Applcation Manual ILab 300 Plus IL Test TM Glucose (Oxidase) Kit Configuration P/N 0018480000 6 x 35 mL GLUCOX Reagent Preparation P/N 0018480000 Using a volumetric pipette add 12 mL of deionized or distilled water to one bottle of GLUC OX. This can be done also by adding the water up to the rim. See the drawing N° 2 page 3 Invert two or three times to mix the solution , then place the bottle into ILab 300 plus reagent rack in the refrigerated area (position from 14 to 33). After daily use cap the bottles and store at 2-8 °C. In Use Stability When used as described in Reagent Preparation the reagent is stable for: 15 days Specimen collection and Preparation Serum Calibration Use ReferrILTM G Calibrator (Cat. No. 0018257000). See calibrator insert sheet for lot and instrument specific concentration. Recalibrate the instrument every 15 days or when a new lot of reagent is used. Reagent blanking should be performed daily or before patient assays if the reagent is used less frequently. Quality Control Serachem® Level 1 (Cat. No. 0018162412) and SeraChem® Level 2 (Cat. No 0018162512). Calculation and Analytical Results The result concentration is automatically calculated by the instrument against the Calibrator. For detailed description, refer to the Instrument settings and to the ILab 300 Plus Operator’s Manual. To convert mg/dL to SI units (mmol/L) multiply by 0.055. Reference Interval Serum: 70 - 105 mg/dL (3.9 - 5.8 mmol/L). It is recommended that each laboratory establish its own reference range since ranges may vary with age, diet, gender, and geographical area. References / Literatur / Bibliografía / Bibliographie / Bibliografia / See package insert inclosed in the kit Performance Characteristics All performance data were obtained on human serum. Limitation/Interfering Substances No interference from lipemia up to sample absorbance of 6.8/cm at 660 nm (1100 mg/dL or 12.4 mmo/L triglycerides). No interference up to 500 mg/dL (0.300 mmol/L) hemoglobin. No interference up to 8 mg/dL (137 µmol/L) bilirubin. 1 For a comprehensive review of interfering substances, refer to the publication by Young. Precision Samples/Runs Mean (mg/dL) CV(%) Mean (mg/dL) CV (%) Within Run 5/10 79 2.2 273 0.9 Total 5/10 79 2.5 273 2.2 Method Comparison Comparison Method (x) Same reagent Comparison Instrument ILab 600 Slope 0.936 y intercept 2.48 Range of x (mg/dL) 52-422 Mean x (mg/dL) 119 Mean y (mg/dL) 114 r 0.999 SE 2.64 N 80 Linearity 0 - 450 mg/dL ( 0.0 - 24.8 mmol/L) Sensitivity milliabsorbance change per 1 mg/dL: 2.3 milliabsorbance change per 1 mmmol/L: 41.8 Minimun Detection Limit 0 mg/dL (0.0 mmol/L). Instrumentation Laboratory SpA – V.le Monza 338 – 20128 Milano (Italy) Instrumentation Laboratory Company - Lexington, MA 02421-3125 (USA) 38 of 75 Applcation Manual ILab 300 Plus IL Test TM Glucose (Oxidase) Instrument Settings Reference Range Description: Unit: Decimals: LIS Code: Unit Factor: Slope: Intercept: Reaction Type: Direction: E.P. Limit (abs): Depl Limit (abs): First Limit (abs): Linear Factor: Fit: Glucose OX mg/dL (mmol/L) 0 (1) ** 1.0 1.00 0.00 End Point Up 0.2500 N/A N/A N/A N/A Lin Limit. Low: 0 (0.0) High: 450 (24.8) Rerun when Over: Yes Male: Female: Children: Rerun: LOW VALUES 0 (0.0) 0 (0.0) 0 (0.0) 0 (0.0) 0 (0.0) 0 (0.0) 0 (0.0) 0 (0.0) 0 (0.0) Low Alert Very Low Low No No Predilut.-> S.+R. 1-> Times (sec): 0 Dil./Rgt. Code: GLU Lot Number: Ratio/Vol. (ul): 1/1 300 Rinse (ul): 0 Sample (ul): 4 RBL Min (abs): 0.0000 Max (abs): 0.6500 HIGH VALUES 70 (3.9) 105 (5.8) 450 (24.8) 450 (24.8) 450 (24.8) 70 (3.9) 105 (5.8) 450 (24.8) 450 (24.8) 450 (24.8) 70 (3.9) 105 (5.8) 450 (24.8) 450 (24.8) 450 (24.8) Normal Values High Very High High Alert No No Parameter Reag. 2 -> Reag. 3 -> 0 0 0 0 0 0 Calculation Model: Standard Factor: N/A Sample Blank: Yes ** N/A Instrumentation Laboratory SpA – V.le Monza 338 – 20128 Milano (Italy) Instrumentation Laboratory Company - Lexington, MA 02421-3125 (USA) Incubation -> Read 80 0 *) = kinetic Filter 1 (nm): 510 Filter 2 (nm): (none) Bicr. Factor: 1.00 RBL Stability (days): 1 Calibration Stab. (days): 15 Dinamic Controls (min): None Operator definable Not applicable to this test 39 of 75
© Copyright 2025 Paperzz