Moisture of Medicine (4) [Lactose] Pharmaceutical Volumetric titration (direct method)by Karl Fischer Moisture Titrator Standard JIS ASTM ISO K 0113 E 203 760 Hydranal manual Japanese pharmacopoeia 1.Abstract Moisture titration using Karl Fischer reagent is popularly practiced water determination worldwide as the most reliable method. The procedure is adopted in many official standards as test method specified in ISO, ASTM, DIN, BS and JIS. The test conducted this time is an example of volumetric moisture titration according to JIS K 0113-2005 for measurement of water content in Lactose, which dissolves with ease in FM extracting medium. 2. Reference 1) JIS K 0113-2005:Standard Test Method by Potentiometric, Amperometrtic, Coulometric and Karl Fischer Titration 2) ASTM E 203-96 Standard Test Method for Water Using Volumetric Karl Fischer Titration 3) ISO 760:1978 Determination of Water-Karl Fischer method (General method) 4) Hydranal manual published by Riedel de Haen 5) Japanese Pharmacopoeia Codex 15 3. Cautions in measurement 1) In order to refrain from the effect of ambient humidity, the test must be conducted in a well air-conditioned room. 2) Handle with care when you work on chemical. 3) Obtain the factor of Karl Fischer reagent using the solvent in advance. 4) Generally FM solvent is used for Saccharide, except for Lactose use FMII. KVX-01504 Ver.01 1/4 4. Post-measurement care After the reagent is drained out and the electrode is cleaned, keep the titration flask filled with extracting solvent. 5.Test equipment Main unit: Karl Fischer moisture titration volumetric system Electrode: Twin platinum electrode for KF titration 6.Reagent Titrant Solvent : : Hydranal Composit 2 and 5 (Riedel de Haen) Extracting medium FMII(for saccharide)(Hayashi Chemicals) 7.Measurement procedure -Pretreatment1) Prepare approximately 30mL extracting medium FMII in titration cell. 2) Dehydrate the measuring cell by performing pre-titration in advance. -Measurement1) Prepare approximately 0.2g sample in a sampler. 2) Weigh the sampler on an electronic balance of which precision is to the nearest 0.1mg. 3) Transfer the sample in sampler to titration cell to dissolve in solvent. 4) Press Start key of titration unit. 5) Weigh the sampler of the above 3). 6) Enter the weight of item 2) for Wt1, and 5) for Wt2. 7) Obtain water content from titration volume, of which EP is detected automatically. 8.Formula Moisture ( % ) = (( Data × F - Blank ) / ( Wt1 - Wt2 ) ) × 0.1 Data F Blank Wt1 Wt2 : : : : : Titration volume ( mL ) Factor of titrant Blank level ( mg ) Sample + Sampler ( g ) Empty sampler ( g ) KVX-01504 Ver.01 2/4 9. Example of measurement -Titration parameterMKA-610 Method No. 1 [Titration] Titr.mode t(stir) t(wait) t(max) t(interval) Max.volume Titr.bur.No. Dose mode Normal 0s 10 s 0s 0s 10 mL 1 Off [Control] End time Final vol. Titr.speed Detect.mode Drift titr. Start mode End level Samp.time Stir.speed 30 s 0.01 mL 3 1 On Manual 75 mV 5s 3 -Calculation parameterMKA-610 [Calculation] Calc.type Blank No. Calc.No. Unit Decimal Fraction Drift comp. Evalution Sample 1 2 % 2 Half adjust Off Off MKA-520 [Titration] Method Titr Mode Titr Buret No. End Time Final Vol. Titr.Speed Detector Mode t(stir) t(wait) t(max) Drift Titr Start Max.Volume Dose mode Oven MKS-500 1 Normal 1 30 s 0.01 mL 3 1 0s 10 s 0s On Manual 10 mL Off Off MKA-520 [Calculation] Calc. Unit Weight [Titration] Method Titr.Speed End Time Final Vol. Detector Mode t(stir) t(max) Drift Titr. Max.Volume Driect 3 30 s 0.01 mL Normal 0s 0s On 10 mL MKS-500 2 % Variable [Calculation] g->% KVX-01504 Ver.01 3/4 -Measurement result- Sample name Water content mg % Sample (g) Extracting medium 0.1539 Solvent FMII 0.90 0.58 0.1840 Solvent FMII 9.24 5.02 Lactose (Anhydride) Lactose (Monohydrate) 10.Summary Lactose is disaccharide of chemical formula C12H22O11 with molecular mass 342.3, contained in mammals whether cow milk or human. The test sample in this application dissolves in solvent FMII, which makes moisture titration performed with ease. Stable measurement of water content is assured by Karl Fischer moisture titration. Water content in Lactose is precisely measured by any of the following models made by Kyoto Electronics (KEM). 【MKA-610】 Awarded Product of Supreme Technology from Kyoto City • Easy key entry by touch panel on 8-inch large color LCD • Two measurements in parallel at a time • Both potentiometric and Karl Fischer moisture titration (coulometric ・ volumetric ) can be performed at a time 【MKA-520】 • Compact 【MKS-500】 and cost performance • Dispenser easy to deliver and discharge extracting medium • One touch burette setup • Low cost and high performance model • User friendly easy to operate • Burette nozzle made of PTFE, standard supplied 2 types (regular and anti-diffusion) KVX-01504 Ver.01 4/4
© Copyright 2026 Paperzz