Journal of Analytical Toxicology 2012;36:112 –115 doi:10.1093/jat/bks002 Article A Gas Chromatography– Thermal Conductivity Detection Method for Helium Detection in Postmortem Blood and Tissue Specimens† Jason E. Schaff1*, Roman P. Karas1 and Laureen Marinetti2 1 Federal Bureau of Investigation Laboratory, Quantico, Virginia, and 2Montgomery County Coroner’s Office, Dayton, Ohio * Author to whom correspondence should be addressed: Jason E. Schaff, FBI Laboratory Chemistry Unit, 2501 Investigation Parkway, Rm 4220, Quantico, Virginia, 22135. Email: [email protected]. In cases of death by inert gas asphyxiation, it can be difficult to obtain toxicological evidence supporting assignment of a cause of death. Because of its low mass and high diffusivity, and its common use as a carrier gas, helium presents a particular challenge in this respect. We describe a rapid and simple gas chromatography –thermal conductivity detection method to qualitatively screen a variety of postmortem biological specimens for the presence of helium. Application of this method is demonstrated with three case examples, encompassing an array of different biological matrices. Experimental Introduction Standards and controls Standards of pure helium and pure air were prepared by filling 250-mL volumetric flasks with DI water, inverting in a DI water bath, and then bubbling in the appropriate gas to completely displace the water. The flasks were capped with rubber “turnover septum” stoppers, providing an airtight seal, before being removed from the water bath. Standards of helium in air were prepared by filling 100-mL volumetric flasks with air, as described, just to the volume mark and then injecting appropriate amounts of the pure helium standard through the stopper into each flask with a 3-cc plastic syringe. Standards were prepared at concentrations of 0.5%, 1%, 2%, 3%, and 4% (v/v), each in a separate flask. The 4% standard was prepared with two consecutive 2-mL injections. Negative control samples of DI water and of whole blood were prepared by measuring 10 mL of the appropriate matrix into 16- 100-mm culture tubes, capping with rubber septa, and then venting the headspace in each tube to vacuum for 5 s through a syringe needle. Positive control samples of DI water and of whole blood were prepared from 10-mL aliquots of matrix measured into 16- 100-mm culture tubes which were then capped with rubber septa. Each sample was two-needle sparged with a gentle flow of helium for 30 min to saturate the matrix. The vent needle was fitted with a 0.2-mm PTFE syringe filter to contain the resulting blood foam, with approximately 0.5 mL of the blood sample lost to foaming. After settling, the headspace of each sample was vented to vacuum for 10 s through a syringe needle to remove any undissolved helium. Though asphyxiation has been a well-known means of suicide for decades, deliberate use of inert gases as a means of asphyxia has become a known practice only fairly recently. The first public discussion of inert gases as a means to commit suicide seems to have been at a 1999 conference of the Self Deliverance New Technology Group, with detailed instructions published three years later in the third edition of Final Exit (1). The usual method recommended by various sources is to feed a tube from the gas source into a bag secured over the victim’s head, although some internet sources recommend use of a breathing mask. Helium is generally recommended as the “ideal” gas for suicide by asphyxia both because of its easy availability in party balloon kits and its low narcotic and hallucinogenic potential relative to other inert gases (2). Despite this fact, there are only a handful of cases of deliberate helium asphyxia reported in the medicolegal literature (3 –8). In all but one of these reports, the cause of death was determined from physical evidence found at the site of death and/or physical findings from autopsy without supporting toxicological data. Auwaerter et al. (3) reported analysis of helium by headspace gas chromatography–mass spectrometry (GC–MS), but their method requires a complex procedure for sampling gas from the lungs at the time of autopsy. Yohitome et al. (9) reported GC– thermal conductivity detection (TCD) analysis for helium in a case of accidental asphyxiation requiring a similar procedure for obtaining gaseous samples at autopsy. † Materials Carrier gas grade helium was obtained from ARCET (Fredericksburg, VA), dry air was taken from the laboratory building compressed air system, and high purity nitrogen was produced in-house via a Domnick-Hunter (Gateshead, U.K.) MaxiGas system. Drug-free human whole blood was obtained from Clinical Controls International (Los Osos, CA), and deionized (DI) water was produced in-house with a Millipore (Millford, MA) Synergy reverse osmosis system. This is publication 11-17 of the Laboratory Division of the Federal Bureau of Investigation (FBI). Names of commercial manufacturers are provided for identification purposes only, and inclusion does not imply endorsement of the manufacturer, or its products or services by the FBI. The views expressed in this article are those of the authors and do not necessarily reflect the official policy or position of the FBI or the U.S. Government. This work was prepared as part of their official duties. Title 17 U.S.C. 105 provides that ‘Copyright protection under this title is not available for any work of the United States Government’. Title 17 U.S.C. 101 defines a United States Government work as a work prepared by an employee of the United States Government as part of that person’s official duties. Published by Oxford University Press 2012. through the electrical tape, and a second piece of electrical tape was placed over the puncture after sampling. When not under analysis, liquid samples were stored at approximately 48C, and solid samples were stored at approximately –208C. Results and Discussion Figure 1. Experimental setups for producing gas standards (A) and matrix positive controls (B). Figure 1 illustrates the process for preparation of standards and positive controls. Analytical procedure CG experiments were performed on an Agilent Technologies (Wilmington, DE) 6890-N system equipped with a 30-m 0.32-mm 12-mm J&W (Wilmington, DE) HP-molesieve capillary column. Carrier gas was high purity nitrogen at 1.0 mL/min. The oven was maintained at a constant temperature of 358C with a 10 min run time. The thermal conductivity detector was maintained at a temperature of 2508C with makeup flow of 5 mL/min and reference flow of 20 mL/min, both of high purity nitrogen. The detector was set for negative polarity operation with a 5 Hz data sampling rate. For the first reported case, a purged-packed inlet was used with a temperature of 2008C. Prior to submission of the second two cases, the instrument was converted to a split/splitless injector, also operated at 2008C, with a 2:1 split ratio. All injections were performed manually using a standard Agilent 10-mL GC syringe with 10-mL injection volume. Standards were directly sampled from the flasks in which they were prepared, drawing gas aliquots through the rubber septa. The controls and samples submitted in either blood collection tubes or crimp-top vials were centrifuged 90 s at 1000 rpm to remove any biological residue from the caps and then warmed for 30 min at 35–378C. Gas aliquots for analysis were drawn through the container septa or caps. Samples submitted in metal specimen cans were allowed to stand for at least 1 h at room temperature. The lid on each can was then punctured with a hammer and metal probe, and the resulting hole was immediately covered with a piece of electrical tape. Gas aliquots were taken Method validation and performance Nitrogen was chosen as the carrier gas for this method because its thermal conductivity is much lower than that of helium while still being greater than that of most common gases. Because the thermal conductivity of the analyte of interest is higher than that of the carrier gas, it was necessary to operate the detector in negative polarity mode. Given that the response factor for a TCD is proportional to the difference between the thermal conductivity of the carrier gas and that of the eluting analyte (10), the only common gases expected to produce a positive signal response, along with their approximate response factor relative to helium at 300 K, are hydrogen (120%), neon (20%), methane (6%), and oxygen (, 1%). These response factors are only approximate because detector conditions do have some effect upon analyte response. A 2% standard of helium in air was used for optimization of instrumental method parameters. Initial development and validation was carried out with the purged-packed inlet. Two peaks were observed: one, based on the work of Yoshitome et al. (9), for oxygen and one, seen in the helium-fortified samples but not in any blank samples, for helium. Flow rates between 1.0 and 2.5 mL/min and isothermal column temperatures between 30 and 508C were investigated, with optimum peak separation obtained at a flow of 1.0 mL/min and a temperature of 358C. Upon conversion to a split-splitless inlet, these flow and temperature conditions were retained, and split ratios from 0:1 (splitless) to 10:1 were investigated. An optimal balance between peak shape and analyte response was obtained at a split ratio of 2:1. Inlet temperature and detector conditions, except for polarity, were taken from default parameters used with other methods on this instrument. The set of five standards of helium in air and an air blank were analyzed three times, twice with the purged-packed inlet and once with the split-splitless inlet. We did not attempt to quantitate any case samples because of an inability to procure or produce adequate quantitative controls of helium in biological specimens. However, results from these curves strongly suggest that accurate quantitation would be possible if this problem could be overcome. For all three curves, R 2 values of 0.99 or greater were observed, and the standard error estimate of the intercept was larger than the calculated value of the intercept. All samples except for the 0.5% standard yielded calculated values within 10%, relative to the target values. For the 0.5% standard, calculated values were all within 20%, relative to the target value. Figure 2 shows chromatograms for the series of standards on both the purged-packed and split-splitless configurations. Retention times for helium and the presumptive oxygen peak were, respectively, 3.15 + 0.03 min and 3.62 + 0.11 min for the purged-packed configuration and 2.79 + 0.01 and 3.34 + 0.01 min for the split-splitless configuration. The negative control water and blood samples showed no detectable signal at the retention time for helium, while the A Gas Chromatography –Thermal Conductivity Detection Method for Helium Detection in Postmortem Blood and Tissue Specimens 113 Figure 2. Chromatograms of standards of helium in air with a purge-packed injector (A) and with a split-splitless injector (B). Figure 3. Chromatograms of water and blood helium controls analyzed with split-splitless injection. The 4% standard of helium in air is included for reference. positive control samples showed strong helium response, much greater than the 4% helium standard. Figure 3 shows sample chromatograms for controls and the 4% helium standard. Note that there was no significant difference in helium response between the water and blood positive control samples, suggesting that there is little or no matrix effect on analyte recovery. No significant change in helium signal, measured as chromatographic peak area, was observed in either the positive controls or the 1% helium standard over the course of four days, indicating that helium is reasonably stable in samples stored under the conditions given in the Experimental section. Example cases Case 1. The decedent was a white male, 39 years old, who was estranged from his wife. His wife received a letter from him two days after the last time he was known to be alive. The letter stated he would be dead by the time she received it. The wife called the local police and requested an officer go to her husband’s residence to check on him. 114 Schaff et al. Figure 4. Chromatograms of a 4% standard of helium in air, a negative control blood sample, and the blood, brain, and lung specimens from case example #2. Responding officers found that he was deceased, seated in a chair. He had a plastic bag over his head, and the bag was secured with a Velcro strap. The decedent had a flexible hose in his mouth and the other end of the hose was attached to a helium tank. The valve on the tank was open. Empty beer cans and an empty Jack Daniels bottle were found near the body. The box for the helium tank was found in a trash can. A purchase receipt was not found. The autopsy results were as follows; generalized visceral congestion with severe congestion and edema of lungs (right lung 725 g, left lung 635 g) and hypertensive and arteriosclerotic cardiovascular disease with cardiomegaly (405 g). The atherosclerosis disease involved the left anterior descending coronary artery, which was 20 –30% narrowed by plaque in its mid section, the circumflex coronary artery, which was 20– 30% narrowed by plaque in its mid section, and the right coronary artery, which was 20 –30% narrowed by plaque in its mid section. Blood in a red-top Vacutainer tube, liver and lung in metal specimen cans, and a portion of brain in a 20-mL crimp-top vial were submitted for helium analysis. Toxicology results showed the presumptive presence of helium in the lung, brain, blood, and liver. Toxicology results were negative for volatiles and drugs of abuse. The cause of death was determined to be asphyxia due to helium inhalation. Case 2. The decedent was a 56-year-old male with a recent job loss and in the process of forced eviction. He was found with a bag over his head with a hose attached to a helium tank. The decedent’s pet cat was found in a similar situation. A copy of Final Exit opened to the proper page was also found at the scene. The autopsy showed signs consistent with asphyxia. The toxicology results were as follows: ethanol in the blood and urine both with a concentration of 0.14 g%, doxylamine in the blood at less than 0.05 mg/mL and positive in the urine, and 35 ng/mL carboxy THC in blood and positive in the urine with no parent THC in blood. Blood in a grey-top Vacutainer tube and brain and lung in metal specimen cans were submitted for helium analysis. Toxicology results showed the presumptive presence of helium in the lung but not in the blood or brain, as shown in Figure 4. Case 3. The decedent was a 20-year-old white male college student. He had some psychiatric problems and had attempted suicide by means of an overdose in the past. The decedent had to drop his classes because of his mental status, and his roommate had also told him he did not want to room with him any longer. He was taking lithium and clonazepam and had been seeing a psychiatrist since his past overdose attempt. He told another student he was going to use a helium tank to take his life. That student called the decedent’s roommate, who was off on spring break, and he in turn called campus police. Campus police checked on the decedent in the morning and spoke with him, and he denied making any threats. Later in the afternoon, campus housing personnel called the decedent’s mother and requested that she ask her son to come home because of his threats and past conduct. She asked if they could explain the situation to him. They went to his apartment on campus, and when they received no response, they used a key to gain entry, found him lying unresponsive in bed, and called 911. Emergency medical services and campus police responded and pronounced him dead. The decedent had a helium tank between his legs with an attached hose that extended to his face. His head was wrapped in a plastic bag with a piece of clothing wrapped around his neck to effect some type of seal. There were several notes found at the scene, one addressed to his parents and one to his roommate. The autopsy results were as follows: an abrasion to the frontal scalp, atherosclerotic cardiovascular disease, and cardiomegaly (460 g) with left ventricular hypertrophy. The atherosclerosis consisted of left anterior descending and right main coronary arteries with 10– 70% narrowing. There was also papillary thyroid carcinoma. Toxicology results included lithium in blood of 0.29 mEQ/L. Blood in a grey-top Vaccutainer tube and portions of liver and brain in 20-mL crimp-top vials were submitted for helium analysis. Helium was not detected in the blood, brain, or liver. The cause of death was determined to be asphyxia. Conclusions A simple and rapid GC –TCD method to screen for the presence of helium in biological samples was developed and applied to a variety of matrices from several postmortem cases. This method does not require that any specialized sample collection techniques be used at autopsy, only that samples be promptly and completely sealed and that they be stored refrigerated or frozen, as appropriate. Not surprisingly, based upon the cases examined, lung tissue appears to be the best matrix for analysis, although detection is clearly possible in other samples. This method is purely for screening purposes, to provide analytical evidence that may or may not support physical evidence found at a death scene. With only a chromatographic retention time, it is not possible to establish an incontrovertible identification for helium. The fact that only a very small number of gases are physically capable of producing a chromatographic peak in this analysis does, however, mean that false-positive results for unputrefied specimens are extremely unlikely. References 1. Humphry, D. Final Exit, 3rd ed. Random House: New York, NY, 2002; pp 130–140. 2. Bennett, P.B.; Roxtain, J.C. Inert gas narcosis. In Physiology and Medicine of Diving, Brubank, A.O.; Neuman, T.S., Eds.; Elsevier Science Ltd., Edinburgh U.K., 2003; pp 300–322. 3. Auwaerter, V.; Perdekamp, M.G.; Kempf, J.; Schmidt, U.; Weinmann, W.; Pollak, S. Toxicological Analysis after Asphyxial Suicide with Helium in a Plastic Bag. Forensic Sci. Int. 2007, 170, 139–141. 4. Schon, C.A.; Ketterer, T. Asphyxial Suicide by Inhalation of Helium Inside a Plastic Bag. Am. J. Forensic Med. Pathol. 2007, 28 (4), 364–367. 5. Gallagher, K.E.; Smith, D.M.; Mellen, P.F. Suicidal Asphyxiation by Using Pure Helium Gas: Case Report, Review, and Discussion of the Influence of the Internet. Am. J. Forensic Med. Pathol. 2003, 24 (4), 361–363. 6. Ogden, R.D.; Wooten, R.H. Asphyxial Suicide with Helium and a Plastic Bag. Am. J. Forensic Med. Pathol. 2002, 23 (3), 234–237. 7. Grassberger, M.; Krauskopf, A. Suicidal Asphyxiation with Helium: Report of Three Cases. Wien Klin. Wochenschr. 2007, 119 (9), 323–325. 8. Ogden, R.D.; Hamilton, W.K.; Whitcher, C. Assisted Suicide by Oxygen Deprivation with Helium at a Swiss Right-to-Die Organization. J. Med. Ethics 2010, 36, 174– 179. 9. Yoshitome, K.; Ishikawa, T.; Inagaki, S.; Yamamoto, Y.; Miyaishi, S; Ishizu, H. A Case of Suffocation by an Advertising Balloon Filled with Pure Helium Gas. Acta Med. Okayama 2002, 56 (1), 53 –55. 10. Colon, L.A.; Baird, L.J. Detectors in Modern Gas Chromatography. In Modern Practices of Gas Chromatography, Grob, R.L.; Barry, E.F., Eds.; John Wiley & Sons: Hoboken, NJ, 2004; pp 289– 298. A Gas Chromatography –Thermal Conductivity Detection Method for Helium Detection in Postmortem Blood and Tissue Specimens 115
© Copyright 2026 Paperzz