PPB Dissolved Oxygen Measurement Calibration and Sampling Techniques Introduction The water is used as a rinse agent in the clean room process, The amount of dissolved oxygen in process water is continually which is why there is a focus on ever-tightening standards for gaining importance in many industries as a critical parameter to controlling dissolved oxygen. Dissolved oxygen, when allowed to be measured and controlled. The semiconductor and power remain in solution, can cause oxidation in the mixed bed resins industries are continuously requiring increased accuracy in and can then be carried over to the clean room chip dissolved oxygen detection as lower levels of oxygen are manufacturing level. Upon start-up of this facility in 1995, there required in the process water. The purpose of the on-line, was a dilemma in establishing a baseline for dissolved oxygen dissolved oxygen measurement is to continuously monitor the removal. Once a baseline was established, it would have to be process water to ensure that the level of oxygen is within accurately verified with a known standard, thus assuring optimum specifications. The dissolved oxygen concentration is typically system performance. measured at multiple points to determine the efficiency of the means used for oxygen removal - typically some form of The vacuum degasification system removes the dissolved mechanical degasification/deaeration and/or chemical injection - oxygen from the process water (Figure 1). Measurements are as well as to detect process upsets or leaks. However, plant taken before and after the degasifiers, as well as on the supply personnel are often frustrated with the process of trying to obtain loops feeding the factory. Although the vacuum degasification quantitative results for the measurement that are useful and system under peak performance is capable of producing water in meaningful. Many of the proper techniques used to actually the 9 parts per billion (ppb) range, many variables must be met in obtain and correctly measure the dissolved oxygen concentration order to maintain these operating specifications. The readings in a representative sample of the process water are not well after Vacuum Degasifier #1 consistently read within the 5-20 ppb understood, and the personnel responsible for the measurement range, while the readings after Degasifier #2 read in the 25-50 are often very distrustful of the instrumentation as a result. When ppb range. Process streams which should have been identical these techniques, consisting of proper calibration and sampling consistently read a difference in dissolved oxygen concentration practices, are learned and understood, they can be used to of more than 20 ppb. The sampling lines and instrumentation effectively discern whether or not the dissolved oxygen levels in were always suspect since consistent and repeatable results their process meets the specifications required. A semiconductor were difficult. In order to ensure that the dissolved oxygen plant in Austin, Texas is one such example of a site where these removal system was working correctly, the meters and sample techniques were effectively utilized to validate their sample lines lines would have to be proved first. To do so, a procedure was and measurement instrumentation, and in doing so discovered a needed to test and prove that the readings were representative of problem with a process component. the true oxygen concentrations. This was done by validating calibration methods and verifying that the sample lines were not The Problem The aforementioned semiconductor facility is one of the leaders in the semiconductor manufacturing industry, and as such focuses on producing the most advanced product with a major emphasis on quality. Ultrapure water is one asset that allows this facility to achieve this with minimal impact to its people and the environment. contaminating the reading. PPB Dissolved Oxygen Measurement - Calibration and Sampling Techniques 2 Calibration The relationship between oxygen partial pressure and total Dissolved oxygen probes designed for measurements in the atmospheric pressure should be understood and incorporated parts per billion range generate a current proportional to the into the air calibration in order to minimize calibration error, which oxygen partial pressure. For the typical levels of oxygen partial could be as high as 5-10% dependent upon altitude and local pressure present in the process and calibration mediums used in weather conditions. Most dissolved oxygen meters that have any the semiconductor industry, the signal produced by the probe is sort of advanced air calibration (such as temperature essentially a linear function of the partial pressure.1 In order to compensation, which will be discussed in a later section) will be obtain an accurate representation of the dissolved oxygen, based upon an atmospheric pressure of 760 mmHg. Most tables current values must be calibrated to the partial pressures which of oxygen solubility are referenced to this value.3,4 Because of they represent. Few dissolved oxygen instruments display the the change in oxygen partial pressure with changes in partial pressures of oxygen, however. Rather, the dissolved atmospheric pressure, a correction must be made when the oxygen concentration, which is directly proportional to partial pressure varies from this value. A simple means of incorporating pressure of oxygen for the range of interest , is the commonly pressure changes is listed in the “correction factor” shown in displayed unit of measure, and will be used in the following Table 1. The value listed is a rough multiplier which can be used discussion along with partial pressure. once the initial oxygen concentration is determined based upon 2 temperature and relative humidity. A more accurate calculation The common calibration method for most dissolved oxygen for incorporating pressure will be discussed after relative humidity systems involve exposing the probe in air, since air will always and temperature effects are investigated. have a known partial pressure of oxygen dependent upon atmospheric pressure as well as temperature-dependent A few newer dissolved oxygen meters contain a pressure humidity, if present. Air-saturated water can also be used for sensing device which provides compensation for pressure effects calibration, but guaranteeing a precise partial pressure of oxygen when an air calibration is performed. Since most meters do not in the water is difficult and problematic. Air calibration remains have this, it is usually necessary to note the average pressure in the tried and true procedure for calibrating dissolved oxygen the local vicinity of the probe, which will be mostly altitude-based, probes. However, air calibration can easily can be done and adjust the calibration using the simple correction factor or the incorrectly, yielding significant error if the characteristics of the air more complex calculation performed later. A mercury barometer in the immediate vicinity of the probe are not taken into located in the immediate vicinity of the meter will give a relatively consideration. It is vitally important that the user understand the accurate measurement of the local atmospheric pressure if an effects caused by variations in any of these variables in order to older meter with no pressure sensor is used. ensure an accurate and consistent air calibration. Relative Humidity and Temperature Pressure The discussion of pressure effects were based upon atmospheric The constituents of air have been well defined, and it is known pressure with dry air (no moisture content). Whenever air that air contains 20.946% oxygen. Since the total pressure in the contains a certain amount of moisture, the atmospheric pressure air is the sum of all of the partial pressures (Dalton’s Law), an contains another source of partial pressure -- water vapor. If a atmospheric pressure of 760 millimeters Mercury (mmHg) in dry comparison of the oxygen partial pressure in air with 100% air will contain a partial pressure of oxygen (pO2) of relative humidity and air with 0% relative humidity is done while approximately 159 mmHg (760 mmHg * 0.20946). Changes in both are at the same atmospheric pressure, the air with 100% atmospheric pressure will cause a directly proportional change in relative humidity will have a lower oxygen partial pressure due to the partial pressure of oxygen in the air. Atmospheric pressures the presence of the water vapor pressure (pH2O). Water vapor will vary depending upon altitude and local weather conditions. pressure in air varies with temperature, and is well defined.3 The Some average pressures for varying altitudes are listed in Table effect of temperature on oxygen partial pressure in moist air is 1. such that higher temperatures yield lower oxygen partial 3 pressure, while lower temperatures yield higher pressures. This relationship is shown in Figure 2. PPB Dissolved Oxygen Measurement - Calibration and Sampling Techniques 3 Note that the effects of relative humidity and temperature can Equation 1 should be used with air with 100% relative humidity, cause errors when air calibration is performed in dry air, since and Equation 2 should be used for air with 0% relative humidity. most of the current tables and meter temperature compensations are based on air containing 100% relative humidity. Table 2 Equation 1 (100% Relative Humidity): S = (S’) * (P - p) / (760 - p) shows both the oxygen concentration, which is linear with the partial pressure of oxygen, that would be present at 100% where: relative humidity and 0% relative humidity. The values only differ by a few percent in ambient air conditions, and thus is generally S = Oxygen solubility at barometric pressure of interest ignored. Most dissolved oxygen meters have temperature compensation for air at 100% relative humidity, and no manual S’ = Oxygen in saturation at one atmosphere (760 mmHg) at a correction is necessary. However, many older meters do not given temperature have temperature compensation included, and therefore this calculation must be done manually. If temperature is not P = Barometric pressure of interest compensated for in the calibration, the error can be as much as 20 to 30 % for every 10 degrees difference from 25º C, and p = Vapor pressure of water at the temperature of interest therefore temperature compensation is standard on most dissolved oxygen meters today. Since the effects of relative Example 1: The user wishes to calibrate a dissolved oxygen humidity is minimal at all but the highest temperatures, no current probe in air at an altitude of 3500 feet. The temperature is 30º C, dissolved oxygen meters incorporate any kind of relative humidity and the relative humidity is 100%. sensing device. At an altitude of 3500 feet, the atmosphere pressure will usually In order to ensure an accurate temperature and current reading, be about 668 mmHg (Table 1). The sample temperature is 30º C, the probe must be exposed to the air for enough time to allow and the relative humidity is 100%. From water vapor pressure thermal equilibrium to occur. There are often significant tables, the water vapor pressure at 30º C is 31.8 mmHg. The temperature differences between the process water and the oxygen saturation level at 760 mmHg and 30º C is 7.54 ppm ambient air. Larger temperature gradients between the two (Table 2). Substituting these values in the above equation gives necessitate additional time for thermal equilibrium to take place. the following: 3 For instance, a 20º C difference between ambient air and process water can cause a calibration delay of about 30 minutes S = (7.54) * (668 - 31.8) / (760 - 31.8) = 6.59 ppm in many probes for the probe to fully equilibrate to ambient temperature. Since most temperature gradients will not be this Example 2: Assume the same conditions as in example 1, but large, allowing approximately 15 minutes is usually a safe with a relative humidity of 0%. In this case, the value used for the assumption. It is common for users to calibrate the unit before oxygen saturation level would be 7.87 (Table 2), not 7.54. The the dissolved oxygen meter is reading the stabilized temperature calculation will change since there will be no water vapor and current value, which can cause significant error since a pressure. difference of even 5º C from actual can cause the reading be off by 5 to 10%. It is often useful to have a calibrated temperature Equation 2 (0% Relative Humidity): S = (S’) * (P) / (760 mmHg) sensor, accurate to 1º C or better, at the calibration location to know when the probe temperature is reading the correct ambient Substituting the above values into the equation yields the air temperature. following: It is useful to have an equation which can be used to determine S = 7.87 * (668) / (760) = 6.92 ppm oxygen concentrations in air based upon temperature, relative 2 humidity, and pressure. Since the full equation is quite lengthy Note that the multiplier of (668) / (760) is actually the simplified and complex, two easier versions are presented to the user, correction factor listed in Table 1 for an altitude of 3500 feet along with Table 2, to determine the correct oxygen (0.88). Table 3 lists calibration values for varying temperatures concentration in air. pressures at relative humidity levels of 100%. PPB Dissolved Oxygen Measurement - Calibration and Sampling Techniques 4 Calibration Procedure Effects such as long recovery times from air calibrations, errors in The following steps can be followed to ensure an accurate comparisons between on-line probes and wet chemistry tests, calibration. and incorrect readings of the actual process dissolved oxygen levels are typical examples which affect more sampling systems 1. 2. 3. 4. 5. Expose the probe to air. Since meters that incorporate temperature for calibration purposes base the compensation on 100% R.H. air, the calibration will be most accurate whenever the air around the probe is high in relative humidity. Wait until the dissolved oxygen and temperature readings stabilize. This usually takes about 5 to 45 minutes, depending upon the temperature differences between process water and air, as well as the oxygen difference between the process and the air. If the meter does not have a temperature reading, it may not have a temperature sensor either, in which case the probe will be ready for calibration as soon as the dissolved oxygen reading stabilizes. When no temperature reading is available from the meter, the user should have an accurate temperature sensor to determine the ambient air temperature for calculation purposes. Compare the temperature and pressure reading on the meter to a calibrated temperature and pressure reading in the same vicinity of the probe. If the meter doesn’t have a pressure sensor available, use a local mercury barometer to determine the atmospheric pressure. If a barometer is not available, just use the value from Table 2 associated with the altitude at the location of the meter. If the pressure and temperature agree with the local readings, then perform the air calibration. If the temperature and pressure reading on the meter differ significantly (pressure by more than about 5 mmHg or temperature by more than about 2º C), recalibrate them, if possible, or adjust the readings manually. If the temperature and pressure readings are not included with the meter, the manual calculations described earlier should be used to determine the correct calibration value. Place the probe back in the process. It is useful to adjust the flow up to 300 ml/min or more for the first few hours in order to flush out any oxygen that may have become entrained as a result of the air calibration. than users know. Sampling lines often have dead legs (spaces where stagnant water or air is present), trapped air bubbles, and/or leaks in them. Deadlegs are found in areas of piping that do not always have flow through them, causing an area of stagnant water to be present. This can result in an area of trapped air, which will cause a seemingly high dissolved oxygen reading. Air bubbles can be found in piping that is flowing in a downward direction. The natural buoyancy of air will cause the bubbles to resist the flow of water in downward flowing piping at lower flow rates. Deadlegs and air bubbles will eventually dissolve (they will dissolve faster at higher flow rates), but may take hours or days to dissolve if the flow rate is low (100 ml/min or less). While air bubbles will eventually dissolve, leaks will continuously cause erroneously high readings. While leaks are usually minimized by higher line pressures due to successive stages of smaller sampling lines, their presence can still cause large errors in readings, especially at the low levels of oxygen present in most power plant and semiconductor process lines. A way of testing for leaks is to increase the flow by 50 to 100% and observe the dissolved oxygen reading for the total length of time that it would take water to travel from the process to the probe. If a drop in reading is noticed, it may be due to a leak. The flow should be returned to normal for 10 to 20 minutes, and the test repeated. If the same drop in reading is noticed again, there is most likely a leak somewhere upstream of the probe. Higher flow rates can help to minimize leaks, since higher flow rates will dilute the oxygen ingress. Leaks can be caused by nonmetallic piping, fittings, rotameters, and even probe mountings. Flowmeters, a common cause of leaks, can often be placed after the probe and perform the same function without introducing a leak until after the probe reading. Non-metallic sample lines have been observed to allow oxygen ingress even when new, and often get worse over time, so they should be replaced with Sampling stainless steel piping. Loose fittings are also a common source Correct sampling is perhaps the most misunderstood for leaks and may need to be redone with additional teflon tape requirement in obtaining valid dissolved oxygen readings at low or sealant of some sort. Incorrect mounting of the probes are ppb levels. common and can cause leak problems as well. The manufacturer of the probe should be contacted to ensure that the correct procedure is being followed for mounting the probe. PPB Dissolved Oxygen Measurement - Calibration and Sampling Techniques Sampling for comparison readings with a portable meter should The sampling lines were entirely stainless steel, but leaks were always be done in parallel lines and never in series, as one probe suspected from fittings and sample line flowmeters. Leak testing will contaminate the downstream reading. Wet chemistry was done with little change in reading, thus indicating that the readings should always be done in parallel lines as well. Most discrepancies between Degasifier #1 and #2 were probably not users have a separate line for these readings which is turned on due to sample line leaks. As a precaution, the sampling system for a few minutes at 200 ml/min or less before the reading is was further pressurized in order to eliminate any potential leaks made. Doing such will often result in a higher-than-actual as a possible point oxygen filtration into the lines. At this point, reading, since even a few hours at this flow rate may not be readings were trended and the data logged to pinpoint possible enough time to totally remove all the entrapped air or air bubbles discrepancies and changes occurring within the system. From in the sample line. In addition, a period of time (approximately 5 this information and a new confidence in the sampling system minutes) is often needed in many sample lines just to ensure that and analyzers, discrepancies between supposedly identical the sample contains current process water, and not process streams pointed to the vacuum pump system supplying the water from the last test. At a minimum, it is recommended that degasifiers. 5 the wet chemistry sample line be running for at least four hours at a high flow rate (300 ml/min or more) before sampling in order to This system is vital in order to maintain levels of vacuum useful ensure a good reading. More extensive work on proper wet to the process, and is assisted by seal water and eduction chemistry sampling has been reported elsewhere. 4,5 venturis. By checking vacuum levels in the degasifiers and proving seal water flow rates to the pumps, all indications point to Results the eductors at the pumps. It was determined that the cause was Based on the information provided regarding proper pressure drops in the eductors, causing ice buildup which measurement system calibration, a procedure was developed to plugged the orifice holes, drastically reducing pump performance. ensure consistent, accurate air calibrations. The meter used Elimination of this problem caused readings at the outlet of employs temperature compensation for air with 100% relative Degasifier #2 to change from the 25 to 50 ppb range to peak humidity. No pressure compensation was available on the model performance of about 10 to 15 ppb. This proved beneficial used, so the air calibration was adjusted manually if pressure through money savings, system balancing, and improved deviated from the norm. Complete electronic calibrations were process quality, which in turn has gained the trust for the also done on the dissolved oxygen analyzers, with several instrumentation measuring such a critical parameter in the discrepancies found and repaired at that time. After full electronic system. As technology continues to advance, even lower levels and air calibrations, the meters and probes sampling the outlets of dissolved oxygen will become more critical to detect for of Degasifier #1 and #2 were switched. The readings at the applications in today’s competitive semiconductor market. This sample points remained the same, verifying that the probes and means that baselines must first be established for allowable meters were working correctly. This pointed to the sampling lines dissolved oxygen concentrations in the process. However, potentially being the source of the problem. specifications concerning process equipment and all controlling variables must be addressed prior to questioning the integrity of the dissolved oxygen analyzers and sample lines at this semiconductor manufacturer’s facility. PPB Dissolved Oxygen Measurement - Calibration and Sampling Techniques 6 Ambient Air Water from UV Sterilizers Eductor 2 Dissolved Oxygen #1 1 Vacuum Degasifier Vacuum Pumps Dissolved Oxygen #2 Water Out 2 Process Water 1 Figure 1: Dissolved Oxygen Sample Points for Vacuum Degasifier Calibration Correction Altitude (ft) Pressure Factor -540 775 1.02 Sea Level 760 1.00 500 746 0.98 1000 732 0.96 1500 720 0.95 2000 707 0.93 2500 694 0.91 3000 681 0.90 3500 668 0.88 4000 656 0.86 4500 644 0.85 5000 632 0.83 5500 621 0.82 6000 609 0.80 (mm Hg) Table 1: Oxygen Value Corrected for Pressure (25º C)4 PPB Dissolved Oxygen Measurement - Calibration and Sampling Techniques 7 180 160 140 Partial Pressure, mmHg 120 100 pH2O 80 pO2, 100% R.H. 60 pO2, 0% R.H. 40 20 0 0 10 20 30 40 Temperature, Degrees Celsius Figure 2: Partial Pressures of Water Vapor and Oxygen vs. Temperature 50 PPB Dissolved Oxygen Measurement - Calibration and Sampling Techniques 8 Temperature DO (100% R.H.) DO (0% R.H.) Temperature DO (100% R.H.) DO (0% R.H.) (Degrees Celsius) (ppm, mg/L) (ppm, mg/L) (Degrees Celsius) (ppm, mg/L) (ppm, mg/L) 0 14.6 14.66 26 8.09 8.37 1 14.19 14.26 27 7.95 8.24 2 13.81 13.89 28 7.81 8.12 3 13.44 13.53 29 7.67 8.00 4 13.09 13.18 30 7.54 7.88 5 12.75 12.85 31 7.41 7.77 6 12.43 12.54 32 7.28 7.66 7 12.12 12.23 33 7.16 7.56 8 11.83 11.94 34 7.05 7.46 9 11.55 11.66 35 6.93 7.37 10 11.27 11.40 36 6.82 7.27 11 11.01 11.14 37 6.71 7.18 12 10.76 10.90 38 6.61 7.10 13 10.52 10.66 39 6.51 7.01 14 10.29 10.44 40 6.41 6.93 15 10.07 10.22 41 6.31 6.85 16 9.85 10.01 42 6.22 6.78 17 9.65 9.82 43 6.13 6.70 18 9.45 9.63 44 6.04 6.63 19 9.26 9.45 45 5.95 6.56 20 9.07 9.27 46 5.86 6.49 21 8.90 9.11 47 5.78 6.43 22 8.72 8.95 48 5.70 6.36 23 8.56 8.80 49 5.62 6.30 24 8.40 8.65 50 5.54 6.24 25 8.24 8.51 Table 2: Dissolved Oxygen Solubility vs. Temperature2,3 PPB Dissolved Oxygen Measurement - Calibration and Sampling Techniques 9 790 775 760 745 730 715 700 685 670 655 0 15.18 14.89 14.60 14.31 14.02 13.73 13.44 13.15 12.86 12.57 1 14.75 14.47 14.19 13.91 13.63 13.34 13.06 12.78 12.50 12.22 2 14.36 14.08 13.81 13.54 13.26 12.99 12.71 12.44 12.16 11.89 3 13.97 13.71 13.44 13.17 12.91 12.64 12.37 12.10 11.84 11.57 4 13.61 13.35 13.09 12.83 12.57 12.31 12.05 11.79 11.53 11.27 5 13.26 13.00 12.75 12.50 12.24 11.99 11.73 11.48 11.23 10.97 6 12.93 12.68 12.43 12.18 11.93 11.69 11.44 11.19 10.94 10.70 7 12.60 12.36 12.12 11.88 11.64 11.40 11.15 10.91 10.67 10.43 8 12.30 12.07 11.83 11.59 11.36 11.12 10.89 10.65 10.41 10.18 9 12.01 11.78 11.55 11.32 11.09 10.86 10.63 10.40 10.17 9.94 10 11.72 11.50 11.27 11.04 10.82 10.59 10.37 10.14 9.92 9.69 11 11.45 11.23 11.01 10.79 10.57 10.35 10.13 9.91 9.69 9.47 12 11.19 10.98 10.76 10.54 10.33 10.11 9.90 9.68 9.47 9.25 13 10.94 10.73 10.52 10.31 10.10 9.89 9.68 9.47 9.26 9.04 14 10.70 10.50 10.29 10.08 9.88 9.67 9.46 9.26 9.05 8.85 15 10.47 10.27 10.07 9.87 9.67 9.46 9.26 9.06 8.86 8.65 16 10.25 10.05 9.85 9.65 9.45 9.26 9.06 8.86 8.66 8.46 17 10.04 9.84 9.65 9.46 9.26 9.07 8.87 8.68 8.48 8.29 18 9.83 9.64 9.45 9.26 9.07 8.88 8.69 8.50 8.31 8.12 19 9.63 9.45 9.26 9.07 8.89 8.70 8.51 8.33 8.14 7.95 20 9.44 9.25 9.07 8.89 8.70 8.52 8.34 8.15 7.97 7.79 21 9.26 9.08 8.90 8.72 8.54 8.36 8.18 8.00 7.82 7.64 22 9.07 8.90 8.72 8.54 8.37 8.19 8.01 7.84 7.66 7.48 23 8.91 8.73 8.56 8.39 8.21 8.04 7.86 7.69 7.52 7.34 24 8.74 8.57 8.40 8.23 8.06 7.89 7.72 7.55 7.38 7.20 25 8.58 8.41 8.24 8.07 7.90 7.74 7.57 7.40 7.23 7.06 26 8.42 8.26 8.09 7.92 7.76 7.59 7.43 7.26 7.10 6.93 27 8.28 8.11 7.95 7.79 7.62 7.46 7.30 7.14 6.97 6.81 28 8.13 7.97 7.81 7.65 7.49 7.33 7.17 7.01 6.85 6.69 29 7.99 7.83 7.67 7.51 7.35 7.20 7.04 6.88 6.72 6.57 30 7.85 7.70 7.54 7.38 7.23 7.07 6.92 6.76 6.61 6.45 31 7.72 7.56 7.41 7.26 7.10 6.95 6.80 6.64 6.49 6.34 32 7.58 7.43 7.28 7.13 6.98 6.83 6.68 6.53 6.38 6.22 33 7.46 7.31 7.16 7.01 6.86 6.71 6.57 6.42 6.27 6.12 34 7.34 7.20 7.05 6.90 6.76 6.61 6.46 6.32 6.17 6.02 35 7.22 7.07 6.93 6.79 6.64 6.50 6.35 6.21 6.06 5.92 36 7.11 6.96 6.82 6.68 6.53 6.39 6.25 6.11 5.96 5.82 37 6.99 6.85 6.71 6.57 6.43 6.29 6.15 6.00 5.86 5.72 38 6.89 6.75 6.61 6.47 6.33 6.19 6.05 5.91 5.77 5.63 39 6.79 6.65 6.51 6.37 6.23 6.10 5.96 5.82 5.68 5.54 PPB Dissolved Oxygen Measurement - Calibration and Sampling Techniques 10 40 6.68 6.55 6.41 6.27 6.14 6.00 5.86 5.73 5.59 5.45 41 6.58 6.44 6.31 6.18 6.04 5.91 5.77 5.64 5.50 5.37 42 6.49 6.35 6.22 6.09 5.95 5.82 5.69 5.55 5.42 5.28 43 6.39 6.26 6.13 6.00 5.87 5.73 5.60 5.47 5.34 5.20 44 6.30 6.17 6.04 5.91 5.78 5.65 5.52 5.39 5.25 5.12 45 6.21 6.08 5.95 5.82 5.69 5.56 5.43 5.30 5.17 5.04 46 6.12 5.99 5.86 5.73 5.60 5.47 5.35 5.22 5.09 4.96 47 6.03 5.91 5.78 5.65 5.53 5.40 5.27 5.14 5.02 4.89 48 5.95 5.83 5.70 5.57 5.45 5.32 5.19 5.07 4.94 4.82 49 5.87 5.75 5.62 5.49 5.37 5.24 5.12 4.99 4.87 4.74 50 5.79 5.66 5.54 5.42 5.29 5.17 5.04 4.92 4.79 4.67 Table 3: Oxygen Concentration (ppm) for varying pressures (mmHg) andtemperatures (Degrees Celsius) at 100% Relative Humidity PPB Dissolved Oxygen Measurement - Calibration and Sampling Techniques More Information For more information on using Dissolved Oxygen Measurement, visit www.honeywellprocess.com, or contact your Honeywell account manager. Honeywell Process Solutions Honeywell 1250 West Sam Houston Parkway South Houston, TX 77042 Honeywell House, Arlington Business Park Bracknell, Berkshire, England RG12 1EB UK Shanghai City Centre, 100 Junyi Road Shanghai, China 20051 www.honeywellprocess.com SO-13-19-ENG January 2013 © 2013 Honeywell International Inc. 11
© Copyright 2026 Paperzz