Report No: L121605937R02 8165 E Kaiser Blvd. Anaheim, CA 92808 www.lightlaboratory.com L121605937R02 Report No: Report Prepared For: Model Number: Test: Issue Date: 2/17/2017 Revolution Lighting 4139 Guardian St. Simi Valley CA 93063 204481-113 Electrical and Photometric tests Standards Used: Appropriate part or all test guidelines were used for test performed: IESNA LM79: 2008 Approved Methods for Electrical and Photometric Measurements of Solid-State Lighting Products ANSI NEMA ANSLG C78.377: 2008 Specification of the Chromaticity of Solid State Lighting Products ANSI C82.77:2002: Harmonic Emission Limits-Related Quality Requirements for Lighting Equipment Description of Sample: Client submitted the sample. Received in working and undamaged condition. No modifications were necessary. Testing Condition: SL 28X2 -I580 120-277W D1 M DRIVER was connected to two led tubes from the driver output. Photometric measurements were measured from a single led tube while the other tube was powered and covered with black velvet to prevent any light pollution. Input power of single measured module is calculated from the total power divided by two. Sample Arrival Date: 12/21/16 Date of Tests: Seasoning of Sample: 1/9/17 - 1/9/17 No seasoning was performed in accordance with IESNA LM-79. Equipment List Equipment Used Chroma Programmable AC Source Yokogawa Digital Power Meter ITECH Fluke Digital Thermometer LLI Type C Goniophotometer System LLI 2M Sphere LLI Spectroradiometer Model No Stock No Calibration Due Date 61604 WT210 IT6122 52k/J RMG-C-MKII 2MR97 SPR-3000 PS-AC02 MT-EL06-S1 PS-DC03-S1 MT-TP02-GC CD-LL04-GC CD-SN03-S2 MT-SC01-S2 -11/28/17 11/28/17 11/28/17 --Before Use *All Results in accordance to IESNA LM-79-2008: Approved Method for the Electrical and Photometric Testing of Solid-State Lighting. Page 1 of 4 Report No: L121605937R02 8165 E Kaiser Blvd. Anaheim, CA 92808 www.lightlaboratory.com Test Summary Manufacturer: Model Number: Driver Model Number: Total Lumens: Input Voltage (VAC/60Hz): Input Current (Amp): Input Power (W): Input Power Factor: Current ATHD @ 120V(%): Current ATHD @ 277V(%): Efficacy: Color Rendering Index (CRI): Correlated Color Temperature (K): Chromaticity Coordinate x: Chromaticity Coordinate y: Ambient Temperature (oC): Stabilization Time (Hours): Total Operating Time (Hours): Off State Power(W): Revolution Lighting 204481-113 SL 28X2 -I580 120-277W D1 M 3511.53 120.00 0.24 28.19 1.00 5% N/A 125 84 3898 0.3869 0.3860 25.0 1:30 2:05 0.00 FIG. 1 LUMINAIRE *All Results in accordance to IESNA LM-79-2008: Approved Method for the Electrical and Photometric Testing of Solid-State Lighting. Page 2 of 4 Report No: L121605937R02 Percent Output 8165 E Kaiser Blvd. Anaheim, CA 92808 www.lightlaboratory.com Spectral Power 1.0 0.8 0.6 0.4 0.2 0.0 380 Wavelength W/m2nm 380 390 400 410 420 430 0.0010 0.0012 0.0026 0.0105 0.0460 0.1373 CRI & CCT x 0.3869 y 0.3860 u' 0.2257 v' 0.5065 CRI 84.40 CCT 3898 Duv 0.00244 R Values R1 82.53 R2 91.35 R3 96.72 R4 82.04 R5 82.45 R6 88.01 R7 86.31 R8 65.40 R9 13.49 R10 79.47 R11 81.07 R12 65.63 R13 84.87 R14 98.36 430 440 450 460 470 480 490 500 480 0.3451 0.8714 0.6945 0.4657 0.3571 0.4145 0.5133 530 510 520 530 540 550 560 570 580 Wavelength(nm) 0.5954 0.6447 0.6879 0.7387 0.7950 0.8509 0.9071 580 590 600 610 620 630 640 630 680 0.9595 0.9943 0.9964 0.9580 0.8824 0.7788 0.6625 650 660 670 680 690 700 710 730 0.5435 0.4330 0.3339 0.2537 0.1897 0.1410 0.1043 780 720 730 740 750 760 770 780 0.0766 0.0564 0.0418 0.0310 0.0233 0.0174 0.0151 CIE 1931 Chromaticity Diagram Y 0.9 520 0.8 530 540 510 0.7 550 560 0.6 570 500 0.5 580 590 0.4 600 0.3869, 0.3860 0.3 610 620 630 640 650 700 490 0.2 480 0.1 470 460 0 0 0.1 440 430 380 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 X *All Results in accordance to IESNA LM-79-2008: Approved Method for the Electrical and Photometric Testing of Solid-State Lighting. Page 3 of 4 Report No: L121605937R02 8165 E Kaiser Blvd. Anaheim, CA 92808 www.lightlaboratory.com Test Methods Photometric Measurements - Goniophotometer A Custom Light Laboratory Type C Rotating Mirror Goniophotometer was used to measure candelas(intensity) at each angle of distribution as defined by IESNA for the appropriate fixture type. Ambient temperature is set to 25oC and is measured from the center of the fixture, within 1ft from the outside of the fixture. Temperature is maintained at 25oC throughout the testing process and the sample is stabilized for at least 30mins and longer as necessary for the sample to achieve stabilization. Electrical measurements are measured using the listed equipment. Spectral Measurements - Integrating Sphere A Sensing Spectroradiometer SPR-3000, in conjunction with Light Laboratory 2 meter integrating sphere was used to measure chromaticity coordinates, correlated color temperature(CCT) and the color rendering index(CRI) for each sample. Ambient temperature is set to 25oC and is measured from the center of the fixture, within 1ft from the outside of the fixture. Temperature is maintained at 25oC throughout the testing process and the sample is stabilized for at least 30mins and longer as necessary for the sample to achieve stabilization. Electrical measurements are measured using the listed equipment. Disclaimers: This report must not be used by the customer to claim product certification, approval or endorsement by NVLAP, NIST or any agency of Federal Government. Report Prepared by : Keyur Patel Test Report Released by: Test Report Reviewed by: Jeff Ahn Engineering Manager Steve Kang Quality Assurance *Attached are photometric data reports. Total number of pages: 10 *All Results in accordance to IESNA LM-79-2008: Approved Method for the Electrical and Photometric Testing of Solid-State Lighting. Page 4 of 4 IES INDOOR REPORT PHOTOMETRIC FILENAME : L121605937R02.IES DESCRIPTION INFORMATION (From Photometric File) IESNA:LM-63-2002 [TEST] L121605937R02 [TESTLAB] LIGHT LABORATORY, INC. [ISSUEDATE] 2/17/2017 [MANUFAC] REVOLUTION LIGHTING [LUMCAT] 204481-113 [LUMINAIRE] 2-LAMP EXTERNAL DRIVER T5 LAMP KIT 28W/LAMP 4000K DIMMABLE [BALLASTCAT] SL 28X2 -I580 120-277W D1 M [LAMPPOSITION] 0,0 [LAMPCAT] N/A [OTHER] INDICATING THE CANDELA VALUES ARE ABSOLUTE AND [MORE] SHOULD NOT BE FACTORED FOR DIFFERENT LAMP RATINGS. [_TEST CONDITION] SL 28X2 -I580 120-277W D1 M DRIVER WAS CONNECTED [MORE] TO TWO LED TUBES FROM THE DRIVER OUTPUT. [MORE] PHOTOMETRIC MEASUREMENTS WERE MEASURED FROM [MORE] A SINGLE LED TUBE WHILE THE OTHER TUBE WAS POWERED [MORE] AND COVERED WITH BLACK VELVET TO PREVENT ANY LIGHT POLLUTION. [MORE] INPUT POWER OF SINGLE MEASURED MODULE IS CALCULATED [MORE] FROM THE TOTAL POWER DIVIDED BY TWO. [INPUT] 120VAC, 28.19W [TEST PROCEDURE] IESNA:LM-79-08 CHARACTERISTICS Lumens Per Lamp Total Lamp Lumens Luminaire Lumens Total Luminaire Efficiency Luminaire Efficacy Rating (LER) Total Luminaire Watts Ballast Factor CIE Type Spacing Criterion (0-180) Spacing Criterion (90-270) Spacing Criterion (Diagonal) Basic Luminous Shape Luminous Length (0-180) Luminous Width (90-270) Luminous Height N.A. (absolute) N.A. (absolute) 3512 N.A. 125 28.19 1.00 Semi-Direct 1.24 1.42 1.48 Hor. Cylinder Along Length 3.67 ft 0.04 ft (Diameter) 0.04 ft (Diameter) Photometric Toolbox Professional Edition - Copyright 2002-2012 by Lighting Analysts, Inc. Calculations based on published IES Methods and recommendations, values rounded for display purposes. Results derived from content of manufacturers photometric file. Page 1 IES INDOOR REPORT PHOTOMETRIC FILENAME : L121605937R02.IES LUMINANCE DATA (cd/sq.m) Angle In Degrees Average 0-Deg Average 45-Deg Average 90-Deg 45 55 65 75 85 5088055 3350228 2010360 939462 184778 51882 45519 39215 33432 28255 41917 38918 35676 32196 28638 Photometric Toolbox Professional Edition - Copyright 2002-2012 by Lighting Analysts, Inc. Calculations based on published IES Methods and recommendations, values rounded for display purposes. Results derived from content of manufacturers photometric file. Page 2 IES INDOOR REPORT PHOTOMETRIC FILENAME : L121605937R02.IES CANDELA TABULATION 0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 100 105 110 115 120 125 130 135 140 145 150 155 160 165 170 175 180 0.0 666.16 663.05 652.83 635.40 611.98 583.67 549.45 509.68 466.92 420.42 362.21 320.69 267.30 212.91 158.43 106.04 60.95 21.51 4.48 2.49 3.32 4.24 3.57 2.49 1.99 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 22.5 666.16 664.25 655.04 639.67 618.67 592.84 561.70 527.49 489.21 448.52 406.05 362.46 316.62 276.72 236.03 198.13 165.82 139.09 118.66 104.25 94.99 88.19 84.70 84.70 85.40 85.61 84.08 83.83 77.27 66.47 56.09 48.00 37.16 19.43 6.93 3.49 0.00 45.0 666.16 663.46 656.57 645.07 629.34 609.66 586.41 561.12 533.26 503.83 473.31 442.46 411.61 381.47 349.79 325.38 298.85 275.06 252.81 232.50 214.69 199.00 184.96 171.68 158.81 150.80 143.03 134.06 122.85 112.27 106.50 95.70 73.61 56.17 33.13 8.06 0.00 67.5 666.16 664.13 659.64 651.84 640.96 627.51 611.69 593.92 574.45 553.02 530.98 507.44 483.56 458.99 434.28 409.58 384.83 361.33 338.04 315.71 295.24 274.93 255.63 237.28 219.72 202.61 186.63 173.75 162.25 151.58 134.48 118.83 106.00 69.00 48.20 12.12 0.00 90.0 666.16 663.63 660.06 653.58 644.45 633.74 621.03 606.50 589.56 572.21 552.11 531.27 509.68 487.01 463.43 439.51 414.77 390.94 367.94 345.18 323.01 301.09 280.83 260.24 240.89 221.87 203.77 186.33 173.13 160.34 145.90 134.27 121.32 94.99 57.13 13.78 0.00 Photometric Toolbox Professional Edition - Copyright 2002-2012 by Lighting Analysts, Inc. Calculations based on published IES Methods and recommendations, values rounded for display purposes. Results derived from content of manufacturers photometric file. Page 3 IES INDOOR REPORT PHOTOMETRIC FILENAME : L121605937R02.IES ZONAL LUMEN SUMMARY Zone Lumens %Lamp %Fixt 0-20 0-30 0-40 0-60 0-80 0-90 10-90 20-40 20-50 40-70 60-80 70-80 80-90 90-110 90-120 90-130 90-150 90-180 110-180 0-180 245.43 526.73 877.64 1653.06 2335.79 2604.86 2541.71 632.21 1018.7 1138.97 682.73 319.18 269.07 415.70 571.08 694.61 852.95 906.67 490.97 3511.53 N.A. N.A. N.A. N.A. N.A. N.A. N.A. N.A. N.A. N.A. N.A. N.A. N.A. N.A. N.A. N.A. N.A. N.A. N.A. N.A. 7.00 15.00 25.00 47.10 66.50 74.20 72.40 18.00 29.00 32.40 19.40 9.10 7.70 11.80 16.30 19.80 24.30 25.80 14.00 100.00 Total Luminaire Efficiency = N.A.% ZONAL LUMEN SUMMARY Zone Lumens 0-10 10-20 20-30 30-40 40-50 50-60 60-70 70-80 80-90 90-100 100-110 110-120 120-130 130-140 140-150 150-160 160-170 170-180 63.15 182.29 281.29 350.92 386.49 388.94 363.55 319.18 269.07 226.18 189.52 155.38 123.53 93.76 64.58 38.05 14.26 1.41 Photometric Toolbox Professional Edition - Copyright 2002-2012 by Lighting Analysts, Inc. Calculations based on published IES Methods and recommendations, values rounded for display purposes. Results derived from content of manufacturers photometric file. Page 4 IES INDOOR REPORT PHOTOMETRIC FILENAME : L121605937R02.IES COEFFICIENTS OF UTILIZATION - ZONAL CAVITY METHOD Effective Floor Cavity Reflectance 0.20 RC RW 80 70 50 30 10 70 70 50 30 10 50 50 30 10 30 50 30 10 10 50 30 10 0 0 0 1 2 3 4 5 6 7 8 9 10 113 113 113 113 100 94 88 84 90 80 72 66 81 69 60 53 74 61 52 44 67 54 45 38 62 48 39 32 57 44 35 28 53 40 31 25 50 36 28 22 46 33 25 20 107 107 107 107 94 89 84 80 84 76 69 63 76 66 58 51 69 58 49 43 63 51 43 36 58 46 37 31 54 41 33 27 50 38 30 24 47 34 27 21 44 32 24 19 97 80 68 59 52 46 42 38 34 31 29 87 71 61 53 47 42 37 34 31 29 26 78 64 54 47 42 37 34 31 28 26 24 74 55 44 36 30 25 22 19 17 15 14 97 76 62 52 45 39 34 31 27 25 22 97 73 58 47 39 34 29 25 22 20 18 87 68 56 48 41 36 31 28 25 23 21 87 66 52 43 36 31 27 23 21 19 17 78 61 51 43 37 32 28 25 23 21 19 Photometric Toolbox Professional Edition - Copyright 2002-2012 by Lighting Analysts, Inc. Calculations based on published IES Methods and recommendations, values rounded for display purposes. Results derived from content of manufacturers photometric file. 78 59 47 39 33 28 24 22 19 17 16 Page 5 IES INDOOR REPORT PHOTOMETRIC FILENAME : L121605937R02.IES POLAR GRAPH 666 500 333 3 167 2 1 Maximum Candela = 666.16 Located At Horizontal Angle = 0, Vertical Angle = 0 # 1 - Vertical Plane Through Horizontal Angles (0 - 180) (Through Max. Cd.) # 2 - Vertical Plane Through Horizontal Angles (90 - 270) # 3 - Horizontal Cone Through Vertical Angle (0) (Through Max. Cd.) Photometric Toolbox Professional Edition - Copyright 2002-2012 by Lighting Analysts, Inc. Calculations based on published IES Methods and recommendations, values rounded for display purposes. Results derived from content of manufacturers photometric file. Page 6
© Copyright 2026 Paperzz