LC/MS Crossroads Jerry Pappas Sales Representative 265 Davidson Avenue Somerset, NJ 08873 [email protected] 732-698-2778 1 Comparison of Quads and Traps Ion Traps Quadrupoles Mass Separation in time Mass Separation in space High sensitivity Full Scan Lower sensitivity Full Scan Lower sensitivity SIM and SRM High sensitivity SIM and SRM Offer multiple stages of MSn Offers Only MS or MS/MS Parent and neutral loss scans 2 LCQ Instrumentation Classic Duo 3 Deca Duo/Deca Comparisons LCQ DUO LCQ DECA • 400um capillary • 500um capillary • 2 octopoles • 1 square quadrupole & 1 octopole •One rotary pump •2 rotary pumps Deca: approximately 10 x better signal than Duo 4 Current LCQ Generation 5 Advantage/XP Comparisons • 450um Ion Transfer Tube Ç • 550um Ion Transfer Tube Ç •Orthogonal Probes •Orthogonal Probes XP: approximately 10 x better signal than Advantage 6 LCQ — MSn Quadrupole Ion Trap LC Pump ESI Quadrupole Ion Trap Ion optics Detector Syringe Pump Quadrupole refers to the shape of the ion confining field inside the trap and not the shape of or number of electrodes in the trap. 7 Mass Spectrometry Simplified Generate Move S elect D etect 8 Ion production Ion optics Mass filter Electron multiplier The Mass Spectrum Red Light, all colors Green Prism Blue 100 Ions, various masses Mass Spectrometer 200 300 9 Trace produced by summing all observed masses in each scan Total Ion Chromatogram or TIC 10 Ionization vs. Fragmentation API Soft No Fragments 11 CI Ionization EI Hard Fragments EI and CI Mass Spectra of Ephedrine 58 100 % Relative Intensity EI 50 0 0 100 12 40 60 80 100 120 CI 20 140 160 180 148 (M+H)+ 166 50 0 0 MW = 165Th 20 40 60 80 100 m/z 120 140 160 180 Ionization Techniques 200,000 ESI 15,000 1,000 Molecular Weight APCI TSP GC Non Polar 13 FAB PBI Polar What is API? Atmospheric Pressure Ionization Source Types: 1. Electrospray Ionisation (ES) – Solution phase process (for the most part). 2. APCI (Atmospheric Pressure Chemical Ionization) - Gasphase process. Source Purpose: 14 1. Ionize the analyte (APCI) or transport ion in solution to the gas phase. 2. Desolvate sample flow for introduction into mass spectrometer. 3. Baffle the first vacuum region of the MS from atmospheric pressure in the source. 4. Pump away neutrals and opposite charged ions which would otherwise interfere with the analysis of the desired polarity. LCQ Classic/Duo/Deca API Probes Electrospray Ionization (ESI) Peek insulator Atmospheric Pressure Chemical Interface (APCI) 15 LCQ Advantage/XP API Probes Electrospray Ionization (ESI) Atmospheric Pressure Chemical Interface (APCI) Orthogonal ESI & APCI probes 16 Chemistry Considerations ESI: Ions formed by solution chemistry Good for Thermally labile analytes Good for Polar analytes Good for Large Molecules (Proteins / Peptides) APCI: Ions formed by gas phase chemistry Good for Volatile / Thermally Stable Good for Non-polar analytes Good for Small Molecules (Steroids) 17 ESI Versatility Advantage/XP (Right: Picture represents a low flow position, <50uL/min) A-F Positions for increased ruggedness 1-5 Positions (Left: Picture represents a high flow position) 18 Electrospray—Basic Layout Heated Capillary ESI Needle Solvent evaporation and ion release +/- 5 kV + + + + + + + + + + + ++ + + ++ + + ++ ++ ++ + + ++ ++ ++ + + + + + + ++ ++ ++ ++ + + Taylor Cone 19 + ++ + + + + + + + ++ + ++ ++ ++ + ++ + + + + + + ++ + + + + + +++ + + + + + + + + + + + + + APCI Position on Advantage/XP In/Out Movable Corona Discharge Pin 20 Total control for any flow rate (200ul/min 2000ul/min) LC Flow Rates ESI: 3 µL/min - 1mL/minute Optimal Flow Rate: 200 µL/min Generally, higher flow rates require higher heated capillary temperatures and higher gas flow rates. APCI: 200 µL/min - 2mL/minute. Optimal Flow Rate: 500 µL/min Generally, higher flow rates require more sheath and auxiliary gas, but do not require higher heated capillary temperatures. 21 Theoretical Increase in Response 2 Conc max = D Column 1 Col. Diameter mm 4.6 3.0 2.0 1.0 Flow Rate ml/min 1 0.5 0.2 0.05 1 2.3 5 Theoretical Increase 22 D 21 2 Column 2 Capillary < 10 µl / min LC Additives ¾Acids Do not use inorganic acids (may cause source corrosion) Formic and acetic acid are recommended ¾Bases Do not use alkali metal bases (may cause source corrosion) Ammonium hydroxide is recommended ¾Surfactants (surface active agents) Detergents and other surface active agents may suppress ionization ¾Trifluoroacetic Acid (TFA) May enhance chromatographic resolution, but causes ion suppression in both negative and positive ion mode ¾Isopropyl Alcohol May Enhance Negative Ion Formation 23 Buffers (pH) Avoid using non-volatile HPLC additives such as: ¾ Alkali Metal Phosphates ¾ Borates ¾ Citrates Keep Buffer concentrations below 20 mM using volatile salts such as ammonium acetate. When using buffers, more frequent cleaning of the heated capillary and API stack will be necessary 24 LC/MS Additives and Buffers Summary Acetic Acid Formic Acid Ammonium Hydroxide Ammonia Solutions Trichloroacetic Acid (< 0.1% v/v) Trifluoroacetic Acid (< 0.1% v/v) Isopropyl Alcohol (10% of organic phase) Ammonium Acetate Ammonium Formate 25 Proton Donors Proton Acceptors Chromatographic Separation Negative ion formation Buffers Common LC/MS Solvents Methanol Acetonitrile Water Isopropanol Dichloromethane Chloroform Hexane 26 Effects of Solvents and Additives on ESI 50/50 ACN/H2O 0.1% NH4OH 50/50 MeOH/H2O 0.1% NH4OH 50/50 ACN/H2O 0.02% TFA 50/50 ACN/H2O 0.05% TFA 50/50 ACN/H2O 0.1% TFA 50/50 MeOH/H2O 0.02% TFA 50/50 MeOH/H2O 0.05% TFA 50/50 MeOH/H2O 0.1% TFA 50/50 MeOH/H2O 10mM NH4OAc 50/50 MeOH/H2O 5mM NH4OAc 50/50 ACN/H2O 0.1% Formic 50/50 ACN/H2O 1% Acetic 50/50 MeOH/H2O 0.1% Formic 50/50 MeOH/H2O 1% Acetic 100 ACN 100 MeOH 100 H2O 50/50 ACN/H2O 50/50 MeOH/H2O Solvent System Tyr-Gly-Gly-Phe-Leu Leucine Enkephalin 0 100000 200000 300000 400000 Counts (protonated ion species) 27 500000 API Stack LCQClassic, LCQDUO, LCQDECA 28 LCQ DECAXP, LCQAdvantage Ion Transfer Tube and Removal Tool Ion transfer tube Removal tool Heated capillary 29 Vent Prevent Mechanism Heated Tube in-situ Heated Tube removed Tungsten Vent Prevent 30 Ion Optics Intermultipole Lens First multipole Lens IONS IN IONS OUT Octapole Mount Vacuum Baffle 31 Second multipole Lens Analyzer Mount Multipole Potential Wells Mass Range Trans Efficiency Octapole Square Quadrupole Round Quadrupole 32 Mass Analyzer (Ion Trap) 33 Vacuum System Every mass analyzer must operate under vacuum in order to minimize both ion/molecule and molecule/ molecule collisions. At atmospheric pressure, the mean free path of a typical ion is only ca. 52 nm and at 1 mTorr, it is 40 m. Without vacuum, the ions produced in the source won’t make it to the detector. The LCQ vacuum is maintained by a both rotary and turbomolecular pumps 34 Ion Optics (Operating Pressures) 760 torr 1.3 torr 1.7x10-3 torr 2.0 x10-5 torr (1.0x10-5 torr He) 3.5x10-3 torr He 60 m3/hr 100 L/sec 35 220 L/sec Steps to Ion Trap Scan Functions Trapping- all scans Isolation- SIM and MSn Excitation- MSn Ejection- all scans 36 Helium as a Damping Gas Without Helium + With Helium + + He collision + 37 + + He He He He + + Discovery of the Effects of Helium 38 Helium as a Damping Gas 39 Ion Trap Resolution–Effect of Damping Gas Traps injected ions by removing kinetic energy Damps ion motion to center of trap – Result... Increase in resolution and sensitivity 40 Helium Effect Helium flowing into trap S#:1 RT:0.00 AV:1 SM:7G NL:2.50E7 T:+ p Full ms S#:23-32 RT:0.71-1.00AV:10 SM:7G NL:5.61E7 T:+ p Full ms 524.3 1522.04 100 Relative Abundance Relative Abundance 100 80 60 60 40 524.26 40 525.3 20 1721.89 1222.14 1821.95 1122.21 20 195.15 0 1621.97 1322.06 80 1921.88 1022.09 0 514 516 518 520 522 m/z 524 526 528 500 1000 m/z 1500 2000 Helium shut off and not flowing into trap Relative Abundance 100 S#:23-32 RT:0.39-0.54AV:10 SM:7G NL:2.80E7 T:+ p Full ms 522.6 523.0 100 Relative Abundance S#:1 RT:0.02 AV:1 SM:7G NL:9.70E6 T:+ p Full ms 80 80 521.8 60 521.2 40 41 514 516 518 520 522 m/z 524 1220.75 40 20 526 528 0 1720.44 1320.95 60 523.9 520.7 20 0 1620.79 1520.26 523.01 1919.96 1120.90 192.17 500 1000 m/z 1500 2000 Ion Trap Stability Diagram The region shaded blue indicates a (DC) and q (RF) values which provide stable trajectories in the r-direction The region shaded yellow indicates the z-stable a and q combinations The green area where the rand z-stable regions overlap indicates the a and q combinations under which ions will be stable in the trap 42 Stability Diagram for Commercial Traps V qz = k ( m / e) 43 LCQ Scan-Out (Ejection) Rates Normal Scan (5500 amu/sec) Common full, SIM, or MSn (SRM and CRM) scanning Resolution (FWHM) = 0.50, Mass Accuracy = ± 0.05 Zoom Scan (280 amu/sec) Increases resolution and mass accuracy across a narrow range (allows charge state determination) Resolution (FWHM) = 0.15, Mass Accuracy = ± 0.02 Turbo Scan (55,000 amu/sec) Decreases total scan time of a full scan, thus increasing number of scans across a chromatographic peak Resolution (FWHM) = 3.0, Mass Accuracy = ± 0.5 Used for better quantitation due to an increase of scans across a chromatographic peak 44 What is AGC and Why Is it Important? Camera AE • Too much light degrades the image stored on film, causing a loss of color and image resolution. • Too little light results in dark picture with no fine details visible. • Cameras with high quality light meters and AE controls produce high quality pictures over a wide dynamic range of lighting conditions. 45 LCQ Series AGC • Controls amount of ions (light) entering the ion trap (film) • Too many ions degrade the spectral quality in the trap, causing loss in mass resolution and mass assignment. Too few ions result in poor sensitivity to low level or minor components. • AGC ensures excellent quality MS, SIM and MS/MS spectra, as well as excellent sensitivity over a wide dynamic range. Automatic Gain Control (AGC) Prescan before the analytical scan - Measures the # of ions in the trap for a pre-defined time (10 ms) -Allows software to determine optimum ion injection time 46 No AGC Spectrum of Ultramark 1621, Caffeine, MRFA Calibration Mixture – space charging 47 Spectrum of Ultramark 1621, Caffeine, MRFA Calibration Mixture with AGC 48 AGC (Ion Population Control) 100 80 60 40 525.3 20 524.4 0 522 60 40 525.4 20 526.3 527.5 0 m/z 530 522 Good Resolution 49 100 80 526.3 m/z ~ 3000 Ions 530 524.5 100 80 60 40 525.5 20 0 522 526.5 527.5 m/z ~ 6000 Ions 530 Relative Abundance 524.3 Relative Abundance Relative Abundance 100 ~ 1500 Ions Relative Abundance ~ 300 Ions 524.8 80 60 525.7 40 526.7 20 0 522 m/z 530 Poor Resolution Calculation of Ion Time Constant During Prescan AGC Prescan Signal = Number of Ions x Multiplier Gain x Prescan Time (3 x 105 counts) (10 ms) Calculated Ion Time = (how long the gate lens is “open”) 50 Target Value AGC Prescan Signal Unscaled TIC (counts) Triplicate Injection of 5 nmol of MRFA (AGC ON) 2.0E+08 1.8E+08 1.6E+08 1.4E+08 1.2E+08 1.0E+08 8.0E+07 6.0E+07 4.0E+07 2.0E+07 0.0E+00 Unscaled TIC 0 100 200 300 400 500 600 400 500 600 400 500 600 Injection Time (ms) Scan Number 60 Injection Time 50 40 30 20 10 0 Scaled TIC (counts) 0 8.0E+08 7.0E+08 6.0E+08 5.0E+08 4.0E+08 3.0E+08 2.0E+08 1.0E+08 0.0E+00 200 300 Scan Number Scaled TIC 0 51 100 100 200 300 Scan Number Isolation of Ions Ion we wish to isolate qz 0.0 0.908 • Ions at different qz values oscillate at different frequencies (ωo) 7 52 ωo ≈ q zΩ 2 2 Isolation Waveforms ~ m/z 200 q axis .908 500 Hz 16 msec q axis 53 .908 Why MS/MS or MSn ? Intensity Signal to Noise Improvement Signal Noise S/N 0 1 2 3 4 Stages of Analysis 54 5 6 MS/MS Parameters • Precursor ion m/z • Excitation qz Intensity • Excitation voltage Precursor ion Σ(Product ions) 0.0 1.0 2.0 3.0 Excitation Voltage (V) 55 4.0 77001-1285 970219 Resonant Excitation qz Value Fragment ions not trapped qz 0.0 0.908 0.225 Product Ion m/z Range 1/4 qz 1/3 0.0 0.30 0.908 qz 1/2 0.0 56 0.45 0.908 Fragmentation Energy Ion Trap Scan Functions 57 1. Collect 3. Fragment 2. Isolate 4. Eject Summary (Ion Trap Functions) 1) Collection 2) Isolation 3) Excitation 4) Ejection 58 For Scans: All By: Ring Electrode Method: Alternating RF frequency (760 kHz) at a set amplitude along with He dampening gas traps and cools the ions to the center of the trap. Summary (Ion Trap Functions) 1) Trapping 2) Isolation 3) Excitation 4) Ejection 59 For Scans: SIM, MSn By: Endcap Electrodes Method: a) Tailored waveform applied to all ions in the trap except ion of interest b) Thus, only ions of interest remain in the trap. Summary (Ion Trap Functions) 1) Trapping 2) Isolation 3) Excitation 4) Ejection 60 For Scans: MSn By: Endcap Electrodes Method: a) Cool ion of interest back to set q value (default = 0.25). b) Apply custom RF waveform in resonance with the set q value, activation time (default = 30 msec), and optimized activation amplitude. Summary (Ion Trap Functions) For Scans: All 1) Trapping 2) Isolation 3) Excitation 4) Ejection 61 By: Ring Electrode Method: Ramp ring RF power to increase the q values of all ions in desired scan range, low mass to high mass. (i.e. Mass Selective Scanning) Also, ramp the RF amplitude on the endcap electrodes to consolidate the ions to a group (Resonance Ejection) Tune Page convection gauge pressure < 1.5 torr? Ion gauge Pressure <2.0x10-5 torr? 62 ESI Calibration Solution Caffeine stock: 1 mg/ml in methanol MFRA stock: Dissolve 3.0 mg MRFA in 1 ml 50:50 methanol:water Ultramark stock: Measure 10 υl of Ultramark 1621, and dissolve it in 10 ml acetonitrile ESI calibration solution Into a clean vial pipette 100 µl of caffeine stock, 5 µl of MRFA stock and 2.5 ml of Ultramark stock. Add 50 µl of glacial acetic acid and 2.34 ml 50:50 methanol:water 63 Ion Trap Animation 64 The Eighth Generation Triple Quad TSQ 15, TSQ 45, TSQ 46, TSQ 70, TSQ 700, TSQ 7000, TSQ, TSQ Quantum 65 Smaller because… 8degrees 25 cm quad 25 cm quad TSQ 7000 90 Degree Square quad collision cell 25 cm quad Quantum 25 cm quad 66 90 degrees HyperQuadsTM Hyperbolic Quadrupoles TSQ 7000 1993 to 2000 r0 = 4 mm L = 250 mm • Forms Pure Quadrupolar Fields • Reduces Fringing Effects • Significantly Improves Resolution • Improves Transmission • Improves Peak Shapes 67 TSQ Quantum 2001 to … r0 = 6 mm L = 250 mm Quadrupole Mass Analyzer + + + + The ion is transmitted along the quadrupole in a stable trajectory Rf field. The ion does not have a stable trajectory and is ejected from the quadrupole. 68 How does the Quadrupole work ? The quadrupole consists of four parallel rods. The opposing rods have the same polarity while adjacent rods have opposite polarity. Each rod is applied with a DC and an RF voltage. Ions are scanned by varying the DC/Rf quadrupole voltages. Only ions with the selected mass to charge ratio will have the correct oscillatory pathway in the Rf field. -ve +ve 69 Effect of Peak Width On Transmission HYPERQUAD T r a n s m % i s s i o n ROUND RODS 100 80 60 40 20 0 2 1.5 0.7 0.5 Peak Width FWHM 70 0.2 0.1 Effect of Peak Width on Resolution 12000 Quantum operating at 0.1 FWHM at m/z 1000 R = 10,000 Quad 0.7 FWHM Quad 0.1 FWHM 10000 Sector Resolution 8000 6000 4000 Quantum, API 4000, Ultima at 0.7 FWHM at m/z 1000 R = 1428 R is relatively flat across m/z 2000 0 0 200 400 600 m/z 71 800 1000 1200 The Power of Resolution z Separation of ions with same nominal m/z value z Unequivocal determination of charge state (ESI) z High resolution precursor ion selection for MS/MS z High resolution product ion for charge state determination 72 Effect of changing resolution on peak shape 1.0 FWHM 0.3 FWHM 73 0.7 FWHM 0.2 FWHM 0.5 FWHM 0.1 FWHM Resolution vs. Intensity 1.0 FWHM 2.5e6 74 0.7 FWHM 2.1e6 0.5 FWHM 1.9e6 Resolution vs. Intensity 0.3 FWHM 1.5e6 75 0.2 FWHM 0.1 FWHM 1.4e6 0.8 e6 Quadrupole Mass Analyzer If one MS scan between m/z 100 and 500 is completed in one second, then each m/z will be allowed to pass for 2.5 ms. 1000 ms 400 amu = 2.5 ms/amu + + + + + + 76 To detector Quadrupole Mass Analyzer And, for the same peak, for example, the quadrupole performs 5 complete scans from 100 – 500 Da each taking 1 sec. 100 – 500 Da 100 – 500 Da 100 – 500 Da 100 – 500 Da 100 – 500 Da 77 5 sec Ion Trap Pre-Scan The length of time that the trap stays open to collect ions is determined by a pre-scan which measures total ion current (prevents space charging, so no ghost peaks…) Pre-Scan 78 5 sec Ion Trap Pre-Scan Cont’d Across a peak 5 sec. wide the trap might fill and empty 5 times. So, a group of ions are collected ca. every 1 sec – each group is then ejected to the detector, smaller ions first. 1 sec 1 sec 1 sec 1 sec 1 sec 79 5 sec Comparison of Quads and Traps For the quadrupole, each m/z is scanned (sequentially) to the detector for 2.5 ms. + + + + 80 But for a trap, each m/z (and all m/z at the same time) is/are collected for ca. 750 ms (taking pre-scan and interscan times into account) and then scanned to the detector. Comparison of Quads and Traps A trap will fill similar to filling a glass with water. All ions entering the trap will be collected until the trap fills with a pre-defined amount of ions (AGC target value). At any one particular instant, a quad will only scan/look for only one m/z. All other m/z will be ignored. In this example, each m/z is scanned for 2.5 ms. + + + + 81 Comparison of Quads and Traps So, for full scan MS, a trap will give better sensitivity – because there are more ions representing each m/z arriving at the detector for each scan. + + + + 82 How Much More??? Accounting for pre-scan/interscan timing, the trap produces ca. 300 times (750 ms/2.5 ms) for the “collection” of each m/z compared to a quadrupole (i.e 2 orders of magnitude). + + + + 83 But… What if I wanted to pass (filter) only one m/z ion to the detector (i.e. SIM or SRM) – then I could spend more time on that ion… + + + + 84 Comparison of Quads and Traps Yes, on a quadrupole, interscan times are relatively short and so the quadrupole remains fixed on that one ion – a duty cycle of close to 100% + + + 85 But a trap will still only collect ions in “batches”and prescan/interscan times afford a duty cycle of about 75% + What is a Duty Cycle? Definition:Time taken to acquire 1 scan and be ready to acquire the next one Duty Cycle Scan 1 ISD Scan 2 Interscan delay (ISD) is the time taken to return all system voltages to the start values and reach a stable state This is ‘dead’ time and should be minimized 86 In Addition… While the beam instrument is continuously detecting one particular m/z – a trap builds a curve from an “average” over each collection time – and the points are least frequent at the most important region for quantitation (the “take off”). 87 For Example… SRM of 5 pg Alprazolam with LCQ Deca gives a %RSD of 6.11 while… SRM of 750 fg Alprazolam with TSQ 7000 gives a %RSD of 1.87. Signal to noise ratio (S/N) are similar (20:1 and 28:1 respectively). 88 RT: 4.26 SN: 20 100 90 Relative Abundance 80 70 SRM 5pg Alprazolam with LCQ Deca %RSD 6.11 60 50 40 30 20 10 0 SRM 750fg Alprazolam with TSQ RT: 4.41 SN: 28 100 90 80 70 60 %RSD 1.87 50 40 30 20 10 0 0.0 89 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 Time (min) 4.5 5.0 5.5 6.0 6.5 7.0 7.5 The Effect of Ion Trap Scan Speeds on Quantitative Performance • In general, faster scanning produces more points across a chromatographic peak, hence better precision and lower LOQ’s. • In fact, for an ion-trap, scan speed refers only to the time taken to scan ions from the trap during mass analysis. • Scan speed does not refer to the total analytical cycle. In an ion trap device, an MSn analytical scan comprises at least four events: 90 The Effect of Ion Trap Scan Speeds on Quantitative Performance 1- AGC Pre-scan 2- Ion Injection (usually the rate-determining step) 3- Isolation and activation of the parent ion within the trap 4- Scanning the ions out of the trap (mass analysis) 91 MS Scan Function Mass Analysis Ion isolation Ion Injection AGC Prescan 92 Analytical Scan Ion Activation Scan Terminology Prescan Mass Analysis 1st Microscan Prescan Mass Analysis 2nd Microscan Complete Scan Cycle 93 Prescan Mass Analysis 3rd Microscan Save Data The Effect of Ion Trap Scan Speeds on Quantitative Performance • The injection time constitutes the majority of the total scan time. • For good quantitative reproducibility, it is necessary to take enough points to precisely determine the chromatographic peak particularly at the take-off point. • Increasing the scan speed in the trap does not significantly increase the number of data points taken across the peak. 94 Scan speeds in ion traps Pre-scan Pre-scan Quantitation m/z 500 Isol scanning 135-510 /Activ Isol /Activ//Download Download Pre-scan Pre-scan 60 375 amu/sec 60msec msec 375amu amu@ @13,000 13,000 amu/sec Isol Injection Isol/Activ /Activ//Download Download 80 Injectiontime time80 Pre-scan 60 msec Pre-scan 60 msec 375 amu @ 13,000 amu/sec 30 Total scan 375 amu @ 13,000 amu/sec 30 Total scantime time Isol/Activ/ 80 Isol/Activ/Download Download Injection time wide Injection time80Scans 500 Scans//10 10sec sec500 widepeak peak 375 amu/sec 70 375amu amu@ @5500 5500 amu/sec 70 Total scan time 670 Total scan time 670 Injection 500 Injectiontime time Scans / 10 sec wide 15 Scans / 10 sec500 widepeak peak 15 Total Totalscan scantime time 710 710 Scans Scans//10 10sec secwide widepeak peak 14 14 95 00msec msec 80 80 30 30 500 500 610 610 16 16 The Effect of Ion Trap Scan Speeds on Quantitative Performance • The only significant way to increase the sampling rate across the peak is to reduce the injection time which can be achieved in two ways: 1- Set a lower max injection time which reduces the number of ions in the trap, hence sensitivity 2- Increase the efficiency of the source and lenses to improve the transmission of ions; I.e filling the trap to the same level in a shorter period of time. 96 Data Dependant Acquisition of MSn Spectra Data Dependant Acquisition: “Intelligent” decision-making software that selects precursor ion for MSn experiments based on user-define criteria. Critical for metabolite screening experiments. Scan event 1 Full-Scan MS Scan event 2 Full-Scan MS2 Software selects most intense ion from scan event 1 as precursor ion for ms2 experiment in scan event 2, provide that its intensity is above a user selected threshold m/z 97 m/z “Dynamic Exclusion”-- MS and MS/MS of Co-Eluters MS MS/MS MS/MS MS MS/MS 377 407 MS MS MS MS MS 452 Threshold 365 231 Time MS 407 m/z 206 255 452 m/z 98 m/z MS/MS 377 m/z Comparison of Quads and Traps Major Strengths of Triple Quads • SRM Sensitivity • Neutral Loss Scan Mode • Parent (Precursor) Scan Mode Major Strengths of the LCQ Deca XP Plus • MSn Scan Mode • Full Scan MS/MS Sensitivity • Consecutive Reaction Monitoring (CRM) 99 What are neutral loss scans ? • Both Q1 and Q3 are scanned together • Q3 is offset by the neutral loss under investigation • The precursor ions collide with Argon gas in Q2 to create fragment ions • Only those compounds which give a fragment having that specific loss are detected • Since both Q1 and Q3 are scanning, neutral loss scan mode is slower than any other mode 100 Neutral loss scans Neutral loss scans are used for screening experiments where a group of compounds all give the same loss H2 N - m/z 84 H2 N N N H2 N H2 N N NH2 - m/z 84 HO N 101 N N N N NH2 H2 N - m/z 84 H2 N N N - m/z 84 HO N N What are precursor ion scans ? • Precursor ion scans also known as parent ion scans • Q1 is scanned • Q3 is set to allow only a fragment ion of one m/z to pass; (Q3 fixed) • ions collide with Argon gas in Q2 to create fragment or product ions • Only those compounds which give that specific fragment ion are detected 102 Precursor ion scans Precursor ion scans are used for screening experiments where a group of compounds all give the same fragment ion m/z 192 NH2 m/z 162 N N H 2N N HO N N N N m/z 192 N m/z 84 H 2N N H 2N N m/z 238 103 N m/z 268 Comparison of MSn in a Triple Quad verses an Ion Trap Instrument. Triple Quad(nonresonant excitation): Acceleration voltage applied equally to all masses. Get a mix of ms2, ms3…msn products. Ion Trap(resonant excitation): Excitation energy is in resonance with only one mass at a time. Fragments, once formed, can not be further excited unless they are purposely selected for next stage of MS. Allows one to take apart a molecule in a controlled, step-wise fashion. 104 Comparison of Quads and Traps Ion Traps Quadrupoles Mass Separation in time Mass Separation in space High sensitivity Full Scan Lower sensitivity Full Scan Lower sensitivity SIM and SRM High sensitivity SIM and SRM Offer multiple stages of MSn Offers Only MS or MS/MS Parent and neutral loss scans 105
© Copyright 2026 Paperzz