Application Note SI-01741 Analysis of Greenhouse Gases by Gas Chromatography Coen Duvekot Varian, Inc. Introduction Nitrous oxide (N2O), carbon dioxide (CO2) and methane (CH4) are considered to be very powerful greenhouse gases. These gases absorb the sun’s rays and thus, influence the global heat budget. Continuous measurement provides insights into the sources of the different gases. A Varian 450-GC based system was custom configured to analyze all three gases in one run in a matrix of atmospheric air containing water vapor. Conditions Sample: Concentrations: Front EFC: Middle EFC: Rear EFC: Front TCD: Middle FID: Rear ECD: Instrumentation Technique: Varian 450-GC Gas Chromatograph Injection: Headspace sampler into a 1041 injector Detection: TCD, ECD and FID Table 1. Valve event table. Vent 1 V 3 Sample loop 500 µL 1041 Sample line purge Helium 10 1 9 Vent 2 8 7 Flowcontroller 5 6 EFC 24 front Helium Column A : Hayesep N , 80 /100 0.5m x 1/8" SS 10 1 9 3 V 1 Vent 3 4 8 7 EFC 23 middle Nitrogen or Argon / Methane Column B : Porapak QS , 80 /100 2m x 1/8" SS 2 3 V 2 6 5 Column C : Hayesep N , 80 /100 0.5m x 1/8" SS TCD Reference TCD DEFC 13 front Carbon Dioxide Sample OUT Sample loop 500 µL 2 Spiked lab air CO2 0.4 % CH4 1775 ppb N2O 320 ppb 17 psi, 40 mL/min total flow 14 psi, 40 mL/min total flow 29 psi, 50 mL/min total flow Filament Temperature: 200 °C Block Temperature: 120 °C Carrier Gas: Helium 300 °C 300 °C Cell Potential: 415 mV Make Up Flow: 2.0 mL/min 4 EFC 24 rear Nitrogen or Argon / Methane Column D : Porapak D , 80 /100 2m x 1/8" SS FID DEFC 11 middle ECD DEFC 13 rear Methane Nitrous Oxide Figure 1. Plumbing scheme for the analysis of greenhouse gases. Time GSV-1 GSV-2 BF to Vent V-4 Initial Fill Fill Forward OFF 0.01 Fill Fill Forward ON 1.00 Inject Inject Forward ON 1.20 Inject Inject Forward OFF 2.05 Inject Inject BF OFF 2.70 Inject Inject BF OFF Results and Discussion When conditions were set, as outlined in Table 1, three chromatograms were obtained. Figure 2 shows the front channel chromatogram where CO2 was analyzed. In essence, two channels were used, one of which was equipped with two detectors. After injection on the first channel, CO2 and methane were separated from air. The TCD detected the CO2 and the FID plumbed in series with the TCD detected the methane. On the second channel, the nitrous oxide was separated from water. Water was backflushed to vent and the nitrous oxide detected on an ECD. Materials and Reagents Packed columns: HayeSep® N 80/100 mesh, 0.5 m x 1/8 in. SST Porapak® QS 80/100 mesh, 2 m x 1/8 in. SST HayeSep® D, 80/100 mesh, 2 m x 1/8 in. SST Application Note SI-01741 CO2 0 1 2 3 min 4 5 6 Figure 2. CO2 analyzed on the front detector (TCD). 1 of 2 www.varianinc.com Methane was analyzed on the middle channel via FID detection (Figure 3). Table 2. Repeatability data on peak area. Methane 0 1 2 3 4 5 6 min Run CO2 CH4 N20 1 13451 2947 900 2 13472 2986 913 3 13477 2928 935 4 13497 2961 929 5 13450 2933 916 6 13482 2996 922 7 13547 2974 875 8 13508 2972 901 9 13535 2967 904 10 13390 2982 904 Average Std Deviation RSD % 13481 45.31 0.34 2965 22.51 0.76 910 17.07 1.88 Figure 3. Methane analyzed on the middle detector (FID). Figure 5 provides a visual impression of the repeatability data with 10 % variation limits. Finally, N2O was analyzed on the ECD detector mounted in the rear position (Figure 4). Peak Area 16000 14000 12000 CO2 CH4 N2O 10000 8000 6000 N2O 4000 2000 0 0 0 1 2 3 4 5 6 2 4 6 8 10 Run Figure 5. Repeatability data with 10 % variation limits. min Figure 4. Nitrous oxide analyzed on the rear detector (ECD). To show system suitability, a sample was tested several times. Repeatability data generated on peak area are shown in Table 2. The headspace sampler permits the analysis of samples of different origin, such as air, water and soil. On the ECD detector, N2O is analyzed. This, however, can easily be expanded with CFCs and SF6. Repeatability data was very good. RSD on peak area below 2 %, for this type of analysis, is excellent. Conclusion The Varian 450-GC based system was custom configured in order to analyze greenhouse gases, carbon dioxide, methane and nitrous oxide, in one run. The headspace sampler permitted the analysis of samples of different origin. Air, soil and water samples can be handled in this way. Repeatability data indicated that the system was perfectly suited to greenhouse gas analysis. This system is highly flexible and the application range can be expanded quite simply to include CFCs and SF6. These data represent typical results. For further information, contact your local Varian Sales Office. Varian and the Varian Logo are trademarks or registered trademarks of Varian, Inc. in the U.S. and other countries. HayeSep and Poropak are registered trademarks of Hayes Separations, Inc. © 2008 Varian, Inc. Application Note SI-01741 Varian, Inc. www.varianinc.com North America: 800.926.3000 – 925.939.2400 Europe: The Netherlands: 31.118.67.1000 Asia Pacific: Australia: 613.9560.7133 Latin America: Brazil: 55.11.3238.0400
© Copyright 2025 Paperzz