handle and press the block firmly on and let the wax harden. The long side of the rectangle should be parallel to the edge of the microtome knife. Sectioning Microtomes are designed to cut paraffin sections in the form of continuous ribbons. Before using the microtome, the feed mechanism should be turned back as far as it will go. The knife is then inserted and secured. It is important to tilt the knife at the correct angle. This is normally very acute (about 5°) so that the knife edge passes freely through the material rather than acting as a scraper. Take great care as microtome knives are extremely sharp. The section thickness gauge is then set to the required position (about 8-12µm). The microtome is operated until a ribbon of sections is formed. The ribbon should be lifted from the knife by means of a metal seeker, and the support maintained until the ribbon is about 30 cm long. The ribbons should be stored in a draught-free place. Attaching the sections to the slides Mount the sections on a clean glass slide using egg albumin as an adhesive. Put 1 drop on the slide and spread it evenly with a clean finger. If the film is too thick, it may subsequently take up stain and make the preparation untidy. Flood the surface of the albumenized slide with distilled water. Single or short ribbons of sections are carefully lifted with a moistened paint brush and laid on to the water, taking care not to trap any air bubbles. The slides are transferred to a hot plate (at about 50°C) and left until the wax expands. When the sections are flat, drain the water and orientate the section in the correct position. The sections are left in an incubator at 37°C to dry completely (12 hrs) before using them. Care should be taken during the process not to melt the wax. The prepared slides may be stored indefinitely in a dust free container. Other methods One disadvantage of the paraffin wax method is that relatively high temperatures are required during preparation. Therefore other methods can be used to avoid heat denaturation. One solution is to use polyester wax, which has a lower melting point than the paraffin wax (about 38°C). The Celloidin method does not require any heat, but impregnation is slow. One further method which deserves mention are frozen sections. Tissues may be frozen and sections of the frozen tissues, which are embedded in ice, cut on a special freezing microtome (cryostat). This method has the advantages of rapid preparation of sections with little fixation being necessary, so the sections can be used for histochemical studies. Frozen sections are not as satisfactory as paraffin or celloidin sections. NON-SECTIONAL METHODS FOR THE PREPARATION OF TISSUES Spreading This method is used for the preparation of membranous tissue, to prevent it from shrinking and wrinkling. For example, the membrane may be stretched between two plastic rings or a coverslip. It is then fixed before proceeding with the preparation. Smears The preparation of smears is used in the study of blood, lymph, spermatic fluid etc. as well as for the study of soft tissues such as testes and lymph glands. Smears are classified as wet or dry. Wet smears are those which are fixed when they are still moist. Dissociation of Cells The technique of dissociation or maceration of cells is of great value for the study of isolated cells. There are many methods, some of which are listed below. 1. Teasing: This is the process of separating tissue using sharp needles. This is used for fibrous tissues such as nerves and tendons, which are not extremely fragile or held together firmly. 2. Dissociation using Chemicals: The tissue is soaked in a suitable solution for a considerable length of time; this loosens the connections between the cells so that they can be easily shaken or teased apart. 3. The use of digestive enzymes: This brings about the removal of certain structures while leaving the rest of the tissue intact. For example pepsin dissolves albumens, collagen, mucin and elasten but has no effect on fats, carbohydrates, chilin and keraten. 4. Corrosion: This is the process by means of which hard structures such as sponge spicules or fish scales are separated from the soft, surrounding tissue. A 10% solution of potassium hydroxide may be employed for this purpose. STAINING The purpose of staining is to enhance natural contrast and make more evident various cell and tissue components. Stains are coloured organic substances capable of being held by the tissue components. They are in fact similar to textile dyes in their chemistry. The chemical structure which confers colour to the stain is known as a chromophore. For example, the nitro group (NO2) in trinitrobenzene gives this compound its yellow colour. Compounds containing chromophores are known as chromogens. Chromogens, though coloured, are not dyes in themselves as they have no affinity to bind to tissue constituents. To be a dye a compound must contain in addition to a chromophore another group which enables the compound as a whole to combine with the tissue, such groups are known as axochromes. Classification of Stains Stains may be classified according to the chromnophoric group present: Basic stains are those having a base as a chromophore combined with acetate, chloride or sulphate groups. Basic stains usually stain acid compounds, of cells (e.g. nuclei). Acidic stains have an acidic organic chromophore and stain basic components of cells (e.g. cytoplasm). Neutral stains contain both types of chromophores. Mordants Certain stains such as haematoxylin, will not stain tissues directly, unless a third compound is present. This compound is known as a mordant and forms a link between the tissue and the stain. The commonest mordants are the salts and alums of aluminium and iron (Fe3+). Accentuators These are substances which increase the selectivity or staining power of certain dyes, which actually combine chemically in any way. Methods of Staining 1. Progressive staining: the tissue is left in the stain until it reaches the required depth of colour, this is determined by viewing the preparation under a microscope. 2. Regressive staining: the tissue is overstained and then decolourised (differentiated) to the required depth of colour. Nuclear stains are frequently used regressively. 3. Counter staining: first part (or all) of the cell is stained with a suitable dye, then another component of the cell is stained with a contrasting dye (or the second dye displaces the first). 4. Double staining: a mixture of two dyes is used. Tissues may be stained in bulk or as sections. PROCEDURE FOR STAINING AND MOUNTING PARAFFIN SECTION The basic procedure is as follows: 1. Dewaxing: this is the removal of the paraffin wax with xylene. The slide with the attached sections is first placed on a hotplate for about 2 minutes, and then placed in the xylene. The wax should be entirely removed and the tissue becomes transparent. 2. Hydration: this is the reverse of dehydration. Most solutions of stains contain water and this is immiscible with xylene. Hence the section must be hydrated gradually by passing it through successive solutions of alcohol of decreasing concentration (absolute alcohol, 90%, 70%, distilled water). Unless hydration is carried out gradually, the cells absorb water too fast and become distorted. 3. Staining: the staining process to be carried out depends on the stain(s) to be used. 4. Dehydration: the slide is passed from distilled water to absolute alcohol through the graded alcohols. 5. Clearing: The absolute alcohol is replaced by the cleaning agent, which is miscible with the mounting medium. Xylene, benzene, oil of cloves or cedarwood oil may be used as the clearing agent. 6. Mounting: the choice of a mounting medium determines the permanency of the preparation. The mounting medium should be stable, and have no effects on the stains. The refractive index of the medium is a very important consideration. The medium should render the specimens transparent enough so that light will pass readily through them, but not to such an extent that important structures become difficult to see. Media having refractive indices close to that of crown glass are suitable. Canada balsam and D.P.X. are two mounting media in common use. A drop of the mounting medium is placed on the centre of the slide, over the section. Great care should be exercised not to trap any air bubbles. Cover with a coverslip by resting the coverslip against a finger and lowering it gently down using a mounted needle. The coverslip should settle by itself, to prevent entrance of air bubbles. Label the slide properly with the name of the tissue, method of fixation and staining and the date. Leave it to dry for at least a week.
© Copyright 2026 Paperzz