Hematopathology / EVALUATION OF THE PENTRA 80 HEMATOLOGY ANALYZER Analytic Performance of the PENTRA 80 Automated Blood Cell Analyzer for the Evaluation of Normal and Pathologic WBCs María E. Arroyo, PhD,1 María D. Tabernero, MD, PhD,2 María A. García-Marcos, MD, PhD,3 and Alberto Orfao, MD, PhD1 Key Words: PENTRA 80; Hematology analyzer; Flow cytometry; Five-part leukocyte differential; Atypical lymphocytes; Large immature cells; Immunophenotyping DOI: 10.1309/6U2T6UTWK10M3NCB Abstract We evaluated performance of the ABX PENTRA 80 (ABX Diagnostics, Montpellier, France) hematology analyzer in enumerating the most frequent subsets of WBCs in peripheral blood, atypical lymphocytes (ALYs), and large immature cells (LICs) by comparing results with those obtained by manual microscopic counts, another hematology analyzer, and flow cytometric immunophenotyping. Identification and enumeration of neutrophils and lymphocytes with the PENTRA 80 showed high correlation with all 3 reference methods (R2 ≥ 0.92 and R2 ≥ 0.88, respectively); quantification of eosinophils showed good correlation with the other analyzer and flow cytometric immunophenotyping (R2 ≥ 0.70); lower correlation coefficients were found for comparisons with conventional microscopy (R2 ≥ 0.50). For monocytes, lower but acceptable correlation and agreement were found; marginal correlation was found for basophils. The PENTRA 80 also showed good performance in detecting LICs but was less effective for the identification of ALYs in relatively low frequencies in abnormal peripheral blood samples. We found good performance of the 5-part leukocyte differential analyses for the PENTRA 80, especially for enumeration of neutrophils, lymphocytes, eosinophils, and LICs. 206 206 Am J Clin Pathol 2005;123:206-214 DOI: 10.1309/6U2T6UTWK10M3NCB The PENTRA 80 automated cell counter is a new 80sample-per-hour hematology analyzer manufactured by ABX Diagnostics (Montpellier, France) that, according to the specifications of the manufacturer, provides CBC counts, including a 5-part WBC differential count; at the same time, it provides enumeration of large immature cells (LICs) and atypical lymphocytes (ALYs). Measurements are based on principles of impedance, cytochemistry, and the measurement of light absorbance through the use of the so-called double-hydrodynamic sequential system (or DHSS) technology1 associated with staining of samples with a cytochemical reagent (Eosinophix, ABX Diagnostics) before analysis of the cell volume and light absorbance. Owing to its recent availability, no studies have been reported so far in which the PENTRA 80 automated cell counter is evaluated in detail in a clinical diagnostic hematology laboratory for its performance in calculating the 5part WBC differential and the identification of the related quantitative and/or morphologic abnormalities to be flagged. Few studies have been reported on the PENTRA 120 (ABX Diagnostics).1-3 According to the National Committee for Clinical and Laboratory Standards (NCCLS),4 evaluation of the total and differential counts of WBCs and WBC-related flags provided by a new hematology analyzer should be based on conventional microscopic analysis. Although conventional microscopy is extremely useful in clarifying complex hematologic samples and verifying the nature of malignant vs reactive cells, it has some disadvantages and limitations once applied to the control of WBC counts: (1) Few cells are available for counting. (2) Experience is required to enumerate different cell types. (3) It is time-consuming. In addition, some cells that normally © American Society for Clinical Pathology Hematopathology / ORIGINAL ARTICLE circulate in the blood, such as dendritic cells, cannot be identified specifically by morphologic examination, and distinct WBC populations can show a different distribution throughout the slide, leading to variations in the differential counts.5 Flow cytometry is an innovative, rapid, sensitive, and less operator-dependent technology6,7 that allows precise multiparameter enumeration of high numbers of cells from different cell populations in a sample in a relatively short period.8 In recent years, it has been suggested that combined flow cytometric analysis of the light scatter and phenotypic characteristics of WBCs might represent a reliable approach to accurately enumerating WBCs and their major subsets in normal and pathologic samples.9 The aim of the present study was to evaluate the performance of the PENTRA 80 blood cell analyzer in the 5-part differential analysis of WBCs and its related quantitative and/or morphologic flags. The results provided by this instrument were compared with those obtained using conventional morphologic examination, multiparameter flow cytometric immunophenotyping, and another independent hematologic cell analyzer. Materials and Methods Peripheral Blood Samples In the present study, a total of 200 peripheral blood (PB) samples were analyzed. Half of them were normal PB samples defined by the absence of quantitative and qualitative morphologic and numeric flags on RBCs, WBCs, and platelets; the other 100 samples were classified as abnormal based on the presence of quantitative and/or qualitative morphologic and/or numeric flags on a routine WBC differential analysis (CELLDYN 4000, Abbott Diagnostics, Santa Clara, CA). The abnormal samples were from patients with nonhematologic disorders (n = 51) or diagnosed with hematologic disease (n = 49); the latter group included myelodysplastic syndromes (n = 4), acute myeloid leukemia (n = 3), chronic myeloid leukemia (n = 2), Bcell chronic lymphocytic leukemia (n = 13), B-cell nonHodgkin lymphoma (n = 20), and multiple myeloma (n = 7). All samples were obtained by venipuncture and placed directly in VACUTAINER tubes containing tripotassium EDTA as the anticoagulant (Becton Dickinson, Franklin Lakes, NJ). In all cases, PB samples were analyzed within 6 hours after being obtained; whenever the sample was not analyzed immediately, it was kept at stable room temperature. Analysis of samples was performed in parallel in the PENTRA 80 and the 3 reference methods: (1) the CELL-DYN 4000 hematology analyzer (n = 200); (2) morphologic analysis of blood smears by optical microscopy (n = 200); and (3) multiparameter flow cytometric immunophenotyping (n = 50; 25 normal and 25 abnormal randomly selected samples). WBC Analysis With the PENTRA 80 All PB samples were analyzed in duplicate in the PENTRA 80 in the “differential and automatic” mode. The mean value derived from duplicate measurements was calculated and used in the evaluation. Information on the following WBC-associated parameters was recorded specifically: absolute WBC count and absolute and relative neutrophil, lymphocyte, monocyte, eosinophil, and basophil counts. Whenever present, information about abnormal WBC populations (eg, ALYs, which include large lymphocytes, activated lymphocytes, lymphoid cells, and small lymphocytes, and/or LICs, which include myelocytes, promyelocytes, metamyelocytes, and large blasts) also was recorded. The mean coefficients of variation obtained between sample replicates were 1.92% and 2.35% for neutrophils, 2.63% and 2.99% for lymphocytes, 4.28% and 4.47% for monocytes, 9.84% and 10.49% for eosinophils, and 9.75% and 9.83% for basophils in relative and absolute numbers, respectively. Morphologic Evaluation of PB WBC Differential Counts For morphologic evaluation of the WBC differential count, 4 smears were prepared and dried for each blood sample. Afterward, 3 of the 4 blood slides were stained with MayGrünwald-Giemsa according to well-established methods. Of the 3 stained slides, 2 were assigned randomly for the evaluation of WBC differential counts. Blood smear examination was performed by 2 independent expert physicians who randomly counted 200 WBCs each, following the NCCLS criteria.4 For each cell population identified in each sample, the average value obtained by the 2 experts was used for comparison with the results obtained with the PENTRA 80. The mean coefficients of variation (and correlation coefficients) obtained for the measurements performed by the experts were 2.91% (R2 = 0.84) for neutrophils, 3.54% (R2 = 0.92) for lymphocytes, 1.86% (R2 = 0.97) for monocytes, 4.7% (R2 = 0.99) for eosinophils, and 14.1% (R2 = 0.93) for basophils. Independent Automated Blood Cell Analysis The CELL-DYN 4000 was used as an independent blood cell counter. This instrument uses multiangle polarized side-scatter technology to calculate the distribution of the populations of WBCs in PB samples.10 All 200 samples studied were processed in the “differential and automatic” mode in the CELL-DYN 4000 by an independent operator as done with the PENTRA 80. Multiparameter Flow Cytometric Immunophenotypic Analysis of WBC Differential Counts Fifty samples were randomly selected from the 100 normal (n = 25) and 100 abnormal (n = 25) sample groups and analyzed by multiparameter flow cytometry. The Extended Hematology Panel11 was used with permission for the staining Am J Clin Pathol 2005;123:206-214 © American Society for Clinical Pathology 207 DOI: 10.1309/6U2T6UTWK10M3NCB 207 207 Arroyo et al / EVALUATION OF THE PENTRA 80 HEMATOLOGY ANALYZER of these samples. Briefly, an exact volume of 100 µL of premixed blood was placed by reverse pipetting using a calibrated pipette in a 12 × 75-mm polystyrene tube containing TRUECOUNT beads (Becton Dickinson Biosciences [BDB], San Jose, CA), to which the Extended Hematology Panel mixture of monoclonal antibodies directly coupled with fluorescent dyes was added.11 The mixture then was incubated for 15 minutes at room temperature in the dark. Once this incubation period was finished, 2 mL of FACS Lysing solution (BDB) diluted 1:10 (vol/vol) in distilled water was added to lyse the nonnucleated RBCs, and another 10-minute incubation at room temperature was performed in the dark. All samples were measured in a FACSCalibur flow cytometer (BDB), and information on a total of 50 × 103 cells plus the corresponding TRUECOUNT beads was acquired and stored using CellQUEST software (BDB). PAINT-A-GATE software (BDB) was used for data analysis, according to the Extended Hematology Panel protocol. TRUECOUNT beads were used to calculate the absolute count per microliter for each cell population identified as previously described.11 Briefly, for the identification of each subset of WBCs the following phenotypic characteristics were used: neutrophils, forward scatter (FSC)int/side scatter (SSC)high/HLADR–/CD33lo/CD14–/lo/CD45+; monocytes, FSCint/SSCint/ HLADR+/CD33hi/CD14hi/CD45int to hi; lymphocytes,FSClo/SSClo/ HLA-DR– to +/CD33–/CD14–/CD45hi; eosinophils, FSCint/ SSCvery high/HLA-DR–/CD33+/CD14–/CD45+; basophils, FSClo/SSClo/HLA-DR–/CD33+/CD14–/CD45+; and dendritic cells, FSCint/SSClo to int/HLA-DRhi/CD33– or +/CD14– or +/CD45int to hi. In addition, in pathologic samples, blast cells were identified as being SSClo/CD45lo events and immature granulocytes as events showing phenotypic characteristics similar to the neutrophils but with higher FSC and/or CD33 expression. ALYs were identified as events similar to lymphocytes with abnormally high FSC and/or HLA-DR expression. Statistical Methods All numeric and coded data were introduced in a database using SPSS 10.0 statistical software (SPSS, Chicago, IL). The Pearson correlation was used for the initial comparison between methods. The Bland-Altman test was used to assess the degree of agreement between methods.12 To establish the best cutoff values for maximum sensitivity, specificity, and efficiency for the PENTRA 80 to identify the presence of a given population of abnormal WBCs, receiver operating characteristic curves were used, according to NCCLS recommendations.4 Results ❚Table 1❚ summarizes the results of correlating the relative and absolute counts for each WBC population analyzed with the PENTRA 80 with those of the 3 reference methods. High correlation was observed for WBC counts between the 208 208 Am J Clin Pathol 2005;123:206-214 DOI: 10.1309/6U2T6UTWK10M3NCB PENTRA 80 and the CELL-DYN 4000 and between the PENTRA 80 and flow cytometric immunophenotyping in normal and abnormal samples (R2 ≥ 0.85 and R2 ≥ 0.93, respectively). As shown in Table 1, a high degree of correlation was observed between results from the PENTRA 80 and those in normal and abnormal samples with the other 3 methods for neutrophils (R2 ≥ 0.92 and R2 ≥ 0.84, respectively) and lymphocytes (R2 ≥ 0.88 and R2 ≥ 0.86, respectively). High correlation also was observed for the relative and absolute number of eosinophils obtained with the PENTRA 80 and with the CELL-DYN 4000 and flow cytometric immunophenotyping in normal (R2 ≥ 0.92 and R2 ≥ 0.95, respectively) and, to a lesser extent, in abnormal samples (R2 ≥ 0.72 and R2 ≥ 0.66, respectively). The correlation between the percentage of eosinophils obtained with the PENTRA 80 and manual microscopic counts was lower in normal (R2 = 0.57) and abnormal (R2 = 0.50) samples. A clearly lower degree of correlation was obtained for the percentage and absolute monocyte counts obtained with the PENTRA 80 and the CELL-DYN 4000 (R2 ≥ 0.42 and R2 ≥ 0.12, respectively), manual microscopic counts (R2 ≥ 0.30), and flow cytometric immunophenotyping (R2 ≥ 0.46 and R2 ≥ 0.15, respectively). Regarding the number of basophils, no statistically significant correlation was found between results from the PENTRA 80 and the other 3 methods, except for a marginal correlation found between the PENTRA 80 cell counts and flow cytometric immunophenotyping in normal PB samples for the relative and absolute basophil counts. On exploring the degree of agreement between results obtained with the PENTRA 80 and the other methods, it should be noted that, in general, higher concordance was observed for normal than for abnormal samples in percentage and absolute cell counts. Accordingly, a high degree of agreement was observed for the WBC count with the PENTRA 80 compared with the CELL-DYN 4000 (n = 200) and flow cytometric immunophenotyping (n = 50) (83% and 94%, respectively). The differential neutrophil counts obtained with the PENTRA 80 were in total agreement with those measured by the CELL-DYN 4000 and optical microscopy for normal PB samples; the values decreased substantially for abnormal samples ❚Table 2❚. For lymphocyte counts, a high degree of agreement also was observed in relative numbers for the PENTRA 80 and the CELL-DYN 4000 and microscopic counts (≥98%) in normal samples; however, the degree of agreement decreased notably in absolute values for the CELL-DYN 4000 in normal and abnormal samples. As could be expected, the degree of agreement between results with the PENTRA 80 and the CELL-DYN 4000 and manual counts was lower for monocytes and eosinophils in normal and, especially, abnormal samples. The degree of agreement observed for basophil counts in normal and abnormal samples was consistently 22% or lower when results obtained © American Society for Clinical Pathology Hematopathology / ORIGINAL ARTICLE ❚Table 1❚ Degree of Correlation Between Percentage and Absolute Values for the Five-Part WBC Differential With Two Analyzers, Optical Microscopy, and Flow Cytometric Immunophenotyping* Samples Neutrophils Methods Normal Monocytes Eosinophils Basophils Abnormal Normal Abnormal Normal Abnormal Normal Abnormal Normal 0.84 1.05 1.77 <.001 0.94 0.90 4.73 <.001 0.88 0.81 8.38 <.001 0.49 0.54 2.39 <.001 0.42 0.77 6.00 <.001 0.94 0.93 0.43 <.001 0.73 0.64 0.71 <.001 0.00 –0.01 0.65 .78 0.02 0.70 2.84 .18 0.92 0.93 0.36 <.001 0.88 0.90 4.06 <.001 0.86 –7.18 1.09 <.001 0.32 0.41 3.63 <.001 0.30 0.88 12.40 <.001 0.57 0.78 1.02 <.001 0.50 0.50 0.91 <.001 0.00 0.02 0.64 .58 0.05 0.72 2.74 .02 0.92 0.94 3.38 <.001 0.96 0.93 4.80 <.001 0.93 0.86 6.31 <.001 0.46 0.48 2.40 <.001 0.58 0.81 3.87 <.001 0.95 0.99 0.58 <.001 0.86 0.81 0.80 <.001 0.32 0.29 0.40 .003 0.00 0.04 2.10 .94 0.99 1.14 –0.791 <.001 0.85 0.91 0.42 <.001 0.97 0.99 1.17 <.001 0.49 0.54 2.39 <.001 0.12 0.77 1.37 <.003 0.92 1.03 0.03 <.001 0.72 1.07 0.08 <.001 0.98 1.05 –1.36 <.001 0.91 0.88 0.35 <.001 0.97 0.91 1.05 <.001 0.65 0.65 0.07 <.001 0.15 0.57 1.03 <.07 0.97 0.96 0.02 <.001 0.66 0.78 0.07 <.002 Percentage CELL-DYN 4000 R2 0.92 Slope 0.93 y-Intercept 2.73 P <.001 Optical microscopy R2 0.92 Slope 0.93 y-Intercept 2.61 P <.001 Flow cytometry R2 0.97 Slope 0.97 y-Intercept 0.25 P <.001 Cells/µL CELL-DYN 4000 R2 0.85 Slope 1.01 y-Intercept 0.31 P <.001 Flow cytometry R2 0.93 Slope 0.98 y-Intercept –0.19 P <.001 * Lymphocytes 0.01 –0.06 0.04 .43 0.26 0.33 0.03 .012 Abnormal 0.02 2.32 0.77 .87 0.01 –0.57 0.57 .87 Methods vs the ABX PENTRA 80 blood cell analyzer. Numbers of samples for the comparisons were as follows: CELL-DYN 4000 hematology analyzer and microscopic-based manual counts, 100 normal and 100 abnormal samples; flow cytometric immunophenotyping, 25 normal and 25 abnormal samples. For proprietary information, see the text. with the PENTRA 80 were compared with those from the CELL-DYN 4000 and microscopy. The overall degree of agreement between results from the PENTRA 80 and flow cytometric immunophenotyping is shown in ❚Figure 1❚ for the 50 samples analyzed. A high degree of agreement between the methods was observed for the relative and absolute counts of neutrophils and lymphocytes. For monocytes, the degree of agreement was acceptable in relative and absolute numbers. By contrast, for eosinophils and basophils, the degree of agreement observed between the PENTRA 80 and flow cytometry was lower (≥24%). For the LIC flag obtained with the PENTRA 80, a relatively acceptable degree of agreement was observed with morphologic counts ❚Table 3❚. Once the “blast,” “band,” and “immature ❚Table 2❚ Percentage of Agreement Between Results Obtained With Two Analyzers and Optical Microscopy* Normal Samples CELL-DYN 4000 Cell Population WBCs Neutrophils Lymphocytes Monocytes Eosinophils Basophils % of Cells — 100 99 54 38 8 Cells/µL 96 95 68 80 39 4 Abnormal Samples Microscopy % of Cells — 100 98 52 29 3 CELL-DYN 4000 Microscopy % of Cells Cells/µL % of Cells — 66 39 53 28 0 79 71 57 46 29 22 — 72 28 44 17 4 * Agreement is between the results with the ABX PENTRA 80 hematology cell analyzer and those with the CELL-DYN 4000 and optical microscopy for enumeration of WBCs, neutrophils, lymphocytes, monocytes, eosinophils, and basophils in 100 normal and 100 abnormal peripheral blood samples. Cutoff values for disagreement were established as the difference between the values obtained with the 2 methods higher or lower than 10% of their mean values. Results are expressed as percentage values. For proprietary information, see the text. Am J Clin Pathol 2005;123:206-214 © American Society for Clinical Pathology 209 DOI: 10.1309/6U2T6UTWK10M3NCB 209 209 Arroyo et al / EVALUATION OF THE PENTRA 80 HEMATOLOGY ANALYZER B –10 –20 0 20 40 60 80 Mean % of Neutrophils 100 20 15 10 5 0 –5 –10 0 58% agreement 10 20 Mean % of Monocytes 30 –2 –4 –6 0 10 20 30 40 Mean No. of Neutrophils/µL 7 6 28% agreement 5 4 3 2 1 0 –1 –2 –3 0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 Mean % of Basophils 20 74% agreement 10 0 –10 –20 0 20 40 60 80 100 Mean % of Lymphocytes G 6 5 4 3 2 1 0 –1 –2 –3 –4 –5 0 76% agreement 1 2 3 4 Mean No. of Monocytes/µL J % Difference of Basophils I % Difference of Basophils 0 F % Difference of Monocytes % Difference of Monocytes E 2 30 % Difference of Lymphocytes 0 86% agreement D 4 28% agreement 3 2 1 0 –1 –2 –3 –0.5 0.0 0.5 1.0 1.5 2.0 Mean No. of Basophils/µL 2.0 1.5 1.0 0.5 0.0 –0.5 –1.0 0 8 6 4 2 74% agreement 0 –2 –4 –10 0 10 20 30 40 50 Mean No. of Lymphocytes/µL H 0.6 24% agreement 0.5 0.4 0.3 0.2 0.1 0.0 –0.1 –0.2 0.0 0.1 0.2 0.3 0.4 0.5 0.6 Mean No. of Eosinophils/µL % Difference of Eosinophils 10 4 % Difference of Lymphocytes 94% agreement C % Difference of Eosinophils 20 % Difference of Neutrophils % Difference of Neutrophils A 24% agreement 2 4 6 Mean % of Eosinophils 8 ❚Figure 1❚ Degree of agreement for the relative and absolute numbers of different subpopulations of peripheral blood WBCs between the ABX PENTRA 80 hematology cell counter and flow cytometric immunophenotyping in 50 peripheral blood samples analyzed. Cutoff values for disagreement were established as the difference between values obtained with the 2 methods higher or lower than 10% of their mean value. Dotted line, zero value; solid line, mean difference; dashed line, 10% cutoff; continuous external lines, ±SD. For proprietary information, see the text. ❚Table 3❚ Degree of Correlation and Agreement Between Relative and Absolute Counts of LICs and ALYs With Two Analyzers, Microscopy, and Flow Cytometric Immunophenotyping* % of Cells Methods CELL-DYN 4000 R2 Slope y-Intercept P Percentage of agreement Optical microscopy R2 Slope y-Intercept P Percentage of agreement Flow cytometry R2 Slope y-Intercept P Percentage of agreement LICs Cells/µL ALYs LICs ALYs 0.46 0.35 1.67 .11 43 0.36 0.43 0.09 <.001 2 0.64 0.43 0.09 .01 35 0.22 4.75 0.54 <.001 11 0.64 0.62 2.62 .01 36 0.12 0.42 3.27 <.001 6 — — — — — — — — — — 0.49 0.45 1.47 <.001 62 0.60 0.31 1.55 <.001 8 0.25 0.49 0.34 .02 58 0.95 0.47 0.16 <.001 40 ALYs, atypical lymphocytes; LICs, large immature cells. Methods vs the ABX PENTRA 80 blood cell analyzer. Numbers of samples for comparisons were as follows: CELL-DYN 4000 hematology analyzer, 100 abnormal samples; microscopic-based manual counts, 100 abnormal samples; flow cytometric immunophenotyping, 25 abnormal samples. For proprietary information, see the text. * 210 210 Am J Clin Pathol 2005;123:206-214 DOI: 10.1309/6U2T6UTWK10M3NCB © American Society for Clinical Pathology Hematopathology / ORIGINAL ARTICLE granulocyte” flags provided by the CELL-DYN 4000 were summed up and compared with the LIC values from the PENTRA 80, an acceptable degree of agreement was observed. In contrast, the presence of ALYs as determined by the PENTRA 80 showed poor agreement when compared with morphologic examination and the CELL-DYN 4000 “Varlym” flag (≤11%). Despite this, it should be noted that when compared with flow cytometry, the LIC and ALY counts obtained with the PENTRA 80 showed statistically significant correlations in relative and absolute numbers, even though the percentage of agreement between methods was acceptable for the LIC flag but much lower for the ALY flag. Because the usefulness of the LIC and ALY flags provided by a hematology cell analyzer depends on qualitative rather than quantitative data, a final goal of our study was to establish the best cutoff values for optimal sensitivity, specificity, and efficiency for both flags. As shown in ❚Figure 2❚, the cutoff values for percentages of LIC and ALY flags showing higher efficiency were 1.0%, 1.5%, and 3.0% for LICs and 20.5%, 12.2%, and 5.0% for ALYs for comparisons done with the CELL-DYN 4000, microscopy, and flow cytometry, respectively. At these cutoff values, the sensitivity and specificity obtained for the LIC and ALY flags given by the PENTRA 80 were 100.0% and 89.3%, 100.0% and 85.7%, and 85.7% and 91.4% for the LIC flag and 16.7% and 100.0%, 33.3% and 98.9%, and 66.7% and 97.9% for the ALY flag compared with the CELL-DYN 4000, conventional microscopy, and flow cytometry, respectively. For the absolute counts of LICs and ALYs, highest efficiency was obtained at cutoff values of 0.14/µL and 6.8/µL B 50 Cutoff = 1.0 Sensitivity = 100.0% Specificity = 89.3% 25 0 100 75 50 25 % Specificity 75 50 25 % Specificity 0 100 50 % Sensitivity 75 Cutoff = 3.0 Sensitivity = 85.7% Specificity = 91.4% 25 0 100 75 50 25 % Specificity 0 75 50 Cutoff = 5.0 Sensitivity = 66.7% Specificity = 97.9 % 25 0 100 100 75 50 25 % Specificity 75 50 Cutoff = 1.5 Sensitivity = 100.0% Specificity = 85.7% 25 0 100 75 50 25 % Specificity 0 75 50 Cutoff = 12.2 Sensitivity = 33.3% Specificity = 98.9% 25 0 100 75 50 25 % Specificity 0 ❚Figure 2❚ Receiver operating characteristic curve plots to evaluate the sensitivity and specificity of the relative values of large immature cell (A, C, and E) and atypical lymphocyte (B, D, and F) flags provided by the ABX PENTRA 80 hematology cell counter compared with the CELL-DYN 4000 automated hematology cell analyzer, manual microscopic counts, and multiparameter flow cytometry. For proprietary information, see the text. F 100 % Sensitivity Cutoff = 20.5 Sensitivity = 16.7% Specificity = 100.0% 25 0 100 0 E 75 50 D 100 % Sensitivity 75 Until now, the performance of a new hematology cell analyzer for determining the WBC 5-part differential counts has been evaluated against manual optical microscopy counts as recommended in the NCCLS H20-A document4,13,14; in many studies, in addition to conventional cytology, side-byside evaluation with another already validated blood cell counter is performed in parallel.10,15,16 More recently, the development of single-platform flow cytometric immunophenotyping assays capable of specifically and simultaneously identifying and enumerating different subpopulations of WBCs has provided a new, accurate, external, independent tool for evaluation of the performance of 5-part differential analysis of a cell counter, which overcomes many limitations of microscopy.7,17,18 The PENTRA 80 automated blood cell counter performs a CBC count and a 5-part WBC differential analysis based on principles of impedance, cytochemistry, and the measurement of light absorbance (DHSS technology). Since its development, no external studies have been reported in which this hematologic blood cell analyzer has been specifically evaluated. Such studies have shown good performance of the PENTRA 80 in comparison with other hematology cell analyzers for the measurement of RBC- and platelet-associated parameters, as also C 100 % Sensitivity % Sensitivity 100 Discussion % Sensitivity A and 0.11/µL and 0.88/µL when compared with the CELLDYN 4000 and flow cytometric immunophenotyping, respectively ❚Figure 3❚. 0 Am J Clin Pathol 2005;123:206-214 © American Society for Clinical Pathology 211 DOI: 10.1309/6U2T6UTWK10M3NCB 211 211 Arroyo et al / EVALUATION OF THE PENTRA 80 HEMATOLOGY ANALYZER 212 212 Am J Clin Pathol 2005;123:206-214 DOI: 10.1309/6U2T6UTWK10M3NCB A B 100 75 50 % Sensitivity % Sensitivity 100 Cutoff = 0.14 Sensitivity = 98.7% Specificity = 87.6% 25 0 100 75 50 25 % Specificity 75 50 Cutoff = 0.11 Sensitivity = 100.0% Specificity = 84% 25 0 100 0 C 75 50 25 % Specificity 0 D 100 100 75 50 % Sensitivity % Sensitivity found in the present study (mean ± SD, 4.14 ± 0.88 × 106/µL [4.14 ± 0.88 × 1012/L] and 204 ± 115 × 103/µL [204 ± 115 × 109/L], respectively [data not shown]). However, none of these studies compared the results of the WBC 5-part differential obtained with the PENTRA 80 on a large series of samples with 2 or more independent reference methods, including flow cytometric immunophenotyping. In the present study, we compared the WBC 5-part differential results obtained with the PENTRA 80 not only with conventional morphologic results but also with 2 other reference methods in parallel using the same samples: another blood cell counter and a 4-color, single-platform flow cytometric immunophenotyping assay. Overall, the PENTRA 80 showed good performance compared with the 3 reference methods in the analysis of normal and abnormal samples. However, it should be noted that the quality of the results obtained varied significantly for different subpopulations of WBCs. Accordingly, performance was extremely good for neutrophils and lymphocytes, even in abnormal samples, but it clearly was lower for monocytes and very poor for basophils. Overall, these results would be in agreement with those obtained in evaluation studies of other cell counters.13,19,20 It is interesting, however, that slight variations also were observed in the results obtained depending on the reference method with which they were compared. Accordingly, the highest and the lowest coefficients of correlation for the 5 cell populations analyzed in normal and abnormal samples were obtained when the PENTRA 80 results were compared with flow cytometry and conventional microscopy, respectively; comparisons with the CELL-DYN 4000 showed intermediate correlation coefficients near those observed with flow cytometry. These observations might be related directly to the use in the new instrument evaluated herein of flow cytometry–based principles for the identification of different populations of WBCs and to the higher precision the blood cell analyzers have because of the greater numbers of cells counted.1,21,22 Despite this, it should be highlighted that when results obtained for individual samples were compared, the degree of agreement between the 2 cell counters tended to be higher than that found for flow cytometry and PENTRA 80 comparisons. As for other cell counters, the identification and enumeration of neutrophils and lymphocytes did not represent a major problem23 with any of the 4 methods used; extremely high correlation coefficients and degrees of agreement were observed consistently. Enumeration of eosinophils by the 2 cell counters and flow cytometric immunophenotyping, despite showing high correlation, consistently displayed a degree of agreement of 39% or less and even lower for comparisons with conventional morphologic examination. These results are in contrast with previous reports that have shown much better correlation for eosinophil counts between cell Cutoff = 6.8 Sensitivity = 22.2% Specificity = 99.5% 25 0 100 75 50 25 % Specificity 0 75 50 Cutoff = 0.88 Sensitivity = 40% Specificity = 100.0% 25 0 100 75 50 25 % Specificity 0 ❚Figure 3❚ Receiver operating characteristic curve plots to evaluate the sensitivity and specificity of the absolute numbers of large immature cell (A and B) and atypical lymphocyte (C and D) flags provided by the ABX PENTRA 80 hematology cell counter compared with the CELL-DYN 4000 automated hematology cell analyzer and multiparameter flow cytometry immunophenotyping. For proprietary information, see the text. counters and conventional cytology.1,14,24 Our findings could directly reflect the relatively low frequency at which eosinophils are present in blood samples because even when a total of 400 cells per sample are counted, the mean coefficient of variation of the microscopic measurement for a mean number of 2% eosinophils per sample is around 14%. By contrast, the decreased degree of correlation found for monocytes is in agreement with previous observations for most blood cell counters25-27 and could not be explained entirely in terms of their relatively low frequency in the samples analyzed. In fact, monocytes are a difficult subpopulation to identify by morphologic and light scatter characteristics; at the same time, they might not be identified in pathologic samples including other abnormal cells such as immature granulocytes or large lymphocytes. Moreover, circulating dendritic cell precursors, which represent up to 1% of all nucleated cells in normal blood,5 also might interfere with the unequivocal identification of monocytes and their accurate enumeration. Despite the efforts made in recent years for improving the identification of basophils in automated blood cell analyzers, the results presented herein once again verify that they represent © American Society for Clinical Pathology Hematopathology / ORIGINAL ARTICLE the individual subpopulation of normal blood WBCs showing the worst results. This could be explained by the fact that basophils not only usually are present at very low frequencies in blood28 but also are difficult to identify with the currently used strategies in blood cell analyzers,29,30 as supported by the low correlation reported herein between the PENTRA 80 and the CELL-DYN 4000 and between the PENTRA 80 and other cell counters evaluated subsequently in our laboratory (R2 ≤ 0.02; data not shown). An additional potential explanation for these poor results could be related to the use of conventional morphologic counts in the calibration of automated blood cell analyzers for the enumeration of basophils in normal blood samples.24 In addition to the aforementioned populations of WBCs present in normal blood and according to the specifications of the manufacturer, the PENTRA 80 also is capable of identifying 2 abnormal cell subpopulations in blood: ALYs and LICs. In the present study, up to a total of 100 abnormal samples showing the presence of these quantitative and/or qualitative morphologic flags were analyzed in parallel by the PENTRA 80 and the other 3 reference methods. For the ALY counts, the correlation observed with the 3 reference methods, except flow cytometry, was low, with a relatively low efficiency for the cutoff values used for flagging. This could be explained because the limit of detection for this subpopulation is rather low (>3.5%), and probably some normal large granular lymphocytes might be identified in the ALY flag. Also by optical microscopy, the limit to identify atypical lymphocytes usually is higher (≥7%)31; thus, some samples flagged by the PENTRA 80 are not considered abnormal by optical microscopy. However, a detailed analysis of our results shows that the PENTRA 80 overestimated the ALY counts compared with the reference methods. This could be related to the relatively arbitrary algorithms established in the cell counters for the distinction between normal and atypical lymphocytes.31-33 In fact, in several abnormal samples we observed this morphologic flag, but no atypical lymphocytes could be detected by optical microscopy or by flow cytometry (data not shown). In this sense, our results suggest that by increasing the flagging cutoff values for ALYs, the efficiency of the PENTRA 80 could be improved significantly but at the expense of sensitivity. It is interesting that relatively good correlation was observed for the percentage of LICs found with the PENTRA 80 and all 3 reference methods. Despite this, the degree of agreement observed for individual samples was acceptable only for the comparisons of the number of LICs obtained with the PENTRA 80 and flow cytometric immunophenotyping. The cutoff values selected in our study as showing the highest efficiency displayed only slight improvement compared with those currently used by the PENTRA 80 instrument (data not shown). However, it should noted that the inability to distinguish LICs corresponding to blasts from immature granulocytes might be a limitation of this instrument. Our results showed good performance of the 5-part leukocyte differential blood cell analyses by the PENTRA 80, especially for the enumeration of neutrophils, lymphocytes, and eosinophils (performance was lower for monocytes and poor for basophils). In addition, the PENTRA 80 also showed good performance for the detection of LICs (promyelocytes, myelocytes, metamyelocytes, and blasts) but was less effective for the identification of ALYs in relatively low frequencies in abnormal PB samples. From the 1General Cytometry Service, Cancer Research Center and Department of Medicine, University of Salamanca; and 2Research Unit and 3Hematology Service, University Hospital of Salamanca, Salamanca, Spain. Supported by a grant from the Ministry of Science and Technology (Ramón y Cajal Program), Madrid, Spain (Dr Tabernero). Address reprint requests to Dr Orfao: Servicio General de Citometría. Laboratorio de Hematología, Hospital Universitario Salamanca, Paseo San Vicente, 58-182, 37007 Salamanca, Spain. Acknowledgments: We thank ABX Diagnostics Iberica Group, Madrid, Spain, and ABX Diagnostics Group, Montpellier, France, for their support with equipment (PENTRA 80) for this study. References 1. Van den Bossche J, Devreese K, Malfait R, et al. Reference intervals for a complete blood count determined on different automated haematology analyzers: ABX PENTRA 120 Retic, Coulter Gen-S, Sysmex SE-9500, Abbott Cell-Dyn 4000 and Bayer Advia 120. Clin Chem Lab Med. 2002;40:69-73. 2. Morillon M, Maslin J, De Pina JJ, et al. Evaluation of the monocyte counting by the ABX Vega: comparison with the manual method and fluoro-flow cytometry. Hematol Cell Ther. 1999;41:47-50. 3. Siekmeier R, Bierlich A, Jaros W. The white blood cell differential: three methods compared. Clin Chem Lab Med. 2001;39:432-445. 4. National Committee for Clinical Laboratory Standards. Reference Leukocyte Differential Count (Proportional) and Evaluation of Instrumental Methods; Approved standard. H20-A. Villanova, PA: NCCLS; 1992. 5. Almeida J, Bueno C, Algueró MC, et al. Comparative analysis of the morphological, cytochemical, immunophenotypical, and functional characteristics of normal human peripheral blood lineage–/CD16+/HLADR+/CD14–/lo cells, CD14+ monocytes, and CD16– dendritic cells. Clin Immunol. 2001;100:325-338. 6. Mandy F, Brando B. Enumeration of absolute cell counts using immunophenotypic techniques. In: Robinson JP, Darzynkiewicz Z, Dean PN, et al, eds. Current Protocols in Cytometry. New York, NY: Wiley; 2000;6.8:1-26. 7. Lebeck LK, Mast B, Houwen BH. Flow cytometry white blood cell differentials: a proposed alternate reference method. Sysmex J Int. 1993;97:352-354. 8. Orfao A, Ruiz-Arguelles A, Lacombe F, et al. Flow cytometry: its applications in hematology. Haematologica. 1995;80:69-81. Am J Clin Pathol 2005;123:206-214 © American Society for Clinical Pathology 213 DOI: 10.1309/6U2T6UTWK10M3NCB 213 213 Arroyo et al / EVALUATION OF THE PENTRA 80 HEMATOLOGY ANALYZER 9. Orfao A, Schmitz G, Brando B, et al. Clinically useful information provided by the flow cytometric immunophenotyping of hematological malignances: current status and future directions. Clin Chem. 1999;45:1708-1717. 10. Van der Meer W, Dinnissen JWB, de Keijzer MH, et al. Comparative evaluation of Abbott Cell-Dyn 3700, Bayer Advia 120, and Sysmex XE-2100 automated leukocyte differentials and flagging. Lab Hematol. 2002;8:102-117. 11. Orfao A, inventor; University of Salamanca, Salamanca, Spain, assignee. Multidimensional leukocyte differential analysis. European patent number 02380098.0. May 2002. 12. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;8:307-311. 13. Langford K, Luchtman-Jones L, Miller R, et al. Performance evaluation of the Sysmex XT-2000i automated hematology analyzer. Lab Hematol. 2003;9:29-37. 14. Buttarello M, Bulian P, Temporin V, et al. Sysmex SE-9000 hematology analyzer: performance evaluation on leukocyte differential counts using an NCCLS H20-A protocol. Am J Clin Pathol. 1997;108:674-686. 15. Picard F, Gicquel C, Marnet L, et al. Preliminary evaluation of the new hematology analyzer COULTER GEN-S in a University Hospital. Clin Chem Lab Med. 1999;37:681-686. 16. Fournier M, Gireau A, Chretien MC, et al. Laboratory evaluation of the Abbott Cell DYN 3500 5-part differential. Am J Clin Pathol. 1996;105:286-292. 17. Alexander T. Absolute CD4 counts obtained by a three-color flow cytometry method without the use of a hematology analyzer. Clin Diagn Lab Immunol. 1998;5:266-269. 18. Deneys V, Mazzon AM, Marques JL, et al. Reference values for peripheral blood B-lymphocyte subpopulations: a basis for multiparameter immunophenotyping of abnormal lymphocytes. J Immunol Methods. 2001;253:23-36. 19. Grimaldi E, Scopacasa F. Evaluation of the Abbott CELLDYN 4000 hematology analyzer. Am J Clin Pathol. 2000;113:497-505. 20. Aulesa C, Pastor I, Naranjo D, et al. Application of receiver operating characteristics curve (ROC) analysis when definitive and suspect morphologic flags appear in the new Coulter LH 750 analyzer. Lab Hematol. 2004;10:14-23. 21. Igout J, Fretigny M, Vasse M, et al. Evaluation of the Coulter LH 750 haematology analyzer compared with flow cytometry as the reference method for WBC, platelet and nucleated RBC count. Clin Lab Haematol. 2004;26:1-7. 214 214 Am J Clin Pathol 2005;123:206-214 DOI: 10.1309/6U2T6UTWK10M3NCB 22. Kondo H, Akiyama T, Tatsumi N, et al. Performance evaluation of complete blood count and white blood cell differential parameters obtained using a Celltac F automated hematology analyzer. Lab Hematol. 2004;10:3-13. 23. Ruzicka K, Veitl M, Thalhammer-Scherrer R, et al. The new hematology analyzer Sysmex XE-2100: performance evaluation of a novel white blood cell differential technology. Arch Pathol Lab Med. 2001;125:391-396. 24. Davis BH, Bigelow NC. Performance evaluation of a hematology blood counter with five-part leukocyte differential capability. Am Clin Lab. 1999;18:8-9. 25. Hubl W, Tlustos L, Erath A, et al. Proposed reference method for peripheral-blood monocyte counting using fluorescencelabelled monoclonal antibodies. Cytometry. 1996;26:69-74. 26. Vives Corrons JL, Besson I, Villamor N, et al. Evaluation of the reproducibility of automated monocyte counting using flow cytometry and monoclonal antibodies [in Spanish]. Sangre (Barc). 1994;39:197-201. 27. Goossens W, Van Hove L, Verwilghen RL. Monocyte counting: discrepancies in results obtained with different automated instruments. J Clin Pathol. 1991;44:224-227. 28. Davies S, Bain BJ. Basophil counts on the Technicon H*1 automated counter. Clin Lab Haematol. 1996;18:35-38. 29. Brigden ML, Page NE, Graydom C. Evaluation of the Sysmex NE-8000 automated hematology analyzer in a high-volume outpatient laboratory. Am J Clin Pathol. 1993;100:618-625. 30. al-Ismail SA, Bond K, Carter AB, et al. Two-centre evaluation of the Abbott CD3500 blood counter. Clin Lab Haematol. 1995;17:11-21. 31. Hoffmann JJ, Hoedemakers RM. Diagnostic performance of the variant lymphocyte flag of the Abbott Cell-Dyn 4000 haematology analyzer. Clin Lab Haematol. 2004;26:9-13. 32. van der Meer W, Scott CS, de Keijzer MH. Automated flagging influences the inconsistency and bias of band cell and atypical lymphocyte morphological differentials. Clin Chem Lab Med. 2004;42:371-377. 33. Ferrero-Vacher C, Sudaka I, Jambou D, et al. Evaluation of the ABX Cobas Vega automated hematology analyzer and comparison with the Coulter STKS. Hematol Cell Ther. 1997;39:149-158. © American Society for Clinical Pathology
© Copyright 2026 Paperzz