Precision Pressure Transducer
Variable Range Force Transducer
The TSD104A measures direct arterial or venous blood pressure in
animals or records pressure changes within an in-vitro circuit (e.g.
Langendorff heart preparation). The TSD104A connects to tubing via
standard Luer Lok fittings. The transducer is small and lightweight
and the slotted base, with integral Velcro® strap, permits easy mounting. The TSD104A interfaces with the DA100C general-purpose transducer amplifier. The transducer is supplied non-sterile but can be
cold sterilized. Replacement transducers (RX104A) can be purchased without the cable, making the TSD104A a very economical
solution for pressure sensing applications.
The TSD105A interfaces with the DA100C general-purpose transducer amplifier to measure a wide range of forces (1mg to 1000g) in a
variety of applications, including very fine in-vitro tissue bath work
as well as larger in-vivo applications. Forces are conveyed to the
strain beam via an adjustable pivot, five-position, precision lever arm
to insure accurate measurements. The transducer comes with a versatile mounting rod assembly to support multiple orientations. When
fine tension and position adjustments are required, the mounting
rod can be clamped to an HDW100A.
Previous
Next
Table of
Contents
Product
Index
Search
Home
Catalog No.
Specifications
Contact
Us
BS4 72-3715
Operational Pressure
-50 to 300 mmHg
Overpressure
-400 to +4000 mmHg
Dynamic Response
100 Hz
$
Model
595.00 TSD105A
Product
Variable Range Force Transducer
Transducer Dimensions 67 mm long, 25 mm wide
Previous
Weight
11.5 g
MRI compatible
Yes (no ferrous parts)
Unbalance
50 mmHg max
Connection Ports
Male Luer (2)
8-hour Drift
1mm Hg after 5 minute warm-up
Isolation
≤5 µ A leakage at 120 VAC/60 Hz
Defibrillation
Withstands 5 discharges of 400 joules in 5 minutes across
a load
Next
Operating Temperature +15 to +40 °C
Storage Temperature
-30 to +60 °C
Combined Effects of
Sensitivity, Linearity
& Hysteresis
1mm Hg (nominal)
Output
5 µ V/mmHg (normalized to 1 V excitation)
Cable Length
3m
Interface
DA100C—see page I106
Catalog No.
$
Model
Product
BS4 72-3704
295.00 TSD104A
Transducer and Cable
BS4 72-3705
95.00 RX104A
Replacement Element
For HDW100A Force Transducer Tension Adjuster,
see page I127.
For TSD125 Series Fixed Range Force Transducer,
see page I127.
I121
11,000 Specialty Products to Enhance Your Bioresearch
U.S. Toll Free: (800) 272-2775 • Fax: (508) 429-5732 • Online: www.harvardapparatus.com