endo® MED - Dr. Weigert

Endoscopy
neodisher
endo®
MED
Detergent and Disinfectant for the
Manual Pre-Treatment of Flexible Endoscopes.
Application and Dosage:
neodisher endo® MED
• neodisher endo® MED is used in immersion baths or
ultrasonic baths. Depending on the desired activity a solution
is prepared in accordance with the application recommen dations. For disinfection place endoscopes or endoscopic
accessories into the solution whilst observing the contact
time. All surfaces must be completely wetted with the solution.
Air bubbles must be removed. When used in ultrasonic baths
the treatment time should not exceed the time given by the
instrument manufacturer.
Aldehyde-free and alkylamine-free detergent and disinfectant for the manual pre-treatment of flexible endoscopes;
with excellent material compatibility. Compatible with
disinfectants containing aldehyde and peracetic acid for
subsequent reprocessing.
• For the disinfecting pre-cleaning of flexible endoscopes
and endoscopic accessories in immersion baths or
ultrasonic baths
5L
1L
1051xLL
1L
500
2 L 0,75 L 1 L ml 1 L 1 L
1 L100
ml
Hy
•
her
neodis
endo®
y
10 x 1 L bottle
4210 45
100 g
om
L p 500
e c0,75
e ml
n ce a n d t
5 L canister
4210 35
n
ie
1L
100rips
50 st
ml
100 g
Disinfecting pre-cleaning
bactericidal, yeasticidal, active against enveloped viruses (incl. e.g. HBV, HIV, HCV)
10 ml/l (1.0 %),
15 min
bactericidal
(according to VAH1, EN 13727, EN
14561), dirty conditions
10 ml/l (1.0 %),
15 min
yeasticidal
(according to VAH1, EN 13624, EN
14562), dirty conditions
10 ml/l (1.0 %),
15 min
active against enveloped viruses
(incl. HBV, HCV, HIV) (according to
RKI/DVV2), dirty conditions
10 ml/l (1.0 %),
15 min
50 strips
te
g
2L
Application Recommendation (20 °C):
Excellent material compatibility –
suitable for endoscopes of all leading
manufacturers
te d saf et
Our Competence – Your Benefit:
neodi
is comsher endo ®
disinfe patible withMED
cting a
a
ctive s ll common
ubstan
ces
• excellent cleaning performance; does
not fixate proteins
• facilitates manual cleaning; optimum pre-treatment
• disinfecting action confirmed in accordance
with EN standards, VAH1 and RKI/DVV2 methods
• increased personnel protection; improves the hygiene safety when
reprocessing; VAH1 listed; included in the IHO4 list of virucidal products
• free of aldehydes and alkylamines
• compatible with disinfectants containing aldehydes and peracetic acid for
subsequent reprocessing steps; no layers/discolourings on endoscopes
• without CMR3 disinfecting active substances
• especially user-friendly
• excellent material compatibility
• suitable for stainless steel, anodized aluminium, non-ferrous metals
(copper and brass), synthetic materials (incl. silicone)
GB 1229 / 05-2016 © Dr. Weigert, Hamburg
105LL
es
5L
neodisher endo® MED – manual disinfecting
pre-cleaning compatible with all active substances.
2
Verbund für Angewandte Hygiene [Association of Applied Hygiene]
Testing guidelines of the Robert Koch Institute and the Deutsche Vereinigung zur Bekämpfung von Viruskrankheiten [German Association
3
4
Carcinogenic, Mutagenic, toxic to Reproduction Industrieverband Hygiene und Oberflächenschutz [German Association for Hygiene and Surface Protection]
for the Control of Virus Diseases] 1
Your expert for microbiological questions: Verona Schmidt · [email protected]
Chemische Fabrik Dr. Weigert GmbH & Co. KG • Mühlenhagen 85 • D-20539 Hamburg • Tel.: +49-40-7 89 60-0 • Fax: +49-40-7 89 60-120 • [email protected] • www.drweigert.com