56 ������ Ion-selective electrodes, ISE’s With glass membrane Built similarly to a pH electrode, but using glass which is sensitive to a certain ion. In practice, only electrodes for sodium ions are manufactured. It is an indicator electrode which must always be used with a reference electrode. See page 43. With solid-state membrane The sensitive membrane is a mixture of salts or a crystal in the case of the F- ISE. The membrane and the internal electrolyte can be replaced. They are indicator electrodes, and should always be used with a reference electrode. See page 43. 96 50, Sodium Na+. Measuring range: 0.02 ppm... 20 g/l Linear measuring range: From 1 ppm Working conditions: pH 8 ... 11, 0 ... 80°C Interferences: Ag+, NH4+, K+, Li+, H+ Electrolyte for the reference electrode: NH4Cl 0.1M Applications: Foodstuffs, biological solutions, the environment, water, etc. 96 51, Bromide Br-. Measuring range: 0.4 ppm ... 80 g/l Linear measuring range: From 5 ppm Working conditions: pH 2 ... 12, 0 ... 80°C Interferences: S2-, CN-, OH-, I-, Cl-, S2O32-. Electrolyte for the reference electrode: KNO3 1M Applications: photographic emulsions 96 52, Chloride Cl-. Measuring range: 0.4 ppm ... 35 g/l Linear measuring range: From 5 ppm Working conditions: pH 2 ... 11, 0 ... 80°C Interferences: S2-, Br-, I-, CN-. Electrolyte for the reference electrode: KNO3 1M Applications: foodstuffs, waters, etc. 96 53, Cyanide CN-. Measuring range: 0.03 ppm ... 2.6 g/l Linear measuring range: From 0.5 ppm Working conditions: pH 11 ... 13, 0 ... 50ºC Interferences: Cl-,I-, Br-, S2O32-, CO32-, CrO42-. Electrolyte for the reference electrode: KNO3 1M Applications: waters, galvanizing baths. 96 54, Ag Sulphide//Ag2S. Measuring range: 0.01 ppm ... 108 g/l as silver, 0.3...32 g/l as sulphide Linear measuring range: From 0.1 ppm Ag+, 0.5 ppm S2Working conditions: Ag+, pH 1...9, S2-, pH 2...14, 0...50ºC. Interferences: There should be no Hg2+ and proteins. Electrolyte for the reference electrode: KNO3 1M Applications: waters � �� � � ��� 96 55, Fluoride F-. Measuring range: 0.02 ppm ... 20 g/l Linear measuring range: From 0.1 ppm Working conditions: pH 5 ... 8, 0 ... 80ºC Interferences: OH-, high concentrations of Lanthanum complexers Electrolyte for the reference electrode: KCl 3M Applications: potable waters, toothpastes, milk, etc. �� �� ��� �� �� ��� �� �� ��� �� �� ��� �� �� ��� �� �� �� �� �� �� �� �� ���� �� �� ���� �� �� �� �� �� ���� �� �� ���� ��������� ���������� ������ ��������� 96 56, Iodide I-. Measuring range: 0.01 ppm ... 127 g/l Linear measuring range: From 0.5 ppm Working conditions: pH 0 ... 12, 0 ... 50ºC Interferences: There should be no Hg2+, Cl-, Br-, S2-, S2O32-, CN-, CO32-, CrO42-. Electrolyte for the reference electrode: KNO3 1M Applications: Plants, milk, waters. Laboratory Catalogue 2004 1st Edition Ion-selective electrodes, ISE’s 96 60, Calcium Ca2+. Measuring range: 0.04 ppm ... 40 g/l Linear measuring range: From 10 ppm Working conditions: pH 3 ... 12, 0 ... 50°C Interferences: Cation detergents, Sr2+, Ba2+, Zn2+, Mg2+, Li+, Cs+, Na+, K+, Rb+ and NH4+. Electrolyte for the reference electrode: KCl 3M Applications: milks, serums, soils, waters. 96 61, Potassium K+. Measuring range: 0.04 ppm ... 39 g/l Linear measuring range: From 5 ppm Working conditions: pH 2 ... 12, 0 ... 50°C Interferences: Cation detergents, Rb+, Na+, H+, NH4+, Cs+, Mg2+, Ca2+, Ba2+, Sr2+. Electrolyte for the reference electrode: Al2(SO4)3 0.9M Applications: potable waters, toothpaste, milk, etc. 96 62, Nitrate NO3-. Measuring range: 0.6 ppm ... 60 g/l Linear measuring range: From 10 ppm Working conditions: pH 3 ... 12, 0 ... 50°C Interferences: Anion detergents, I-, Br-, F-, Cl-, SCN-, MnO4-, HCO3-, NO2-, ClO4-, SO42-. Electrolyte for the reference electrode: : Al2(SO4)3 0.9M Applications: fertilizers, waters, meats, vegetables. 96 63, Ammonium NH4+. Measuring range: 0.02 ... 18 g/l Linear measuring range: From 5 ppm Working conditions: pH 4 ... 7, 0 ... 50°C Interferences: Cation detergents, K+, Na+, Li+, H+, Rb+, Cs+, Mg2+, Ca2+, Ba2+, Sr2+. Electrolyte for the reference electrode: Al2(SO4)3 0.9M Applications: foodstuffs, fertilizers, boiler waters. Description Electrodes Specify code and name of the ion to be measured IMPORTANT: Solid-state and polymer-membrane selective electrodes are supplied with an electrolyte solution. Spare membranes 96 70 Bromide (Br-), 1 membrane 96 71 Chloride (Cl-), 1 membrane 96 72 Cyanide (CN-). 1 membrane 96 73 Ag Sulphide (Ag2S), 1 membrane 96 74 Fluoride (F-), 1 membrane 96 75 Iodide (I-), 1 membrane 96 76 Copper (Cu2+), 1 membrane 96 80 Calcium (Ca2+), 1 membrane 96 81 Potassium (K+), 1 membrane 96 82 Nitrate (NO3-), 1 membrane 96 83 Ammonium (NH4+), 1 membrane Spare electrolytic solutions 96 87 Bromide (Br-), 10 ml flask, 2 units 96 88 Chloride (Cl-), 10 ml flask, 2 units 96 89 Cyanide (CN-), 10 ml flask, 2 units 96 90 Ag Sulphide (Ag2S), 10 ml flask, 2 units 96 91 Fluoride (F-), 10 ml flask, 2 units 96 92 Iodide (I-), 10 ml flask, 2 units 96 93 Copper (Cu2+), 10 ml flask, 2 units 96 94 Calcium (Ca2+), 10 ml flask, 2 units 96 95 Potassium (K+), 10 ml flask, 2 units 96 96 Nitrate (NO3-), 10 ml flask, 2 units 96 97 Ammonium (NH4+), 10 ml flask, 2 units ������ ���������� ������������� With polymer membrane The polymer membrane is made up of several ion-selective substances, depending on the ion to be measured. The membrane and the internal electrolyte can be replaced. Used together with a reference electrode. Ordering information Code ���������� ������ 96 57, Copper Cu2+-. Measuring range: 0.1 ppm ... 6 g/l Linear measuring range: From 1 ppm Working conditions: pH 1 ... 6.5, 0 ... 50°C Interferences: Pb2+ and Cd2+. Electrolyte for the reference electrode: KNO3 1M Applications: galvanising baths. 57 Ionic Strength Adjusters (ISA’s) Aluminium Sulphate Al2(SO4)3 0.9M. This is the ionic strength adjuster and the reference electrolyte for potassium-, nitrate- and ammoniumselective electrodes TISAB III This is the ionic strength adjuster for the fluoride selective electrode. Laboratory Catalogue 2004 1st Edition Ionic strength adjusters, ISA 99 00 TISAB III, for F- electrode, 250 ml flask 99 02 Aluminium sulphate, for K+,NO3-,NH4+, 250 ml flask 58 Ion-selective electrodes, ISE’s (continued) No reference electrode is required. ������ Etc, … DIN 96 65, Ammonia NH3 Measuring range: 0.1 ppm... 17 g/l Linear measuring range: From 1 ppm Working conditions: pH 4...13, 0... 50ºC Interferences: volatile amines. At pH < 11, CO2, SO2, Cl2, H2S. Applications: waste waters, Kjeldahl. 96 66, Carbon dioxide CO2 Measuring range: 2.6 ppm ... 0.9 g/l Linear measuring range: From 5 ppm Working conditions: pH 1...5, 0 ... 50ºC Interferences: volatile acids Applications: mineral waters, wines, bacteriological cultures. Spare parts USA ��������� Note: These electrodes are manufactured with fixed cable and BNC connector. They can also be supplied, under demand, with another connector. See page 60. Price does not vary. �� ���������� ������������� With membrane for gases These are combined electrodes integrating a pH electrode. This pH electrode is immersed in an electrolyte and separated from the solution to be measured by a gas-permeable membrane. The membrane and the internal electrolyte can be replaced. That is why they are supplied in boxes of 20 membranes and a flask of 50 ml of electrolyte. The membrane for these electrodes is PTFE and is common for both NH3 and CO2. �Membranes �������� (20 ���u.) ��� ����������� ��� Electrolyte (50 ��� ml) � �� � � ��� Ordering information Code Description Electrodes 96 65 Combined electrode (NH3), BNC connector 96 66 Combined electrode (CO2), BNC connector IMPORTANT: Selective electrodes with membranes for gas are supplied with 20 membranes and 1 electrolyte solution. Spare membranes (common for both electrodes) 96 85 Set of 20 membranes for NH3 or CO2 electrodes Spare electrolytic solutions 96 65 NH3 96 66 CO2 96 98 Ammonia (NH3), 50 ml flask 96 99 Carbon dioxide (CO2), 50 ml flask Laboratory Catalogue 2004 1st Edition
© Copyright 2026 Paperzz