TO R N I E R BLUEPRINT ™ 3D Planning + PSI SC AN PROTOCOL V1.5 Contents 2 3 Introduction 3 Patient preparation 3 Scanning instructions 4 Image instructions 5 Scanning parameters 6 Technical instructions BLUEPRINT™ 3D Planning Software Scan Protocol v1.5 Scan Protocol Introduction This document describes the guidelines for a Shoulder CT-Scan to be processed by BLUEPRINT™ 3D Planning Software. Following this scanning protocol will optimize the likelihood of successful scan reconstruction in BLUEPRINT™ 3D Planning Software. Note: Please contact the helpline if you require further information. Figure 1: Artifacts generated by metallic implants Patient Preparation » Verify that metal is not present within the scanned portion of the operative side (Figure 1) Note: Pacemakers, dental and cervical hardware can cause artifact if too close to the GlenoHumeral joint. If artifact too intense, do not scan. » Do NOT use contrast material » Place the patient supine on the scanner table, Humerus along the trunk (Figure 2) Figure 2: Position of the patient in the machine Note: Position must be craniocaudal oriented (head first) and iso-centered, with the hand in a neutral position. » It is strongly recommended to place a sponge or a small pillow between humerus and trunk Scanning instructions » No gantry tilt » Limit the field of view to the desired/operative shoulder Figure 3: Example of incomplete projections » Avoid incomplete projections of soft-tissue wrap by having all shoulder softtissue within the field of view (causes pseudo-metal artifact) (Figure 3) » It is strongly recommended to perform a full scan of the scapula (Figure 4) Note: Scan must include several slices about the AC joint and several slices (around 1 cm) below the inferior angle. » All slices must have the same field of view and the same slice spacing » DO NOT use auto-mA setting Figure 4: Full scan of the scapula BLUEPRINT™ 3D Planning Software Scan Protocol v1.5 3 Images instructions » Use only axial images, no other reconstructions or reformations in different planes (sagittal or coronal) » The DICOM file format is defined in the Transfer Syntax (0002, 0010) Tag field, the supported formats are the following: Uncompressed Transfer Syntax UID Description 1.2.840.10008.1.2 Implicit VR – Little Endian 1.2.840.10008.1.2.1 Explicit VR – Little Endian 1.2.840.10008.1.2.2 Explicit VR – Big Endian Lossless compressed Transfer Syntax UID Description 1.2.840.10008.1.2.5 RLE Lossless 1.2.840.10008.1.2.4.57 JPEG Lossless, Non-Hierarchical (Process 14) 1.2.840.10008.1.2.4.70 JPEG Lossless, Hierarchical, First-Order Prediction 1.2.840.10008.1.2.4.90 JPEG 2000 Image Compression (Lossless Only) Lossy compressed 4 Transfer Syntax UID Description 1.2.840.10008.1.2.4.91 JPEG 2000 Image Compression BLUEPRINT™ 3D Planning Software Scan Protocol v1.5 Scanning parameters » Recommended parameters Parameter Recommended Modality CT Kernel / Algorithm Bone or Bone+ kVp 120 or 140 kVp mA Manual 240 mA DO NOT use auto-mA Slice Increment 1.2 mm Image Thickness - Detector Coverage should be maximum - Helical Thickness 0.625 mm or 1.25 mm - Pitch 0.9 or less - Rotation time 1 sec or less Exposure time 1000 ms Note: Using the manual-mA setting of 240 mA will provide compatible images. Not doing so will produce images that are incompatible with BLUEPRINT™ 3D Planning Software. » Display Field Of View (DFOV) Parameter Recommended DFOV (axial plane) (Figure 5) Please include the entire scapula 25 cm or no more 32 cm DFOV (frontal plane) (Figure 6) Please include the entire scapula More than 10 cm Matrix size 512 x 512 Figure 5: DFOV (axial plane) BLUEPRINT™ 3D Planning Software Figure 6: DFOV (frontal plane) Scan Protocol v1.5 5 Technical Instructions » GE HEALTHCARE In order to define the recommended mA, please follow the next steps: 1. Click on mA button 2. Set manual-mA to 240 mA and dose reduction 0% 3. Set the desired slice thickness for helical CT by clicking on Slice Thickness: - Set the detector coverage to maximum depending on each machine - Set the helical thickness to 0.625 mm or no more than 1.25 mm - Set the pitch to 0.9 or less - Set the rotation time to 1.0 sec » SIEMENS This example works for Somatom Sensation 64 but the user interface is the same in most of Siemens systems. You have two main tabs to configure the scan protocol on Siemens machines: 1. On the Reconstruction tab: - Set the Kernel/Algorithm to Bone 70s, sharp or very sharp - Set the window on Bone - Set the reconstruction increment to 0.6 mm or less 2. On the Scan tab: - Set the quality ref mAs to 300 mAs - Set the kVp to 120 or 140 kVp - Set the rotation time to 1.0 sec - Set the slice thickness to 1.25 mm or less - Set the pitch to 0.9 or less » OTHER SCANNER BRANDS For any other brand of CT scan machine please refer to the standard recommendations in pages 4, 5 and 6. 6 BLUEPRINT™ 3D Planning Software Scan Protocol v1.5 Notes BLUEPRINT™ 3D Planning Software Scan Protocol v1.5 7 BLUEPRINT™ 3D PLANNING SOFTWARE HELPLINE Email: [email protected] Phone: 855.378.1459 1023 Cherry Road Memphis, TN 38117 800 238 7117 901 867 9971 www.wright.com 10801 Nesbitt Avenue South Bloomington, MN 55437 888 867 6437 952 426 7600 www.tornier.com Proper surgical procedures and techniques are the responsibility of the medical professional. This material is furnished for information purposes only. Each surgeon must evaluate the appropriateness of the material based on his or her personal medical training and experience. Prior to use of any Tornier implant system, the surgeon should refer to the product package insert for complete warnings, precautions, indications, contraindications, and adverse effects. Package inserts are also available by contacting Wright. Contact information can be found in this document and the package insert. and ® denote Trademarks and Registered Trademarks of Wright Medical Group N.V. or its affiliates. ©2016 Wright Medical Group N.V. or its affiliates. All Rights Reserved. ™ CAW-7230 Rev B ECN 161103 05-Aug-2016
© Copyright 2026 Paperzz